ODM RP 450-02 Quick Start Manual

M-UM037-01 Page 1
RP 450-02 Optical Power Meter
Quick-Start Guide
The RP 450-02 optical power meter is an essential tool for verifying the proper function
of ber optic networks. This document will serve as an overview of the major features and functions of the device.
1 2
3 4
1. Connector Adapter - Interchangeable
The RP 450 comes with a 2.5mm universal adapter. Contact ODM for additional adapter options.
2. Low Battery Indicator
“BATT” will appear on the LCD when the CR2 3V battery needs to be replaced.
3. 2kHz Tone Indicator
The RP 450 will beep and show “2kHz” on the LCD when a 2kHz modulated signal is recognized.
4. dB/dBm Button
Press this button to toggle between dB and dBm modes. Hold this button down when in dBm mode to set a reference on the RP 450.
5. LCD Screen
This screen displays the measured value, currently-selected wavelength, and status indicators.
6. Measurement Mode Indicator - dB/dBm
dB is a loss measurement. dBm is an absolute power measurement.
7. Wavelength Button
Hold this button to switch the RP 450 between the 850, 1300, 1310, 1490, 1550, 1611, and 1625nm wavelengths. Press the button once to immediately display the current wavelength. The currently-selected wavelength will briey appear onscreen every 8 seconds.
8. Power Button
Press this button once to turn the RP 450 on. Press once again to turn the RP 450 o. The RP 450 is set to turn o after 5 minutes. For continuous usage, press and hold the power button for 3 seconds when powering on.
Device Overview
M-UM037-01 Page 2
Unpacking and Inspection
The RP 450 comes supplied with a soft carry case, 2.5mm adapter (on the unit), 3.0V CR2 battery, and instruction card. Contact ODM immediately if any part of the unit or packaging is damaged or otherwise unsatisfactory.
Low Battery
The RP 450 will provide over 1000 operations under normal conditions. When the BATT indicator is shown on the device screen, the CR2 battery should be replaced.
Optical Connector Interface
The RP 450 is equipped with a universal connector adapter which allows the interface of
many popular style connectors. This connector adapter interface should be kept covered and protected from contamination. Care must be taken to avoid objects that could damage the glass surface of the detector mount. If scratches or breaks occur on the surface, please contact ODM for proper service.
Warranty and Calibration
All ODM products come with a 2 year warranty. ODM recommends recalibration every two years to ensure adherence to NIST measurement standards. The rst recalibration is free within the warranty period. Contact ODM for return information.
Maintenance
Specications
OPTICAL POWER METER
Detector Type -02: InGaAs / -04: Filtered InGaAs
Measurement Range -02: +6 to -70dBm / -04: +23 to -45dBm
Wavelength Range 850nm to 1650nm
Selectable Wavelengths 850 / 1300 / 1310 / 1490 / 1550 / 1611 / 1625 nm
Resolution 0.01dB
Absolute Accuracy ± 0.25 dB
Optical Interface Universal 2.5mm (Additional Adapters Available)
Display LCD
Tone ID 2kHz
Power Push Button Toggle / Auto OFF
Battery CR2
Dimensions 6.1″ x 0.94″ x 0.75″ (15.5 x 2.38 x 1.9 cm)
Weight 3 oz (85.4 g)
Contact ODM
Phone: 603-524-8350
Email: tech.support@odm.ripley-tools.com
Web: www.odm-inc.com
Loading...