Omax M825 User Manual

Page 1
User Manual
Binocular Compound
Microscope
Model M825
MicroscopeNet.com
Page 2
Table of Contents
i. Caution……..……………………………………………………………1
ii. Care and Maintenance……….………………………………….….…2
1. Component Illustration………………………….…………………..…3
2. Installation ……………………………………………………………...4
3. Operation..…………….……….………………………….…………....5
4. Specifications…………………….………………………………….....7
5. Troubleshooting Guide…................................................................8
Page 3
www.microscopenet.com
i. Caution
1. Find the “ UP” sign and place the St yrofoam container on your table or bench so that the arrow is pointing upward. Open the shipping carton carefully to prevent any accessory items (i.e. objectives or eyepieces) from dropping and being damaged.
2. Do not discard the molded Styrofoam container. The container should be retained should the microscope ever requires reshipment.
3. Keep the instrument out of direct sunlight, high temperature or humidity, and dusty environments. Ensure that the microscope is located on a smooth, level and firm surface.
4. If any specimen solutions or other liquids splash onto the stage, objectiv e or any other component, disconnect the power cord immediately and wipe up the spillage. Otherwise, the instrument may be damaged.
5. All electrical connectors (power cord) should be inserted into an electrical surge suppressor to prevent damage due to voltage fluctuations.
6. Confirm that the input voltage indicated on your microscope corresponds to your line voltage. The use of a different input voltage other than that as indicated will cause severe damage to the microscope.
- 1 -
Page 4
www.microscopenet.com
ii. Care and Maintenance
1. Do not attempt to disassemble any component including eyepieces, objectives or focusing assembly.
2. Keep the instrument clean; remove dirt and debris regularly. Accumulated dirt on metal surfaces should be cleaned with a damp cloth. More persistent dirt should be r emoved using a mild soap solution. Do not use organic solvents for cleansing.
3. The outer surface of the optics should be inspected and cleaned periodically using an air stream from an air bulb. If dirt remains on the optical surface, use a soft cloth or cotton swab dampened with a lens cleaning solution (available at camera stores). All optical lenses should be swabbed using a circular motion. A small am ount of absorbent cotton wound on the end of a tapered stick makes a useful tool for cleaning recessed optical surfaces. Avoid using an excessive amount of solvents as this may cause problems with optical coatings or cemented optics or the flowing solvent may pick u p grease making cleaning more difficult. Oil immersion objectives should be cleaned immediately after use by removing the oil with lens tissue or a clean, soft cloth.
4. Store the instrument in a cool, dry environment. Cover the microscope with the dust cover when not in use.
- 2 -
Page 5
www.microscopenet.com
1 Components Illustration
1
2
3
4
9
10
11
12
13
5
6
7
8
1. Eyepiece
2. Eyepiece Tube
3. Diopter Ring
4. Viewing Head
5. Slide Holder
6. Stage
7. Condenser
8. Light Collector
9. Set Screw
10. Nosepiece
11. Objective
12. Stand
14
15
16
17
13. Coarse Focus Knob
14. Fine Focus Knob
15. Translation Knobs
16. Intensity Dial
17. Base
- 3 -
Page 6
www.microscopenet.com
2 Installation
2.1 Installation of the binocular viewing head
1) Loosen the set screw on the top of the stand and remove the plastic cover on the top.
2) Remove the cap on the dovetail of the binocular viewing head.
3) Seat the dovetail of viewing head into the socket on the top of the stand and tighten the set screw.
Caution: Do not release the head from your hand grip until you are sure the head is
installed securely.
2.2 Installation of the eyepieces
1) Remove the protective caps from the eyepiece tubes.
2) Insert the eyepieces into the eyepiece tubes.
2.3 Installation of the objectives
1) Adjust the coarse focus knob until the mechanical stage is at its lowest position.
2) Install the 4x objective into the nosepiece. Then in a clock-wise direction, rotate the nosepiece and install each succeeding higher magnification objective.
Note:
Inspect the objectives frequently for dirt or oil; clean if necessary. Use the 10X objective to initially focus the image of your specimen. When changing the objective magnification, rotate the objective nosepiece until you
hear a “click” sound. This ensures the objective is centered in the optical light path.
2.4 Installation of the light filter
1) Swing out the filter holder which is located underneath the condenser.
2) Insert the light filter into the holder, swing the holder back in.
3 Operation
- 4 -
Page 7
www.microscopenet.com
3.1 Adjusting illumination
1) Insert the power cord into the power socket on the microscope and connect it to the power outlet.
2) Turn on the main switch at the back of the microscope. 3) Rotate the intensity dial to increase or decrease the brightness.
3.2 Placing specimen
1) Place the slide on the mechanical stage.
2) Use the slide holder to gently secure the slide.
3) Turn the X and Y translation knobs to position the specimen for viewing.
Caution: Be sure not to allow an objective to touch a specimen slide when chan ging
objectives.
3.3 Focusing
1) With the 10X objective in position, raise the mechanical stage using the coarse focus knob until the specimen is close to the objective.
2) Turn the coarse focus knob until the specimen is in focus. Then use the fine focus knob to obtain a sharp image. You may now switch to another magnification objectiv e.
Tips:
To prevent your specimen slide from making contact
with an objective, turn the 100X objective in position and rotate the focus stop lever clock-wise all the way to the end so that the 100X objective will not contact the specimen while the stage is adjusted to its highest position.
3.4 Adjusting interpupillary distance
While observing with both eyes, hold the left and right eyepiece tubes then s wing them around the center axis. The interpupillary distance is correct when the left and right fields of view converge completely into one image.
3.5 Adjusting eyepiece diopter
1) Using the 10x objective and your left eye only, observe your specimen through the eyepiece and bring it into focus by adjusting the focus knobs.
2) Then observe the specimen with your right eye only through the right eyepiece. If the specimen is not in focus, turn the diopter ring on the eyepiece tube until a sharp image is obtained.
3.6 Adjusting condenser
1) Turn the condenser focus knob to raise or lower the condenser.
2) The condenser is raised when using high magnification objectives and lowered when using low magnification objectives.
Note:
The centering of the condenser and the light axis of the objective are factory adjusted.
- 5 -
Page 8
www.microscopenet.com
Do not attempt re-adjustment.
The highest position of the condenser has been factory adjusted. Do not attempt to
re-adjustment.
3.7 Adjusting iris aperture diaphragm
Move the iris diaphragm lever left or right to adjust the aperture size.
Note:
The iris diaphragm is designed to adjust the aperture
size, not to adjust brightness. Generally, opening the diaphragm to 70-80% of the N.A. value of the respective objective will provide an image of acceptable quality. If you want to observe the image of the iris diaphragm, remove one eyepiece and look through the tube. You will see a dark circle encroaching on the bottom of the tube.
3.8
Adjusting focus knob tension
The tightness of the focus knob tension has been pre-set at the factory. If the mechanical stage drops by itself, rotate the tension adjustment ring located inside the focus knob on the power switch side until the tension is in maintained. Tension Ring
Tension Ring
4 Specifications
- 6 -
Page 9
www.microscopenet.com
Model M825 Total Magnification 40X, 64X, 100X, 160X, 400X, 640X, 1000X, 1600X Objective Tube Length 160mm
30º inclined, 360º swiveling Siedentopf binocular viewing head
Viewing Head
Interpupillary distance 55-75mm Adjustable diopter on right
eyepiece tube Eyepieces 1 pair of WF10X 1 pair of WF16X Nosepiece Reversed revolving quadruple Objectives Achromatic 4X, 10X, 40X(spring), 100X(spring, oil)
Condenser
Focus Mechanism
Stage
Abbe, NA=1.25, w/ iris aperture diaphragm and filter holder
Rack and pinion adjustment
Coaxial coarse and fine focusing knobs on both sides w/ focus
stop, tension adjustable
Double layer mechanical stage Dimension: 5-1/2” x 5-1/2”
(140mmx140mm) Translation range: 3” x 2” (75mm X 50mm) Illumination Transmitted: 6V/20W Halogen, Variable intensity Power Supply AC 110V, 60HZ (US and Canada plug) Dimension 12-1/2” x 11-1/2” x 15-3/4” (32cm x 29cm x 40cm) Net weight 13 lb (6.34 kg)
5 Troubleshooting Guide
- 7 -
Page 10
www.microscopenet.com
OPTICAL PROBLEMS
Problem Cause Solution
Darkness at the periphery or
Revolving nosepiece not in click stop position
uneven brightness in the field of view
The bulb of light source is not at the center
Dirt or dust on the view
Dirt or dust on the eyepiece, condenser, objective, collector lens or specimen
No slide cover attached to the slide
Slide cover is too thick or thin
Slide may be upside down
Poor image quality Immersion oil is on a dry
objective (especially the 40x) No immersion oil used with 100x
objective Air bubbles in immersion oil Remove bubbles
Revolve the nosepiece to click-stop position by swinging the objective correctly into the optical path
Adjust the position of the bulb if possible
Clean the lens with a camera cleaning kit
Attach a 0.17mm slide cover Use a slide cover of the appropriate
thickness (0.17mm) Turn slide over so the cover-glass
faces up Check the objectives, clean if
necessary Use immersion oil
Problem Cause Solution
Lamp does not light when switched on
Condenser aperture is closed or open too much
Open or close properly
Condenser is positioned too low Position the condenser upward
ELECTRICAL PROBLEMS
No electrical power Check power cord connection Lamp bulb burnt out Replace bulb Fuse blown out Replace fuse
- 8 -
Page 11
www.microscopenet.com
IMAGE PROBLEMS
Problem Cause Solution
Specimen rises from stage
Image moves while focusing
surface Revolving nosepiece is not in
the click-stop position Blue filter not used Use daylight blue filter
Image tinged yellow
Lamp intensity is too low
Image is too bright Lamp intensity is too high
Lamp intensity is too low
Insufficient brightness
Aperture diaphragm closed too far
Condenser position too low Position the condenser upward
IMAGE PROBLEMS
Problem Cause Solution
Secure the specimen in the slide holder
Revolve the nosepiece to the click-stop position
Adjust the light intensity by rotating the intensity control dial
Adjust the light intensity by rotating the intensity control dial
Adjust the light intensity by rotating the intensity control dial
Open to the proper setting
Image will not focus with high power objectives
High power objective contacts slide when changed from low power objective
Slippage of focus when using the coarse focusing knob Fine focus is ineffective
Slide upside down
Turn the slide over so the cover glass faces up
Cover glass is too thick Use a 0.17mm cover glass Slide upside down
Turn the slide over so the cover
glass faces up Cover glass is too thick Use a 0.17mm cover glass Diopter adjustment is not set
properly Tension adjustment is set too
low Tension adjustment is set too
high
Readjust the diopter settings
Increase the tension on the
focusing knobs
Loosen the tension on the focusing
knobs
- 9 -
Loading...