ODA OPE, OPE-S, OPE-Q, OPE-DI, OPE-QI User Manual

...
www.odacore.com
Programmable DC Power Supply
OPE-QI Series
User Manual
Manual Part NO. 018OPE-QI-2.0
Legal Notices
Warranty Certification
The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, ODA disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. ODA shall not be liable for errors or for incidental or be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should ODA and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control.
Assistance
This product comes with the standard product warranty. Warranty options, extended support contacts, product maintenance agreements and customer assistance agreements are also available. Contact your nearest ODA Technologies. Sales and Service office for further information on ODA Technologies. full line of Support Programs. Refer to below information.
www.odacore.com
oda@odacore.com
82-2-1800-8644
Waste Electrical and Electronic Equipment
The affixed product label (see right) indicates that you must not discard this electrical/electronic product in domestic household waste. Do not dispose in domestic household waste. To return unwanted products, contact our local ODA distributors, or call us for more information.
Manual Editions
Manual Part Number: 018OPE-QI-2.0 Edition 2, February, 2018 Printed in ROK Reprints of this manual containing minor corrections and updates may have the same printing date. Revised editions are identified by a new printing date.
Safety Notices
p
d
p
y
The following general safety precautions must be observed during all phases of operation of this instrument. Failure to comply with these precautions or with specific warnings or instructions elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. ODA Technologies assumes no liability for the customer's failure to comply with these requirements.
eneral Do Not Modify the Instrument
G Do not use this product in any manner not specified by the manufacturer. The protective features of this product may be impaired if it is used in a manner not specified in the operation instructions.
Do not install substitute parts or perform any u nauthorized modification to the product. Return the product to an ODA Sales and Service Office for service and repair to ensure that safety features are maintained.
round the Instrument
G This product is a Safety Class 1 instrument (provided with a protective earth terminal). To minimize shock hazard, the instrument unintended o chassis and cover must be connected to an qualified service personnel. electrical ground. The instrument must be connected to the ac power mains through a grounded power cable, with the ground wire firmly connected to an electrical ground (safety ground) at the power outlet. Any attention to an operating procedure, practice, or interruption of the protective (grounding) the like that, if not correctly conductor or disconnection of the protective to, could result in damage to the product or loss earth terminal will cause a potential shock hazard that could result in personal injury. C Before Applying Power are fully understood and met. Verify that all safety precautions are taken. Make all connections to the unit before W applying power. Note the instrument's A WARNING notice denotes a hazard. It calls external markings described under attention to an operating procedure, practice,
Fuses death. Do not proceed beyond a WARNING The instrument contains an internal fuse, which notice until the indicated conditions are fully is not customer accessible.
n Case of Damage
I Instruments that appear damaged or defective should be made inoperative and secured against
eration until they can be repaired b
AUTION
C
AUTION notice denotes a hazard. It calls
A C
erformed or adhere
of important data. Do not proceed beyond a
AUTION notice until the indicated conditions
ARNING
rofrep yltcerroc ton fi ,taht ekil eht ro "slobmyS ytefaS" med or
adhered to, could result in personal injury or
understood and met.
D
o Not Operate in an Explosive Atmosphere Do not operate the instrument in the presence of flammable gases or fumes.
D
o Not Remove the Instrument Cover Only qualified, service-trained personnel who are aware of the hazards involved should remove instrument covers. Always disconnect the power cable and any external circuits before removing the instrument cover.
Safety Symbol
Direct current
Both direct and alternating current Three phase alternating current
Earth (ground) terminal
Frame or chassis terminal
Neutral conductor on permanently installed equipment
Alternating current
Protective earth ground terminal.
Terminal is at earth potential.
Line conductor on permanently installed equipment.
Standby supply. Unit is not completely disconnected from ac mains when switch is off
Out position of a bi-stable push switch
Caution, hot surface
ylppus ffO ylppus nO
In position of a bi-stable push switch
Caution, risk of electric shock
Caution, refer to accompanying documents
CONTENTS
1. General Information
------------------------------------------------------------------------------
4
1-1. Feature
------------------------------------------------------------------------------------------------------
4
General Feature Remote Interface
Calibration
1-2. Accessories & Options
------------------------------------------------------------------------------
5
Accessories Options Homepage and Reference(www.odacore.com)
1-3. Check
----------------------------------------------------------------------------------------------------------
6
Instrument Check Electrical Check
1-4. Operation Conditions
----------------------------------------------------------------------------------------------------------
6
1-5. Check Before Power On
----------------------------------------------------------------------------------------------------------
7
Output Terminal Check Power Cord Check Input Power Line Check
1-6. Check After Power On
----------------------------------------------------------------------------------------------------------
9
Display Procedure on the LCD Default Setting Values
1-7. Installation
----------------------------------------------------------------------------------------------------------
10
Cooling Bench Operation Rack Mounting
2. Front Panel, Rear Panel Composition & Function
----------------------------------------------------------------------------------------------------------
12
2-1. Front-Panel
----------------------------------------------------------------------------------------------------------
12
2-2. Front-Panel Setting
----------------------------------------------------------------------------------------------------------
20
2-3. Rear Panel Composition
----------------------------------------------------------------------------------------------------------
21
2-4. Output Check
----------------------------------------------------------------------------------------------------------
22
Voltage Output Check Current Output Check
- 2 -
3. Front-Panel Operating
---------------------------------------------------------------------------------------------
23
Overview 3-1. Constant Voltage Operation (CV)
------------------------------------------------------------
24
3-2. Constant Current Operating (CC)
------------------------------------------------------------
25
3-3. I/O & LOCAL
---------------------------------------------------------------------------------------------
26
RS232C Setting RS485 Setting (Option) RS232C & RS485 Communication Specification
RS232C Connection Drawing
RS485 Connection Drawing
3-4. STORE
30
3-5. RECALL
30
3-6. OUTPUT ON/OFF
--------------------------------------------------------------------------------------
31
3-7. V/I & LMT DISPLAY
31
3-8. P1/P2 Select & TRACKING
--------------------------------------------------------------------
32
3-9. Remote Voltage Sensing (option)
--------------------------------------------------------------------
34
3-10. Last Memory (option)
-------------------------------------------------------------------------
34
4. SCPI Command
-------------------------------------------------------------------------------------------
35
4-1. Commands Syntax
---------------------------------------------------------------------------------------------
35
4-2. Commands
----------------------------------------------------------------------------------------------------
35
4-3. Channel Select Command
-----------------------------------------------------------------------------
36
4-4. Output Voltage / Current Setting & Operating Command
-----------------
37
4-5. Protection Command (option)
----------------------------------------------------------------------------------
37
4-6. Measure Command
----------------------------------------------------------------------------------
38
4-7. System Command
-----------------------------------------------------------------------------------
38
5. Caution
--------------------------------------------------------------------------------------------------------
40
---------------------------------------------------------------------------------
- 3 -
------------------------------------------------------------------------------------------------------
-----------------------------------------------------------------------------------------------------
1. General Information
1-1. Feature
ODA Technologies' OPE-Series Single & Dual Output products are high performance programmable DC Power
supply with RS232C or RS485 interface based on SCPI (Standard Commands for programmable Instruments)
protocol and the combination of bench-top and system features in these power supplies provides versatile
solutions for your design and test requirements in the industrial fields, R&D institute center and education fields.
General Feature
Easy-to-use knob control settings2-Line * 16Char LCD Display(Ch1 ,Ch2 Voltage/Current Display)Output voltage & current On/OFFAlarm beep when events occurExcellent Load Regulation & Line RegulationSave & Recall 5 Operating StateSize(2U * 19inch Half-Rack , 19inch Rack Compatibility)Safety power down mode when detecting system malfunction
Remote Interface
RS232C or RS485 InterfaceSCPI(Standard Commands for Programmable Instruments) CompatibilityHigh speed setting & measumentAbundant Commands I/O setup easily done from front-panel
Insulation from instrument and floating logic embodiment
Built-in SCPI Programming error check function
Calibration
Software Calibration, no internal adjustments requiredEasy calibration by using PC interface
- 4 -
1-2. Accessories and Options
Accessories
1 AC Input Power CordOutput 1 (+), (-) Terminal. (Part number : OE-LW-BCW-2.0)User's Manual 1 piece.
Option
RS485 Module RS485 to RS232 ConvertorRS232C Cross Cable 1M, 2M, 4M, 10M100V, 110V, 115V, 230V ± 10% , 50~60Hz AC Power InputRear Output TerminalRack mount Support (Bracket and Shelf)
Homepage and Reference (www.odacore.com)
Windows application demo version included manual
National instrument company's LabVIEW VI supplyTechnical information for Power Supply included.
- 5 -
1-3. Check
When you receive your power supply, inspect it for any obvious damage that may have occurred during shipment. If any damage is found, notify the carrier and the nearest ODA Sales Office immediately. Warranty information is shown in the front of this manual. Keep
the original packing materials in case the power supply has to be returned to ODA
Technologies in the future. If you return the power supply for service, attach a tag identifying the owner and model number. Also include a brief description of the problem.
Mechanical Check
Check the broken key, encoder switch, power switch.
Check the broken output terminals.
Check the panel surfaces are free of dents and scratches.Check the cabinet is free of scratches.Check the LCD display is not scratched or cracked.
Electrical Check
When turning on the power, it shows instrument model and ODA website at first.Check the model Number is matched with displayed model number.After it shows "**OUTPUT OFF**" message, verifies to a high level of confidence
that the power supply is operating in accordance with its specifications.
Service Center : 82-70-5032-2926 Home page : www.odacore.com
1-4. Operating Conditions
The instrument is designed for following environmental conditions in order to use optimized condition.
Environmental Temperature : 0 ∼ 40℃
Relative Humidity : 80% RH
Operating altitude : ≤ 2000m
No vibration
Avoid the electric magnetic field
- 6 -
Note
1-5. Check Before Power On
Output Terminal Check
Check the variable output and fixed output terminals.
Front Output Rear Output
< figure 1-1> OPE-S Series
Front Output Rear Output
< figure 1-2> OPE-DI Series
OPE-QI Series
OPE-Q Series
< figure 1-3> OPE-QI , Q Series
Floating the power supply output more than ±60 Vdc from the chassis presents an
electric shock hazard to the operator.
Power Cord Check
Your power supply is equipped with a 3-wire grounding type power cord; the third conductor being the ground. The power supply is grounded only when the power­ line cord is plugged into an appropriate receptable. Do not operate your power supply without adequate cabinet ground connection.
WARNING
- 7 -
Input Power Line Check
You can operate your power supply from a nominal 198-242 V single phase AC power source at 47 to 63 Hz. AC 100V, 110V, 115V, 230V input power is optional.
Refer to chapter "1-2, Accessories & Options"
In order to prevent the instrument severe damage from overload, fuse is installed in inlet case. If the fuse is also repeatedly when power turns on, check the input power line or broken breaker and then call to nearest ODA Technologies A / S Center Input power connection is following.
< figure 1-3> Output capacity 90W ~ 300W or 400W ~ 600W Input Power
< figure 1-4> Output Capacity Over 400W Customized Input Power
- 8 -
1-6. Check After Power On
When turning on the power switch, the front-panel display will light up briefly while the instrument performs its power-on default value setting. And also keep the ex-remote interface setting mode, voltage value is zero and current value is max value.
Display Procedure on the LCD
Display "OPE-303S V1.0" and website address
Visit ODA Technologies website. Get the manuals, windows application software demo, and upgrade
information & technical support.
Display "**OUTPUT OFF**"
By using front panel key and encoder switch, set the voltage/current and functions.
Default Setting Values
Output Voltage : 0V Cursor location : Default voltage.Output Current : Limit setting max value Voltage : 1V UnitOutput Select : Voltage Current : 0.1A UnitRemote Interface : ex-remote interface setting modeAfter standby mode : Display "**OUTPUT OFF**"
The RS232C is attached in the instrument when the power supply is shipped from the
factory for remote interface configuration and baud rate is set 9600bps at first time.
In case of choosing RS 485 interface, address no. is 05 when this is shipped from the factory.
For your safety, Setting Voltage and Current will not be set up after turn-on. After power on, power supply will be in "**OUTPUT OFF**" Mode.
Note1
Note2
- 9 -
1-7. Installation
Cooling
The power supply can operate without loss of performance within the
temperature range of 0 to 40℃, and with derated output current from
40℃ to 55℃. A fan cools the power supply by drawing air through the rear
panel and exhausting it out the sides. Using an ODA rack mount will not
impede the flow of air.
< figure 1-6 Bottom view>
Bench Operation
Your power supply must be installed in a location that allows sufficient space
at the sides and rear of the power supply for adequate air circulation.
< figure 1-7 > 2U * 19inch Half
< figure 1-8> 19inch
- 10 -
ventilation opening
Rack Mounting
Power supply is designed for 2U * 10 inch Rack ( Half Size ) remove the rubber on the bottom and put the rack bracket
< figure 1-9 > 2U * 19inch Half Setting
This is designed for 19 inch size rack. Before installment, remove the rubber on the bottom and put the rack bracket
< figure 1-10 >19inch fixed product