Turns On or Off display and fan.
2
When pressed, sends Print command.
3
When pressed, either starts or stops the drying process.
4
When pressed, performs Tare function.
5
Leveling indicator
6
Leveling Feet
Used to level the Moisture Analyzer.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
General hazard
Electrical shock
Hot surface
Inflammable or
explosive substance
Toxic substance
Acid / Corrosion
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
It is not permitted to use the instrument in explosive atmosphere of gases, steam, fog, dust and flammable dust (hazardous environments).
A lab coat should be worn.
A suitable eye protection such as goggles should be worn.
Use appropriate gloves when handling chemicals or hazardous substances, checking their integrity before use.
WARNING Risk of electric shock
Your instrument is supplied with a 3-pin power cable with an equipment grounding conductor. Only extension cables which meet this relevant standards and also have an equipment grounding conductor may be used. Intentional disconnection of the equipment grounding conductor is prohibited.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CAUTION The Halogen Moisture Analyzer works with heat!
a) Ensure sufficient free space around the instrument to avoid heat accumulation and overheating (approx. 1 m free space above the heating module). b) The vent over the sample must never be covered, plugged, ta ped over or tampered with in any other way. c) Do not place any combustible materials on, under or next to the instrument since the area around the heating module may be hot. d) Exercise caution when removing the sample. The sample itself, the sample chamber, the draft shield and any sample vessels used may still be very hot. e) During operation, you should never open the heating module itself as the ring-shaped heating reflector or its protective glass can reach 400 °C! If you have to open the heating module e.g. for maintenance, disconnect the instrument from the power supply and wait until the heating module has cooled down completely. f) No modifications must be made within the heating module. It is particularly dangerous to bend any components or remove them or to make any
other changes.
Certain samples require special care!
With certain types of samples, there is a possibility of danger to personnel or damage of property. Please note that the user always has the responsibility and liability for damage caused by use of any types of samples!
CAUTION Fire or Explosion
Flammable or explosive substances.
Substances containing solvents.
Substances which evolve flammable or explosive gases or vapors when
heated. a) In cases of doubt, perform a careful risk analysis. b) Work at a drying temperature that is low enough to prevent the formation of flames or an explosion. c) Wear protective goggles.
d) Work with small amounts of sample.
e) Never leave the instrument unattended!
WARNING Substances which contain toxic or caustic components
Toxic gases produced during drying could cause irritations (eyes, Skin, breathing), illness or death.
Such substances may be dried only in a fume cupboard.
CAUTION Corrosion
Substances which evolve corrosive vapors when heated (e.g. acids). – Work with small amounts of samples as the vapor can condense on cooler
housing parts and cause corrosion.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
DO NOT install the Moisture Analyzer:
• Next to open windows or doors causing drafts or rapid temperature changes.
• Near air conditioning or heat vents.
• Near vibrating, rotating or reciprocating equipment.
• Near magnetic fields or equipment that generate magnetic fields.
• On an unlevel work surface.
• In confined areas, allow sufficient space around the instrument for ease of operation and keep away from radiating heat sources.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Lift the cover straight up and Install the tray pan in the base of the heating chamber.
Install the draft shield (only one position possible) on top of the tray pan.
Install the pan support into position. Turn the pan support until it engages. In the locked position, the arm of the pan support points directly towards the front of the analyzer.
Adjust the leveling feet of the Moisture Analyzer until the air bubble in the indicator is centered. The level indicator is located under the cover towards the rear of the Moisture Analyzer.
Attention: The instrument should be leveled each time its location is changed.
Leveling
Indicator
Leveling
Feet
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
WARNING: Risk of electric shock.
- Use only the 3-pin power cord with equipment grounding connector which was supplied with your instrument. Only connect the power cord to a 3-pin ground outlet.
- Only extension cords which meet the relevant standards and also have an equipment grounding conductor may be used.
Attention:
Check to ensure the voltage stated on the Moisture Analyzer’s data label matches your local line voltage. If this is not the case, do not connect the Moisture Analyzer to the power supply, contact your responsible OHAUS dealer. The halogen dryer unit is designed to operate at a specific line voltage (90V AC or 240V AC). The dryer unit is installed at the factory and is matched to the particular line voltage of the country of destination.
Connection to a line voltage that is too high can lead to burning out the halogen heater, whereas, a supply voltage that is too low will prolong the drying process and the instrument may not operate properly.
Connect the power cord to the power supply socket located at the rear of the Moisture Analyzer and to the power supply outlet. The Moisture Analyzer becomes operational as soon as power is
applied. The display will remain off until the On/Off button is
pressed.
Allow the Moisture Analyzer to warm up for at least 30 minutes to
enable it to adapt itself to the ambient conditions. If the Moisture Analyzer has been stored in a very cold environment before installation, it may require several hours for it to stabilize.
WARNING: If the power cable supplied is not long enough, use only a proper 3-pin extension cable with an equipment grounding connector.
To switch the Moisture Analyzer on, press the On/Off button. An internal diagnostic test is
performed; the display lights and the Home screen appears followed by the initial display.
To switch the Moisture Analyzer off, press the On/Off button. After the analyzer has been switched
off, it is in the standby mode. If you wish to perform a
test, press the On/Off button again.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Note:
As your Moisture Analyzer needs no warm-up time when in the standby mode and is immediately ready for testing samples, we advise you to switch the display off by use of the On/Off button and not to disconnect it from the power supply. This also ensures that the Moisture Analyzer is always in thermal
equilibrium.
If the display is not visible and the unit is plugged into a power source, press the On/Off button.
Directly start a measurement.
Open the cover on the Moisture Analyzer. Clean the pan. Place the pan handle in the sample chamber. Ensure that the tongue of the pan handle fits exactly in the slot of the draft shield element.
Place the empty sample pan and glass fiber pan in the pan handle. The sample pan must lie flat in the pan handle.
Attention: It is advisable for you to work with the pan
handle at all times. The pan handle is ergonomic, safe and provides protection against possible burns due to a hot sample pan. Close the cover. This sets the Moisture Analyzer to zero automatically.
Open the cover on the Moisture Analyzer.
Wet the glass fiber pan with a 0.5 to 1 gram of water.
Attention: Minimum sample size must be greater than 0.5 gram.
Starting the test
Close the cover, the Moisture Analyzer starts the drying and measurement process automatically.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Drying and measurement
You are now able to follow the progress of the drying cycle on the display. This first display comes up automatically when the test was started.
Observing the test progress
There are five displays available when the Moisture Analyzer is running a test.
(Showing %MC, %DC, g, %RG or a curve. )
When pressing the data area while %MC is shown, the display will cycle through each of the five displays.
The display indicates the Method Name, Temperature/Time, actual temperature, elapsed time duration of the test, moisture and curve. The test automatically stops at the end of the measurement. If you want to end the test sooner, press the button.
Congratulations! You have just performed your first
measurement with your moisture analyzer.
CAUTION: Danger of burns.
Sample, sample pan and sample pan holder may still be hot.
Open the cover and carefully remove the sample pan handle from the drying area.
The analyzer comes with a default method preinstalled. The settings of the default method are shown on the display. To edit the settings, press on the area where the settings are displayed as shown to the left.
Press the setting items to edit them. Please see section 7 for more details.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press the Method library button to access the Method library menu.
The test library contains 2 methods. All methods which were previously entered can be recalled and the parameters for the recalled method when selected will be repeated.
Press the Test results button to access the
test results menu.
The Test results contains up to 100 test results. All results can be recalled, sorted and displayed statistically.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press the Setup button to access the Setup
menu.
Press the Help button to access the Help
menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Setup button to access the Setup
menu.
Press Back to return to Home screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press the Adjustment button to access adjustment menu.
Press Back to return to previous step.
Press the Weight adjustment-external button
The display highlights the action you should take. Follow the instructions on the display. Remove the sample pan. Place the required weight on the pan support. The display indicates if the adjustment was successful.
Press Setup button to access the adjustment menu.
Press Back to return to Home screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press adjustment button.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Temperature adjustment button.
You are now prompted to remove the pan support. Place a temperature calibration kit and close the upper housing.
Press the Start to initiate the temperature
adjustment process. Follow the screen prompts throughout the process.
The dryer unit is heated to a temperature of 100°C. You can observe the progress on the display as the dryer temperature and count down period are displayed. After 15 minutes, read the thermometer through the inspection window on the cover and enter this temperature. Adjust the display reading to agree with the thermometer, then press the
Enter button. You have 15 minutes to make
this adjustment, otherwise the adjustment is
terminated.
This is a two-point adjustment (100°C and 160°C). The dryer unit now heats to the second temperature (160°C). Adjustment of the temperature is defined by two points. Proceed exactly as you did for the first temperature. After you have set the display to agree with the thermometer, then press
the Enter button.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
CAUTION: Danger of burns.
Be careful when removing the temperature calibration kit from the dryer unit as it can be very hot. Allow it to cool down by opening the cover before removing. Remove the calibration kit. Replace the pan support in its proper positions.
Press Setup button to access the Setup
menu.
Press adjustment to access the adjustment
menu.
Press Back return to Home screen.
Press Weight adjustment record to list the
weight adjustment record data.
Press Back to previous step. Press Exit return to Home screen.
Press Setup button to access the Setup
menu.
Press Back to return to Home Screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Weight and temperature adjustment button to access the adjustment screen.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press the Temperature adjustment history button. Here you will find the last temperature adjustment result.
Press Back to return to previous screen.
Press Setup button to access the Setup screen.
Press Back to return to Home screen.
Press User settings to access the user
settings menu.
Press Language to set language.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Select LANGUAGE.
Note: If you select English as the dialog
language, the date format is changed and appears on all records in the MM/DD/YYYY format.
Press Setup button to access the Setup menu.
Press Back to return to Home screen.
Press User settings button to access the
User settings menu.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Brightness button.
Adjust the brightness of the screen to desired level.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Setup button to access the Setup menu.
Press Back to return to previous screen.
Press User settings button to access the User settings menu.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Sound button.
Adjust the volume of the beep to desired level.
.
Press Setup button to access the setup menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Back to return to previous screen.
Press Device settings to access the Device settings menu.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press the Date-Time button.
Press the follow items to set the date and time.
Press Back to return to previous screen. Press Exit to return to Home screen.
Set date format.
Press Back to return to previous screen.
Set date. Press + and – to set correct date, month and year.
Set time format.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Set time. Press +and – to set correct time.
Press Setup button to access the Setup menu.
Press Back to return to Home screen.
Press Device settings button to access the Device settings menu.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Peripherals button.
Select the following items to set RS232 port when RS232 port link to printer or other RS232 compatible device.
Press Back to return to previous screen. Press Exit to return to Home screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Setup button to access the Setup
menu.
Press Back to return to Home screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Device settings button.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Print settings button to access the Print settings menu.
Press Print adjustment result automatically to switch ON or OFF. Press Print measurement result automatically to switch ON or OFF. Press Intermediate results print interval to
set the print interval.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press the Intermediate results print interval button. Choose the interval seconds or minutes.
Press Setup button to access the setup menu.
Press Back to return to Home screen.
Press Device settings button to access the Device menu settings.
Press Back to return to previous screen. Press Exit to return to Home screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press GLP and GMP Data to access the GLP and GMP Data menu.
Press Back to return to previous screen. Press Exit to return to Home screen.
Press Company name, Department or Instrument ID buttons to define related information.
Press Setup button to access Setup menu.
Press Back to return to Home screen.
Press Device settings to access the device
settings.
Press Touch screen adjustment to adjust
the touch screen.
Press the Start button to start the
adjustment.
Please follow the process to adjust the screen.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
When the adjustment complete, press OK to
exit.
Press Setup button to access Setup menu.
Press Back to return to Home screen.
Press System and data management button to access the System and data management menu.
Press Back to return to previous step. Press Exit to return to Home screen.
Press Export and import settings and methods to access the export and import
menu.
6.5.1.1 Export data
Press Export data to USB flash drive to
select location and export data to USB drive.
Press Back to previous step. Press Exit to return to Home screen.
Press File name to change file name. Press Location to select the export location. Press Export selection to select the export
items.
Press Cancel to previous step.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press the items to select.
Press Cancel to cancel data export. Press OK to confirm and back to previous
step.
After selecting the location, Export button is
available.
Press Cancel to cancel data export. Press Export to execute.
6.5.1.2 Import data
Press Import data from USB flash drive to
select location and import data from USB drive.
Press Back to previous step.
Press Import from file to select the file to
import.
Press Cancel to cancel data import.
After select the import file, the Import button
is available.
Press Cancel to cancel data import. Press Import to execute.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
6.5.2 Backup and restore
6.5.2.1 Backup system
Press Backup and restore to backup or
restore the system settings.
Press Backup system on USB flash drive
to back up system.
Press Back to previous step. Press Back to return to previous step.
Press File name to change file name. Press Location to select backup location. Press Cancel to cancel backup.
After select backup location, the Backup
button is available
Press Cancel to cancel the backup. Press Backup to execute.
6.5.2.2 Restore system
Press Restore system from selected back
up to restore system.
Press Back to previous step.
Press Restore from file to select the restore
file.
Press Cancel to cancel the restore.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
After select the restore file, the Restore
button is available.
Press Cancel the restore. Press Restore to execute.
6.5.3 Export event logs
Press Export event logs to access the
export event logs menu.
Press File name to change file name. Press Location to select the export location. Press Cancel back to previous step.
After select the export location, the Export button is available.
Press Cancel to cancel export. Press Export to execute.
6.5.4 Factory reset
Press Factory reset button to reset to factory default settings.
Press Cancel to return to previous screen
without resetting.
Press Reset to perform a reset.
CAUTION:
All methods and test results will be deleted when performing a factory reset.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
After factory reset, press OK to restart
instrument.
6.5.5 Software update
Press Software update to access the
update the software.
Press update from file to select file to
update.
Press Cancel to return to previous screen.
After select the update file, the Update
button is available.
Press Update to execute.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Method library in Home screen.
Press method name to check method settings and load to current method.
Press Back to Home screen. Press New to create a new method. Press Export to export methods. Press Import to import methods.
Enter the Method 1 to check the settings. Press Back to previous screen. Press Delete to delete this method. Press Edit to edit method settings. Press Load to load this method as current
method.
Press Method Library on the Home screen. The method library is displayed.
Complete information on how to use the library is covered in this section.
Press New button to start a new test.
The screen will change to key in the new method name.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Enter either a method name or identifying number and press the OK button.
Press the Drying profile button.
Standard Profile
Fast Profile
Press the Drying Temperature button to access the temperature setting menu.
Enter the desired drying temperature and press the OK button.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
The following settings can be selected for the switch-off criterion: — Timed switch-off — Auto switch-off
Timed switch-off
With this switch-off criterion selected, the measurement lasts until the preset drying time has elapsed (the display provides you with continuous information on the drying time).
Using the +/- buttons, enter the desired
drying time and press the OK button.
Press the Displayed result button.
Select %MC, %DC, %RG or g.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Select the Target Weight.
Enable the Target weight and press ON button.
Set the target weight and press OK button.
Press Target weight tolerance to set the
tolerance of target weight.
Set the target weight tolerance in %.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
The check weighing bar indicates the sample weight.
Press the Starting mode button.
Automatic
This mode can be used for most kinds of samples. When closing the sample chamber, the weight of the sample is recorded and the measurement is started. It is the factory default setting.
The analyzer comes with a default method preinstalled. Follow the instruction message to start a measurement directly. Press parameters area to edit.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
After change the settings, the Save button
available.
Press Save to save to default method.
Press Export to export methods.
Press File name to change export file name. Press Location to select the export location. Press Export selection to select methods to
export.
Press Cancel to previous screen.
After select the export location and methods, the
Export button is available. Press Export to execute.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Import to import methods.
Press Import from file to select the file to import. Press Cancel to previous screen.
After select the import file, the Import button is
available.
Press Location to change the import location. Press Import selection to select methods to
import.
Press Cancel to previous screen.
Press Results button on the Home screen to access the results menu.
Select the method name to list all the test results of the method.
Press Back to return to Home screen.
8.1.1 Result data and curve
Press result item to access the detail data.
Press Back to previous step. Press Delete to delete the results. Press Export to export these results. Press Statistics to access statistics menu.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press Back to previous step. Press Delete to delete current result. Press Export to export current result. Press Curve to show result curve.
Press Back to results list. Press Delete to delete current result. Press Export to export current result. Press Data to show detailed result data.
8.1.2 Statistics
Select the method name to list all the test results of the method.
Press Statistics to access statistics menu.
Press result item to select.
Press to select all items.
Press Cancel back to previous step.
After select the items, the OK button is available. Press result item to remove.
Press to remove all items.
Press Cancel back to previous step. Press OK to execute.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
8.1.3 Result export
Press Export to export all the results.
Press File name prefix to edit file name. Press Location to select export location. Press Export selection to select export
results.
Press Cancel to previous step.
After select the export location, the Export
button is available.
Press Cancel to previous step. Press Export to export results.
Press result item to check one result data and export it.
Press Export to export this result.
Press Export to export this result.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Press File name to edit file name. Press Location to select export location. Press Cancel to previous step.
After select the export location, the Export button is available.
Press Cancel to previous step. Press Export to the export result.
Press Help button on the Home screen to access the help information menu.
Select Level Assist button to access it. Please adjust the level feet accordingly.
Press screen to exit.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Weight adjustment printout
Temperature adjustment printout
WEIGHT ADJUST EXTERNAL
Weight Adjustment end 19.Dec 2017 11:20
Type MB90 SNR 1234567
Nominal Weight 50.000 g Actual Weight 50.001 g Difference 0.001 g Cell Temperature 27.19 °C Adjustment Done
Weight ID:
............................................................
Signature:
............................................................
------------END---------------
-TEMPERATURE ADJUSTMENT-
Temperature Adjustment end 19.Dec 2017 17:00
Type MB90
SNR 1234567
Temp1 target 100.0 °C Temp1 actual 100.0 °C Temp2 target 160.0 °C Temp2 actual 166.0 °C
Adjustment Done
Temp Kit ID:
......................................................
Signature:
.....................................................
------------END----------------
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Test data printout
Statistics printout
---MOISTURE DETERMINATION---
Type MB90 SNR 1234567 SW(Drying Unit) 1.30 SW(Terminal) 1.10.01
Company ABC Department QC Instrument ID 01
Method Name 01 Drying Prog Standard Drying Temp 100 °C Switch Off Auto
Start Weight 10.098 g 00:00:30 9.17 %MC 00:01:00 12.35 %MC 00:01:30 15.28 %MC 00:02:00 21.94 %MC
00:03:13 24.64 %MC
00:05:00 29:36 %MC 00:05:21 31.94 %MC Total Time. 00:05:21 Dry Weight 6.873 g
Moisture Content 3.225 g
End Result . 31.94 %MC Note Signature:
………………………………………
Measurement end 19.Dec 2017 15:35
-----------------------END--------------------
--- STATISTICS DATA ---
Halogen Moisture Analyzer Type MB90 SNR(Drying Unit) 1234567 SNR(Terminal) SW(Drying Unit) 1.30 SW(Terminal) 1.10.01
Sample Number 6 Last Data 88.12 %MC Mean Value 87.01 %MC Standard Deviation 2.06 %MC Minimum Value 83.47 %MC Maximum Value 88.98 %MC
Signature:
………………………………………
---------------END------------------
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Command
Character
Description
V
Print SR Version
TIM
Print Current Time
DAT
Print Current Date
1
N/C 2 <–
Date Out (TXD)
3
–>
Date Out (RXD)
4&6
N/C
5
Ground
7
–>
Clear to send (CTS)
8
<–
Request to send (RTS)
9
N/C
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
WARNING: Electric Shock Hazard. Disconnect the equipment
from the power supply before cleaning.
Open the cover and remove the pan handle, sample pan, draft shield, and heat shield from the instrument before cleaning.
Use a lint-free cloth for cleaning.
Clean the exterior of the instrument and the drying compartment components with a mild cleaning agent. Although the housing is extremely rugged and resistant to solvents, never use abrasive cleaning agents or solvents!
Ensure that no liquid enters the interior of the instrument.
Replace the components after cleaning.
Cleaning Temperature Sensor and Protective Glass
Check the protective glass and the temperature sensor for debris which could impede the operation. If the glass appears dirty, clean the surface facing the compartment using a commercial glass cleaner. If the sensor is dirty, clean using a mild cleaning agent.
WARNING: Do not spray the glass cleaner
directly on the glass. Instead, spray it on a cloth and use the cloth to clean.
Draf t Shield
Pan Support
Heat Shield
Pan Handle
Sample Pan
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Removing Glass for Cleaning
If the inside of the glass is dirty, lift up the heating chamber, release the glass for cleaning by pressing both sides of the glass holder.
Attention: Caution when releasing the glass from the
glass holder.
Reassemble after cleaning.
WARNING: Do not touch or clean the
Halogen heater.
Cleaning air inlet
The air inlet of the fan is located at the rear of the instrument and its exterior should be cleaned from time to time to free it from any dust deposits.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
WARNING: Electric Shock Hazard. Disconnect the
equipment from the power supply before cleaning.
Using a test pen, turn the fuse holder to the left (counterclockwise) and remove the fuse.
Check the condition of the fuse. Replace blown fuse by those of the same type with the same rated value (6.3A 250VAC for 100-120VAC power supply or 2.5A 250VAC for 200-240VAC power supply according to the heating element).
Attention: If the fuse is good and power is available at the
outlet, the cord or instrument may be defective. Try a new cord. If this does not work, the instrument should be sent back for servicing.
The use of a fuse of a different type or with a different value, or bridging or shunting the fuse is not allowed and can possibly cause a hazard to your safety and lead to instrument damage!
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Description Ohaus Part No.
Reusable Sample Pan 80252478 Security Locking Cable 80850043 Printer SF40A 30045641 Sample Pans 90mm diameter 80850086 Glass Fiber Pads 80850087 In-use Cover 30284478 Temperature Calibration Kit 11113857
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Model
MB90
Capacity
90
Readability
0.01%/0.001g
Repeatability (Std Dev) (g)
0.18% (3g sample)
0.02% (10g sample)
Moisture range
0.01% to 100%
(0.01% to 1000% for regain mode)
Heating Element
Halogen
Drying Programs
Standard, Fast
Temp range
40°C - 200°C
Switch-off Criteria
Timed, Auto, manual
Adjustment
External adjustment mass - 50g
Power
100 V – 90 VAC 5A 50/60 Hz
200 V – 240 VAC 2.5A 50/60 Hz
Operating temperature range
50° to 104°F / 10° to 40°C
Display type
4.3’, QVGA, TFT touch screen
Display results
%moisture, %solids, %regain, time, temperature, weight, method name, drying curve and statistics
Pan size (mm)
90
Interface
RS232, USB host, USB device,
Adjustable Feet and Level
Yes
Dimensions (WxHxD) (cm)
21x18x35
Net wt. (kg)
5.23
Shipping wt. (kg)
8.35
Marking
Standard
This product conforms to the EMC directive 2004/108/EC, the Low Voltage Directive 2006/95/EC. The complete Declaration of Conformity is available online at www.ohaus.com.
AS/NZS 61000.6.1, AS/NZS 61000.6.3
CAN/CSA C22.2 No. 61010-1, UL Std No. 61010-1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
The Batteries Directive 2006/66/EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States. To comply with this Directive, this device has been designed for safe removal of the batteries at end -of-life by a waste treatment facility.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment.
If you have any questions, please contact the responsible authority or the distributor from which you purchased this device.
Should this device be passed on to other parties (for private or professional use), the content of this regulation must also be related.
For disposal instructions in Europe, refer to www.ohaus.com/weee Thank you for your contribution to environmental protection.