Omega AQM-101 Instruction Sheet

FEATURES
1) Super large LCD simultaneously display of HCHO level, Temp., Relative humidity, Calendar (Y/M/D) and Time (clock).
2) Three (3) smiley icons indicate indoor air quality levels (0/0.08/0.8ppm), easy to understand HCHO concentration.
3) Stable Electrochemical type sensor for HCHO detection.
5) Manual 0ppm calibration.
6) Audible/Visible LED warning for high HCHO concentration
1
o
ppm
AIR
TEMP
Humidity
M D
LCD DISPLAY
Symbols:
1)ppm HCHO unit
2)
icon
:
0ppm~0.08ppm
3) icon
:
0.08ppm
~
0.8ppm
4) icon
:
0.8ppm
5)Air Temp.: Ambient Temperature
6)MD : Month / Date
7)HM: Hour / Minute
8)TWA: Time Weighted Average (8 hours)
9)STEL: Short-Term Exposure Limit (15 minutes weighted average )
10)Logging: When it is i n the recording mode
11)VOC, NO2, WBGT, hpa, mbar ,EXT TEMP,lps%,cfm/p are not
available in these models
:
INTRODUCTION
Thank you for purchasing desktop HCHO meter. It is used t o measure HCHO concentration,air temperature and relative humidity with visible and audible alarms. This HCHO meter is an ideal instrument for indoor air quality (IAQ) diagnosis and HVAC system performance verification.
MATERIAL SUPPLIED
(1) Meter (2) Universal Adaptor (3) Operation manual (4) 4 pcs batteries. (5) USB cable (6) White Box
POWER SUPPLY
The meter is powered by Super heavy duty batteries AAA x 4 pcs (Recommended) or an AC adaptor(9V/0.5Aoutput)
2
KEYPAD (CONTROLS)
1) MODE : HCHO mode,MAX,TWA,STEL
2) SEL./R : Edit alarm function Logger mode: Long press to key start/off .
3) TRH.M. : Temp. Humidity mode. Setting/ Next
4) UNIT : Select Temp. unit
5) POWER: Turn meter on/off , alarm on/off.
OPERATION
Warning: Remove batteries when not in use to prevent battery leakage (Out of warranty) and power consumption.
(a) POWER ON/OFF
Put battery or plug in adaptor, meter turns o n automatically . LCD shows current HCHO, RH & Temperature, Date and Time. Three(3) smiley icons indicate the indoor air quality level and appear on the top of first layer display.
Note:
(1) When low battery, meter will restart and logger will stop if
user switch to adaptor.
(2) Please use properly power source, input voltage:
100~240 VAC,50-60Hz
Output voltage: DC 7.5~9.0V Output current: 0.5A,or the meter will be damaged.
(3) Battery life approx. 120 hours when continuous used
(b) TAKING MEASUREMENT
The meter starts the measurements when powered on and response time is less than 3 0 sec.
Do not hold the meter close to any other source o f HCHO.
Note:
(c) HCHO MODE
MODEIn the HCHO Mode, press button to switch HCHO
displa
y. The top layer of display will cycle from HCHO,
Maximum, TWA, STEL.
3
Note 1:
1)
MAXIMUM:
The unit automatically records maximum reading and minimum reading after powered on.(MINIMUM is always 0.00ppm for HCHO meter)
2) TWA : ( Time Weighted Average 8 hours) The meter keeps updating the reading every minute. If the meter has been powered on for less than 8 hours, the TWA value will be the weighted average of readings
since powered on.
3)STEL: (Short-Term Exposure Limit in 15 minutes weighted
average)The meter keeps updating readings every minute. If the meter has been powered on for less than 15 minutes, the STEL value shows the weighted average of reading taken since powered on.
4)RESET MAX:
Press button to get MAX reading andMODE switch T WAo r STEL.Under MAX mode, Long press button to clear all the values.MODE
ppm
MAX
ppm
TWA
STEL
ppm
ppm
(d) TRH MODE : Temperature / Relative Humidity mode
TRHIn the Mode, the meter displays both
related humidity and air Temperature simultaneously
.
(e) Temperature unit (°C or °F) selection
Press button to toggle theUNIT
temperature unit (°C or°F).
o
AIR
TEMP
HUMIDITY
+ TEMPERATURE
HUMIDITY
o
AIR
TEMP
HUMIDITY
o
AIR
TEMP
HUMIDITY
4
MAINTENANCE
Cleaning and storage:
1). The meter should be cleaned with a damp cloth and mild detergent when necessary.
2). Store the meter in an area with moderate temperature and humidity.
MANUAL CALIBRATION TO ZERO
Before start HCHO manual calibration make sure the area is ventilation and no HCHO environment. Ensure the calibration completed, to use adaptor during calibration is recommended.
HCHO meter is completed calibration in factory before
packed. Manual calibration is only for meter inaccuracy.
When power on, long press button moreUNIT than 6 seconds, the value will turn into “00000" and flash seconds.60
After calibration completed, display will show full LCD icon for a second. Meter will restart automatically to return normal measurement.
BATTERY INDICATION
When battery is low (HCHO consumes power ),the reading of HCHO will be terribly high, could b e couple thousand in ppm. “LO BAT” i s flashing at the bottom of the screen, near calendar and time. Replace new battery when LO BAT shows.
SETTING DATE & TIME
a) Real time and date:
The meter shows Mo./Date and Hr/Min on the 3rd layer of display, and each cycle is 16 seconds.
b) Month/Date and Hour/Mi
n
Pres
s button to enter the real date andMODE+POWER
time setting. The time default is 24-hours format.
M D
5
HCHO Hi alarm setting.
Enter HCHO alarm setting.+SEL/R POWER- ->
- ->
- ->
- ->
Change digit.UNIT
Number goes up.SEL/R
Number goes down.TRH/M
After set up + to save and exit setting.SEL/R POWER
ALARM SETTING
Set up alarm function:
In normal measurement, press buttons ,SEL/R+POWER bell will appear alarm function. Once set up, the red LED will flash if the data exceeding the setting data.
SETTING DATE & TIME
The meter interchangely shows current date and time on the 3rd layer of display, and each cycle i s 6 seconds. The time default is 24-hours format.
+ -->Enter date/time modeMODE POWER
--> Number goes up.SEL/R
--> Number goes down.TRH/M
-->Edit Month/date or Hour/Minute.MODE
-->Select to edit Month or date, Hour o r Minute.UNIT
After set up, press + to save and exit.MODE POWER
HCHO alarm on
HCHO alarm off
o
ppm
AIR TEMP
Humidity
M D
o
ppm
AIR TEMP
Humidity
M D
HCHO alarm ON/OFF: In normal function, short press bell icon appearPOWER alarm on, short press again to turn off .POWER
o
AIR
TEMP
Humidity
H M
TEMP. HUMIDITY OFFSET
Make sure the meter is of f, press buttonSEL/R+TRH/M then plug in the adaptor at the same time t o enter offsetting mode.
WARNING: DO NOT USE THIS FUNCTION WITHOUT STANDARD INSTRUMENT.
Example: Standard instrument is 30.2°C, meter shows 30.0°C.Enter temp. offset mode, press
till 0.2 appear.TRH/M
Offset range of Temp.and RH a s below: Temp.:-9.9°C to +9.9°C / -9.9°F to +9.9°F RH%:-9.9% to +9.9%
a) Humidity offset:
Increase value Decrease value
b) Air temp. offset
Increase value
Decrease value
After setting up short press to save setting and, POWER “%” icon will flash a second. Then replug adaptor to return normal
measurement.
Logger setup:
When using datalogger function, please use batteries and adaptor at the same time to avoid power outages and data loss.
A. Install logger driver for Windows
The PL2303_Prolific_DriverInstaller_v110.exedriver is for
Windows operation system, follow the installation steps
hereunder:
1.Put the CD into CD driver, click PL2303_Prolific_DriverInstaller_v110.exe for installation the driver.
2.The set up status.
3.Click “NEXT” to continue.
4.After complete the installation, press “Finished”.
6
7
B. Install the logger software for Windows X P
(win7/win8/VISTA)
Please follow the following steps to install the software
1. Put CD into C D driver, click “Setup.exe” to install
2.Select the destination folder you are go to save b y click Browse button
3.Once you finish selecting ,press Next ”.
4.Select “I accept the License Agreemen:(s)” and press NEXT
button.
5.Click NEXT button when you see the following.
6.Press FINISH button to complete the installation.
7.Restart the Windows XP system
C Run the software from Windows XP.
1. Plug the USB cable into USB port, since you have already
Installed the USB driver, computer will automatically detect logger ,
2. Find “Datalogger” in S TART PROGRAMS
3.Select logger software folder.
Reminder:
Software is installed completed,”Get Datalog’s Identifier”shows next page(pic.1), then press SETTING button to enter Log setting
(Pic.1)
Note: If logger is not plugged i n USB port, the warning message “Reminder: Data Logger is not plugged in USB port!” shows next page(pic.2) Make sure logger is plugged in and choose the correct COM port.
MADE IN TAIWAN
AQM-101
IAQ Formaldehyde (HCHO) Logger
M5443/1114
INSTRUCTION
SHEET
Shop online at: www.omega.com e-mail: info@omega.com For latest product manuals: www.omegamanual.info
(Pic.2)
Logger Setting:
(Pic.5)
(Pic.3)
(Pic.4)
8
NOTE:
If software message box show as Pic2.Follow the next steps.
1.Click My computer Content in hardware content, click ”” Device manager to find the listed (COMx) listed under Ports
(COM & LPT as below with red line. If you may Prolific USB-to-Serial Comm. Port , it means the logger is connected
to computer properly.
2. Remember the Comm port number and go back t o software.
3 Click “Setting” ”Comm port“select the port number a s you.
see from step 1, then press OK button.
9
BASIC:(Pic.3)
1:ID setting
Maximum 20 characters , name your logger press OK button to confirm
Please type in capital. Space is unacceptable.
2:Current Date/Time
Clock setting, the system automatically shows current date and
time of your pc. Press OK t o confirm.
(Note: Schedule date & time refer to current pc date and time)
Please click OK to sync the meter ’s date and time
every time.
ALARM&SENSOR:(Pic.4)
1.Appear the parameter.
HCHO model default with Temp. and RH sensor.
SETTING:(Pic.5)
1. Set sample point (K=1,000 multiple basis within total memory points).
2.Set sample rate, start mode to record:
Select “hour(1~6), “minute(0~59)”,“second(0~59)” time interval. Select 5 start modes from: Immediate (Logger start recording immediately) Real-time(PC is always
connected with the logger)
Key
start/off (Start/stop recording by long pressing
SEL/R button)
Roll-over (when memory are full, logger covers the earliest memory
automatically)
Schedule (Select the date and time meter will start recording.)
3.Select Unit :
Select Temperature unit °C/°F for recording, press OK to confirm.
Long press
10
Schedule start mode:
Select the date and time meter will start recording. The recording start time as software’s current time not meter’s . Please set up the current time correctly.
Real-time start mode:
1.After set start mode at real-time,datalogger starts login records. When logger records, the green fan at left corner shows it i s under recording.(Meter should be connected with pc all the time, under this mode.
)
2.In real-time mode, it shows “STOP” button instead o f “DOWNLOAD” at the right corner. Press ST OP to stop real-time recording.
Before exit software data has to be downloaded.
Key-start/off mode:
Select the mode to record once long press
button for 3 seconds. Long press R/LESR/LES
button again to stop recording. Data has to b e downloaded before next key-start. LCD display appear “Logging” at left corner.
After key off the recording, the data has t o downloadNOTE:
before restart the Key-Start function.
Roll-over mode:
After Roll-over mode is selected. Press OK, the meter will start to record. When the data is full, it will cover the earliest memory. After download the data , long press button to recordSEL/R without reset up each parameter an
d selection.
Logger Status:
Before download data, press to show ID,LOGGER status Records, Sample rate, Unit.
Once button is pressed, recording isNOTE: Download
inactivated.
(pic.6)
11
Logger function:
Download data:
Press DOWNLOAD button, data download in few seconds. Graph shows automatically
Please do not press DOWNLOAD button if you won’tNOTE: stop record. You can press LOGGER STATUS button to see record points.
Retrieve file:
Press RETRIEVE FILE button to select file and show the graph.
Save data:
Press SAVE button t o save txt.or csv.file. After save as txt. may print as pdf. file if needed.
View Data Table:
Press View button to get data table with details.(pic.6)
Print graph:
Press PRINT button and print graph. Follow the print setting below to print full screen graph. Press OK t o print pdf file.(pic.7)
Print Table Data:
Press table data, the screen shows the record range, enter the start number to the end number (For example: 1~500 or 200~1000...or leave it blank for printing all data ).(pic.8) Save each pdf. files with different names.
√ √
(pic.7)
(pic.8)
=
Screen graph zoom in
=
Ctrl
+
Screen graph zoom out
Mouse
Hold
Click
Shift
Click
Mouse
=
Move screen graph
Press
or
= Read each measurement .
Ctrl
+
Hold
Click
Ctrl
+
Hold
Mouse
12
Clear Data:
The data is stored in the memory till next setting start, meter
will clean previous data automatically. So if you start next data logging, there is no way you can find the data you haven't stored.
Line color selection:
1.Click the color beside the reading(Temp. RH. HCHO),
color box appear.
2.Move mouse to review and select color.
3.Click to save, the color appear on line and bar.
Zoom in/out&Move
NOTE: May follow the line color selecting, the graph size will
go back to default.
Meas. Range
Audible alarm
Interfacce
LED(Green.Red) Temp. range
RH%range
Resolution Accuracy
Accuracy(HCHO)
Response time Memory Sensor life time
YES (Adjus table)
0.00-5.00 ppm (m g/m
3
)
10 ~ 90%RH
YES
AQM-101
USB
T90: Less than 30 se conds
Expected3years
0~50°C (32~122°F )
0.01ppm(HCHO) ,1°C/°F , (1%RH)
±1°C /2°F (Temperature), ±5% (10~90%RH)
±5% of reading + 0.03ppm (mg /m3 )
( Under vaccum ,ventilated standard gas cham ber
). MeetNOISH acceptance criterion within +/-25%
of the true value atth e95% confidence level.
Example: OSHAForm aldehyde regulation
0.10ppm ( EPA 0.08ppm), accuracy within
0.025ppm ( 0.075~0.125ppm)
32,000
points
13
SPECIFICATIONS
TROUBLE SHOOTING
(1) When meter appears break character, please find
out if meter ever dropped to the floor. I f “yes”, please contact with local distributor for technical service.
(2) Error codes:
Er1 : HCHO/Temp./RH sensor is failed Er2 : HCHO/Temp./RH Over range
14
WARRANTY
This instrument is warranted for two years from the date of
purchase(One year limited warranty applies to cables).AReturn Authorization letter must be issued before returning for any reason. This warranty does not apply to defects resulting from action of the user such as misuse, abuse, alteration neglect, improper wiring, improper maintenance or repair, or unauthorized modification, damage resulting from leaking batteries, operation outside of specification.
During the warranty period the manufacturer reserved the right to decide either to repair or replace the product. The meters are t o be returned along with good packing to prevent any damage in shipment and insured against possible damage or loss. The two years’ warranty doesn’t apply to:
Accessories and batteries (not covered by warranty). Claims is not acceptable for improper use (including
adaptation to particular applications not foreseen in the instructions manual) or improper combination with incompatible accessories or equipment, or b y previous attempts for repair carried out by none skilled or unauthorized personnel. Sensor damaged by user improper use
(Watering,Burning,Condensation....)
HCHO LEVELS AND GUIDELINES
Non-enforced reference levels: a) 0.03 ppm-Average outdoor level. b) 0.10 ppm-Recommend upper limit for residences by
ASHRAE , ANSI,EPA,NIOSH for STEL. c) 0.40 ppm-Recommended upper limit for manufactured homes. d) 0.50 ppm-OSHA workplace limit. e) 0.75 ppm-OSHA TWA limit. f ) 0.80 ppm-Level at which most people first detect odor. g) 2.00 ppm-OSHA STEL limit.
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in
materials
and workmanship for a period of 13 months from date of purchase. OMEGA’
s
WARRANTY adds an additional one (1) month grace period to the normal one
(1)
year product warranty to cover handling and shipping time. This ensures
that
OMEGA’s customers receive maximum coverage on each product. If the unit malfunctions, it must be returned to the factory f or evaluation
.
OMEGA’s C ustomer Service Departme nt will issue an Authorized Retur
n
(AR) n umber immediately upon phone or written request. Upon examinatio
n
by OM EGA, if the unit is found to be defective, it wil l be repaired o r replace
d
at no charge. OMEGA’s WARRANTY does not apply to defects resulting fro
m
any ac tion of th e purchaser, including but not limited to mishandling
,
improper interfac ing, o peration outsid e of design limits, improper repair, o
r
unauthorized modification. This WARRANTY is VOID if the unit shows evidence o
f
having been tamper ed with or shows evidence of havi ng been damaged as
a
result of excessive corrosion; o r current, heat, moisture or vibratio n; imprope
r
specification; misapplication; misuse or other op erating condition s outside o
f
OMEGA’s control. Components in which wear is not warranted, include but are no
t
limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products
.
However, OMEGA neither assumes responsibility for any omissions or error
s
nor assumes liability for any damages that result from the use of its product
s
in accordance with information provided by OMEGA, either verbal or written
.
OMEGA warrants only that the parts manufactured by the company will be a
s
specified and free of defects. OMEGA MAKES NO OTHER WARRANTIES O
R
REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED O
R
IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIE
S
INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FO
R
A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION O
F
LIAB ILITY: The rem edies of purch aser s et forth herein are exclusive, a nd the total liability of OMEGA with respect to this order, whether based on c ontract, wa rranty, neglige nce, indemnification, strict liability or otherwi se, shall not exceed the purchase price of th
e
component upon which liability is based. In no event shall OMEGA be liabl
e
for consequential, incidental or special damages.
CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall i
t
be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with an
y
nuclear installation or activity; or (2) in medical applications or used on humans
.
Should any Produ ct(s) be used in or with any nucle ar inst allation or activity
,
medical application, used on humans, or misused in any way, OMEGA
assumes
no responsibility as set forth in our basic WARRANTY/DISCLAIMER language, and
,
additionally, purchaser will indemnify OMEGA and hold OMEGA harmless fro
m
any liability or damage whatsoever arising out of the use of the Product(s) in suc
h
a manner.
Loading...