Olympus BH2 MJLT, BH2 MJLA User manual [gb]

IC INSPECTION MICROSCOPE
1-
This instruction manual has been prepared for use with the Olympus
IC Inspection Microscope Models putting the microscope into operation, it is recommended that you read this manual carefully and familiarize yourself with the use of this microscope in order to obtain the best performance.
WARNING
BH3-MJLT
and BH3-MJLA. Before
OLVMPUS
_.I_-
_
.
_
_-,_
I
;omradl
I.
1.
This microscope is a precision instrument. Always handle it with the care it deserves and avoid abrupt movements or
c.1'
.
on,
-1.L
. .
-
.-
;
;
~.
-.
,,-.:
~
--
--
--
-
.
-...-
--
-
-.
_-
_
;..
_
__
,
,:.
.'
.r.,
"
-.,.--.-..
,
{,,
,,,:
.
L
-.
.
I:
.
,
L..
,.I
.
-
-1
..---
-I
-.---
,'
.--.-
-
.
' , '
1.-
:,
.
'
;-
I
.'c---,--j:
'i,"
.
lA1.'
---. .
.
:
I
---I-----
shock.
2.
Do not subject the microscope to direct sunlight, high temperature and humidity, dust, or excessive vibration. The operating conditions are: Ambient temperature:
10°C - 35OC (50°F - 95°F)
Relative humidity: 15% - 85%
3.
Prevent dust from accumulating on the lenses and avoid leaving fingerprints on the optics.
4.
Unplug the power cord before replacing the light bulb or fuse.
5. Connect the grounding wire of the microscope.
6.
Use the tension adjustment ring to adjust the rotational tension of the coarse focus adjustment knobs.
7.
Confirm that the line voltage selector switch at the bottom of the microscope base is set to your local mains voltage.
I
Ao~licable Arms
and
Focus Units
I
.
Arm
Focus unit
Stage
8"
BH3-SIC6 BH3-SIC8 BH3-SIC106 BH3-SIC128
*
BH3-SIC10
*
BH3-SIC1410
The items marked with
l~sted above, use of different models of mechanical stages require different models of focus units and horizontal
8. As
0
0
0
0
1
'*'
are scheduled for future sale.
arms. Please consult your local
9.
To power the vertical illuminator and the transmitted light illuminator simultaneously, a power supply unit VH3) for the
+lom
-,
I
0
10
agentldealers for further details on the selection and/or exchange of these models.
F02
0
F04
0
0
0
-
,-
+FOG
0;
0
Halogen lamp housing
(for transmitted
BHS-LSHSO
BH3-LSH50-10
0
0
101
I01
light)
0
0
halogen light source is required.
UV
10. The breath shield and
protective shield mounting pins (paragraph 3-3
)
permit attachment of the shield alternatingly.
The pins have a central tap to accept the clamping screws of the breath shield. Exchange the pins without central tap.
1.
Lens surfaces must always be kept clean. Fine dust on lens surfaces should be wiped off gently with gauze pads. Carefully wipe off oil or fingerprints on the lens surfaces with a gauze pad moistened with a small amount of or a
3
:
7
mixture of alcohol and ether.
2.
Do not use organic solutions to wipe the surfaces of various components. Plastic parts, especially, should be cleaned
xylene.
w~th neutral detergent.
3. Fine dust on the half-mirror units of the universal illuminator should be lightly blown off.
If
dust still remains, contact
an Olympus office, as listed on the back page of this manual.
4.
Do not disassemble the microscope for any reason. Only authorized Olympus service personnel should make repairs.
5. When not in use, cover the microscope with the vinyl dust cover provided and keep it out of areas of high heat and humidity.
1.
STANDARD CONFIGORATIONS
.-
".,
A
-,.-
2.
NOMENCLATURE'
......
-.
.....
;
.
-
---.
3.
ASSEMBLY
_
.
_.
3-1
Assembiy._Outline
3-2
'
Preparation fpr Assembly
3-3
Detailed Assembly Procedure
4.
OPERATION
...........
1
5-1
I I
1
1
I
1
Reflected Light ~rl~htfieldl~arkflelij 0bsedtl.on..
5-2
Reflected
5-3
Refie~ted~Light
5-4
~rensrnitted
_
-'
I._
Light
-
Light
--.
-
.
,..
-
. - --.
:
.
--
......
,
,
,
. %-I.
-
.
,
>
-
,
.
.
.
-
1.
.-
.
-.
'-.
_..--
.
'.
-
.-..
.
.
, ,
_-~-
".
.,
...
.
.
,..
.._.
-.
.
.-
.
Nomarski Differential Interference Contrast Obsepation
Slmp!e,
Observation in Polarlze,d Light
Obsetyatibn
-
--
A
.
-
,
. 7 .
,
L.
..-A
.
....----.
.
,
I--
!:'.
-~.
g
,
..
..
.
.
.
.
--
9
12
13
25
33
34
35,
36.
-
..
1
]
1
1
1
-.
.
1
.5-5
6.
......
Transmitted Light Simple Observation in Polerized.Light
PHOTOMICROGRAPHY
__
_;_._
&_._
.....-
.............
37
1
39.
-
.-
. .
- . -
L
-.i.-
...
..-
--
---
.
-:
The following shows only typical examples of the standard configurations. Check your configuration against the price list, catalogs, etc.
IC Standard Configurations
Model
BH3-MJ
Observe- Photo- Photo- Observa- Phote Observa- Photo-
tion micro- Observation only micro-
only
Component 163B 863U 1838 ' 1830
BH3-MJW-F
Microscope
Stand
-
Power cord UYCP
Binocular
Tlltlng b~nocular tube BH2-TBI Tnnocular tube BH2-TR30 Tllt~ng tr~nocular tube
Motorized
nosepiece
Motor~zed
nosepiece
uves
Un~versal vert~cal lllurn~nator
Br~ghtf~eld half-m~rror unlt BH2-U6F
Darkf~eld half-m~rror unlt BH2-UDF ND f~lter sl~der. ~nclud~ng 8H2-UND
NO12 Tlnt plate, lnclud~ng BH29TP530
NO12
Polarizer BH2-UP0
Analyzer BH2-UAN
~cof
Breath shleld Yellow f~lter Llght balanc~ng f~lter 20LBD3-W
Contrast f~lter 201F550-W
Halogen lamp houslng BH2-ULSH45/50
(50W
Halogen bulb Halogen lamp hous~ng BH2-HLSH451100
(1
Collector lens BH2-HCL
tube BH2-B130
5
revok~ng BH3-5MRE-M
for
M
objectives
5
revolving
for Neo objec- BH3-5NRE-M
f11ter
and
shutter
frlter
cover
OOWI
BH3-MJLT2-F BH3-MJ LA4-F BH3-M JLT4-F
BH2-?lR
BH2-UMA
I
(
BH3-UCO
I
BH3-BSH
1
2OY48-W
45LBD2-N
431F550-W45
IZVSOWHAL-L-L-LI
!
-
-
-
-
I
-
-
I
I
-
-
ll/l!l
1
-
-
-
-
Reflected
W
gnphy
tight
Version Version
BH3-MJLA4 BH3-M J LT2 BH3-MJLT4
-
-
-4,
- -
1
1
1
1
-
-
-
1
-
1
1 1
1 1 1
1
1 1 1
1 1
- -
---
- - -
-
1
-
1
-
-
-
-I-
Ill
-
11-
->
-1-
1
2
I
-
-
1
1
2 2
- -
-
1 1 1
--
-
-
-
-
1
-
--
-
-
graphy
88311
i
--
-
1
1 1
1
I
1
---
I
I
Reflected and Trensm'kted
tion
on!y
1638 86311
rmcw
graphv only
-
-
,
'
-
-
- -
1
-
-
1
-
-
1
1
1
7
-
-
1
-
1
1
-
-
-
1
-
--
1 1 1
1
1
1
-
I
lI-I
1
1/111,
1
-
l
1
/
1 1 111
11
1
-1
-
1
1
-
-
1
1
Light
t~on mcro-
graphy
1830 88311
-1-
-
-I-
1 1
1
1,
1
-
- -
- -
-
-
1
1'1
1
1
-
1
1
I
-
-1
1
-1
I
1
Ill
)I1
-
-
2
1 1
-
-
-I1
L
IC Standard Configurations
Model
Component Halogen bulb
Halogen lamp hous~ng
(50WI
for
transm~tted l~ght
Halogen bulb JCl 2V5OWHAL-L Mechan~cal stage wlth
right-hand low drlve controls
Stage gnp for nght-hand s~de
Wafer holder plate
6'
and
5'
rlng Wafer holder plate.
lncludlng wafer holder rlng
5' x 5'
7' Nomarsk~ prlsm
attachment
Screwdr~ver ~llen" wrench
Ob~ect'ves
Standard mlrror for color temperature determlnat~on
LB eyeplece 10X LB eyeplece 10X
with
mask LB photo eyepiece
3 3X
wafer holder
8'
and
mask holder
x
7' mask holder
MSPL
achromatic
achromatic
Nee
achromat~c
universal
verscon
bulk-ln 35tT~n
JC12V1
OOWHAL-L
-
I
BHS-LSH-50
+-
'
BH3SIC6-R 1 1 BH3-SlC8-R BH3-SGR-R
BH3-WHP6 BH2-MrHR65
6'
BH3-WHPR86
BH3-MH5 BH3-MH7 U-NICS-N
U-NIC 10-N U-NIC20-N U-NIC50-N U-NIC 100-N
AA7213 MA7578 MS
MSPLlOX-T MS PL20X-T MS PL5OX MS PL100X
Neo SPL5X Neo SPLIOX-T
SPL
Neo SPL2OX-T Neo SPLlOOX
SPL
Neo SPL5X-NIC Neo SPLlOX-NIC Neo SPL20X-NIC ~eo
,
M-SM2
.-
PL5X
SpL50X
SPLSOX-NIC
I
~~~lox
35WHK1 OX
NFK
3 3x
LD
BH3-MJLA2
Observa-
tion only
1638
-
- -
--
-.
,
-
-
I
-
I
-
-
-
-
-
each
-
-
-
1
I
I
-
-
1-
Reflected Light Version
Photo­micro-
graphy
86311
2
-
-
1
-
--
1
1 1
1
1
-
1
-
1
-
each
1 1
-
-
1
-
each
11-
1 12
1
Reflected end Transmitted Light
BHBMJLA4
Photomicrography
1838
-
-
1830
-
-
Photo­micro-
gmphy
88311
2
-
BH3-MJLTZ
Observa-
tion
only
1638
-
'
1 1
!
I
-
-
2!2
1
1
I
I
I
1
1
-
-
-
1 1 1
-
1
-
I I I1
-
-
1
1111
-1-1
*"TI
1
Photo­micro-
graphy
863U
2
1
-
-
Version
BH3-MJLT4
Observa-
tlon
only
183D
-
2
-
-
- -
1;1
Photo­micro-
graph~
883U
2 1
I
2
-
1
1
I
-
I
I
I
I
ekh
-
1
-
-
-
-
I
each
j
-
-
-
-
-
-
1
1
-
-
-
1-
-
1-
each
each
I
'
I
-
-
-
1
1
-
1
1 1
-
-
-
-
1-
each
1
-
1-
each
-
1 1
-
-
-
I
-
lrl
1
j-
-
-I
2
-
-
1-
each
I
--
1, 1
-
1
-
/-
r-
-
*
I
-
-
1-
each
1
I1
1
1
-
-
21
-
-1
each
I
-I-'-
~
I I
I
each
11
I
I
-
-
-
1
-
1
1 1
1-
1
1
I
IC
Fluorescence Standard Configurations
Model
Component 363F 383F 363F 383F
I
BH3-MJLA2-F
r
Microscope
Power cord UYCP Trinocular tube BH2-TR30 Tilting trinocular tube BHZ-STR
Motorized
nosepiece for M objectives Un~versal venical illuminator / BH2-UMA Brightfield hallf-mirror unit BH2-UBFL
ND05 filter
Roof cover Breath shield Light balancing fllter Fluorescence supplementary
unit consisting of slider for supplementary exciter filter. adapter to fit extension tube 8H2-URF
BH3-UET8 with BH2-UMA, supporting block and UV protective shield
Extension tube BH3-UET8
Stand
-
5 revolving BH3-5MRE-M
BH3-MJLA4-F BH3-MJLT2-F BH3-MJLT4-F
--A
I
I
BH2-UCO
/
BH3-BSH
I
45BD2-N
I
1
-
BH3-MJLA2 BH3-MJLA4 BH3-MJLT2 BH3-MJT4 BH3-MJL
1
-
-
-
1
-
-
.
-
-
-
-
2
-
.
.
-
-.
1 1
--
1
1
1 1
1
1 1 1 1 1
1
-
1
--
1
-
1
1
-
1
-
1
1
-
1 1
1
1 1
-
1
1 1
1
1
I
lI1-
ll11-
1
1-11 1
I
-
I
I
I
1
i
I
F
L-SET
-
-
-
1
-
-
-
-
1
1
I
-
1
IC Fluorescence Standard Configurations
BH3-MJLAZ BH3-MJLA4 BH3-MJLT2 BH3-MJT4 BH3-MJL
Component
Fluorescence lamp hous~ng
-
-
Mercury burner
---
-
Power supply Center~ng screen Halogen lamp houslng
(for transm~tted I~ght) Halogen bulb
--
Mechan~cal stage w~th r~ght-hand low dr~ve controls
Stage grtp for nght-hand s~de
Wafer holder plate
6'
and
5'
wafer holder rlng
Wafer holder plate, ~ncluding
8'
and
6'
wafer holder nng
5'
x
5'
mask holder
7'
x
7'
mask holder
,
ob~ectlve
1
LB
eyeplece
LB
eyeplece
In
35
LB
photo eyeplece
1 OX
mm mask
(50W) BHS-LSH-50
MS PL
achrornatc
10X WHKlOX
mth
bullt-
3 3X NFK 3.3X
1
BH2-LSRF
-
USHlO2D
-­BH2-RFL-T3
.
-
BH2-SGRF
JC
I
ZVSOWHAL-L
BH3-SICSR BH3-SIC8-R BH3-SGR-R BH3-WHP6 BH3-WHR65 BH3-WHPR86
BH3-MH5 BH3-MH7
MS
PL5X
MS
PLlOX
MS
PL2OX
MS
PL50X
MS PLlOOX
35WHK1 OX
LO
-~
363F
I
1
--
1
1
--
-
-
1
-
1 1
--
1
-
1
I--
1-each 1-each
1
11/
1
383F
-
I
-
-
I--
1
1
-
-
1
1
1
--A
-
-
-1,-
A
-
363F 383F FL-SET
1
­1 1
:
2
-
,-A
lI-1-
-
1 1
1
*
I
-
1
-each
1
-
-
1
-each
-
2
1
-
1
1
1
1
-
-
-
-
-
-
-
-
-
-
1
-
1
1
1
1
1
-
LCD
Standard Configurations
Component
1
Microscope
Power cord
--
--
Tlltlng
Motorized
noseplece for M objectlves
Motorized
nosep~ece for 5 Un~versal venlcal ~llum~nator BH2-UMA Br~ghtf~eld half-m~rror unlt I BH2-UBF Darkf~eld half-rn~rror unlt Polanzer BH2-UP0
Tint
plate. ~nclud~ng Polar~zor for lransm~rred Rotatable analyzer BH3-UAN360 Llght
Roof cover Breath sh~eld Halogen lamp
for transm~tted llght
Halogen bulb Halogen lamp Halogen bulb
Collector lens
Stana
--
--
binocular
5
revolving
-
revolving
Neo object~ves
balancing
housing
housrng
tube
ND12
light
fllter 20LBD3-W
(50W
(100W
BH3-MJLT4-F BH3-MJLT4S-F
1
UYCP
I
BH2-TBI BH3-SMRE-M
BH3-5NRE-M
I
BH2-UDF
f~lter
BH2-UTP530
BH3-POL
,
45LBD2-N
,
BH3-UCO BH3-BSH BHS-LSH-50
BH3-LSH50-10
JC12V!XWHAL-L
BH2-HLSH45/100
I
JClZVlOOWHAL-L
BHZ-HCL
\
Model
-
BH3-MJLT4
-033
U
1
-
--
---.-a
1
1
-
-
A-
1 1
1
-.,-
1 1 1 1 1
--
--
1 1 1 1 1
-
2
1
22,2]2
1
BHSMJLT4
-1
B
1
-
--
-
-
-
1
--
1
--
1
-- - -
-
1
-11
,--.
-
-
1
1
1
-
1 1 1
--.-.
1
-
.
--
2 1
1
06
U
1
-
-
-
-
1
--
-
--
1
---
-
-++-j.f]IT
1
j1;
!lI1l!
1
l!-
1
-
1
-.
1
1 1 1 1
-
2
1
1
BHSMJLT4
B
1
-
1 1
:-I1
1-
1
lil 1
1 1 1 1
-
2
111
1
-1
28
I
BWLT4SBH3MJLT4S
-1 01
U
1
-
1 1
U
-
7'
1
R-~
17
111
lI1
1
111 1
1
1 1 1 1
-11
1
1
1
1 1 1
-
212
1
I
?I2
1
111
-141
U
-
1-
--
1
-'
1 1
1
I
1
1
-
1 1 1
-
1
2
1
LCD
Standard Configurations
Component
t
LWD MSPL100X-LCD
Neo SPL100X-NIC
ULWD
Neo
SPLlOOX
I
U-NIC
Nomarskl
Screwdrcver Allen" wrench LB eyepiece 1 OX WHKlOX
Mechanical
r~ght-hand low drlve
controls BH3-SIC106-R
set
U-NIC set
stage wlth
U-NIC10-N 1 U-NICUL2O-N
U-NICUL50-N U-NICUL80-N
AA7213 1
MA7578 1 2
BH3-SIC8-R 1
Model
BH3-MJLT4
-088
-
--
2
BH3-MJLT4 BH3-MJLT4
-1
06
U
I----
-1
28
B
1
1
1
-I-
1
BH3-MJLT4S BH3WLT4S
-1
01
U
U
-1-
-
1
1
2
1
111
1
-141
U
1
-
1
-
1
2
Stage plate glass
Stage
gnp
for nght-hand s~de
Rubber damper to absorb shock
BH3-SICl28-R 1 1 BH3-SICIO-R
]
BH3-SIC1410-R
BH3-SPG 106 1 BH3-SPG128 BH3-SPG 10 BH3-SPG 141 BH3-SGR-R BH3-DMP
I
-I_
-
-
1
I
_
0
-
1
-
-
-
1
-
-
-
-
1
ljl
-1-
-
1
-
-
-
-
-
1
-
-
-
-
1
-
- -
-
I
!
-
l1-
-
1
1 111
1 1
I
'The illustration below shows the
BH3-MJLT4-883U
combination.
Eyepiece
-
Observation tube
,Vertical illuminator
Objective magnification
Vertical illuminator connecting cord
button
Levelling
Lamp housing for
transmitted light
screw1 Stage clutch knob
evolving
nosepiece control box socket
Revolving nosepiece remote control box
(optional)
Universal Vertical llluminator
ktandard Accessories for universal Illuminator
Filter insertion slot
Field iris diaphragm centering knob
1
.I.
_
-. .
Half-mirror unit
indicator dial
Analyzer insertion
UV-protective shield1
breath shield mounting pins
Half-mirror unit
selector knob
I
Accessories for Universal Venical llluminator
slot
Huxlllary Ins
d
*\\-'.-
/
'-;ertion knob
'I
ND filterhint plate \ \PoIarizer insert~on
insertion slot
I
I
\
Eimlrl
I
,V,"
iric
rlianhranm
I,,"
".",...."J...
.
lever
.---.
.
.
.
slot
Brightfield half-mirror unit
BH2-UBF
Universal analyzer slider Universal tint plate slider Universal polarizer slider
BH2-UAM BH2-UTP530 BH2-UP0
Nomarski prism attachment Nomarski prism mount Brightfield objective adapter
U-NIC NRE-MAD
Darkfield half-mirror unit
BH2-UDF
3
3-1
Assembly
Outline
The dlagram below dlagram covers all modules, select the steps pertinent to your modules and assemble fluorescence
+r
Before assembly, wipe off all glass and connecting surfaces with a soh cloth or gauze in order to remove any dust or dirt and carefully follow the assembly procedure illustrated below.
"
Modules necessary for the BH3-MJLT version
"
Optional modules
illustrates
comblnatlon, refer to the Instruction manual for the reflected llght fluorescence attachment @HZ-URFL)
(available
the assembly procedure and the numbers lndlcate the order of assembly. Although the
accordingly.
for some versions)
Motorized revolving
nosepiece
When assembling the
Wafer holder ring
Wafer holder plate
Mask holder
-
I
,
A
chip
-1
!-1
II*
Stage plate Stage plate glass
0
7
'9
Stage grip
I
I
.
DarK~eld objective
adapter
..
0
I
t
Nomarski prism attachment"
Brightfieldldarkfield (Neo) objective
Q
El
II
Polarizer for transmitted light illumination
GI
3
3
t
2
;
:I
i
1
i
i
f
i
Eyepiece
Brightfield Darkfield
half-mirror half-mirror
I
:
I
t.
Slider with
360"
rotatable
Roof cover
Filter for universal
verrical illumintor
I
1
I
!
i
1
-,,
Breath shield
Breath shield mounting bracket
Tint plate sliderO
Halogen lamp housing for universal vertical illuminator
Polarizer
0
slider
Collector lens
Microscope stand
Halogen bulb
12V
50W
(JC
12V
Power cord
halogen bulb
50W
HAL-L)
Halogen lamp housing
for transmitted light
m
Stages and Related Accessories
Wafer holder ring
For
4", 3" wafers (BH2-WHR43)
For
Y,
4"
wafers (BH2-WHR54)
For 6",
5"
wafers (BH2-WHR65)
I
Silica chip
Q
(BH3-HCH)
Water holder plate
(BH3-WHPR86)
I
Wafer holder plate
(BH3-WHP6)
I
5" x 5"
(BH3-MH5)
4" x 4"
(BH3-MH4)
6" x 6"
(BH3-SP6)
mask holder
mask holder
stage plate
7" x 7" mask holder
(BH3-MH7)
I
6"
x
6"
mask holder
(BH3-MH6)
8"
x
8" stage plate
(BH3-SP8)
Stage plate glass
(BH3-SPG8)
Stage plate glass
(BH3-SPG6)
6"
unit
BH3-SIC8
-l
for
6"
x
6*
stage plate
(BH3-SP106) (BH3-SP128)
Stage plate glass
(BH3-SPG106) (BH3-SPG128)
12" x 8" stage plate
12" x 8" stage
IIBH3-SIC128)I
10"
x
10"
stage
14* x 10" stage
34
Preparation for
Assembly
Fig.
The stage will begin to slide at an inclination of
clutch disengaged, because the free-floating gearless rail system is ex-
it
tremely smooth. Therefore, before installation of the microscope.
is very important to properly level
i
1.
Gently tip the microscope cover, as shown In Fig.
2.
Determrne microscope base
selector
1
After placing the microscope stand, individually turn both levelling screws
@at the bottom of the microscope base until they make contact with the table surface. Then turn the screws a half or one full turn further to stabilize the microscope.
rf
the l~ne voltage selector switch @ at the bottom of the
IS
set to your local voltage. If not, set the line voltage
swltch to the correct posltion using a
lo
(71400
--
-
-I-
1.
screwdriver
mm) with the
the
table
i~i~~<
..
.-
or slm~lar tool
Fig.
2
Detailed
Back
Assembly
Procedure
1.
2.
3.
4.
5.
6.
7.
8.
Remove the transport locking plates @ on the rear of the stage using the
Allena wrench provided. Turn the coarse focus adjustment knob @ to bring the focus unit
down to its lowest position.
Place the stage gently onto the focus unit and align the mounting holes
@
on the stage with the screw holes @ on the focus unit.
Attach the stage to the focus unit by lightly tightening the four hex screws
with the Push in the stage as far as it will go, making sure that the stage does not
make contact with the pillar. Securely clamp the stage with the Allen" wrench. Remove the transport locking plates @ on the front of the stage with the
Alleng wrench. Attach the caps to the four screw holes.
*
*
*
Allene wrench provided.
Lower the focus unit as much as possible before removing the stage.
The stage can also be rotated a full
Keep the transport locking plates removed from the front and back of the stage in an appropriate place and use for reclamping the stage when it is removed.
180"
for mounting.
@
Fig.
Fig.
3
*
For
BH3-MHL
Interference Contrast Attachment, perform the procedures discussed
in paragraphs
*
When mounting the brightfield objective to the darkfield revolving
nosepiece, perform the procedure in paragraph
1.
Paste the labels provided near the objective apertures.
2.
Screw the objectives into the apertures marked ascending or descending order of magnification.
4
1.
Loosen the clamping screw @ of the Nomarski prism attachment, shown in Fig.
6,
with the Allena wrench and remove the Nomarski prism mount
@,
shown in Fig.
2.
Screw the Nomarski prism mount @ securely into the nosepiece with the mounting tool provided.
combinations provided with the Nomarski Differential
Q
and
.
.
'A'
through
5,
from the Nomarski prism attachment.
'E",
in either
Fig.
5
Fig.
Fig.
6
7
Similar to the objective mounting procedure discussed in Section
the
Nomarski prism attachments onto the circular dovetails of the Nomarski prism mounts, and temporarily clamp the prism clamping screw @ with the Allen8 wrench.
Ir
Mount each Nomarski prism attachment in such a manner that the straight line passing through the centers of the nosepiece and the optical axis of each objective (Fig.
screws only after all attachments are mounted correctly.
Ir
When using brightfield objectives, screw the brightfield objective adapters (NRE-MAD) into the
mounting the objectives.
1.
Gently insert the motorized revolving nosepiece @ along the slide dovetail
@
as shown in Fig.
Ir
DO not force the nosepiece when tilted.
Ir
A slight resistance may be felt at the bottom of the slide dovetail due to the engagement with the connector. Insert the nosepiece as far as it will go.
2.
To clamp the dovetail slide, use the 3 mm Allena wrench provided. to lightly tighten the hex screw
7,
pressing the nosepiece against one side of the slide dovetail.
3.
Clamp the revolving nosepiece by tightening the four hex screws @ shown in Figs. the Allena wrench.
7,
8
(short screws for the top and long screws for the front) using
7.
@
6).
Tighten the prism clamping
Nomarski prism attachments before
on the side of the arm as shown in
2,
mount
Fig.
?
I
i
Fig.
I
8
Align the breath shield mounting bracket with two pins on the front of the vertical illuminator and insert it. Securely clamp the screws with the Allens wrench.
Fig.
9
I
Fig. 10
Mounting.4he Universal Vertical Illuminator
Lamp
\Housing
1.
Fully loosen the vertical illuminator clamping screw @ as shown in Fig.
10
with the Allenm wrench, and attach the vertical illuminator to the top of the motorized revolving nosepiece. Clamp the unit lightly so that it is parallel to the arm. (Fig.
+
Fully tighten the clamping screw @ only after having attached the
lamp housing. (Fig. 10)
1
1)
--
and
-
?Figs.
-
9,
-
30)
..
5i
!!
Fig. 11
Fig. 12
2.
Mount the collector lens and lamp housing.
[
For the 50W Halogen Lamp Housing (BH2-ULSH)
Screw the BH2-UCLW collector lens @ provided with the BHZ-ULSH lamp housing into the universal vertical illuminator, insert the lamp housing
@as far as it will go, and then completely tighten the lamp housing clamping screw
1
Screw the BH2-HCL collector lens @ provided with the BH2-HLSH lamp
housing into the universal vertical illuminator, insert the lamp housing as far as it will go, and then completely tighten the lamp housing clamping screw
I
For funher details, refer to the instruction manual for the BH2-URFL reflected light fluorescence attachment.
*
*
3.
After mounting of the lamp housing is completed, adjust the universal vertical illuminator so that it is parallel with the arm and completely tighten the illuminator clamping screw
@.
For the lOOW Halosen Lamo Housina (BH2-HLSH)
8.
For Fluorescence Lam~
The fluorescence lamp housing cannot be mounted to the 10"
Use the BH3-UET8 extension tube especially designed for the BH3-
MJL.
HOUS~~~(BHP-LSRF)
@.
(Figs.
-
9,
10)
(Fig. 11)
(Fia. 12)
--
]
1
@
1
arm.
Fig. 13
Fig. 14
~Mou~t1n~g;MakIalpppn
1.
!
*
The BH3-MJL is available with two types of light sources for the
universal vertical illuminator
(
Mounting the Bulb (Fig. 14)
1.
Pressing the spring levers @ in the direction of the ARROWS, insert the pins of the halogen bulb the socket holes
2.
Release the spring levers to hold the bulb in position.
*
Never put the lOOW bulb in the 50W lamp housing.
*
DO not touch the bulb with bare fingers. If any fingerprints or din are on the bulb, wipe clean.
*
Before replacing the bulb during microscopy, unplug the power
cord or turn
I
Mounting the Socket and Cord Assembly
1.
Loosen the bulb socket clamping knob
2.
Aligning the bulb socket clamping knob with the positioning groove @ of the lamp housing, and tighten the bulb socket clamping knob.
-
.
@.
off
the power and wait until the bulb cools.
insen the socket and cord assembly into the lamp housing,
Bulb,and
-.A----.&
(JC
12V
Bull@ike$for
-A
:
-
50W or lOOW halogen.
50W HAL-L or
12V
@.
-
-
100W
the
VerticalI\
=
HAL-L)
(Fig.
into
15)
J
1
]
Fig. 15
16
Fig.
*
Use the BHS-LSH5O for
1
Mounting the Bulb
1.
Loosen the lamp housing clamping knob @ of the lamp housing for the transmitted light illuminator and remove the bulb socket.
2.
Pressing the spring levers
pins of the halogen bulb
8"
arm and the BH3-LSH50-10 for
@I
in the direction of the ARROWS, insert the
(JC
12V
50W
HAL-L)
into the socket holes
10"
arm.
1
Q.
Fig. 17
I
Mounting the Bulb Socket and Lamp Housing
1. Fully insert the bulb socket into the lamp housing for the transmitted light illuminator and tighten the lamp housing knob. (See Fig.
2.
Insert the lamp housing into the lamp housing mounting port on the micrsocope stand as far as it will go.
*
When the
10) into the mounting poiat the bottom
1. Fully loosen the observation tube clamping knob @ of the universal vertical
illuminator.
2.
Mount the observation tube on the universal vertical illuminator and tighten the observation tube clamping knob securely, after aligning the eyepiece sleeves to
10"
arm is in use, insert the lamp housing (BH3-LSH50-
bf
mint forward.
16
the arm.
on page
17.)
4
EM
Fig.
18
Fig. 19
lnsen the eyepieces into the eyepiece sleeves
--
When Using the Superwide Field Observation Tube
(BHP-SWTR) (Fig.
1. Aligning the positioning pin of the field-of-view eyepiece with the positioning groove @ at the upper edge of the right eyepiece sleeve, insert the eyepiece.
2.
Aligning the positioning pin of the eyepiece (SWHKlOX) with the position-
ing groove
eyepiece.
[
When Using the Trinocular Tube
1.
Aligning the positioning pin of the field-of-view eyepiece with the positioning groove @ of the right eyepiece sleeve, insert the eyepiece.
2.
Insert the eyepiece (WHKlOX) into the left eyepiece sleeve.
@
at the lower edge of the left eyepiece sleeve, insert the
(BH~-T~O/BH~-TTR)
@.
(35
(~=9(
(35
WHKlOX)
SWHlOX)
18)
Fig.
20
Mo-unting
1.
Remove the right and left half-mirror unit compartment cover plates (Fig.
19)
fingertips.
.k
Half-mirror unit compartment cover plates are attached with a magnet. Set them aside carefully.
2.
Remove the auxiliary mounting knob @ provided with the universal illuminator shown in Fig. 21 and attach it to the brightfield half-mirror unit Looking at Fig. brightfield is heard.
+
A
click will operate as soon as the half mirror unit is fully inserted
and its side panel is flush with the body of the vertical illuminator.
.k
Half-mirror units can be inserted from either the right or left side of the universal vertical illuminator.
.k
For brightfieldidarkfield and universal versions, mount the darkfield
half-mirror unit
3.
Remove the auxiliary mounting knob from the half-mirror unit and reattach
to the universal vertical illuminator.
4.
Reattach the half-mirror unit compartment cover plates.
5.
Set the half-mirror unit indicator dial half-mirror
.k
When the brightfield half-mirror unit is mounted on the right-hand side of the illuminator, set the left-hand side half-mirror unit indicator
dial to the
the
Half-mircxunie
from the universal vertical illuminator by grasping them with your
22,
hold the auxiliary knob @ and carefully insert the
half:mirror unit into the half-mirror unit compartment until a click
(BH2-UDF) following procedure
unit, to the 'B.F.' or 'D.F.' position.
*B.F.* position.
-
-
(Figs.
20,
(BH2-UBF).
2.
@,
located opposite the mounted
21,2;!?
@
Fig.
Fig.
21
22
Insert the connecting cord plug @ of the vertical illuminator socket and cord assembly
*
When connecting, align the connector pin of the connecting cord with the groove on the microscope base.
Into the receptacle at the rear of the microscope base.
Fig.
23
L
=-TP
Typical example
of
3-pin grounding
Fig. 24
Fig. 25
plug
Connecting
Connect the primary cord with a 3-pin plug (it will fit into a ground type power outlet, and no need to connect it to any other grounding device)
to
a
receptacle (wall mains outlet). (Fig.
*
If a 2-pin grounding plug is used, ground the microscope to a properly grounded device (except a gas pipe). If necessary, use an extension cord.
Ir
This microscope incorporates a noise filter in the electric circuit
built-in the microscope base, bleeding a very low voltage current in order to reduce effect of any external noises. Therefore, if the conductive part of the stand is touched without grounding, an electric shock may sometimes be felt depending upon the humidity conditions of the hands and foot wears.
1.
Turn
ON
2.
If
the LED light of an objective magnification button is blinking, rotate the
revolving nosepiece manually until it clicks.
3. If the lit LED of the objective magnification button and the engaged objec­tive match. the Setting is complete.
4.
If not, depress the lit button for 2 seconds. The lit button will change from
"A'
through to 'E". Release the button when it matches the engaged objec-
tive.
5.
Attach the labels showing the magnification powers of the mounted objectives to the convex sections buttons.
_.
_->__
the
Power
_
the power switch.
@
Cord
-.__
-
_
_
_
_
.___
__
-
-__
(Fig.34)-
24)
below the respective objective magnification
m
D
$
3
m
2
Fig.
Fig.
26
27
*
Lower the stage as much as possible to prevent it from hitting the
objective.
I
Mounting the Wafer Holder Plate
1.
Insert the edge of the wafer holder plate @ into the stage.
2.
Pressing from the top, make sure that the wafer holder plate is firmly in position.
3. Tighten the holder clamping screw
1.5
with the
4.
Move the stage towards the right as far as it will go.
5.
Mount the wafer holder ring @ onto the mounting pin @ of the wafer holder plate.
mm AllenE wrench.
Q
on the left-hand side of the stage
(Figs. 25,
26)
1
Fig.
28
I
Mounting the Mask Holder (for BH3-MJLT)
1.
Gently place the mask holder @ on the stage and check that it is firmly in position.
2.
Tighten the holder clamping screw @ at the left-hand side of the stage with the
1.5
mm AllenB wrench.
(Fig.
28)
]
Fig. 29
Fig. 30
-
(Fia. 29)
---
(Fig. 30)
@
---.
I
Mountina Other Holder Plates
Insert the holder into the stage and tighten the holder clamping screw
at the left-hand side of the stage with the
*
Do not tighten the screw with excessive force when mounting the stage plate glass so as not to break the glass.
MquntJn-g,
I
For Reflected Light DIC Observation
With the inscription 'NIC' facing the operator, insert the universal polarizer
@
slider
*
Engage the polarizer slider into the light path. (First click)
I
For Simple Polarizing Observation
With the inscription slider into the polarizer insertion slot of the vertical illuminator.
*
Since the polarizer slider has two click stops, push it all the way into the light path. (Second click)
the
universal
into the polarizer insertion slot of the vertical illuminator.
'PO'
Polarizer
facing the operator, insert the universal polarizer
1.5
mm Allen3 wrench.
Slider
'
1
1
Fig.
31
1.
Remove the dust slider from the vertical illuminator.
2.
With the inscription facing up, push the universal analyzer slider @ into the insertion slot until it stops. The analyzer slider can be engaged or disengaged by pushing or pulling until a click is heard.
*
To prevent dust from entering the vertical illuminator, be sure to insert the dust slider when the universal analyzer slider is not in use.
*
When performing transmitted light simple polarized light observation,
use the analyzer slider with
360° rotatable analyzer (BH3-UAN360).
L
Mounting.
ND
I.
i
I
Fig. 32
Filter Slider
Insert the universal tlnt plate slider (or the ND fllter sl~der) lnto the ND filterltnt plate Insertion slot on the vertical ~llum~nator. (See the
*
Engage the empty aperture of the tint plate slider into the light path.
*
with the inscription facing the operator, insert the ND filter slider so that the ND silter
;
M~~uMingrth~ Sglarizer IllUmipatJop+
The polarizer for transmitted light illumination should be mounted only when performing a transmitted light simple polarized light observation. Mount the polarizer positioning dots.
*
Use the 360" rotatable analyzer (BH3-UAN360).
the
Universal Tint Plate Slider
is
engaged into the light path.
fon
T~q~srnit-d, Ught
1
@
-.
onto the condenser mounting port by aligning the
or
-
-
--
-+
(Fig,
32)
ARROW.)
L
'(Fig.
-
@)
Fig. 33
Fig. 34
5
Mounting>%
.
. -.
1.
Loosen the two clamping screws on the top surface of the breath shield mounting bracket (Fig. screws. Tighten the screws securely.
Breath
34)
-.
Shield
-
-
and insert the breath shield underneath the
.
.
-
.-
(Fig.
344
Fig.
Fig.
35
36
After adjusting the aperture iris diaphragm and field iris diaphragm according to image requirements and inserting the necessary filter, place the
roof cover @ on the arm.
+
Depending on the observation tube in use, the cover may be difficult
a
to mount. If that is the case, mount it in
of
the
direction
+
The pinhole unit
DMPLIOOXPC.
With the inscription facing the operator, insert the pinhole unit into the
filter insertion slot of the universal vertical illuminator.
ARROW.
is
used with
100X
titled fashion in the
objectives such as
UMA
LW-
Fig.
Fig.
37
38
box
Plug the nosepiece remote control on the rear left-hand side of the microscope base.
Mount the stage grip below the projection at the rear of the stage and tighten the two screws with the
cable into the multi-pin socket
2.5
mrn Allena wrench provided.
@
Two
vibration
damping pads
(marked
with
'F')
I
Fig. 39
Fig.
40
I
Mo~*g.th.
Ir
Do not use the vibration damping pads for
Ir
Vibration damping pads are used to prevent external vibrations from being transmitted to the microscope stand. Use as necessary.
Place
the vibration damping pads at the bottom of the microscope base
after placing the microscope stand on the table as shown in the drawing.
:
Mo~nting!~the
-
---
1.
When mounting a brightfield objective to the motorized revolving nose-
(BH3-5NRE-M), first mount the objective adapter (NRE-MAD) in the
piece threaded hole of the revolving nosepiece, and into the adapter. (Fig.
2.
If one or more Nomarski prism attachments @ are mounted on the motor-
ized revolving nosepiece
focal distance when the objective adapter is in use, first mount the objec­tive adapter
41)
(Fig.
3.
If
the brightfieldidarkfield objective is to be used under the same condi-
tions, first mount the objective adapter
the brightfieldldarkfield objective. (Fig.
\?bWon
Oyective . Adapter
40)
(NRE-MSP2) @ and then mount the brightfield objective.
'?am~k!9-.P~d!_-
.-
(BH3-ENRE-M) and you want to obtain the same
.
.
insert the bright-field objective
(NRE-NESP2) 0 and then mount
41)
-
-
6"
stage (BH3-SIC6).
-
-
-
(Fip,
--
(Fig.
40,41!
391
]
1
;
Fig.
41
--T
Rqlaqi.ms_tofthe,
Since the focus unit and arm of your microscope are replaced only by the manufacturer, contact your nearest Olympus office (see back page).
Fop~Inlt:~mjd
the@rm1=
'
11
.--a
Fig.
Fig.
42
43
In accordance with the half-mirror unit indicator dial
Q
mirror unit selector knob
*
Slide the half-mirror unit selector knob until it stops.
1.
Confirm that any one of the objective magnification buttons is lit.
*
If any button is blinking, the corresponding objective is not engaged.
a
Press
*
Do not rotate the revolving nosepiece manually, as memory setting may change or a malfunction may result.
2. Press the button for the objective of choice to engage.
[
Alternative Objectives
Two objectives can be interchanged alternately.
button other than the blinking button.
to the desired position.
@,
slide the half-
I
1.
Select the first objective and press the corresponding objective magnification button.
2.
Pressing the second objective's button completes the setting.
3.
By pressing the recall button
I
Emergency Termination of Nosepiece Revolution
If an objective magnification button is pressed by mistake, immediately press the recall button. The revolving nosepiece will immediately return to the original position.
1.
Turn the voltage adjustment knob @ on the right-hand side of the micrsocope base towards the front ('low voltage' position) and turn on the power switch
2.
To increase the light intensity, slide the voltage adjustment knob backward ('high voltage' position).
*
@
on the rear panel.
To adjust the light intensity without affecting the color temperature during photomicrography, etc., use an For further details, refer to paragraph
@,
only the two objectives can be alternated.
(Fig.
45)
ND
filter provided.
4
,
"Filter Insertion".
1
Fig.
44
Switching from Vertical Illuminator to Transmitted Light Illuminator
(Fins.
44, 45)
Fig.
45
Press the illuminator selector switch
vertical illuminator and the transmitted light illuminator.
EPI: Reflected light DIA: Transmitted light
*.
When switching between the vertical and transmitted light il-
luminators during observation, adjust the transmitted light intensity
ND
with an
adjustment knob, in order to obtain equal brightness in reflected light and transmitted light.
*
To use on the vertical illuminator and the transmitted light illuminator
simultaneously, a separate power supply unit illuminator is needed.
I
Fuse Replacement
1.
Turn off the power switch and unplug the power cord from the AC outlet.
2.
Insert a Phillips screwdriver into the fuse holder @ at the rear of the microscope base, turn counterclockwise and remove the fuse holder.
3.
Remove the fuse from the fuse holder and replace with a new fuse.
Applicable fuse:
filter and frosted filter without touching the voltage
@I
(Fig.
44)
to switch between the
(TH-3) for the vertical
D
2
z
Fig.
46
.
.
,,.
,,
I
Fo..lli.a:
The specimen can be brought into focus by raising and lowering the stage
using the coarse focus adjustment knobs knobs
@I.
in the direction of the
approach each other).
The rotational tension of the coarse focus adjustment knobs is adjustable
by rotating the outer tension adjustment ring increases the tension, while turning it counterclockwise decreases the tension.
+
If loosened excessively, the stage may drop or defocus even after the specimen is brought into focus with the fine focus adjustment knob. adjustment ring clockwise.
,.
,
..
.-;:
..
__
When the fine and coarse focus adjustment knobs are turned
ARROW,
If
that is the case, increase the tension by rotating the tension
the stage
.
.-
.
.-:.
_
a
will
rise (the objective and specimen
6).
.
,
.
,
^
__--I--.-
and fine focus adjustment
Rotating the ring clockwise
..
,:,
:.
.
(Fig.
461
.
L.
:
Fig.
47
Fig.
48
10
Use
of
Coarse
The coarse focus stopper is provided to prevent the objective from hitting
the specimen and to speed focusing. Locking the coarse focus stopper lever
@
by turning it in the direction of the
focus adjustment knob's upper limit.
*
The fine focus adjustment knob's movement is not affected.
*
Even when replacing the specimen by lowering the stage once, the specimen can be brought into an approximate focus by raising the stage up to the upper limit specified by the coarse focus stopper.
Then, by making fine focus adjustment with the fine focus adjustment
knob, the specimen can quickly be brought into focus.
1.
Press the objective magnification button to engage the
2.
Looking through the eyepieces with both eyes, adjust the interpupillary distance by sliding the knurled dovetail slides @ of the right and left eyepiece sleeves until perfect binocular vision is obtained.
*
The knurled dovetail slides are provided with an interpupillary
distance scale
may wish to use the scale for rapid adjustment of your I.P.D.
Focus
@.
Stopper
ARROW
Once you know your interpupillary distance, you
will define the coarse
10X
(Fig.
objective.
rQ8)
Fig.
Fig.
49
50
1.
Looking through the ri' into focus using the coarse and fine focus adjustment knobs.
2.
Looking through the left eyepiece with the left eye, rotate the diopter adjustment ring
When a Field-of-view Eyepiece Trinocular Tube
1. Looking through the field-of-view eyepiece with the right eye, rotate the
knurled collar
511 can be sharply distinguished as two separate lines. Then bring
(Fig. the specimen into focus using the coarse and fine focus adjustment knobs.
2.
Looking through the left eyepiece with the left eye, rotate the diopter adjustment ring
@
ght eyepiece with the right eye, bring the specimen
@
to bring the specimen into focus.
(I35)WHKlOX) is used
.
(BH2-TR3-
on the eyepiece (Fig.
@
(Fig. 50) to bring the specimen into focus.
--
--
O/BHZ-TTR) (Figs. 50, 51)
50)
until the double reticle lines
1
Fig.
51
Fig.
52
When Using the Superwide Field Trinocular Tube
1.
Looking through the right field-of-view eyepiece with the right eye, rotate
@
the knurled collar
can be sharply distinguished as two separate lines.
2.
Adjust the coarse and fine focus adjustment knobs so that both the reticle
lines and the specimen are sharply in focus.
3.
Looking through the left eyepiece, rotate the knurled collar @ on the eyepiece to bring the specimen into focus.
on the eyepiece until the double reticle lines (Fig.
(BH2-SWTR)
(Fig.
52)
51)
Light
path
selector
knob
Indication
--
Light
path
Applica­tion
Fig.
53
I
!
IN
i
I
i
V
(white)
/
1.00%
binocular
1
tube
,
Obsewa-
/
tion of with focusing specimens
dark
specimens
Halfway OUT
1
C
-V 2%
to
binocular tube
80%
raphy
through
/
binocular
j
tub
(green)
to to
C
100% to
photo
of
'
(red)
dark
Tilting the Obs-ervation Tube (for
The observation tube can be tilted to an angle for viewing in the most comfortable position. Grasp the observation tube and raise or lower it to the desired position.
Salectingithe
1.
Slide the light path selector knob @ on the right-hand side of the trinocular tube to the desired position (Table
'
.
Light
Path
(for Trinoculgr Tube)
2d
BH2-lTRJlBl
.
-
- -
.
-..
:
,{Fig.
1)
until a click is heard.
,
--
-
only)
--
52,
(Fig.
Tabla
5311
--
.-I-
l)d
,
:
2
A
m
%
rn
n
z
0
z
Table
1
There are two positions to choose from:
Pushed in: For observation only (100% to binocular) Pulled out: 20% to binocular tube, 80% to photo tube
Fig.
54
Reduced image
9el@ Iris Diaphragm Centratlon fqr Vertical Illuminatol:
--
.
(Figs.
54,
1.
Press the objective magnification button to engage the
the specimen on the stage, and bring it into focus.
2.
Turn the field iris diaphragm lever @ of the vertical illuminator clockwise
stopdown the field iris diaphragm
to
3.
Rotate the right and left field iris diaphragm centering knobs @ of the
vertical illuminator counterclockwise to bring the reduced diaphragm image
into the center of the field of view.
4.
Rotate the field iris diaphragm lever counterclockwise in order to widen the opening of the field iris diaphragm until the polygonal image of the diaphragm is barely touching the rim of the circle indicating the field of view. If the image is off center, repeat the field iris diaphragm centering operation.
5.
Continue widening the opening of the diaphragm until the reduced image circumscribes the field of view.
to
a minimum.
10X
objective, place
-
66)
-
Fig.
Fig.
55
56
+permre
Since it is pre-aligned, aperture iris diaphragm centering adjustment is un­necessary.
1.
Engage the
2.
Remove one of the eyepieces from the eyepiece sleeve and look at the
objective pupil through the empty eyepiece sleeve.
3.
Turn the centering screws @ of the pinhole unit to bring the pinhole to the center of the objective pupil.
This step adjusts the beam of light on the specimen to achieve the optimum
image contrast.
iris
Diaphragm Centration forvertcal llluminatpr
100X
or higher power objective.
Fig.
57
(
Reflected Light Brightfield Observation (Fig.
Depending
of the unrversal vertrcal lllumlnator clockw~se to reduce the openlng unt~l the reduced Image c~rcumscr~bes the fteld of vlew and shuts out unnecessary
use.)
(
Reflected Light Darkfield Observation
When performing fully by turning the field iris diaphragm lever of the universal vertical illuminator counterclockwise.
on the
objective
Ilght.
the
In use, turn the f~eld Iris d~aphragm lever
(This
settlng varies wlth the
darkfield observation, open the field irls diaphragm
objective
58)
@
diaphragm
power In
(
1'
a
t
H.
8
'
F
-F
.&
B
This step adjusts the numerical aperture of the vertical illuminator which influences the resolution, contrast, and focal depth of the observed image.
* When observing under reflected light, a glare can occur due to an
effect of the condenser for transmitted light depending upon the
(Fig.
the
58)
specimen to be observed. Be sure to reduce the opening of
disphragm to its minimum by turning the aperture iris diaphragm adjustment ring fully clockwise.
I
Reflected Light Brightfield Observation
]
Fig.
Fig.
58
59
Remove one of the eyepieces from the eyepiece sleeve and turn the aperture iris diaphragm lever diaphragm opening, while looking in the empty eyepiece sleeve. Turning the aperture iris diaphragm lever clockwise reduces the opening of the diaphragm.
Generally, reducing the diaphragm opening to
objective N.A. will provide an image of favorable quality.
(
Reflected Light Darkfield Observation
In case of reflected light darkfield observation, open the aperture iris dia­phragm fully by turning the aperture iris diaphragm lever counterclockwise.
*
Depending upon the specimen to be observed, slighly reduce the
diaphragm opening to obtain an excellent
minimal glare. Unlike in brightfield observation, the objective exit pupil is not visible even after the eyepiece is removed. Adjust the diaphragm while observing the image until an optimum image with
minimal glare is obtained.
1
Transmitted Light Brightfield Observation
Turn the aperture iris diaphragm adjustment ring @ of the condenser to adjust the opening of the diaphragm. Rotating the aperture iris diaphragm
adjustment ring clockwise reduces the size of the opening.
@
on the vertical illuminator to adjust the
70
-
80%
darkfield image with
(Fig.
of the
59)
1
I
Fig.
60
Disengage the path.
1.
Slide the half-mirror unit selector knob to bring the brightfield half-mirror unit into the light path.
2.
Press the objective magnification button to engage the
3.
Place a highly reflective specimen such as a mirror on the stage, and bring it into rough focus.
4.
Remove the right or left eyepiece, and adjust the bulb centering knob and socket clamping knob @ so that the filament image visible through the observation tube is located in the center of the objective pupil.
Rightneft: Turn the bulb centering knob Upldown: Loosen the socket clamping knob @ and rotate in the
5. Engage the universal analyzer slider, polarizer slider, and filters.
analyzer slider, polarizer slider and filters from the light
10X
objective.
@.
direction of the ARROW. After adjusting, tighten the knob.
@
Fig. 61
Fig. 62
I
Reflected Light Observation (Fig. 61)
Generally, to change the color temperature of halogen light to daylight, open the filter cover 20LBD3-W
20LBD3-W filter: Changes the color temperature of the illumination 201F550-W filter:
20ND6-W filter:
20ND25-W filter: Attenuates the light intensity.
*
If the filter is difficult to insert or remove, spread the sides of the
filter cover
*
When the filter is not in use, replace the filter cover to prevent dust from entering.
I
Transmitted Light Observation
Mount the filter @ in the filter holder @ provided with the lamp housing
for the transmitted light illuminator and then insert the filter holder in the
filter insertion slot tion.
45LBD3 filter: Changes the color temperature of the illumination 451F550 filter: Adds contrast to the observed image.
45ND6 filter: Attenuates the light intensity. 45ND25 filter: Attenuates the light intensity.
filter according to the purpose of observation.
@
of the universal vertical illuminator and insert the
to daylight. Adds contrast to the observed image. Attenuates the light intensity.
(light transmission: (light transmission: 25%)
@I
and remove.
@.
Choose the filter according to the purpose of observa-
to daylight.
(light transmission: 6%) (light transmission: 25%)
6%)
(Fig.
62)
1
(
Fig. 63
1.
Turn the clutch knob @ on the bottom right of the stage to the
64)
(Fig.
2.
The stage can be moved vertically (Y-axis direction) by rotating the upper
0,
section of the stage control knob rotating the lower section of the knob.
*
Clutch knob should be set so that the mark faces the operator.
and horizontally (X-axis direction) by
'A'
position.
(
Use of Stage Clutch
(
Clutch knob (A)
I
(Figs.
64,
65)
1
@
@
Fig.
Fig.
64
65
0
0
When the clutch knob is set to and Y axes, and fine travel of stage can be conducted by following the previous procedure.
*
If the clutch abrupt movement may cause the stage to slip.
I
Clutch knob
When the clutch knob is set to
X and Y axes, and stage travel is not possible with the stage control knob. (The stage will not move when the stage control knob is turned.) Since the stage travels lightly and smoothly, distance by hand. of the stage or the optional stage grip
[
Clutch knob (C)
When the clutch knob is set to
and disengaged stage control knob. Since there is no X-axis travel, the specimen needing repeated patterns can be observed conveniently by fixing the X-axis and moving along the Y-axis.
[
Clutch knob
When the clutch knob is set to and disengaged for the X-axis. This allows stage travel reversed to the
setting of the clutch knob.
*
Never move the stage manually with the clutch engaged, as this may wear out or damage
is
engaged, turn the stage control knob slowly since
(8)
I
It
is convenient to grip the travel pins @ at the front
I
for
the Y-axis. This enables X-axis travel only with the
(Dl
I
'A",
the clutch is engaged for both the X
'B',
the clutch is disengaged for both the
it
can be moved over a large
@
when moving the stage.
'C',
the clutch is engaged for the X-axis
'D',
the clutch is engaged for the Y-axis
the
stage.
'C'
09
rn
n
5
0
Fig.
66
I
Tilt Adjustment of Holders and Plates
When placing a holder (BH3-WHPR86, etc.) or plate (BH3-SPG8, etc.) on
the stage, parallel adjustments of the stage and the top surfaces of the holder
or
plate are not necessary since the assembly has been preset.
However, to make fine adjustments, apply a drop of alcohol to the screws
@,
unlock the screws and adjust them vertically using a slotted screw-
driver from the underside of the stage.
*
The stage can support up to 2 kgf
(4.4
Ib) including holders.
(Fig.
66)
1
The
BH3-MJLTJMJLA
simple polarizing, and fluorescence observation.
*
For reflected light fluorescence observation, refer to the instruction manual for the reflected light fluorescence attachment
5-1
Reflected Light Brightfield/DarMield Observation
(
Preparation for Observation
allows a quick change between the brightfieldidarkfield. Nomarski differential interference contrast,
(BH2-URFL).
Fig.
67
~a~f-mirror
Confirm that the brightfield half-mirror unit and darkfield half-mirror unit are inserted into the half-mirror unit compartment and slide the half-mirror
unit selector knob path.
Engage the ND filter minimizes changes in light intensity (glare) when changing the observation method from darkf ield to brightfield.
If light intensity is insufficient during brightfield observation, or a reduced exposure time is required for photomicrography, engage the empty aper­ture of the
UniKSeMng
@
to engage the desired half-mirror unit into the light
@
of the ND filter slider into the light path. This
NO filter slider.
-
,
,
I
.
-
(~ig.
67)!
68
Fig.
1
Observation Procedures
Engage the desired half-mirror unit with the half-mirror unit selector knob, according to the purpose of observation. p. Disengage the universal analyzer slider, polarirer slider, tint plate slider and Nomarski prism attachment.
Press the illuminator selector switch to *EPIu and switch
1 4 1
1
Place the specimen to be observed on the stage.
Engage the
(
1 6 1 1
Adjust the aperture and field iris diaphragms.
-1
.
Check that the bulb is centered correctly.
lnsen the desired filter into the universal vertical illuminator.
10X
objective and focus.
-
.
ON
the miscroscope main switch, turning
ON
the halogen
-.
1 9 1
I
Engage the desired objective and focus again.
-1
-,
111(1
--
I
Adjust the light intensity with the voltage adjustment knob. Turn the aperture iris diaphragm adjustment ring fully clockwise to reduce the opening of the diaphragm.
Brightfield observation
Darkfield observation
Adjust the field and aperture iris diaphragms. Open the field and aperture iris diaphragm. pp.
pp.
p.
P.
p.
P
p.
29, 29,
p.
P. p.
p.
p.
33
34
26
30
27
27
30
31
26 25
30
30 30
1
1
1
I
1
,
1
I
:
&'
5-2
Reflected Light Nomarski Differential Interference Contrast Observation
(
preparation for Observation
1
1
Fig.
69
Half-mirror
Slide the half-mirror unit selector knob to the brightfield half-mirror unit into the light path.
Confirm that the universal analyzer slider correctly pushed all the way into the universal vertical illuminator.
*
Make sure that the "NIC" inscription of the universal polarizer is
facing the operator.
*
When using the BH3-UAN360 analyzer, position the polarizer with
its
"NIC inscription facing the operator and turn the analyzer rotation
@
dial
Unit
Setting
'B.F.'
position to engage the
@
and polarizer slider
so that the dot is also facing the operator.
are
interference
color Black
Magenta
Fig. 70
71
Fig.
Observation features
,
Darkfield-like observation The highest gray sensitivity
observation
I
Even minute optical differences (refraction, thickness, height, etc.) can be observed as a color change.
@
Move the prism control lever 'IN' position to bring the prism into the light path.
Turn the polarizer rotation dial @ of the universal polarizer slider until the desired background color is obtained for maximum contrast of the specimen
under observation.
*
To obtain a magneta background, put the universal tint plate slider
into the ND into light path.
*
The interference color can be changed continuously from black to
(0 - 700 nm)
blue
1.
Move the prism control knob to the
from the light path.
2.
Disengage the universal analyzer slider, polarizer slider and tint plate slider.
filterhint plate insertion slot and push it all the way
of the Nomarski prism attachment to the
'OUT
position to disengage the prism
Table
2
(
Observation Procedures
-.
Confirm that the half-mirror unit selector knob is at the
'B.F."
position.
-
Engage the universal analyzer slider, polarizer slider, and Nomarski prism attachment.
(
Press the illuminator selector switch to "EPI', and switch ON the microscope main switch, turning ON the halogen
I
bulb.
(Place the specimen to be observed on the stage.
-
15
1
Engage the
-
-
Perform diopter and interpu~illary distance adjustments.
(6
Check that the bulb is centered correctly.
-,
)8)
1
Insert the desired filter into the universal vertical illuminator.
-
191)
Engage the desired objective and focus again.
I
Adjust the light intensity with the voltage adjustment knob.
111
11
Adiust the aperture and field iris diaphragms.
---
Adjust interference color and contrast
[plate
53
Reflected
slider.
10X
objective and focus.
-
-
Light
Simple Observation in Polarized
by
turning the polarizer rotation dial and inserting or removing the tint
Light
PP.
341
p.
34
p.
p.
26
30
p.
27
p. p.
27 30
p.
31
p.
26
p.
25
p.
29,30
p.
34
1
1
1
1
I
1
1
,
1
1
I
I
Preparation for Observation
Fig.
72
Ha!f:mitro;r.
Sl~de the half-mirror unlt selector knob to the br~ghtf~eld half-m~rror unit Into the l~ght path.
8
,
Cl!e~king-@g Analyzer aa,dcPolarher
Confirm that the universal analyzer slider @ and polarizer slider @ are
correctly pushed all the way into the universal vertical illuminator.
*
Make sure that the "PO" inscription of the universal polarizer is facing the operator.
*
When the slider so that its "PO" inscription is facing the operator and turn the analyzer rotation dial so that the dot is also facing the operator (see Fig.
Engeging,
.~mchmed:
L
-
When the Nomarski prism attachment is mounted, swing the prism control
lever
@
of the Nomarski prism attachment to the 'OUT position to disengage
the prism from the light path.
unit.
set tie^
BH3-UAN360
70
on page
34).
a@$r[)Isengsgin@
-.-
--
-
.
.
--
-
--
-
-
-
-
-
- - -
analyzer is
the
---,
.
-
--
-
- - - - - - -
'B.F.'
posit~on to engage the
,
-
.
-
.
in
use, position the polarizer
NqmamllE~~Prism
I---
-.
(n&
.
IEal72)
---
-
-l
:I
,
-I
$$!;
Fig.
73
Setting
Turn the polarizer rotation dial @ of the universal polarizer slider so that
the dot faces to the side in order to obtain a crossed filter position.
Fig.
74
I
Observation Procedures
I
Confirm that the half-mirror unit selector knob is at the Engage the universal analyzer slider, polarirer slider, and Nomerski prism attachment.
Disengage the Nomarski prism attachment.
Press the illuminator selector switch to "€PIe, and switch ON the microscope main switch, turning ON the halogen bulb.
Place the specimen to be observed on the stage.
)
Engage the
10X
objective and focus.
thecrossed
'B.F.'
position.
Filter
Position
_
_
-_
-
-
(Fig,
73
p.
34
p.
34
1
p.
34
p.
26
p.
30
p.
27
Perform diopter and interpupillary distance adjustments.
Check that the bulb is centered correctly.
Insert the desired filter into the universal illuminator.
110
1
I
Turn the polarizer rotation dial until a crossed filter position is obtained. Engage the desired objective and focus.
Adjust the light intensity with the voltage adjustment knob. Adjust the aperture and field iris diaphragms. pp.
5-4
Transmitted Light Observation
Preparation for Observation
Select transmitted light by pressing the illuminator selector switch @ to
'DIA' position.
the
29,
p. p.
p.
P.
p.
p.
27 30 31
36 26
25 30
1
Fig.
75
If
a
half-mirror unit is mounted, slide the half-mirror unit selector knob to
the
'B.F."
position to engage the brightfield half-mirror unit into the light
path.
*
If no half-mirror unit is mounted, observation is possible without
moving the half-mirror unit selector knob.
I
Observation Procedures
Slide the brightfield half-mirror unit into the light path (or confirm that no half-mirror unit is inserted into the half-mirror unit compartment) p.
El
Press the illuminator selector switch to 'DIA", and switch
transmitted light halogen bulb.
El
Disengage the universal analyzer slider end Nomaraki prism attachment.
1
-
Place the specimen
1
Engage the Perform diopter and interpupillary distance adjustments.
Insert the desired filter into the filter insertion slot on the transmitted light lamp housing. Engage the desired objective and focus again.
to
be observed on the stage.
10X
objective and focus.
ON
the microscope main switch, turning
'
ON
the
-
Adjust the light intensity with the voltage adjustment knob.
~djun the aperture iris diaphragm.
1
5-5
Transmitted
Light
Simple
Observation in
-
--
Polarized
Light
p.
p. p.
p.
p. p.
p. p. p.
34
26
34
1
33
27
27
31
26
1
25
30
I
Preparation for Observation
Fig.
76
Selecting
.
-
.
.
- + -
Select transmitted light by pressing the illuminator selector switch @ to the
'DIN position.
If the brightfield half-mirror unit is mounted, slide the half-mirror unit selector knob to the light path.
.k
If no half-mirror unit is mounted, observation moving the half-mirror unit selector knob.
Ir
Use the BH3-UAN360 analyzer and the BH3-POL polarizer.
1.
Insert the polarizer into the condenser frame so that their positioning dots
@
are aligned with each other.
2.
Turn the analyzer rotation dial until a crossed filter position is obtained with the dot facing the operator (see Fig.
the
-
-
lllumina~n-Method-
'B.F.'
position to engage the brightfield half-mirror unit into the
70
on page
.
-
---
is
possible without
34).
(Fig.
1
;!I61
Fig.
77
L
I
Observation Procedures
Slide the brightfield half-mirror unit into the light path (or confirm that no half-mirror unit is mounted in the
half-mirror unit compartment) p. 34 Press the illuminator selector switch to
transmitted light halogen bulb. Engage the universal analyzer (BH3-UAN360) and the polarizer for transmitted light. p. 34 Disengage the Nomerski prism attachment.
1
Turn the analyzer rotation dial until a crossed filter position is attained.
1 6 1
1
Place the specimen to be observed on the staae.
I
Engage the
1 8 1
I
Perform diopter and inter~upillaw distance adiustment.
1OX
objective and focus.
'DIA', and switch On the microscope main switch, turning ON the
--
p.
26
p.
35
p. 36
P.
30
p.
27
P.
27
1
1
1
(
Insert the desired filter into the filter insertion slot on the transmitted light lamp housing.
110
(1
Engage the desired objective and focus again.
I
Adjust the aperture iris diaphragm.
p. 31
P.
26
p.
30
I
Fig.
78
With respect to handling photographic equipment, also refer to the instruc­tion manual for the camera in use.
Use
NFK
photo eyepieces specially designed for photomicrography. Insert the photo eyepiece tube.
@
into the photo tube of the trinocular observation
Light
knob
Indication
Light
path
Applio­tion
Fig.
Pulled out halfway
I
,
C
.V
(green)
20%
into
binocular tube
80%
into photo tube
photo tube
/
(focusing through
,
the binocular tube)
I
79
Pulled out fully
/
C
(red)
100%
into
Wunting.
Place the photomicrographic equipment on the photo tube of the trinocular tube. Align the dots on the trinocular tube and the photomicrographic equipment and clamp the photomicrographic equipment with the clamping screw
With the BH2-TR30 Trinocular Tube or BH2-SWTR3O Superwide Field Trinocular Tube
Pull out the light path selector knob @ on the right-hand side of the trinocular tube to the
the-Photqm_!c_rpgmphicCEqu1~mment
O.
'C.V
or
"C'
position.
(Fig.
79,
-
(Fig.
Table
7%)
3)
Table 3
Fig.
80
[
With the BH2-TTR Tilting Trinocular Tube (Fig. 80)
Pull out the light path selector knob @ on the right-hand side of the tilting trinocular tube to the
20%
light is directed to the binocular tube and
'C'
position in order to direct light to the photo tube.
80%
to the photo tube.
1
?
7
C
B
i
j;
i
Fig.
81
Focusing
Use the binocular tube of the trinocular tube for focusing on the film plane.
1.
If the viewer of the photomicrographic equipment is not to be used, remove the viewer and attach the dust cap.
1.
Confirm that the field-of-view eyepiece is inserted into the right eyepiece sleeve.
2.
Each field-of-view eyepiece has a focusing front lens and a reticle with four frames, with each frame indicating the area covered by the specific photo eyepiece: The image in the focusing front lens and the image on the film plane are in focus at the same time.
*
Several types of field-of-view eyepieces are available according to the camera back in use. standard field-of-view eyepieces.
Field-of-view
)35)
WHKlOX (or
-
135)
SWHKlOX) are
(Figl
81)
Camera
use
in
35
camera
mm
back
Sheet film or Intermediate
3-114' x 4-114'
Polaroidg back
adapter
4' x 5'
for
film
WHKlOX
(Field number
20)
Type MSPlan
Magnification
N.A. 0.13 0.30 0.46 0.80 0.95
W.D.
Focal length Imm)
Resolving power (p)
--
(mm)
5X 1 OX 20X 50X
21.10 9.00 3.00 0.47 0.27
35.9 18.0 9.0 3.6 1.8
1.12 0.73
1
Fox
Total Focal
rnagnificati0+$~ depth Ip)
Field of view (mm)
Trr,
I
I
1
MDPlan 1 NeoDPlan
0.95
1
0.90
0.42
1.3
0.4 0.2 0.13 0.13
1 000X 0.67 1500X 0.55 l500X 0.60
1.2
Type
Magnification
N.A.
W.D.
Focal length (mm)
Resolving power
WHKl
OX
(Field Focal depth number
WHKI
20)
OX
(Field Focal depth number
WHKlOX
(Field number
20)
20)
Total magnification
Field
Focal
Resolving power (p)
Total magnification Field of view (mm)
Focal length (mm)
Resolving power
Total magnification Field of view (mm)
(mm)
of
view (mml
Type ULWDMSPlanlULWDNeoSPlan
Magnification
N.A.
W.D.
(mm)
length (mm)
Type
Magnification
N.A.
W.D.
(mm)
Focal depth
(p)
(p)
(p)
(p)
(y)
.
NeoSPlanlNeoSPlanNlC
5X
0.13J0.13 0.30J0.25 0.46/0.40 0.80/0.70 0.90/0.90
10.00 3.00 0.50 0.30
35.9 3.6 1.8
2.58f2.58 0.4210.48
---
50X 200X 500X 1 000X
70.1 5.1 1.3 0.73 4
20X
0.4
11
.OO
9.0 3.6 2.25 1.80 3.60 0.72
0.84 0.61 0.45 0.42 0.35 0.37
200X
6.09
1
LWDNeoSPlanApo LWDMSPLZOX-LCD LWDMSPLSOX-LCD LWDMSPL100X-LCD
250X 20X 50X 1 OOX
0.90 0.40 0.60 0.80
0.80 7.87 3.02
0.72 9.0
0.37 0.84 0.56 0.42
2500X 200X 500X 1 OOOX
0.50 6.09 1.93 0.87
0.08 1 0.4
1
OX 20X 50X 1 OOX
0.3710.37
,
1
50X 8OX 1 OOX 5OX
0.55 0.75 0.80 0.90
8.00 4.10 3.20
0.4 0.2
LWDMSPlanApo
1
250X
:
0.80
I
500X 800X
2.18 1.07
0.4 0.25
1 OOOX
0.87
0.2
500X
1.04 0.50
0.4 0.08
3.60
I
2500X
1.04
1
.80
0.2
Note: All the objectives are dry type. Resolving power is calculated with the aperture iris diaphragm fully opened.
Technical terms:
W.D.
(Working Distance): The distance between the specimen and the nearest point of the objective.
N.A.
(Numerical Aperture]: The numerical aperture represents a performance number which can be compared to the
Resolving power:
Focal depth:
Field number:
of
Field
view diameter
relative aperture (f-number) of a camera lens. the resolving powers of all types of objectives. The larger the power.
The ability of an objective to register small details. The resolving power of an objective is
by
measured The distance between the upper and lower limits
optical system. The smaller the opening of the aperture diaphragm or the objective larger the focal depth.
This number represents the diameter in millimeters of the image of the field diaphragm that is formed by the lens in front of it.
The actual size of the field of view in millimeters.
its ability to separate two points (resolution).
N.A.
values can be used for directly comparing
N.A.,
the higher the resolving
of
sharpness at an image formed by
N.A.,
an
the
LW
If you are unfamiliar with any aspect of this model or feel that performance is less than 100%. check the Items on the follow~ng table.
.,
1.
Optical
-
.
System
Ttgublb
.
.
'
,;
.
,
.
-
!&Ius~~
.. .
,.
...
.
.
.
.,
.
...
,
.
.
.
.
'-..:~tj&&$'<
.
.
. . .
,
'
i
_/
1'
:
.
,
a. The bulb is burning, but image 1 The field iris diaphram is not opened Open until the diaphra m image cir-
cannot be seen or is dark.
cumscribes the field
o? view. (p. 291
The aperture iris dia hragm is stopped Open fully. (p. 30) down with the dark!eld half-mirror unit engaged (for reflected light
brightfield/
The light path selector knob of the trinocular tube is between settings.
'
Pull it out to
'C.V' or
'V'
position.
(p. 28)
Set it to match the local line voltage.
incorrectly.
I
The half-mirror unit is not correctly inserted into light path.
Enga e the half-mirror unit properly
with
8e halfmirror unit selector knob.
(p. 20)
(p. 25)
b. Image is partially obscured or
correctly.
Center the bulb.
(p. 30)
evenly illuminated.
Insert the universal analyzer, polarizer and tint plate slider as far as it will
90.
(PP.
21. 22)
Pull it out to 'C.V or 'V position.
(p. 28)
Engage the half-mirror unit properly
with the half-mirror unit selector knob.
(p.
25)
Clean.
c.
Stains ordust can be seen in the field of
wew.
The universal analyzer, polarizer or tint plate slider is not inserted far enough.
The light path selector knob of the trinocular tube is between settings.
The half-mirror unit is not correctly inserted into the light path.
Dirty objectives Dirty half-mirror units
,
Insufficient contrast Image details lack definition
f.
One side of the image is out of focus.
Dirty specimen
I-
Dirty condenser top lens
.-
Dirty objective front lens Dirty eyepieces Dirty prism on the bottom of the ob-
servation tube Dirty stage plate glass
is stopped
--
correctly.
Dirty condenser
j
lens Designated
and/or objective front
IC (infinity corrected) ob-
jectives and eyepieces are not used. The specimen is not positioned cor-
rectly.
The objective is not engaged correctly.
,
Open the aperture iris diaphragm as required.
(P.
Click the nosepiece into position.
(p. 30)
Clean.
Use designated
IC objectives and eye-
pieces. Reposition the specimen on the stage.
Click the nosepiece into position
30)
.~ .
:
.
.%&ble
-
,
:,
..
,,
.
'
-
,
&use.
-:
Normaski Differential Interference Contrast Observation
g.
No interference color
.
.
~-
. .
,
.
.
..
-
'
R~*.Y
Place the universal analyzer and polarizer correctly.
----
(p.
21)
-
-
--
--
h.
Interference color is uneven
--
ferential
low. used.
2. Observation Tube
interpupillary distance
Incomplete binocular vision
The adjusted correctly.
pp
Diopter adjustment
3.
Stage
The stage is not clamped
The image goes out of focus cessively when the stage is touched.
4.
Coarse and Fine Focus Adjustment Knobs
a. The coarse focus adjustment knob
ex-
--
Stage clutch is disengaged. Engage the clutch. (p. 32)
The tension adjustment ring is tighten-
movement is stiff. ed excessively.
--
b. Poor focus during observation due
--
'
to unexpected lowering of the
is locked at the upper limit. The tension adjustment ring is too
loose.
stage.
c. The stage cannot be raised com-
pletely with the coarse focus
The coarse focus stopper is locked at the lower limit.
adjustment knob.
--
-.
is not
-
.
IS
incomplete.
-
7
-.
-
focus adjustment stopper
.-
-.
--..
--
I
Set correctly interpupillaw distance.
1
!
Complete diopter adjustment.(~. 27)
(p. 27)
(p.
(p. 14)
-
,
Loosen appropriately. (p. 26)
Release the coarse focus stopper.
I
Tighten appropriately.
(p. 27) (p. 26)
Release the coarse focus stopper.
(p.
181
27)
5.
Electrical System
a. The bulb flickers.
b.
Although an objective LED lights up, the bulb does not come on.
c. Light intensity at maximum voltage
adjustment range is insufficient.
d. Bulb does not burn.
"'Itage
and
Bulb
Fuse
Consum~tion
Power
The bulb is approaching the end of its life.
Loose electric connections.
The bulb is burned out.
Loose electric connections.
The line voltage selector switch is not
set to your local line voltage. Fuse has blown.
1
5V
AC,
10011
20012201240V
For reflected light: For transmitted light:
AC
12V. 100W halogen bulb JC12V1 OOWHAL-L
12V.
50W halogen bulb JC12V5OWHAL-L
12V, 100W halogen bulb JC12VSOWHAL-L
4A
150VA (max.)
-.
-
50160
Replace the bulb.
-
Connect cords and plugs correctly.
Replace the bulb. (p. 17) Connect cords and plugs correctly. Switch to your local line voltage.
Replace the fuse. (p.
Hz
(p. 17)
(pp. 19, 20)
(p.
13)
26)
-
OLYMPUS
OLYMPUS OPTICAL CO., LTD.
San-Ei Building,
22-2.
Nishi Shinjuku 1-chorne, Shinjuku-ku, Tokyo, Japan
OLYMPUS OPTICAL CO., (EUROPA)
Wendenstrasse 14-1
6,
Postfach 104908,2000 Hamburg
1,
OLYMPUS CORPORATION
4
Nevada Drive, Lake Success,
OLYMPUS OPTICAL CO.
2-8 Honduras Street,
London
N.Y.
11042-1179,
EClYOTX,
(U.K.)
United
U.S.A.
Kingdom
GMBH.
Germany
LTD.
The design of the product is under constant review
.
,
..
.
- - -
4---
-L-
-:-k*
:-
LOM-,~~
and
whilst every effort is
t~
+ha-
~nwifir~.
Loading...