Olympus PEF-V User Manual

ENDOSCOPE DISINFECTION GUIDE
VISERA™ PEF-V Transnasal Esophagovideoscope
Refer to companion ENDOSCOPE CLEANING GUIDE for cleaning information
HIGH-LEVEL DISINFECTION
Injection Tube
Fill basin with disinfectant. Attach channel plug and injection tube to previously cleaned endoscope. Immerse completely in disinfectant solution.
Channel Plug
4
Use syringe to completely fill air, water and suction channels with disinfectant solution.
RINSING
5
While the endoscope is immersed, disconnect cleaning accessories and use a clean lint-free cloth to wipe away any bubbles clinging to external surfaces.
6
CAUTION: Follow disinfectant
manufacturer's rinsing instructions.
Soak all equipment according to the disinfectant manufacturer's recommended time and temperature. Use a timer to measure exact contact time.
7
Connect channel plug and injection tube to endoscope. Use a syringe to flush water through air, water and suction channels.
Connect channel plug and injection tube. Inject air to remove disinfectant from air, water and suction channels. Remove endoscope from disinfectant.
DRYING
8
Remove endoscope from rinse water. Use syringe to inject air through air, water and suction channels, expelling the rinse water.
Disconnect channel plug and injection tube. Immerse endoscope and detached parts in water. Thoroughly rinse all external surfaces.
9
Alcohol
Use syringe to flush 70% alcohol through air, water & suction channels. Inject air to remove retained alcohol.
10
Disconnect all accessories from the endoscope. Dry endoscope and accessories with an alcohol­moistened cloth. Make sure equipment is completely dry prior to storage.
TRANSPORTATION
Hold all parts of the endoscope when carrying. Do not let the ends of the endoscope swing freely.
IMPORTANT
Use a compatible disinfectant.
Test for proper concentration per manufacturer's instructions.
Autoclave all autoclavable
accessories.
Store endoscope hanging
vertically with valves and caps removed, and with angulation lock in a free position.
Clean endoscope meticulously
prior to disinfection.
2003 Olympus America, Inc. / Printed in USA - 08/03/ PWC-100002