Ohaus Corporation, 29 Hanover Road, Florham P ark, New Jers ey , 07932, USA
Declaration of Conformity We, Ohaus Corporation, declare under our sole responsibility that the balance models listed below marked with
“CE” - are in conformity with the directives and standards mentioned.
Konformitätserkärung Wir, die Ohaus Corporation, erklären in alleiniger Verantwortung, dass die untenstehenden Waagentypen,
gekennzeichnet mit “CE” - mit den genannten Richtlinien und Normen übereinstimmen.
Déclaration de conformité Nous, Ohaus Corporation, déclarons sous notre seule responsabilité, que les types de balance ci-dessous cité munis de la mention «CE» - sont conformes aux directives et aux normes mentionnées ci-après.
Declaración de Conformidad Nosotros, Ohaus Corporation, declaramos bajo responsabilidad exclusiva que los modelos de balanzas
indicados a continuación - con el distintivo ,CE’ - están conformes con las directivas y normas citadas.
Dichiarazione di conformità Noi, Ohaus Corporation, U.S.A, dichiariamo sotto nostra unica responsabilità, che i tipi di bilance specificati di
seguito - contrassegnati con la marcatura “CE” - sono conformi alle direttive e norme citate.
Balance Type/Waagentyp/Type de balance/Modelo de balanza/Tipo di biliancia Navigator
Marked with:DirectiveStandard
Gekennzeichnet mit:RichtlinieNorm
Munis de la mention:DirectiveNorme
Con el distintivo:DirectivaNorma
Contrassegnati con laDirettivaNorma
Marcatura:
EU 73/23 Low Voltage
EU 73/23 Niederspannung
EU 73/23 Basse tension
EU 73/23 Baja tensión
EU 73/23 Bassa tensione
EU 89/336, 92/31, 93/68
Electromagnetic compatibility
EU 89/336, 92/31, 93/68
Elektromagnetische Verträglichkeit
EU 89/336, 92/31, 93/68
Compatibilité électromagnétique
EU 89/336, 92/31, 93/68
Compatibilidad electromagnética
EU 89/336, 92/31, 93/68
Compatibilità elettromagnetica
IEC1010-1 & EN60950:1992 Safety Regulations
IEC1010-1 & EN60950:1992 Sicherheitsbestimmungen
IEC1010-1 & EN60950:1992 Consignes de sécurité
IEC1010-1 & EN60950:1992 Disposiciones sobre seguridad
IEC1010-1 & EN60950:1992 Prescrizioni . di sicurezza
EN55022:1987 Emissions
EN50082-1:1992 Immunity
NOTE: The displayed value may be adversely affected under
extreme electromagnetic influences, eg. when using a radio unit in the
immediate vicinity of the device. Once the interference has been
rectified, the product can once again be used for its intended purpose.
EN55022:1987 Funkstörungen
EN50082-1:1992 Immunität
Hinweis: Unter extremen elektromagnetischen Einflüssen z.B. bei
Betreiben eines Funkgerätes in unmittelbarer Nähe des Gerätes kann
eine Beeinflussung des Anzeigewertes verusacht werden. Nach Ende
des Störeinflusses ist das Produkt wieder bestimmungsgemäss
benutzbar.
EN55022:Emissions parasites
EN50082-1:1992 Immunité
Remarque: Dans des conditions d'influences électromagnètiques
extrêmes, par exemple en cas d'exploitation d'un appareil radio à
proximité immédiate de l'appareil la valeur d'affichage risque d'être
influencée. Une fois que l'influence parasite est terminée, le produit
peut être de nouveau utilisé de manière conforme aux prescriptions.
EN55022:1987 Radiointerferencias
EN50082-1:1992 Inmunidad
Nota: Bajo influencias electromagnèticas extremas, p.ej. cuando
funciona una radio en las inmediaciones del aparato, se pueden
alterar los valores del display. Cuando concluye el efecto perturbador,
el producto puede ser utilizado de nuevo, de acuerdo con lo
estipulado.
EN55022:1987 Radiointerferenze
EN50082-1:1992 Immunità
Nota: ll valore visualizzato può essere influenzato negativamente dalla
presenza di forti interferenze elettromagnetiche, per esempio quando
viene usata una radio in prossimità della bilancia. Eliminata la fonte
dell'interferenza, il prodotto può essere nuovamente utilizzato per le
funzioni cui è preposto.
James Ohaus
President
2
ISO 9001 Certificate for Ohaus Corporation.
BVQI, and was awarded the ISO 9001 certificate. This certifies that Ohaus Corporation, USA, has a quality system that conforms with the international standards for
quality management and quality assurance (ISO 9000 series). Repeat audits are carried out by BVQI at intervals to check that the quality system is operated in the proper
manner.
ISO 9001-Zertifikat für Ohaus Corporation.
und erhielt das ISO 9001 Zertifikat. Dieses bescheinigt, dass Ohaus Corporation, USA über ein Qualitätssystem verfügt, welches den internationalen Normen für
Qualitätsmanagement und Qualitätssicherung (ISO 9000er-Reihe) entspricht. Anlässlich von Wiederhol-Audits durch das BVQI wird periodisch überprüft, ob das
Qualitätssystem zweckmässig gehandhabt wird.
Certificat ISO 9000 pour Ohaus Corporation.
et a obtenu le certificat, degré ISO 9001. Celui-ci atteste que Ohaus Corporation, USA, dispose d’un système qualité correspondant aux normes internationales pour
la gestion de la qualité et pour I’assurance qualité (degré ISO 9000). Des audits réguliers effectués par la BVQI vérifient si le système qualité est appliqué de facon
appropriée.
Certificado ISO 9001 para Ohaus Corporation.
(BVQI) y ha obtenido el certificado ISO 9001. Esto acredita que Ohaus Corporation, USA, dispone de un sistema de calidad que cumple las normas internacionales para
gestión y garantfa de calidad (ISO serie 9000). Con ocasión de las inspecciones de repetibilidad por parte de la BVQI, se comprueba periódicamente si el sistema de
calidad se manipula de forma correcta.
Certificato ISO 9001 per la Ohaus Corporation.
1994 dall Bureau V eritas Quality International BVQI, e così fomice la dimostrazione che il suo sistema die Garanzia Qualità soddisfa i massimi requisite. ll sistema della
garanzia della qualità Ohaus Corporation viene verificato periodicamente dall BVQI, dando cosi evidenza di.
Ohaus Corporation, USA, was examined and ev aluated in 1994 by the Bureau V eritas Quality International,
Die Firma Ohaus Corporation, USA, wurde 1994 durch das Bureau V eritas Quality International BVQI geprüft,
La société Ohaus Corporation, USA, a été contrôlée en 1994 par Bureau V eritas Quality International BVQI
La firma Ohaus Corporation, USA, ha sido inspeccionada por la Bureau V eritas Quality International
ll sistema di garanzia della qualità della Società Ohaus Corporation, USA è certificato ISO 9001 sin dal
NOTE: THIS EQUIPMENT HAS BEEN TESTED AND FOUND T O COMPLY WITH THE LIMITS FOR A CLASS B
DIGIT AL DEVICE, PURSU ANT TO PART 15 OF THE FCC R ULES.
THESE LIMITS ARE DESIGNED T O PRO VIDE REASONABLE PRO TECTION AGAINST HARMFUL INTERFERENCE
WHEN THE EQUIPMENT IS OPERATED IN A COMMERCIAL ENVIRONMENT . THIS EQUIPMENT GENERATES,
USES, AND CAN RADIA TE RADIO FREQUENCY ENERGY AND, IF NOT INSTALLED AND USED IN ACCORDANCE
WITH THE INSTR UCTION MANUAL, MAY CAUSE HARMFUL INTERFERENCE T O RADIO COMMUNICA TIONS .
OPERA TION OF THIS EQUIPMENT IN A RESIDENTIAL AREA IS LIKELY T O CA USE HARMFUL INTERFERENCE
IN WHICH CASE THE USER WILL BE REQ UIRED T O CORRECT THE INTERFERENCE AT HIS OWN EXPENSE.
THIS DIGITAL APP ARATUS DOES NOT EXCEED THE CLASS B LIMITS FOR RADIO NOISE EMISSIONS FROM
DIGIT AL APPARA TUS AS SET OUT IN THE INTERFERENCE-CA USING EQUIPMENT STANDARD ENTITLED
“DIGIT AL APPARA TUS”, ICES-003 OF THE DEPARTMENT OF COMMUNICA TIONS.
CET APPAREIL NUMERIQUE RESPECTE LES LIMITES DE BRUITS RADIOELECTRIQUES APPLICABLES A UX
APP AREILS NUMERIQUES DE CLASSE B PRESCRITES D ANS LA NORME SUR LE MATERIEL BROUILLEUR :
“APP AREILS NUMERIQUES”, NMB-003 EDICTEE PAR LE MINISTRE DES COMMUNICA TIONS.
Unauthorized changes or modifications to this equipment are not permitted.
3
TABLE OF CONTENTS
OV ERVIEW OF CONTR OLS ........................................................................................................................ 6
OVER VIEW OF DISPLA Y INDICATORS ...................................................................................................... 7
1. GETTING TO KNOW Y OUR BALANCE .............................................................................................................. 8
2.4 Installing Subplatform and Pan ........................................................................................................................... 9
2.5 Connecting Power and Comm unications ........................................................................................................... 10
2.6 Leveling the Balance ........................................................................................................................................ 1 0
3. OPERA TING YOUR BALANCE ......................................................................................................................... 11
3.1 The Menu (Basic Settings of the Instrument) ................................................................................................... 1 1
3.2 T urning On the Balance..................................................................................................................................... 12
3.3.5 User Calibration ...................................................................................................................................... 18
3.3.7 Calibration Test ....................................................................................................................................... 20
3.6 Parts Counting.................................................................................................................................................. 2 4
4. SETTING UP Y OUR BALANCE........................................................................................................................ 28
4.1 Setting Date and Time ...................................................................................................................................... 28
4.3 Good Laboratory Practices (GLP) Data............................................................................................................. 30
4.4 Good Laboratory Practices (GLP) Set .............................................................................................................. 30
4.10Custom Unit ............................................................................................................................................... 37
5. CARE AND MAINTENANCE ............................................................................................................................ 41
5.4 Information Messages ...................................................................................................................................... 45
5.5 Service Information......................................................................................................... .................................. 46
5.6 Replacement Parts ........................................................................................................................................... 46
1 I On/Off buttonPower on/off button.
2Mode buttonSelects standard weighing, percent, parts counting, animal weighing and checkweighing
modes.
3Units buttonSelects weighing units.
4Setup buttonSelects various menus: calibration, date, time, readout, GLP data, GLP set, print,
RS232, mode, units, global, custom and auto shut-off.
5 ->O/T<- buttonWhen pressed, performs tare or rezero function.
9
10
6 buttonWhen pressed, travels up through menus.
7 buttonWhen pressed, travels to the left through menus.
8 buttonWhen pressed, travels to the right through menus.
9 buttonWhen pressed, travels down through menus.
10Enter/Print buttonWhen in menus, selects item on displa y, otherwise prints data.11Leveling indicatorIndicates leveling position of the balance.
1
6
OVERVIEW OF DISPLAY INDICATORS
4
5
9
6
8
7
10
3
2
1
No. Function
1Use (Pointer Group) key to c hange - used to
prompt the user while navigating through the menu
system.
2Standard (7) segment numeric characters. Six
characters are available and are used for display-
ing weight values.
3Center of Zero - Not used 4Stable - Indicates that the measured value has
become stable.
5Change - Is displayed together with Mode, Units
or Setup signifying that a change to balance
settings is being performed.
6Mode - Is displayed when the Mode button is
pressed. Allows the user to know what area of
the balance menu is being addressed.
(14) segment alpha numeric characters. Six
7
characters are used to present features and
functions.
Units - Is displayed when the Units button is
8
pressed. Allows the user to know what area of the
balance menu is being addressed.
9
B/G & NET - Not used
11
12
13
14
No. Function
11Custom - The user can input a f actor to meet
unique unit measuring applications.
12Symbols for weighing units and modes, include:
mg - Milligrams
g - Grams
kg - Kilograms
dwt - Penn yweight
ct - Carats
oz - Ounces
oz t - Ounces troy
GN - Grain
t - Taels. Taels are available in three types;
Hong Kong, Singapore , and Taiwan
m - Mommes
lb - Pounds
N - Newton
ti - Tical
% - Percent w eighing
PC - Parts counting
tola - tola (appears on 14 segment display)
13- Not used
14press Enter - Used as a prompt to the user to
press the Enter button. The menu item displayed
is accepted/selected.
18
10Setup - Is displayed when the Setup button is
10
pressed. Allows the user to know what area of the
balance menu is being addressed.
1. GETTING TO KNOW YOUR BALANCE
Please read through this section carefully, as it contains important information for safe and economical operation of
your NavigatorTM balance.
1.1Introduction
Thank you for deciding to purchase a Na vigatorTM balance from Ohaus. This v ersatile, portable balance offers a high
level of operating convenience and useful functions to make accurate measurements usually obtained from higher
priced laboratory instruments. A custom LCD panel has a large 6 digit, 7 segment display which indicates the weight
value of an item being measured and a 6 character 14 segment display which spells out items selected in the menus.
In addition, the display contains text which indicates the active operating mode of the balance. Arrow indicators in
the display prompt the user as to what panel keys are to be pressed to initiate a change. Built-in RS232 communication is standard on all models and allows communication with printers or computers.
Panel controls are clearly marked as to their function with large zero/tare buttons on either side of the front panel.
Operation and setup of the balance is straightforw ard and easy. Setup is simple with 4 cursor k eys , straightf orw ard
menus, and prompts that easily guide you through the process.
NavigatorTM is designed with contours for easy cleaning and has a sealed front panel and spill gutter to direct liquids
off the balance. For le v eling purposes, all models are equipped with adjustable f eet and a le vel indicator.
Behind your instrument stands OHAUS Corporation, a leading manufacturer of precision w eighing equipment. An
Aftermarket Department with trained instrument technicians is dedicated to provide you with the fastest service
possible in the ev ent your instrument requires servicing. OHA US Corporation also has a Customer Service Department to answer any inquiries regarding applications and accessories.
To ensure you make full use of the possibilities offered by y our NavigatorTM balance, we advise you to read through
these operating instructions very carefully .
2. INSTALLATION
2.1 Unpacking and Checking the Standard Equipment
Open the package and remove the instrument and the accessories. Check the completeness of the delivery . The
following accessories are part of the standard equipment of your new NavigatorTM balance.
Equipment Capacities
••
•Pan 3" Diameter •
••
••
•Pan Support, round 3" •
••
••
•Pan 4.75 Diameter • •
••
••
•Pan Support, round 4.75" • •
••
••
•Pan 6"x5.5" • •• •
••
••
•Pan Support, rectangular • •• •
••
••
•AC Power Adapter • • • • • • •
••
••
•Scoop •
••
••
•Windshield • • • •
••
••
•30g Calibration mass •
••
••
•Glass Draft shield •
••
••
•Draft shield cover •
••
••
•Instruction Manual • • • • •• •
••
••
•Warranty Card
••
• Remove packing material from the instrument.
32g 210 410g 810g 2100g 4100g 8100g
• • • • •• •
• Check the instrument for transport damage. Immediately inform your Ohaus dealer if you ha v e complaints or
parts are missing.
• Store all parts of the packaging. This packaging guarantees the best possible protection for the transport of your
instrument.
8
2.2 Selecting the Location
The balance should always be used in an en vironment which is free from e xcessive air currents, corrosives, vibration,
and temperature or humidity extremes. These factors will aff ect displa y ed w eight readings.
DO NOT install the balance:
• Next to open windows or doors causing drafts or
rapid temperature changes.
• Near air conditioning or heat vents.
• Near vibrating, rotating or reciprocating equipment.
• Near magnetic fields or equipment that generates
magnetic fields.
• On an unlevel work surface .
• Allow sufficient space around the instrument for
ease of operation and keep awa y from radiating heat
sources.
2.3 Installing Wind Shield
On 810g to 8100g balances, a wind shield is provided to
reduce the possibility of air currents from disturbing the
pan. When the wind shield is in place, air currents are
deflected up over the pan. Place the wind shield on top
of the balance making sure the small alignment tab is
Wind Shield
located properly on the balance.
2.4 Installing Subplatform and Pan
Pan
Subplatform
Windshield
All balances are supplied with a subplatform and pan.
The subplatform fits through the hole on top of the
balance and rests on a cone assembly which is part of
the measuring load cell.
The bottom of the subplatform contains a slotted keyway
for installation purposes. Place the subplatform in place
aligning the slot with the pin on the cone assembly.
Install the pan on the subplatform. The illustration shows
the rectangular pan.
9
2.5 Connecting Power and
Communications
Security Bracket
Rear of Balance
AC Adapter Connection
RS232 Connector
AC Power Operation
Connect the cord from the AC Adapter supplied to the
connector located at the rear of the balance. Connect the
AC Adapter to an appropriate power source.
Battery Operation
The balance can be operated using 8 AA alkaline batteries (not included). The battery compartment is located on
the bottom of the balance. T o install batteries, proceed as
follows:
• Remove the pan, subplatform, and wind shield.
• T urn the balance over . Do not rest the balance on
the cone.
• Press the tab on the battery cover inward and lift the
cover off.
• Remove the battery holder.
• Install the 8 AA alkaline batteries in the battery holder,
orienting the plus (+) and minus (-) ends as indicated
on the holder.
• Replace the battery holder in the compartment. Make
sure the snap connector is connected.
• Replace the battery cover .
• T urn the balance over and replace the wind shield,
subplatform and pan.
Bottom Vie w of Balance
2.6 Leveling the Balance
Exact horizontal positioning and stable installation are
prerequisites for repeatable results. To compensate small
irregularities or inclinations at the location, the instrument
can be leveled.
For exact horizontal positioning, the balance is equipped
with a level indicator located at rear .
Position the balance in the intended
operating location. Adjust the leveling
feet until the air bubble in the indicator
is centered.
NOTE: The instrument should be leveled each time its
location is changed.
Communication Connections
When an optional printer or computer is going to be used
with the balance, connect an Ohaus RS232 Interface
Cable to the connector at the rear of the balance and the
external device. Refer to paragraph 5.7 Accessories for
cable information and part numbers.
NOTE: A standard RS232 cab le cannot be used as the
pin connections are different.
The balance is now ready for operation.
Security Bracket
A security bracket (molded in base) is provided at the
rear of the balance which allows the balance to be
secured by the optional cable and lock accessory.
10
3 OPERATING Y OUR BALANCE
3. 1 The Menu (Basic Settings of the Instrument)
The balance has three basic menus; each is selected by front panel buttons marked Mode, Units and Setup.
Mode Button
The Mode button, when pressed, permits the selection of five weighing modes which are: weigh, percent, count, animal
weighing and checkweigh. These modes are controlled by an on or off selection made in the Setup menu under the
Mode menu.
Units Button
The Units button, when pressed, allows you to select or change units.
Setup Button
The Setup button, when pressed, allows entry into menus which allow you to set the balance for specific operating
parameters. Each of the menus contain settings which are user selectab le. The tables below illustr ates the various
menus and the functions which are selectable. The items shown on the menu, which are bolded, are the factory default
settings. If the Setup menu is not entered, the balance would function in the basic manner shown by the various
settings which are bolded. The setup menus shown belo w are arranged in the order as displayed in the balance .
ST ART
CAL (CALIBRATION)
InCalTM Calibration (No Lock)*
Span Calibration
User Calibration
Linearity Calibration
Calibration Test (No Lock)
Calibration Now ON/OFF
Calibration Adj ± 100 (0)
Lock ON/OFF
Exit
*(If option is installed)
READOUT
Stable.5d, 1d, 2d, 5d
Auto 0OFF, .5d, 1d, 3d
Filter-0-, -1-, -2-, -3-
LockON / OFF
Exit
PRINT
Auto PrintOFF, Cont, Inter, On Stb
IntervalEnter 1 seconds
StableON/OFF
NumericON/OFF
GLP Con ON/OFF
GLP T ar ON/OFF
Reference ON/ OFF
LockON / OFF
Exit
UNITS
Units (grams only)ON/OFF
LockON/OFF
Exit
SETUP SUBMENUS
DATE
Type m/d/y, dMy, yMd, Myd,ydM,
or dyM
Set
Exit
GLP DATA
User NumberEnter 6 digits
Project Number Enter 6 digits
LockON/OFF
Exit
The balance is ready to operate after the installation procedures are performed. When the balance is first turned on
and it completes its checks, it can be used to weigh or tare materials without setting the menus.
It is recommended that you read this manual carefully and set the balance to operate for your specific applications
using the procedures in Chapter 4 Setting up Your Balance and calibrate the balance before using.
In this section, you will enter the menu for the first time. Do not worry if you are unfamiliar with the function of the
buttons on the panel, the display provides the necessary coaching as you go along.
Power On/Off
To turn the balance ON, press the ON/OFF button (circled
button with an I inside) located at the upper left-hand
corner of the panel once. To turn OFF , press b utton
again.
NOTE: If the balance has been battery operated for a
long period of time, and the batteries are low, the balance
displays LO WB A T. Turn the balance OFF. Remo ve the
batteries and replace.
Stabilization
Before initially using the balance, allow time for it to
adjust to its new environment. Recommended warm up
period is 5 minutes.
Calibration
Refer to paragraph 3.3 and calibrate the balance before
proceeding.
3.3 Calibration
Navigator balances offer a choice of fiv e calibration methods: Internal Calibration (InCALTM), Span Calibration, User
Calibration, Linearity Calibration, and CalTestTM.
•
•
•User
•Linearity
•Cal Test
TM
InCal
Span
-Span calibration ensures that the balance reads correctly within specifications using two
-User calibration is a method where the balance can be calibrated using a mass of known
-Linearity calibration minimizes deviation between actual and displayed weights within the
-Calibration test allows the stored calibration data to be tested against the current mass being
Internal calibration (InCALTM) of the balance is accomplished by an internal mass (If option
is installed). When CAL NOW is selected and set ON, CAL NOW is displayed when the
balance requires calibration. When CAL NO W is set OFF, the message CAL NOW is not
displayed.
weight values: zero and a weight v alue at 50% of full capacity and or 100% of the balance’ s
full capacity.
value and entering that numeric value into the balance.
balance’s weighing range . Three weight v alues are used: zero , a weight value at midpoint of
the balance's weighing range, and a weight value at or near the balance’ s specified capacity.
used for the test.
•
Lock
-Can be set on or off. When set on, Span, User and Linearity calibration are locked out and
cannot be used.
12
3.3 Calibration (Cont.)
Calibration Menu Protection
NOTES:
• Calibration may be locked out to prevent unauthorized
personnel from changing calibration. If calibration has been
locked out, you can only access Cal Test.
• To lock out calibration menu, after calibration, refer to the
section titled Menu Lock-Out Protection, paragraph 4.12.
Calibration Masses
Before beginning calibration, make sure masses are
available. If you begin calibration and realize calibration
masses are not availab le , e xit the men u. The balance will
retain previously stored calibration data. Calibration should
be performed as necessary to ensure accurate weighing.
Masses required to perform the procedures are listed in
the adjacent table and are available as accessories.
NOTE:
Any of the calibration modes can be terminated
time
by pressing either the Mode, Units or Setup buttons.
at any
CALIBRA TION MASSES
LINEARITY SPAN ONLY
CAPACITY MASSES MASSES
32g 10g and 30g 30g
210g 100g and 200g 100g or 200g
410g 200g and 400g 200g or 400g
810g 400g and 800g 400g or 800g
2100g 1000g and 2000g 1000g or 2000g
4100g 2000g and 400g 2000g or 4000g
8100g 4000g and 8000g 4000g or 8000g
It is recommended that masses meet or exceed ASTM
Class 4 Tolerance. Models 410g and 4100g use Class
2. Calibration masses are available as accessories.
13
3.3.1Internal Calibration (InCALTM)
On NavigatorTM balances equipped with the InCalTM feature, calibration can be accomplished using the internal calibration mass. InCalTM requires use of the AC adapter supplied with the balance. When the balance requires calibration, a
screen prompt of CAL NOW appears. This display can be turned off as described under Calibration Message. Also , a
software feature is incorporated which permits the internal calibration mass to be adjusted by +100 divisions. The
adjust feature is described under Calibration Adjust. Internal calibration can be performed at any time providing the
balance has warmed up to operating temperature.
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is displayed.
• Press Enter button, CAL TYP InCAL is displa y ed.
• Press Enter button, INCAL is displayed.
IMPORTANT !
DO NOT DISTURB THE BALANCE DURING
CALIBRATION. IF THE MESSAGE UNSTBL
IS DISPLA YED , THE BALANCE W AS UNABLE
TO ACQ UIRE ST ABLE D ATA DURING INTERNAL CALIBRATION. THE BALANCE WILL
CONTINUE TO PERFORM INTERNAL CALIBRA TION UNTIL READINGS ST ABILIZE. THE
BALANCE WILL THEN COMPLETE THE INTERNAL CALIBRATION FUNCTION.
TO EXIT INTERNAL CALIBRATION MODE
BEFORE COMPLETION, PRESS ENTER OR
SETUP BUTTONS.
STABILITY CAN BE AFFECTED BY TEMPERATURE CHANGES, AIR CURRENTS, VIBRA TION, ETC...
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
The internal mass is positioned several times during
calibration and then removed, then after a few seconds,
CAL SET is displayed indicating a successful calibration. The displa y returns to WEIGH mode.
14
3.3.2Calibration Message
On NavigatorTM balances equipped with the InCalTM feature, a screen prompt of CAL NOW appears when the balance
requires calibration. This display can be turned off if it is desired not to have the balance indicate that calibration is
required. Turning the display off has no effect on the basic balance operation.
Procedure
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is display ed.
• Press button, CAL NOW is displayed.
• Press Enter button, CAL NOW ON or OFF is displayed.
• Press or button and select either ON or
OFF. When OFF is selected, the CAL NOW message
will not appear in the display.
• Press Enter button, SAVED is momentarily display ed,
then the display indicates CAL ADJ.
NOTE: At this point you may continue with the calibra-
tion adjust procedure on the next page or exit.
The calibration adjust procedure is only used when it is
desired to calibrate the internal calibration mass to a
known external mass if a difference exits between the
known mass and the InCalTM value.
• Press Enter button, LOCK is displayed.
• Press or button and select either ON or OFF.
• Press Enter button, SA VED is momentarily display ed,
then the display indicates EXIT.
• Press Enter button, displa y returns to WEIGH mode.
15
3.3.3Calibration Adjust
Balances with InCalTM contain software which allows the
internal calibration value to be adjusted by + 100 divisions
at full scale capacity using an external mass which is
traceable to a certified standard.
Procedure
• Perform the internal calibration procedure.
• Press >O/T< button to zero the balance.
• Place a mass equal to the
the balance on the pan. Note the reading on the balance and see if the balance indicates the exact
weight or indicates a higher or lower reading. If the
reading is higher or lower , proceed.
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is display ed.
• Press button until , CAL ADJ is displayed.
• Press Enter button, CAL ADJ 0 should be displayed (0 is factory setting).
NOTE: Balance will retain last CAL ADJ setting.
• Press or button until desired number is
displayed.
• Press Enter button, SAVED is momentarily display ed,
then display indicates LOCK.
• Press Enter button, LOCK ON or LOCK OFF is
displayed.
• Press or button and select either LOCK ON
or LOCK OFF.
• Press Enter button, SA VED is momentarily displayed,
then EXIT is displayed.
• Press Enter button, displa y returns to WEIGH
mode. Note weight reading and remove mass from
pan.
• Perform the internal calibration procedure. The v alue
entered as an adjustment is now stored. Place the
calibration mass on the pan and check. Repeat procedure if further correction is required.
• T o return to factory setting, follow procedure above
and set CAL ADJ to 0.
span calibration value
of
16
3.3.4Span Calibration
Span calibration utilizes two calibration points , one at zero and a choice of either half capacity or full capacity. All
balances will prompt for a particular weight which is shown in the table. If the prompt calls for a half capacity mass,
you can use a full capacity mass, the balance will accept it. If the prompt calls for a full capacity mass, you can use
a half capacity mass. This versatility allows calibration at other than full capacity. Sample display below illustrates a
210g balance.
Procedure
• Clear the pan.
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is displayed.
• Press Enter button to select SPAN calibration, CAL
TYP SPAN is displayed.
• Press Enter button, BUSY is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
• Display changes to PUT WT 200g. The display ed
weight is the value of the mass to be used. See the
following table for masses that may be used with the
different model balances. V alues shown in bold are
displayed.
SPAN CALIBRATION MASSES
MODEL 32g 210g 410g
MASS 30g 100g, 200g 200g, 400g
MODEL 810g2100g4100g8100g
MASS 400g, 800g 1kg, 2kg 2kg, 4kg4kg, 8kg
• Place the calibration mass on pan.
NOTE: The PUT WEIGHT message indicates the
calibration mass that is on the pan.
• Press Enter button, BUSY is displayed. After a
few seconds CAL SET is displayed, the display then
returns to WEIGH mode.
• Span calibration is completed.
• Remove calibration mass from the pan.
17
3.3.5User Calibration
User calibration is used when it is desired to calibrate the balance using a mass of known value. User calibration limits
are no less than the 1/2 nominal capacity point and no more than the display over limit. For example, on a 210g
balance, it's no less than 100.00g and no more than 210.09g. To use this calibration feature, proceed as follows:
Procedure
• Clear the pan.
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is displayed.
• Press Enter button.
• Press and scroll to select USER calibration,
CAL TYP USEr is displayed.
• Press Enter button, the display indicates the last
calibration mass value which was entered with the
first digit flashing. (Sampleillustrates 200g).
• Press to change value of digit and
button to advance to next digit. Enter the desired
mass value for y our balance . This number must be
at least 50% of the full span value.
(FLASHING)
• After entering value, press Enter button, BUSY is
displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
• Display changes to PUT WT 200 g.
• Place specified calibration mass on pan.
• Press Enter button, BUSY is displayed. After a
few seconds CAL SET is displayed, the display then
returns to WEIGH mode.
• User calibration is completed.
• Remove calibration mass from the pan.
18
3.3.6Linearity Calibration
Linearity calibration utilizes three calibration points, one at zero, center span and full span. This method minimizes
deviation between actual and displayed weights within the balance's weighing range. Sample display illustrates a 210g
balance.
Procedure
• Clear the pan.
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is displayed.
• Press Enter button.
• Press button and scroll to select Linearity
calibration, CAL TYP Lin is displa y ed.
• Press Enter button, BUSY is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
• Display changes to PUT WT 100 g. The displayed
weight is half the capacity of the balance.
• Place specified calibration mass on pan.
• Press Enter button, BUSY is displayed. After a
few seconds displa y changes to PUT WT 200 g.
The displayed weight is the full capacity of the
balance.
• Place specified calibration mass on pan.
• Press Enter button, BUSY is displayed. After a
few seconds CAL SET is displa y ed, the displa y then
returns to WEIGH mode.
• Linearity calibration is completed.
• Remove calibration mass from the pan.
19
3.3.7Calibration Test
Calibration test feature allows a check of a known calibration mass against the last stored calibration information in the
balance. Sample display illustrates a 210g balance.
Procedure
• Clear the pan.
• Press the Setup button, CAL is displayed.
• Press Enter button, CAL TYP is displayed.
• Press Enter button.
• Press button and scroll to select CALtSt
calibration, CAL TYP CALtSt is displayed.
• Press Enter button, BUSY is displayed.
NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.
• Display changes to PUT WT 200 g. The displayed
weight is the full capacity of the balance.
• Place specified calibration mass on pan.
• Press Enter button, BUSY is displayed. After a
few seconds , DIFF is displa yed. The display now
indicates the actual diff erence in weight between
what value was just placed on the pan and the previous weight value which was stored in the balance.
After approximately 8 seconds, the displa y returns to
the WEIGH mode.
• Remove calibration test mass from the pan.
20
3.3.8Calibration GLP Printout
If any option in the GLP Set Menu is turned On, GLP automatically prints data after calibration is completed.
Span Calibration Printout
When performing Span calibration with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
Linearity Calibration Printout
When performing a Linearity calibration with all GLP
options turned on, a printout is automatically made after
the calibration is completed.
- - - - - SPAN CAL - - - - - -
12/01/981:00:00 PM
Bal Id 1234
Cal:200.00g
Old:200.00g
Dif: 0.00g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - LIN CAL - - - - - -
12/01/981:00:00 PM
Bal Id 1234
Cal:200.00g
Old:199.80g
Dif: 0.20g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
Calibration T est Printout
When performing a Calibration Test with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
User Calibration Printout
When performing a User Calibration with all GLP options
turned on, a printout is automatically made after the
calibration is completed.
- - - - - END - - - - -
- - - - - CAL TEST - - - - - -
12/01/981:00:00 PM
Bal Id 1234
Cal:200.00g
Act:200.20g
Dif: 0.20g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -
- - - - - USER CAL - - - - - -
12/01/981:00:00 PM
Bal Id 1234
Cal:200.00g
Old:200.20g
Dif: 0.20g
Wt.Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
21
- - - - - END - - - - -
3.4 Weighing
NavigatorTM balances are shipped with grams only enabled. When the balance is to be used with other units of measure, the desired unit must be enabled. Refer to paragraph 4.8 to enable other measuring units. 32g models are
shipped ready to weigh in g, oz t, dwt and ct.
Procedure
• Pressto rezero the display.
• Press Units button to select measuring unit.
• Press or button for desired measuring unit.
• Press Enter button, balance is now ready for weigh-
ing.
• Place the object(s) or material to be weighed on the
pan. Example illustrates a 200 gram weight on a high
capacity model.
• Wait for the stability indicator to appear before readin g the weight.
Zero/Tare
When weighing material or objects that must be held in a container, taring stores the container weight in the balance’ s
memory separate from the weight of the material in the container.
(
Example
Procedure
• Presswith no load on the pan to set the
balance to zero.
• Place an empty container on the pan. Its weight is
displayed.
• Press. The display blanks until stable weight
readings are received, then indicates zero . The
container’s weight is stored in memory.
• Add material to the container. As material is added,
its net weight is displayed.
Container 200g)
• Removing the container and material from the platform will cause the balance to display the container’s
weight as a negative number . The tared weight will
remain in memory untiliis pressed again or
the balance is turned off.
• Pressing resets the balance to zero.
(
Example
Material 1620g)
22
3. 5Percent Weighing
Percent W eighing is
permits you to place a reference load on the balance, then view other loads as a percentage of the ref erence. The load
you place on the pan as a reference may be displayed as any percentage you select from 5% to 100% (in 1% increments). One hundred percent does not necessarily have to represent the reference load. Subsequent loads, displayed
as a percentage of the reference are limited only by the capacity of the balance . The default setting is Ref erence
100%. Refer to paragraph 4.7 to enable percent weighing.
enabled only
when Percent is turned ON in the Mode menu under Setup . Percent weighing
Procedure
• Press the Mode button, WEIGH is display ed.
• Press button until PERCNT is displayed.
• Press Enter button, PUTPAN 100% is displayed. If
a container is used, the balance can be tared at this
point. The % display momentarily blanks while the
balance is taring out.
• Put the reference load on the pan.
• Press or button and select reference
weight percentage (Percent Range 5 to 100). Hold
button down for fast change.
• Press Enter button to save setting, BUSY is displayed while the balance calculates the reference
weight. The balance shows the 100% equivalentreference weight for two seconds in the selected measuring unit, then displays the percentage which was
entered.
• T o view the weight of the ref erence load, press the
UNITS button, then press button until unit of
measurement such as GRAMS is displayed, then
press the Enter button.
• T o return to percent weighing, press the UNITS button
and then press button until PERCNT is dis-
played, then press the Enter button.
• Remove the reference weight from the pan and replace it with another load. The second load is displayed as a percentage of the ref erence . The displa y
illustrates 100% of the reference value.
• T o return to weighing mode, press the MODE button
and then press button until WEIGH is displa y ed,
then press the Enter button.
23
3.6 Parts Counting
Parts Counting is
mode, the balance displays the quantity of parts you place on the pan. Since the balance determines the quantity
based on the average w eight of a single part, all parts must be reasonably uniform in weight. Ref er to paragraph
4.7 to enable parts counting.
enabled only
when Count is turned ON in the Mode menu under Setup. In the parts counting
Procedure
• Press the Mode button.
• Press button until COUNT is displayed.
• Press Enter button. PUTPAN 10 is displayed
(default setting). Balance will retain last sample size
saved.
• Put a container on the pan and pressto tare.
• Press and hold or button and select sample
size. Sample size may be 5 to 1000 parts.
• Place sample on the pan.
• Press Enter button to continue, display indicates
BUSY.
Balance displays the reference weight of an individual
part for five seconds and then displays the total number
of parts on the pan.
• Place parts to be counted on the pan. Balance displays number of parts.
• T o view the total w eight of the parts, press the UNITS
button, then press to select desired weighing
unit, then press the Enter button.
• T o return to parts counting, press the UNITS button,
then press button until COUNT is displayed,
then press the Enter button.
• T o begin counting with a new sample, repeat procedure.
Update
Update is a function which permits placing additional
samples which are greater than the value of the original
sample but less than three times the value. This action
increases the accuracy of the measurement.
• Place sample on the pan which is at least one but not
more than three times the original sample size.
• Press Mode button, COUNT is displayed.
• Press Enter button, UPDATE is displayed.
• Press Enter button, BUSY is displayed then the
reference weight followed by the new sample size.
24
3. 7 Animal Weighing
Animal Weighing is
4.7 to enable animal weighing.
enabled only
when Animal is turned ON in the Mode menu under Setup. Ref er to parag raph
Procedure
• Press the Mode button.
• Press button until ANIMAL is displayed.
• Press Enter button to continue, LEVEL is displayed.
• Press or button to change sampling level,
0, 1, 2 or 3. 0 level is used for an inactive subject, 3
is used for a very active subject.
• Press Enter button to continue, A U TO is displayed.
• Press or button to select A U T O ON or OFF.
• Press Enter button to continue.
When the AUTO function is set ON, different subjects can
be weighed one after another without pressing any
buttons. When the balance displays READ Y, simply
place the subject on the pan.
Starting Animal Cycle
• Place animal container (if used) on pan.
• Pressto tare the container.
• Place subject on pan.
• The animal cycle will automatically start if AUTO w as
set to ON.
• Press Enter button to start animal cycle if AUTO was
set to OFF.
During Animal Cycle
• Display shows countdown ending with A W 0, then
displays the subjects weight.
Completed Animal Cycle
• Balance displays weight until subject is removed from
the pan.
25
3.8 Checkweighing
Checkweighing is
permits you to set a target nominal weight, and over and under weight limits . This type of weighing is used where
individual items must be checked against preset parameters. Refer to paragraph 4.7 to enable checkweighing.
enabled only
when Check is turned ON in the Mode menu under Setup. Checkweighing mode
Procedure
• Press the Mode button.
• Press button until CHECK is displayed.
• Press Enter button to continue, NOM WT is displayed. The n umerical display indicates 0000.00 with
the first digit flashing. The sample display indicates
0200.00g
• Press to change the digit and to
go to the next digit. Enter the desired nominal value.
• Press Enter button to continue. OVR WT is displayed. The n umerical display indicates 0000.00 with
the first digit flashing. Enter the over weight value by
NOTE: Samples shown below are for a 200g nominal
setting. The tolerances were set for 1g over and under.
pressing to change the digit and
to go the next digit. The sample display indicates
0201.00g.
• Press Enter button to continue. UND WT is dis-
played. The n umerical display indicates 0000.00 with
the first digit flashing. Enter the under weight value
by pressing to change the digit and
to go the next digit. The sample display indi-
cates 0199.00g.
• Press Enter button to continue.
• Place item to be checked on the balance pan.
ACCEPT is displayed when the item is within limits.
OVER and/or UNDER is displa y ed if the item is out of
the tolerance set in the limits.
UNDER is displayed for weights less
than or equal to the UND WT limit
T ypical Nominal Weight Displa y
26
OVER is display ed f or weights greater
than or equal to the OVR WT limit
3.9 Printing Data
Printing data to an external computer or printer requires that the communications parameters in the Setup menu, print
options and communication parameters be set first. Refer to page 23 Print menu settings and page 25 for RS232
communication settings.
Procedure
• Press the Print button. Printing to an external printer
or computer will occur each time the Print button is
pressed unless the autoprint feature is turned on in
which case, printing can occur in a continuous fashion, at specified intervals or each time a stable reading is achieved.
Sample printout is shown below with time turned on.
SAMPLE PRINTOUT
12/01/9812:01:37 AM
429.5 g
For a review of of printing samples, refer to Section 4
Setting Up Your Balance. What is printed is controlled
by the GLP Set Menu and the selection of GLP
Continous or GLP Tare in the Print Menu.
27
4.SETTING UP YOUR BALANCE
4. 1 Setting Date and Time
Your NavigatorTM balance provides date and time data which can be viewed on a computer or printed out on an external
printer. When you put y our new instrument into operation f or the first time, y ou should enter the current date and the
time. These settings are retained as long as the balance remains connected to an A C power source using the A C
Adapter. Date and time are not retained with battery operation.
Date
Date is a feature which enables the balance to be set to a
U.S . date standard or European date standard. U .S.
standard has the month, day, followed by the y ear, each
separated by (/) in the printout. The European date
standard has the day first, followed by the month and
then the year; each separated b y a period. The def ault
setting is Mdy.
Time
Time is a feature which enables the balance to be set to
the current time in either 12 hour or 24 hour periods. The
default setting is 12 hour .
Procedure
• Press the Setup button, CAL is displayed.
• Press button and select Date from the menu.
• Press Enter button, TYPE is display ed.
• Press Enter button, SET M d y, d M y, y M d, M y d,
y d M, or d y M is displayed.
• Press or button and select type of date.
• Press Enter button, SAVED is displayed, then SET is
displayed.
• Press Enter button, date is displayed.
• Using arrow buttons, enter the correct date.
• When the correct date is entered, press Enter button,
SA VED displa ys momentarily and EXIT appears.
• Press Enter button, balance returns to a weighing
mode.
Procedure
• Press the Setup button, CAL is displayed.
• Press button and select Time from the menu.
• Press Enter button, TYPE is displayed.
• Press Enter button, TYPE 12 hr is display ed.
• Press or button and select 12 hr or 24 hr.
• Press Enter button, SAVED is displa yed momentarily
then SET is displayed.
• Press Enter button,SET with time is displayed.
• Using arrow buttons, enter the correct time.
• When the correct time is entered, press Enter button,
SA VED displays momentarily and ADJUST appears.
• Press Enter button, 0 appears.
Adjustments up to +60 seconds per day can be made
to the balance internal clock. Using arrow buttons,
enter time correction and press Enter button. SA VED
displays momentarily and EXIT appears. Press Enter
button, balance returns to a weighing mode.
28
4.2Readout
The Readout menu is used to adapt the balance to environmental conditions. It contains four menus: Stable, Auto 0,
Filter, Lock and Exit. Loc k, when set ON, locks the Readout settings entered, pro viding the Lock s witch underneath
the balance is also set ON.
Stability
The stability range can be set to a preset tolerance limit. When the balance reading is stable within preset limits, the
stability indicator remains ON. When a displa y ed weight changes bey ond the allowab le range, the stability indicator
turns OFF, indicating an unstable condition. Factory default setting is shown in bold type .
Procedure
To select any of the items in the Readout menu, proceed
as follows:
• Press the Setup button, CAL is displayed.
• Press button until READO T is display ed.
• Press Enter button to continue, STABLE is displayed.
• Press or button until either STABLE, AUT O
0, FILTER, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press or button and select the desired menu
setting. Review settings available below.
• Press Enter button, SAVED is displa y ed.
• Press or button to continue to LOCK and
EXIT.
• Press Enter button to continue.
.5 dSmallest range: stability indicator is ON only
when displayed weight is stable within .5 divisions.
1 dReduced range.
2 dNormal range.
5 dLargest range, stability indicator is ON when displayed weight
changes up to 5 divisions.
When the RS232 interface is configured to print stable data only, the stability range also governs data output. Displayed data will only be output if it is within the selected stability range.
Auto-Zer o
Auto-Zero minimizes the eff ects of temperature changes and shift on the z ero reading. The balance maintains the zero
display until the threshold is exceeded. Factory default setting is shown in bold type.
OFFT urns Auto-Zero OFF.
.5 dSets threshold to .5 divisions.
1 dSets threshold to 1 division.
3 dSets threshold to 3 divisions.
Filter
Filter compensates for vibration or excessive air currents. Default settings are shown bold.
-0- minimum filtering, fastest stabilization time
-1- reduced filtering, fast stabilization time
-2- normal filtering, normal stabilization time.
-3- maximum filtering, slowest stabilization time.
Lock
Lock ON/OFF can only be changed when the hardware Lock Switch is set OFF/disabled. A menu is locked when the
menu lock is set ON and the Lock switch is ON. Lock, when selected and turned on, locks all of the entries made
under the Readout menu. In the loc ked condition, items ma y be looked at b ut not changed in the men u. When set
OFF, entries may be changed. OFF is the default setting. Ref er to paragraph 4.12 f or Menu Loc k-Out Protection.
29
4.3 Good Laboratory Practices (GLP) Data
The GLP Data menu enables the storage of a user identification number (6 digits) and/or a project number (6 digits).
When entered into the balance, the identification number and project number are available when printing, provided they
are turned on in the GLP Set menu. A lock setting is also available which locks in the user identification and project
number.
Procedure
To enter data in the GLP Data menu, proceed as follows:
• Press the Setup button, CAL is displayed.
• Press button until GLPDAT is displayed.
• Press Enter button to continue, USER NO is displaye d.
• Press Enter button to continue.
• Press buttons to change digit values and
buttons to advance to next digit and enter a
6 digit number for the user number .
• Press Enter button to save setting, PROJNO is displayed.
• Press Enter button to continue.
• Press to change digit values and
to advance to next digit as directed by thedisplay
and enter a 6 digit number for the project number .
• Press Enter button to save setting, LOCK is displayed.
• Press Enter button, press to select ON or OFF.
• Press Enter button to save setting, SAVED is dis-
played then EXIT.
• Press Enter button, balance returns to weigh mode.
4.4Good Laboratory Practices (GLP) Set
Good Laboratory Practices (GLP) Set menu allows the selection of and will permit printing of Time, Balance Identification, User Number, Project Number , Diff erence and Name data. The default setting is OFF. When an e xternal printer is
used with the balance calibrated, and all items are set ON, the printer will print out calibration data for audit trail
purposes and will indicate date, and time. (It should be noted that the User number and Project number must be
entered in the GLP Data menu before printed data is available).
Procedure
To select any of the items in the GLP Set menu, proceed
as follows:
• Press the Setup button, CAL is displayed.
• Press button until GLPSET is displayed.
• Press Enter button to continue.
• Press or button until either TIME, BAL ID
(see note), USER NO, PROJ NO, DIFF , NAME,
LOCK or EXIT is displayed.
• Press Enter button to continue.
NOTE: Balance ID (BAL ID) represents the actual serial
number of the balance and cannot be entered or changed.
Setting BAL ID ON allows printouts to display the serial
number of the balance during GLP data.
• Press or button and select either ON or
OFF for each entry .
• Press Enter button to save setting.
• Press or button to continue or EXIT.
• Press Enter button, balance returns to weigh mode.
30
Loading...
+ 77 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.