OKmeter Cameo OK-P User Manual

SIDE
Read instructions before use.
OK-P
10-62-2965-0003
V1-FEB11
OKP010001
Before You Begin
PLEASE READ THIS BEFORE USING.
The following basic safety precautions should always be taken.
1. Close supervision is necessary when the device is used by, on, or near children, handicapped persons or invalids.
2. Use the device only for the intended use described in this manual.
3. Do not use test strips and control solutions which are not supplied by the manufacturer.
4. Do not use the device if it is not working properly, or if it has suffered any damage.
5. Before using any product to test your blood glucose, read all instructions thoroughly and practice the test. Do all quality control checks as directed and consult with a diabetes healthcare professional.
KEEP THIS USER GUIDE WITH YOU.
Intended Use
The system is intended for use outside the body (in vitro diagnostic use only). It
should be used only for self-testing blood glucose (blood sugar) and only with fresh capillary whole blood samples. The system is intended for use in the home and in clinical settings. The system should not be used for the diagnosis of diabetes or for the testing of newborns.
Principle of Measurement
Blood glucose is measured by an electrical current that is produced when a blood samples mixes with the reagent (special chemicals) of the test strip. The electrical
Getting To Know Your System
The OKmeter Cameo Blood Glucose Monitoring System.
The OKmeter Cameo system uses the latest technology to provide you with easy and
comfortable testing. The system requires only a 0.7 μL of blood sample to complete
the testing in only 6 seconds.
Important :
Use only OKmeter Cameo test strips and control solutions with your OKmeter Cameo meter. Use other test strips and control solutions with this meter can produce inaccurate results.
current changes with the amount of glucose in the blood sample. The OKmeter Cameo meter measures the strength of the electrical current, calculates your blood glucose
level and then displays your result in either milligrams of glucose per deciliter (mg/dL) or millimoles of glucose per liter (mmol/L).
Caution
1. The user should not take any decision of medical relevance without rst consulting
his or her medical practitioner.
2. Call your doctor immediately if you experience symptoms that are not consistent with your blood glucose test results.
3. High altitudes above than 3,402 meter (11,161 ft) may affect the test results.
4. Temperatures outside the range of 10°C to 40°C (50°F to 104°F) may affect the test results. Do not test beyond of temperature range.
IMPORTANT HEALTH-RELATED INFORMATION
1. Apply only capillary whole blood sample to test your blood glucose. Applying other substances or plasma, serum will cause wrong results.
2. Severe dehydration and excessive water loss may cause false low results. If you believe you are suffering from severe dehydration, consult your healthcare professional immediately.
3. Test results below 60 mg/dL (3.3 mmol/L)*1 indicates low blood glucose (hypoglycemia). Test results greater than 240 mg/dL (13.3 mmol/L)*2 indicates high blood glucose (hyperglycemia).If your results are below 60 mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), repeat the test, and if the results are still below 60 mg/dL (3.3 mmol/L) or above 240 mg/dL (13.3 mmol/L), consult your healthcare
professional immediately.
4. Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate low results may occur for individuals experiencing a hyperglycemic-
hyperosmolar state, with or without ketosis. Critically ill patients should not be tested with blood glucose meters.
5. Abnormal red blood cell counts (hematocrit level below 20% or above 60%) may cause false results. Please consult your healthcare professional if you do not know your hematocrit level.
6. Interference: Reducing substances occurring in the blood naturally (uric acid, bilirubin) or from therapeutic treatments (ascorbic acid, acetaminophen) will not
signicantly affect OKmeter Cameo test results. However, elevated concentrations
of these substances may affect test results. The compounds listed in the table were found to have no affect at the concentration indicated.
Compounds Highest concentrations
tested at which no interference occured
Compounds Highest concentrations tested
at which no interference
occured Acetaminophen≦ 8.0 mg/dL (0.53 mmol/L) Gentisic Acid≦ 5.0 mg/dL (0.32 mmol/L) Ascorbic Acid ≦ 5.0 mg/dL (0.28 mmol/L) Hydroxyurea≦ 3.0 mg/dL (0.39 mmol/L) Aspirin
60 mg/dL (3.33 mmol/L) L-dopa≦ 10 mg/dL (0.51 mmol/L)
Bilirubin
90 mg/dL (1.54 mmol/L) Maltose≦ 900 mg/dL (26.3 mmol/L) Cholesterol≦ 500 mg/dL (12.9 mmol/L) Methyldopa≦ 3.0 mg/dL (0.13 mmol/L) Creatinine≦ 5.0 mg/dL (0.44 mmol/L) Tolbutamide≦ 400 mg/dL (14.8 mmol/L) Dopamine≦ 2.0 mg/dL (0.11 mmol/L) Triglycerides≦ 2,000 mg/dL (22.6 mmol/L) Galactose≦ 900 mg/dL (50 mmol/L) Uric Acid≦ 8.0 mg/dL (0.48 mmol/L)
REFERENCE:
* 1: Kahn, R. and Weir, G.: Joslinis Diabetes Mellitus, 13thed Philadelphia : Lea and Febiger (1994), 489.
* 2: Krall, L.P. and Beaser, R. S.: Joslin Diabetes Manual. Philadelphia : Lea and Febiger(1989), 261-263.
1. DATE:
Display month and date.
2. TIME:
The meter is programmed for a 12
hour period format.
3. AVERAGE:
Appears when the meter is in the memory mode while recalling 7/14/28 day test averages.
4. MEMORY SYMBOL:
Appears when in
the memory mode.
5. THERMOMETER SYMBOL:
Appears when ambient temperature is above or below the acceptable range needed for testing.
6. CONTROL SOLUTION TEST SYMBOL:
Appears when you are in Control Solution
Mode. Your test result will not be stored in meter memory.
7. UNIT OF MEASURE:
Appears with the test result either in mg/dL or mmol/L.
8. BATTERY SYMBOL:
Appears when battery is weak.
9. BLOOD DROP SYMBOL:
Flashes when sample should be applied.
10. STRIP SYMBOL
:
Appears when the meter is turned on and waiting for inserting a
test strip.
SETTING TIME AND DATE
Please install battery rst and set correct time and date before you begin to test.
STEP 1. Enter Setting Mode
Start with the meter off. Open the battery cover. Press SET BUTTON located at the right side of battery compartment. The meter will turn on and enter the setting mode.
STEP 2. Set the Year
The year appears with the number ashing. Press and release the M BUTTON to obtain the desired year. Press SET BUTTON to conrm
and move to next MONTH setting.
STEP 3. Set the Month
The month appears with the number flashing. Press and release the M BUTTON to obtain the desired month. Press SET BUTTON to
conrm and move to next DATE setting.
STEP 4. Set the Date
The date appears with the number ashing. Press and release the M BUTTON to obtain the desired date. Press SET BUTTON to conrm
and move to next HOUR setting.
STEP 5. Set the Hour
The hour appears with the number ashing. Press and release the M BUTTON to obtain the desired hour. Press SET BUTTON to conrm
and move to next minutes setting.
STEP 6. Set the Minute
The minute appears with the number flashing. Press and release the M BUTTON to obtain the desired minute. Press SET BUTTON to
conrm and the meter will enter to next unit setting.
STEP 7. Set the Unit of Measurement
The existing unit (for example, mg/dL) will appear and ashing. If you
do not want to change the unit, Press SET BUTTON to skip this step. If you want to change the unit, press and hold the M BUTTON for 3
seconds to change the unit to mmol/L. Press SET BUTTON to conrm
and the meter will enter to next delete all memory setting.
STEP 8. Delete Memory
When the dEL symbol and the flashing memory symbol appear on
the display, you can choose to clear the memory. If you do not want to clear the memory, press the SET BUTTON again to skip this step.
If you want to clear ALL memory, press and hold M BUTTON for 3 seconds. The “---“ image will appear on the LCD screen to indicate
that all memory has been deleted.
STEP 9. Complete Setting
After deleting memory, the meter will display “OFF” before shut down. The meter setting is now completed.
IMPORTANT :
1.The time, date and unit of measurement can ONLY be changed in the setting
mode. Therefore, when you perform a blood glucose testing, it is not possible to change those parameters.
2.Your meter displays 7, 14 and 28-day averages which you can access from the meter memory. These averages are calculated from your latest result obtained during the last 7, 14 and 28 days.
3.Your meter can display test result in milligram per deciliter (mg/dL) or millimoles per liter (mmol/L). Your meter has been preset at the factory with the standard unit measurement used where you live (mg/dL) . When you turn off the meter or
replace the battery, the unit will not be changed.
4.Use of the wrong unit of measure may cause you to misinterpret your blood glucose level, and may lead to incorrect treatment. Please always consult with your healthcare professionals before you reset the unit of measure.
1.
A
1
. Getting To Know Your System
2. Prepare For Blood Sampling
3. Performing Blood Test
4. Memory Recall
5. Control Solution Testing
6. Caring For Your Meter And Test Strip
7. System Specifications
8. About Alternative Site Testing (AST)
9. Display Messages And Problem-Solving Guide
SIDE
A :
SIDE
B :
Meter
Test Strip
ABSORBENT CHANNEL:
Apply a drop of blood here.
CONTACT BARS:
Insert this end into strip slot.
TEST STRIP HANDLE :
Hold this part to insert the test strip into the meter slot.
CONFIRMATION WINDOW:
This is where we conrm if enough blood is
applied for has been drawn into strip.
Prepare For Blood Sampling
Adjustable Lancing Device
Your lancing device and lancets are used for obtaining
blood samples from the puncture site.
Lancet
Protective cap
2
.
Important Lancing Device and Lancets Information
1. Lancet is for single use only.
2. Keep lancing device and lancets clean.
3. Use caution when removing the used lancet from the device and when disposing the used lancet.
IMPORTANT :
The meter and lancing device are for single patient use. Do NOT share them with anyone including other family members ! Do NOT use on multiple patients !
REPLACING THE BATTERY
The meter comes with ONE Lithium batteries (battery type CR2032) Before you start,
please install battery with the plus(+) side up. Battery life will vary depending on usage, so always keep a spare battery on hand. The battery should last about 1000 tests or 12 months when testing 3 times a day. When the battery symbol appears on the meter display, battery is getting low. You will
still be able to test with low battery, but you should replace it as soon as possible. When battery symbol appears together with E-b on the display, the meter will no longer
give results and you must replace the battery immediately. Please always have one spare battery with you to ensure that you can replace the battery anytime.
How to replace the battery
1. Make sure the meter is turned off. Let the front of the meter
rest in the palm of your hand. Press the buckle on the battery cover and lift up to open the cover. Remove the battery.
2. With the plus (+) side up, place the battery (battery type CR2032) in the compartment and slide it upwards until it is locked into place. You should hear a beep to indicate the battery installed correctly. If not, please replace the battery correctly.
3. Close battery cover.
NOTE:
1. Replacing the battery does not affect the test result stored in memory. However the time and date may need to re-set.
2. As with all small objects, the battery should be kept away from small children as a safety precaution. If the battery is swallowed, seek medical assistance immediately.
3. Batteries might leak chemicals if not used for a long time. Remove the batteries if you are not going to use the device for extended period(i.e., 3 months or more ).
4. Please discard the product or the batteries properly according to the regulations
of your country.
For
in vitro
diagnostic use. Temperature limitation / Store at.
Please consult instructions for use.
Use by / Expiry date.
Do not reuse.
Manufacturer.
Lot number. EU representative.
Caution, consult accompanying document.
This product fulls the requirements of Directive 98/79/EC
in vitro
diagnostic medical device.
Performance Characteristics: Accuracy: ≧ 95% of the individual glucose results fall within ± 15 mg/dL(0.83 mmol/L) at glucose concentration < 75 mg/
dL(or 4.2 mmol/L) and within ± 20% at glucose concentration ≧75 mg/dL(or 4.2 mmol/L), compared with glucose analyzer
YSI 2300. PRECISION: CVs (%) of intermediate precision and repeatability were less than 5%.
DISPLAY :
Shows test results and messages.
METER LABEL :
Each meter has its unique number on it. Do not alter or tear the label off.
SET BUTTON :
Located in the battery
compartment; press button to enter meter setting. Refer to "
SETTING TIME AND DATE
"
section for details.
BATTERY COMPARTMENT :
Holds ONE 3V Lithium battery
(battery type CR2032).
STRIP SLOT :
Insert the test strip here. The meter will turn on automatically.
M BUTTON :
Main button, press button to turn the meter on or perform other functions described in User Guide.
DATA PORT :
USB port, allows you to transfer the information stored
in the meter to a computer to view, analyze and print.
1 5 62
3
8
10
7
9 4
Blood Glucose Monitoring System
I
t
'
s
t
i
m
e
t
o
t
r
a
v
e
l
l
i
g
h
t
OK Biotech Co., Ltd. No. 91, Sec. 2, Gong Dao 5th Road,
30070 Hsinchu City, TAIWAN
TEL: +886-3-5160258
FAX: +886-3-5160028 www.okbiotech.com E-mail: service@okbiotech.com
MDSS GmbH Schiffgraben 41, 30175 Hannover,
GERMANY
V2-DEC14
Setting your Lancing Device
1. Screw off the cap of lancing device. Insert a lancet into the lancet holder and push down until it is fully seated.
2. Twist off the protective cap until it separates from the lancet.
3. Replace the lancing device cap and set the puncture depth to the desired number.
NOTE: THE DEPTH SELECTOR OFFERS 5 LEVELS OF SKIN
1-2 : FOR SOFT OR THIN SKIN
3 : FOR AVERAGE SKIN
4-5 : FOR THICK OR CALLOUSED SKIN
4. Pull back the cocking control until it makes a click, and then release. If it does not click, the device may have been cocked when the lancet was inserted.
Lancing Device Cap
Lancet Holder
Depth Selector
Release Button
Cocking Control
SIDE
Control Solution Testing
OKmeter control solutions contain a known amount of glucose that reacts with OKmeter Cameo test strips. By testing your control solution and comparing the test results with the expected range printed on the test strip vial label, you can make sure that the meter and the test strips are working properly together as a system and that you are performing the test correctly. It is very important that you do this simple check routinely to make sure you get accurate results.
Why perform a control solution test?
1. To ensure that your meter and test strip are working properly together.
2. To allow you to practice testing without using your own blood.
When should the control solution test be performed?
1. When you first get your OKmeter Cameo meter. Before use this system to test your blood, you can practice the procedure by using control solution. When you can do three tests in a row that are within the expected range, you are ready to test your blood.
2. Once a week (to make sure that you continue to get accurate results)
3. When you begin using a new vial of test strips.
4. Whenever you suspect that the meter or test strips are not working properly.
5. When your blood glucose test results are not consistent with how you feel, or when you think your results are not accurate.
6. When test strips are exposed to extreme environmental conditions.
7. If you drop the meter.
8. After changing the batteries.
Important Control Solution Information
1. Check the expiration date on the control solution bottle. Do not use if expired.
2. Control solution, meter, and test strips should come to room temperature (68-77 ºF/20-25ºC) before testing.
3. Shake the bottle before use, discard the first drop of control solution after
squeezing, wipes off the dispenser tip to avoid contaminations. These steps
ensure you will get a good sample and an accurate result.
4. Record the discard date on the bottle when you open a new bottle of control solution.
NOTE :
1. There are two levels of control solution (normal and high) available to purchase. Please contact with your local distributor when required.
2. The control solution range printed on the test strip vial is for OKmeter Control Solution only. It is used to test meter and strip performance. It is not recommended range for your blood glucose level.
How to Perform a Control Test
NOTE :
1. DO NOT APPLY THE CONTROL SOLUTION DIRECTLY TO THE TEST
STRIP! Overdosed solution may give inaccurate result.
2. Repeat test if test result falls outside the control range stated on the test strip label. If subsequent test remains to produce unacceptable result, the meter or test strip may be faulty. DO NOT use the system. Contact us or your local distributor for help.
1.
Insert Test Strip :
Insert a new test strip into the strip slot,
the meter will activate.
2.
Mark as a Control Solution Test :
After the blood symbol( ) appears, press M BUTTON and a CONTROL SOLUTION TEST SYMBOL( ) shows up, now you are in the Control Solution Mode. The meter will not store your test result in the memory when you preset the test as a control solution test. If you decide not to perform a control
solution test, press M BUTTON again and the CONTROL SOLUTION TEST SYMBOL ( ) will disappear.
3.
Squeeze a drop of Control Solution :
Shake control
solution bottle well. Remove the cap. Squeeze the bottle, discard the rst drop and wipe off the dispenser tip with a clean tissue paper or cotton. Squeeze the bottle again to
get the second drop onto a clean non-absorbent surface
or on your ngertip rst.
4.
Apply Control Solution :
Apply the drop to the opening
of the strip absorbent channel until the confirmation
window is lled. The meter begins to count down.
5.
Check if the test result is in range:
After the meter counts down from 6 to 1, the test result shows up. Compare the test result with the range printed on the test strip vial. The result should fall within the printed range.
5
.
About Alternative Site Testing
(AST )
There are important limitations for doing AST. Please consult your healthcare professional before you perform AST.
What is AST?
Alternative Site Testing (AST) means you can use parts of the body other than your
ngertips to check your blood glucose levels. The system allows you to test from the palm, forearm, upper arm, calf or thigh, with equivalent results to ngertip testing.
What is the advantage?
Fingertips feel pain more readily because they are full of nerve endings (receptors). At other body sites, nerve endings are not so numerous and you will not feel as much
pain as you will experience at the ngertip.
When to use AST?
Food, medication, illness, stress and exercise can affect blood glucose levels.
Capillary blood at ngertip reects these changes faster than capillary blood at other
sites. Therefore, if you are testing blood glucose level during or immediately after
meal, physical exercise or stressful event, take the blood sample from your ngertip
only.
Use AST only:
1. In a pre-meal or fasting state (more than 2 hours since the last meal).
2. Two hours or more after taking insulin.
3. Two hours or more after exercise.
4. During steady state blood glucose conditions.
Do NOT use AST if:
1. You have reason to believe you have hypoglycemia or hyperglycemia.
2. Your routine glucose results are often uctuating.
3. You are pregnant.
How to increase the accuracy?
Stimulating blood perfusion by rubbing the puncture site prior to blood extraction has
a signicant inuence on the glucose value obtained.
Blood from the site without rubbing exhibits a measurably different glucose
concentration than blood from the ngertip. When the puncture site was rubbed prior to blood extraction, the difference was signicantly reduced.
IMPORTANT :
To increase the accuracy when using AST, rub the puncture site more than 20 seconds before extracting blood.
8
.
The following is a summary of some display messages and symbols. These messages help to identify certain problems but do not appear in all cases when a problem has occurred.
Improper use may cause an inaccurate result without producing an error message. In the event of a problem, refer to information under ‘‘action to take’’.
DISPLAY DESCRIPTION ACTION TO TAKE
Display check.
If some parts of the display are not working. Contact your local distributor for help.
Moving Strip.
The meter is waiting for test strip be insert.
Blinking Blood.
The meter is ready for blood applying into test strip.
Deleting memory. Deleting is complete.
Display Messages And Problem-Solving Guide
9
.
B
Caring For Your Meter And Test Strip
To avoid the meter and test strips getting dirt, dust or other contaminants, please wash and dry your hands thoroughly before use.
Cleaning
Your meter does not require special maintenance. As long as no blood or control solution comes in direct contact with the meter, there is no special cleaning required.
To clean the meter exterior, wipe with a cloth moistened with tap water or a mild cleaning agent, then dry the device with a soft and dry cloth. Do not flush with water.
Do not use organic solvents to clean the meter. Your meter is a precision instrument. Please handle it with care.
Storage
1. Meter Storage
* Storage condition: -20ºC~50ºC (-4ºF~122ºF), below 90% relative humidity. * Avoid dropping and strong impact. * Avoid direct sunlight and humidity.
2. Strip Storage
* Storage condition: 4ºC~40ºC (39ºF~104ºF), and 10~85% relative humidity. Do
not refrigerate.
* Store your test strips in their original vial only. Do not transfer to other
container.
* Store test strip packages in a cool and dry place. Keep away from direct
sunlight and heat.
* After removing a test strip from the vial, immediately replace the vial cap and
close it tightly.
* You may touch the test strip anywhere with clean, dry hands when removing it
from the vial or inserting it into the meter.
* Use each test strip immediately after removing it from the vial. * Do not bend, cut, or alter a test strip in any way. * Keep the strip vial away from children since the cap and the test strip can be
a potential choking hazard. If swallowed, please seek medical assistance
immediately.
3. Control solution storage
* Storage condition: Store the control solution tightly closed at temperatures
between 4°C (39°F) and 30°C (86°F). Do not refrigerate.
6
.
System Specifications
Model Name OK-P Assay Method Electrochemical biosensor Test Sample Capillary whole blood Test Result Referenced to plasma glucose value Alternative Site Testing YES (palm, forearm, upper arm, calf, and thigh)
Sample Size 0.7 μL
Measuring Time 6 seconds Measuring Range 20 – 600 mg/dL (1.1-33.3 mmol/L) Acceptable Hematocrit Range 20~60% Operating Condition
10°C~40°C(50°F~104°F), between 10-85% R. H.
Altitude ≦ 3,402 m (11,161 ft) above the sea level
Strip Storage/ Transportation Condition
4ºC~40ºC (39ºF~104ºF), 10~85% R.H.
Memory Capacity 120 test results with time and date Average Calculation 7, 14, and 28 days Power Supply One 3-volt Lithium Battery (battery type CR2032) Battery Life Approximately 1,000 tests Dimensions 90 mm(L) × 35 mm(W) ×21 mm(H) Weight 37 g
7
.
8
.
Performing Blood Test
1.
Wash Your Hands and the Puncture Site :
Wash your
hands in warm, soapy water. Rinse and dry completely.
Warm your ngers to increase blood ow.
2.
Insert Test Strip :
Remove a new test strip from vial. Be sure to tightly replace vial cap after removing test strips. Insert test strip immediately into strip slot as illustrated. The meter turns on automatically. When the blood symbol blinking, you are ready to perform a test.
3.
Select and Lance a Puncture Site
*For Fingertip
Hold the prepared lancing device rmly against the side of
your ngertip. Press release button.
*For AST
Please refer to the “About Alternative Site Testing (AST)”
Section. Please consult your healthcare professional
before obtaining blood from site other than your ngertip.
4.
Obtain a Blood Sample :
Gently massage your finger
or puncture site to obtain the required blood volume. To
perform the test, you need only 0.7 μL of blood sample.
Do not smear the blood sample. To obtain best accurate
result, wipe off the rst drop of blood and gently squeeze
another drop of blood.
5.
Apply Blood Sample :
Apply the blood sample to the
opening of absorbent channel of test strip until the
conrmation window is fully covered with blood. Then the
meter begins to count down and displays the test result in 6 seconds.
6.
Read Your Result :
The meter will display your blood
glucose value after 6 seconds.
7.
Remove Strip to Turn Meter Off :
Your blood glucose result is automatically stored in the meter memory. Turn the meter off by removing the test strip. Discard the used strip and lancet safely in a puncture resistant container.
8.
Secure the Used Lancet :
Twist off the lancing device cap,
and push the exposed tip of the lancet into its protective cap.
Incorrect Correct
3
.
Memory Recall
The OKmeter Cameo automatically stores 120 test results, letting you review them in order from the most recent to the oldest. The meter also calculates and displays 7, 14 and 28-day averages. You can review the individual or average test result by entering the memory mode.
Recall the Memory
STEP 1. Enter the Memory Mode
While the meter is turned off, press M BUTTON to turn on the meter. When the strip symbol blinks in the display, press M BUTTON again to enter memory mode.
STEP 2. Recalling Average Test Results
When entering the memory mode, the 7-day average will appear. If you continue to press the M BUTTON, the 14-day and 28-day averages will appear in order.
STEP 3. Recalling Individual Test Results
After 28-day average, the most recent test result with date and time will be shown. Press M BUTTON once and the next most recent test result will appear. Each time you press and release the M BUTTON, the meter will recall up to your last 120 test results in order. When the memory is full, the oldest result is dropped as the newest is added. After reaching the last set of result, the meter will display 7-day average again.
STEP 4. Exit the Memory Mode
I
f you want to exit the memory mode, press and hold M
BUTTON for three (3) seconds to turn off the meter.
4
.
Test result is higher than
600 mg/dL (33.3 mmol/L).
If this is not conrmed by the way
you feel, review proper testing procedure and perform a control test. Repeat blood test, if the display still appears, please call medical assistance immediately.
Test result is lower than 20
mg/dL (1.1 mmol/L).
If this is not conrmed by the way
you feel, review proper testing procedure and perform a control test. Repeat blood test, if the display still appears, please call medical assistance immediately.
Battery is weak. Replace battery soon.
Battery is dead. Replace battery now.
Maybe:
1. Used strip or moistened strip
2. Defective meter.
You have to:
1. Repeat test with a new test strip.
2. Contact your local distributor for help.
Temperature is out of the operating range.
The meter is not working. Move to an area with temperature between 10ºC to 40ºC (50ºF - 104ºF) and wait at least 30 minutes. Do not
articially heat or cool the meter.
System error
Contact your local distributor for help.
No responses when the test strip is inserted into the meter.
There are multiple potential causes :
1. Battery is dead.
2. Wrong strip was inserted.
3. Meter is defective.
Depending on the cause, please take these given actions :
1. Replace the battery.
2. Insert the test strip correctly.
3. Contact us for help.
No responses when blood sample is applied to the test strip.
Maybe:
1. Blood sample is not
sufcient.
2. Meter is defective.
Please repeat the test with sufficient blood volume. If you still see the same results, please contact us for help.
IMPORTANT :
1.The OKmeter Cameo only requires a tiny sample of blood to perform a test. Choose a different spot each time you test. Repeat puncture in the same spot may cause soreness and calluses.
2.The rst drop of blood usually contains tissue and serum, which may affect the test
result. It should be discarded.
3.The blood should be completely filled the confirmation window before the meter
begins to count down. If you nd that the conrmation window is not lled with blood
when the meter is counting, NEVER try to add more blood to the test strip. Discard the test strip and retest with a new one.
4.If you do not apply a blood sample within three (3) minutes, the meter will automatically turn off. You must remove the test strip and re-insert it again to turn on the meter and restart the test procedure.
5.If you have trouble filling the test strip, please contact your local distributor for assistance.
NOTE:
1.
The control solution results are NOT stored in the memory if you preset the control test
correctly. (Please refer to the “CONTROL SOLUTION TEST” Section for details). The list of
past results and the result average are for blood glucose results only.
2.
When using the meter for the rst time, ‘‘----‘’ is displayed. When you recall the test results or
review the average result, it means that there is no test result in memory.
3.
The averages are calculated from your latest result obtained during the last 7, 14 and 28 days.
4.
Anytime in memory mode, you can press M BUTTON for three (3) seconds to exit and turn off the meter.
5.
While in the memory mode, if you leave the meter alone without any action for one (1) minute, the meter will turn off automatically.
Loading...