
TABLE
OF
CONTENTS
Precautions
Introduction
System
Software
JO
DESCEIPETON
1.1
General
1.2
Model
1.3
Accessories
1.4
Specifications
2.0
Theory
2.1
Power
2.2
Microcomputer:
2.3;
1€D
2.4
Backrighte
2.5
Front Panel
2.6
Sensor
2.1
D/A - A/D
2-8
Audio
2.9
Watchdog
2.1
0
MAM
ZE
1
Analog
3.0
Test
3.1
Special
3.2
Functional
3.3
Calibration
3.4
Final
(Warnings
Installation
Versions
Features
Variations
of
Operation:
Supply
Display
Input
Alarm
OULPUE
Output
and
Calibration
Tools
Test
es
Circuit
2
. . .
Switches
Circuit
Converter
and
Timer
Required
Checkout
..
and
and
。
(Modulus
ое.
>.»
e
Alarm
о
να.
оо
. o
Sensor
Cautions)
.
II
Anesthesia
us.
e e o
sii
6
es
JL
sc
со
es
ее
e
oa
ER
. . S
Indicators
ее
ов
оао
for
Calibration
Procedure
PSE
Linearity
..
. ©
>
©
e —
e
cio
a
..
с
. .
・
Check
System)
p
ロビ
1
ロロ
ロロ
<
<<<
ロロ
の
よ
の
BES
Troubleshooting,
4.1
Troubleshooting
550:
Repals
5.1
How
5.2
How
5.3
How
5.4
How
6.0
Parts
5120
Accessories:
Monitor
Control
Control
Display
Display
Display
Mist
of
Procedures
to
Remove
to
Replace
to
Replace
to
replace
List
and
Oxygen
Assembly
Board
Board
Board
Board
Board
Tiustrations,
ο ο
Procedure:
Ca
the
Back
the
the
Battery
Schematics
Monitor
Standard
Assembly
Schematic
Assembly
Schematic
Assembly
dis.
νο
Control
Display
ως
S
q
Housing
Board
Board
Terminals
.....
.
and
Po ce ee
“optional
©.
(for
. . .
(for
. . .
S-
Sa
socketed
edge-
-connected
or:
e
ら
E
<
display)
LCD
display)
1
Hi
σι σι
1
σι
|;
σι
À
ui
οι
の
の
1
の の の
SENAR
の の の
の
OJMU
a 1 の
Service
Manual
Page
iii
5

INTRODUCTION
This
manual
competent
in
proper
to
restore
If
you
Operation
monitor
The
monitor
damage
by
ESD,
System
The
carried
makes
the
When
internal
System,
connector.
contains
individual
operating
it
are not
and
used
is
contains
by
electrostatic
service
Installation
5120
Oxygen
from
it
an
Modulus
used
batteries.
the
to
meet
familiar
Maintenance
what
and
the
monitor
Monitor
place
ideal
II
Anesthesia
as
a
stand-alone
system
information
to
maintain
order
and,
published
with
Manual
routine
electronic
discharge
(Modulus
is
to
place
module
When
provides
when
the
only
a
as
for
System.
unit
used
power
that
the
specifications.
5120
will help
Ohmeda
service
Oxygen
(0178-1757-000)
maintenance
components
(ESD).
at
a
II
Anesthesia
stand-alone
To
static
required.
use
in
larger
the
monitor
in a Modulus
through
5120
is
Monitor,
can
that
quard
control
System)
unit
Its
systems
operates
II
the
a
technically
Oxygen
required,
read
learn
to
performed.
be
are
susceptible
against
workstation.
that
can
compact
such
Anesthesia
Sensor
Monitor
the
how
damage
be
size
as
from
input
the
to
To
determine
(section
installed
and
repair
3.2)
in
require
Software Versions
This
Where
with
Notable
-
-
-
manual
applicable,
earlier
The
original
masked
the
microprocessor
In
units
flashing
for
the
For
Rev.
fall
below
is
revision
differences
ROM
memory
with
indicator
duration
3
Software,
18%,
the
monitor's
can
be
performed
the
system;
that
based
on
separate
monitors
level
between
microprocessor
for
program
includes
Rev.
3
software,
(LED)
of
the
a
rather
condition,
however,
the
monitor
notes refer
monitors.
Rev.
used
EPROM
to
cease
alarm
Low
02
than
10%
while
calibration,
using
3.0
and
in
storage.
memory,
silencing
flashing
silence
alarm
as
found
functional
the
be
removed
Software
to
differences
earlier
the
5120
Beginning
period.
sounds
monitor
Oxygen
instead.
of
an
(on
whenever
in
earlier
tests
is
still
troubleshooting,
from
Version
the
3.0.
encountered
Software
Monitor
with
alarm,
Rev.
causes
continuously)
the
revision
system.
revisions
includes
3.0
any
oxygen
software.
|
|
are:
software
level
Service
Manual
Page
v

To
compensate
calibrates
into
(linearity)
(greater
All
that
The
alarm
The
is
Checkout
troubleshooting.
use.
displays
is
alarm
silence
monitor
used
The
than
part
indicator
to
Procedure;
for
reduced
the
monitor
accuracy
can
be
checked
99%
is
specified).
are
produced
of
the
Display
LEDs
switch
includes
verify proper
and
a
section
output
to
room
of
the
using
on a single
board.
(light
must
self-test
operation
3.2)
of
the
air
(21%)
displayed
a
single
emitting
be
replaced
feature
of
and can
sensor
whenever
readings
concentration
liquid
diodes)
as
(System
the
monitor
be
as
crystal
a
unit.
used
it
ages,
the
monitor
over
the
display
are
part
Test
as
Mode)
(Functional
an
of
aid
the
full
oxygen
of
which
in
operator
is
range
(LCD)
the
put
Page
1-2
Section
1.0
1ο

output
output
input
input
output
output
output
output
input
input
input
input
output
input
output
output
output
output
output
output
output
|
65
cé
CT
output
output
output
D/A
bit
D/A
bit
Sensor
Battery
Analog
Analog
Analog
Analog
Bit
0
of
of
1
Bit
of
2
Bit
of
Bit
Yellow
3
External
Watchdog
Alarm
D/A
D/A
D/A
D/A
or
D/A
or
D/A
or
D/A
or
D/A
or
output,
bit
bit
bit
bit
selects
bit
selects
bit
selects
bit
selects
bit
selects
8
9
voltage
voltage
switch,
switch,
switch,
switch,
BCD
BCD
BCD
BCD
LED;
Alarm
Timer
つの
3,
ων
input
4,
LSD
5,
MSD
6,
LSD
7,
MSD
comparator
comparator
connects
grounds
connects
to
sample/hold
switch
switch
switch
switch
on,
is
1
Silence,
output
1
=
switches
of
of
of
of
sensor
input
reference
amplifier
inputs
inputs
inputs
inputs
is
0
rising
alarm
low
low
high
high
condition
limit
limit
limit
limit
output
output
to
off
$201,
to
amplifier
amplifier
voltage
edge
exists
S202
tricaered
「
output
output
output
output
output
input
output
output
output
Figure
Theory
of
Input/Output
7.
Operation
Clock
Data
Switch
Strobe,
Switch
External
Red
Oscillator
Tone;
for
to
for
for
LED;
0
Ports.
display
display
comparators;
latches
backlight;
supply
1
is
on/off
152
on,
kHz,
shift
shift
data
sense
0
control;
Ii
register
register
0
is
into
is
às 1 kHz
0
shift
is
off
on,
off,
0
is
1
is
register
1
is
on
off,
1
off
is
outputs
on
Page
2-5

ce
+5
E
Autos
100K
B3
,
ANIL
uloz
P
F
|
>
BI
ao
as
400K
5
400K
4001
100K
RN401
RMOt
i
1
1 1 1 ㅣ
т
1 に
1
I
1 1 1 + 1
1
i
|
i
À
1
A |
LO
02
MSD
STROBE
To
ie
し
で
1
ne
απ
v
ili
站
1002
LINES
D/A
CONVERTER
に
<
in
1 1
Teens
T
hel
Ea
1
C
i |
LSD
1
te!
し
<
T
1
nr
lest
Jů
Oe
lei
at
1
el
156
HIGHOZ
т
MSD
GR402
a
i
1
απ
må.
πα,
T
leet
1,
jar
i
ns
ii
1
r
μη
+
ЕЯ
HIGHO2
LSD
CRIOU
к
rol
5
cam
5202
ALARM
cR10
BATTERY
TEST
ser
CIRCUIT
TEST
68109
SILENCE
Figure
2.5
Front
Operator
(for
Circuit
BCD-encoded
The two
pushbutton
The
switches
microprocessor
one
row
diodes
The
switch
Note
and
the
function
or
the
10.
inputs
High
Test)
test
of
is
that
Digital-to-Analog
is
digitial/analog
Front
Panel
Ο2
and
and
thumbwheel
switches
switches.
are
switches
are
closed.
the
output
activated
Panel
Switches
to
the
oxygen
Low
02
an
alarm
switches,
and
connected
output
used
lines
and
eliminate
to
lines
at
converter
L
Switches
monitor consist
alarm),
silence
alarm
in a matrix
(C3
inputs
(C2
converter.
time,
a
am
munewmet
two
each
silence
to
their
false
to
either
is
test
switch.
having
C7).
status
C7)
This
being
sunrcu
switches
switch
and
The
readings
are
shared
works
the
used.
αἱ
of
two
(Batt
The
two
limit
8
output
are
momentary
scanned
microprocessor
(BO
to
R3).
when
between
because
switches
limit
Test
switches
connections.
(strobed)
more
are
than
the
only
being
switches
and
bv
the
selects
one
switches
one
scanned
are
Theory
Operation
of
Page
2-9

2.7
D/A - A/D
Converter
2
The
digital-to-analog
performs
described
converter
Integrated
a
current
input
a
voltage
1.138
The
The
The
reference
reference
reference
output
two
in
as
which
terminals.
by
volts.
diode
comparators
sensor
The
amplifier
outputs
microprocessor
The
microprocessor
to
convert
10-bit
digital
functions:
section
described
Circuit
U104
is
proportional
The
amplifier
voltage
(VR102)
voltage
of
the
D/A
(U107).
from
One
and
each
(U102).
uses
the
battery voltage
number.
converter
it
2.11
and
here.
is
output
U105.
(U104-15)
and
(Vref)
converter
comparator
the
other
comparator
a
"successive
(U104
generates
is
a
10-bit
to
current
The
output
is
an
adjustment
is
set
(U105-7)
comparator
are
or
the
part
CMOS
the
of
generated
to
measures
connected
the
and
its
associated
analog
of
the
analog-to-digital
D/A
converter
digital
the
D/A
voltage
by
resistor
-1.138
is
volts
connected
the
measures
approximation
sensor
amplifier
output
value
unit
range
a
fixed
voltage
to
inputs
voltage
that
present
is
converted
is
voltage
(R138).
(section
to
from
battery
conversion
voltage
circuitry)
as
produces
at
from
two
0
3.3).
voltage
the
voltage.
on
the
routine"
to
its
to
to
a
©
Resistors
outputs
To
minimize
powered
is
turned
and
turned
Auto-Zero
The
"auto-zero
voltage
sensor
(using
sensor output
output,
determine
into
the
the
offset
R118
to
when
on
Logic
of
voltage.
U109,
and
the
inverting
help
power
the
when
off
when
the
switch
to
finally,
sensor
is
always
and
R121
eliminate
drain
conversion
the
completed.
logic"
sensor
The
microprocessor
A)
the
input
provide
oscillations.
on
the
analog
is
a
software
amplifier
and
measures
amplifier
subtracts
voltage.
of
positive.
a
batteries
routine
to
digital
(U111)
(using
the
A
negative
the
amplifier
slight
is
routine
grounds
the
offset
hysteresis
the
in
software
and
offset
U109,
offset
comparators
effect.
that
subtracts
the
input
voltage,
switch
from
via
the
R131
to
Transistor
routine
measures
it
of
B)
output
voltage
to
the
are
is
the
from
the
connects
and
measures
voltage
is
ensure
comparator
only
Q106
executed,
offset
the
amplifier
the
the
to
injected
that
Theory
of
Operation
Page
2-13

FROM
WATCHDOG
TIMER
ALARM
OUT
Mode
3103-3
"a
10K
uloz
PP
87
ALARM
EXT
|
Figure
2.10
When
watchdog
about
static
Refer
When
When
pulses
conditions
sent.
low.
OUTPUT
Alarm
an
1
discharge
to
no
an
corresponding
When
This
when
15.
Output
alarm
timer,
mA
of
Figure
alarm
alarm
are
pulses
allows
SILENCE
Jr03-4
individual
Alarm
condition
the
drive
protection,
16,
condition
condition
prioritized
are
a
+5
ne
=
Output
ALARM
current,
Figure
to
not
simple
alarm
태조.
WOK
is
detected
OUTPUT
17,
exists,
occurs,
the
alarm
so
being
alarm
+5
CRile
имам.
crus
14914
line
while
and
ALARM
ALARM
condition
that
transmitted,
indicator
decoding
by
is
CR119,
Figure
OUTPUT
OUTPUT
only
is
the
microprocessor,
pulled
CR120,
18.
is
emits
which
the
highest
ALARM
to
be
not
necessary.
"|:
low.
high
connected
and
a
exists.
priority
OUTPUT
U108
R123
(>
series
or
provides
provide
4.5
Alarm
remains
to
by
volts).
of
10
code
ALARM
the
ms
is
logic
External
An
input
connected
alarm
The
maintain
A
allowed.
Page
input
maximum
2-16
just
Alarm
is
provided
between
as
is
an
alarm
input
Silence:
pin
the
"alarm
high-to-low
silence.
voltage
for
4
remote
and
pin
silence"
sensitive,
The
of
5.5
alarm
3,
switch
input
volts
silencing.
when
closed,
on
the
a
continuous
is a transient-protected
and
a
minimum
A
normally
will
front
low
of
silence
of
the
input
-0.5
open
the
monitor.
will
CMOS
volts
Section
switch
audible
not
input.
are
2.0

3.4
Final
Test
and
Sensor
Linearity
Check
After
monitor's
concentration
required
NOTE:
the
1.
Set
HIGH
2.
Turn
(see
3.
Introduce
and
Adjust
Remove
oe
The
21%
sensor
6.
To
(or
NOTE:
enable
monitor
readings
to
verify
the
High
02
alarm.
the
note
allow
the
the
display
+/-
1
If
the
cartridge
to a desired
has
of
oxygen
02
monitor
below).
the
the
CAL
probe
must
digit.
display
the
HIGH
been
throughout
greater
the
linearity
alarm
On
and
oxygen
display
Control
is
sensor
from
stabilize
is
not
linear
O2
High
calibrated,
its
measurement
than
of
limit
to
until
the
not
alarm,
O2
switch
allow
into
stabilize
the
"100%"
within
within
and
set
alarm
verify
99%
the
it
to
a
display
02
three
the
must
the
limit).
the
range
("100%").
oxygen
to
stabilize
"100%"
(approximately
and
+/-
be
High
sensor.
"00"
to
concentration
indicates
expose
minutes
1%
replaced
O2
tolerance
accuracy
using
This
disable
for
five
two
it
to
and
(see
alarm
of
a
test
99% 02.
limit
is
the
minutes.
of
minutes).
room
indicate
range
note
the
oxygen
air.
the
below).
to
"30"
Before
and
minutes.
the
monitor
test.
installed,
cartridge
just
If
five
removed
allowed
disconnected
of
layer
(if
of
If in
time
is
linear.
additional
and
testing
allow
sensor
In
before
the
sensor
minutes
12
hours.
that
disconnected
the
cell.
doubt,
of
successful,
repeat
the
the
sensor
The
five
(cartridge)
for
at
least
cases
was
from
to
disconnect
However,
where
the
five
removed
testing.
assembly
or if
its
stabilize
(or
The
accumulates
allow
the
time
the
test.
monitor,
to
stabilize
minute
assembly
24
hours
a
new
minute
from
has
the
new
protective
for
removed
stabilization
on
and
exposed
the
monitor
and
perform
sensor
if
the
test
for
the
monitor
turn
stabilization
sensor
stabilization
its
been
sensor
packaging,
as
many
from
the
can
the
for
has
immediately
cartridge
sealed
disconnected
cartridge
hours
its
packaging),
is
required
galvanic
to
oxygen)
to
stabilize
this
be
assumed
is
not
to
monitor
at
least
time
been
test.
successful,
stabilize
connected
assumes
protective
the
as
cell
impeding
to be
on
five
assumes
preceding
has
been
for
has
been
monitor
the
sensor
up to
to
remove
of
the
for
the
If
the
allow
(up
that
to
the
this
that
packaging
more
the
test
to
the
than
previously
must
has
a
maximum
an
sensor
output
presumed
24
hours)
be
been
oxide
Page
3-10
Section
3.0
Qu

6.0
PARTS
For
External
For
Accessories
For
Monitor
For
Control
For
Display
For
Display
LIST
Components
Assembly
Board
Board
Board
AND
,
SCHEMATICS
...
.
.
. . :
with
with
socketed
edge-connected
LCD
display
LCD
dis
play
see
see
see
see
see
see
Figure
Figure
Figure
Figure
Figure
Figure
27
28
29
30
32
34
Item
pero»
Ohmeda
Description
Sensor
Assembly
(does
not
Sensor
Batteries
Front
O-ring
O-ring
include
Cartridge
(3),
Housing
(outside)
(inside)
5120
en
Monitor
Item
size
C,
with
2)
alkaline
O-rings
Stock
No.
0237-2030-700
0237-2034-700
Obtain
locally
0237-2035-800
0210-0503-300
0210-0499-300
and
5120
Schematics
Figure
Parts
27.
List
Oxygen
Monitor
Page
6-1
_.

LIST
OF
ILLUSTRATIONS
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Figure
Ohmeda
Standard
wi
Optional
5120
UA
Power
Microcomputer
JO
Input/Output
LCD
DO
Backlight
+
o
Front Panel
II
Sensor
125
Digital-to-Analog
Audio
13.
14.
Watchdog:
so
Alarm
16.
Remote
17.
Alarm
18.
Alarm
19.
Analog
20.
Tools
21
Test
22.
Battery
How
23.
How
24.
25.
How
26.
How
27.
5120
Accessories:
28.
29%
Monitor
Control
30.
Control
31.
Display
Display
33%
Display
34.
5120
Oxygen
Accessories
Accessories
Oxygen
Supply
Monitor,
Circuit
Si
Ports.
Display
epi
еек
Switches
Input
Alarm
Output
Alarm
Output
and
Timer’
Circuit
Converter
Alarm
«2
не
Pulse
Waveforms
Interface Circuitry
Output
and
Points
to
Remove
to
Replace
to
Replace
to
Replace
Oxygen
Test
and
Test
Monitor
Assembly
Board
Board
Board
Board
Board
Eo
Eguipment
Calibration
Verification
the
Back
the
the
the
Standard
.....
Assembly
Schematic
Assembly
Schematic
Assembly
Monitor:
. . .
.
. .
Block
.
.
<
500
. . .
. .
Indicators
eso
la
Code
.
ek
o в
Housing
Control
Display
Control
..
and
.
(for
(for
Diagram
o
do
。
.
.
ODO
on
tro
Board
Board
Board
Optional
.
・
socketed.
5
eage-connected
à
πο.
LCD
display)
LCD.
displ
ay
|
ロビ
ピロ
ee
ロロ
いら
1
1
ららら
Io
3=2
の の
の
ロロ
の
の
の の
の
の の
の
の の
나
da
UN
00
N
ль
©
NH
-
UEL
NH
UE
IO
O
Parts
List
and
Schematics
Page
6-9
/

Ohmeda
355
Invemess
Englewood
303 799
Rapidfax
Telex
4947238
A
Division
Regional
Eastern
Fort
Lee
One
Executive
Fort
Lee
800 922
201
947
Midwestern
2101 S Arlington
Suite
145
Arlington
312 364
800 372
Pacific
5635 W Las
Suite
406
Pleasanton
800 227
800 772
Southeastern
4565
Winters
Atlanta
800
241
404
449
Drive
CO
8333
303 799
of
The
Service
Region
Executive
NJ
07024
0443
Outside
6100
In
Region
Heights
7090
5867
Region
Positas
CA
2054
Outside
3517
In
Region
Chapel
GA
30360
4300
Outside
7990
In
80112
6614
BOC
Drive
NJ
Heights
IL
94577
CA
GA
South
5810
Group
Offices
Park
NJ
Rd
60005
Blvd
CA
Rd
GA
Inc
International
Europe,
Elizabeth
Harlow
Essex
England
Telex
Latin
10700 N Kendall
Suite
Miami
Telex
Southeast
#06
5611 N Bridge
Singapore
Telex
Middle
Elizabeth
Harlow
Essex
England
Telex
Africa
Way
CM19
817566
America,
301
FL
33176
810
8486057
Asia
01
Eng
Cheong
0719
RS35276
East,
Nigeria
Way
CM19
817566
5AB
Caribbean
Dr
Road
5AB
Tower
National
Ohmeda
4507 C Mills
Atlanta
800
404
Canada
Service
172
Rexdale
416 243
Telex
Service
Place
GA
241
691
30336
6442
3635
Centre
Belfield
Road
Ontario
9533
06989362
Center
M9W
1H1
Southwestern
831
Greenview
Grand
Prairie
800
433
800 772
BOC
5070
5420
Health
Region
Drive
TX
75050
Outside
In
TX
Care
TX
Bk
Cvr
12
87
0178
04 87 B 01
Printed
©
The
BOC
Chg
0164
in
000
0102
USA
Group
Inc