Ohio Medical MiniOX 200, MiniOX 200E User guide

Page 1
®
®
®
®
MiniOX 200
BAT
CAL
®
MiniOX 200
MINIOX® 200
MiniOX® 200E
®
MiniOX 200 / 200E
OPERATING MANUAL &
INSTRUCTIONS FOR USE
600002 Rev. 1
Page 2
Page 3
TABLE OF CONTENTS
WARNINGS, CAUTION
1.0 SYSTEM OVERVIEW ..........................................1
1.1 Base Unit Description ............................................1
1.2 Symbol Guide ..................................................1
1.3 Component Identification .........................................2
1.4 Component Description...........................................2
1.5 MiniOX Oxygen Sensor ...........................................3
2.0 OPERATING INSTRUCTIONS ...................................3
2.1 Getting Started .................................................3
2.1.1 Protect Tape .................................................3
2.1.2 Automatic Calibration...........................................3
2.2 Calibrating the MiniOX 200/200E Oxygen Analyzer ......................4
2.2.1 In-line Calibration (Flow Diverter - Tee Adaptor) .......................4
2.2.2 Direct Flow Calibration (Barb) ....................................5
3.0 FACTORS INFLUENCING ACCURATE READINGS .....................5
3.1 Elevation Changes...............................................5
3.2 Temperature Effects .............................................5
3.3 Pressure Effects ................................................6
3.4 Humidity Effects ................................................6
4.0 CALIBRATION ERRORS AND ERROR CODES ........................6
5.0 CHANGING THE BATTERIES....................................7
6.0 CHANGING THE OXYGEN SENSOR ...............................8
6.1 MiniOX 200 Model ..............................................8
6.2 MiniOX 200E Model .............................................9
7.0 CLEANING AND MAINTENANCE .................................9
8.0 SPECIFICATIONS...........................................10
8.1 Base Unit Specifications .........................................10
8.2 Sensor Specifications ...........................................10
9.0 MiniOX 200/200E SPARE PARTS AND ACCESSORIES ................10
9.1 Included with your Unit..........................................10
9.2 Standard Replacement Parts......................................11
9.3 Mounting Options ..............................................11
CLASSIFICATION, WARRANTY ....................................12
This manual describes the function, operation and maintenance of the Ohio Medical Model MiniOX 200/200E oxygen analyzer. The MiniOX 200/200E Oxygen Analyzer utilizes the Ohio Medical MiniOX-250 oxygen sensor and is engineered for fast response, maximum reliability and stable preformance. The MiniOX 200/200E is designed as a tool for use by qualifield personnel to spot­check or messure oxygen concentration of delivered air/oxygen mixtures. The MiniOX 200/200E
Analyzers are not intended for use in continuous monitoring of oxygen delivery to a patient.
600002 Rev.1
Page 4
WARNINGS
!
Indicates a potentially hazardous situation, if not avoided, could result in death or serious injury.
»
Never install the sensor in a location that will expose the sensor to patient’s
exhaled breath or secretions, unless you intend to dispose of the sensor, flow diverter and tee adapter.
» Improper use of this device can cause inaccurate oxygen readings which can lead to
improper treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user manual.
»
Not for use in an MRI environment.
»
Device specified for dry gas only.
»
Never allow an excess length of tubing, lanyard or sensor cable near the patient’s head or
neck, which may result in strangulation. »
Before use, all individuals who will be using the MiniOX 200/200E must become thoroughly
familiar with the information contained in this Operation Manual. Strict adherence to the operating instructions is necessary for safe, effective product performance. This product will perform only as designed if installed and operated in accordance with the manufacturer’s operating instructions.
»
Use only genuine Ohio Medical accessories and replacement parts. Failure to do so may
seriously impair the analyzer’s performance. Repair of this equipment must be performed by a qualified service technician experienced in repair of portable hand held medical equipment.
»
Calibrate the MiniOX 200/200E weekly when in operation, or if environmental conditions change
significantly. (ie, Elevation, Temperature, Pressure, Humidity — refer to Section 3.0 of this manual).
»
Use of the MiniOX 200/200E near devices that generate electrical fields may cause erratic readings.
»
If the MiniOX 200/200E is ever exposed to liquids (from spills or immersion) or to any other
physical abuse, turn the instrument OFF and then ON. This will allow the unit to go through its self test to assure everything is operating correctly.
»
Never autoclave, immerse or expose the MiniOX 200/200E (including sensor) to high temperatures
(>70°C). Never expose the device to pressure, irradiation vacuum, steam, or chemicals.
»
This device does not contain automatic barometric pressure compensation.
»
Although the sensor of this device has been tested with various anesthesia gases including nitrous
oxide, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane and found to have acceptably low interference, the device in entirety (including electronics) is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. Only the threaded sensor face, flow diverter, and “T” adapter may be allowed to contact such a gas mixture.
»
NOT for use with inhalation agents. Operating the device in flammable or explosive atmospheres
may result in fire or explosion.
CAUTION: Indicates a potentially hazardous situation, if not avoided, could result in minor or moderate injury and property damage.
»
Federal Law (USA) restricts this device to sale by or on the order of a physician.
»
Replace the batteries with recongnized high quality AA Alkaline or Lithium batteries.
DO NOT use rechargeable batteries. »
If the unit is going to be stored (not in use for 1 month), we recommend that you remove the
batteries to protect the unit from potential battery leakage.
»
The Ohio Medical MiniOX oxygen sensor is a sealed device containing a mild acid electrolyte,
lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste constituents and should be disposed of properly, or returned to Ohio Medical for proper disposal or recovery.
DO NOT use ethylene oxide sterilization. DO NOT immerse the sensor in any cleaning solution, autoclave or expose the sensor to high temperatures.
»
Dropping sensor can adversely affect its performance.
»
The device will assume a percent oxygen concentration when calibrating. Be sure to apply
100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
NOTE: This product is latex free.
600002 Rev.1
Page 5
1.0 SYSTEM OVERVIEW
BAT
!
1.1 Base Unit Description
The MiniOX 200/200E analyzer provides unparalleled performance and reliability due to an advanced design that includes the following features and operational benefits.
»
Extra-life oxygen sensor of approximately 1,500,000 O2 percent hours (2-year warranty). »
Durable, compact design that permits comfortable, hand-held operation and easy to clean. »
Operation using only two AA Alkaline batteries (2 x 1.5 Volts) for
approximately 5000 hours of performance with continuous use. For extra extended long life, two AA Lithium batteries may be used.
»
Oxygen-specific, galvanic sensor that achieves 90% of final value in approximately 15 seconds at room temperature.
»
Large, easy-to-read, 3 1/2-digit LCD display for readings in the 0-100% range. »
Simple operation and easy one-key calibration. »
Self-diagnostic check of analog and microprocessor circuitry.
»
Low battery indication.
»
Calibration reminder timer that alerts the operator, using a calibration icon on the LCD display, to perform a unit calibration.
1.2 Symbol Guide
The following symbols and safety labels are found on the MiniOX 200/200E:
Follow instructions for use.
Warning
Meets ETL standards
Do not throw away. Follow local guidelines for disposal.
CAL
Calibration required
Federal Law (USA) restricts this device
only
to sale by or on the order of a physician
Do Not
Contains acid
IPx1
EC REP
On/off Button
Calibration Button
Low Battery
Percent
%
Manufacturer
Serial Number
SN
Lot code/Batch code
LOT
Drip Proof
Catalog Number
REF
Authorized Representative in the European Community
1600002 Rev.1
Page 6
1.3 Component Identification
CAL
MiniOX 200
®
®
LCD Display
Keypad
Sample Inlet Connection
FIGURE 1
1.4 Component Description
LCD Display
3 1/2-Digit Display - The 3 1/2 digit liquid crystal display (LCD) provides
direct readout of oxygen concentrations in the range of 0 – 105.0% (100.1% to 105.0% used for calibration determination purposes). The digits also display error codes and calibration codes as necessary.
Low Battery Indicator - The low battery indicator is located at the top of the display and is only activated when the voltage on the batteries is below a normal operating level.
“%” symbol - The “%” sign is located to the right of the concentration number and is present during normal operation.
Calibration symbol - The calibration symbol is located at the bottom of the display and is timed to activate when a calibration is necessary.
Keypad
ON/OFF Key - This key is used to turn the device on or off.
Calibration Key - This key is used to calibrate the device. Holding the key
for more than three seconds will force the device to enter a calibration mode.
Sample Inlet Connection - This is the port at which the device is connected to determine oxygen concentration.
2 600002 Rev.1
Page 7
1.5 MiniOX 250 Oxygen Sensor
®
MiniOX 200
®
AB
FIGURE 2
The MiniOX oxygen sensor offers stability and extra life. The MiniOX 250 is a galvanic, partial pressure sensor that is specific to oxygen. It consists of two electrodes (a cathode and an anode), a teflon membrane and an electrolyte. Oxygen diffuses through the teflon membrane and immediately reacts at a gold cathode. Concurrently, oxidation occurs electrochemically at the lead anode, generating an electrical current and providing a voltage output. Electrodes are immersed in a unique gelled weak acid electrolyte which is responsible for the sensors long life and motion insensitive characteristis. Since the sensor is specific to oxygen, the current generated is proportional to the amount of oxygen present in the sample gas. When no oxygen is present, there is no electrochemical reaction and therefore, negligible current is produced. In this sense, the sensor is self-zeroing.
2.0 OPERATING INSTRUCTIONS
2.1 Getting Started
2.1.1 Protect Tape
Prior to turning on the unit, a protective film covering the threaded sensor face must be removed. After removing the film, wait approximately 20 minutes for the sensor to reach equilibrium.
2.1.2 Automatic Calibration
After the unit is turned on it will automatically calibrate to room air. The display should be stable and reading 20.9%.
CAUTION: The device will assume a percent oxygen concentration when calibrating. Be sure to supply 100% oxygen, or ambient air concentration to the device during calibration or the device will not calibrate correctly.
To check the oxygen concentration of a sample gas: (after the unit has been calibrated)
1. Connect the tubing to the bottom of the analyzer by threading the barbed adapter onto the oxygen sensor. (FIGURE 2, B)
2. Attach the other end of the sample hose to the sample gas source and initiate flow of the sample to the unit at a rate of 1-10 liters per minute (2 liters per minute is recommended).
3. Using the “ON/OFF” key, make sure the unit is in the power “ON” mode.
4. Allow the oxygen reading to stabilize. This will normally take about 30 seconds or more.
3600002 Rev.1
Page 8
2.2 Calibrating the MiniOX 200/200E Oxygen Analyzer
NOTE: We recommend use of medical grade USP or >99% purity oxygen when
calibrating the MiniOX 200/200E.
The MiniOX 200/200E Analyzer should be calibrated upon initial power-up. Thereafter, Ohio Medical recommends calibration on a weekly basis. To serve as a reminder, a one week timer is started with each new calibration. At the end of one week a reminder icon “
CAL
” will appear on the bottom of the LCD. Calibration is recommended if the user is unsure when the last calibration procedure was performed, or if the measurement value is in question. Start calibration by pressing the key for more than 3 seconds. The MiniOX 200/200E will automatically detect if you are calibrating with 100% oxygen or 20.9% oxygen (Normal air).
DO NOT attempt to calibrate to any other concentration.
For hospital and home care a new calibration is required when:
» The measured O2 percentage in 100% O2 is below 97.0% O » The measured O2 percentage in 100% O2 is above 103.0% O » The CAL reminder Icon is blinking at the bottom of the LCD. » If you are unsure about the displayed O2 percentage. (See Factors influencing accurate readings.)
2.
2.
For ID testing, (or optimum accuracy) a new calibration is required when:
» The measured O2 percentage in 100% O2 is below 99.0% O » The measured O2 percentage in 100% O2 is above 101.0% O » The CAL reminder Icon is blinking at the bottom of the LCD. » If you are unsure about the displayed O2 percentage. (See Factors influencing accurate readings.)
2.
2.
A simple calibration may be made with the sensor open to static ambient air. For optimum accuracy Ohio Medical recommends that the Sensor be placed in a closed loop circuit where gas flow is moving across the sensor in a controlled manner. Calibrate with the same type of circuit and flow that you will use in taking your readings.
2.2.1 In Line Calibration (Flow Diverter – Tee Adapter)
1. Attach the diverter to the MiniOX 200/200E by threading it on to the bottom of the analyzer or sensor.
2. Insert the flow diverter into the center position of the tee adapter (FIGURE 2, A).
3. Attach an open-ended reservoir to the end of the tee adapter. Then start the calibration flow of oxygen at two liters per minute.
» Six to 10 inches of corrugated tubing works well as a reservoir. A calibration oxygen flow to the MiniOX 200/200E of two liters per minute is recommended to minimize the possibility of obtaining a “false” calibration value.
4. Allow the oxygen to saturate the sensor. Although a stable value is usually observed within 30 seconds, allow at least two minutes to ensure that the sensor is completely saturated with the calibration gas.
5. If the MiniOX 200/200E is not already turned on, do so now by pressing the analyzer “ON” button.
4 600002 Rev.1
Page 9
6. Press the Cal button on the MiniOX 200/200E until you read the word CAL on the analyzer display.This can take approximately 3 seconds. The
analyzer will now look for a stable sensor signal and a good reading. When obtained, the analyzer will display the calibration gas on the LCD.
NOTE: Analyzer will read “Cal Err St” if the sample gas has not stabilized.
2.2.2 Direct Flow Calibration (Barb)
1. Attach the Barbed Adapter to the MiniOX 200/200E by threading it on to the bottom of the analyzer or sensor.
2. Connect the tube to the barbed adapter. (FIGURE 2, B)
3. Attach the other end of the clear sampling tube to a source of oxygen with a known oxygen concentration value. Initiate flow of the calibration gas to the unit. Two liters per minute is recommended.
4. Allow the oxygen to saturate the sensor. Although a stable value is usually observed within 30 seconds, allow at least two minutes to ensure that the sensor is completely saturated with the calibration gas.
5. If the MiniOX 200/200E is not already turned on, do so now by pressing the analyzer “ON” button.
6. Press the Cal button on the MiniOX 200/200E until you read the word CAL on the analyzer display. This can take approximately 3 seconds. The analyzer will now look for a stable sensor signal and a good reading. When obtained, the analyzer will display the calibration gas on the LCD.
3.0 FACTORS INFLUENCING ACCURATE READINGS
3.1 Elevation/Pressure Changes
» Changes in elevation result in a reading error of approximately 1% of reading per 250 feet. » In general, calibration of the instrument should be performed when elevation at which
the product is being used changes by more than 500 feet.
» This device does not automatically compensate for changes in barometric pressure
or altitude. If the device is moved to a location of a different altitude, it must be recalibrated before use.
3.2 Temperature Effects
The MiniOX 200/200E will hold calibration and read correctly within ±3% when at thermal equilibrium within the operating temperature range. The device must be thermally stable when calibrated and allowed to thermally stabilize after experiencing temperature changes before readings are accurate. For these reasons, the following is recommended:
» For best results, perform the calibration procedure at a temperature close to the
temperature where analysis will occur.
» Allow adequate time for the sensor to equilibrate to a new ambient temperature.
CAUTION: “CAL Err St” may result from a sensor that has not reached thermal equilibrium. »
When used in a breathing circuit, place the sensor upstream of the heater.
5600002 Rev.1
Page 10
3.3 Pressure Effects
Readings from the MiniOX 200/200E are proportional to the partial pressure of oxygen. The partial pressure is equal to the concentration times the absolute pressure.Thus, the readings are proportional to the concentration if the pressure is held constant. Therefore, the following are recommended:
»
Calibrate the MiniOX 200/200E at the same pressure as the sample gas.
»
If sample gases flow through tubing, use the same apparatus and flow rates when
calibrating as when measuring.
3.4 Humidity Effects
Humidity (non-condensing) has no effect on the performance of the MiniOX 200/200E other than diluting the gas, as long as there is no condensation. Depending on the humidity, the gas may be diluted by as much as 4%, which proportionally reduces the oxygen concentration. The device responds to the actual oxygen concentration rather than the dry concentration. Environments where condensation may occur are to be avoided since moisture may obstruct passage of gas to the sensing surface, resulting in erroneous readings and slower response time. For this reason, the following is recommended:
» Avoid usage in environments greater than 95% relative humidity. » When used in a breathing circuit, place the sensor upstream of the humidifier.
HELPFUL HINT: Dry sensor by lightly shaking moisture out, or flow a dry gas at two liters per minute across the sensor membrane.
4.0 CALIBRATION ERRORS AND ERROR CODES
The MiniOX 200/200E analyzers have a self test feature built into the software to detect faulty calibrations, oxygen sensor failures, and low operating voltage. These are listed below, and include possible actions to take, if an error code occurs.
E02: No sensor attached
MiniOX 200: Open unit and disconnect and reconnect sensor. Unit should
perform an auto calibration and should read 20.9%. If not, contact Ohio Medical Customer Service for possible sensor replacement.
MiniOX 200E: Disconnect and reconnect external sensor. Unit should perform an auto calibration, and should read 20.9%. If not, contact Ohio Medical Customer Service for possible sensor replacement or cable replacement.
E03: No valid calibration data available Make sure unit has reached thermal equilibrium. Press and hold the Calibration Button for three seconds to manually force a new calibration.
E04: Battery below minimum operating voltage Replace batteries.
CAL Err St: O2 Sensor reading not stable Wait for displayed oxygen reading to stabilize, when calibrating the device at 100% oxygen.
6 600002 Rev.1
Page 11
Wait for unit to reach thermal equilibrium, (Please note that this can take up to one half hour, if the device is stored in temperatures outside the specified operating temperature range).
CAL Err lo: Sensor voltage too low Press and hold the Calibration Button for three seconds to manually force a new calibration. If unit repeats this error more than three times, contact Ohio Medical Customer Service for possible sensor replacement.
CAL Err hi: Sensor voltage too high Press and hold the Calibration Button for three seconds to manually force a new calibration. If unit repeats this error more than three times, contact Ohio Medical Customer Service for possible sensor replacement.
CAL Err Bat: Battery voltage too low to recalibrate Replace batteries.
5.0 CHANGING THE BATTERIES
Batteries should be changed by service personnel.
» Use only brand name batteries. » Replace with two AA batteries and insert per orientation marked on the device.
Should the batteries require changing, the device will indicate this in one of two ways:
» The battery icon on the bottom of the display will begin to flash. This icon will continue to flash until the batteries are changed. The unit will continue to function normally for approximately 200 hours. » If the device detects a very low battery level, an error code of “E04” will be present on the display, and the unit will not function until the batteries are changed.
To change the batteries, begin by removing the three screws from the back of the device. A #1 Phillips screwdriver is required to remove these screws. Once the screws are removed, gently separate the two halves of the device.
The batteries can now be replaced from the back half of the case.
Be sure to orient the new batteries as indicated in the embossed polarity on the a back case.
NOTE: If the batteries are installed incorrectly the batteries will not make contact and the device will not operate.
Carefully, bring the two halves of the case together while positioning the wires so they are not pinched between the two case halves.
The gasket separating the halves will be captured on the back case half. Reinsert the three screws and tighten until the screws are snug. (FIGURE 3).
FIGURE 3
7600002 Rev.1
Page 12
The device will automatically perform a calibration and begin displaying % of oxygen.
FIGURE 4
LOCK LEVER
HELPFUL HINT: If unit does not function, verify that the screws are tight to allow proper electrical connection.
HELPFUL HINT (MiniOX 200E): Before closing the two case halves together, verify that the keyed slot on top of the coiled cable assembly is engaged on the small tab located on the back case.
This is designed to position the assembly in the correct orientation and prevent it from rotating. Improper positioning could hinder the case halves from closing
and prevent operation when tightening the screws.
6.0 CHANGING THE OXYGEN SENSOR
6.1 MiniOX 200 Model
Should the oxygen sensor require changing, the device will indicate this by presenting “Cal Err lo” on the display after initiating a calibration.
To change the oxygen sensor, begin by removing the three screws from the back of the device.
A #1 phillips screwdriver is required to remove these screws.
Once the screws are removed, gently separate the two halves of the device.
Disconnect the oxygen sensor from the printed circuit board by pressing the un-lock lever first and then pulling the connector out of the receptacle. The oxygen sensor can now be replaced from the back half of the case.
HELPFUL HINT: Be sure to orient the new sensor by aligning the red arrow on the sensor with the arrow in the back case. A small tab is located on the back case that is designed to engage the sensor and prevent it from rotating within the case. (FIGURE 4)
NOTE: If the oxygen sensor is installed incorrectly, the case will not come back together and the unit may be damaged when the screws are reinstalled.
NOTE: If the new sensor has red tape over the outside, remove it, then wait 30 minutes before calibrating.
Reconnect the oxygen sensor to the connector on the printed circuit board. Carefully bring the two halves of the case together while positioning the wires to ensure they are not pinched between the two case halves. Make sure the sensor is fully inserted and in the proper orientation.
Reinsert the three screws and tighten until the screws are snug. Verify the unit operates properly. The device will automatically perform a calibration and begin displaying % of oxygen.
8 600002 Rev.1
Page 13
6.2 MiniOX 200E Model
!
Should the oxygen sensor require changing, the device will indicate this by presenting “Cal Err lo” on the display.
Unthread the sensor from the cable by rotating the thumbscrew connector counterclockwise and pull the sensor from the connection. Replace the new sensor by inserting the electrical plug from the coiled cord into the receptacle on the oxygen sensor. Rotate the thumbscrew clockwise until snug. The device will automatically perform a calibration and begin displaying % of oxygen.
7.0 CLEANING AND MAINTENANCE
Store the MiniOX 200/200E analyzer in a temperature similar to its ambient environment of daily use.
The instruction given below describes the methods to clean and disinfect the instrument, sensor and its accessories (e.g. flow diverter, tee adapter):
Instrument Cleaning:
» When cleaning or disinfecting the exterior of the MiniOX 200/200E analyzer, take
appropriate care to prevent any solution from entering the instrument.
DO NOT immerse unit in fluids.
» The MiniOX 200/200E analyzer surface may be cleaned using a mild detergent and a
moist cloth.
» The MiniOX 200/200E analyzer is not intended for steam, ethylene oxide or radiation
sterilization.
Oxygen Sensor:
WARNING: Never install the sensor in a location that will expose the sensor
to patient’s exhaled breath or secretions, unless you intend to dispose of the sensor, flow diverter and tee adapter after use.
» Clean exterior surface of the the sensor with a cloth moistened with isopropyl alcohol
(65% alcohol/water solution).
»
Ohio Medical does not recommend use of spray disinfectants because they can
contain salts, which can accumulate in the sensor membrane and impair readings.
»
The oxygen sensor is not intended for steam, ethylene oxide or radiation sterilization.
Accessories:
»
Sterilizing -The flow diverter and tee adapter. Because of the variability of the cleaning, disinfecting and sterilizing processes, Ohio Medical cannot provide specific sterilization instructions, nor can the sterility of the item be ensured.
Ohio Medical recommends that operators carefully examine the flow diverter and tee adapter after sterilization and prior to use to verify that the item is fit for use. The operator should verify that there are no cracks and the item does not show any indication of material changes or physical damage that may compromise its effective use.
9600002 Rev.1
Page 14
8.0 SPECIFICATIONS
8.1 Base Unit Specifications
Measurement Range:.................................................... 0-100%
Resolution:.............................................................. 0.1%
Accuracy and Linearity: ..................1% of full scale at constant temperature, R.H. and
pressure when calibrated at full scale
Total Accuracy: .....................±3% actual oxygen level over full operating temp range
Response Time: ....................90% of final value in approximately 15 seconds at 23˚C
Warm-up Time: ................................................... None required
Operating Temperature: .....................................15˚C - 40˚C (59˚F - 104˚F)
Storage Temperature:.......................................-15˚C - 50˚C (5˚F - 122˚F)
Atmospheric Pressure:............................................800-1013 mBars
Humidity:................................................ 0-95% (non-condensing)
Power Requirements:..............................2, AA Alkaline batteries (2 x 1.5 Volts)
Battery Life: ..............................approximately 5000 hours with continuous use
Low Battery Indication:...................................."BAT" icon displayed on LCD
Sensor Type: .............................. Ohio Medical MiniOX series galvanic fuel cell
Expected Sensor Life:........................... > 1,500,000 O2 percent hours minimum
(2-year in typical medical applications) Dimensions:
200 Model Dimensions: ..............3.0”(W) x 4.0”(H) x 1.5”(D) [76mm x 102mm x 38mm]
200 Weight: ......................................................0.4 lbs. (170g)
200E Model Dimensions: .............3.0”(W) x 3.75”(H) x 1.5”(D) [76mm x 95mm x38mm]
200E Weight: .....................................................0.6 lbs. (285g)
Drift of Measurement: .......................< +/-1% of full scale at constant temperature,
pressure and humidity
8.2 Sensor Specifications
Type: .............................................. Galvanic fuel sensor (0-100%)
Life: ................................................2-years in typical applications
Height excludes external cable
9.0 MiniOX 200/200E SPARE PARTS AND ACCESSORIES
9.1 Included With Your Unit
Part Number Item 200 Model 200E Model
606904 Lanyard x x
606903 Threaded Adapter with Tubing x
606902 Flow Diverter x x
606901 Tee Adapter x x
606905 Dovetail x
10 600002 Rev.1
Page 15
9.2 Standard Replacement Parts and Accessories
Part Number Item 200 Model 200E Model
606930 MiniOX 250 Oxygen Sensor x
606940 MiniOX 250E Oxygen Sensor x
606911 Gasket x x
606912 #4-40 Pan Head Stainless Steel Screw x x
606913 Front Assembly (Includes Board & LCD) x x
606914 Back Assembly x x
606910 Coiled Cable Assembly x
606915 Overlay x x
9.3 Mounting Options (requires dovetail 606905)
Part Number Item
606906 Pole Clamp 606907 Wall Mount
606909 Rail Mount
NOTE: Repair of this equipment must be performed by a qualified service technician experienced in repair of portable hand held medical equipment.
11600002 Rev.1
Page 16
CLASSIFICATION
Protection against electric shock: ..........................Internally powered equipment.
Protection against water: ....................................................IPX1
Mode of Operation: ...................................................Continuous
Sterilization: ......................................................See section 7.0
Flammable anesthetic mixture: Not suitable for use in presence of a flammable anesthetic mixture
Product Disposal Instructions:
The sensor, batteries, and circuit board are not suitable for regular trash disposal. Return sensor to Ohio Medical for proper disposal or dispose according to local guidelines. Follow local guidelines for disposal of other components.
WARRANTY
The MiniOX 200/200E Analyzer is designed for medical oxygen delivery equipment and systems. Under normal operating conditions, Ohio Medical warrants the MiniOX 200/200E Analyzer to be free from defects of workmanship or materials for a period of 2-years from the date of shipment from Ohio Medical, provided that the unit is properly operated and maintained in accordance with Ohio Medical’s operating instructions. Based on Ohio Medical product evaluation, Ohio Medical’s sole obligation under the foregoing warranty is limited to making replacements, repairs, or issuing credit for equipment found to be defective. This warranty extends only to the buyer purchasing the equipment directly from Ohio Medical or through Ohio Medical 's designated distributors and agents as new equipment.
Ohio Medical warrants MiniOX oxygen sensors in the MiniOX 200/200E Analyzers to be free from defects in material and workmanship for a period of 2-years from Ohio Medical’s date of shipment in a MiniOX unit. Should a sensor fail prematurely, the replacement sensor is warranted for the remainder of the original sensor warranty period.
Routine maintenance items, such as batteries, are excluded from warranty. Ohio Medical and any other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential damages or equipment that has been subject to abuse, misuse, mis-application,
alteration, negligence or accident.
THESE WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
Manufactured By: Maxtec, LLC 2305 South 1070 West Salt Lake City, UT 84119 U.S.A.
Authorized Representative:
EC REP
QNET BV Kantstraat 19 NL-5076 NP Haaren The Netherlands
12 600002 Rev.1
Distributed By: Ohio Medical, LLC 1111 Lakeside Drive Gurnee, IL 60031 TEL 866.549.6446 www.ohiomedical.com U.S.A.
Page 17
Page 18
®
© 2017 Ohio Medical, LLC.
This document contains information that is proprietary and confidential to Ohio Medical, LLC. Use of
this information is under license from Ohio Medical, LLC.
Any use other than that authorized by Ohio Medical, LLC is prohibited.
MiniOX, Ohio Medical and the Ohio Medical logo are registered trademarks of Ohio Medical, LLC.
Loading...