Ohmeda Giraffe, OmniBed Service Manual

Giraffe
®
OmniBed
Service Manual
®
Important
The information contained in this service manual pertains only to those models of products which are mar­keted by Ohmeda Medical as of the effective date of this manual or the latest revision thereof. This service manual was prepared for exclusive use by Ohmeda Medical service personnel in light of their training and experience as well as the availability to them of parts, proper tools and test equipment. Consequently, Ohmeda Medical provides this service manual to its customers purely as a business convenience and for the customer’s general information only without warranty of the results with respect to any application of such information. Furthermore, because of the wide variety of circumstances under which maintenance and repair activities may be performed and the unique nature of each individual’s own experience, capacity, and qualifi­cations, the fact that a customer has received such information from Ohmeda Medical does not imply in anyway that Ohmeda Medical deems said individual to be qualified to perform any such maintenance or repair service. Moreover, it should not be assumed that every acceptable test and safety procedure or method, precaution, tool, equipment or device is referred to within, or that abnormal or unusual circumstances, may not warrant or suggest different or additional procedures or requirements.
This manual is subject to periodic review, update and revision. Customers are cautioned to obtain and consult the latest revision before undertaking any service of the equipment.
CAUTION w Servicing of this product in accordance with this service manual should never be undertaken
in the absence of proper tools, test equipment and the most recent revision to this service manual which is clearly and thoroughly understood.
This static control precaution symbol appears throughout this manual. When this symbol appears next to a procedure in this manual, static control precautions MUST be observed. Use the static control work station (Stock No. 0175-2311-000) to help ensure that static charges are safely conducted to ground and not through static sensitive devices.
Technical Competence
The procedures described in this service manual should be performed by trained and authorized personnel only. Maintenance should only be undertaken by competent individuals who have a general knowledge of and experience with devices of this nature. No repairs should ever be undertaken or attempted by anyone not having such qualifications. Genuine replacement parts manufactured or sold by Ohmeda must be used for all repairs. Read completely through each step in every procedure before starting the procedure; any exceptions may result in a failure to properly and safely complete the attempted procedure.
Definitions
Note: A note provides additional information to clarify a point in the text. Important: An Important statement is similar to a note, but is used for greater emphasis. CAUTION: A CAUTION statement is used when the possibility of damage to the equipment exists.
WARNING: A WARNING statement is used when the possibility of injury to the patient or the operator exists.
m x
y
~
Type B Electrical equipment Protective ground Functional Ground Alternating Current (AC) Static Control Precaution European Union Representative
Table of Contents
Chapter 1 – Functional Description
1.1 Control Board .............................................................................................1-1
1.2 Relay Board ...............................................................................................1-3
1.3 Display Driver Board/EL Display.................................................................1-4
1.4 LED Board .................................................................................................1-4
1.5 Power Supply .............................................................................................1-4
1.6 Peripheral Components..............................................................................1-4
1.61 Rail and Heater Door Switches ..........................................................1-5
1.7 DataLink Option .........................................................................................1-5
1.8 Servo Controlled Oxygen Option ................................................................ 1-6
Chapter 2- Service Checkout
2.1 Setup..........................................................................................................2-1
2.2 Mechanical checks.....................................................................................2-1
2.3 Controller checks .......................................................................................2-2
2.4 Humidity check...........................................................................................2-3
2.5 Servo Controlled Oxygen check .................................................................2-3
2.6 Accessory checks ......................................................................................2-4
2.5 Cable Connections and Mechanical Controls ............................................. 2-5
Chapter 3- Calibration and Maintenance
3.1 Maintenance schedule ...............................................................................3-1
3.2 Special tools...............................................................................................3-1
3.3 System Calibration .....................................................................................3-3
3.4 Line Volt age Calibration..............................................................................3-3
3.5 Humidifier Calibration .................................................................................3-4
3.6 Servo Controlled Oxygen Calibration .........................................................3-4
3.7 Scale Calibration ........................................................................................3-5
3.8 Leakage Current ........................................................................................3-5
3.9 Ground Resistance ....................................................................................3-5
Chapter 4- Troubleshooting
4.1 Service Screen...........................................................................................4-1
4.2 Alarm Messages ........................................................................................ 4-5
4.3 Error Codes................................................................................................4-8
4.4 Troubleshooting Table ................................................................................4-13
4.5 Additional Troubleshooting Tips..................................................................4-16
4.6 Servo Controlled Oxygen ...........................................................................4-19
Servo Controlled Oxygen Service Screen..................................................4-19
Servo Controlled Oxygen Alarm Messages ...............................................4-20
Servo Controlled Oxygen Troubleshooting Tips .........................................4-22
Chapter 5- Repair Procedures
5.1 Canopy removal for replacement ...............................................................5-1
5.1 1 Replacing the seals............................................................................ 5-2
5.12 Porthole Door Replacement...............................................................5-3
5.2 Canopy Lift Assembly Repair Procedures ..................................................5-4
5.21 Removing the right upright (motor side) ............................................. 5-4
5.22 Removing the left upright ...................................................................5-7
5.23 Right rail internal repairs ....................................................................5-8
5.231 Removing the inner rail assembly..............................................5-8
5.232 Replacing rollers and tension springs ........................................ 5-8
5.233 Replacing the micro-switches....................................................5-9
5.234 Separating the lift rail from the belt channel...............................5-9
5.235 Replacing the drive belt .............................................................5-10
5.236 Replacing the rail buoyancy springs ..........................................5-10
5.24 Re-assembling the right upright .........................................................5-11
5.25 Left rail internal repairs.......................................................................5-13
6600-0343-000 08/27/03 i
Table of Contents
5.251 Removing the inner rail assembly..............................................5-13
5.252 Replacing the rail buoyancy springs ..........................................5-14
5.253 Replacing rollers, tension springs and cable carrier links ..........5-14
5.26 Re-assembling the left upright ...........................................................5-14
5.3 Radiant Heater Assembly Rep air Procedures ............................................5-16
5.31 Removing the Canopy/Heater Assembly............................................5-16
5.31 1 Re-aligning the canopy ..............................................................5-17
5.32 Heating Element ................................................................................5-17
5.33 Heater doors ......................................................................................5-18
5.34 Heater door cable adjustment............................................................5-18
5.4 Compartment Probe repairs .......................................................................5-19
5.5 Top rail end cap replacement .....................................................................5-20
5.6 Lower Unit Repairs.....................................................................................5-21
5.61 Removing the chassis cover with the storage door in place...............5-21
5.62 Incubator fan/motor/optical sensor.....................................................5-21
5.63 Cartridge heater replacement ............................................................5-21
5.64 Elevating base ...................................................................................5-23
5.65 Chassis replacement .........................................................................5-24
5.66 Elevating footswitch ...........................................................................5-25
5.67 Canopy footswitch..............................................................................5-25
5.68 Canopy finger switch..........................................................................5-25
5.69 Humidifier heater components ...........................................................5-25
5.7 Bed Tilt Brake Shoe Replacement..............................................................5-25
5.8 Castor Replacement ..................................................................................5-26
5.9 Humidifier Repairs......................................................................................5-27
5.10 Controller and Display Module repairs......................................................5-29
5.1 1 1 Display module ..........................................................................5-29
5.1 12 Probe panel ...............................................................................5-30
5.1 13 Controller components...............................................................5-31
Control Board ...................................................................................5-31
Relay Board......................................................................................5-32
Solid State Relays ............................................................................5-32
Power supply....................................................................................5-32
Battery..............................................................................................5-32
Transformers....................................................................................5-32
Circuit breakers, power switch and outlets .......................................5-33
5.1 14 Servo Controlled Oxygen Service Procedures...........................5-33
5.1 141 Inst alling oxygen sensors......................................................5-33
5.1 142 Replacing the vent screen .................................................... 5-33
5.1 143 Sensor housing rep airs.........................................................5-34
5.1 114 Valve housing repairs ............................................................5-35
5.1 115 Endcap Safety valve .............................................................5-36
Chapter 6- Illustrated Parts
6.1 Exploded Views..........................................................................................6-1
6.1 1 Probe housing, display module and electrical enclosure ....................6-1
6.12 Bed and side panels ..........................................................................6-8
6.13 Radiant heater and canopy ................................................................6-14
6.14 Chassis..............................................................................................6-19
6.15 Humidifier ..........................................................................................6-22
6.16 Elevating base ...................................................................................6-24
6.17 Uprights and lift rail components........................................................6-26
6.18 Compartment probe...........................................................................6-34
6.19 Servo Controlled Oxygen ...................................................................6-35
6.20 Storage drawer and shelves...............................................................6-39
6.2 Accessories................................................................................................6-44
6.3 Labels ........................................................................................................6-45
6.4 PCB layouts ...............................................................................................6-48
6.5 Wiring diagrams .........................................................................................6-52
ii 08/27/03 6600-0343-000
Table of Contents
6600-0343-000 08/27/03 iii
Appendix
Compartment and Skin Probe Characteristics .................................................A-1
Specifications...................................................................................................A-2
Power requirements...................................................................................A-2
Operating Environment..............................................................................A-2
Storage Conditions ....................................................................................A-2
User Control Settings.................................................................................A-2
Alarms .......................................................................................................A-2
Performance..............................................................................................A-3
System ................................................................................................A-3
Humidity ..............................................................................................A-3
Servo Controlled Oxygen ....................................................................A-3
Mechanical Specifications..........................................................................A-4
RS-232 Serial Data ..........................................................................................A-5
Data S tream...............................................................................................A-5
Nurse Call..................................................................................................A-6
List of Figures
1-1 Block Diagram ...........................................................................................1-2
2-1 Cable connections and mechanical controls..............................................2-5
3-1 Control board test points............................................................................3-2
4-1 First service screen ...................................................................................4-1
4-2 Second service screen ..............................................................................4-2
4-3 Status menu...............................................................................................4-2
4-4 Switch diagnostic diagram .........................................................................4-3
4-5 Pedal screen .............................................................................................. 4-3
4-6 First service screen- diagnostics................................................................4-4
4-7 Servo O2 service screen............................................................................4-19
5-1 Heater housing cover and soffit .................................................................5-1
5-2 Canopy , bracket and heater housing..........................................................5-2
5-3 Canopy seals and extrusions .....................................................................5-3
5-4 Disconnecting door cable...........................................................................5-4
5-5 Upright decorative strips, end caps and wire covers .................................. 5-5
5-6 Removing the lift motor ..............................................................................5-5
5-7 Display module disassembly ...................................................................... 5-6
5-8 Right upright assembly............................................................................... 5-8
5-9 Rollers and tensioning spring .....................................................................5-9
5-10 Replacing the drive belt............................................................................5-10
5-1 1 Replacing the buoyancy springs...............................................................5-11
5-12 Left inner rail ............................................................................................5-12
5-13 Left rail springs and spools ......................................................................5-13
5-14 Re-attaching the upright...........................................................................5-14
5-15 Radiant heater disassembly .....................................................................5-15
5-16 Canopy alignment ....................................................................................5-16
5-17Heater door cable adjustment ...................................................................5-18
5-18 Compartment probe disassembly.............................................................5-19
5-19 Top rail end cap........................................................................................5-19
5-20 Bed disassembly......................................................................................5-20
5-21 Fan motor ................................................................................................5-21
5-22 Heat sink and fan .....................................................................................5-22
5-23 Elevating base .........................................................................................5-23
5-24 Chassis bottom cover ..............................................................................5-24
5-25 Replacing the tilt brake............................................................................. 5-26
5-26 Humidifier parts........................................................................................5-28
iv 08/27/03 6600-0343-000
5-27 Display module ........................................................................................5-30
5-28 Probe panel.............................................................................................. 5-30
5-29 Electronics enclosure...............................................................................5-31
5-30 Installing sensors ..................................................................................... 5-33
5-31 Sensor housing ........................................................................................5-34
5-32 V alve housing...........................................................................................5-35
5-33 Endcap safety valve .................................................................................5-36
6-1 Probe Panel Assembly...............................................................................6-1
6-2 Display Module...........................................................................................6-3
6-3 Electrical enclosure....................................................................................6-5
6-4 Humidifier Transformer and Rs232 option..................................................6-7
6-5 Bed ............................................................................................................6-9
6-6 Side panel (E/W)........................................................................................6-11
6-7 Rear (north) and front (south) panel...........................................................6-13
6-8 Radiant Heater...........................................................................................6-15
6-9 Canopy ......................................................................................................6-17
6-10 Heater doors ............................................................................................6-18
6-1 1 Upper chassis ..........................................................................................6-19
6-12 Lower chassis ..........................................................................................6-21
6-13 Humidifier................................................................................................. 6-23
6-14 Base and elevating column ......................................................................6-25
6-15 Right (East) upright, motor side................................................................6-27
6-16 Belt channel -1 (lift motor side).................................................................6-28
6-17 Belt channel -2 (lift motor side).................................................................6-29
6-18 Lift rail (both sides)...................................................................................6-30
6-19 Rail End caps...........................................................................................6-31
6-20 Left (west) upright ....................................................................................6-33
6-21 Compartment Air Probe............................................................................6-34
6-22 Servo Control Oxygen Sensor Housing assembly....................................6-35
6-23 Servo Control Oxygen V alve Housing ......................................................6-36
6-24 Expansion Chamber/Heatsink Vent..........................................................6-37
6-25 Servo Control Oxygen Cooling Fan..........................................................6-37
6-26 Servo Control Oxygen PC board ..............................................................6-38
6-27 Manifold Endcap Safety V alve..................................................................6-38
6-28 Strorage Drawer.......................................................................................6-39
6-29 Instrument Shelf.......................................................................................6-40
6-30 Monitor Shelf............................................................................................6-40
6-31 E Cylinder holder......................................................................................6-41
6-32 Tubing Management Arm.........................................................................6-41
6-33 Dovetail mount DIN rail ............................................................................6-42
6-34 Silo Support .............................................................................................6-42
6-35 Rotating IV Pole Assy. ..............................................................................6-43
6-36 Dovetail extension....................................................................................6-44
6-37 Control board ...........................................................................................6-48
6-38 Display driver board .................................................................................6-49
6-39 Relay board (Rev 10 or higher) ................................................................6-50
6-40 Relay board (Rev 9 or lower) ...................................................................6-51
6-41 Wiring Diagram Control Board .................................................................6-52
6-42 Wiring Diagram Elevating Base and Canopy Lift Rails.............................6-53
6-43 Wiring Diagram Electrical Enclosure........................................................6-54
6-44 Wiring Diagram Graphics Display ............................................................6-55
6-45 Wiring Diagram Incubator (Relay board rev 10 or higher) ........................6-56
6-46 Wiring Diagram Incubator (Relay board rev 9 or lower) ...........................6-57
6-47 Wiring Diagram Radiant heater (Relay board rev 10 or higher)................6-58
6-48 Wiring Diagram Incubator (Relay board rev 9 or lower) ...........................6-59
6-49 Wiring Diagram Servo Humidifier (Relay board rev 10 or higher)............. 6-60
6-50 Wiring Diagram Servo Humidifier (Relay board rev 9 or lower) ................6-61
6-51 Wiring Diagram Servo Control Oxygen .................................................... 6-62
6600-0343-000 08/27/03 v
Precautions
wWarnings
Before using the OmniBed, read through this entire manual. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. This device should only be operated by personnel trained in its operation under the direction of qualified medical personnel familiar with the risks and benefits of this type of device. Additional precautions specific to cert ain proce-
dures are found in the text of this manual.
Complete the “Pre-operative Checkout Procedures” section of the Operator’s manual before putting the unit into operation. If the incubator fails any portion of the checkout procedure it must be removed from use and repaired.
Do not use the OmniBed in the presence of flammable anesthetics; an explosion hazard exists under these conditions.
Always disconnect the power before performing service or maintenance procedures detailed in this manual. Apply power only if you are specifically instructed to do so as part of the procedure.
Thoroughly air dry the incubator after cleaning it with flammable agents. Small amounts of flammable agents, such as ether, alcohol or similar cleaning solvent s left in the incubator can cause a fire.
wCautions
Only competent individuals trained in the repair of this equipment should attempt to service it as detailed in this manual.
Detailed information for more extensive repairs is included in the service manual solely for the convenience of users having proper knowledge, tools and test equipment, and for service representatives trained by Ohmeda Medical.
vi 08/27/03 6600-0343-000
Chapter1- Functional Description
This functional description is divided into four sections representing each of the four boards. The reader should also reference the block diagram and wiring diagram when studying this section.
1.1 Control Board
The Intel 80C188EC microcontroller is an enhanced X86 processor with many on-board peripheral features, such as a interrupt controller, DMA controller, peripheral chip select driver, programmable timers, etc. The two programmable timers are used to control the two heaters (bed and radiant). The input to these timers is line frequency . This allows the control signal to be synchronized with the line frequency to better control the zero­crossing solid state relays. The on-board interrupt controller has several interrupts: analog-to-digital converter (ADC) conversion ready signal, overtemperature comparator output, watchdog output, power fail signal, and module interrupt signal from the system data bus. The microcontroller external bus is a multiplexed address and data bus.
The system memory consists of a programmable read-only memory (PROM) and static random access memory (SRAM). The EEPROM is used for calibration values and infrequently changing variables. This memory holds the data even after power is turned off.
The RS-485 integrated circuit converts the RS-232 TTL signals from the microcontroller to RS-485 signals for the bus. This bus is the main communications bus from the control board to all other boards with processors.
There are two isolation transceivers used to isolate the circuits powered by +5V and the circuits powered by +5VSTBY (battery backup).
The board contains a 16 channel multiplexer. There are seven temperature measurement channels. These channels measure the two patient probes with two thermistors each, the two air temperature thermistors used for display and control, and an additional thermistor used to measure the heat sink temperature. Additional channels include the humidity sensor (RHIN), LINE COMP & LINE COMP2, 5 V olt s, Motor current, Vthref, VDAC, and 1.2Vind.
Attached to the environmental probe connection is the relative humidity signal conditioning circuitry. The 1V reference that is used for the analog circuitry is also the maximum input voltage and the offset voltage for the ADC. This yields a purely ratiometric system.
The overtemperature circuit compares the air temperature to a reference level, generates an interrupt, and turns off the heat if the air temperature is higher than the reference level. The overtemperature circuit requires varying its voltage levels to accommodate various thermistor measurements. This is because the calibration is digital (no potentiometer).
The watchdog circuitry monitors the 80C188 microprocessor, and monitors the +5V and +5VSTBY voltages. It generates the interrupt signal and power failure signal to the 80C188 microprocessors. The audio circuit includes a 8752 microcontroller that reads a wavetable located in a PROM and sends the table to a digital audio circuit and amplifier. The high priority (HP) and other alarm signal lines select an output at the correct frequencies.
Three OR gates are combined to generate the error signal. The inputs to the circuit are overtemperature, power failure, and system failure. This circuit generates an error signal that turns off the heater and sounds the HP alarm. This circuit is independent of the microcontroller.
6600-0343-000 08/27/03 1-1
Chapter 1- Functional Description
LED Board
EL Display
Microprocessor
PROM
SRAM
Microprocessor
PROM
SRAM
Audio
Time
Display Driver Board
Temp
Channels
Overtemp
Watch
Dog
Membrane Switches
Knob
Figure 1-1
Block Diagram
Control Board
Relay Board
RS485
Future Options
and
Modules
Power Supply
1-2 08/27/03 6600-0343-000
Chapter1- Functional Description
1.2 Relay Board
The Relay Board includes 2 safety relays, which close to supply mains power to the heater and motor circuits. Safety relay 1 is wired in series with the primary coil of the isolation transformer for the incubator and radiant warmer heaters. Safety relay 2 closes the mains supply to the humidifier isolation transformer and the trans­former for the e-base and canopy motors. Control signals for the two relays originate on the Control Board.
The Relay Board interfaces the DC control signals to the two chassis mounted solid-state relays (SSRs), which control the incubator and radiant warmer heaters individually . The control signals for the two heater SSRs originate on the Control Board.
The Relay Board includes a SSR for the humidifier. The SSR output is wired in series with the humidifier heater. The humidifier SSR control signal originates on the Control Board.
There is one current sense circuit for the incubator and radiant warmer heaters and an additional one for the humidifier heater. These circuits consist of a small signal transformer that produces a current proportional to the current through the heater circuits. The current is rectified and measured. The subsequent comparator then generates a digital level based on a specified current level. This results in a signal to the Control Board representing the state of the heater (on or off).
The two line compensation circuits consist of a signal transformer connected to the mains voltage. The secondary of this transformer feeds a full wave rectifier and capacitor . The resulting DC voltage is proportional to mains voltage, and it is measured on the Control Board.
The line frequency circuit consists of a full wave rectifier and a comparator . This circuit generates a digit al pulse with frequency twice that of the line frequency (50 or 60 Hz). The output signal is provide to the Control Board.
The Relay Board provides the +5v standby power supply to the entire Giraffe system. A +5V regulator gener­ates the +5V standby from the diode OR combination of the system +12V power supply or the backup battery . If there is no mains power, then +12v is not present, and the battery will generate the +5V st andby. When +12V is present, the battery is biased out of the circuit with the diode and is merely being trickle charged though a resistor.
The motor driver circuit turns the DC motor coils in the incubator airflow fan motor on and off based on feed­back from the hall effect position sensors. This integrated circuit can also vary the speed and brake the motor based on input signals from the Control Board.
The airflow sensor consists of an opto-coupler that outputs a clocking pulse proportional to the fan movement. The signal is AC coupled to eliminate of fset voltages and drif t s. The resulting pulse is half wave rectified and stored in a capacitor to yield a DC voltage proportional to the fan speed. If the fan stops or there is no fan, this DC voltage becomes zero. The output signal is provide to the Control Board to indicate proper airflow motor operation.
The hood lift and elevating base circuit consists of a series of relays that apply voltage to the hood lift motor or the elevating base motor. The hood lif t motor is driven at 30V going up and 15V going down. The e-base motor is always driven at 30 volts. The motor current sense circuit consists of a small signal transformer that produces a current proportional to the motor current. The transformer output current is converted to a voltage and filtered. An output volt age indicative of the motor current amplitude is provided to the Control Board. A subsequent comparator then generates a digital level based on a specified current level. This results in a signal to the Control Board indicating whether or not the e-base motor is stalled.
The Relay Board interfaces the user and system status input switch signals to the Control Board. Switch signals include, e-base & canopy activation, canopy and heater door position, and humidifier reservoir and water level status.
6600-0343-000 08/27/03 1-3
Chapter 1- Functional Description
1.3 Display Driver Board / EL Display
The Display Driver board contains the same Intel microcontroller as the Control board. The processor on the display board is used to control the EL display contents and monitor user inputs received from the membrane switch panel and rotary encoder knob.
There are two groups of digital inputs: membrane switch panel and rot ary encoder knob. The membrane switches are pulled high; pressing the switch grounds the input. The encoder also has a switch, and two optically isolated lines that pulse out of phase with each other. The number of pulses represents the number of steps the knob rotates. The phase of the pulses represents the direction of the knob rot ation.
The display board system memory consists of a programmable read-only memory (PROM) and static random access memory (SRAM).
The RS-485 integrated circuit converts the RS-232 TTL signals from the microcontroller to RS-485 signals for the bus.
The timekeeping RAM has a battery integrated into the chip so that the time and date run are kept current even with the power off. The battery has a minimum life of 10 years.
The graphics controller is an S-MOS VGA controller. The graphics controller interfaces the data from the video RAM to the EL display. The controller also synchronizes the display using a horizontal pulse (LP) and a vertical pulse for the whole display frame (YD). The controller handshakes with the 80C188 using the READY line to eliminate any lost data during display refreshes.
1.4 LED Board
The LED Board contains five display banks and two display drivers. One of the display drivers controls the patient temperature and air temperature display banks. The other driver controls the patient set temperature, air set temperature, warmer bar graph, and the mode and override indicators.
This allows the two large displays (patient and air temperature) to be multiplexed at a slower rate than the other LEDs. This results in brighter large displays. Each driver has a brightness potentiometer that is preset at the factory and should not be adjusted in the field.
1.5 Power Supply
The universal input switching power supply converts the line voltage to +5V DC and +12V DC. This supply can source up to 75 watts. The 5 volts powers the electronics and the 12 volts is used by the EL display and for future boards.
1.6 Peripheral Components
There are several peripheral components. The isolation transformer isolates the radiant heater from the line voltage.
The toroidal transformer bucks the line voltage to the range of the elevating base and the canopy lift drive system.
The humidifier isolation transformer isolates the humidifier heater from the line voltage. The solid state relays mounted to the chassis are used to control the radiant and bed heaters.
1-4 08/27/03 6600-0343-000
Chapter1- Functional Description
1.61 Rail and Heater Door Switches There are seven switches used to determine the position of the canopy and the heater doors. Two normally open switches on each heater door determine the position of the doors. Each switch is wired in
series with the corresponding switch on the other door. One p air of switches closes only when the doors are fully open and the other pair closes only when the doors are fully closed.
There are 3 normally open switches in the right upright which are used to determine the position of the canopy . The top switch closes when the canopy reaches the upper position. The middle switch detects downward movement of the canopy . The bottom switch closes when the lowest position is reached.
The radiant heater will turn on only if the two heater door open switches and the top rail switch are all closed. The system will control as an incubator only if the bottom rail switch and the two heater door closed switches
are all closed. As the canopy lowers the system senses the closure of the middle rail switch, then looks at the heater door
closed switches. If they are not closed the canopy will return to the highest position. This insures that the canopy will not lower to the lowest position if the heater doors are not closed.
1.7 DataLink Option
The DataLink option allows direct output of serial data to various remote monitoring systems, such as a computer or commercial RS-232 monitor. The Dat aLink option board cont ains the electronic circuitry neces­sary to provide a 2500 VRMS isolated serial interface to meet the logic levels specified by EIA RS-232D and CCITTV.28.
The MAX250 and MAX251 (U1 and U2), together with two 6N136 optocouplers and transformer TR1, form an isolated RS-232 transmitter and receiver. The MAX250 connect s to the non-isolated or “logic” side of the interface, translating logic signals to and from the optocouplers, while the MAX251 resides on the isolated or “cable” side, translating data between the optocouplers and RS-232 line drivers and receivers. In addition to the optocoupler drivers and receivers, the MAX250 also contains isolation transformer drive circuitry which supplies power to the isolated side of the interface, and the MAX251.
The transmit signal is input to the MAX250 driver (U1 pin 4) whose output (U1 pin 3) drives optocoupler U4. The optocoupler output (U4 pin 6) is then fed into the MAX251 driver (U2 pin 3). The output of the MAX251 driver (U2 pin 12) is at the logic levels conforming to EIA RS-232D and CCITTV.28. Conversely, the receive signal enters the MAX251 driver (U2 pin 10) and is stepped down to CMOS/TTL levels at U2 pin 5. This logic level drives optoisolator input (U3 pin 3) whose output is fed into U1 pin 10. The output (U1 pin 9) signal is then available to the control printed circuit board.
A slide switch SW1 is used as a “self test” for the RS-232 interface. In the closed position, the J30-1 transmit signal is sent through the MAX250/MAX251 transmitter and back into the receiver portions. The signal can be read at J30-2 and verified to be correct. Any external cable connection must be removed for this self test to function. CR1 and CR2 provide transient protection for MAX251. In normal operation SW1 should be in the open (OFF) position.
The nurse call signal is input at J30-5 as a TTL logic level. In the “no alarm” st ate, this signal is a logic high, which turns on Darlington Q1, energizing relay K1. This results in contact closure between J31-1 and J31-2. In the “alarm” state, J30-5 is a logic low , which turns of f Q1, de-energizes K1 and results in cont act closure between J31-2 and J31-3. K1 provides 2500 VRMS isolation between the relay coil inputs and contact outputs.
6600-0343-000 08/27/03 1-5
Chapter 1- Functional Description
1.8 Servo Controlled Oxygen Option
The Giraffe Servo Control Oxygen System consists of an oxygen sensing circuit, Servo Oxygen circuit board, and an oxygen delivery system.
The sensing circuit is located beneath the bed and consists of a pair of fuel cell oxygen sensors, a three-way solenoid calibration valve, and a calibration fan. In normal operation the calibration valve is closed and allows the Giraffe fan to circulate gas from the infant compartment across the sensors.
The unit must be calibrated at least every 24 hours when servo oxygen is in use. Af ter 24 hours have elapsed the system prompts the user to perform calibration. Once the operator initiates calibration, the calibration valve opens and the calibration fan is turned on. This draws ambient air across the sensors until a stable reading is obtained. This 21% oxygen reference value is then used to calibrate the measuring algorithm. After calibration 100% oxygen is briefly delivered to the system to ensure there are no occlusions. When calibration is complete the unit will resume controlling oxygen based on the last set point.
The system must have two sensors present to operate. One sensor is always used for control and the other is used for a redundant check and display . The sensors generate a volt age of about 40 millivolt s at 21% oxygen concentration and about 200 millivolts at 100% oxygen concentration. The voltage is directly proportional to the concentration of oxygen. Humidity and temperature sensors located in the sensor plug are used for voltage compensation. A fan mounted to the sensor-housing door is activated when the temperature reaches 50 degrees C. This fan circulates air to keep the sensors below the maximum allowable operating tempera­ture, about 55 degrees C.
The Servo O2 board is located in the Giraffe controller enclosure. The microcontroller and integrated EPROM on the board perform the following:
Convert sensor output from analog to digital Activates oxygen alarm conditions. Two-way communications via 485 bus with the Giraf fe control board. Controls the calibration valve to select calibration mode. Controls the two supply valves to maintain the desired oxygen set point. Opens the safety relay , which removes power to the three-way valve and the supply valves in case of
a system failure. The oxygen delivery system consists of two solenoid supply valves, and a regulator assembly . The preset
regulator regulates the oxygen supply to 345 kPa (50 psi). T wo supply valves, controlled by the Servo Oxygen board, control flow to the infant compartment. Both valves are opened until the measured level gets close to the desired set point then one valve is closed. One valve is then cycled on and off as needed to maintain the desired oxygen levels in the infant compartment. The valve selected is alternated so both valves cycle about the same number of times. There are 2 fuses between the Servo O2 board and the supply valves that prevent high current from the board entering the valve housing should a short occur in the supply valves. A safety valve that shuts off oxygen flow whenever the canopy is raised actuates mechanically by the movement of the canopy support rail. When the canopy is down the valve is open (canopy up/valve closed). The safety valve actuates independently of the solenoid type supply valves.
1-6 08/27/03 6600-0343-000
Chapter 2- Service Checkout
2.1 Setup
The OmniBed is shipped with the canopy in the locked down position. Before the canopy can be raised the rail shipping locks must be released. They are located in both sides of the OmniBed near the outside bottom of the uprights. An orange tear-away label marks their location. The lock consists of a socket head cap screw in a slot. Tightening the screw disengages the screw head from the slot and releases the lif t rail. Using the 4 mm hex key provided with the unit, turn the screw clockwise about 8 rotations until the screw securely seats in its hole. Remove the tear away label.
WARNINGS w Do not perform the preoperative checkout procedure while the patient occupies the
unit.
w Complete the preoperative checkout procedure section of this manual before
putting the unit into operation. If the equipment fails any portion of the checkout procedure it must be removed from use and repaired.
2.2 Mechanical checks
1. Disconnect the power cord for the mechanical portion of the preoperative checkout procedure.
2. Examine the power cord for any signs of damage. Replace the cord if damage is evident.
3. Check that both plug retaining brackets are in place.
4. Examine the unit overall for any damaged or missing parts.
5. Check that all the casters are in firm contact with the floor and that the unit is stable. Lock the caster brakes and check that they hold the unit in place. Release the brakes and check that the unit moves smoothly.
6. Check the operation of the two side doors. Open the doors and check that they swing all the way down and hang perpendicular to the bed. Check that the doors are securely attached to the unit and that the hinge pins are properly seated. Check that the inner walls are securely fastened to the doors. Close the doors and check that the latches hold the doors securely shut. The orange latch open indicators should not be visible when the latches are engaged. Check that the top of the doors meet the canopy seal.
7. Check the portholes. Open the portholes by pressing on the latch. The cover should swing open. Close the porthole and check that the latch holds the cover securely shut and that the cover seals tightly against the porthole gasket. Check that all the porthole seals are in place and are in good condition.
8. Check that the tubing access covers in the four bed corners and the large slot grommet at the head of the bed are in place and are in good condition.
9. Check the operation of the bed. The bed should rotate easily without binding. If the bed is properly seated and locked in place, the mattress should be level. With the bed rotated back into the straight position, check to see that the bed platform extends and stops when it is pulled out on either side. Check the operation of the bed tilt mechanism. Squeeze the tilt control and push down on the foot of the bed. The head of the bed should raise easily , and should stay in position at any angle along it s tilt p ath when the tilt control is released. Push down on the head of the bed. The foot of the bed should raise easily, and should stay in position at any angle along its tilt path when you the tilt control is released.
6600-0343-000 08/27/03 2-1
Chapter 2- Service Checkout
2.3 Controller checks
WARNING w Do not use the OmniBed in the presence of flammable anesthetics: an explosion
hazard exists under these conditions.
1. Connect the OmniBed power cord to a properly rated outlet.
2. Connect the patient probe to jack 1 on the probe panel.
3. Switch on the power at the mains switch on the back of the unit, and at the standby switch on the jack panel, while holding in the override button (>37) on the control panel during power up until the software revision screen appears. Release the button and the first service screen will appear.
4. Scroll to “Down” and select it to bring up the second service screen. Select Status to see Status screen. Check status of the software self tests. These include: incubator heater on (INCHTRON), warmer heater on (WRMHTRON), incubator/warmer heater off (I/WHTROFF), humidifier heater on (RHHTR), remote monitoring data stream (RS232LOOP), incubator fan on (F ANON), and incubator fan off (FANOFF). All test should say P ASS except RS232 LOOP (the connector pins must be shorted to get the PASS mes­sage).
5. Using the standby switch turn off the unit, then turn it back on. Verify the following:
All the displays and indicators light The software revision appears The prompt tone begins
Note: If the unit has been used in the last 2 hours, the patient history query appears.
6. Adjust the set temperature to silence the prompt tone.
7. Check the patient probe. If the probe is below 30 C, the display will show -L-. Warm it by placing it between your fingers, and verify that the baby temperature reading increases.
8. Unplug the patient probe and check that both visual and audio alarms trigger in the Baby control mode.
9. Check the canopy lift mechanism. Push the lift pedal and verify the canopy raises smoothly in one continu­ous movement to its upper limit, the heater doors open, and the unit shifts into warmer operation. Check that the pedals on both sides of the unit raise the canopy .
10. With the canopy raised, check the operation of the panel at the foot of the bed. Check that after you lift up on the panel, it swings down and hangs perpendicular to the bed. Check that the panel is securely at­tached. Check that it swings back up and seats in the closed position.
1 1 . Lower the canopy and verify that it stops when you remove your finger from the hand switch. Check that it
stops automatically at its lower limit, and that the canopy seal makes contact with all four bed sides, and that the unit shifts into the incubator operating mode. Check that the raise and lower buttons at the head of the bed on both sides of the unit raise and lower the canopy .
12. If so equipped, check the operation of the bed elevating system. Raise and lower the bed along its entire travel range, checking that the mechanism operates smoothly . Check that the pedals on both sides of the unit raise and lower the bed.
13. Check the power failure alarm and the battery backed up memory. Make note of the current control mode and temperature settings and wait one minute, then unplug the OmniBed from the wall outlet. An alarm
2-2 08/27/03 6600-0343-000
Chapter 2- Service Checkout
should sound and the power failure indicator should light. Wait one to two minutes and plug the OmniBed back in. Verify that the alarm cancels and that the OmniBed returns to the same control mode and tempera­ture settings it displayed before the power interruption.
Note: A fully charged battery should supply the power failure alarm for approximately 10 minutes. If the alarm is tested for the full 10 minutes the OmniBed must be run at least two hours to recharge the battery before it is used with a patient. Total recharge time is 8 to 10 hours.
14. Perform the Leakage Current and Ground Resistance checks in Chapter 3 of this manual.
2.4 Humidity check
Turn on the Giraffe unit and verify that the Servo Humidity icon is on the screen. Set the Humidity to 65%. Wait for 4 minutes. If no alarms are seen (except for a possible “Add Water” message) the humidifier is operational.
Note: It is not necessary to have water in the reservoir to perform this test.
2.5 Servo Controlled Oxygen check
Leak Check
This test checks for leaks between the O2 sensors and the Heat sink vent fitting.
1. Remove translation deck, tilt platform, upper pan, and fan.
2. Cover the heat sink vent near the fan shaft with adhesive tape. Make sure the tape will not interfere with fan
rotation. Reinstall the fan, upper pan, tilt platform, and translation deck. Note: If the conical shaped rubber grommet was removed with the fan, when reinstalling fan be sure that rubber grommet clicks into groove on fan shaft.
3. Power up the unit, hold down the ‘Air curtain’ button and press the ‘Down’ button to force the Giraf fe into low
fan speed.
4. Set O2 set point to 21%; the display will show the actual concentration in the larger numerals next to the set
point. Open the doors until the actual concentration reaches 21% (ambient).
5. Run the Servo O2 calibration routine and wait for the calibration complete message. In approximately 20
seconds, the “Check O2 Supply” alarm should sound. If there is no alarm, the tubing between the sensor housing and the heat sink vent fitting has a leak or is disconnected. Repair the leak and repeat steps 1 through 4 of this procedure.
6. After performing the test, power off the unit. Remove the translation deck, tilt platform, upper p an, and the
fan, and remove the adhesive tape. Be sure to remove any residue on the heat sink left by the tape.
7. Reassemble the system and run the calibration routine one final time.
Pre-use Checkout
This test checks for leaks between the chassis vent fitting and the O2 sensors.
1. Connect an acceptable hose from an oxygen supply to the oxygen inlet fitting on the unit. Supply pressure
should be between 310 kPa (45 PSI) and 586 kPa (85PSI).
2. Select wrench icon on display screen to bring up setup menu. Select Cal Oxygen on the setup menu to
initiate calibration.
6600-0343-000 09/09/03 2-3
Chapter 2- Service Checkout
3. When calibration is completed, 100% oxygen is delivered for approximately 20 seconds to ensure there are no occlusions in the system. Do not turn off the unit or disconnect the oxygen supply during this brief period after the ‘Calibration Complete’ screen appears. Exit calibration screen.
4. Select O2 icon on display screen to bring up Servo Control Oxygen menu. Hold down the ‘Air curtain’ button and press the ‘Down’ button to force the Giraffe into low fan speed.
5. Set O2 set point to 65%. Start timer and verify that unit reaches 60% in less than 10 minutes. If rise time is longer than 10 minutes check all tubing between the O2 sensors and the chassis vent fitting.
Note: The chassis vent may be identified by its mushroom cap shaped cover .
Supply Valve Leak Test
1. Connect oxygen supply to Servo Oxygen fitting.
2. Disconnect the 10mm hose from the expansion chamber .
3. Power up unit in Service Mode.
4. The canopy should be closed for this step and step 5. Scroll to Servo Oxygen service screen, and open V1 & V2 and verify that gas flows audibly .
5. Close V1 & V2 and place the 10mm hose in a cup of water. Verify that no more than 10 bubbles appear over a one minute period. If unit fails, replace supply valves.
6. Raise the canopy a couple of inches and open V1 & V2. Verify that no more than 10 bubbles appear over a one minute period. If unit fails replace two-way valve or spring assembly .
When test is completed, reattach 10 mm hose to expansion chamber.
2.6 Accessory checks
1. Check that all accessories are securely mounted and out of the path of the canopy .
2. Check the operation of any accessories with reference to their appropriate operation manuals.
3. Setup any required suction or gas supply systems. Check them for leaks as described in their respective operation manuals.
2-4 08/27/03 6600-0343-000
Chapter 2- Service Checkout
2.6 Cable Connections and Mechanical Controls
Numeric Temperature Displays
Temperature Regulation Controls
Tubing Grommets
Standby Power Switch (I/O)
Drainage
Hanger Caster Brake
Graphics Screen
Control Knob
Canopy Compartment Probes
Side Door Latch
Tubing Grommets
Humidifier Reservoir (air intake filter located behind reservoir)
Accessory Power Outlets
Ventilator Slot
Side Door Latch
RS 232 Connector
Probe Jacks
Controller Cover
Mains Power Switch
FRONT
Figure 2-1
Connections and controls
6600-0343-000 08/27/03 2-5
O2 connection for Servo Controlled Oxygen option
Plug retaining brackets not shown for clarity
BACK
Power Cord Inlet
Chapter 2- Service Checkout
Portholes
Canopy
Control Panel
Upright
Side Door Latch
Drainage Hanger
Storage Drawer
Raise Canopy Pedal
SIDE
Dovetail rail
Canopy Raise/Lower Switches
Pleural Drainage Hanger
Elevating Column
Bed Height Pedal
2-6 08/27/03 6600-0343-000
Chapter 3- Calibration and Maintenance
Use Static Control W ork Station to help ensure static charges are safely conducted to ground. The velostat material is conductive; do not place electrically powered boards on it. Whenever this symbol appears beside a procedure, take static control precautions.
WARNING w Af ter performing any rep air or calibration, always perform the Service Checkout Proce-
dure before putting the unit back into service.
3.1 Maintenance schedule
The unit should be maintained in accordance with the procedures detailed in this manual. Service mainte­nance must be performed by a technically competent individual.
Service maintenance
This schedule lists the minimum frequencies. Always follow hospital and local regulations for required frequencies.
Annually
Perform the electrical safety and calibration procedure as described in the service manual. Calibrate the scale. If unit is equipped with Servo Controlled oxygen:
Replace vent screen. Perform supply valve leak test. Replace sensors*. It’s recommended both sensors be replaced at the same time.
*
Sensor life of one year is approximate. If the sensor is used often at high oxygen concentrations, sensor life will decrease.
Every T wo Years
Replace the battery . Note: The battery is used to sound the power failure alarm and to power memory circuits during a
power failure.
Every Three Y ears
Calibrate the humidifier.
3.2 Special Tools
The following tools (or their functional equivalents) are required to complete the recommended service procedures:
Digital Multimeter , 4-1/2 digit Leakage Current Tester PLCC Extractor for removing socketed chips Static Control Work Station (recommended)
Light gray touch-up paint (Munsell .16GY8.56-0.44 chroma)-18ml ........6600-0714-200
Servo Humidity Calibration Kit................................................................6600-0048-850
Scale calibration weight- 5kg .................................................................6600-0209-800
6600-0343-000 08/27/03 3-1
Chapter 3- Calibration and Maintenance
J1
J4
J2
Calibration jumper
NORMAL MODE CAL MODE
Test Point 1
Dipswitch
Test Point 2
12345678
Figure 3-1
Control board test points
3-2 08/27/03 6600-0343-000
Chapter 3- Calibration and Maintenance
3.3 System Calibration
Note: If Only performing line voltage calibration, follow instuctions in next section, 3.4. Important: Be sure to perform System Calibration after replacing a control board.
1. Turn power off.
2. Remove electrical enclosure back panel.
3. Unplug the temperature sensors and from J1, J4 and J2 on the control board.
4. Move jumper JP1 to the CAL MODE position on the control board (see Figure 3-1). Be sure to orient the jumper correctly so pins 5-6 and 7-8 are shorted.
5. Turn power on. After running the power-up testing INITIALIZING will be displayed and dots will move across the top of the screen.
6. After about 90 seconds CALIBRATION MODE, Enter “VREF”, and Enter “Mains” will be displayed.
7. Using a 4 ½ digit DVM (capable of measuring to 0.1 millivolt) measure VREF at TP1, pins 1 and 6 (pin 6 is ground) on the control board. Measure to the nearest 0.1 millivolt.
8. Dial in VREF using the control knob. Press the knob to enter the value.
9. Measure the Mains Voltage at the AC connectors at the bottom of the electronics enclosure cover .
10. Dial in the Mains voltage using the control knob. Press the knob to enter the value.
11. After a few seconds the dots will stop moving across the screen and DONE will appear on the top right of the display . Do not shut of f the unit until the DONE message is displayed or the new calibration values will not be stored.
If the message “Mains voltage calibration failed. Please enter the mains voltage again” appears this indicates the dialed in line voltage is 20% different than the measured value (not the nominal value).
12. Power down the unit and move the calibrate jumper to the NORMAL MODE position. Be sure to orient the jumper correctly so pins 1-2 and 3-4 are shorted. Reconnect J1, J4 and J2 on the control board.
3.4 Line Voltage Calibration
Important: Be sure to perform line voltage calibration after replacing a relay board.
1. Hold the overide button (>37) while powering up the unit to enter the service screen.
2. On the second service screen select CAL LV.
3. Measure the line voltage at the AC connectors located at the bottom of the electrical enclosure.
4. On the CAL LV screen, dial in the mains voltage value that you measured and push the knob to enter it.
5. When calibration is completed screen will say Mains Volt age Calibration Complete. If the message “Mains voltage calibration failed. Please enter the mains voltage again” appears this
indicates the dialed in line voltage is 20% different than the measured value (not the nominal value).
6600-0343-000 08/27/03 3-3
Chapter 3- Calibration and Maintenance
3.5 Humidifier Calibration
Important: Be sure to re-calibrate the humidifier whenever either the sensor or the control board is
replaced. Important: In order for the water in the calibration bottle (6600-0048-850) to be completely saturated, most
of the salt should not be dissolved. There should be as little standing water above the salt line as possible to minimize the response time. The salt in the calibration bottle may only be used for a period of one year after it’s initial mix with water then the kit should be discarded.
1. Take the cap off the humidity calibration bottle and add one half cap full of distilled water to the bottle. Shake the bottle to thoroughly mix the salt and water solution. Place the smaller end the elbow over the bottle.
2. Slide the elbow over the humidity sensor (mounted on the back wall) until it stops. This creates a 75% RH environment for the sensor.
3. Hold the override key while powering up to enter the service screen.
4. On the second service screen, select Cal RH. The screen will prompt “Push knob when RH reading is stable.” Wait for 20 minutes or until the RH display st abilizes (does not change by more than 1% in 5 minutes).
5. Depress the control knob to complete the calibration. On software revision 1.3 and higher , you will have the option to select ST ABLE, SET TO DEFAULT or EXIT. STABLE initiates calibration. SET T O DEFAUL T resets calibration values back to factory default settings. If you started calibration by mistake (without the calibration bottle in place, for example) you would select SET TO DEFAULT and then calibrate the unit. If you have entered the calibration routine by mistake, select EXIT to leave without initiating calibration.
6. If “RH Sensor Calibration Completed.” is displayed the calibration is complete. Depress the knob to exit the Cal RH routine.
7. If “RH Sensor Calibration Failed. Try Again.” is displayed verify your setup and press the knob to try the calibration again. This message appears if the signal from the RH sensor is outside the values expected from the sensor at 65-85% RH. If the failure persists it means the readings are out of this range and either the calibration bottle or the RH sensor may be defective.
3.6 Servo Controlled Oxygen Calibration
1. Select Set Up icon (wrench) to bring up Set Up screen.
2. Scroll down and select Cal Oxygen to initiate calibration. Calibration is automatic and takes less than five minutes. A bar graph indicates progress toward completing calibration. If for any reason you wish to discontinue calibration before it is completed, turning the control knob in either direction will cause the word Cancel to appear on the calibration screen. Pushing in the control knob will discontinue calibration.
When calibration is completed, 100% oxygen is delivered for approximately 20 seconds to ensure there are no occlusions in the system. Do not turn off the unit or disconnect the oxygen supply during this brief period after the ‘Calibration Complete’ screen appears.
Note: The servo control oxygen system prompts for calibration every 24 hours, but the system may prompt for calibration if there is a large leak in the system (for example if a door is open) for half an hour.
CAUTION w The servo-control system must be calibrated at the same atmospheric pressure in which it is
to be used. Operation at atmospheric pressures other than that present during calibration may result in readings outside the stated accuracy for the unit.
3-4 08/27/03 6600-0343-000
Chapter 3- Calibration and Maintenance
3.7 Scale Calibration
NOTE: The scale is calibrated using a Class F calibration weight between 1 kilogram and 8 kilograms
(accuracy of 0.01%).
1. Place the test weight on the center of the bed.
2. Hold the override key while powering up to enter the service screen.
3. On the second service screen, select Cal Scale.
4. Remove the weight and push the knob at the screen prompt “REMOVE THE WEIGHT AND PUSH
KNOB”. The screen will prompt “INITIALIZING......” for a few seconds.
5. Replace the weight and push the knob at the screen prompt “PLACE TEST WEIGHT AND PUSH KNOB”.
The screen will prompt “MEASURING ....” for a few seconds
6. When the screen prompts “ENTER TEST WEIGHT” Dial in the test weight to the nearest gram. Press the
knob to enter. The screen will prompt “CALCULATING.” for a few seconds.
7. When the screen prompts:
SA VE AND EXIT EXIT ONLY RESTORE DEFAULT
Select and enter “SA VE AND EXIT”
8. Turn off the power to exit the service mode.
3.8 Leakage Current
Use approved equipment and techniques to test the unit’s leakage current and ground continuity. Follow the directions supplied by the test equipment manufacturer to verify the following:
1. Less than 300 microamperes measured at any exposed metal surface for equipment rated at 120 Vac, 50/
60 Hz.
2. Less than 500 microamperes measured at any exposed metal surface for equipment rated at 220 Vac, 50/
60 Hz or 240 V ac, 50/60 Hz.
3.9 Ground Resistance Check
Measure the resistance between the ground pin on the line cord plug and exposed metal of the electronic enclosure. The ground resistance must be less than 0.2 ohms.
6600-0343-000 08/27/03 3-5
Chapter 3- Calibration and Maintenance
3-6 08/27/03 6600-0343-000
Chapter 4-Troubleshooting
4.1 Service Screen
To access the service screen, hold in the override button (>37) during power up until the software revision screen appears. Release the button and the first service screen will appear.
Figure 4-1
DAC Volt 0.000
Language English Temp U C Volume Maximum Pat Alarm 1.0C Elevate Enable Pat Ctrl Both Pat Algo Cascade Preheat 25% Canopy Enabe Scale U gms Scale R 10g Comfort Enable Set Time View Mods Down
HFS 1500 LFS 1000 RH 50 SR 1548 SC 21279
Last Cal: 1/1/02 Fri 3/3/03 9:54am
ADT 22.66 ACT 22.65 P11 327.67 P12 327.67 P21 327.67 P22 327.67 HSP 18208
LV1 117.4 LV2 117.8 LF 60 MC 0.001 TV 1.648 5V 5.059 VR 1.233 DV 0.000 BV 0.000
First service screen
Default options that may be selected from this screen appear along the left side of the screen
DAC Volt Digital/Analog Converter voltage. This is the over temperature voltage that is used by the
system to verify the computer independent circuitry is working. To manually test this circuit enter voltages from 0 to .5 V. The DV value at the bottom right corner of the screen should match this value within 10mV.
Language English is the default language that appears on the EL screen, but you can select French,
Spanish, etc., depending on what software is installed.
Temp U Changes temperature displays to show Fahrenheit, Celsius or Celsius Only so the
Fahrenheit option is not present on the user Set-up screen (Celsius is factory set default).
Volume Select one of four volume settings. 1 is minimum, 4 is maximum.
Pat Alarm Set the default Baby Hot/Baby Col alarm to activate when either 0.5ºC or 1.0ºC difference
is read between a set temperature and the baby probe temperature.
Elevate Disable or enable the elevating column foot pedal switches. If the pedals are disabled on
the service screen they cannot be enabled on the user setup screen (wrench icon).
Pat Ctrl Allows you to disable patient control.
Pat Algo To be used for future software options.
Preheat Select from 10 to 50% radiant heater power to preheat without alarms; 25% is the default.
NOTE: Reseting maximum preheat level to above 25% may result in noncompliance to
device standard IEC 601-2-21
Canopy Disable or enable the canopy foot pedal control; hand switches will remain active. If the
pedals are disabled on the service screen they cannot be enabled on the user setup screen (wrench icon).
Scale U Select from Grams, Pounds or Grams Only so that the pounds option is not present
on the user Scale screen (grams is the factory default).
6600-0343-000 08/27/03 4-1
Chapter 4- Troubleshooting
Scale R Select from 2 scale resolution settings; 10 grams or 5 grams (10 grams is the factory
default).
Comfort Allows you to remove the Comfort Screen feature.
Set Time Set real time clock for time, day and date. Choose how date is displayed (North American
or European). Choose a 12 hour (AM/PM) or 24 hour time display.
View Mods Display the current software revision of the options installed on this specific unit (Humidi-
fier, Scale, SPO2, etc.)
Down Go to second service screen.
Up Return to previous screen.
Figure 4-2
Second service screen
Up
Status Switches Errors Hours Run Cal LV Man temp al Enable Pedals BatLoad Disable Cal RH ServO2 Cal Scale
HFS 1153 LFS 887 RH 20
Last Cal: 1/1/00 Fri 3/3/02 9:54am
ADT 22.66 ACT 22.65 P11 327.67 P12 327.67 P21 327.67 P22 327.67 HSP 18208
LV1 117.4 LV2 117.8 LF 60 MC 0.001 TV 1.000 5V 5.059 VR 1.233 DV 0.002 BV 0.000
Status Check status of all self test the software runs continuously. These include: incubator
heater on (INCHTRON), warmer heater on (WRMHTRON), incubator/warmer heater off (I/WHTROFF), humidifier heater on (RHHTR), remote monitoring data stream (RS232LOOP), incubator fan on (FANON), and incubator fan off (FANOFF).
If the RS232 option is not installed RS232LOOP will display N/A. To test the circuit if the option is installed, short pins 2 & 3 on the 9 pin connector on the back of the electrical enclosure.
Up
Status
Switches Errors Hours Run Cal LV Man temp al. Enable Pedals BatLoad Disable Cal RH ServO2 Cal Scale
HFS 1500 LFS 1000 RH 20
INCHTRON PASS WRMHTRON PASS I/WHTROFF PASS RHHTR PASS RS232LOOP N/A FANON PASS FANOFF PASS
TV 1.000 5V 5.059 VR 1.233 DV 0.002 BV 0.000
Figure 4-3
Status menu
Last Cal: 1/1/00 Fri 3/3/00 9:54am
Switches Select to bring up a diagnostic diagram of the unit that displays the status of all the
switches. If the circle next to the switch is lit, the switch is closed; if its not lit, the switch is open. Also, while the switch status screen is active, you can hold down the alarm silence button to light the alarm light, system failure light and all LED segments to test them.
4-2 08/27/03 6600-0343-000
Chapter 4-Troubleshooting
Heater doors open
Canopy up
Heater doors closed
Overide control
Canopy middle
Canopy hand switch up
Canopy hand switch down
Canopy down
Elevating base pedal down
Elevating base
Figure 4-4
pedal up
Switch diagnostic diagram
Up Status
Switches
Errors Hours Run Cal LV Man temp al. Pedals BatLoad Cal RH ServO2 Cal Scale
Canopy pedal
c
37
LastCal: 1/1/02 Fri 3/3/03 10:10am
Humidifier reservoir engaged
Baby mode control
Increase control
Manual mode control
Decrease control
Air mode control
Boost air curtain control
Humidifier add water
Errors Lists a chronological log of the last 8 system errors that occurred. Shows error number
and date. It is possible to clear the list.
Hours Run Brings up a screen with 2 hour meters; one that is resetable and one that is not. To reset
the resetable hour meter, highlight CLEAR and push the control knob. The non-resetable meter will roll to 0 at 199,999.
Cal LV Use to calibrate line voltage at the factory. To calibrate line voltage follow the procedure in
calibration section (chapter 3) of this manual.
Man temp al. Use to enable or disable the patient temperature alarms in the manual mode (revision
1.40 and higher software only). If disabled on the service screen it cannot be enabled on the user setup screen (wrench icon).The alarm will not activate if a patient set point has not been entered.
Up Status Switches Errors
Figure 4-5
Hours Run Cal LV Man temp al. Enable
Pedals
BatLoad Disable Cal RH ServO2 Cal Scale
Last Cal: 1/1/00 Fri 3/3/02 9:54am
Canopy Up Off
Bed Up Off Bed Down Off EXIT
HFS 1500 LFS 1000 RH 20
TV 1.000 5V 5.059 VR 1.233 DV 0.002 BV 0.000
Pedal screen Pedals In the event of an elevating base or canopy pedal switch failure, selecting pedals allows
the canopy to be raised or the bed to be raised or lowered (revision 1.6 software and higher).
BatLoad Used for manufacturing only, not for service use.
Cal RH Brings up humidifier calibration screen, if servo humidifier is installed. See calibration
section for information on how this screen is used.
Cal Scale Brings up scale calibration screen, if scale is installed. See calibration section for informa-
tion on how this screen is used.
6600-0343-000 08/27/03 4-3
Chapter 4- Troubleshooting
A number of diagnostic readings appear on the right side of the service screens.
Figure 4-6
First service screen Diagnostics
DAC Volt 0.000
Language English Temp U C Volume Maximum Pat Alarm 1.0C Elevate Enable Pat Ctrl Both Pat Algo Cascade Preheat 25% Canopy Enabe Scale U gms Scale R 10g Comfort Enable Set Time View Mods Down
HFS 1500 LFS 1000 RH 50 SR 1548 SC 21279
Last Cal: 1/1/02 Fri 3/3/03 9:54am
ADT 22.66 ACT 22.65 P11 327.67 P12 327.67 P21 327.67 P22 327.67 HSP 18208
LV1 117.4 LV2 117.8 LF 60 MC 0.001 TV 1.648 5V 5.059 VR 1.233 DV 0.000 BV 0.000
HFS High fan speed. Should be 1500 +/- 100 (measured at power up only)
LFS Low fan speed. Should be 1000 +/- 100 (measured at power up only)
RH Relative Humidity. % humidity read in the patient chamber
SR Scale counts raw
SC Scale counts corrected
ADT Air display temperature. Temperature read by the first thermistor in the compartment probe.
Should be +/- 0.3ºC of ACT temperature.
ACT Temperature read by second thermistor in the compartment probe. Should be +/- 0.5ºC of
ADT temperature.
P11 Reading from the first thermistor in patient jack 1. Should be +/- 0.5ºC of P12 temperature.
P12 Reading from the second thermistor in patient jack 1. Should be +/- 0.5ºC of P11 temperature.
P21 Reading from the first thermistor in patient jack 2. Should be +/- 0.5ºC of P22 temperature.
P22 Reading from the second thermistor in patient jack 2. Should be +/- 0.5ºC of P21 temperature.
HSP Heat sink probe resistance. Should be 20000 ohms @ 25ºC. see section 4.5 for resistance
verses temperature values.
LV1 Line voltage in first mains circuit. Should be +/- 4V of LV2
LV2 Line voltage in second mains circuit. Should be +/- 4V of LV1
LF 60Hz or 50Hz
MC Motor current. Display number x 7 equals current drawn by the canopy lift or e-base motor
TV Thermistor voltage. Voltage of thermistor circuits located on the mother board.
5V Power supply voltage. Should be +/- 0.25V of 5V
VR Voltage reference. Independent voltage reference. Should be 1.235V +/- 1%
DV DAC output voltage. It should match the DAC volt value within 10 mV.
BV Used for manufacturing only, not for service use.
Across the bottom of the screen the date of last time the temperature and line voltage calibration was performed appears plus the current time settings of the unit.
4-4 08/27/03 6600-0343-000
Loading...
+ 124 hidden pages