Omega 400B-CV, 402B-CV User guide

Page 1
TABLE OF CONTENTS
Model: 400B-CV and 402B-CV
M3778/0602
SECTION 1
1.1 GENERAL DESCRIPTION
1.2 OPTIONS SECTION 2 SPECIFICATIONS SECTION 3 INSTALLATION
3.1 UNPACKING/ REPACKING
3.2 CONFIGURING
3.4 WIRING SECTION 4 CALIBRATION AND SCALING
Note: This manual is for the enhanced version of the Model 400 and 402 units. For models 400 and 402 before 3-2002 with serial numbers without the letter “N” please refer to manual part number M0035.
Page 2
SECTION 1 DESCRIPTION
1.1 DESCRIPTION.
This manual contains installation, operation, and calibration information for Series 400 digital process indicators. These solid-state instruments are designed for accurate, reliable, and trouble-free display of voltage or milliamps. All instruments feature a dual-slope A/D converter digital linearization, and a large 0.8” (20mm) easy-to-read 4-digit LED display. Input voltage or current can be scaled to Engineering units from –999 to +9999 counts.
The displays are in-line, uniplanar, 7-segment red LEDS, easy to read through the front Plexiglas panel. A -OL displayed on the read-out indicates a negative polarity overload. Positive readings are inferred (no sign displayed). OL indicate a positive overload condition. All units operate from 100, 120, 220, or 240VAC 50-60 Hz power source. A power connector and ground screw for shield are located at the rear of the instrument. The instrument case is made of solid die-cast aluminum, which is rugged enough to withstand the most hostile environments. It is designed to mount in a panel. The optional rack mounts can accommodate up to 3 units. Bench mount option 400-W5 provides for convenient bench top use.
1
Page 3
Multiple Input Options, Models 402 and 405
Model 402 can accept five inputs of the same input either voltage or current. Buttons numbered 1-5 are located below display. Individual input is selected by pressing appropriate button. The Model 405A provide 10-switchable inputs to a single indicator. Model 405A are housed in the same rugged case design as the basic indicator. Each may be rack-mounted beside another 405A or indicator. Pushbuttons are in two rows of five and are numbered 1-10, 11-20, etc. Each row is interlocked to prevent simultaneous input selection.
1 3 4 52
Model 402
2
Page 4
Figure 2. Input Sensor Connections 402A
+V -V
-I +I
-
+
3
+
-
VOLTAGE CURRENT
Copper Wire
-
+
SENSOR WIRES 402 5 PL
+
12345
-
+
-
+
-
Page 5
USING THE EXTERNAL SWITCHING UINTS (MODELS 405
The External Switching Units provide 10 switchable inputs to a single indicator. Inputs are selected by two rows of front panel pushbuttons labeled 1-10. (For special numbering contact factory). Pressing a particular pushbutton selects that input point for display. Each row of pushbuttons is interlocked so that only one pushbutton within that row is selected at once time.
1 3 4 52
6 8 9 107
Model 405
4
Page 6
SECTION 2 SPECIFICATIONS
Input Range:
mA: 0-20mADC
Voltage: 0-10 VDC
Sensitivity:
Maximum: 200uV/count; 0.4uA/count
Minimum: 1 count
Input Impedance:
MA: 5Ω
Voltage: > 1MΩ
A/D Read Rate:
2 per second nominal
Power:
115VAC ± 10%, 50-60 Hz
230VAC ± 10%, 50-60 Hz
30mA Max input current @ 115VAC
Operating Ranges:
Temperature: 0 to 50°C
Relative Humidity: 10-80% RH non-condensing
5
Page 7
Storage Range:
-40 to 65°C
Size:
Case: 144mm W x 72mm H x 173 mm D
(5.768” x 2.84” x 6.82”)
Maximum Weight:
1.7 kg / 2.5 lb
Panel cutout:
68mm H x 138mm W
2.68” H x 5.44” W
PERFORMANCE SPECIFICATIONS
Reference Operating Conditions (ROC):
±10% line voltage
23 ±2 °C ambient temperature
<80% RH non- condensing
Accuracy (at ROC):
0.02% RDG +/- 1 count
6
Page 8
Noise Rejection:
NMRR: ≥ 60 db @ 50/60 Hz, ± 0.1Hz
CMRR: ≥ 120 db@ 50/60 Hz, ± 0.1Hz with 250Ω unbalance.
Overload Protection:
Power Lead to ground: 1500VDC or AC RMS
Voltage: Up to 250VDC or VAC for I minute, V+ to V-
mA: Up to 150mADC mAAC for 1 minute, +I to -I
Stability With Temperature:
Zero: 1µV/°C
Span: 0.01% rdg /°C
Stability With Time:
10 counts/year max
Repeatability:
±1 count
7
Page 9
SECTION 3 INSTALLATION
3.1 UNPACKING / REPACKING
The Series 400 indicators are rugged, but they must be properly packed. Instruments are shipped in a custom­designed carton for shipping, but damage may occur. When you receive your instrument, look for evidence of transit damage. If damage is found, ask the carrier to prepare a Damage Inspection Report and notify our Instrument Repair Department immediately. If your instrument has arrived in good condition, you may perform the functional test described in 3.2.2 to verify proper operation.
REPACKING AND RETURNING THE INSTRUMENT
The original shipping container should be retained in case the instrument must be returned for repair or modification. When returning an instrument for any reason, advise us of the model number, serial number, your name, billing address, shipping address, phone number and a description of the problem. This information will enable our Instrument Repair Department to expedite the return of your instrument. Instruments being returned to the factory are required to be shipped freight prepaid.
Instruments being returned for warranty service must also refer to the original purchase date on packing lists and
8
Page 10
purchase orders. Instruments without this information will be processed as a non-warranty repair at current service rates.
If the original shipping container has been discarded, pack your instrument for shipping as follows: Select a strong cardboard box of sufficient size to allow an inch of packing material around all sides of the unit.
a. Ensure that the printed circuit boards are secured
and front and rear panels are firmly in place.
b. Wrap the instrument in plastic or strong paper.
Place it centrally in the shipping container, and pack poly foam, bubble pack, or rubberized hair around all six sides of the instrument.
c. Tape the carton flaps securely and label the
container “FRAGILE, DELICATE INSTRUMENT”. Ship the instrument, freight prepaid (do not ship by U.S mail)
9
Page 11
Input Voltage Selection (requires removal)
Input voltage 115 or 220VAC, is selectable by cutting and adding jumpers on the main board as shown on the next page. Locate JP18 on the printed circuit board. Cut and make connection as shown on table depending on input voltage.
REMOVING THE ELECTRONICS FROM THE ENCLOSURE
1- Make sure to disconnect power from the unit
2- Remove the three back cover screws and remove
the rear panel
3- Remove the screw located directly above the AC
plug
4- Remove two screws from the front panel lens and
remove lens
5- Remove two screws from the front holding the
Board assembly and pull it out of the case.
6- Reinstall the assembly when finished.
10
Page 12
Input Voltage JP18
120 VAC 220 VAC
1 2 3 4 5 6
1-2, 4-5 2-3, 5-6
JP18
11
Page 13
3.3 INSTALLATION Panel Mounting
Panel mounting hardware is supplied standard with the instrument. The instrument bezel butts against the front of the mounting panel; the mounting bracket fits over the instrument rear panel. The bracket screws force it against the rear of the mounting panel, locking the instrument in place. Panel cutout dimensions are 68mm x 138mm (2.68” x 5.44”).
BRACKET
SCREW 2 PL
REAR COVER
SCREW
REAR PANEL
PANEL
MOUNTING
BRACKET
PLASTIC PLUG
12
Page 14
3.4 WIRING THE INSTRUMENT.
All signal input connections are made to the instrument using screw terminals. To gain access to the recessed terminal strip, the rear cover panel must be removed. The four signal input terminals are located in the upper-left corner, and are labeled +I, +V, -V, and -I.
REMOVE THE POWER CORD BEFORE WIRING
Basic Single Channel Wiring
+V -V
-I +I
VOLTAGE
CURRENT
SENSOR INPUT TERMINALS
13
Page 15
Model 402 Switching Unit Wiring
+V -V
-I +I
-
+
14
+
-
VOLTAGE CURRENT
Copper Wire
-
+
SENSOR WIRES 402 5 PL
+
12345
-
+
-
+
-
Page 16
SECTION 4
CALIBRATION AND SCALING
4.1 APPLYING POWER TO THE INSTRUMENT
Power is applied to the instrument as long as the instrument is plugged into an active source. To remove power, unplug the power cord from the instrument or AC outlet.
To eliminate shock hazard or possible instrument damage, always remove the power cord either from the instrument or power source before configuring the indicator.
4.2 Selecting mA or Volts Input Mode
To select between mA or Volts input mode, remove the front panel lens to gain access to the display board. Behind the display is a DIP switch used to set the indicator’s input measuring mode. See the drawing on the following page showing DIP switch location as well as the location of the SET and ENTER scaling buttons.
15
Page 17
PROCESS/SCALING
16
SET ENTER
SCALING
BUTTONS
Page 18
PUSH DOWN HERE
FOR PROCESS
MEASURING MODE
DISPLAY
PUSH DOWN HERE TO CONFIGURE INSTRUMENT FOR mA INPUT
DIP SWITCH
PUSH DOWN HERE
FOR SCALING MODE
PUSH DOWN HERE TO CONFIGURE INSTRUMENT FOR VOLTAGE INPUT
17
Page 19
Calibrating and Scaling the Indicator
Scaling the indicator to display in your desired engineering units requires simulation of two different input VALUES (typically the lowest input and highest input) and using the two “SET” and “ENTER” buttons to scroll the digital display.
With the front lens removed and the mA/V switch
In the position compatible with your input… Step 1: Hook up a DC milliamp or Voltage source Calibrator or
actual transducer, capable of supplying a low scale and
full scale input, to the appropriate rear panel screw
terminals on the indicator. Step 2: Set the “PROCESS/SCALING” switch to the SCALING
position. The word “Lo” will appear on the display. Step 3: a) Simulate the low scale input (e.g. 4 mA)
b) Push the “ENTER” button.
c) Use the “SET” button to change the value of the
flashing digit. When the flashing digit is correct, push the “ENTER” button. The flashing digit will
18
Page 20
now move to the next right hand digit. Continue until all digits are correct with the right most digit still flashing.
d) Push both “SET” and “ENTER” buttons at the same
time to program in this scale factor. In other words, when the indicator receives a process input signal identical to the simulated (calibration) one it will display the same value shown now. (e.g. 4mA=0000)
NOTE:While the indicators calibrating itself “oo” will appear in the display. After a few seconds it will return to display “Hi” (go to step 4) or “ERR” (see error message table).
Step 4: With “Hi” displayed change the input to simulate +Full
scale (High e.g. 20mA). Step 5: Repeat steps as shown in Steps 3b, 3c, 3d changing
the digits to represent the Full Scale desired display
(e.g. 7500). When complete, the indicator will then go
to the decimal point position.
19
Page 21
Step 6: With decimal points displayed…
a) Push “SET” button until desired position is displayed
b) Push “SET” and “ENTER” buttons at the same time
to program into memory Step 7: Return the “PROCESS/SCALING” switch to the
PROCESS position. Replace the front lens. Remove
the calibrator from the input terminals. NOTES:
1) Pushing both the SET and ENTER buttons at the same
time always causes the indicator to recalibrate itself to the given input at that moment. As a protective measure, if scaling changes are made and not terminated this way no recalibration will occur, previous values will remain.
2) As a further protective measure, the “SET” and “ENTER”
buttons are not functional unless the PROCESS/SCALING switch is in the SCALING position.
20
Page 22
Button Function Legend
Use "SET" key to change value of flashing digit or decimal point
"SET"
Lo
*
Hi
* *
21
"ENTER"
* * * * * *
* *
Same time SET
& ENTER
Recalibrate to the displayed value for the existing Input Voltage or Current
*
Page 23
ERROR MESSAGE TABLE
DURING CALIBRATION/
Message Displayed
SCALING MODE
CAUSE CURE CAUSE CURE
Slope error Same values entered for both Hi and Lo
Slope error. Too many display counts for too little input
Overloaded input or display. Input exceeds specification.
Internal A/D overload
"Lo" and "Hi" values
must be different. Push "set" button
and re-enter for "Lo"
and "Hi" values
Push "SET" button. Reduce number of counts for given input.
Check input voltage or current. Must be within 0-10 VDC 0-20mA
Turn power off, wait 25 seconds, and turn power on again.
22
DURING NORMAL PROCESS
MEASURING MODE
N/A
N/A
Input exceeds maximum specifications or display is beyond
-999 or 9999
Internal A/D overload
Check input voltage or current for over range or open circuit
Turn power off, wait 25 seconds, and turn power on again.
Loading...