EFFORTSONALINEOFNON-LUBRICATEDAIRCOMPRESSORSANDVACUUMPUMPSANDANICHE
WASFOUNDSERVICINGTHEPRINTINGPRESSINDUSTRY. UTILIZINGTHISEXPERTISE, THECOMPANY
BEGANTOSUPPLYCUSTOMPUMPINGPACKAGESFORAVARIETYOFAPPLICATIONSANDINTHE
EARLY 1950’S SQUIRE-COGSWELLSOLDITSFIRSTHOSPITALVACUUMPUMPPACKAGE, THE
CORNERSTONEOFTODAY’SBUSINESSESANDPRODUCTLINES.
IN 1962, MR. HERBERT SCHIFTER, MR. DEVEREAUX’SSON-IN-LAWTOOKOVERAS PRESIDENT.
MR. SCHIFTER WAS THE ORIGINALINVENTOROFTHE “DIAMOND ONE” GASOUTLETWHICHHE
ERIESOFMEDICALGASPIPELINEEQUIPMENTINCLUDINGTHE MEDICAL GAS OUTLETSAND
EDICAL GAS ALARM SYSTEMS.
M
SQUIRE-COGSWELL/AEROS INSTRUMENTS INC. ARETHEONESOURCEFORALLMEDICALGAS
PUMPING, MEDICALGASPIPELINEEQUIPMENTANDMEDICALASPIRATIONEQUIPMENT. FROM
CUSTOMDESIGNEDPUMPINGSYSTEMSTOTHEMOSTMODERNLINEOFASPIRATIONEQUIPMENT
ANDMEDGASPIPELINEAVAILABLEANYWHERE, SQUIRE-COGSWELL–AEROSDOESITALL. NO
OTHERSUPPLIERSURPASSESOURCONTINUITYOFEXPERIENCEANDLEADERSHIPINTHEMEDICAL
GASPUMPING, PIPELINEANDASPIRATIONMARKETS.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 5
1.0 Responsibilities
The information contained in this Installation and Operation Maintenance Manual, pertains only to the
Ohio Medical Corporation medical gas alarm system. The medical gas alarm system will operate in
conformance with the descriptions outlined in this manual only when it is operated, maintained and
serviced properly and in compliance with the instructions contained herein.
1.1 Installer Responsibilities
All contents of the alarm are shipped in ISTA approved packaging. The alarm should be handled,
installed, and tested per the recommended practices in this manual. Should such repair or
replacement become necessary, contact Ohio Medical Customer Service (800-448-0770) for original
equipment replacement parts.
1.2 User Responsibilities
The alarm should be tested and examined periodically. Any parts which are found to be damaged,
corroded, contaminated, etc. should be replaced. Possible replacement items are listed on pages 26
to 27 in this manual. Should such repair or replacement become necessary, contact Ohio Medical
Customer Service (800-448-0770) for original equipment replacement parts.
2.0 System Description & Function
The Ohio Medical Corporation Medical Gas Alarm System is designed to continuously monitor the
status of source equipment and gas line pressure of medical gases. The system is comprised of Area
Alarm Modules which monitor and indicate the gas pressure or vacuum level of a specific gas at a
specific point in the hospital and Master Alarm Modules which monitor source equipment such as
oxygen systems, air compressors or vacuum pumps.
The Alarm System monitors gas line pressure with sensors. The Alarm’s Annunciator determines the
status of the pressure and the pressure is displayed on an area alarm module. The alarm module also
has trend light LED’s to show that the displayed pressure is either in a normal, caution, or a high or
low alarm condition.
The Alarm System also has a master alarm module to monitor 10 condition points from source
equipment. Each point represents a condition that the source equipment might have. A dry contact
version of the master alarm is also available to relay conditions of the source equipment to a building
management system.
More detailed descriptions of each module is explained on pages 5 through 8.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 8
2.2 Annunciator Panel
The Annunciator Panel is the main processing board for the Medical Gas Alarm System. The
Annunciator Panel accepts information by two different means. For gas pressure, the input
signal comes from the Sensor Module and goes to the annunciator to evaluate the pressure
being observed. This information is then transmitted directly to the corresponding Area Alarm
Module where the pressure or vacuum is displayed via the ribbon cable. For source
equipment, a condition is relayed to a point on a master alarm module first and then
transferred to the Annunciator via the ribbon cable. In normal operation, the Annunciator
continuously monitors all pressure, vacuum, and source equipment for alarm conditions.
When an alarm condition is detected, an audible alarm (approximately 80 decibels measured
at 3 feet) will sound and the appropriate red LED will illuminate from the corresponding Area
or Master Alarm Module. The audible alarm will sound under one of the four conditions listed
below.
■ When the pressure input increases above a specified high set-point
□ When the pressure or vacuum decreases below a specified low set-point
■ When an alarm condition is detected on the Master Alarm Module
■ When one of the Sensor Module’s wiring has become disconnected
There are two momentary push buttons on the Annunciator Panel, which are labeled as
“TEST” and “ALARM SILENCE/RESET”. To stop the audible alarm from sounding, press the
“ALARM SILENCE” button. However, the alarm will re-sound again after the specified repeat
time has elapsed unless the alarm condition has been cleared. When a new alarm condition
is detected from another alarm panel, the audible alarm will sound again. Each time the
“Alarm Silence” button is pressed the alarm repeat time will reset. To review if the alarm’s
displays and audible alarm are in working order, press the “TEST” button for approximately 3
seconds. While testing, all display segments and independent LED’s will be illuminated and
the audible alarm will sound.
The Annunciator has a relay for wiring an external buzzer to a remote location. This relay
energizes the external buzzer when the audible alarm sounds.
A communications port is provided on the Annunciator Panel that allows for integration with a
LonWorks® network system allowing the medical gas alarm system to communicate with a
LonWorks® capable building management system.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 9
2.3 Area Alarm Panel
The Area Alarm Panel provides a digital display of the actual gas
line pressure or vacuum being monitored. The module has three
different color trend lights to indicate three different conditions of
the gas pressure. These colors are green, yellow and red for
normal, caution, and high or low pressure indications
respectfully. The trend lights and how they correspond to range
in pressure is explained in more detail while programming in
Chapter 5.2 on page 21. When an alarm detects a “LOW” or
“HIGH” condition the corresponding red LED on the Area Alarm
Panel will be illuminated and an audible alarm will sound from
the Annunciator Panel. The red LED will stay illuminated until the
alarm condition has been corrected and the “Reset” button is
pressed.
The two LED’s above the colored trend lights indicates which
unit of measure is being displayed. The units are English
(Standard) or Metric units for pressure or vacuum accordingly
and can be changed to whichever units when programming.
The Area Alarm Module has dry contacts on the backside to
connect to a pair of contacts on a Master Alarm Module for
remote monitoring purposes.
2.4 Master Alarm Panel
The Master Alarm Panel monitors dry contacts from source
equipment or area alarm panels. Each individual Master Alarm
Panel can monitor up to 10 signals. Each pair of contacts is
either configured for normally open or normally closed operation.
The default mode is normally closed per NFPA code. The
Master Alarm panel may be easily labeled to indicate each of the
conditions being monitored. A Dry Contact Master Alarm Module
is also available for computer interface or service to a building
management system.
In operation, the unit continuously monitors for alarm conditions.
Except for the green “POWER” indicator, all Master Alarm Panel
LED lamps will be off when there are no alarm conditions
present. When an alarm condition is detected, the
corresponding LED on the Master Alarm Panel will flash and the
audible alarm from the Annunciator Panel will sound. The LED
will continue to flash until the alarm condition has been corrected
and the alarm has been reset. If two or more alarm condition
occur at the same time, the most recent alarm will flash and the
other existing alarms will remain in constant illumination. Any
alarm will remain lighted until the respective alarm condition has
been corrected and the alarm has been reset.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 10
3.0 Labeling Alarm Modules
Each application needs to describe the medical gas conditions and locations that are unique
to each installation. To provide clear and precise representation of this information, the area
and master alarm modules allocate a sufficient size area on their displays to label this
Information. To simplify in labeling the area or master alarms, please try to label these
assemblies prior to connecting leads to the dry contacts.
3.1 Labeling Area Alarm Modules
For ease of referencing what specific location the area alarm is representing, a window on
the bottom of each alarm is provided to label this information. See Labeling Module
Diagram on figure 3.
1. Make sure the alarm assembly is turned off prior to servicing.
2. Referencing figure 9, disconnect the cable ribbons leading to and from adjacent PCB/
Weldment assemblies.
3. Unfasten top and bottom hex nuts only. Do not unscrew the button head screws attaching the circuit board to the weldment plate.
4. Extract the area alarm PCB/Weldment assembly from the front panel assembly.
5. Either slide required label through pocket provided in mylar screen or affix required label
directly over screen per Figure 3.
6. Refasten the master alarm assembly to the front panel with the hex nuts.
7. Reconnect cable ribbons as demonstrated in figure 6 on page 14 of your manual.
3.2 Labeling Master Alarm Modules
For ease of referencing what specific alarm conditions are present, each LED on the master
alarm module can be labeled to identify what condition it is represent. See Labeling Module
Diagram on figure 3.
1. Make sure the alarm assembly is turned off prior to servicing.
2. Disconnect cable ribbons from master alarm board.
3. Unfasten top and bottom hex nuts only. Do not unscrew the button head screws
attaching the circuit board to the weldment plate.
4. Extract master alarm PCB/Weldment assembly from the front panel assembly
5. Either slide required label through pocket provided in mylar screen or affix required
label directly over screen per Figure 3. Note that the “MA 1” dry contact will activate
the top LED, the “MA 2” contact activates the second from the top, and then follows
on down in that consecutive order.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 12
4.0 Installation
For ease of installation, the Ohio Medical Corporation medical gas alarm is delivered in two
separate assemblies; the box assembly and the front frame assembly. All box assemblies
include the box itself, mounting brackets and hardware, the power supply, the terminal block
and the fuse holder module. When specified for “local” installation, per the customer
requirement, the appropriate sensor modules equipped with gas-specific DISS demand check
units and type K copper stubs are also part of the box assembly. When the customer has
specified a “remote” sensor installation, the sensors are not part of the box assembly and will
be shipped loose.
The front-frame assembly consists of the frame, hinge, the Annunciator module and the
individual Area or Master Alarm modules as specified by the customer. The front-frame
assembly is fully wired and tested at the factory. To insure product functionality, leave the
front panel in the anti-static bag(s)) and packaging until it is ready to be installed.
4.1 Mounting Instructions (Prior to Drywall & Plaster Installation)
The box assembly includes adjustable brackets to allow for varying wall thickness and for
ease of installation.
1. Remove the power supply cover.
2. Install the brackets to the left and right sides of the box assembly using the hardware
provided. Do not tighten hardware at this time.
3. Adjust the brackets so that the front edge of the box assembly will be flush with the
finished wall.
4. Tighten the hardware securing the mounting brackets to the box assembly.
5. Mount the box securely to the wall studs at the desired height.
6. The copper pipe stubs may now be silver-brazed to the appropriate gas pipeline. The
pipe stubs are clearly marked with color coded labels indicating the correct gas. Take
extreme caution not to overheat the DISS check units during the brazing process by
wrapping the lower part of the pipe stub in a cool, damp cloth. In remote sensor
installations, the installer must braze a tee with a ¼” NPT Female port at a convenient
location. The Remote Sensor Module can then be installed into the appropriate gas
pipeline.
7. Per NFPA-99 and/or local standards, pressure testing of all joints must be inspected for
leaks. When spraying leak-testing solution on the check units adjacent to the sensor
modules, it is recommended to do so conservatively and with a solution consisting of
liquid dishwashing detergent and distilled water only. Verify that each gas stub is
connected to a pipeline which carries the same gas labeling as the sensor module does.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 16
4. Connect the DC power from the power supply to the Annunciator Panel. The five-wire
connection cable is used to make this connection. This cable can only be installed
one way, it makes the connection from on the power supply to position “J8” of the
Annunciator Panel. This connection will provide the voltage needed for operation of all
panels and sensors.
5. After all interconnections have been made, the AC power may be connected to the
power terminal strip. The hot wire is connected to the fused terminal block, the neutral
wire to the neutral terminal block and the ground wire is connected to the green/yellow
terminal block.
6. Re-install the power supply cover.
4.3 Area Alarm Remote Output Signal Connections
Included on each Area Alarm Panel are contacts to be used to transmit an alarm condition to
a Master Alarm Panel or to other slave alarm device, if required. These contacts may be
used to transmit either a “High Alarm” or “Low Alarm” condition and may be used in either a
normally open or normally closed state. The usual connection scheme when connected to a
master alarm panel is to make these connections in the normally open mode using the relay
connection terminals marked “N.O.” and “C”. These contacts are clearly marked on the Area
Alarm circuit board. See Figure 7 for more details. NOTE: Do NOT wire the sensor modules
directly to the area alarm modules
Output Contacts to
a Master Alarm
for High Alarm
Condition.
Contact Positions:
1. Normally Open
2. Normally Closed
3. Common
Input connection
from the
annunciator or the
previous area alarm
module
Output Contact Positions
1 2 3 4 5 6
Output Contacts to a
Master Alarm for Low
Alarm Condition.
Contact Positions:
4. Normally Open
5. Normally Closed
6. Common
Output connection going
to the next area alarm
module or a master
alarm module.
Figure 7: Area Alarm Module
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 17
4.4 Master Alarm Connections from Source Equipment
Each Master Alarm Panel may accommodate up to 10 alarm inputs. Dry Contacts from
source equipment are connected directly to the Master Alarm Panel circuit board at positions
“MA1” through “MA10”. Each of these contact points is clearly marked on the circuit board
and includes both positive (+) and negative (-) terminals. The dry contacts from the source
equipment must be rated for at least 12 VDC.
CONSIDEREDTOBEPARTOFTHESYSTEMCONFIGURATIONUNLESSTHEEQUIPMENT
INQUESTIONCOMPLIESWITHTHESTANDARDFORHOSPITALANDSIGNALINGAND
NURSECALLEQUIPMENT, UL 1069
The “Maintenance Mode” switch needs to be on to disable the audible alarm while wiring
contacts to the master alarm module. To make the Maintenance switch active, move the
switch at position “S11” on the Master Alarm Module toward the row of switches labeled “S1”
to “S10” as shown in Figure 8A. The maintenance lamp will be illuminated on the front of the
panel. Moving the switch away from the row of switches will restore the Master Alarm to
normal operation.
Switch settings on the Master Alarm Panel must be set to configure each alarm point for
normally open or normally closed operation. As shown on Figure 8A, the switches are on
the headers labeled “S1” through “S10”. The switches at position “S1” corresponds to the
alarm connected to the first pair of contacts “MA1”, “S2” to “MA2”, “S3” to “MA3”, and so on.
If one of the S1 through S10 switches is positioned toward the “S11” switch, the circuit loop is
normally open. It will cause an alarm when a closed contact is applied to the input terminal
strip (MA1-MA10). The switches at positions S1 & S3-S10 are shown in this position in
Figure 8A. This is the factory default setting. Any pair of contacts that are not connected to
source equipment should have their corresponding switch positioned toward the
maintenance switch.
If one of the S1 through S10 switches is positioned away from the S11 switch, the circuit loop
is normally closed. The contact will create an open circuit loop when an alarm condition
exists. The S2 switch is shown in this position in figure 8A. This is the setting for input
contact that is closed during normal operation but open during an alarm condition.
Per NFPA 99, information relayed from the source equipment is to be sent to two separate
master alarm panels at separate locations inside the medical facility. Wiring from the
source equipment is to be ran in parallel to each master alarm module and can not to be
slaved from one master alarm panel to the another.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 18
4.5 Dry Contact Master Alarm Output Connections
CAUTION: When testing the output relays, make all connections with the power off.
Shorting the terminals together while the relay is energized can cause permanent
damage to the relay.
Dry Contact Master Alarm Panel is similar to the Master Alarm Panels except that they also
provide 10 alarm outputs that connect to building management systems. These additional
connections are connected directly to the Dry Contact Master Alarm Panel circuit board at
position “J1” through “J10” as shown in Figure 8B. Each of these pairs of contacts are clearly
marked on the circuit board and includes both positive (+) and negative (-) terminals. The dry
contacts going to the other alarm devices or a building management system must be rated for
at least 12 VDC.
Figure 8A:
Wiring from Source Equipment to a
master alarm position
(On either master alarm modules)
Wiring from the source equipment to the second pair of
contacts on the bottom row (MA2) on the master.
Figure 8B:
Wiring from a master alarm Position to a
Building management System
(On Dry Contact Master Modules Only)
Second pair of
contacts on
the top row
(J2) is wired to
a building
management
system
Corresponding
switch to the second
pair of contacts, S2,
on the master
showing “Normally
Closed Circuit Loop”
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Maintenance Switch, S11, in
the “ON” position
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Input Connection from
Annunciator or previous alarm
module.
Output Connection to the next
master alarm module.
Page 19
4.6 Remote Audible Alarm Installation
The “medical gas alarm system includes two provisions to provide remotely mounted audible
or visual alarms which will operate in parallel with the audible alarm on the Annunciator Panel
and will sound each time that the alarm sounds. The connections for the remote alarm are on
the Annunciator panel.
The +24V and ground terminals allow a buzzer similar to the one mounted on the
Annunciator Panel to be used as an remote buzzer. If a different type of buzzer is desired for
external mounting, the relay contacts could be used in conjunction with another power source
to supply the necessary power.
To connect a remote buzzer, which is, similar to the one installed on the Annunciator
panel:
1. Connect the +24V terminal to the positive (+) power terminal of the remote buzzer.
2. Connect the negative ( - )terminal of the buzzer to the COM terminal of J7.
3. Connect a jumper across the N.O. terminal of J6 to the GND terminal of J7.
To connect an independent 24V relay-operated device:
1. Connect the positive (+) terminal of the external power supply to the positive (+) terminal
of the remote buzzer.
2. Connect the ground terminal of the external power supply to the normally open (N.O.)
terminal of “J6” on the Annunciator panel.
3. Connect the common (COM) terminal of “J7” on the Annunciator panel to the negative
terminal (-) of the external buzzer.
Figure 9: Annunciator
Sensor
Plug
Connection
Power
Supply
Connection
Programming Buttons
LonWorks®
Connection
Remote Audible Alarm
Connection (J7)
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Remote Audible Alarm
Connection (J6)
Page 20
4.7 Initial Start-Up Checks
Perform the following checks PRIOR to turning the power switch to the “ON” position.
1. For Area alarms, the number of sensors installed must match the number of area
alarm
modules.
2. Gas lines must have normal pressure.
3. Alarm box and front panel assemblies must be free from plaster dust, tools, and
anything else that is not an Ohio Medical component or is not necessary for the
proper operation of the alarm.
5.0 Programming
NOTE: Ohio Medical Corporation Medical Gas Alarms, which are purchased as factory
assembled, do NOT
the default settings.
need to be field programmed unless the user desires to change
If the standard default values do not match the desired configuration, the user can change
program configuration data and save them. All programming is accomplished by using the 3
programming buttons on the Annunciator circuit board. The buttons are labeled (proceeding
from left to right) “UP”, “S” (Save) and “E” (Exit).
To enter the annunciator’s program, press the 2 outside keys as shown in Figure 10A for
about 3 seconds. The upper left-hand module will display “SEnr” as shown in figure 10B.
Follow the steps in the Programming Sequencing Chart on Page 20 till the desired changes
are completed and saved. Tables A & B on page 21 also provide information such as the
range setting and the default settings that are used for a certain gases. An example of
changing a program setting is shown on page 22 as a programming reference.
Advances to next moduleAdvances to step 6 for theExits Program, goes to step 11
changes type of unitsSaves Selection, returns to step 6Exits Module, returns to step 5
skips routine & goes to step 7Enters Routine, goes to step 6AExits Module, returns to step 5
Skips routine & goes to step 9Enters Routine, goes to step 8AExits Module, returns to step 5
changes parameter typesaves selection, goes to step 8B or 8CExits Module, returns to step 5
changes HSPsaves selection, returns to step 8Exits Module, returns to step 5
changes parameter typesaves selection, goes to step 9B or 9CExits Module, returns to step 5
no change, shows default HSPsaves selection, returns to step 8Exits Module, returns to step 5
Skips routine & goes to step 10Enters Routine, goes to step 9AExits Module, returns to step 5
changes LSPsaves selection, returns to step 9Exits Module, returns to step 5
no change, shows default LSPsaves selection, returns to step 9Exits Module, returns to step 5
Skips routine & returns to step 6Enters Routine, goes to step 10AExits Module, returns to step 5
Programming Sequence
4
1
3
2
PA" for Metric Pressure Units [see Table B on page 21
et Point Programming Routine
larm High Set Point Programming Routine
Setting type - "FAC" for utilizing Factory default Setting
or "USr" for user to utilize own perferred Setting
HSP, if "FAC" was selected in step 8A, See Tables A or B for Default HSP
HSP, if "USr" was selected in step 8A, See Tables A or B for Default HSP
Setting type - "FAC" for utilizing Factory default Setting
or "USr" for user to utilize own perferred Setting
* = Number of Gases being Programmed (1 to 8)
To enter program to modify parameters, press the outside 2 buttons on the annunciator simultaneously (as shown in Fig. 9A) for about 3 seconds.
AlarmProgramming Field Description
1SEnrInitial Screen (No Changeable parameters)Advances to step 2Advances to step 2Exits Program, goes to step 11
2SEn*
StepDisplay
3AlrnProgram Name (No Changeable Parameter)Advances to step 4Advances to step 4Exits Program, goes to step 11
(*- module specified, I.E. PSn1 = First module)or returns to the 1st modulearea alarm module specified
Units of Measure for Pressure Programming Routine
or
Area Alarm Module Indication
4Ar**** = alarm resound Interval: ranging 0-99 min. (30 min. default) increases time by 1 min. Saves Selection, Advances to step 5Exits Program, goes to step 11
5PSn*
Types of Units: "PS" for English Pressure Units [see Table A on Page 21]
6UnS
6APS
or See Table B if Program is set for Metric Units in Step 6
7rAn9Corresponding Sensor's Pressure Range Programming RoutineSkips routine & goes to step 8Enters Routine, goes to step 7AExits Module, returns to step 5
8HSP
8AFAC
7A****range setting - See Table A if Program is set for English Units in Step 6 changes pressure rangesaves selection, returns to step 7Exits Module, returns to step 5
8B****
8C****
Alarm Low S
9LSP
9AFAC
LSP, if "FAC" was selected in step 9A,, See Tables A or B for Default LSP
LSP, if "USr" was selected in step 9A, See Tables A or B for Default LSP
[See Chapter 5.3 for Calibration Precautions before continuing.]
NOTES:
1) Number of gases being programmed should match the number of sensors and area alarm modules used in the alarm for proper operation.
2) Modules progress top to bottom on each bay. Bays progress left to right.
10A****Calibration Value (a fixed value ranging between 190 - 210) no changeable parameter
11SAuEExit Alarm Programno changeable parameterExits program with changes saved
3) Steps 6 - 10 are displayed on module selected in step 5
4) Only enter routine if existing alarm is to monitor an additional gas or annunciator is being replaced.
5) If gas module has been preprogrammed from the factory, Hit "Exit" to avoid changing setting.
Page 22
5.2 Default Settings Tables
These 2 tables are showing which range setting, the default High Set Point (HSP), and the
default Low Set Point (LSP) that the annunciator should be programmed to for the gas
specified. Each table shows the default parameters when using either English (standard) or
Metric pressure units. The tables also provide the working pressure range that the LSP and
the HSP can be set for in the “USr” function sub-routine. When changing the set points, the
following limits exist:
1. When changing the HSP number, a value lower than the LSP can never be selected.
Likewise for the LSP number, a value higher than the HSP can not be selected.
2. The “Normal Pressure” may not be the midpoint between the LSP and the HSP settings.
3. For Vacuum and WAGD gases only, HSP can be set, but the alarm condition will not be
acknowledged if the vacuum reaches the HSP setting.
Table A - English Pressure Units (Step 6)
RangeWorkingDefaultDefaultDefaultGas Types
SettingPressureHSPLSPNormal
(Step 7)Range(Step 8)(Step 9)PressureWith Range Type
2
Corresponding
150-30" Hg181215" HgVACUUM or W.A.G.D.
Vacuum
see note 3
Vacuum
1000-100 PSI604050 PSIOXY, N2O, AIR, or C O2
2500-250 PSI195140160 PSIN ITROGEN
Table B - Metric Pressure Units (Step 6)
RangeWorkingDefaultDefaultDefaultGas Types
SettingPressureHSPLSPNormal
2
Corresponding
(Step 7)Range(Step 8)(Step 9)PressureWith Range Type
1000-776 mmHg414305381 mmHgVACUUM or W.A.G.D.
Vacuum
see note 3
Vacuum
7000-689 kP a414276345 kP aOXY , N2O, AIR, or CO2
17200-1723 kP a13459651103 kPaN ITROGEN
Table C – Relationship between the Area Alarm Module Trend Lights and the Pressure Range
Low Normal High
Pressure Pressure Pressure
Low AlarmLow Ala rm CautionHigh Alarm Ca utionHi gh Alarm
Red LEDYel l ow LEDYellow LEDRed LED
LSP11% Below11% AboveHSP
MidpointMidpoint
Default Range: 11-19%Deafult Range: 11-19%
Below MidpointAbove Midpoint
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
For this example, the second gas is nitrous oxide and the high set point is the factory default of 60 PSI
and needs to be changed to 65 PSI.
1. Press the “UP” and “E” keys at the same time for 3 seconds to enter program.
(“SENr” is displayed and program is at step 1)
2. Press the “S” key to advance to step 2.
3. If the number in the last digit displayed matches the number of sensors and area alarm modules
the alarm is using, press the “S” key.
4. Press the “S” key to advance to step 4.
5. Press the “S” key to advance to step 5 if resound interval is not being changed.
6. Press the “UP” key once till “PSn2” is displayed.
7. Press the “S” key to advance to step 6, which “UNS” will be displayed on the second module.
8. Press the “UP” key twice to bypass changing both units of measure (step 6) and sensor pressure
range (step 7) to reach the High Set Point programming routine (step 8).
9. Press the “S” key to enter the HSP routine. “FAC” will be displayed (step 8).
10. Press the “UP” key once till “USr” is displayed (step 8A).
11. Press the “S” key to advance to step 8C. “60” is displayed.
12. Press the “UP” key 5 times till “65” is displayed.
13. Press the “S” key to temporarily save the HSP selected and return to step 8.
14. Press the “E” key to return to step 5.
15. Press the “E” key to advance to step 11. “SauE” is displayed.
16. Press the “S” key to permanently save all the changes and exit the program.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 24
5.5 Review Area Alarm Programmed Parameters
To review the programmed parameter of area alarms, press “S” and the “UP”
simultaneously for approximately half a second and the programmed information will then be
displayed. To cancel reviewing at any time, press the “Test” or the “Reset” buttons on the
annunciator. The first five items of the review program show up on the first area alarm
module as listed below. Information for each individual gas is then displayed on its
respective area alarm module.
Displaying Information Programming
Order Description Step
1. The program revision N/A
2. The revision date of the program N/A
3. The number of sensors being used in the alarm 2
4. The program name, which is “Apn1” N/A
5. The alarm interval, which is displayed in this format: “AI##” 4
The following information is displayed for each area alarm.
Displaying Information Programming
Order Description Step
6. Units of measure (“PS” or “PA” is shown) 6
7. Range (first shown is “rAn9” and then variables shown 7
table A or B in chapter 5.2)
8. High set point (first shown is “HSP”, second is either “FAC” 8
or “Usr” and then the high set point value being implemented)
9. Low Set Point (First shown is “LSP” , second is either “FAC” 9
or “Usr” and then the low set point value being implemented
10. Calibration 10
5.6 Master Alarm Programming Mode
NOTE: NO MASTER ALARMPROGRAMMINGISNECESSARYUNLESSTHEUSERDESIRESTO
1. The program mode is accessed by simultaneously pressing and holding the “UP” and
“E” buttons. The Master Panel programming mode is only displayed on Master Alarm
Panel 1. The maintenance LED will flash and the number of LED lamps lighted on the
Master Alarm Panel 1 will indicate the alarm repeat time. Each LED lamp represents an
alarm repeat time of approximately 5 minutes. The maximum alarm repeat time is 50 and
the minimum is 0. When no lamps are illuminated the repeat time is zero minutes
meaning that the repeat feature is disabled.
2. When the desired number of minutes are selected, then press “E” and all of the lights on
Master Alarm Panel 1 will flash.
3. To save the changes before exiting, press the “S” button or press the “E” button again to
exit without saving any changes
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
.
Page 25
6.0 LonWorks® Capabilities
The Medical Gas Alarm System is fully LonWorks® compatible.
In cases where the building management system is LonWorks® capable, the Medical Gas
Alarm System may be incorporated into the overall building management system by the
building management system installer or systems integrator.
Connections to a LonWorks® capable building management system are made with two wires
connected to the appropriate terminals at position “J4” on the Annunciator panel. Device-
specific documentation required by your LonWorks® integrator is available from the factory
upon request.
7.0 Specifications Summary
■ Cable Requirement for
a. Area alarm module to remote sensor;
b. Master alarm module to source equipment
c. Dry contact master alarm module to building management system
Distance: Maximum 5,000 feet (1,500 m)
Cable: 18 - 22 gauge shielded, twisted pair. Multi-conductor twisted pair cable can be
used when connecting multiple sensors.
■ UL & CSA Listed
■ NFPA 99 compliant
■ The Medical Gas Alarm System complies with the EMC (electromagnetic compatibility)
requirements of all of the following:
■ FCC Part 15
■ CISPR11 (EN55011)
■ EN60601-1-2
■ Canada, Department of Communications
■ LonWorks® Interface node to communicate to compatible building management system
■ Dry Contact for interface with building management system
q 12 VDC, 120 mA rating
■ Power Rating: 100-240VAC, 50/60HZ, 1.0A
■ Replacement Fuse: ¼” x 1-1/4” (6.3 x 32 mm), Time Delay, 1.25 amp max.
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 26
8.0 Troubleshooting
Display does not light, no “POWER” indications
■ Check electrical connections
■ Check fuses
■ Check power supply output
No indication on a single panel of a multi-panel alarm
■ Check electrical connections
Audible alarm does not sound or Repeat Alarm does not sound
■ Check programming to insure that the resounding interval is not set to zero
■ Replace Annunciator panel
Display reads in incorrect units
■ Check programming, Units of Measure
■ Check wiring connection from sensors
Pressure Reading fluctuates back and forth
■ Clean out transducer on sensor board
■ Replace sensor board
One or more displays have readings far from the expected pressure or vacuum
■ Sensors are connected to the Annunciator in the wrong sequential order
■ Check programming, Sensor’s Pressure Range is not correct for sensor board being used
“SENr” is displayed when alarm is turned on
■ Check if the number of sensor wired to the annunciator matches the number area alarm
modules daisy chained to the annunciator in the alarm.
■ Check the wiring from the annunciator to the corresponding sensor board for bad
connections or if wiring is in bad repair.
■ Check Ribbon Cables are connected firmly to modules and are in good repair.
“CAL” is displayed on an area alarm module
■ Check if corresponding gas line has pressure.
■ Check if the sensor module has been disconnected from its check unit.
■ Check wiring from the annunciator to the corresponding sensor board for bad connections
or if wiring is in bad repair.
■ Replace sensor module
Calibration value is not between “190” to “210”.
■ Check if corresponding gas had pressure in the line while calibrating the annunciator.
■ Check if the potentiometer screw on the sensor board has been tampered with.
■ Replace sensor board
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 27
9.0 Cleaning
The surface of the MEDICAL Gas Alarm is made from a one piece mylar with minimum
penetrations. It may be cleaned/disinfected using common surface cleaners and a soft cloth.
261834 Mounting Bracket Kit
261580 Restraining Cable
261253 Bushing/Copper Tube Assembly
261371 Locknut for Bushing/Tube Assembly
263222 Hinge, Piano, for 4-bay or 6-bay boxes
261584-X Front Panel Assembly, 4 bay Alarm
261585-X Front Panel Assembly, 6 bay Alarm
ELECTRICAL COMPONENTS
263299 Replacement Fuse
263506 Power Supply
261874 Power Supply Cable Assembly
261508 Ribbon Cable
261634 Wire, Local Sensor
261599 Ferrite, Local Sensor Wiring
261998 Plug, Connector, Single Position
263267 Plug, Connector, 8 Position
754042 Switch, I/O, Rocker, Panel Mountable, DPST
263297 Fuse Holder, W/ 1 Amp fuse
263519 Line Filter Assembly
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 28
SENSOR MODULES
261849-1 Local Sensor w/Demand Check, OXYGEN
261849-2 Local Sensor w/Demand Check, VACUUM
261849-3 Local Sensor w/Demand Check, N2O
261849-4 Local Sensor w/Demand Check, AIR
261849-5 Local Sensor w/Demand Check, N2
261849-6 Local Sensor w/Demand Check, EVAC
261849-7 Local Sensor w/Demand Check, CO2
261752 Demand Check Unit, EVAC
261753 Demand Check Unit, CO2
261729 Demand Check Unit, AIR
261730 Demand Check Unit, N2O
261731 Demand Check Unit, N2
261732 Demand Check Unit, VACUUM
261733 Demand Check Unit, OXYGEN
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 29
11.0 Limited Warranty
1. OHIO MEDICAL CORPORATION warrants the Medical Gas Pipeline Equipment to be free from functional
defects in material and workmanship for a period of twenty-four (24) months from the date of shipment or twelve
(12) months from the date of start-up, whichever occurs first. Within said period Ohio Medical Corporation will repair
or replace any part or component which is proven to be defective in either material or workmanship.
2. To obtain service within the warranty period, first contact the Ohio Medical Service Department.
3. Ohio Medical Corporation's responsibility under this warranty shall be limited to providing at Ohio Medical
Corporation's sole discretion, new or replacement parts to replace any component found to be defective within the
warranty period. Installation of user replaceable items will be the user's responsibility.
4. Labor to repair any part or component proved to be defective within the warranty period will be provided at no
charge for any item returned to our factory adequately packaged and insured with shipping costs prepaid.
Standard surface freight shipping cost to return the repaired part or component to the user will be paid by Ohio
Medical Corporation.
a. Before returning any part or component to the factory, proper return authorization must first be obtained
from Ohio Medical Service Department.
b. The user will be required to issue a purchase order for replacement items. Upon receipt of the defective
items, Ohio Medical Corporation will issue a credit to the user in the amount equal to the purchase order.
5. This warranty is valid only when the product has been properly installed according to Ohio Medical Corporation
specifications, used in a normal manner and serviced according to factory recommendations. The warranty does
not cover failures due to damage which occurs in shipment or failures which result from accidents, misuse, abuse,
neglect, mishandling, alteration, misapplication or damage that may be attributable to acts of God.
6. Ohio Medical Corporation shall not be liable for incidental or consequential damages resulting from the use of this
product. There are no expressed or implied warranties which extend beyond the warranties set forth above. Ohio
Medical Corporation makes no warranty of merchantability or fitness for a particular purpose to equipment and to its
parts and components.
7. THE CONDITIONS OF THE BUYER'S RESPONSIBILITY ARE:
1. The equipment is stored properly before installation;
2. The equipment is installed according to Ohio Medical Corporation's specifications and installation
procedures;
3. The equipment is properly maintained and not altered unless by an authorized representative of Ohio
Medical Corporation.
573078 (REV.2) 580273
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 30
NOTES:
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
Page 31
NOTES:
Ohio Medical Corporation., 1111 Lakeside Drive, Gurnee, IL 60031-4099 (800/448-0770) www.ohiomedical.com
Medical Gas Alarm Manual ■ 255098 (Rev.8) 07/08
For 24/7 Technical Support,
call 847-855-6234 for assistance
LonWorks® is a registered trademark of the Echelon Corporation
OHIO MEDICAL CORPORATION
1111 Lakeside Drive
Gurnee, IL 60031-4099
Phone: 800-448-0770 or 847-855-0500
Fax: 847-855-6300
Email: info@ohiomedical.com
http://www.ohiomedical.com
Medical Gas Alarms
255098 (Rev.8) 07/08
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.