USER
RESPONSIBILITY
This
contained
inserts,
accordance
checked
that
should
become
request
Service
repaired
by
altered
Department.
responsibility
faulty
other
TRACEABILITY:
this
registration
Ohmeda.
6050-0000-134.
Product
periodically.
are
be
necessary,
for
Center.
Ohmeda
without
maintenance,
than
equipment.
when
broken,
other
If
will
in
this
assembled,
with
replaced
service
than
and
by
The
for
Ohmeda.
Federal
card
additional
perform
operating
the
instructions
missing,
immediately.
Ohmeda
advice
This
in
Ohmeda
the
prior
user
any
improper
Please
included
in
operated,
A
defective
plainly
recommends
Product
accordance
trained
written
of
this
malfunction
law
fill
cards
conformity
manual
be
made
or
any
personnel.
Product
repair,
in
the
out
with
this
are
and
accompanying
maintained
provided.
Product
worn,
Should
that
to
the
of
its
with
approval
U.S.A.
the
required,
written
shall
which
damage,
self-addressed
product
with
distorted
such
the
and
This
should
repair
a
telephonic
nearest
parts
The
of
Ohmeda's
have
results
or
requires
and
order
description
labels
repaired
Product
not
be
or
contaminated
or
replacement
or
Ohmeda
should
instructions
Product
the
from
alteration
traceability
return
Stock
not
must
Safety
sole
improper
traceability
Number
thereof
and/or
in
must
used.
written
Regional
be
provided
not
by
it
to
be
Parts
be
use,
anyone
of
CAUTION:
sale
by
Federal
or
on
the
law
order
in
U.S.A.
a
of
and
licensed
Canada
medical
restricts
practitioner.
this
device
to
PRECAUTIONS
A
WARNINGS:
e
Connect
the
that
monitor.
If
the
Checkout
corrected.
A
When
moisture
Use
immerse
or
Do
not gas
Do
not
Following
(front
only
output
is
lower
See
monitor
Procedure,
caurions:
in
use,
build-up
the
recommended
autoclave
sterilization
half
quarantine
absorbed
recommendations
by
a
high
connector.
than
page
fails
Pages
See
the
sensor
monitor.
the
sterilize
the
only)
parts
the
to
material.
for
impedance
the
24,
to
do
29
its
on
cleaning
the
monitor.
of
the
allow
procedure
Improper
corresponding
respond
not use
36.
and
should
sensing
solution
Page
See
monitor.
See
with
tee
ethylene
manifold,
dissipation
Follow
and
device
loading
as
it
until
always
surface.
36.
See
Page
sterilizer
aeration
(at
least
will
value
described
the
be
pointing
sparingly;
Page
36.
oxide
or of
of
residual
10
result
displayed
in
the
malfunction
Page
See
do
36.
of
the
the
sensor
manufacturer's
period.
K
ohms)
in
an
output
by
the
Preoperative
has
down
31.
not
probe
to
saturate
housing
cartridge,
ethylene
See
Page
to
pin
oxygen
been
reduce
oxide
37.
5
of
voltage
gas
©
Only
should
Detailed
service
knowledge,
competent
trained
After
the
the
replacing
Preoperative
monitor
stabilize
After
fresh
Procedure
properly.
Page
ii
individuals
attempt
information
manual solely
by
the
in
See
to
tools
Ohmeda.
the
is
working
sensor;
batteries
Section
Page
and
Checkout
service
for
for
test
See
sensor
refer
are
3.1
46.
trained
it.
more
the
equipment,
Page
properly.
39.
cartridge
Procedure
to
installed,
to
verify
in
See
Page
extensive
convenience
and
or
in
Allow
the
NOTE
perform
that
the
39.
repairs
for
the
Section
at
on
the
repair
of
users
service
sensor
least
page
the
monitor
of
this
is
included
having
assembly,
3.1
to
five
28.
See
Preoperative
is
equipment
in
the
proper
representatives
perform
verify
minutes
Page
that
to
43.
Checkout
working
5120
Oxygen
Monitor
6
1.0
GENERAL
INFORMATION
This
illustrates
of
1.1
The 5120
throughout
nursery,
The
(LCD)
100
easy
section
operation
Description
Oxygen
or
5120
Oxygen
that
percent.
viewing.
describes
the
(1.3),
the
hospital:
wherever
indicates
The
Additional
- A switch-operated
- A digital
set
points.
Setting
tones
does
alarm
switch
the
and
not
accept
whenever
regardless
the
accessories
and
Monitor
a
measurement
Monitor
oxygen
digits
backlight
to
select
Low
02
a
continuous
low
the
of
how
general
supplied
lists
(refer
in
anesthesia,
features
concentrations
are
large
features
High
alarm
visual
alarm
displayed
far
below
features
detailed
to
Figure
of
a
LIQUID
(1.3
include:
for
viewing
02
limit
limits
02
18%
with
it
specifications
1
on
in
respiratory
oxygen
CRYSTAL
from
cm,
in
and
Low
below
indicator
18%
below
reading
the
Low
of
the
monitor
(1.2),
page
concentration
2)
digital
zero
0.5
inches,
subdued
O2
alarm
results
(vellow).
18%.
is
02
The
below
alarm
includes
(1.4).
can
therapy,
to
high)
light(*).
limit
in
The
monitor
18%
limit
(1.1),
a
be
used
is
required.
DISPLAY
for
three
monitor
will
02
is
theory
in
the
beep
set.
Setting
High
procedures.
-
Visual
concentration.
concentration
For
silenced
than
Alarm
duration
remain
The
(*)
external
General
the
O2
alarms,
and
audible
a
Low
for
Low
02,
Silence
of
active
backlight
source;
Information
High
alarms
The
is
less
O2
alarm
approximately
the
switch,
the
throughout
always
is
example,
for
02
a
feature
alarms
than
condition,
audible
alarm
alarm
to
warn
respond
or
alarm
and
condition.
any
when
on
a
limit
useful
of
greater
the
30
seconds.
will
alarm
the
Modulus
switch
during
a
low
whenever
than
audible
mav
he
remain
The
condition.
monitor
II
to
"00"
100%
or
a
the
the
selected
alarm
For
conditions
silenced
silenced
visual
is
Anesthesia
disables
oxygen
high
displayed
oxygen
can
with
for the
indicators
powered
System.
the
flush
oxygen
limits.
be
other
the
an
by
Page
3
ATTENTION
SENSOR
LIFE
EXPECTANCY
The
oxygen
time
percent
to
This
approximately
For
one
at
to
life
This
atmosphere
Do
prior
This
which
conditions.
increased
prevent
and,
Placing
refrigerator
should
DO
service
and
while
sensor
example,
year
100%
C02
of
sensor
not
to
sensor
will
thus,
NOT
sensor
oxygen
oxygen.
may
the
puncture
the
unopened
extend
FREEZE.
life
is
conditions,
the
in
service.
will
438,000
it
at
50%
also
sensor.)
STORAGE
is
packaged
to
maximize
or
placing
has
vary
relative
enhance
depending
Reduced
sensor
(Temp.
the
of
this
affected
as
sensor
typically
percent
would
oxygen
(Continuous
shorten
the
a
finite
sensors
shelf-life.
typically
or
shelf-life.
open
sensor
temperatures
humidity
from
its
shelf-life.
6 C +/- 3 C)
high-quality
by
storage
well
the
in
the
shelf-life
drying
in
as
is
exposed
function
hours.
6
months
exposure
expected
an
inert
package
in
on
storage
will
a
the
last
service.
and
help
out
for
for
General
Information
Page
5
\\
1.2
Accessories
Several
monitor;
mounting
system
A
dome
as
optional
A.
Standard
The
22
airway
If
the
provides
If
the
a
system),
to
a
convenient
Standard
refer
clamp
such
adapter
mm
circuit.
monitor
monitor
as
accessories;
Accessories:
Tee
a
convenient
use
Accessories
to
Figure
are
the
Modulus
kit
Manifold
is
to
is
to
the
universal
shelf
not
and
be
be
2.
included
a
sensor
refer
adapts
used
place
used
or
pole.
are
Note
II
Anesthesia
to
the
as
to
keep
in
mounting
supplied
that
with
cord
a
a
monitors
extension
Figure
15
mm
portable
the
"permanent"
the
sensor
sensor
clamp
with
probe
Svstem.
3
on
unit,
to
the
oxygen
holder
that
page
probe
probe
location
attach
kit
are
are
7.
the
and
built
available
to
the
probe
when
(not
the
the
into
patient
holder
not
in
built
monitor
&
universal
a
use.
into
22
mm
Tee
0212-0763-100
Manifold
©
6)
Universal
0237-2032-700
Probe
Holder*
0237-2031-800
Mounting
Clamp*
©
Figure
Page
2.
6
Standard
Items
system
Accessories
marked
installations.
with
asterisk
(*)
are
not
å
supplied
Section
with
6
|
One
V
と
1.3
Theory
of
Operation
The
Oxygen
sensor
a
voltage
pressure)
The
sensor
(chemical
current
that
at
diffuses
reaction)
(electrical)
(concentration)
gradually
The
output
consumed
temperature
sensor
The
analyzing
housing
monitor
circuitry
corresponding
to
programmed
The
block
between
diagram
the
assembly
is
its
cartridge
through
of
voltage
of
the
compensates
assembly
percent
limits
circuit
contains
proportional
detecting
is
an
a
a
base
is
proportional
the
oxygen.
(worn
from
out)
the
monitored
contains
which
oxygen
and
in
Figure
components
an
oxygen
to
surface.
electrochemical
membrane
metal
electrode.
The
by
the
sensor
gas
mixture.
for
temperature
sophisticated
converts
display.
if
required,
4
on
page
described
sensing
the
oxygen
into
to
base
chemical
cartridge
the
sensor
The
appropriate
9
concentration
device
the
cell
The
the
partial
metal
is
A
thermistor
changes,
signal
displayed
details
below.
cartridge
(galvanic
and
resultant
pressure
electrode
process.
affected
processing
signal
alarms
the
relationship
which
oxidizes
oxidation
is
by
in
the
into
a
value
are
produces
(partial
cell).
the
and
is
compared
generated.
6
SENSOR
ASSEMBLY
A/D
CONVERTER
MICROCOMPUTER
DISPLAY
SWITCH
INPUTS
(LCD)
-
produces
oxygen
-
converts
to
-
receives
operator-controlled
circuitry.
-
compares
-
updates
-
generates
concentration.
a
digital
measured
the
indicators
-
generates
converts
-
microprocessor
equivalent.
provide
-
an
microprocessor.
a
signal
the
signal
form
inputs
digital
an
audible
if
an
output
digital
operator
proportional
from
required
from
the
switches
values
Display.
alarm
alarm
signals
information
to
its
interface
the
by
oxygen
to
and
condition
for
visual
to
the
sensor
the
microprocessor.
sensor,
and the
set
limits.
visual
is
remote
from
the
alphanumeric
the
to
monitored
(analog)
the
power
alarm
sensed.
monitoring.
Page
8
Section
One
M
나
(7
—
Switch
Inputs
Display
Backlight
Suyu
Alarm
Silence
Sensor
Assembly
Figure
>
scemi,
4.
5120
Analog
Eri
тя
Foa
Calibration
Control
sm
hr.
=
O”
Sensor
Substitute
Circuit
©
b
Size
Alkaline
Oxygen
Converter
Power
ve
Monitor
Ao
Circuitry
Block
a
Miro
Diagram
Alarm
Circuitry
watchdog"
Timer
Alarm
z
~
0
ALARM
"WATCHDOG"
DISPLAY
POWER
SENSOR
General
CIRCUITRY
BACKLIGHT
CIRCUITRY
CIRCUITRY
TIMER
SUBSTITUTE
Information
-
generates
indicators
and
"Watchdog"
=
issues
failure
-
improves
light.
-
regulates
source
produces
-
concentration
circuit
an
is
LCD
(such
a
test.
audible
based
alarm
sensed.
Display
battery
as
signal
of
alarms
on
microprocessor,
circuit
condition
power
a
Modulus
that
oxygen
and
inputs.
if
visibility
or
power
II
simulates
for
the
visual
a
system
Anesthesia
alarm
operator
in
low
from
a
high
operator-initiated
(monitor)
ambient
an
external
System).
1.4
Specifications
DISPLAY
RANGE:
DISPLAY
DISPLAY
RESPONSE
DRIFT
MONITOR
OPERATING
TEMPERATURE
COMPENSATION
SENSOR
ALARM
RANGE:
RANGE:
LIFE:
SYSTEM:
TYPE:
RESOLUTION:
UPDATE:
TIME,
LINEARITY:
TEMPERATURE
TYPICAL:
RANGE:
LCD
0
to
1%
3
times
90%
in
+/-
+/-
#9
415
12
egual
(Liquid
100%
of
full
of
less
1%
1%
C
to
€
months,
to
a
total
than
over
of
110
to
50%
Crystal
Oxvgen
scale
second
change
15
an 8 hour
full
40
typical;
scale
С
ς
at
Display)
reading
in
seconds
т
(9ο
assumes
+25
oxvgen
at
period
во
r
to
G
(77.
concentration
25
104
-1045
average
E)
©
e)
F)
(77
%
E)
02
VISUAL
AUDIBLE
ALARM
Note:
SILENCE
Setting
tones
the
monitor
18%,
--
Alphanumeric
out
--
Alphanumeric
BATT
--
Continuously
indicators
determined
ーー
--
continuous
--
intermittent
--
3-beep
determined
--
For
for
--
For
for
silence
the
and
Low
02
issues
regardless
of
OK
lating
Low
30
conditions
the
Low
a
continuously
alarm
a
selected
and
LOW
(red
by
two-tone
low
low-tone,
by
02
condition,
seconds.
duration
key
is
O2
alarm
limit
LOW
02
of
the low
messages
limits:
messages
BATT.
lit
or
or
alarm
tone
low
tone,
alarm
other
of
pressed.
limit
lit
can
be
alarm
limit
indicating
HIGH
indicating
intermittently
yellow
condition.
(2
(1
Light
kHz/1
kHz),
or
a
condition.
the
audible
than
the
alarm
physically
whenever
Low
alarm
below
indicator
switch
O2
and
Emitting
kHz),
02,
condition
18%
results
set
the
setting.
oxygen
LOW
battery
flashing
alarm
the
(yellow).
to
02
level
concentration
O2.
condition:
alarm
Diodes
can
be
alarm
below
if
in
is
the
three
falls
-
silenced
silenced
alarm
beep
Although
18%,
helow
LEDs)
the
Page
10
Section
One
No