Ohio Medical l-4-2-3-96-service-manual-for-amvex-120hl-high-low-flow-air-o2-blender O2 Blender Service Manual

L 4.2.3-96 Rev 2
120H/L High-Low
Flow Air/O2 Blender
Service Manual
Contact Us for Product Assistance
.Tenacore Holdings Inc. 1525 E. Edinger Avenue Santa Ana, CA 92705 USA Tel: 800.297.2241 Fax: 866.776.3101 www.tenacore.com
Distributed by Ohio Medical 1111 Lakeside Drive Gurnee, IL 60031 www.ohiomedical.com
866.549.6446
L 4.2.3-96 Rev 2
AMVEX 120H/L HIGH-LOW AIR/O2 BLENDER SERVICE MANUAL
TABLE OF CONTENTS
SECTION 1: Introduction……………………………………………………………………………………….01
SECTION 2: Overview of the 120H/L High-Low Flow Air/Oxygen Blender Operation……………………...02
SECTION 3: Warnings, Cautions & Notes……………………………………………………………………...05
SECTION 4: Setup Procedure…………………………………………………………………………………...07
SECTION 5: Performance Checks…………………………………………………………………..………......08
SECTION 6: Service and Repair…………...……………………………………………………………………09
SECTION 7: Calibration………………………………………………………………………………………...24
SECTION 8: Cleaning & Sterilization…..………………………………………………………………………29
SECTION 9: Maintenance & Service Policy………………………………………………………………........30
SECTION 10: Disposal Instructions…………………………………………………………………………….31
SECTION 11: Replacement Parts….……………………………………………………………...………….....32
SECTION 12: Product Specifications..…………………………………………………………...……………..35
SECTION 13: Troubleshooting………………………………………………………………………………….37
SECTION 14: Warranty………………………………………………………………………………………....39
NOTE: This manual is intended as a guide
for the servicing and calibration of the 120H/L High- Low Flow Air/O2 Blender.
Page 1 of 40
L 4.2.3-96 Rev 2
SECTION 1: INTRODUCTION
The Amvex 120H/L High-Low Flow Air/Oxygen Blenders are intended for non-invasive mixing and delivery of a ratio of medical air and medical oxygen as stand-alone devices or as components of a respiratory support apparatus implemented by trained medical professionals. The main body of the device is constructed of aluminum, and primary flow channels are contained within or between the aluminum pieces. Various internal components that are or can be in contact with the gas or gases include diaphragms, o-rings, check valves, and foam filters. All components are safe for oxygen use.
The Amvex 120H/L High-Low Flow Air/Oxygen Blender can be used in conjunction with:
Blender Wall Mount
Blender Mount with Pole Clamp
Blender Mount with Eclipse® or Fairfield® Equipment Rail Bracket
Blender Mount with Unimount® Equipment Rail Bracket
Blender Water Trap with Two 90-Degree Fittings
Medical Gas Hose
INDICATIONS FOR USE:
The Amvex 120H/L High-Low Flow Air/Oxygen Blender is intended to deliver air and oxygen with a respiratory setting. Oxygen concentrations can be adjusted from 21% to 100%. The blender is designed to operate from a 50 PSIG (3.52kg/cm2) air source and a 50 PSIG (3.52kg/cm2) oxygen source.
Use of this device is restricted to prescription or direction by a licensed medical practitioner.
NOTE:
Remove from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact Ohio Medical (refer to “Contact Us” page).
Page 2 of 40
L 4.2.3-96 Rev 2
SECTION 2: OVERVIEW OF THE 120H/L High-Low Flow AIR/OXYGEN BLENDER OPERATION
The 120H/L High-Low Flow Air/O2 blender mixes medical grade compressed air and oxygen to provide a pressurized gas source ranging from 21% to 100% oxygen.
Figure 1
GAS INLETS (FIGURE 2)
The two 30-75 PSIG gas sources enter through the diameter indexed (DISS) air and oxygen inlet connectors located on the bottom, rear of the blender. Each inlet connector incorporates a 30 micron particulate filter. After being filtered, the gases travel through duckbill check valves which prevent possible reverse gas flow to either the air or oxygen supply systems.
Oxygen Connector
DISS Air Connector
Figure 2
BALANCE MODULE (FIGURE 3)
The two gases then enter the two-stage Balance Module. The purpose of this module is to equalize the operating pressure of the air and oxygen gas sources before entering the Proportioning Module. The diaphragm responds to the difference in pressure and directs the movement (stroke) of each ball valve assembly contained within the air and oxygen chambers. The movement of each ball valve adjusts the amount of gas flowing through the Balance Module, equalizing the air and oxygen pressures to the lower of the two pressures.
Figure 3
Amvex Air/O2 Blender Diagram
PROPORTIONING MODULE (FIGURE 4)
From the Balance Module the gases flow into the Proportioning Module and are mixed according to the oxygen percentage selected on the external control knob. This Module consists of a double ended valve positioned between two valve seats.
Page 3 of 40
L 4.2.3-96 Rev 2
SECTION 2 OVERVIEW OF THE 120H/L High-Low Flow AIR/OXYGEN BLENDER OPERATION
One valve seat controls the passage of air and the other valve seat controls the passage of oxygen into the 120H/L High-Low Flow Air/O2 blender outlet. At this point, the two gases have been blended according to the oxygen percentage selected on the oxygen blender control knob.
(1.41Kg/cm²) differential is not seen, there will not be an audible alarm. If either source gas pressure drops, the output pressure of the blender will drop similarly, since the source gases are always balanced to that of the lower pressure. The bypass function operates in unison with the alarm. The alarm bypass poppet communicates directly with the air supply on one end and the oxygen supply on the other.
Figure 4
With the control knob at the full counter clockwise position (21%), the double ended valve will completely close off the flow of oxygen, allowing only the air to flow. By adjusting the control knob in the full clockwise (100%) position the flow of air is blocked, permitting only the flow of oxygen through the blender outlet.
ALARM/BYPASS (FIGURE 5)
The alarm feature provides for an audible alarm if source pressures differ by 20 PSI (1.41 Kg/cm²) or more. The primary purpose of the alarm is to audibly warn the operator of an excessive pressure drop or depletion of either source gas. The alarm will also activate in the event of elevation of either source gas when a difference of 20 PSI (1.41 Kg/cm²) or more is detected. Should both gas pressures (oxygen of medical air) increase or decrease simultaneously, and a 20 PSI
Figure 5
When the two source gases are near equal in pressure, the alarm bypass poppet is positioned over the bypass channel, blocking the flow of both gases. The poppet will remain seated for unequal pressures up to 20 PSI (1.41Kg/cm²). Once a 20 PSI (1.41Kg/cm²) difference is sensed by the poppet, the higher gas pressure will overcome the spring force and pressure the poppet at its opposite end, thus creating a path for gas (air or oxygen) to flow into the alarm channel. The gas with the higher pressure will also flow directly to the blender outlet port bypassing the Balance and Proportioning Modules. The gas is also directed to the bottom of the unit to the reed alarm, thus creating an audible warning. The oxygen concentration will be that of the gas at the higher pressure. The blender in the alarm/bypass mode will deliver the oxygen (100%) or air (21%) until the bypass mechanism resets when source gas pressure is restored to a differential of approximately 6 PSI (0.42 kg/cm²).
Page 4 of 40
L 4.2.3-96 Rev 2
SECTION 2 OVERVIEW OF THE 120H/L High-Low Flow AIR/ OXYGEN BLENDER OPERATION
If the 120H/L High-Low Flow Air/O2 Blender is set at 21% and the OXYGEN source pressure is reduced sufficiently to produce a 20 PSI (1.41Kg/cm²) or greater differential, the unit will not alarm because it will continue to deliver 21% concentration according to the setting. If the control is moved slightly from the 21% setting, the alarm will sound. Similarly, if the 120H/L High-Low Flow Air/O2 Blender is set to deliver 100% concentration and AIR source pressure is reduced or lost, the unit will not alarm because it will continue to deliver the selected 100% concentration. The 120H/L High-Low Flow Air/O2 Blender should be disconnected when not being used. If left connected to source gases while not being used (i.e. no output flow or bleed flow), the unit will not alarm if a 20 PSI (1.41Kg/cm²) or greater pressure differential develops. If the blender is not in use, an alarm under these conditions may be an unnecessary distraction or nuisance.
GAS OUTLETS (FIGURE 6 and FIGURE 7)
The primary gas outlet is utilized for unmetered low flow applications in the range of 15-120 LPM. The flow of gas is automatically initiated by an attachment of a pneumatic device to the outlet port. A check valve is unseated upon connection allowing the mixed gases to flow through the primary outlet.
Figure 6
Figure 7
The auxiliary outlet is located on the right side of the 120H/L High-Low Flow Air/O2 Blender and is designed to deliver metered gas through a flowmeter. Mixed gas may be delivered within specified accuracy tolerance from this outlet at 2 LPM and above. When a connection is made to the auxiliary outlet a 10-12 LPM bleed of mixed gas to the atmosphere is attained. The bleed flow exits the blender through a bleed port located on the bottom as shown in Figure 6.This bleed allows for applications utilizing low flows down to 5 LPM.
AUXILIARY
OUTLET
PRIMARY
OUTLET
Page 5 of 40
L 4.2.3-96 Rev 2
SECTION 3: WARNINGS, CAUTIONS & NOTES
The 120H/L High-Low Flow Air/O2 Blender should be operated by trained, qualified medical personnel under direct supervision of a licensed physician. Before clinical application, the WARNINGS, CAUTIONS and NOTES should be read and understood.
Safety Symbol
Content
WARNING
Indicates potential serious adverse reactions and safety hazards which if not avoided could result in death or serious injury
CAUTION
Defines any special care to be exercised by the user for the safe and effective use of the device and the care necessary to avoid damage to the device that may occur as a result of use or misuse
NOTE
Provides additional information to clarify an explanation or instruction for the safe and effective use of the device
Use this 120H/L High-Low Flow Air/Oxygen Blender only for its Intended Use as described in
this manual.
Identify existing conditions that could adversely affect the operator or patient.
DO NOT use this device in an MRI environment.
Always use medical grade gasses that meet the requirements of ANSI/CGA G7.1-2011 Grade D or
better and have a dew point of 5°F (2.75°C) or more below the lowest temperatures to which the air distribution (piping) system is exposed.
An air inlet filter/ water trap is recommended for use with the 120H/L High-Low Flow
Air/Oxygen Blender. Filters can become restricted in a manner that can cause insidious reduction in the flow capability of the blender and lead to malfunction.
DO NOT occlude or obstruct the bleed port or muffler on the bottom of the blender.
Adjustment of oxygen concentrations must be verified using an oxygen analyzer.
DO NOT use this 120H/L High-Low Flow Air/Oxygen Blender when the alarm is sounding.
Use recommended lubricants sparingly as lubricant may migrate to unintended parts of the blender
and cause it to malfunction.
The oxygen selector knob does not rotate 360 degrees. Rotating the dial beyond the endpoint
settings will damage the 120H/L High-Low Flow Air/Oxygen Blender.
WARNINGS
Page 6 of 40
L 4.2.3-96 Rev 2
SECTION 3: WARNINGS, CAUTIONS & NOTES
U.S. Federal law restricts this device to sale by or on the order of a licensed medical practitioner.
DO close off gas supply sources when the 120H/L High-Low Flow Air/Oxygen Blender is not in
use and the auxiliary outlet is connected. The continuous gas bleed may drain compressed gas tanks empty.
Store the 120H/L High-Low Flow Air/Oxygen Blender in a clean, dry area when not in use.
Ensure all connections are tight and leak free.
CAUTIONS
Allow equilibration time for fractional concentration of inspired oxygen (FIO2) changes before
analyzing gas.
Use pressure regulators that display regulated pressures.
Compressed gas can contribute to deposits that can obstruct filters. Filters can become restricted in
a manner that can cause insidious reduction in the flow capability of the blender and lead to malfunction. It is important to perform preventative maintenance as recommended and more frequently if the gas supply is not known to be clean and/or free of condensed water.
NOTES
Page 7 of 40
L 4.2.3-96 Rev 2
SECTION 4 SETUP PROCEDURE
The 120H/L High-Low Flow Air/Oxygen Blender should be tested in a system which closely duplicates the conditions of use for which the blender was designed. Illustrated below is a schematic diagram of the system that should be used to test the 120H/L High-Low Flow Air/ Oxygen Blender.
Page 8 of 40
L 4.2.3-96 Rev 2
SECTION 5: PERFORMANCE CHECKS
Prior to placing the 120H/L High-Low Flow Air/O2 Blender into clinical use, always perform the following test:
BLENDER ALARM BYPASS CHECK
BLENDER ADJUSTMENT
BLENDER RESPONSE
1. Connect 50±5 PSIG (3.52±0.35 kg/cm2)
air/oxygen source gases. Adjust control knob to 60%. Connect flowmeter to auxiliary outlet, set flow to 5 LPM minimum.
1. Alarm/Bypass should not activate.
2. Connect an oxygen flowmeter to auxiliary
outlet to activate auxiliary bleed and disconnect 50PSI (3.52 kg/cm2) air source from Blender.
NOTE: The Blender must be flowing gas for the alarm to activate.
2. Audible alarm.
3. Reconnect 50PSI (3.52 kg/cm2) air source to
Blender.
3.Audible alarm stops. Verify oxygen concentration (57% to 63%) with an oxygen analyzer.
4. Disconnect 50PSI (3.52 kg/cm2) oxygen
source from Blender.
4.Audible alarm.
5. Reconnect 50PSI (3.52 kg/cm2) oxygen source
to Blender.
5.Audible alarm stops. Verify oxygen concentration (57% to 63%) with an oxygen analyzer)
6. Verify oxygen flowmeter is set at 5 LPM.
6.Oxygen analyzer should read 57% to 63% when measured from the flowmeter outlet.
WARNINGS:
If the 120H/L High-Low Flow Air/O2 Blender does not function as
described below. Contact Ohio Medical (Refer to “Contact Us” page).
Do NOT use the blender until the correct performance is verified.
Page 9 of 40
L 4.2.3-96 Rev 2
31
15
6
34
32
33
31
29
35
30
29
32
PART NO. 10-0701-090002 REV.A
AMVEX
HIGH / LOW FLOW BASE ASSEMBLY
WITH DISS INLETS
35
33
34
16
6
15 31
15
6
16
32
34
33
29
31
35
32
30
29
32
15
6 1 2
14
13
12
11
13
17
13
18
19
19
6
20
21 15
36
22
23
26
37
25
23
22
35
24
33
16
34
20
6
18
15
16
3
1
32
5
39
4
40
27
41
28
5
8
7
8
9
10
38
Assembly
Drawing
120HL
High/Low
Flow
Blender
ITEM
NO.
PART NUMBER
QTY.
1 10-0700-080008
2
2 10-0700-020001
1 3 10-0700-040001
1 4 10-0700-030006
1
5 10-0700-060003
2 6 10-0700-060010
7 7 10-0702-010002
1 8 10-0700-060005
1 9 10-0700-010004
1 10 10-0700-060007
1 11 10-0700-030004
1 12 10-0700-060013
1 13 10-0700-060002
3 14 10-0700-010002
1 15 10-0700-020005
6 16 10-0700-060001
8 17 10-0700-020004
1 18 10-0700-030005
2 19 10-0700-060009
2 20 10-0700-060008
2
21 10-0700-070001
1 22 10-0700-060006
2 23 10-0700-010001
2 24 10-0700-020003
1
25 10-0700-090005
1 26 10-0700-090002
1 27 10-0700-080004
1 28 10-0700-080002
1 29 10-0700-020002
4 30 10-0700-080007
2 31 10-0700-030008
4 32 10-0700-030009
8 33 10-0700-080003
4 34 10-0700-030007
4
35 10-0700-010010
4
36 10-0700-010008
1 37 10-0700-090004
1 38 10-0700-090001
1 39 10-0701-010001
1 40 10-0701-030001
1
41 10-0701-030002
1
Page 10 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
WARNING:
THE 120H/L HIGH-LOW AIR/OXYGEN FLOW BLENDER SHOULD BE SERVICED AND/OR CALIBRATED BY A QUALIFIED AND TRAINED TECHNICIAN.
CAUTION: Before attempting to service/repair
the 120H/L High-Low Air/O2 Flow Blender, the service person should first be familiar with its design and operation as explained in Section 1 of this manual.
A. SERVICE/CALIBRATION TOOLS
Special tools (Figure 8) will be required for the service disassembly and assembly of the 120H/L High-Low Flow Blender. These products may be obtained from Ohio Medical.
Figure 8
Additional tools and supplies recommended for service/repair
5/32” Allen wrench
1/8” Allen wrench
9/32” Hex nut driver
5/32” Allen driver
1/8” Allen driver
3/4” Deep socket
11/16” Deep socket
3/4” Open end or adjustable wrenches (2)
11/16” Open end or adjustable wrench
7/32” Allen wrench
Small needle nose pliers
Torque wrench (to 120in/lb)
0-15 LPM Flowmeter (1/2 LPM increments)
A Maintenance Blender Kit may be ordered by specifying DAOBA-00000-05. This kit includes all parts necessary for periodic preventive maintenance.
Pin Wrench #TAOBA-00000-77
Alarm Tool #TAOBA-00000-78
Alignment Tool #TAOBA-00000-79
Page 11 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
1.0. BALANCING BLOCKS DISASSEMBLY/REASSEMBLY PROCEDURE (FIGURE 9 and 10)
BALANCE BLOCK DISASSEMBLY
-TOP OF 120H/L High-Low Flow BLENDER
1.1 With a 5/32” Allen wrench, remove the top
four screws securing the two balance block assemblies to the valve block.
NOTE: The balance block assemblies are identical and interchangeable. For ease of assembly, the blocks may be labeled.
1.2 Using the pin wrench (P/N: TAOBA-00000-
77), remove the block caps (2 each per balance block assembly). Remove O-rings and discard.
NOTE: Poppet spring and ball will be loose following removal of balance block cap. Remove components and set aside.
1.3 With a 5/32” Allen wrench, remove each of
four (4) screws securing each pair of blocks. Remove the diaphragms and O-rings and discard.
Clean all parts with an ultrasonic cleaner. Ensure all passages are blown completely dry before beginning reassembly. Be sure that the poppet seat areas are perfectly clean.
Balance block assembly replacement parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060001
8
O-Ring
10-0700-080007
2
Diaphragm
10-0700-060010
4
O-Ring
BALANCE BLOCK REASSEMBLY
1.4 Holding Alignment Tool (P/N: TAOBA- 00000-79) in hand, place the “A” Block diaphragm onto the alignment tool with the diaphragm cavity facing up.
1.5 Place Balance Block Diaphragm (P/N: 10-
0700-080007) into cavity.
NOTE: Make sure poppet pin on diaphragm seats into alignment tool.
1.6 Place “B” block on top of assembly with
diaphragm cavity facing down.
NOTE: Align block assemblies for proper gas flow. Three holes on each block (bottom) must be aligned.
1.2 Insert second diaphragm alignment tool into
Block “B”, making sure the poppet pin on
diaphragm seats into alignment tool.
TAOBA-00000-79
TAOBA-00000-79
10-0700-020002
10-0700-080007
10-0700-020002
A
B
Figure 9
Page 12 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
1.8 Fasten Block “A” and “B” together loosely with
two (2) 0.75” socket head cap screws (P/N 10­0700-030009).
1.9 Holding the two (2) alignment tools in place, lay the entire assembly with one of its surfaces on a flat surface. This will align the diaphragm blocks for mating with proportion block.
1.10 Using a 5/32” Allen wrench, tighten the
previously installed two (2) 2.25” socket head screws holding (P/N: 10-0700-030008) blocks “A” and “B” together. Torque to 60 in/lb.
1.11 Install and tighten, using a 5/32” Allen wrench,
remaining two (2) 0.75” socket head screws (P/N 10­0700-030009) to opposite side of “A” and “B” block assembly. Torque to 60 in/lb.
1.12 Remove both diaphragm alignment tools and
place “A” and “B” block assembly on its side.
1.13 Lightly lubricate O-Ring (P/N 10-0700-060010)
with lubricant and install on Balance Block Cap (P/N 10-0700-020005).
1.14 Place a very small amount of lubricant grease on one end of spring (P/N 10-0700-030007), then install lubricated end into balance block cap.
Figure 10
NOTE: Lubricant on spring end ensures adherence of
spring to cap during placement into block assembly.
1.15 Install block cap and spring into block assembly and tighten in place using pin wrench (P/N: TAOBA-00000-77).
1.16 Repeat steps 1.13 to 1.16 for the opposite side of
block assembly “AB”.
1.17 Reassemble blocks “CD” using the same
procedure for blocks “AB” beginning with step
1.4.
1.18 Lightly lubricate O-Rings (P/N 10-0700-
060001) with lubricant.
1.19 Place block assemblies and O-Rings aside of final assembly.
2.0. OXYGEN INLET DISASSEMBLY/
REASSEMBLY-BOTTOM REAR OF 120H/L High-Low Flow BLENDER (FIGURE 11)
NOTE: This assembly threads into the block with a
left-handed thread. A single groove on nut indicates left hand thread.
2.1 With a 3/4” open end wrench, remove the oxygen inlet assembly from the proportion block.
2.2 Using a second 3/4” open end wrench, separate the O2 inlet base connector and O2 inlet center connector.
2.3 Use a 3/4” open end wrench to stabilize the O2 inlet
center connector, then remove the O2 tail piece using a
1/8” Allen wrench.
10-0700-020005
10-0700-060010
10-0700-030007
10-0700-020002
10-0700-080003
Page 13 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
2.4 Remove the filter, duckbill check valve, O-Rings
and washer and discard. NOTE: The filter may have to be grasped with pliers
to remove.
Clean all parts with an ultrasound cleaner. Ensure all passages are blown completely dry before beginning reassembly.
OXYGEN INLET REASSEMBLY Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060012
1
O-Ring
10-0700-060010
1
O-Ring
10-0700-060004
2
O-Ring
10-0700-060014
1
Duckbill Check Valve
10-0700-040007
1
Step Washer
10-0700-080006
1
Inlet Cone Filter
Figure 10
2.5 Lightly lubricate O-Ring (P/N 10-0700-060010)
with lubricant and assemble inside of the O2 inlet center connector (P/N 10-0700-070006).
2.6 Lightly lubricate two (2) O-Rings (P/N 10- 0700-060004) with lubricant and assemble to O2 inlet center connector (P/N 10-0700-070006) and O2 inlet base connector (P/N 10-0700-070005).
2.7 Insert duckbill check valve (P/N 10-0700-
060014), step washer (P/N 10-0700-040007) and inlet cone filter (P/N 10-0700-080006) into O2 inlet base connector.
NOTE: Step on washer fits into duckbill check valve.
2.7 Using two (2) 3/4” open end wrenches, tighten the
O2 inlet center connector to the O2 inlet base connector. Torque to 120in /lbs.
2.8 Take unlubricated O-Ring (P/N 10-0700-060012) and assemble to O2 inlet nipple (P/N: 10-0700-07008).
2.9 Insert O2 inlet nipple (P/N: 10-0700-07008) into O2 inlet DISS nut (P/N 10-0700-070007) and
using a 1/8” Allen wrench, tighten to O2 inlet center
connector. Torque to 120 in./lbs.
2.10 Set O2 inlet aside for final assembly to proportion block.
3.0 AIR INLET DISASSEMBLY/REASSEMBLY-
BOTTOM REAR OF 120H/L High-Low Flow BLENDER (FIGURE 12)
3.1 With a 3/4” open end wrench, remove the air inlet assembly with O-Ring. Remove and discard O-Ring.
3.2 Remove the inlet cone filter located in valve block and discard.
NOTE: The filter may have to be grasped with
pliers to remove.
10-0700-060014
10-0700-060004
10-0700-040007
10-0700-070007
10-0700-070008
10-0700-080006
10-0700-070006
10-0700-060004
10-0700-060012
10-0700-070005
Page 14 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
3.3 Next, remove the washer and duckbill check
valve from the valve block assembly. Discard check valve and washer.
Clean all parts with ultrasonic cleaner. Ensure all passages are blown completely dry before beginning reassembly.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060004
1
O-Ring
10-0700-060014
1
Duckbill Check Valve
10-0700-040007
1
Washer
10-0700-080006
1
Inlet Cone Filter
AIR INLET REASSEMBLY
Figure 12
3.4 Install lightly lubricated O-Ring (P/N 10-
0700-060004) with lubricant on air inlet connector (P/N 10-0700-070009)
3.5 Place inlet cone filter (P/N 10-0700-080006) inside air inlet.
3.6 Set air inlet aside with duckbill check valve (P/N 10-0700-060014) and washer (P/N 10-0700-
040007) for final assembly to valve block.
4.0 PRIMARY OUTLET DISASSEMBLY/
REASSEMBLY-BOTTOM, FRONT OF 120H/L High­Low Flow BLENDER
4.1 With an 11/16” open end wrench, remove
the primary outlet from the valve block assembly.
NOTE: Small spring is loose and may fall out of cavity.
4.2 Remove poppet from primary outlet, then remove O-Ring from poppet and discard.
4.3 Remove O-Ring from primary outlet and discard.
Clean all parts with an ultrasound cleaner and rinse with clean, warm water. Ensure all passages are blown completely dry and that poppet seat areas are perfectly clean before beginning reassembly.
Replacement Parts:
PRIMARY OUTLET REASSEMBLY
Figure 13
PART NO.
QTY.
DESCRIPTION
10-0700-060001
1
O-Ring
10-0700-060010
1
O-Ring
10-0700-060004
10-0700-070009
10-0700-040007
10-0700-080006
10-0700-060014
10-0700-060001
10-0700-070002
10-0700-030001
10-0700-060010
10-0700-070003
Page 15 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
4.4 Lightly lubricate O-Ring (P/N 10-0700-
060010) with lubricant and install on check valve housing (P/N 10-0700-070003).
4.5 Lightly lubricate O-Ring (P/N 10-0700-
060001) on primary check valve poppet (P/N 10­0700-070002).
4.6 Insert poppet into the outlet housing.
4.7 Set assembly with spring (P/N 10-0700-030001) aside for final assembly.
5.0 AUXILIARY OUTLET DISASSEMBLY/ REASSEMBLY-RIGHT SIDE, FRONT OF
120H/L High-Low Flow BLENDER (FIGURE 14)
5.1 Using an 11/16” open end wrench, remove
the auxiliary outlet from the valve block assembly.
NOTE: Small spring is loose and may fall out of cavity.
5.2 Remove poppet from auxiliary outlet then remove O-Ring from poppet and discard.
5.3 Remove O-Rings from auxiliary outlet housing and discard.
Clean all parts with an ultrasonic cleaner. Ensure all passages are blown completely dry and poppet seats are perfectly clean before beginning reassembly. Ensure that the orifice on Auxiliary outlet (P/N: 10­0702-070001) is not occluded.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060001
1
O-Ring
10-0700-060010
1
O-Ring
10-0700-060004
1
O-Ring
AUXILIARY OUTLET REASSEMBLY
Figure 14
5.4 Lightly lubricate O-Rings with lubricant.
Install lubricated O-Rings (P/N 10-0700-060004 and 10-0700-060010) on low flow auxiliary outlet connector (P/N 10-0702-070001) and O-Ring (P/N 10-0700-060001) on auxiliary check valve (P/N 10-0700-070004).
5.5 Insert the auxiliary check valve into the auxiliary outlet connector.
5.6 Set assembly with spring (P/N 10-0700-030002) aside for final assembly.
PROPORTIONING MODULE
6.0 CONTROL KNOB/ FRONT AND REAR SEAT VALVE, DISASSEMBLY/ REASSEMBLY-FRONT OF 120H/L High-Low Flow BLENDER
6.1 With a thin blade screwdriver, remove the knob cover by gently separating it from the knob assembly.
6.2 Using a 9/32” nut driver, loosen the nut just
enough to remove the knob assembly from the front seat valve stem.
10-0700-060010
10-0700-060004
10-0700-060001
10-0700-030002
10-0702-070001
10-0700-070004
Page 16 of 40
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
6.3 Using an 11/16” open end wrench, remove front
seat lock nut. Remove and discard O-Ring from front seat locknut.
6.4 Using a thin blade screwdriver, remove the gray cover plate by gently separating it from the block assembly.
6.4 Using a pin wrench, remove the front seat (incorporating valve stem).
6.5 Separate the rear valve seat from the front valve seat. Remove and discard O-Rings.
6.6 Carefully remove poppet valve and spring from rear valve seat.
6.7 Using a 1/8” Allen wrench, remove the rear
valve seat and O-Ring from the valve body. Remove O-Ring from rear valve seat and discard.
Clean all parts with an ultrasonic cleaner. Ensure all passages are blown completely dry before beginning reassembly.
CAUTION: Valves and seats should be handled
carefully to avoid any damage.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060010
1
O-Ring
10-0700-060005
1
O-Ring
10-0700-060003
1
O-Ring
10-0700-060007
1
O-Ring
10-0700--060011
1
O-Ring
6.8 Inspect valve seat faces carefully. They should have a sharp edge void of chamfer, nicks or wear. Replace as needed.
CAUTION: Any damage to valve seats may
prevent proper calibration.
6.9 Lightly lubricate O-Rings (P/N 10-0700­060010, 10-0700-060003, and 10-0700-
060005) with lubricant and install on front valve seat (P/N 10-0701-010002).
6.10 Lubricate threads on front valve stem (P/N 10-0701-030002) with lubricant and carefully install front valve stem into front valve seat. Rotate valve stem clockwise until light contact is made with valve seat.
10-0701-030001
Figure 15
10-0700-060003
10-0700-060011
10-0700-030006
10-0701-010001
10-0700-010004
10-0700-060005
10-0701-030002
10-0701-010002
10-0700-060007
10-0700-060010
Page 17 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
6.11 Rotate front valve stem clockwise until
light contact is made with front valve seat.
CAUTION: Do NOT over tighten the front
valve stem and valve seat assembly as valve seat damage could occur.
6.12 Set control knob (P/N 10-0700-040009), knob cover (P/N 10-0700-040006) cover plate (P/N 10­0700-040002) , front seat assembly, spring (P/N 10­0700-030006), and O-Ring (P/N 10-0700-060007) aside for final assembly.
7.0 BYPASS DISASSEMBLY/REASSEMBLY­BOTH SIDES, REAR OF 120H/L High-
Low Flow BLENDER (FIGURE 16)
7.1 Using a slender, pointed probe, remove tamper resistant sticker from both side caps.
7.2 Using a 1/8” Allen wrench, unscrew the bypass adjuster from left and right hand side bypass seats.
7.3 Remove and discard O-Ring from each bypass adjuster.
NOTE: A small spring is contained in each
assembly and might remain in poppet bypass valve housing after adjuster is being removed.
7.4 Using a pin wrench, unscrew bypass sleeve from left rear side of valve block. Remove and discard O­Rings.
7.5 From right rear side of block, unscrew bypass seat. Remove and discard O-Ring.
NOTE: If spring(s) are still in cavity, carefully remove them.
7.6 Carefully push bypass poppet valve through bypass sleeve.
NOTE: Use a blunt slender probe to push poppet valve out of enclosure. Use care to avoid scratching surface of cylinder in which poppet valve operates.
7.7 Remove and discard O-Ring from poppet valve.
CAUTION: Carefully inspect internal
surface of sleeve for any signs of wear and damage to the special impregnated Teflon® coating.
Clean all parts with an ultrasonic cleaner. Ensure all passages are blown completely dry before beginning reassembly.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060006
2
O-RING
10-0700-060008
2
O-RING
10-0700-060009
2
O-RING
10-0700-060002
2
O-RING
Page 18 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
ALARM BYPASS REASSEMBLY
Figure 16
7.8 Using lubricant, lubricate (2) O-Rings (P/N 10-
0700-060008). Install one (1) O-Ring on bypass seat cap (P/N 10-0700-020003) and one (1) O-Ring on alarm sleeve (P/N 10-0700-070001).
7.9 Using PFPE Oil, lightly lubricate two (2) O- Rings (P/N 10-0700-060009) and install on alarm sleeve (P/N 10-0700-070001).
7.10 Lightly lubricate two (2) O-Rings (P/N 10-0700-060006) and install one (1) on each bypass adjuster (P/N 10-0700-010001).
7.11 Thoroughly lubricate two (2) O-Rings (P/N 10- 0700-060002) with lubricate grease and install one (1) in each groove at end of alarm shuttle (P/N 10­0700-020004).
7.12 Set bypass seat assembly, alarm sleeve assembly, bypass poppet assembly, alarm sleeve springs (P/N 10-0700-030005) and adjuster assemblies aside for final assembly.
8.0 OUTLET CAP DISASSEMBLY/ REASSEMBLY-LEFT SIDE, FRONT OF 120H/L High-Low Flow BLENDER
8.1 Using pin wrench, remove outlet cap and
O-Ring from valve block. Remove and discard O-Ring.
OUTLET CAP REASSEMBLY
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060010
1
O-RING
8.2 Lightly lubricate and install O-Ring (P/N 10-0700-060010) on block cap (P/N 10-0700-
020005). Set aside for final assembly.
9.0 MUFFLER DISASSEMBLY-BOTTOM MIDDLE OF 120H/L High-Low Flow BLENDER
9.1 With a small screwdriver, carefully lift star
washer from the bottom of the valve block. Discard star washer.
9.2 Remove the muffler bleed carefully with a pointed probe and discard.
MUFFLER ASSEMBLY
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-080002
1
Star Washer
10-0700-080004
1
Muffler Bleed
9.3 Set muffler (P/N 10-0700-080004) and star washer (P/N 10-0700-080002) aside for final assembly.
10-0700-060002
10-0700-010001
10-0700-030005
10-0700-060006
10-0700-060008
10-0700-020003
10-0700-060009
10-0700-020004
10-0700-070001
Page 19 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
10.0 ALARM CAP DISASSEMBLY/REASSEMBLY-
BOTTOM, MIDDLE OF 120H/L HIGH­LOW FLOW BLENDER (FIGURE 17)
10.1 Using blender alarm tool (P/N: TAOBA-
00000-78),unscrew alarm cap
10.2 Remove diffuser foam and discard.
10.3 Remove spring.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-080005
1
Foam Diffuser
10-0700-040003
1
Alarm Cap
ALARM CAP REASSEMBLY
Figure 17
10.4 Install spring (P/N 10-0700-030003) with
its wide base on top of reed inside alarm cap (P/N 10-0700-040003).
10.5 Place foam diffuser (P/N 10-0700-080005) into alarm cap above spring.
10.6 Check alarm assembly for proper audible function.
10.7 Set alarm assembly aside for final assembly.
11.0 ALARM CHECK VALVE DISASSEMBLY/REASSEMBLY-REAR OF
120H/L HIGH-LOW FLOW BLENDER (FIGURE 18)
11.1 With pin wrench, remove block cap from rear of valve block.
11.2 Remove and discard O-Ring from block
cap.
11.3 Using a 5/32” Allen wrench, remove
checkball retainer, rubber checkball and spring.
11.4 Remove and discard O-Ring and rubber checkball.
Clean all parts with an ultrasonic cleaner. Ensure all passages are blown completely dry before beginning reassembly.
Replacement Parts:
PART NO.
QTY.
DESCRIPTION
10-0700-060013
1
Rubber Checkball
10-0700-060002
1
O-Ring
10-0700-060008
1
O-Ring
ALARM CHECK VALVE REASSEMBLY
Figure 18
10-0700-040003
10-0700-030003
10-0700-080005
10-0700-060013
10-0700-030004
10-0700-010002
10-0700-060002
10-0700-060008
10-0700-020005
Page 20 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
11.5 Lubricate O-Ring (P/N 10-0700-060008) and install on block cap (P/N 10-0700-020005).
11.6 Lubricate O-Ring (P/N 10-0700-060002)
and install in groove on checkball retainer (P/N 10-0700-010002).
11.7 Inspect new rubber checkball (P/N 10-0700-
060013) to ensure that it is spotless, clean and not damaged by scratches, nicks or flat spots, lubricate lightly.
11.8 Set rubber checkball, checkball retainer assembly, spring (P/N 10-0700-030004) and cap assembly aside for final assembly.
12.0 VALVE BLOCK
12.1 Clean valve block with an ultrasonic cleaner.
Ensure all passages are blown completely dry before assembly. Inspect for any sign of excessive wear, damage, or any condition that may affect proper function.
SEQUENCE/INDEX OF FINAL ASSEMBLY
13.0 ALARM CHECK VALVE ASSEMBLY
14.0 FRONT/REAR SEAT ASSEMBLY
15.0 ALARM BYPASS ASSEMBLY
16.0 MUFFLER
17.0 OUTLET CAP
18.0 AUXILIARY OUTLET ASSEMBLY
19.0 PRIMARY OUTLET ASSEMBLY
20.0 BALANCE BLOCK
21.0 AIR INLET ASSEMBLY
22.0 OXYGEN INLET ASSEMBLY
23.0 ALARM CAP ASSEMBLY
120H/L HIGH-LOW FLOW AIR/OXYGEN BLENDER FINAL ASSEMBLY
CAUTION: Lightly lubricate all threaded
components with the exception of the Rear Valve Seat (P/N 10-0702-010001).
13.0 ALARM CHECK VALVE ASSEMBLY
13.1 Position Blender resting on front surface,
with rear of proportion block (P/N 10-0700-
020001) facing upwards.
13.2 Install spring (P/N 10-0700-030004); place rubber checkball (P/N 10-0700-060013) on spring.
13.3 Using a 5/32” Allen driver, install checkball retainer (P/N 10-0700-010002). Verify that O­Ring (P/N 10-0700-060002) is lightly lubricated prior to installing retainer. Torque to 30 in/lb.
13.4 Using pin wrench, install block cap (P/N 10- 0700-020005), verify O-Ring (P/N 10-0700-
060010) is lightly lubricated. Torque to 30 in/lb.
14.0 FRONT AND REAR VALVE SEAT ASSEMBLY-FRONT OF VALVE BLOCK
14.1 Position blender with front surface facing
upwards
14.2 Install rear valve seat (P/N 10-0702-
010001) with lubricated O-Ring (P/N 10-0700-
060011) through front port.
14.3 Using a 1/8” Allen driver torque wrench, secure rear seat in place. Torque to 25 in/lb.
Page 21 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
14.4 Lightly lubricate one end of spring (P/N 10-
0700-030006) with lubricant and insert lubricated end into rear valve stem (P/N 10-0701-030001).
14.5 Using small, slim needle nose pliers, carefully place rear valve stem with spring into rear valve seat (P/N 10-0701-010001).
14.6 Verify that front valve stem is hand-tightened until light contact is made with front seat. Place front valve seat assembly (P/N 10-0701-010002) into the threaded hole. With a pin wrench, exert slight pressure and rotate COUNTER-CLOCKWISE until initial thread engagement is found.
This initial thread engagement is very important for the initial calibration, rotate assembly CLOCKWISE three (3) full turns. Then rotate assembly COUNTER-CLOCKWISE one and a half (1.5) turns. The o-ring should not be visible.
14.7 Align the dowel pins in the front cover plate (P/N 10-0700-040002) with the two holes in the valve block surface. Gently push front cover plate against blender front surface
14.8 Loosely hand fasten nut (P/N 10-0700-
010004) with wider shoulder against front plate.
NOTE: DO NOT install O-ring (P/N 10-0700-
060007) in groove at this time.
14.9 Install control knob with black pointer against 21% O2 stop, at left side of valve block.
15.0 ALARM BYPASS ASSEMBLY-
LEFT/RIGHT SIDE, REAR OF VALVE BLOCK
15.1 Position blender resting on its top side.
Verify that the alarm shuttle (P/N 10-0700-
020004) with two (2) lubricated O-rings (P/N 10­0700-060002) is inside alarm sleeve.
15.2 After verifying all O-rings are lubricated, insert assembled alarm sleeve (P/N 10-0700-
070001) into valve block. Using pin wrench, secure alarm sleeve into valve block. Torque to 30 in/lb.
CAUTION: Be extremely careful not to
damage O-rings (P/N 10-0700-060009) during alarm sleeve (P/N 10-0700-070001) installation.
15.3 Install one spring (P/N 10-0700-030005) through alarm sleeve into bypass poppet valve port.
NOTE: Ensure spring (P/N 10-0700-030005) is positioned into alarm shuttle (P/N 10- 0700-020004).
15.4 Using an 1/8” Allen wrench, screw bypass adjuster (P/N 10-0700-010001) with lightly lubricated O-ring (P/N 10-0700-060006) into alarm sleeve until bypass adjuster is slightly recessed into alarm sleeve.
15.5 Position blender on alarm sleeve side and install bypass seat cap (P/N 10-0700-020003) with lubricated O-ring (P/N 10-0700-060007). Using a pin wrench, secure seat to valve block. Torque to 30 in/lb.
WARNING:
DO NOT PRESSURIZE THE SYSTEM UNLESS
THE FRONT VALVE SEAT HAS THREE (3) FULL TURNS OF THE THREADS INSIDE THE VALVE BLOCK. SEAT CAN BE FORCEFULLY EJECTED. DO NOT EXCEED THREE (3) FULL TURNS AS THIS MAY DAMAGE THE REAR VALVE SEAT.
Page 22 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
15.6 Carefully insert 1/8” Allen wrench through
bypass seat cap and into alarm shuttle. Push alarm shuttle against spring (P/N 10-0700-030001) and check for smooth movement and recoil action.
15.7 Install spring (P/N 10-0700-030001) through bypass seat cap (P/N 10-0700-020003) into hole in alarm shuttle (P/N 10-0700-020004).
15.8 Using an 1/8” Allen wrench, screw bypass adjuster screw (P/N 10-0700-010001) with lubricated O-ring (P/N 10-0700-060006) into bypass seat, until adjuster is slightly recessed into valve seat.
16.0 MUFFLER ASSEMBLY-BOTTOM, CENTER OF VALVE BLOCK (FIGURE 19)
16.1 Install one (1) muffler bleed (P/N 10-0700-
080004) into valve block bleed port.
16.2 With a small screwdriver, secure the star
washer (P/N 10-0700-080002) over the muffler bleed.
Figure 19
17.0 OUTLET CAP-BOTTOM, FRONT OF VALVE BLOCK
17.1 Using pin wrench, secure outlet cap (P/N 10-
0700-020005) with lightly lubricated O-ring (P/N 10-0700-060010) into valve block. Torque to 30 in/lb.
18.0 AUXILIARY OUTLET- RIGHT SIDE. FRONT OF VALVE BLOCK
18.1 Position blender assembly on its side
(auxiliary outlet port facing up), then install spring (P/N 10-0700-030002) into bottom of recess in auxiliary outlet valve block port.
NOTE: Auxiliary outlet spring (P/N 10­0700-030002) is longer than primary outlet spring (P/N 10-0700-030001).
18.2 Install auxiliary outlet connector (P/N 10- 0702-070001) and poppet into auxiliary outlet port on valve block and hand tighten in place.
18.3Using an 11/16” open end wrench, tighten assembly to valve block. Torque to 120 in/lb.
19.0 PRIMARY OUTLET ASSEMBLY LEFT SIDE, FRONT OF VALVE BLOCK
19.1 Position blender assembly on its side, with
primary outlet port facing upwards. Install primary spring (P/N 10-0700-030001) in center of primary outlet cavity bottom.
NOTE: Primary outlet spring is shorter than auxiliary outlet spring.
10-0700-080004
10-0700-020005 10-0700-060010
Page 23 of 39
L 4.2.3-96 Rev 2
SECTION 6: SERVICE & REPAIR
19.2 After verifying O-rings are lubricated, install
primary check valve housing (P/N 10-0700-070003) and poppet (P/N 10-0700-070002) into primary outlet port on valve block and hand tighten in place.
19.3 Using 11/16” wrench, tighten assembly to
valve block. Torque to 120 in/lb.
20.0 DIAPHRAGM BLOCK ASSEMBLY-TOP OF VALVE BLOCK
20.1 Assemble four lubricated O-rings (P/N 10-
0700-060001) each, to two diaphragm block assemblies.
20.2 Using a 5/32” Allen wrench and a torque wrench, secure the two diaphragm block assemblies to the proportion block with four screws (P/N 10-0700-030008). Tighten screws to 60 inch-pounds.
NOTE: Align diaphragm block assemblies squarely with valve block prior to tightening in place.
21.0 AIR INLET ASSEMBLY-BOTTOM, LEFT REAR OF VALVE BLOCK
21.1 Install duckbill check valve (P/N 10-0700-
060014) in air inlet port with bill facing inside cavity.
21.2 Place washer (P/N 10-0700-040007) on top of duckbill check valve. Raised step fits into duckbill.
NOTE: Lightly lubricate both sides of washer to prevent binding or twisting between duckbill check valve and nylon cone inlet filter.
21.3 Place large diameter end of nylon cone inlet filter (P/N 10-0700-080006) into air inlet port on valve block.
NOTE: Install air inlet fitting (P/N 10-0700-
070009) with lubricated O-ring (P/N 10-0700-
060004) into air inlet port on valve block and hand tighten in place.
21.4 Using a 3/4" wrench, secure air inlet into valve block. Torque to 120 in/lb.
22.0 OXYGEN INLET ASSEMBLY BOTTOM REAR, RIGHT OF VALVE BLOCK
22.1 Hand-tighten oxygen inlet assembly
into valve block port.
NOTE: This assembly threads into the block with a left hand thread. Turn COUNTERCLOCKWISE to tighten. Single groove on nut indicates left hand thread.
22.2 With a 3/4” wrench, secure assembly to valve block. Torque to 120 in/lb.
23.0 ALARM CAP ASSEMBLY
23.1 From below valve block, carefully install
alarm cap assembly into valve block. Hand­tighten only.
Page 24 of 39
L 4.2.3-96 Rev 2
SECTION 7: CALIBRATION
1. CALIBRATION TOOLS/EQUIPMENT
Thin Bladed Screwdriver or Knife1/8” Allen Wrench1 1/8” Open End Wrench7/8” Open End Wrench11/16” Open End Wrench9/32” Nut DriverPin Wrench (P/N TAOBA-00000-77)Oxygen Regulator (2 Stage, Adjustable 0-80
PSIG [0-5.60 kg/cm²])
Air Regulator (2 Stage, Adjustable 0-80
PSIG).
Oxygen Flowmeter Adapter (0-15LPM)Oxygen Analyzer (analyzer should read in
tenths to ensure accuracy of calibration)
Oxygen Sampling HoseOxygen Sensor
2. AIR/OXYGEN SETUP
The gas supplies must be clean and dry and
have the ability to generate 80 PSIG (0-
5.60 kg/cm²) for both air and oxygen inlet pressures.
When high pressure tanks are utilized,
blow potential debris from the valve; quickly open and close each valve to prevent debris from entering the test equipment.
Connect recommended adjustable air and
oxygen regulators to each gas supply,
securing with a 1 1/8” open end wrench.
Turn the oxygen and air regulator control
knobs to full counterclockwise closed position.
Secure the air and oxygen high pressure
hoses to each regulator using applicable wrenches.
3. OXYGEN ANALYZER SETUP/
CALIBRATION
The accuracy of the calibration of the
120H/L High-Low Flow Blender will depend heavily upon the accuracy of the oxygen analyzer.
The oxygen analyzer should have a
response time of 10 seconds or less. The analyzer should read in tenths and ideally be of the digital type.
Calibrate the oxygen analyzer according
to the manufacturer’s procedure.
4. TEST EQUIPMENT SETUP
Install the built-in post bracket on
the120H/L High-Low Flow Blender into the post bracket on the pole.
Using 7/8” and 11/16” open end
wrenches, secure the air and oxygen high pressure hoses to the 120H/L High-Low Flow Blender inlets.
Attach and secure flowmeter, in
upright position, to side outlet on adapter.
NOTE: Ensure flowmeter is turned OFF.
Attach one end of connecting hose to
flowmeter and other end to bifurcation. Ensure one way valve is secured into remaining large opening of bifurcation. Attach remaining outlet of bifurcation to oxygen analyzer probe.
The system is now ready for calibration.
Page 25 of 39
L 4.2.3-96 Rev 2
SECTION 7: CALIBRATION
CALIBRATION PROCEDURE 120H/L HIGH­LOW FLOW OXYGEN BLENDER
1.0 FRONT AND REAR SEAT LEAK TEST
1.1 Connect gas supply source(s),0-80 PSIG (0-
5.62 kg/cm²) pressure regulator(s), pressure test tube(s), leak test tube, and sampling beaker half way filled with water to valve block as shown in Figure
20.
NOTE: Compressed air may be utilized for seat leak test.
1.2 Adjust O2 pressure regulator between 10-15 PSIG (0.70 – 1.05 kg/cm²). Air regulator remains OFF (i.e. CLOSED).
1.3 Slowly turn blender valve stem (P/N 10­0702-010003) clockwise, until bubbles appear on water surface.
1.4Then turn valve stem counterclockwise until bubbles just stop. A maximum leak of 4 bubbles per minute is acceptable.
NOTE: If bubbling continues at a rate greater than 4 bubbles per minute, either the front seat or valve stem are damaged and should be replaced.
1.5 Carefully install control (or O2 selection) knob with black pointer against left side stop (i.e. 21% O
2
), and secure knob collet nut with 9/32” nut
driver. Turn Oxygen supply source OFF (i.e. CLOSED).
Figure 20
Compressed O
2
Supply
Adjustable
Regulator 0-80PSIG
For 21%
closed position, adjust front valve seat
assembly.
Air
Adjustable
Regulator 0-80PSIG
O
2
For 100%
closed position, adjust
valve stem.
Sampling Device
Gauge
Gauge
Compressed Air
Supply
Page 26 of 39
L 4.2.3-96 Rev 2
SECTION 7: CALIBRATION
1.6 Carefully rotate O2 selection knob clockwise
until black-pointer rests against right side stop (i.e. 100% O2).
CAUTION: Extreme care must be taken during
this procedure. Any resistance to rotation is likely to be valve contact with rear seat. Further turning of knob will force the valve into the rear seat making subsequent calibration difficult.
NOTE: If unable to fully rotate knob to right side stop (i.e. 100% O2), return the knob to the left side stop (i.e. 21% O2), and remove the knob. With pin wrench, rotate the front valve seat (P/N 10-0702-
010002) counterclockwise approximately one half (1/2) turn, and return to step 1.2
1.7 Carefully remove O2 selection knob, make certain not to upset or change valve stem position.
1.8 Adjust air pressure regulator between 10-15 PSIG (0.70 – 1.05 kg/cm²).
1.9 Using pin wrench, carefully and slowly, turn front valve seat (P/N 10-0702-010002) clockwise until bubbling at water surface just stops. STOP adjusting seat, the very moment bubbling ends. Be aware that bubbling may continue after seat and valve have made contact. A maximum leak of 4 bubbles per minute is acceptable.
NOTE: If bubbling continues at a rate greater than 4 bubbles per minute, either the rear seat or valve stem are damaged and should be replaced.
1.10 Using a 11/16” wrench, secure nut (P/N 10­0700-010004) checking the seat position has not changed during tightening of nut, by ensuring that bubbling has just stopped and does not reappear. Torque nut to 40 in/lb.
1.11 Install O-ring (P/N 10-0700-060007) on nut. Carefully reinstall O2selection knob with pointer at right side (i.e.100% O
2)
stop. Torque
collet nut to 4 in/lb.
1.12 Rotate knob counter clockwise to verify left
hand (i.e. 21% O2) top is bubbling at a rate of less than 4 bubbles per minute. Install cap on knob.
2.0 PROPORTIONING VALVE
NOTE: Proportioning valve endpoints are set by
following Front & Rear Seat Leak Test.
2.1 Turn air and oxygen sources ON. Adjust both regulators to a static 50 PSIG and adjust flowmeter to 3.5 LPM.
2.2 Rotate control knob counterclockwise to 21% stop, allow analyzer to stabilize.
2.3Perform the following checks:
Knob
Pressure
Setting
Oxygen/Air
% Concentration
21
*50 PSI/50 PSIG
21.0%/22.0%
30
*50 PSI/50 PSIG
27.0%-33.0%
60
*50 PSI/50 PSIG
57.0%-63.0%
90
*50 PSI/50 PSIG
87.0%-93.0%
100
*50 PSI/50 PSIG
97.0%-100%
* = 3.52 kg/cm² / 3.52 kg/cm²
Knob
Pressure
Setting
Oxygen/Air
% Concentration
30
*50 PSI/50 PSIG
27.0%-33.0%
60
*50 PSI/50 PSIG
57.0%-63.0%
90
*50 PSI/50 PSIG
87.0%-93.0%
* = 3.52 kg/cm² / 2.81 kg/cm²
30
*50 PSI/50 PSIG
27.0%-33.0%
60
*50 PSI/50 PSIG
57.0%-63.0%
90
*50 PSI/50 PSIG
87.0%-93.0%
* = 3.52 kg/cm² / 4.22 kg/cm² If concentrations do not meet specifications, repeat Front and Rear Seat Leak Test
Page 27 of 39
L 4.2.3-96 Rev 2
SECTION 7: CALIBRATION
2.4 Adjust O2 selection knob at 60% oxygen
and set air/oxygen pressures each at 50 PSIG.
2.5 Remove flowmeter adapter and flowmeter from primary outlet of blender.
2.6 Disconnect flowmeter from primary outlet, then connect flowmeter to auxiliary outlet (i.e. right side outlet) on blender .
2.7 Adjust flowmeter to 0.5LPM and check oxygen analyzer. Reading should be between
57.0%-63.0%.
2.8 Insert bleed test tube assembly into bleed port at bottom of valve block, flow tube should read
2.5-3.5 LPM.
NOTE: Should blender not meet either specification, 2.7 or 2.8, inspect bleed orifice in
auxiliary outlet or bleed orifice in valve block for occlusion.
2.9 Increase air pressure from 50 PSIG (3.52 kg/cm²) to 60 PSIG (4.22 kg/cm²). O2% on oxygen should read 57%-63%.
2.10 Lower air pressure from 60 PSIG (4.22 kg/cm²) to 40 PSIG (2.81 kg/cm²). O2% on oxygen analyzer should read 57%-63%.
2.11Repeat steps 1.8 and 1.9 by increasing or decreasing oxygen supply pressure.
2.12 Remove bleed test tube assembly from bleed port.
3.0 ALARM CALIBRATION The Alarm system is designed to sound an audible tone if the inlet pressures are different by 20 PSI (1.41 kg/cm²) or more, such as if either source gas failed.
When the 120H/L High-Low Flow Blender is in the alarm phase, the remaining or higher pressure gas is routed to the blender outlet. Some gas will also flow through the alarm reed valve creating an audible tone.
3.1 Ensure air and oxygen regulators are adjusted to a static 50 PSIG (3.52 kg/cm²), align control knob indicator with 60%, and ensure flowmeter is set to 15 LPM
3.2 Reduce air pressure until the audible alarm sounds. The air pressure should read 30 ± 2 PSIG (2.11 ± 0.14 kg/cm²).
If the alarm sounds above this pressure, rotate adjuster clockwise, left side of blender, with a
1/8” Allen wrench until alarm sounds as 30 ± 2
PSIG (2.11 ± 0.14 kg/cm²). If the alarm sounds below this pressure, rotate
adjuster counterclockwise, left side of blender, with a 1/8” Allen wrench until alarm sounds at 30 ± 2 PSIG (2.11 ± 0.14 kg/cm²).
3.3Raise air pressure slowly. Alarm bypass should reset to normal function when pressure reaches 44 PSIG (3.10 kg/cm²) or above.
3.4 Restore air pressure to 50 PSIG (3.52 kg/cm²) and reduce oxygen pressure until the audible alarm sounds. The oxygen pressure must be 30 ± 2 PSIG (2.11 ± 0.14 kg/cm²).
If alarm sounds above this pressure, rotate
adjuster clockwise, right side of blender,
with a 1/8” Allen wrench until alarm
sounds at 30 ± 2 PSIG.
If alarm sounds below this pressure, rotate
adjuster counterclockwise, right side of
blender, with a 1/8” Allen wrench until
alarm sounds at30 ± 2 PSIG (2.11 ± 0.14 kg/cm²).
Page 28 of 39
L 4.2.3-96 Rev 2
SECTION 7: CALIBRATION
3.5 Raise oxygen pressure slowly. Alarm bypass
should reset to normal function when pressure reaches 44 PSIG (3.10 kg/cm²) or above.
4.0 OUTLET FLOW TEST
4.1 With 0 – 35 LPM flowmeter connected to
auxiliary outlet fitting, set O2 selection knob at 60% and oxygen/air supply sources at 50 PSIG (3.52 kg/cm²) static.
4.2 Turn flowmeter control knob completely open. Flowmeter should read a minimum of 30 LPM.
4.3 Reduce air pressure to 0 PSIG (0 kg/cm²). Flowmeter should read a minimum of 30 LPM. Audible alarm should sound.
4.4 Increase air pressure to 50 PSIG (3.52 kg/cm²) and rotate O2 selection knob against 21% stop. Flowmeter should read a minimum of 30 LPM.
4.5 Rotate O
2
selection knob against 100% stop.
Flowmeter should read a minimum of 30 LPM.
4.6 Set O2 selection knob at 60%. Reduce O
2
pressure to 0 PSIG (0 kg/cm²). Flowmeter should read a minimum of 30 LPM. Audible alarm should sound.
4.7 Remove flowmeter form auxiliary outlet and connect to primary outlet fitting. Repeat steps 4.2 through 4.6.
4.8 Remove flowmeter from primary outlet fitting.
5.0 INSTALLATION OF CONTROL KNOB
5.1 Rotate control knob fully to 100%
position. Check O2 concentration.
5.2 Using a 9/32” nut driver, loosen
control knob nut (P/N 10-0700-010007) and remove control knob.
5.3 Install O-ring (P/N 10-0700-060007) on front valve seat nut (P/N 10-0700-
010004).
5.4 Push control knob onto front valve stem, seating it fully on the front valve seat nut O-ring with the black pointer at the 100% position. Be careful not to rotate valve stem.
5.5 Tighten the control knob nut, using a 9/32” socket. Torque to 4 in/lb.
5.6 Recheck 100% - 21% O2 concentration.
5.7 Snap control knob cover (P/N 10-
0700-040006) into control knob.
Page 29 of 39
L 4.2.3-96 Rev 2
SECTION 8: CLEANING AND STERILIZATION
Use an all-purpose liquid cleaner on exterior. For other general cleaning, use a 70% isopropyl alcohol solution.
WARNINGS:
Do NOT steam autoclave or otherwise subject 120H/L High-Low Flow
Air/Oxygen Blender to temperatures over145°F (62°C).
Do NOT immerse assembled 120H/L High-Low Flow Air/Oxygen
Blender into liquid decontamination agents.
Do NOT use any strong solvent or abrasive cleaners on labels.
Page 30 of 39
L 4.2.3-96 Rev 2
SECTION 9: MAINTENANCE AND SERVICE POLICY
WARNING: The 120H/L High-Low Flow
AIR/OXYGEN BLENDER should be serviced or calibrated by a qualified and trained service technician.
The 120H/L High-Low Flow AIR/OXYGEN BLENDER should be subject to a regular maintenance and service program, including periodic accuracy checks between normal overhauls. Although the frequency of these tests will vary depending on degree and severity of service, it is recommended that they be performed at least once every six (6) months under the best of conditions.
Elastomer components such as diaphragms and O­Rings are designed to function satisfactory for a minimum of two (2) years. The need for cleaning and replacement depends on gas line conditions and is indicated by the blender not meeting its specified performance. The blender may malfunction due to excessive dirt and debris. OHIO MEDICAL recommends that unit be overhauled every two (2) years. Elastomer components will not function indefinitely, and the probability of their causing malfunctions increases progressively after two (2) years of service.
Page 31 of 39
L 4.2.3-96 Rev 2
SECTION 10: DISPOSAL INSTRUCTIONS
This device and its packaging contain no hazardous materials. No special precautions need to be taken when disposing the device and/or its packaging.
Page 32 of 39
L 4.2.3-96 Rev 2
SECTION 11: REPLACEMENT PARTS
Part Number
Description
Quantity
Required
10-0700-020001
Proportion Block
1
10-0700-040002
Cover Plate – Dark Grey
1
10-0700-020002
Diaphragm Block
4 10-0700-030005
Alarm Sleeve Spring
2 10-0700-020003
Bypass Seat Cap
1
10-0700-010001
Bypass Adjuster
2
10-0700-020004
Alarm Shuttle
1 10-0700-070001
Alarm Sleeve
1 10-0700-030001
Primary Check Valve Spring
1
10-0700-070002
Primary Check Valve Poppet
1
10-0700-070003
Check Valve Housing
1
10-0700-030002
Auxiliary Check Valve Spring
1 10-0700-070004
Auxiliary Check Valve
1
10-0701-070001
High flow Auxiliary Outlet Connector
1
10-0700-070006
Oxygen Inlet Center Connector
2
10-0700-070007
Oxygen Inlet DISS
1
10-0700-070008
Oxygen Inlet DISS Nipple
1 10-0700-080002
Starwasher
1
10-0700-030003
Alarm Cap Spring
1
10-0700-030004
Rear Cap Spring
1
10-0700-010002
Checkball Retainer
1 10-0700-070009
Air Inlet Connector
1
10-0700-010004
Front Seat Nut
1
10-0700-040009
Control Knob
1
10-0700-010005
Knob Collet
1
10-0700-010006
Control Knob Washer
1
10-0700-010007
Control Knob Nut
1
10-0700-040006
Knob Cover
1
10-0700-030008
Socket Head Cap Screw, 2.25”
4
10-0700-080003
Diaphragm Block Ball Bearing
4 10-0700-030007
Diaphragm Block Spring
1 10-0700-030009
Socket Head Cap Screw, 0.75”
8
10-0700-060001
O-Ring, Balance Blocks and Check Valve
10
10-0700-060002
O-Ring, Alarm Sleeve and Block Cap
1 10-0700-060003
Front Valve Seat O-Ring
1 10-0700-060004
Inlet/Outlet Connector O-Ring
3
Page 33 of 39
L 4.2.3-96 Rev 2
SECTION 11: REPLACEMENT PARTS
Part Number
Description
Quantity Required
10-0700-060005
Front Valve Stem O-Ring
1
10-0700-060006
Bypass O-Ring
2
10-0700-060007
Front Seat O-Ring
1
10-0700-060008
Alarm Cap O-Ring
2
10-0700-060009
Alarm Sleeve O-Ring
2
10-0700-060010
O-Ring, Connectors and Caps
4
10-0700-060011
Rear Seat O-Ring
1
10-0700-080004
Muffler Bleed
1
10-0700-080005
Foam Diffuser
1
10-0700-040003
Blender Alarm Cap
1
10-0700-060013
Check Ball
1
10-0700-060014
Duck Bill Check Valve
3
10-0700-040007
Step Washer
2
10-0700-080006
Inlet Cone Filter
2
10-0700-080007
Balance Block Diaphragm
2
The Quantity required column indicates the number of parts actually required for one 120H/L High-Low Flow Air/O2 Blender.
Maintenance Kit (P/N DAOBA-00000-05) is packaged with the actual quantity required.
Page 34 of 39
L 4.2.3-96 Rev 2
SECTION 11: REPLACEMENT PARTS
)
120H/L HIGH-LOW FLOW AIR/O2 BLENDER KIT (P/N DAOBA-00000-05)
Part
Description
Quantity
Number
Required
10-0700-130006
Caution/Warning Label
1
10-0700-080002
Starwasher
1
10-0700-030006
Control Knob Spring
1
10-0700-080003
Diaphragm Block Ball Bearing
4
10-0700-060001
Balance Blocks and Check Valve O-Ring
11
10-0700-060002
Alarm Sleeve and Block Cap O-Ring
3
10-0700-060003
Front Valve Seat O-Ring
1
10-0700-060004
Inlet/Outlet Connector O-Ring
4
10-0700-060005
Front Valve Stem O-Ring
1
10-0700-060006
Bypass O-Ring
2
10-0700-060007
Front Seat O-Ring
1
10-0700-060008
Alarm Cap O-Ring
2
10-0700-060009
Alarm Sleeve O-Ring
2
10-0700-060010
Connectors and Caps O-Ring
10
10-0700-060011
Rear Seat O-Ring
1
10-0700-060012
O2 Inlet O-Ring
1
10-0700-080004
Muffler Bleed
1
10-0700-080005
Foam Diffuser
1
10-0700-060013
Check Ball
1
10-0700-060014
Duck Bill Check Valve
3
10-0700-040007
Step Washer
2
10-0700-080006
Inlet Cone Filter
2
10-0700-080007
Balance Block Diaphragm
2
Page 35 of 39
L 4.2.3-96 Rev 2
SECTION 12: PRODUCT SPECIFICATIONS
120H/L HIGH-LOW FLOW AIR/O2 BLENDER
Size (not including inlet and outlet fittings)………… Height3 ½” (9cm)
Width– 2 ¼” (5.6cm) Depth– 4 ½” (11.5cm)
Weight…………………………………………….. 2.75 lbs (1.4 kg)
Gas Supply Pressure (Air & O2)……………………. 30 – 75 PSIG (2.11 kg/cm² - 5.27 kg/cm²). The
blender will maintain stated accuracy at supply pressures provided the differential between supply pressures does not exceed 10 PSIG (0.70 kg/cm²). Output flow rate will be diminished if either supply pressure is below 50 PSIG (3.52 kg/cm²) and will be increased if both supply pressures are above 50 PSIG (3.52 kg/cm²).
Knob Adjustment Range…………………………….
Primary Outlet……………………………………….
Primary Outlet Flow Range…………………………
Maximum Flow @ 60% knob setting, 50 PSIG (3.52 kg/cm²) inlet pressure………………..
Flow @ 21% or 100% knob setting, 50 PSIG (3.52 kg/cm²) inlet pressure………………..
Bypass flow (loss of air or O2) 50 PSIG (3.52 kg/cm²) inlet pressure
of remaining gas……………………………………..
Auxiliary Outlet……………………………………..
Auxiliary Outlet Flow Range………………………..
Accuracy – with inlet gases within 10 PSIG (0.70 kg/cm²) and each gas pressure greater than 30 PSIG (2.11 kg/cm²), but less than 75 PSIG (5.27 kg/cm²)………………………………….
21 to 100%
Bottom Port
15 to 120 LPM (no bleed flow)
>120 LPM
>90 LPM
>90 LPM
Right Side Port (Facing unit)
2 to 100 LPM (Bleed 10-12 LPM)
± 3% of full scale over the stated flow ranges (i.e., 3 percentage points at any reading
Page 36 of 39
L 4.2.3-96 Rev 2
SECTION 12: PRODUCT SPECIFICATIONS
120H/L HIGH-LOW FLOW AIR/O2 BLENDER
Alarm/Bypass Activation…………………………… 20 ± 2 PSIG (1.41 ± 0.14 kg/cm²) when inlet gas
pressures differ by a nominal 20 PSIG (1.41 kg/cm²) or more provided maximum pressure of either supply gas does not exceed 75 PSIG (5.27 kg/cm²) and minimum pressure of one supply gas remains at 40 PSIG (2.81 kg/cm²) or above. In other words, one supply gas must remain at 40 PSIG (2.81 kg/cm²) or above to provide enough gas pressure to operate the alarm in the event the other supply pressure falls to 20 PSIG (1.41 kg/cm²) or below. There will be no alarm or bypass if the control knob is set to 21% and source oxygen pressure is reduced or turned off. Similarly, if control is set to 100% there will be no alarm if air pressure is reduced or turned off. In either case, the unit will continue to deliver the selected concentration of 21% or 100%. There will be no alarm under condition of 20 PSIG (1.41 kg/cm²) or greater source pressure differential if unit is not in use (i.e. no output flow or bleed flow).
Alarm Sound Generator…………………………….. Vibrating Reed
Alarm/Bypass Reset………………………………… When inlet pressure differential is 6 PSIG (0.42
kg/cm²) or less.
Pressure Drop……………………………………….. Less than 6 PSIG (0.42 kg/cm²) at 50 PSIG (3.52
kg/cm²) inlet pressure and 40 LPM.
Page 37 of 39
L 4.2.3-96 Rev 2
SECTION 13: TROUBLESHOOTING
PROBLEM
POTENTIAL CAUSE
CORRECTIVE ACTION
Oxygen concentration discrepancy between blender setting and analyzer (greater than 3%)
1. Flow requirements are outside the specified LPM range
2. Analyzer out of calibration
3. Blender is out of calibration
4. Low flow bleed muffler is obstructed causing restriction of fixed bleed
5. Gas supply contaminated
6. Air entrained into circuit by ventilator or accessory device.
1. Correct the flow. Verify that the
correct outlet port is being used. Each outlet port has a different flow range. Auxiliary outlet (2 LPM or more). Primary outlet (15 LPM or more)
2. Calibrate analyzer
3. Recalibrate or service further as
necessary.
4. Remove obstruction and verify
bleed flow is within tolerance.
5. Correct the contaminated gas
supply. Check source gases with calibrated O2 analyzer to confirm O2 is 100% and AIR is 21%
6. Contact Ohio Medical for
further information.
Alarm Sounding
1. Inlet pressure differences of 20 PSI (1.41 kg/cm2) or more.
2. Alarm module is not calibrated properly.
3.Inlet gas contamination alarm module malfunction.
1. Correct the pressure difference.
2. Recalibrate or service further as
necessary.
3. Disassemble, clean, reassemble,
calibrate, install inlet filter/water trap on air line, and correct cause of gas contamination.
Page 38 of 39
L 4.2.3-96 Rev 2
SECTION 13: TROUBLESHOOTING
PROBLEM
POTENTIAL CAUSE
CORRECTIVE ACTION
120H/L High-Low Flow Air/O2 Blender in bypass- no alarm
1.Reed plate is improperly installed or damaged.
2.Alarm gas orifice is obstructed.
1.Remove and replace.
2.Remove obstruction from orifice. If problem persists, contact Ohio Medical for repair.
120H/L High-Low Flow Air/O2 Blender Accurate only when inlet gas pressures are equal
1. Balance module not functioning properly.
2.Both air and oxygen gas sources are below 30 PSIG.
1.Disassemble balance module, clean, replace diaphragm(s). Reassemble and test.
2.Correct the low pressure condition.
Page 39 of 39
L 4.2.3-96 Rev 2
SECTION 14: WARRANTY
PRODUCTS MANUFACTURED BY TENACORE HOLDINGS INC. ARE WARRANTIED TO BE FREE FROM DEFECTS IN MATERIALS AND WORKMANSHIP
The liability of OHIO MEDICAL under this warranty is limited to replacing, repairing or issuing credit. Determination will be made by OHIO MEDICAL. Such determination will be made in relation to the parts that become defective or fail to meet published specifications during the warranty period. OHIO MEDICAL will not be liable under this warranty unless (A) OHIO MEDICAL is promptly notified in writing by Buyer upon discovery of defects or failure to meet specifications; (B) the defective unit or part is returned to OHIO MEDICAL, transportation charges prepaid by Buyer; (C) the defective unit or part is sent by buyer to OHIO MEDICAL prior to the last day of the warranty period; and (D) OHIO MEDICAL’s examination results in a determination that the unit has suffered such defects or failures as a result of misuse, neglect, improper installation, unauthorized repair, alteration or accident.
OHIO MEDICAL warranties as described above and set forth shall not be accretive, diminished or affected by, and further no obligation or liability shall arise or grow from the disclosure of technical advice or service by OHIO MEDICAL or its agents in connection with Buyer’s orders.
LIMITATIONS OF LIABILITIES
In no event shall OHIO MEDICAL be liable to Buyer for loss of profits, loss of use, consequential damage or damages of any kind based upon a claim of breach of warranty, other than the purchase price of any defective product covered within.
This warranty does not cover normal maintenance such as cleaning, adjustment or lubrication and updating of equipment or parts. This warranty shall be void and shall not apply if the equipment is used with accessories or parts not manufactured by TENACORE or authorized for use in writing by OHIO MEDICAL, or if the equipment is not maintained in accordance with a prescribed schedule of maintenance.
The warranty stated above shall extend for a period of six months from the day of delivery, with the following exceptions:
1. Elastomeric components and other parts or components subject to deterioration over which OHIO MEDICAL has no control are warranted for sixty (60) days from date of receipt.
The foregoing is in lieu of any other warranty, expressed or implied, including, without limitation, any warranty of merchantability, except as to title, and can be amended only in writing by a duly authorized representative of OHIO MEDICAL.
1111 Lakeside Drive,
Gurnee, IL 60031
www.ohiomedical.com
Loading...