Omega 3200, 3400ST User Manual

FMA 3200/3200ST/3400/3400ST Series
Thermal Mass Flow Controllers
READ THIS MANUAL COMPLETELY BEFORE ATTEMPTING TO CONNECT OR OPERATE YOUR FLOW SENSOR. FAILURE TO DO SO MAY RESULT IN INJURY TO YOU OR DAMAGE TO THE FLOW CONTROLLER.
TABLE OF CONTENTS
A. Introduction .............................................................................................4
1. Unpacking...........................................................................................4
2. Product Overview And Principle Of Operation..........................................4
B. Installation...............................................................................................5
1. General Considerations.........................................................................5
2. Mounting The flow controller ...............................................................7
3. Tubing Connections..............................................................................7
5. Electrical Connections ...........................................................................8
a) Overview ....................................................................................8
b) Connecting The 6 Pin Mini Din Connector............................................9
c) Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable ........... 10
d) Connections For The 9 Pin D Sub Connector ......................................11
e) Connections For The 15 Pin D Sub Connector.....................................12
f) Using a 0-5VDC Output / Input Power Adapter Package....................... 13
C. Operation .............................................................................................14
1. Warm-Up .......................................................................................... 14
2. Verification Of Zero ............................................................................14
3. Flow Readings....................................................................................14
4. Changing The Flow Rate Set-Point (Using An External Voltage Source) ....... 15
5. Changing The Flow Rate Set-Point – FMA3400 / 3400ST Only.................16
6. Power Save Mode............................................................................... 16
7. Zero Adjustments................................................................................17
8. Recalibration...................................................................................... 17
9. Changing The Calibration Gas – FMA3400 / 3400ST Only.....................17
D. Maintenance And Product Care................................................................ 18
1. General............................................................................................. 18
2. Returning Units For Repair Or Recalibration............................................18
E. Specifications......................................................................................... 19
F. Dimensions............................................................................................20
G. Gas K Factors ........................................................................................22
H. Trouble Shooting Guide .......................................................................... 23
M-4271/0707, pg. 3 of 26
A.
Introduction
1. Unpacking
All units are suitably packaged to prevent damage during shipping. If external damage is noted upon receipt of the package, please contact
Omega Engineering immediately.
Open the package from the top, taking care not to cut too deeply into the package. Remove all the documentation and contents. Take care to remove all the items and check them against the packing slip. The products should also be checked for any concealed shipping damage. If any shortages or damage is noted, please contact Omega Engineering to resolve the problem.
Typical Contents of Box: Controller, Calibration Certificate & Manual
FMA 3200/3200ST shown; FMA 3400/3400ST have an integrated display.
Caution: Take care not to
installation section of this manual before providing power or tubing connections to the unit. Any damage caused by improper installation or careless handling will not be repaired under warranty (see limited warranty on page 25 for more details).
2. Product Overview and Principle of Operation
The FMA 3200/3400 Series Mass Flow Controllers from Omega Engineering any clean, dry gas as low as 0-20 sccm or as high as 0-10 l/min. Repeatable results are achieved using a patented thermal mass flow
are capable of measuring and controlling the flow of virtually
drop
your controller. Read the
M-4271/0707, pg. 4 of 26
sensor design. This proven design minimizes zero drift while maintaining fast response and linear outputs with virtually no maintenance.
The FMA 3200/3400 Series utilizes thermal flow sensing technology. A portion of the gas flowing through the unit is redirected into a small sensor tube. This tube has two coils on the outside. The first coil introduces a small amount of heat into the gas stream. As the gas passes through the tube heat is transferred from one coil to the other. The flow rate is proportional to the amount of heat transfer. Smart electronics analyze the amount of temperature change in the
second coil and provide a linearized analog output. A patented system insures that the zero remains stable and the sensor is extremely repeatable.
Flow in the FMA 3200/3400 Series is controlled by a proportional solenoid valve with active servo electronics. The flow measurement signal is analyzed by micro-processor controlled electronics and compared to a set­point. Adjustments are then made to the valve in order to achieve the required flow rate. The set point can be either externally input via a 0­5VDC signal or in the case of the FMA 3400/3400ST Series it can be input manually on the unit.
The output of the thermal mass flow sensor is directly related to the specific heat characteristic of the gas being measured. A sensor is calibrated for one gas but may be used with other gases by applying a correction factor to the output. The calibration gas for each specific flow controller is detailed on the product label.
B.
Installation
Caution: Do not exceed the pressure, temperature or power
operating ranges detailed in the SPECIFICATIONS section of this manual. Omega Engineering shall not be liable for any damage or injury caused by incorrect operation of their products.
1. General Considerations
It is recommended that a safety shut-off valve be installed upstream (before) of the controller.
All wetted parts should be checked for compatibility with the gas to be used. If there are any incompatibilities eg. highly corrosive gas, then the
M-4271/0707, pg. 5 of 26
unit may be damaged or fail prematurely. Such damage will not be repaired under warranty.
Units should be installed in a clean, dry environment with an ambient temperature that is as stable as possible. Avoid areas with strong magnetic fields, strong air flows or excessive vibration.
In order to operate the differential pressure across the controller should be in the range 15-45psid (1-3 bar). For optimum performance a differential pressure of 25psid is recommended.
For Example, consider the following system:
The differential pressure across the flow controller in this system would be 100 psi – 14 psi = 86 psid. Consequently the flow controller would NOT be able to control flow. For the unit to operate at optimum performance the supply pressure from the gas cylinder would need to be lowered to 39 psig to give 39 psi -14 psi = 25 psid.
M-4271/0707, pg. 6 of 26
2. Mounting the Flow Controller.
The FMA 3200/3400 Series controllers have no particular orientation or installation requirements so may be mounted in any convenient position.
It is recommended that units be fixed to a suitable substrate using the two 4-40 mounting holes provided.
Mounting View from Bottom
(mounting hardware not included with sensor)
3. Tubing Connections
All tubing must be clean, dry and purged with clean dry air before installation of the FLO-CONTROLLER
If the gas to be used may contain particles then a filter (20 microns or less) should be installed upstream of (before) the unit.
When connecting the sensor to the tubing, take care not to over-tighten the fittings or leaking may occur.
Caution: Only use the fittings factory installed on the unit. If
the fittings are removed the calibration of the unit may be effected and leaking may occur. If different fittings are required please contact the Omega Engineering Customer Service Department for assistance.
®
.
M-4271/0707, pg. 7 of 26
4. Electrical Con ne ctions
Caution: Incorrect wiring may cause severe damage to the unit.
Applying an AC voltage (115VAC or 230VAC) directly to the unit will cause damage. Read the following instructions carefully before making any connections.
a) Overview
The FMA 3200/3400 Series provides a 0-5VDC analog output proportional to the flow rate. This output may be connected to a display, data acquisition system or voltmeter with an impedance of greater than 2.5 k (kilo ohms).
The flow controller needs to be supplied with a 0-5 VDC set point signal to enable control. On the FMA 3400/3400ST Series this may be generated internally by altering the set-point potentiometer on the front panel of the unit.
A stable D.C. power supply is required to operate the unit. The voltage and current requirements depend on the configuration of the unit. Full details may be found in the Specification section of this manual.
Connecting wires should be as short as possible to avoid voltage drops. Twisted conductor cable should be used if the length of the wiring is to be longer than 2 meters.
Units are supplied with either a 6 pin mini DIN type connector (requires mating cable assembly), a 9 Pin D Sub connector or 15 Pin D Sub connector.
Caution: Cutting off the integrated connectors on the unit IS
NOT RECOMMENDED and will void the product warranty. Mating cables should be ordered along with each unit.
Electrical connections to the units are made as detailed in the following sections.
M-4271/0707, pg. 8 of 26
b) Connecting The 6 Pin Mini Din Connector
Using a suitable mating connector the pins of the integrated connector should be wired as follows:
Connecting To The Integrated 6 Pin Connector
Pin Out of Integrated
Connector
Pin 2 should be connected to the Positive of the power source.
Pin 6 should be connected to the Negative (Ground) of the power source.
Pin 3 provides the signal output and should be connected to the positive
terminal of the display, data acquisition system or voltmeter.
Pin 1 is the signal negative (ground) and should be connected to the
negative (Ground) terminal of the display, data acquisition system or voltmeter.
Pin 4 provides the input signal and should be connected to the positive
terminal of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low impedance source.
Pin 5 is the input signal negative (ground) and should be connected to
the negative (Ground) terminal of the voltage source.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the input/output signal wires or allow them to contact the power wires at any time. DAMAGE WILL
RESULT!
M-4271/0707, pg. 9 of 26
c) Connecting The 6 Pin Mini Din Connector & FMA 3000C Cable
The two mating connectors should be pushed together and the pigtail leads wired as follows:
Connecting To The Integrated 6 Pin Connector Using A FMA 3000C Cable
The RED wire should be connected to the Positive of the power source.
The BLACK wire should be connected to the Negative (Ground) of the
power source.
The ORANGE wire provides the signal output and should be connected to
the positive terminal of the display, data acquisition system or voltmeter.
The BROWN wire is the signal negative (ground) and should be
connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.
The YELLOW wire provides the input signal and should be connected to
the positive terminal of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low impedance source.
The GREEN wire is the input signal negative (ground) and should be
connected to the negative (Ground) terminal of the voltage source.
The wire colors above describe the pigtail leads of the FMA 3000C cable assembly and may not correspond with the internal wiring of your flow sensor.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the input/output signal wires or allow them to contact the power wires at any time. DAMAGE WILL RESULT!
M-4271/0707, pg. 10 of 26
d) Connections For The 9 Pin D Sub Connector
Using a suitable mating connector the pins of the integrated connector should be wired as follows:
Connecting To The Integrated 9 Pin Connector
Pin Out of Integrated
Connector
PIN 3 should be connected to the Positive of the power source.
PIN 4 should be connected to the Negative ( Ground ) of the power
source.
PIN 2 provides the signal output and should be connected to the positive
terminal of the display, data acquisition system or voltmeter.
PIN 8 is the signal negative (ground) and should be connected to the
negative (Ground) terminal of the display, data acquisition system or voltmeter.
Pin 6 is the input signal and should be connected to the positive terminal
of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low impedance source.
Pin 7 is the input signal negative (ground) and should be connected to
the negative (Ground) terminal of the voltage source.
Pins 1, 5, and 9 are not used.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the input/output signal pins or allow them to contact the power connections at any time. DAMAGE WILL RESULT!
M-4271/0707, pg. 11 of 26
e) Connections For The 15 D Sub Conne c tor
Using a suitable mating connector the pins of the integrated connector should be wired as follows:
Connecting To The Integrated 15 Pin Connector
Pin Out of Integrated
Connector
PIN 7 should be connected to the Positive of the power source.
PIN 5 should be connected to the Negative ( Ground ) of the power
source.
PIN 2 provides the signal output and should be connected to the positive
terminal of the display, data acquisition system or voltmeter.
PIN 10 is the signal negative (ground) and should be connected to the
negative (Ground) terminal of the display, data acquisition system or voltmeter.
Pin 8 is the input signal and should be connected to the positive terminal
of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low impedance source.
Pin 1 is the input signal negative (ground) and should be connected to
the negative (Ground) terminal of the voltage source.
Pins 3, 4, 6, 9, 11, 12, 13, 14 and 15 are not used.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the output signal pins or allow them to contact the power connections at any time. DAMAGE WILL RESULT!
M-4271/0707, pg. 12 of 26
f) Using a 0-5VDC Input / Output Power Adapter Package.
An optional 0-5VDC Input / Output Power Adapter Package is available for use with the FMA 3200/3400 Series. This consists of a power source (115VAC or 230VAC), a connection hub and two cable assemblies with pig­tail (soldered wire) ends. This should be assembled as shown in the following diagram.
Assembling a FMA 3215PW Power Adapter Package
(the FMA 3223PW Power Adapter Package is similar)
The RED connector should be inserted in the RED socket on the connection hub. The WHITE connector should be inserted in the WHITE
socket on the connection hub.
The cable with a RED connector provides the input signal. The RED wire
of this cable should be connected to the positive terminal of the voltage source. The (0-5VDC) voltage control signal should be supplied from a low
impedance source. The bare wire of this cable assembly is the input
signal negative (ground) and should be connected to the negative (Ground) terminal of the voltage source.
The cable with a WHITE connector provides the signal output. The WHITE wire should be connected to the positive terminal of the display,
data acquisition system or voltmeter with an impedance of greater than
2.5 k (kilo ohms). The bare wire of this cable assembly is the signal
negative (ground) and should be connected to the negative (Ground) terminal of the display, data acquisition system or voltmeter.
Caution: Avoid high voltage static discharges to the input signal
connection. Do not short the output signal wires or allow them to contact the power wires at any time. DAMAGE WILL RESULT!
M-4271/0707, pg. 13 of 26
C.
Operation
1. Warm Up
Before applying power to the unit check all tubing and electrical connections. Once correct installation is verified switch on the power. The unit should then be allowed to warm up for 5 minutes before gas pressure is applied.
2. Verificat ion o f Ze ro
Flow through the unit should be stopped by sealing or capping the inlet of the controller. It is not adequate to only stop flow by turning off the gas supply or closing a valve as there may be a leak in the system. This would give a false reading.
After 5 minutes, the zero should be stable when there is no flow through the unit. If after 10-15 minutes the output is still not zero volts (within ±0.05 volts) the unit should be adjusted as detailed in section C part 6.
It should be noted that power supply voltage variations and changes in ambient temperature can have an effect on zero readings.
3. Flow Readin gs
Each controller is factory calibrated for a specific flow range and gas (or gas mixture). The calibration gas and flow range are shown on the unit’s label and calibration certificate.
By monitoring the voltage output signal it is possible to determine the flow rate of the gas. Units are configured so that an output signal of 5.0VDC is provided when the maximum flow (i.e. Full Scale flow) is passing through the unit. The output signal is linear and scaleable enabling calculation of flow rates with in the sensor’s range. For example:
For a flow range of 0-500sccm:
At 500sccm the output signal would be 5VDC
If the output signal were 3.5VDC then the flow rate would be:
500 ÷ 5 × 3.5 = 350sccm
If the maximum flow rate is exceeded non-linear and inaccurate readings will result.
Units may be used for gases other than the calibration gas. In this case a “K Factor” would need to be applied and a corrected value calculated using the following formula:
M-4271/0707, pg. 14 of 26
Q
/ Q2 = K1 / K2
1
Q1 is the flow rate of the new gas Q
is the flow rate of the original calibration gas
2
is the K factor of the new gas
K
1
K
is the K factor of the original calibration gas
2
= (K1 / K2) Q2
Q
1
If K
is larger than K1 then linear results will only be achieved if the unit
2
does not exceed 5(K
Example 1
For a 0-200sccm unit calibrated for air the flow at 5.0VDC would be 200sccm. The K factor for air is 1. If the unit is used with Helium (K factor
1.454 relative to air) then the flow at 5VDC (i.e. the maximum flow) would be (1.454/1)200 = 290.8 sccm
Example 2
For a 0-10.0 l/min unit calibrated for Argon the flow at 5.0VDC would be
10.0l/min. The K factor for Argon is 1.45. If the unit is used with Carbon Dioxide (K factor 0.74) then the flow rate 5.0VDC would be (0.74/1.45)10.0 = 5.10l/min
The accuracy of readings using K factors is not as good as that achieved for the calibration gas. The accuracy obtained (typically ±3% for K factors similar to the calibration gas) depends on the gas being used and the flow rate.
For a list of common K Factors see Section J.
4. Changing The Flow Rate Set-Point (Using An External Voltage Source)
The required flow rate is selected by adjusting the set-point voltage. The normal control signal voltage is 0-5VDC with 0VDC corresponding to zero flow and 5VDC being equivalent to the maximum rated flow of the unit. This input is linear and scaleable allowing different flow rates within the range of the unit to be selected. For example:
For a flow range of 0-500sccm:
/ K2)VDC for the full scale output.
1
A 5 VDC Input Signal would correspond to a flow rate of 500sccm
If a flow rate of 300sccm were required then the set-point would
be:
(300 ÷ 500) × 5 = 3.0VDC
M-4271/0707, pg. 15 of 26
If a gas other than the calibration gas is used then the adjusted maximum (full scale) flow for the unit should be calculated using the K Factor for that gas (see section C3 above).
A zero or negative set-point voltage will cause the solenoid valve to close fully. Whilst closed, the valve is configured to withstand pressures up to 60 psig (higher pressures on request).
Caution: The flow controller valve will open if the pressure
exceeds 60psig. For safety it is recommended that a separate positive shut-off valve is installed upstream of the controller.
5. Changing The Flow Rate Set-Point – FMA3400/3400ST Series Only
On the FMA 3400/3400ST Series the set-point may be input from an external source or be supplied internally.
For an external set-point, dip switch 1 should be OFF and dip switch 2 ON. See section C4 above for details of how to adjust the set-point using
an external voltage source.
For an internal set-point, dip switch 1 should be ON and dipswitch 2 OFF.
Adjustment of the internal set-point is made by turning the coarse and fine set-point potentiometers on the front panel of the display, with the gas is flowing, until the desired flow rate is achieved.
FMA 3400/3400ST Series Set-Point Potentiometers
6. Power Save Mode.
To improve valve performance and reliability over time, the FMA 3200/3400 Series features a Power Save Mode. This is activated after a prolonged application of a zero or negative set-point. When a control voltage greater than 0 VDC is applied after the Power Save Mode has been initiated there may be a short delay (1-2 secs) before the valve actuates.
M-4271/0707, pg. 16 of 26
7. Zero Adjustments
The zero should be checked as detailed in section C part 2. If an adjustment is needed the Zero Potentiometer should be carefully turned until the output (VDC) becomes zero.
Caution: Do NOT adjust the Gain Potentiometer when adjusting
the zero or the unit will need to be recalibrated.
Making Zero Adjustments Using a Small Flathead Screwdriver
Care should be taken to only make small adjustments to the zero potentiometer. If too much of an adjustment is made and difficulties are being experienced in achieving a zero reading then turn the potentiometer fully anti-clockwise and begin making small clockwise adjustments until a zero reading is obtained.
8. Recalibration
If recalibration is required please contact the Omega Engineering Customer Service Department.
9. Changing the Calibration Gas – FMA 3400/3400ST Series Only
The FMA 3400/3400ST Series may be calibrated for up to three gases. These gases, their corresponding flow ranges and accuracy specifications are detailed on the calibration certificate.
The calibration gas required is determined by selecting the corresponding dip switch on the front panel of the display.
M-4271/0707, pg. 17 of 26
FMA 3400/3400ST Series Dip Switches
D.
Dip switch 4 is allocated to the primary calibration gas. Dip switch 5 is allocated to the second calibration gas (if applicable). Dip switch 6 is allocated to the third calibration gas (if applicable).
To select the gas, the dip switch should be turned ON. All other switches
allocated to gases (i.e. 4, 5 or 6 except the required switch/gas) should be
set to OFF.
Maintenance and Product Care
1. General
Inlet filters should be periodically checked and cleaned or replaced as necessary.
Regularly check all electrical and process connections for damage or deterioration.
If the sensor is to be stored, keep both the inlet and outlet ports sealed.
Do not allow any liquid or moisture to enter the sensor or damage will occur.
2. Returning Units for Repair or Recalibration
To return a unit for repair or recalibration please contact the Omega Engineering Customer Service Department. An Authorized Return (AR) number will then be issued. The AR number should then be noted on the outside of the package and on any correspondence. Further details may be found on page 25 of this manual.
M-4271/0707, pg. 18 of 26
E.
Specifications
Series FMA 3200 FMA 3400 FMA 3200ST FMA 3400ST
Accuracy (including linearity)
Repeatability ±0.25% Full Scale*
Pressure Rating 150 psig (10.3 bar) 500 psig (34.5 bar)
Pressure Sensitivity ±0.02% Full Scale* per psi (per 69 mbar)
Temperature Rating Operating Range: 5 to 55ºC
Temperature Sensitivity
Valve Normally Closed
Body Leak Integrity 1x10-7 sccs of He
Wetted Materials Aluminum
O-Ring Material Viton®
Fitting Material Choose from acetal, brass, or stainless steel
±1.5% of
Full Scale*
Recommended Range (for best performance) : 10 to 40ºC
304 Stainless Steel 316 Stainless Steel
±1.5% of
Full Scale*
Second and
third gases
±3.0% F.S.*
Storage Range: 0 to 70ºC
±0.15% F.S.* or less per ºC
Positive Shut-off up to 60psig (4 bar)
±1.5% of
Full Scale*
303 Stainless Steel 304 Stainless Steel 316 Stainless Steel
±1.5% of
Full Scale*
Second and
third gases
±3.0% F.S.*
Epoxy
Recommended Filtration
Compatible gases Clean, dry gases compatible with wetted materials
Output Signal 0-5VDC, Impedance greater than 2.5 K
External Set-point Signal
Internal Set-point Signal
Warm-Up Time Less than 5 minutes
Integrated Display N/A 3½ digit N/A 3½ digit
Typical Power Consumption
Peak Power Consumption
Electrical Connections
Certifications CE Approved
N/A Front Panel
Standard: 12 VDC @ 250 mA (12.5-15 VDC)
“E” Suffix: 24 VDC @ 130 mA (22-25 VDC)
Standard: 12 VDC @ 400 mA (12.5-15 VDC)
“E” Suffix: 24 VDC @ 260 mA (22-25 VDC)
Integrated 36” (92 mm) cable, terminated with:
Standard: 6-pin Mini-DIN male (PS/2 Style)
89/336/EEC (EN 55011 & EN 50082-1)
20 microns or less
Optional inline filters available
0-5VDC, Integrated 2M load
N/A Front Panel
Adjustment
D1 Option: 9-pin D-Sub male
D2 Option: 15-pin D-Sub male
73/23/EEC Low Voltage Directive
Adjustment
*Specifications from 10-100% of rated flow. Linearity is best fit straight line. All calibrations performed with air unless otherwise stated on calibration certificate.
M-4271/0707, pg. 19 of 26
F.
Dimensions
ALL DIMENSIONS IN INCHES (MILLIMETERS IN BRACKETS)
FMA 3200/3200ST Series - 1/4” Stainless Fittings Shown
M-4271/0707, pg. 20 of 26
FMA 3400/3400ST Series - 1/4” Stainless Fittings Shown
M-4271/0707, pg. 21 of 26
G.
Gas K Factors
Gas
Acetylene C2H2 0.589
Air - 1.000
Argon Ar 1.438
Butane C4H10 0.260
Carbon Dioxide CO2 0.739
Deuterium D2 1.000
Ethylene C2H4 0.598
Freon 11 CCL3F 0.330
Freon 12 CCL2F2 0.354
Freon 13 CCLF3 0.385
Freon 14 CF4 0.420
Freon 22 CHCLF2 0.460
Germane GeH4 0.570
Helium He 1.458
Chemical
Symbol
K Factor
Hydrogen H2 1.011
Krypton Kr 1.440
Methane CH4 0.721
Neon Ne 1.443
Nitric Oxide NO 0.990
Nitrogen N2 1.000
Nitrous Oxide N2O 0.710
Oxygen O2 0.991
Ozone O3 0.446
Propane C3H8 0.383
Sulfur Dioxide SO2 0.690
Xenon Xe 1.437
These K Factors are given for reference only and are not intended as a recommendation of application suitability. Accuracy and response will be affected depending on the gas and flow range. Check the compatibility of all wetted materials before using any gas other than the calibration gas for the unit.
M-4271/0707, pg. 22 of 26
H.
Troubleshooting Guide
Symptom Possible Cause Method of Correction
Clicking noise from controller
No response
Inaccurate control
Unit in error mode Check there is sufficient
pressure and that the flow path is not restricted or blocked
Unit wired incorrectly Check wiring is according to
Section B5
Loose connection Check all connectors and
wiring
Damaged connector pins Contact Omega Engineering
Blocked flow path Check flow path for
obstructions.
Piping leak before sensor Check all piping and
connections.
Insufficient power Check the power supply
output and increase if necessary
Output load resistance too low Ensure the voltmeter or
data acquisition system or display has an impedance of greater than of 2.5kohm
Flow too low for the unit Ensure that the flow being
measured is within the capabilities of the unit
Unit damaged or faulty Contact Omega Engineering
Particles in flow path Add filtration before the
sensor.
Flow path obscured Remove any debris or
blockage in the flow path eg. PTFE tape.
Unit calibrated for a different gas
Gas composition is variable Contact Omega Engineering
Fittings have been changed Replace the factory installed
Moisture in gas Ensure gas is clean and dry
Insufficient warm-up period Allow the unit to warm-up
Zero drift Verify the zero and adjust
The gain potentiometer has been adjusted
Unit needs recalibration Contact Omega Engineering
Flow too high for the unit Ensure that the flow being
Check calibration certificate and apply a “K” Factor to readings if necessary.
fittings
for at least 5 minutes.
as necessary as explained in Section C
Contact Omega Engineering
measured is within the capabilities of the unit
M-4271/0707, pg. 23 of 26
Symptom Possible Cause Method of Correction
Inaccurate control
Problems with rezeroing
Insufficient or varying power Check the power supply
output and increase if necessary
Insufficient pressure Ensure the pressure (an
differential pressure) is high enough o operate the unit.
Varying pressure Check the stability of the
pressure regulation and improve if necessary.
Ambient temperature too high or too low
Output load resistance too low Ensure the voltmeter or
Gas temperature too high or too low
Unit damaged or faulty Contact Omega Engineering
Gas flow through unit not completely stopped
Severe fluctuations in the ambient temperature e.g. unit in direct sunlight
Unstable power supply Check the stability and
Insufficient warm-up period Allow the unit to warm-up
Place the unit in a suitable environment
data acquisition system or display has an impedance of greater than of 2.5kohm
Ensure the gas temperature is within the recommended operating range
Ensure there is no flow through the unit. The easiest way to do this is to plug both the inlet and outlet.
Carry out the rezero procedure in a stable environment
suitability of the power source
for at least 5 minutes.
M-4271/0707, pg. 24 of 26
.
M-4271/0707, pg. 25 of 26
M-4271/0707, pg. 26 of 26
Loading...