All rights reserved. No part of this publication may be reproduced,
stored in a retr ieval system, or tr ansmitted, by any means, elect r onic,
mechanical, photocopying, recording, or otherwise, without written
permission from Ocean Optics, Inc.
This manual is sold as part of an order and subject to the condition that it
shall not, by way of trade or otherwise, be lent, re-sold, hired out or
otherwise circulated with out the prior consent of Ocean Optics, Inc. in any
form of binding or cover other than that in which it is published.
Trademarks
Microsoft, Windows, Windows 95, Windows 98 and Windows NT are
either registered trademarks or trademark s of Microsoft Corp or ation.
Limit of Liability
Every effort has been made to make this manual as complete and as
accurate as possible, but no warranty or fitness is implied. The information
provided is on an “as is” basis. Ocean Optics, Inc. shall have neither
liability nor responsibility to any person or entity with respect to any loss or
damages arising from the information contained in this manual.
Table of Contents
Introduction
Quick Start
Installing the PC2000 and OOIChem Software
Using OOIChem Software
Appendix C: Calibrating the Wavelength of Your Spectrometer
....................... 37
Introduction
Ocean Optics miniature fiber optic spectrometers and accessories have revolutionized the analytical
instrumentation market by dramatically reducing the size and cost of optical sensing systems. More than
10,000 Ocean Optics spectrometers have been sold worldwide -- striking evidence of the far-reaching
impact of low-cost, miniature components for fiber optic spectroscopy. Diverse fields such as research and
development, industrial process control, medical diagnostics, environmental monitoring and o f course,
education have benefited from access to Ocean Optics technology.
In fact, Ocean Optics has its roots in education. It formed in 1989 when Florida university researchers
developed a fiber optic pH sensor as part of an instrument designed to study the role of the oceans in global
warming. The researchers soon formed Ocean Optics, Inc. and their ingenious work earned a Small
Business Innovation Resea rch grant from the U.S. Department of Energy. While d esigning the pHmonitoring instrument, the researchers wanted to incorporate with their sensor a spectrometer small enough
to fit onto a buoy and were surprised to discover none existed. So, they built their own.
In 1992, the founders of Ocean Optics filled a substantial need in the research community and changed the
science of spectroscopy forever by introducing a breakthrough technology: the S1000 Miniature Fiber
Optic Spectrometer, nearly a thousand times smaller and ten times less expensive than previous systems.
Due to this dramatic reduction in size and cost of optical sensing systems, applications once deemed too
costly or impractical using conventional spectrometers were now feasible.
CHEM2000 and CHEM2000-UV-VIS Spectrophotometer Systems
The CHEM2000 and CHEM2000-UV-VIS Spectrophotometer Systems utilize the optical bench of our
second-generation miniature fiber optic spectrometer, the S2000, by mounting it onto an A/D converter and
turning the system into a PC plug-in spectrometer.
CHEM2000
The CHEM2000UV-VIS also consists of four basic elements:
the PC2000-UV-VIS PC Plug-in Fiber Optic Spectrometer
(200-850 nm), a miniature deuterium tungsten light source
with integrated cuvette holder, a 300-µ
resistant optical fiber, and OOIChem operating software.
The light source supplies light to the sample. The light
transmitted through the sample is collected and sent to the spectrometer via the fiber. The spectrometer
measures the amount of light at each wavelength in the sampled spectrum. The A/D converter, on which
the spectrometer is mounted, transforms the analog data from the spectrometer into digital information that
is passed to a computer. Finally, the software performs basic acquisition and display functions on your data.
m solarization-
The CHEM2000 consists of four basic elements: a
PC2000 PC Plug-in Fiber Optic Spectrometer (350900 nm), a tungsten-halogen light source with
µ
integrated cuvette holder, a 400OOIChem operating software.
m optical fiber, and
CHEM2000-UV-VIS
PC2000 PC Plug-in Fiber Optic Spectrometer
The PC2000 is 2048-element linear CCD-array fiber optic spectrometer mounted on a 1 MHz A/D card that
fits easily into an ISA-bus slot in the PC. The spectrometer has a 25-µ
nm (FWHM). The 12-bit, 8-channel A/D card is single-ended and half-length. Though the standard
CHEM2000 and CHEM2000-UV-VIS are single-channel systems, additional slave spectrometer channels
can be purchased at any time to expand wavelength range or perform multiple tasks.
m slit and provides a resolution of 1.5
Light Source/Sample Holder
The ISS-2 Integrated Sampling System that comes with the CHEM2000 is a versatile light source/sample
holder with a tungsten-halogen bulb, a built-in filter slot for color-correcting and other uses and a sample
holder for 1-cm square cuvettes. The light source/sample holder is SMA-terminated for easy coupling to
optical fibers and has a 5-mm diameter f/2 collimating lens.
The ISS-UV-VIS Integrated Sampling System comes with the CHEM2000-UV-VIS. This light
source/sample holder has deuterium and tungsten bulbs, and a cuvette holder for 1-cm cuvettes. It occupies
7-1/2" x 4" and weighs <1 lb. The cuvette holder attaches directly to the light source and has a 5-mm
diameter f/2 collimating lens to collect the light and funnel it to the solarization-resistant optical fiber.
Optical Fiber
Included as part of the CHEM2000 is a 400-µm diameter, single-strand UV/VIS optical fiber that couples
easily via SMA terminations to the PC2000 and light source/sample holder. This silica-core, silica-clad
fiber is 2 meters in length. For the CHEM2000-UV-VIS, the included optical fiber is a 300solarization-resistant patch cord that carries light from the sample to the spectrometer.
µ
m diamter
OOIChem Spectrometer Operatin g Software
OOIChem is our standard Windows operating software for the CHEM2000 and CHEM2000-UV-VIS
systems. It is a basic acquisition and display software that provides a real-time interface to a variety of
spectral-processing functions. OOIChem allows users to perform basic spectroscopic measurements such as
absorbance, transmission, relative irradiance, and concentration. OOIChem operates with Windows 95,
Windows 98 and Windows NT. Visit our web site at www.OceanOptics.com/Software_Updates.asp to
download free OOIChem upgrades.
OOIBase32 Spectrometer Operating Software
OOIBase32 is our standard spectrometer operating software that we provide free of charge to all customers.
While OOIChem is a basic acquisition and display program, OOIBase32 is user-customizable and a much
more advanced acquisition and display program. With OOIBase32 you have the ability to control all
system parameters; collect data from up to 8 spectrometer channels simultaneously and display the results
in a single spectral window; perform reference monitoring and time acquisition experiments; and use
numerous editing, viewing and spectral processing functions. At any time, users can receive free
OOIBase32 updates from our web site at www.OceanOptics.com/Software_Updates.asp.
These are the standard components for the CHEM2000 and CHEM2000-UV-VIS systems. However, we
offer a complete line of light sources, sampling holders, in-line filter holders, flow cells, and other sampling
devices; an extensive line of optical fibers and probes; and collimating lenses, attenuators, diffuse
reflectance standards and integrating spheres. All components have SMA terminations so that changing the
sampling system is as easy as unscrewing a connector and adding a new component or accessory.
This modular approach -- components are easily mixed and matched -- offers remarkable applications
flexibility. Users pick and choose from hundreds of products to create distinctive systems for an almost
endless variety of optical-sensing applications
2
Packing List
A packing list comes with each order. It is located inside a plastic bag attached to the outside of the
shipment box. The invoice is mailed separately. The items listed on your packing slip include all of the
components i n your order. However, some items on your packing list are actually i t ems installed into your
spectrometer, such as the grating and slit. The packing list also includes important information such as the
shipping address, billing address, and components on back order.
Wavelength Calibration Report
In your shipment box, you will find your spectrometer in a silver-gray anti-static bag. Your software
diskettes and Wavelength Calibration Report are packaged with this manual. A Wavelength Calibration
Report comes with every spectrometer channel, contains important information specific to your
spectrometer, and should be retained. Use the calibration coefficients and intercept numbers on your
Wavelength Calibration Report to calibrate the wavelength of your spectrometer.
Upgrades
Customers sometimes find that they need Ocean Optics to make a change or an upgrade to their system. In
order for Ocean Optics to make these changes, the customer must first contact us to obtain a Return
Merchandise Authorization (RMA) number and to receive specific instructions when returning a product.
3
Quick Start
The CHEM2000 and CHEM2000-UV-VIS systems are easy to set up, allowing the user to start collecting
data within minutes. This Quick Start provides brief instructions on installing your PC2000, installing and
configuring the soft ware, and setting up your system. To find detailed directions on a specific component,
Table of Contents
see the
software instructions begin on page 9.)
Step 1: Interface the PC2000 to your PC
The PC2000 is our fiber optic spectrometer mounted onto an A/D converter. The entire instrument is
installed into your PC. But first you must find available Base Address and IRQ values.
1. The default settings for our A/D products are a Base Address (or Input/Output Range) of 768 decimal
and an IRQ of 7. You will need to match Base Address and IRQ settings on the A/D card to available
settings in your computer. First determine which settings are not being used by other hardware devices.
•If you have Windows 95/98, go to
icon. Choose the
devices. Under
Remember that these I/O settings are expressed in hexadecimal and correspond to our Base
Address, which is given in decimal, followed by the hexadecimal equivalent in parenthesis.
•If you have Windows NT, go to
NT Diagnostics
Select the
2. Note these available settings. When you first run OOIChem, you must enter these values in the
“Configure Hardware” dialog box.
3. Turn off the computer and take off the computer cover. Ground yourself to the computer chassis or
power supply and remove the PC2000 (spectrometer-A/D combination) from its static-shielded bag.
4. If necessary, change the position of the switches on the A/D converter. For the PC2000, there is only one
bank of switches on the A/D converter: the Base Address may be changed via the first 6 switches and the
IRQ may be changed via the last 3 switches. (See
5. Insert the PC2000 into an ISA-bus slot. Make sure the connections are snug and restart your computer.
. (Detailed PC2000 installation instructions begin on page 6. Detailed OOIChem
Device Manager
View Resources
. Click on the
I/O Port
button. Find an available I/O Range (Base Address).
Start | Settings | Control Panel
tab and double-click on “Computer” at the top of the list of
, note available settings -- numbers unassigned to hardware.
Start | Programs | Administrative Tools (Common) | Windows
Resources
tab. Select the
Appendix A
IRQ
on page 34 for switch positions.)
. Double-click the
button. Find an available IRQ.
System
Step 2: Install OOIChem Software
Before installing OOIChem, make sure that no other applications are running.
1. Insert “Disk 1” into your floppy drive. (When prompted, insert Disks 2 and 3.) Execute
2. At the “Welcome” dialog box, click
3. At the “Destination Location” dialog box, you can choose
Next>
Click
4. At the “Backup Replaced Files” dialog box, select either
you select Yes, you can choose
5. Select a Program Manager Group. Click
6. At the “Installation Complete” dialog box, choose
7. When prompted to do so,
.
Step 3: Configure OOIChem Software
After you restart your computer, navigate to the OOIChem icon and select it. Now that the PC2000 and
software ha ve been installed, you ne ed to configure your software. The first time you run OOIChem after
installation, you must follow several prompts to configure your system before taking measurements.
Next>
.
Browse
restart your computer
to pick a destination directory. Click
Next>
. At the “Start Installation” dialog box, click
Browse
Yes
Finish>
when the installation is complete.
4
.
to pick a destination directory.
or No. We recommend selecting Yes. If
Next>
Setup.exe
.
Next>
.
.
Hardware Configuration
Configure Hardware
The
box are usually set only once -- when OOIChem is first installed and the software first opens.
1. Under
2. Under
3. Under
4. Under
5. For your setup, only these parameters apply to your system. Click OK. You can always change these
Spectrometer Type
A/D Converter Type
Base Address
available setting you found in your computer). Remember that the Input/output Range (Base Address)
you selected was expressed in hexadecimal. In this dialog box, the base address is given in decimal,
followed b y the hexadecimal equivalent in parent hesis. For example, “
address as 768 decimal and 300 hexadecimal.
IRQ
, choose the same setting as the dip switches on the A/D converter (and the same available
setting you fou nd in yo ur computer).
settings once OOIChem is fully operational by selecting
dialog box opens when you first run OOIChem. The parameters in this dialog
, choose S2000/PC2000.
, choose ADC1000/PC2000.
, choose the same setting as the dip switches on the A/D converter (and the same
768 (0x0300)
Spectrometer | Hardware Configuration
” gives the base
.
Spectrometer Configuration
At this point, OOIChem should be acquiring data from your spectrometer. There should be a dynamic trace
responding to light near the bottom of the displayed graph. Now that OOIChem is running, you need to
configure your system. Select
Coefficients.
•
coefficients for your CHEM2000 or CHEM2000-UV-VIS under
Trigger mode.
•
spectrometer for synchronizing data with an external event.
Graph and chart display mode.
•
spectrometer channel. Choose
of the graph and to view a chart displaying your kinetics experiment in the bo ttom half of the graph.
Flash Delay.
•
VIS systems do not come with a strobe light source.
Color Temperature.
•
irradiance measurements. For CHEM2000 users, your lamp has a color temperature of 3100 Kelvin.
For CHEM20000UV-VIS users, your light source does not have a known color temperature.
From the Wavelength Calibration Report that came with your system, enter the
Select No E xternal Trigger, unless yo u have wired an external triggering device to the
This function is for use with a strobe light source. The CHEM2000 and CHEM2000-UV-
Spectrometer | Spectrometer Configuration
Choose
Spectrum & Kinetics
Enter the color temperature of your reference light source used in relative
Spectrum Only
to view both real-time live spectra in the top half
to only view live spectra from one
Master
from the menu.
.
Acquisition Parameters
Set data acquisition parameters by choosing an integration period and selecting averaging and boxcar
smoothing values.
Text Box
Enter the operator name, or any other identifying text here. This text appears in your data files. You can
edit this text at any time.
Step 4: Connect Sampling Optics
If you have a CHEM2000, take your 400-µm fiber that came with your system, screw one end into the
SMA connector on the light source and screw the other end of the fiber into SMA connector on your
PC2000, which you have already installed in your computer. If you have a CHEM2000-UV-VIS, take your
µ
m solarization-resistant fiber that came with your system, screw one end into the SMA connector on
300the light source and screw the other end of the fiber into SMA connector on your PC2000.
Step 5: Receive Data
Run OOIChem in Scope Mode and take a reference spectrum and a dark spectrum (see the
Tutorial
irradiance mode to take your sample measurements.
section beginning on page 27 for details). Choose the absorbance, transmission, or relative
Experiment
5
Installing the PC2000 and OOIChem Software
PC2000 PC P
The
spectrometer mounted onto an ADC1000 A/D card. The
analog-to-digital converter card that connects our miniature fiber optic spectrometers to desktop PCs. The
ADC1000 is a 12-bit, 8-channel, single-ended A/D card. This sturdy spectrometer-A/D card combination
fits easily into an ISA-bus slot in the PC. The following directions apply to both CHEM2000 and
CHEM2000-UV-VIS users.
Each device in or connected to your computer is assigned specific settings; it’s similar to giving each
device its own name so that your computer will know what to call and how to recognize the device. In
order for your PC2000 to work as a device in your computer, it has to be assigned a Base Address setting
and an IRQ setting. The default settings for each are:
Base Address (I/O Range)768 decimal (300 hexadecimal)
IRQ (Interrupt Request)07
These default values are set on the A/D converter. There are dip switches on the A/D converter and their
positions determine the values. These default values are set in the operating software as well. Most of the
time, these default settings will work with your computer. However, if you have many devices installed in
your computer, you may have a conflict; other devices may be using these settings. If there is a conflict
with another device in your computer, you must change the positions of the switches on the A/D converter.
For the PC2000, there is only one bank of switches on the A/D converter: the Base Address may be
changed via the first 6 switches and the IRQ may be changed via the last 3 switches. To first check your
computer to see which settings are available, follow the instructions for the Windows system that you use.
LUG-IN SPECTROMETER
has our 2048-element linear CCD-array fiber optic
ADC1000 A/D C
ARD
is a high-speed ISA-bus
Windows 95 and Windows 98 Users:
Find Available Base Address and IRQ Settings
1. Go to
2. Choose the
3. Under
Start | Settings | Control Panel
Device Manager
your computer.
View Resources
available settings for both the Interrupt request (IRQ) and the Input/output (Base Address). When you
first run OOIChem, you must enter these values in the “Configure Hardware” dialog box. (Remember
that Input/output settings are expressed in hexadecimal.)
tab and double-click on “Computer” at the top of the list of devices in
, find available settings -- numbers unassigned to hardware. Note these
and double-click on the
System
icon.
6
For most computers, the default settings on our A/D converters work well. In the picture at left on
page 6, it appears that the Printer occupies IRQ Setting 07, but for most computers, our A/D
converters can share the IRQ 07 setting with a printer and conflicts will not arise. All computers have
multiple Base Address (Input/output) settings from which to choose.
Windows NT Users:
Find Available Base Address and IRQ Settings
1. Go to
2. In the “Windows NT Diagnostics” dialog box, click on the
3. Select the
4. Select the
5. Note these available settings. When you first run OOIChem, you must enter these values in the
Start | Programs | Administrative Tools (Common) | Windows NT Diagnostics
Resources
IRQ
button. Find an available IRQ -- a number unassigned to a device.
I/O Port
numbers unassigned to a device. (T he number is in hexadecimal.)
“Configure Hardware” dialog box.
With Windows NT, devices cannot share IRQs; each device must be assigned a unique IRQ.
button. Find an available I/O Range (Base Address) -- a number or range of
tab.
.
1. Turn off the computer and remove the computer cover.
2. Ground yourself to the computer chassis or power supply.
3. Remove the PC2000 (spectrometer-A/D card combination) from its static-shielded bag.
4. If necessary, change the position of the switches on the A/D converter. Position the switches to match
the availab le settings you found in the previous section -- numbers not bei ng used by other hardware
devices. See
5. Find an open ISA-bus slot and remove the slot protector.
6. Insert the PC2000 into an available expansion slot on the motherboard connector by gently rocking the
card into the slot. Make sure the card is fully seated in the motherboard before screwing the tab on the
PC2000 to the computer. Do not bend the card or move it from side to side once it is seated in the slot.
7. If you have a CHEM2000, take your 400-
the SMA connector on your PC2000. If you have a CHEM2000-UV-VIS, take your 300-µ
solarization-resistant fiber that came with your system and screw one end into the SMA connector on
your PC2000. Reinstall the cover.
Appendix A
Install the PC2000
on page 34 for switch setting positions.
µ
m fiber that came with your system and screw one end into
7
m
Install OOIChem
The resolution of your computer's monitor must be 800 x 600 or better to use OOIChem software. Before
installing OO IChem, make sure that no other applic ations are running.
.
Setup.exe
.
1. Insert “Disk 1” into your floppy drive. (When prompted, install Disks 2 and 3.) Execute
2. At the “Welcome” dialog box, click
3. At the “Destination Location” dialog box, you can choose
Next>
Click
4. At the “Backup Replaced Files” dialog box, select either
you select Yes, you can choose
5. At the “Installation Complete” dialog box, choose
6. When the installation is complete,
.
Next>
.
Browse
Yes
or No. We recommend selecting Yes. If
Browse
to pick a destination directory. Click
Finish>
you must restart your computer
.
to pick a destination directory.
Next>
.
Run OOIChem
After you restart your computer, navigate to the OOIChem icon and select it. The first time you run
OOIChem after the installation, you must enter data into the Configure Hardware dialog box.
Configure Hardware
Configure Hardware
The
in this dialog box are usually set only once -- when OOIChem is first installed and the software first opens.
1. Under
2. Under
3. Under
Spectrometer Type
S2000/PC2000.
A/D Converter Type
ADC1000/PC2000.
Base Address
the dip switches on the A/D converter (and the
same availab le setting you found in your
computer).
dialog box opens automatically the first time you run the software. The parameters
, choose
, choose
, choose the same setting as
Remember that the Input/output Range (Base
Address) you selected was expressed in hexadecimal.
In this dialog box, the base address is given in
decimal, followed by the hexadecimal equivalent in
parenthesis. For example, “
base address as 768 decimal and 300 hexadecimal.
4. Under
5. Only these parameters apply to your system.
IRQ
, choose the same setting as the dip
switches on the A/D converter (and the same
available setting you found in your computer).
(Ignore the other settings; they apply to other
A/D converters.) Click OK. You can always change these settings once OOIChem is fully operational
by selecting
If you do not see the “Configure Hardware” screen, exit the software. Then select
type c:\windows\ooidrv.ini for Windows 95/98 systems or c:\winnt\ooidrv.ini
for Windows NT systems. Notepad will open. Edit this file for our device driver by finding the
“Initialized” entry and making sure this line reads Initialized=0. Save the OOIDRV.INI file and
exit Notepad. Restart OOIChem. You should now see the “Configure Hardware” dialog box.
Spectrometer | Hardware Configuration
768 (0x0300)
” gives the
from the menu.
Start
|
Run
, and
8
Using OOIChem Software
HEM OPERATING SOFTWARE
OOIC
signal processing, display, and measurement functions. The 32-bit software program operates with Windows
95/98 and Windows NT and performs basic absorbance, emission and kinetics functions, which are presented
in an easy-to-follow format that is ideal for undergraduate-level students and beginning spectroscopists.
OOIChem software provides users with five different modes of operation: Scope, Absorbance,
Transmission, Relative Irradiance, and Concentration. In addition, the software allows users to control data
acquisition features such as integration period, averaging, and boxcar smoothing -- directly from the
spectral graph display. Users can acquire data by taking manual single scans or by running continuous
scans, and add into the spectral window as many as 8 previously saved overlay spectra.
provides users with a real-time interface to a variety of acquisition,
Users can also perform kinetics
experiments, whic h all o ws user s
to monitor and report up to 4
single wavelengths or up to 2
calculated values from a
combination of wavelengths -for example, an absorbance value
of 400 minus an absorbance value
of 700. A kinetics chart displays
the time series. When the
experiment is complete, the data
can be exported to an ASCII file
for additional processing.
Another exciting feature of OOIChem
is that users can control the
parameters for all system functions
such as acquiring data, designing the
graph display, and configuring the
cursor. Additional features of
OOIChem include the ability to save
data as ASCII files and to store and
retrieve sample spectra.
OOIChem software will be updated
and improved continuously. To
obtain free upgrades, visit out web
www.oceanoptics.com/Software_Updates.asp
site at
the password found on the back of your Software and Resources Library CD.
If you find that you need more advanced spectrometer operating software, we have included, free of charge,
OOIBase32, a sophisticated, 32-bit, user-customizable advanced acquisition program. See the included
OOIBase32 manual for a list of functions and features.
. In order to download free upgrades, you will need
9
Display Functions
Several functions are accessed not through the menu but through buttons and task bars directly on the display
screen, on the top and to the right of the graph area. (The resolution of your computer's monitor must be 800 x
600 or better to view OOIChem software.) From the display screen, you can choose a mode of operation,
configure the cursor, configure the graph, enter acquisition parameters, choose a mode to acquire data, take
reference, dark, and sample scans of your sample, and scale the graph.
Spectrometer Channel Selection
The Spectrometer Channel area allows you to animate the window for a
specific spectrometer channel. Though the standard CHEM2000 and
CHEM200-UV-VIS are single spectrometer channel systems, you can
purchase additional channels at any time. For now, select
Mode of Operation
Scope
The signal graphed in Scope Mode is the raw voltage coming out of the A/D converter. Once you open
OOIChem and it begins to acquire data, you see the raw voltage from the detector expressed in A/D
counts. This spectral view mode is unique to Ocean Optics. It allows you to control signal processing
functions before taking absorbance, transmission, and relative irradiance measurements. Be careful
when using this mode, as it represents a combination of several factors:
the intensity of the light source, the reflectivity of the grating and the
mirrors in the spectrometer, the transmission of the fibers, the response
of the detector, and the spectral characteristics of the sample.
Master
.
Scope Mode should primarily be used when selecting signal
acquisition parameters such as integration period, averaging and
boxcar smoothing; and when taking dark and reference scans.
Absorbance
Selecting this mode switches the spectral window into Absorbance Mode. Before this can occur, both a
dark and reference scan must be stored in Scope Mode. (See the
page 27.) Absorbance is calculated by the following equation. When this equation is evaluated for each
pixel of the detector, the absorbance spectrum is produced.
S
- D
λ
Aλ = - log
where S is the sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the
reference intensity at wavelength λ.
10
(
Rλ - D
λ
)
λ
Experiment Tutorial
beginning on
Transmission
Selecting this mode switches the spectral window into Transmission Mode. This is also the spectral
processing mode used for Reflection spectroscopy, as the math necessary to compute reflection is
identical to transmission. Before this can occur, both a dark and reference scan must be stored in Scope
Mode. (See the
calculated by the following equation:
%Tλ =
where S is the sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the
reference intensity at wavelength λ.
Experiment Tutorial
Sλ - D
λ
x 100%
- D
R
λ
λ
beginning on page 27.) The transmission of a solution is
10
Relative Irradiance
Selecting this mode switches the spectral window into Relative Irradiance Mode. The reference
spectrum must be made in Scope Mode with a blackbody of known color temperature. (CHEM2000UV-VIS users cannot make relative irradiance measurements because the light source that comes with
the system is not a blackbody source with a known color temperature.) A dark spectrum is usually
obtained by preventing light from entering the fiber that connects to the spectrometer. (See the
Experiment Tutorial
of a light source relative to a reference emission source. Relative irradiance is calculated by the
following equation:
Iλ = B
λ
(
where B is relative energy of the reference calculated from the color temperature in Kelvin, S is the
sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the reference intensity
at wavelength λ.
beginning on page 27.) Relative irradiance spectra are a measure of the intensity
S
- D
λ
λ
Rλ - D
)
λ
Concentration
Concentration is the amount of a specified substance in a solution. Graphs of absorbance vs.
concentration are known as Beer’s Law plots. These are calculated by first measuring the light that is
absorbed from a series of solutions with different known concentrations. The length of the sample -such as the path length of our cuvette holder -- and the wavelength chosen for monitoring the amount
of light absorbed, are constants. Then a linear plot derived from the scans of these standard solutions
with known concentrations is obtained. The plot is then used to determine the unknown concentrations
of solutions. (See the
The absorbance of a solution is related to the concentration of the species within it. The relationship,
known as Beer’s Law, is:
ε
A
=
λ
λ
where A is the absorbance at wavelength λ, ε is the extinction coefficient of the absorbing species
at wavelength λ, c is the concentration and l is the optical pathlength.
Experiment Tutorial
l
c
beginning on page 27.)
Cursor Function Bar
Sign
+
When the + is selected, the pointer becomes a crosshair symbol, enabling you to drag the cursor around
the graph.
Magnify Symbols
When the magnify symbol is selected, you can choose from among 6
magnify functions. The function chosen will remain in use until another
magnify icon or the crosshair symbol is selected. Clockwise, beginning with
the top left symbol, the magnify icons perform the following functions:
1. magnifies a specific area by clicking and dragging a box around an area
2. zooms in on the horizontal scale, but the vertical scale remains the same
3. zooms in on the vertical scale, but the horizontal scale remains the same
4. zooms in approximately one point vertical and horizontal, click once or press continuously
5. zooms out approximately one point vertical and horizontal, click once or press continuously
6. reverts to the last zoom function
Cursor Diamond
To move the cursor left or right in small increments in the graph area, click on the left and right
sections of the move cursor diamond. The top and bottom sections of the dia mond will send the
cursor to the next or previous channel in your system. (The CHEM2000 and CHEM2000-UV-
VIS are single channel systems. However, additional channels can be purchased at any time.)
11
Cursor Properties
In this bar, you can label the cursor and monitor the
cursor’s X value and Y value. To the right of the X and Y values
of the cursor is a cursor selection button that allows you to choose
a cursor style and a point style. You can also choose a color for
the cursor and whether or not to display the name of the channel
the cursor is currently re porting. Fina l ly, you can bring the cursor
to the center of the spectrum or center the spectrum around the
cursor’s current position.
Text Box
This box allows you to enter an operator name and any other text to identify your experiment. This text
appears in your data files. You can edit this text at any time.
Acquisition Parameters
Integration Period
Enter a value to set the integration period in milliseconds for an active spectrometer channel. The
integration period of the spectrometer is analogous to the shutter speed of a camera. The higher the
value specified for the integration period, the longer the detector “looks” at the incoming photons. If
your scope mode intensity is too low, increase this value. If the intensity is too high, decrease the value.
While watching the graph trace in Scope Mode, adjust the integration period and other acquisition
parameters until the signal intensity level is approximately 3500 counts.
Average
Enter a value to implement a sample averaging function that averages the
specified number of spectra. The higher the value entered the better the signal-tonoise ratio. The S:N improves by the square root of the number of scans averaged.
Boxcar Smooth
Enter a value to implement a boxcar smoothing technique that averages across
spectral data. This method averages a group of adjacent detector elements. A
value of 5, for example, averages each data point with 5 points (or bins) to its left and 5 points to its
right. The greater this value, the smoother the data and the higher the signal-to-noise ratio. However, if
the value entered is too high, a loss in spectral resolution results. The S:N impro ves by the square root
of the number of pixels averaged.
Reference Scan
Selecting the Reference button activates a prompt to make sure your light is on. You then must choose to
Store
either
blank in the sampling region. Storing a reference spectrum is requisite before the software can calculate
absorbance, transmission, and relative irradiance spectra. This command merely stores a reference spectrum.
To permanently save the reference spectrum to disk, select
or
Cancel
your reference scan. A reference spectrum is taken with the light source on and a
File | Save Spectral Values
from the menu.
Dark Scan
Selecting the Dark button activates a prompt to make sure the light path is blocked.
You then must choose to either
taken with the light path to the spectrometer blocked. Storing a dark spectrum is
requisite before the software can calculate absorbance, transmission, and relative
irradiance spectra. This command merely stores a dark spectrum. To permanently save
the reference spectrum to disk, select
Store
Cancel
or
File | Save Spectral Values
your dark sca n
.
A dark spectrum is
from the menu.
12
Subtract Dark
Selecting this box subtracts the current dark spectrum from the spectra being displayed. This command is
useful if you are trying to look at a change in an emission spectrum or are trying to eliminate from the
spectra fixed pattern noise caused by a very long integration period. The subtract dark spectrum function
only acts on spectra displayed in Scope Mode.
Acquire Data Modes
Chart Active
Spectrum & Kinetics
When
Configuration
Scan button. This function is responsible only for the Kinetics chart. By
deselecting this function, users can use the scan button and collect data in just
the spectrum section of the graph. If the function is enabled, choosing the scan
button results in the collection of data in both the spectrum section of the graph
and the kinetics section of the graph.
dialog box, the
is chosen as the
Chart Active
Scan/Stop Button
When in
Single mode, click on the Scan button to take a scan. The button depresses and
Stop replaces Scan. The button will stay depressed until the scan ha s b een completed (the time set in
the Integration Period box).
Single
mode, the Scan button acts as a snapshot. After selecting the
Graph and chart display mode
function becomes visible in the display area above the
in the
Spectrometer
When in
another scan will immediately begin. The button depresses and Stop rep laces Scan. Click on Stop to
halt the scanning process and discontinue acquiring data.
Continuous
mode, the Scan button continuously takes scans. After each integration cycle,
Scaling the Graph
You can change the vertical and/or horizontal scales of the graph by simply clicking on an X and Y
endpoint and manually typing in a value. The graph will then resize itself.
13
File Menu Functions
Save Spectral Values
File | Save Spectral Values
Select
acquisition parameters are included in the headers of these files. You can then use these files as overlays or
import them into other software programs, such as Microsoft Excel.
Save Kinetics Values
File | Save Kinetics Values
Select
parameters are included in the headers of these files. You can then import them into other software
programs, such as Microsoft Excel.
Open Spectrum Overlay
File | Open Spectrum Overlay
Select
previously saved spectrum and to open it as an overlay (a static spectrum) while still acquiring live data.
You can open up to 8 overl ays in the graph.
Open Kinetics Values
File | Open Kinetics Values
Select
previously saved kinetics chart.
from the menu to save the current spectrum. Text box entries and
from the menu to save kinetics data. T e xt bo x entries and acquisitio n
from the menu to open a dialog box that allows you to open a
from the menu to open a dialog box that allows you to open a
Printer Setup
File | Printer Setup
Select
kinetics data.
from the menu to select and configure a printer for printing graphical spectra or
Print Spectra and Kinetics
File | Print Spectra
Select
menu to print kinetics data.
from the menu to print a spectrum, or select
Exit
File | Exit
Select
want to exit the software.
from the menu to quit OOIChem. A message box appears asking you if you are sure you
File | Print Kinetics
from the
14
Edit Menu Functions
Clear Spectrum Ov erlays
Edit | Clear Spectrum Overlays
Select
Clear Kinetics Values
Edit | Clear Kinetics Values
Select
to clear the kinetics traces. A message box then appears, asking if you are sure you want to clear the
kinetics chart.
Autoscale X
Autoscale X
The
horizontal spectrum fills the display area.
function automatically adjusts the horizontal scale of a current graph so the entire
Autoscale Y
Autoscale Y
The
spectrum fills the display area.
function automatically adjusts the vertical scale of a current graph so the entire vertical
Show Kinetics Values
When setting up your kineti cs experiment, you
must first select
Configuration
chart display mode
Configuration
will be displayed above the kinetics chart if this function is enabled.
Spectrometer | Spectrometer
from the menu and make sure that
. Then configure your experiment by selecting
from the menu. When you select your wavelengths, the values from these wavelengths
from the menu to remove static spectra from the graph.
from the menu to clear both the kinetics values from the chart and
Spectrum & Kinetics
Spectrometer | Kinetics
is selected next to
Graph and
Show Legends
Edit | Show Legends
Select
overlays, and kinetics traces. When the legends are displayed, you can opt to configure
the traces by simply clicking on the legend trace you want to configure. You have the
opportunity to choose from several aesthetic functions such as: the plot design of the
spectrum, the point style used in the spectrum, the line style and width desired, color of
the plot, and a bar plot design. You can also choose to fill the baseline in the spectrum.
Utilize this function to differentiate one spectral trace from another.
to enable or disable the legends for the spectral trace,
15
Spectrometer Menu Functions
Scan
When the
selecting the Single mode, select
When the
continuously take scans.
Select Concentration Wavelength
This function is used when calculating the unknown concentration of a substance in a solution. You select
this function after you take an absorbance measurement of a standard solution with a known concentration.
Choose the wavelength of the highest peak in your absorbance spectrum. Then select
Calculate Calibration Curve
pages 31-32 for step-by-step instructions on calculating concentrations.
Calculate Calibration Curve
Concentration is the amount of a specified substance in a solution. In order to calculate concentration, you
must take absorbance measurements of a series of solutions with different known concentrations. The
length of the sample and the wavelength chosen for monitoring the amount of light absorbed are constants.
Then a linear plot from taking these scans is obtained. This Calibration Curve is used to determine the
unknown concentrations. See pages 31-32 for step-by-step instructions on calculating concentrations and on
using this dialog box.
Single
mode is selected in the display screen, the Scan menu function acts as a snapshot. After
Continuous
Spectrometer | Scan
mode is selected in the display screen, select
from the menu and complete the rest of your concentration experiment. See
from the menu to take one scan of the sample.
Spectrometer | Scan
from the menu to
Spectrometer |
Enable Strobe
This function allows you to enable or disable the triggering of external strobes through the spectrometer.
You would only select
strobe source. The CHEM2000 and CHEM2000-UV-VIS systems do not include a strobe light source.
However, the ISS-UV-VIS that comes with the CHEM2000-UV-VIS can be turned off and on through the
software and this function.
Spectrometer | Strobe Enable
from the menu if you were operating an external
Spectrometer Configuration
Intercept
From the Wavelength Calibration Report that
came with your system, enter the
under the
CHEM2000 and CHEM2000-UV-VIS come
with only a Master spectrometer channel,
additional spectrometer channels can be added
at any time. Contact an Applications Scientist
for more information.
Master
column. T hough the
First, Second and Third Coefficients
From the Wavelength Calibration Report that
came with your system, enter the
Coefficient, Second Coefficient
Coefficient
the CHEM2000 and CHEM2000-UV-VIS come
with only a Master spectrometer channel,
additional spectrometer channels can be added
at any time.
under the
Master
Intercept
First
Third
and
column. Though
16
Trigger Mode
With the CHEM2000 or CHEM2000-UV-VIS and OOIChem Operating Software, you have two
methods of acquiring data. In the
“free running.” That is, the spectrometer is continuously scanning, acquiring, and transferring data to
your computer, according to parameters set in the software. In this mode, however, there is no way to
synchronize the scanning, acquiring and transferring of data with an external event.
No External Trigger Mode
(or Normal Mode), the spectrometer is
To synchronize data acquisition with an external event, the
available. It involves connecting an external triggering device
External Software Trigger Mode
!!!!
to the spectrometer and then applying
is
an external trigger to the spectrometer before the software receives the data. In this mode, the
spectrometer is “free running,” just as it is in the Normal Mode. The spectrometer is continually
scanning and collecting data. With each trigger, the data collected during the integration period is
transferred to the software. All acquisition parameters, such as the integration period, are still set in the
software. You should use this mode if you are using a continuous light source and its inte nsity is
constant before, during and after the trigger.
!!!!
In order for you to use the External Software Triggering option, it is imperative that you know the
specifications and limitations of your triggering device. The design of your triggering device may
prevent you from using the external software triggering mode as it is described here.
To use the External Software Trigger Mode:
1. Supply a line from your triggering device to Pin 3 of the J2 Accessory Connector on the PC20000 to
provide the positive voltage +5VDC to the spectrometer. (See figure for pin location.) We do not
advise using an outside source to supply the voltage, as it is based on a referenc ed ground and your
reference may be different from ours. Using Pin 3 to supply voltage ensures that the spectrometer
will receive the appropriate voltage for the trigger event.
2. Supply a line from Pin 8 of the J2 Accessory Connector
to your triggering device. (See figure for pin location.)
3. Set your acquisition parameters in the software.
4. Select
5. Once you select
Spectrometer | Spectrometer Configuration
from the menu and choose
External Software Trigger
External Software Trigger
.
, it will
appear on your computer that your spectrometer is
unresponsive. Instead, i t is waiting for the trigger.
Activate your triggering device.
J2 (D-SUB-15) Accessory Connector (female)
Graph and chart display mode
If you choose
can be displayed in the graph area. If you choose
spectrometer channel and up to 8 overlays can be displayed in the in the top half of the graph area. A
kinetics chart monitoring values from up to 4 wavelengths and 2 of their arithmetic calculatio ns ca n be
displayed in the in the bottom half of the graph area, if the
Spectrum Only
, spectra from one real-time spectrometer channel and up to 8 overlays
Spectrum & Kinetics
Chart Active
, spectra from one real-time
function is enabled.
Flash Delay
The value entered here sets the delay, in milliseconds, between strobe signals sent out of the
spectrometer. This function allows you to enable or disable the triggering of external strobes
through the spectrometer. You would only enter a value in this box if you were operating an
external strobe source. The CHEM2000 and CHEM2000-UV-VIS systems do not include a strobe
light source.
Color Temperature
This box allows you to enter the color temperature (in Kelvin) of your light source. In order to calculate
relative irradiance, the reference spectrum must be made in Scope Mode with a blackbody light source
of known color temperature. This data is necessary for the software to complete calculations for relative
irradiance measurements. For CHEM2000 users, your lamp has a color temperature of 3100 Kelvin.
(CHEM2000-UV-VIS users cannot make relative irradiance measurements because the light source that
comes with the system is not a blackbody source with a known color temperature.)
17
Kinetics Configuration
Spectrometer | Kinetics Configuration
Select
a kinetics experiment. In the
of time, from up to 4 single wavelengths and up to two mathematical combinations of these wavelengths.
Kinetics Configuration
from the menu to configure and establish the parameters for
dialog box, you can collect spectral data as a function
Data from a kinetics experiment will not be displayed in the graph unless you c hoo se
Spectrometer Configuration
display mode
area, you will also still see real-time spectra in the top half of the graph area.
. This way, not only will your kinetics experiment be d isp layed in the bottom half of the graph
from the menu and select
Spectrum & Kinetics
Spectrometer |
next to
Graph and chart
Preset Duration
Enter a value to set the length of
time for the entire kinetics
process. Be sure to select hours,
minutes and se conds. Your
kinetics experiment can not
exceed a duration of 24 hours.
Preset Sampling Interval
Enter a value to set the frequency
of the data collected in a kinetics
process. Be sure to select hours,
minutes and seconds.
Wavelength
Enter the single wavelengths from which you wish to collect data. You can collect data from up to 4
a, b, c
single wavelengths, characterized as
, and d.
Display
If you want the data graphed from these single wavelengths, or from the mathematical calculations of
these wavelengths (described below), select the display box to the right of your values.
Mathematical Calculations
In the boxes next to
from the single wavelengths you specified as
which represents the calculation used in the box next to
x =
and
y =
, you have the opportunity to perform calculations on the data collected
Hardware Configuration
This command opens the Configure Hardware
dialog box. It allows you to configure the specific
type of hardware connecting the spectrometer to
the computer. For the CHEM2000 and
CHEM2000-UV-VIS systems, you only need to
select from
Base Address (I/O Range)
Request)
for other Ocean Optics systems. The parameters in
this dialog box are usually set only once -- when
OOIChem is first installed and the software first
starts. There is no reason to change these
parameters unless you need to alter the IRQ or
Base Address values. (See pages 6-8 for details
about changing the IRQ and Base Address values
and about this dialog box.)
Spectrometer, A/D Converter Type
IRQ (Interrupt
and
. The remaining hardware parameters are
,
a, b, c
, and d. Also in the box next to
x =
.
y =
, you can use x,
18
CHEM2000 Sampling Components
ISS-2 Integrated Sampling System
The CHEM2000 comes with a light source/sample holder combination called the ISS-2 Integrated Sampling
System. It is a fully integrated tungsten halogen light source and 1-cm square cuvette holder that couples to
the PC2000 with optical fiber to create a small-footprint system from ~350 to 900 nm. The ISS-2 has a 900hour bulb with a color temperature of 3100 Kelvin and has a 5-mm diameter f/2 collimating lens. In the ISS2, the fan is not enclosed in the base; it is exposed and should be handled with care.
Parts Included
• ISS-2 light source and cuvette holder assembly
• 12VDC wall transformer for providing power
• 1-cm square, plastic cuvette for holding samples
• Screwdriver for adjusting the fit of the cuvette
• Allen wrench for adjusting the collimating lens
Caution!
The light source becomes
"
DO NOT insert plastic or flammable materials in the filter slot. The materials could melt or ignite.
"
The unit could ignite flammable materials that come in contact with the metal housing.
"
There is an exposed fan on the bottom of the light source/sample holder. Handle with care.
"
HOT
during operation. Handle with care.
Operation
Adjusting the Fit of the Cuvette for the ISS-2
The ISS-2 is designed to hold 1-cm square cuvettes. When properly adjusted, the cuvette should fit snugly
into the holder.
1. Locate the two ball plunger screws.
2. Use the screwdriver to loosen the two ball plunger screws until the ball end of each screw is just
visible in the holder.
3. Insert your cuvette into the holder.
4. Gently tighten the screws until the ball contacts the cuvette and starts to compress.
over-tighten.
Turning On the ISS-2
1. Attach the 400-µm optical fiber that came with you CHEM2000 to the SMA connector on the front of
the sample holder and attach the other end of this fiber to the SMA connector on the PC2000 installed
in your computer .
2. Plug the wall transformer into a standard 110 V outlet. Plug the 12 V output into the back of your ISS-2.
For users of European-version wall transformers, plug the transformer into a standard 220 V outlet.
3. Find the on/off switch on the back of the ISS-2. Turn the lamp on.
4. Allow ~30 minutes to stabilize before using.
Do not
19
Using the Filter Slots in the ISS-2
The filter slots (one at the light source, a second in the sample holder) can be used to hold filters or light
blocks. The most useful filters include an FG-3 blue filter for increasing the relative energy near 400 nm
and 800 nm compared to 600 nm, an IR cutoff filter to reduce stray light below 750 nm, and a 550 nm
high-pass filter to eliminate second-order effects on Shortwave NIR measurements.
1. To install a filter at the light source, simply insert the filter into the filter slot. The slot accommodates
filters up to 3 mm thick and accepts color-correcting and longpass filters (round or square). There is no
clamp to hold the filter in place.
2. To install a filter in the sample holder filter slot, loosen the filter clamping screw with an Allen wrench
and insert the filter into the filter slot. The filter slot can accommodate filters up to 6 mm thick. Clamp
the filter in place by gently tightening with an Allen wrench.
Replacing the Bulb on the ISS-2
1. Order a replacement bulb.
2. Turn off the lamp and detach the power supply. Allow the lamp to cool.
3. Remove the fan from the bottom of the lamp by loosening the four screws securing it to the base. Two
of the four screws also hold the two front legs in place. Once the screws are loosened, gently pull the
fan, along with the legs, away from the lamp.
4. Now that the fan is removed, use an Allen wrench to loosen the set screw that is underneath the fan.
This screw holds the bulb in place. You do not need to remove the set screw -- loosening it is sufficient.
5. Locate the two set screws at the back of the lamp, one above each back leg. These two screws keep the
two halves of the lamp together. Remove the two screws.
6. Gently separate the two halves of the lamp and pull the bulb out of its housing.
7. Detach the wire and socket from the lamp leads. Remove bulb unit and discard.
8. Plug the new bulb into the socket and slide it forward into the front of the lamp as far as it will go.
9. Tighten the set screw on the bottom of the lamp to hold the bulb in place.
10. Close the two halves of the lamp, being ca reful not to pinch the wire s.
11. Replace the two screws at the back of the lamp and secure the fan to the bottom of the lamp.
Spectral Output of the ISS-2
This graph represents a blackbody curve for the 3100 K color-temperature, tungsten-halogen light source.
The actual spectral output of the lamp will vary due to the spectrometer configuration, grating efficiency,
the sampling optics used, and the detector’s efficiency.
1.2
1
0.8
0.6
0.4
Normalized Intensity
0.2
0
30050070090011001300150017001900
Wa vel ength (nm)
900-hour bulb / 3100K
20
ISS-2 Specifications
Path length:1 cm
Collimating lens:BK 7 glass (~360 nm - 2 µm*), 5 mm diameter, f/2
Collimating lens termination:SMA 905
Filter slot:accepts filters up to ¼" (6 mm) in thickness
Base material:aluminum
Spectral range:360 nm - 2 µm*
Dimensions:
Power input:12 VDC/800 mA, 7-20 VDC/0.5-2 amps
Power output:6.5 watts
Bulb life:900 hours
Bulb color temperature:3100K
Output to bulb:5 volts/1.3 amps
Output regulation:0.2% voltage
Time to stabilized output:
*The useable range of the ISS-2 is limited to the wavelength range of the spectrometer to which it is
coupled -- i.e., though the spectral range of the ISS-2 goes to 2 µm, it can only "see" to 900 nm when
used with the PC2000 that came with your CHEM2000.
**The amount of time it takes for the lamp to achieve a stabilized output will vary. A fan cools the lamp;
therefore the lamp may take much less than 30 minutes to stabilize.
9.0 cm x 5.0 cm x 3.2 cm (LWH)
3.5" x 2.0" x 1.25" (LWH)
~30 minutes
**
21
µµµµ
400-
The 400-µm diameter optical fiber that comes with your CHEM2000 is a single-strand optical fiber. The
active part of the fiber consists of a silica core, surrounded by a silica cladding material. The fiber is very
fragile, and if not protected by a suitable buffer material, would be nearly useless for most applications.
The buffer materials are polymer coatings that provide mechanical strength. Cabling further protects the
buffer-coated fiber. Our standard laboratory cabling is blue PVC. The ends of the fibers are cleaved,
epoxied into the SMA connectors, and polished. Follow the guidelines below when handling your fiber.
m Diameter Optical Fiber
Care and Use Guidelines
When fibers break, they stop transmitting light. Inspect fibers by eye to determine if light is being
!!!!
transmitted.
Bending the fiber will cause attenuation. To minimize this effect, add extra strain relief to both ends
!!!!
of the fiber.
Do not exceed the temperature specifications for the materials involved: 2000C for the fiber, 1000C
!!!!
for PVC cabling, 100
Do not allow the fiber to be bent at a sharp angle. A bending radius of less than 1" is dangerous.
!!!!
Keep connectors and probe tips covered when the fibers are not being used.
!!!!
Clean the ends of the fibers with lens paper and distilled water, alcohol or acetone. Avoid scratching
!!!!
the surface.
Do not immerse fiber ends in caustic materials or other solutions that can damage glass.
!!!!
0
C for standard epoxy.
Specifications
Jacket:
Sheathing/cabling:PVC (standard)
Connector/termination:SMA 905
Fiber core:pure fused silica
Cladding:doped fused silica
Fiber profile:step-index multi-mode
Numerical aperture:0.22
Recommended minimum bend radiu s:momentary = 10 cm, long term = 16 cm
Nylon (temperature range -40 to 100° C)
22
CHEM2000-UV-VIS Sampling Components
ISS-UV-VIS Integrated Sampling System
The ISS-UV-VIS Integrated Sampling System that comes with the CHEM2000-UV-VIS is a combination of
a RF deuterium source with a tungsten halogen bulb connected to a cuvette holder for 1-cm cuvettes. The
cuvette holder attaches directly to the light source and has a 5-mm diameter f/2 collimating lens to collect the
light and funnel it to the solarization-resistant optical fiber that also comes with the CHEM2000-UV-VIS.
The ISS-UV-VIS can be operated manually or through the software.
• Power cord for connecting the ISS-UV-VIS to outlet
• 15-pin accessory cable for software control of the ISS-UV-VIS
Caution!
The beam emerging from the ISS-UV-VIS produces low levels of ultraviolet radiation. Direct eye contact
"
could cause eye injury. Safety eyewear is recommended.
Never look directly into the light source, sight down the beam into the source or stare at the diffuse
"
reflected beam.
Dangerous voltages present. NO serviceable parts inside unit. To replace bulbs, contact Ocean Optics.
"
This instrument should not be used for any clinical or diagnostic purposes.
!!!!
Handle with care. Dropping the instrument may cause permanent damage.
!!!!
Operation
Adjusting the Fit of the Cuvette for the ISS-UV-VIS
The ISS-UV-VIS is designed to hold 1-cm square cuvettes. When properly adjusted, the cuvette should fit
snugly into the holder.
1. Locate the two ball plunger screws.
2. Use the screwdriver to loosen the two ball plunger screws until the ball end of each screw is just
visible in the holder.
3. Insert your cuvette into the holder.
4. Gently tighten the screws until the ball contacts the cuvette and starts to compress.
over-tighten.
Setting Up the ISS-UV-VIS
1. Attach the 300-µm solarization-resistant optical fiber that came with you CHEM2000-UV-VIS to the
SMA connector on the front of the sample holder and attach the other end of this fiber to the SMA
connector on the PC2000 installed in your computer.
Do not
23
2.Plug the wall transformer into a standard 110 V outlet. Plug the 12 V output into the back of your
12V
ISS-UV-VIS above the
transformer into a standard 220 V outlet. At this time, the green LED indicator light on the front of the
lamp will be lit. This indicator light only means that the lamp is receiving power, not that the
deuterium and tungsten bulbs are on.
label. For users of European-version wall transformers, plug the
Operating the ISS-UV-VIS Manually
1. Find the switch on the back of the ISS-UV-VIS.
2. There are three positions:
position. There can be up to a 1.5 second delay between switching the lamp to on and the bulbs igniting.
If the lamp has not been used recently, the deuterium bulb may take up to 60 seconds to ignite.
3. For 0.3% peak-to-peak stability, allow 30 minutes warm-up time before taking your measurements.
On, Off
, and
Remote
. For Manual operation, move the switch to the
On
Operating the ISS-UV-VIS through Software
1. Take the 15-pin accessory cable and plug one end into the PC2000 installed into your computer.
Connect the other end of the accessory cable into the back of the ISS-UV-VIS.
2. Find the switch on the back of the ISS-UV-VIS. There are three positions:
For Software operation, move the switch to the
position enables you to control the lamp through the software, whether you are using OOIChem
or OOIBase32.
3. When using OOIChem, sele ct
in the ISS-UV-VIS on and off.
4. When you want to control the ISS-UV-VIS through our more advanced OOIBase32 Spectrometer
Operating software, select or deselect the
bar above the graph area to turn the light source in the ISS-UV-VIS on and off. Refer to the
OOIBase32 Spectrometer Operating Software Manual
4. There can be up to a 1.5 second delay between turning the bulbs on via the software and the bulbs
igniting. If the lamp has not been used recently, the deuterium bulb may take up to 60 seconds to ignite.
5. For 0.3% peak-to-peak stability, allow 30 minutes warm-up time before taking your measurements.
Spectrometer | Enable Strobe
Remote
Strobe Enable
position. Moving the switch to the
from the menu to turn t he light sourc e
box in the Acquisition Parameter dialog
for more details.
On, Off
, and
Remote
Remote
.
Disabling the Tungsten or Deuterium Bulb
It is possible to disable the deuterium or the tungsten bulb in the ISS-UV-VIS. Both bulbs are enabled at
the time of manufacture. In order to disable the deuterium or tungsten bulb, you must remove the casing of
the ISS-UV-VIS.
Disabling the Deuterium Bulb
Jumper block JA of the ISS-UV-VIS's circuit board controls the deuterium bulb. Short pins 2-3 (that is,
place a jumper over pins 2-3 of JA) to enable the deuterium bulb. Short the pins 1-2 (that is, place a
jumper over pins 1-2 of JA) to disable the deuterium bulb.
Disabling the Tungsten Bulb
Jumper block JB of the ISS-UV-VIS's circuit board controls the tungsten bulb. Short pins 2-3 (that is, place
a jumper over pins 2-3 of JB) to enable the tungsten bulb. Short pins 1-2 (that is, place a jumper over pins
1-2 of JB) to disable the tungsten bulb.
Bulb Replacement
The deuterium and tungsten bulbs cannot be replaced by unauthorized personal. To replace a bulb in the
ISS-UV-VIS, contact Ocean Optics.
Power input:12 V
Power requirement:12 VDC/420 mA
Power consumption:5 watts (deuterium 3.8 watts, tungsten 1.2 watts)
0.3% in 30 minutes
1.0% in 10 minutes
25
µµµµ
300-
The 300-µm Diameter Solarization-resistant Optical Fiber that comes with your CHEM2000-UV-VIS is a
single-strand optical fiber. The active part of our solarization-resistant fibers consists of a silica core,
surrounded by a silica cladding material. The fiber is then coated in aluminum. Our standard laboratory
cabling is blue PVC. Our solarization-resistant fibers are best used for applications where exposure to longterm UV light occurs.
Solarization is the loss of transparency in glass due to exposure to ultraviolet radiation. If you are using a
UV light source, the UV radiation degrades the silica in a standard patch cord fiber over time, resulting in
increased overall absorption values and invalid data. This degradation is called so larization, which is the
formation of “color centers” or silica molecules with an unbound pair of electrons. These color centers have
an absorption band that is centered at 215 nm. In regular HOH fibers (our standard fibers), the
concentration of these color centers grows until nearly all UV light in the 215 nm region is blocked.
Other solarization-resistant fibers are hydrogen-loaded but their solarization-resistant properties do not
last as the hydrogen eventually leaks out of the silica -- meaning that these hydrogen-doped solarizationresistant fibers have a “shelf life.” With Ocean Optics solarization-resistant op tical fibers, there is no
shelf life.
m Diameter Solarization-resistant Optical Fiber
Care and Use Guidelines
When fibers break, they stop transmitting light. Inspect fibers by eye to determine if light is being
!!!!
transmitted.
Bending the fiber will cause attenuation. To minimize this effect, add extra strain relief to both ends
!!!!
of the fiber.
Do not exceed the temperature specifications for the materials involved: 2000C for the fiber, 1000C
!!!!
for PVC cabling, 100
Do not allow the fiber to be bent at a sharp angle. A bending radius of less than 1" is dangerous.
!!!!
Keep connectors and probe tips covered when the fibers are not being used.
!!!!
Clean the ends of the fibers with lens paper and distilled water, alcohol or acetone. Avoid scratching
!!!!
the surface.
Do not immerse fiber ends in caustic materials or other solutions that can damage glass.
!!!!
0
C for standard epoxy.
Jacket:
Sheathing/cabling:PVC (standard)
Connector/termination:SMA 905
Fiber core:pure fused silica
Cladding:doped fused silica
Fiber profile:step-index multi-mode
Numerical aperture:0.22
Recommended minimum bend radius:momentary = 8 cm, long term = 16 cm
Specifications
Nylon (temperature range -40 to 100° C)
26
Experiment Tutorial
When you are ready to begin your experiment, you should have already installed the PC2000, installed
OOIChem, set up your light source/sample holder, and connected your fiber from the PC2000 to your light
source/sample holder.
Now you are ready to take your measurements. Because of the components making up your CHEM2000 and
CHEM2000-UV-VIS, your system is ideal for absorbance and transmission. The CHEM2000 can also make
relative irradiance measurements. If, however, you wish to utilize your system for other measuring funct ions,
additional products might be required. Contact an Ocean Optics Application s Scientist for options.
Absorbance Experiments
Absorbance spectra are a measure of how much light is absorbed by a sample. The software calculates
absorbance (
A
λ
) using the following equation:
S
- D
λ
Aλ = - log
where S is the sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the
reference intensity at wavelength λ.
10
(
Rλ - D
λ
)
λ
Common applications include the quantification of chemical concentrations in aqueous or gaseous samples.
To take an absorbance measurement:
1. Select
2. While still in
3. Begin an absorbance measurement by first making sure the sample is in place and nothing is blocking
Scope
Adjust acquisition parameters so that the peak intensity of the reference signal is about 3500 counts. Take
a reference spectrum by first makin g sure nothing is blocking the light path going to your spectr ometer.
The analyte you want to measure must be absent while taking a reference spectrum. Take the reference
reading b y clicking the
reference spectrum. To save a spectrum, you must select
Storing a reference spectrum is requisite before the software can calculate absorbance spectra.
your spectrometer. (If possible, do not turn off the light source . If you must turn off your l ight source
to store a dark spectrum, make sure to allow enough time for the lamp to warm up before continuing
your experiment.) Ta ke the dark rea ding by clicking the
(This command merely stores a dark spectrum. To save a spectrum, you must select
Spectral Values
absorbance spectra.
the light going to your sample. Then select
display area. Click on the
scan will be taken. If
click on the
!!!!
If at any time any sampling variable changes -- including integration period, averaging, boxcar
smoothing, distance from light source to sample, etc. -- you must store a new reference and
dark spectrum.
Stop
Mode of Operation
under
Reference
Scope Mode
from the menu.) Storing a dark spectrum is requisite before the software can calculate
button. To save the spectrum, select
, take a dark spectrum by first completely blocking the light path going to
Scan
button in the display area to take a scan. If
Continuous
in the software display area. Make sure the signal is on scale.
button in the software display area. (This command merely stores a
File | Save Spectral Values
Dark
button in the software display area.
Absorbance
is selected, the spectrometer will continuously take scans until you
File | Save Spectral Values
Mode of Operation
under
Single
from the menu.)
File | Save
in the software
is selected, only one
from the menu.
27
Transmission Experiments
Transmission is the percentage of energy passing through a system relative to the amount that passes
through the reference. Transmission Mode is also used to show the portion of light reflected from a sample.
Transmission and reflection measurements require the same mathematical calculation. The transmission is
λ
λ
- D
- D
%T
λ
) relative to a standard substance (such as air). The software calculates
λ
x 100%
λ
expressed as a percentage (
%R
λ
(or
) by the following equation:
S
%Tλ =
R
where S is the sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the
reference intensity at wavelength λ.
Common applications include measurement of transmission of light through solutions, optical filters,
optical coatings, and other optical elements such as lenses and fibers.
!!!!
For transmission of light through solutions, we offer a transmission dip probe with screw-on,
removable tips in 2-mm, 5-mm or 10-mm path lengths. Contact Ocean Optics for more information.
To take a transmission measurement:
1. Select
Scope
Mode of Operation
under
in the software display area. Make sure the signal is on scale.
Adjust acquisition parameters so that the peak intensity of the reference signal is about 3500 counts.
Take a reference spectrum by first making sure no thing is blocking the light path going to your
spectrometer. The analyte you want to measure must be absent while taking a reference spectrum. Take
the reference reading by clicking the
merely stores a reference spectrum. To save a spectrum, you must select
Reference
button in the software display area. (This command
File | Save Spectral Values
from the menu.) Storing a reference spectrum is requisite before the software can calculate
transmission spectra.
2. While still in
Scope Mode
, take a dark spectrum by first completely blocking the light path going to your
spectromete r. (If possib l e, do not turn off the light source. If you must turn off your light source to store a
dark spectrum, make sure to allow enough time for the lamp to warm up before continuing your
experiment.) Take the dark reading by clicking the
command merely stores a dark spectrum. To save a spectrum, you must select
Values
from the menu.) Storing a dark spectrum is requisite before the software can calculate
Dark
button in the software display area. (This
File | Save Spectral
transmission spectra.
3. Begin a transmission measurement by first making sure the sample is in place and nothing is blocking
the light going to your sample. Then select
display area. Click on the
scan will be taken. If
click on the
Stop
button. To save the spectrum, select
Scan
button in the display area to take a scan. If
Continuous
is selected, the spectrometer will continuously take scans until you
Transmission
File | Save Spectral Values
Mode of Operation
under
in the software
Single
is selected, only one
from the menu.
%T
λ
!!!!
If at any time any sampling variable changes -- including integration period, averaging, boxcar
smoothing, distance from light source to sample, etc. -- you must store a new reference and
dark spectrum.
28
Reflection Experiments
Reflection is the return of radiation by a surface, without a change in wavelength. The reflection may be:
• Specular, in which the angle of incidence is equal to the angle of reflection.
• Diffuse, in which the angle of incidence is not equal to the angle of reflection.
Every surface returns both specular and diffuse reflections. Some surfaces may return mostly specular
reflection, others more diffuse reflection. The glossier the surface, the more specular the reflection.
Reflection is expressed as a percentage (
our WS-1 white reference for a diffuse reflection measurement):
S
- D
λ
λ
%Rλ =
R
λ
where S is the sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the
reference intensity at wavelength
- D
x 100%
λ
Common applications include measuring the reflection properties of mirrors, anti-reflection coatings, and
measuring the visual properties of the color in paints, graphic arts, plastics, and food products.
!!!!
The components that came with the CHEM2000 and CHEM2000-UV-VIS will not allow the user to
make reflection measurements. A reflection probe is necessary. We offer several components for
reflection measurements such as a variety of reflection probes, a reflection probe holder, diffuse
reflectance standards and an integrating sphere. Contact Ocean Optics for more information.
%R
λ
) relative to the reflection from a standard substance (such as
λ.
To take a reflection measurement:
1. Select
Scope
Mode of Operation
under
in the software display area. Make sure the signal is on scale.
Adjust acquisition parameters so that the peak intensity of the reference signal is about 3500 counts. Take
a reference spectrum by first makin g sure nothing is blocking the light path going to your spectr ometer.
The analyte you want to measure must be absent while taking a reference spectrum. Take the reference
reading b y clicking the
reference spectrum. To save a spectrum, you must select
Reference
button in the software display area. (This command merely stores a
File | Save Spectral Values
from the menu.)
Storing a reference spectrum is requisite before the software can calculate reflection spectra.
2. While still in
Scope Mode
, take a dark spectrum by first completely blocking the light path going to
your spectrometer. (If possible, do not turn off the light source . If you must turn off your l ight source
to store a dark spectrum, make sure to allow enough time for the lamp to warm up before continuing
your experiment.) Ta ke the dark rea ding by clicking the
(This command merely stores a dark spectrum. To save a spectrum, you must select
Spectral Values
from the menu.) Storing a dark spectrum is requisite before the software can calculate
Dark
button in the software display area.
File | Save
reflection spectra.
3. Begin a reflection measurement by first making sure the sample is in place and nothing is blocking the
light going to your sample. Then select
Transmission
Mode of Operation
under
in the software
display area. The mathematics required to calculate reflection measurements are identical to those
necessary to compute a transmission spectrum. Click on the
scan. If
continuously take scans until you click on the
Spectral Values
!!!!
Single
is selected, only one scan will be taken. If
Stop
button. To save the spectrum, select
from the menu.
If at any time any sampling variable changes -- including integration period, averaging, boxcar
smoothing, distance from light source to sample, etc. -- you must store a new reference and
dark spectrum.
Scan
button in the display area to take a
Continuous
is selected, the spectrometer will
File | Save
29
Relative Irradiance Experiments
Irradiance is the amount of energy at each wavelength from a radiant sample. In relative terms, it is the
fraction of energy from the sample compared to the energy collected from a lamp with a blackbody
energy distribution, normalized to 1 at the energy maximum. Relative irradiance is calculated by the
following equation:
S
- D
λ
Iλ = B
λ
(
where Bλ is relative energy of the reference calculated from the color temperature in Kelvin, S is the
sample intensity at wavelength λ, D is the dark intensity at wavelength λ, R is the reference intensity
at wavelength λ.
Common applications include characterizing the light output of LEDs, incandescent lamps and other
radiant energy sources such as sunlight. Also included in irradiance measurements is fluorescence, in which
case the spectrometer measures the energy given off by materials that have been excited by light at a
shorter wavelength.
!!!!
The components that came with the CHEM2000-UV-VIS will not allow the user to make relative
irradiance measurements. Only CHEM2000 users can perform relative irradiance measurements.
In order to make relative irradiance measurement, the reference spectrum must be made in Scope
Mode with a blackbody light source of known color temperature. This color temperature is needed
in order to calculate relative irradiance. The light source that comes with the CHEM2000-UV-VIS is
not a blackbody light source with a known color. To purchase a blackbody light source, contact
Ocean Optics.
Rλ - D
λ
)
λ
To take a relative irradiance measurement:
1. Select
2. Select
3. While still in
4. Begin a relative irradiance measurement by first positioning the fiber at the light or emission source
Spectrometer | Spectrometer Configuration
the color temperature in Kelvin of the reference lamp you are going to use is entered here. The color
temperature of the tungsten halogen light source that came with your CHEM2000 has a color
temperature of 3100 Kelvin. Click OK.
Scope
by adjusting acquisition parameters. Take a reference spectrum of your reference lamp. Take the
reference reading by clicking the
merely stores a reference spectrum. To save a spectrum, you must select
from the menu.) Storing a reference spectrum is requisite before the software can calculate relative
irradiance spectra.
spectrometer. Take the dark reading by clicking the
command merely stores a dark spectrum. To save a spectrum, you must select
Values
from the menu.) Storing a dark spectrum is requisite before the software can calculate relative
irradiance spectra.
you wish to measure. Then select
display area. Click on the
scan will be taken. If
click on the
!!!!
If at any time any sampling variable changes -- including integration period, averaging, boxcar
smoothing, distance from light source to sample, etc. -- you must store a new reference and
dark spectrum.
Stop
Mode of Operation
under
Scope Mode
button. To save the spectrum, select
, take a dark spectrum by first completely blocking light from going to your
Scan
button in the display area to take a scan. If
Continuous
in the software display area. Make sure the signal is on scale
Reference
Relative Irradiance
is selected, the spectrometer will continuously take scans until you
button in the software display area. (This command
from the menu. Next to
Dark
button in the software display area. (This
Mode of Operation
under
File | Save Spectral Values
Color Temp
File | Save Spectral Values
File | Save Spectral
in the software
Single
is selected, only one
from the menu.
, make sure
30
Concentration Experiments
The absorbance of a solution is related to the concentration of the species within it. The relationship, known
as Beer’s Law, is:
ε
Aλ =
where A is the absorbance at wavelength λ, ε is the extinction coefficient of the absorbing species
at wavelength λ, c is the concentration and l is the optical pathlength.
Concentration is the amount of a specified substance in a solution. Graphs of absorbance vs. concentration
are known as Beer’s Law plots. These are prepared by measuring the light absorbed by a series of solutions
with different known concentrations. The length of the sample -- such as the path length of our cuvette
holder -- and the wavelength chosen for monitoring the amount of light absorbed are constants. A linear
plot from taking scans of these standard solutions with known concentrations is then obtained. The plot is
then used to determine the unknown concentrations of substances in solutions.
In order to discover the unknown concentration of a substance in a solution, you must first take spectral
scans of a series of solutions with different known concentrations of the same substance. You begin this
process by taking an absorbance spectrum of the solution with the highest known concentration.
1. Select
2. While still in
3. Take the solution with the highest known concentration and put it in the cuvette holder. Make sure
4. Now select the wavelength for monitoring the concentration of your solutions by choosing
Scope
Adjust acquisition parameters so that the peak intensity of the reference signal is about 3500 counts.
Take a reference spectrum by first making sure no thing is blocking the light path going to your
spectrometer. The solution with the highest known concentration you want to measure must be absent
while taking a reference spectrum. Take the reference reading by clicking the
software display area. To save the spectrum, select
your spectrometer. (If possible, do not turn off the light source . If you must turn off your l ight source
to store a dark spectrum, make sure to allow enough time for the lamp to warm up before continuing
your experiment.) Ta ke the dark rea ding by clicking the
save the spectrum, select
nothing is blocking the light going to your sample. Then select
Operation
To save the spectrum, select
| Select Concentration Wavelength
the spectrum of the solution with the highest known concentration and choose
c
l
λ
Mode of Operation
under
Scope Mode
. Click on the
in the software display area. Make sure the signal is on scale.
Reference
File | Save Spectral Values
, take a dark spectrum by first completely blocking the light path going to
Dark
button in the software display area. To
File | Save Spectral Values
Scan
button in the display area to take a scan. Make sure
File | Save Spectral Values
from the menu. Move the cursor to the highest absorbance peak of
from the menu.
Absorbance
from the menu.
from the menu.
under
Select
button in the
Mode of
Single
.
is selected.
Spectrometer
5. Remove the solution with the highest known concentration. Select
Calibration Curve
begin taking scans of the r est of your series of standard solutions with known concentrations, from the
lowest known concentration to highest, all while working in this dialog box.
6. If you wish, you can take a new reference and a new dark scan for each solution by choosing
Dark
necessary. If no reference or dark scan is taken at this point, the software will use the reference and
dark scans taken in Steps 1 and 2 to calculate absorbance.
7. Take the solution with the lowest known concentration and put it in the cuvette holder. Enter the known
concentration of the standard solution in the chart in the
Scan | Reference
and
from the menu. The Calculate Calibration Curve dialog box opens. Now you will
Spectrometer | Calculate a
from the menu of this dialog box. However, in this case, it is not
31
Concentration
column, next to
Solution #1
Scan |
.
8. Click the
Scan
button or select
Scan | Solution
from the menu. The absorbance value will appear next
to the concentration for Solution #1. At any point, you can select
the dialog box of all data.
Edit | Clear
from the menu to clear
9. Take Solution #1 out of the cuvette holder and put in another standard solution with the next highest
known concentration. Enter the known concentration of the standard solution in the chart in the
Concentration
10. Click the
column, next to
Scan
button or select
Solution #2
.
Scan | Solution
from the menu. The absorbance value will appear next
to the concentration for Solution #2.
11. You may continue to scan solutions with known concentrations. You must scan at least 2 in or d e r to
achieve a calibration curve.
12. When you have completed taking scans of your solutions with known concentrations, click the
Calibration Curve
ε
necessary to compute Beer’s Law and to find the unknown concentration of a solution.
button. You will then ha ve the
13. At this time, you may also select a label for your concentration values, such at
Concentration Units
14. Select
Edit | Show Legend
box. This is only a label and does not affect the data in any way.
from the menu to display the legend for the calibration curve. The legend
Intercept
and
Slope
of your curve. The Slope is the
Moles per Liter
allows you to choose a plot design, point style, line style, line width and plot color. Select
Palette
from the menu to display a variety of options for configuring the curve. The palette provides
View
, in the
Edit | Show
features such as autoscaling, graph formatting, value precision, mode mapping, and graph positioning.
15. You can select
current calibration curve data, select
16. Select
File | Close
File | Print
from the menu for the dialog box to print the dialog box. To save the
File | Save
from the menu for the dialog box.
from the menu for the dialog box to close this dialog box and return to the
main windo w.
17. When a message box asks if you would like to use this calibration curve when calculating
concentration values, select
Yes
.
18. Now that you are back in the main display window, place the solution with the unknown concentration
of a substance in the cuvette holder.
19. Under
Mode of Operation
, select
Concentration
. Click the
Scan
button to receive your
concentration values.
32
Kinetics Experiments
Spectrometer | Kinetics Configuration
Select
a kinetics experiment. In the
of time, from up to 4 single wavelengths and up to two mathematical combinations of these wavelengths.
A kinetics experiment will not be
displayed in the graph unle ss you
Spectrometer |
choose
Spectrometer Configuration
from the menu and choose
Spectrum & Kinetics
Graph and chart display mode
This way, not only will your
kinetics exp eriment be displayed
in the bottom half of the graph
area, you will also still see a
spectrometer channel’s real-time
spectra in the top half of the
graph area.
To run a kinet i cs experiment:
1. In the graph area, select acquisition parameters, such as integration period, averaging, and boxcar
smoothing values. Do not change these parameters for the duration of the kinetics experiment.
2. Select
3. Enter a
4. Enter a
5. Under
6. If you want the data graphed from these single wavelengths, or from the mathematical calculations
7. In the boxes next to
8. Click OK to confirm the parameters and close the dialog box.
9. Click the
10. Click the
11. Select
Spectrometer | Kinetics Configuration
dialog box.
Preset Duration
hours, minutes and seconds. The duration of your kinetics experiment cannot exceed 24 hours.
Preset Sampling Interval
Be sure to select hours, minutes and seconds. Data from a timed acquisitio n is stamped with a time that
is accurate to 1 millisecond.
Wavelength
data from up to 4 single wavelengths, characterized as a, b, c, and d.
(described in the next step), select the
from the single wavelengths you specified as
represents the calculation used in
Scan
button to begin the kinetics experiment. Make sure that
top half of the graph displays a real-time full wavelength spectrum. The bottom half of the graph
displays the data for the selected wavelengths and their derived arithmetic calculations. Each data set is
stored with a time stamp.
Stop
button to stop the experiment. However, if a
experiment will automatically stop after the designated time has passed.
File | Save Kinetics Values
spectrometer’s serial number, active channel and acquisition parameters in a header. This file can be
opened with any text or spreadsheet editor.
Kinetics Configuration
next to
.
value to set the length of time for the entire kinetics process. Be sure to select
value to set the frequency of the data collected in a kinetics process.
, enter the single wavelengths from which you wish to collect data. You can collect
x =
and
y =
, you have the opportunity to perform calculations on the data collected
x =
from the menu to save a tab-delimited ASCII file with the
from the menu to configure and establish the parameters for
dialog box, you can collect spectral data as a function
from the menu to open the Kinetics Configuration
Display
.
box to the right of your values.
a, b, c
, and d. In the
y =
box, you can also use x, which
Continuous
Preset Duration
is selected. The
time was selected, the
33
Appendix A: Changing the Settings on the
PC2000
Base Address Settings
To change the Base Address settings on the PC2000, find the bank of switches on the A/D converter. The
Base Address may be change d via the first 6 switches (the IRQ may be changed via the last 3 switches).
Switches in the OFF position have the decimal values shown. Switches in the ON position have a value of
zero. The Base Address is the sum of the values of the switches. In the default setting, switches 5 and 6 are
added to give a value of 768. A few of the many combinations for Base Address settings are below. After
you have changed the switches, reinstall the card and change the software settings to match the hardware
settings. (See pages 6-8 for instructions.) The gray block indicates the position of the switch.
Decimal equivalent163264128256512
Value as shown256512
Example: 784 decimal = Hex310 = 0x310
Switch #123456
ON
OFF
Value as shown16256512
Example: 800 decimal = Hex320 = 0x320
Switch #123456
ON
OFF
Value as shown32256512
Example: 816 decimal = Hex330 = 0x330
Switch #123456
ON
OFF
Value as shown1632256512
34
Interrupt Request Settings
To change the IRQ settings on the PC2000, see the bank of switches on the A/D converter. The IRQ may
be changed via the last 3 switches. The following matrix defines the different IRQ settings by sw itch positions
7, 8, and 9. In the default setting, the IRQ is set to 7. Other combinations for IRQ settings are belo w. After
you have changed the switches, reinstall the card and change the software settings to match the hardware
settings. (See pages 6-8 for instructions.) The gray block indicates the position of the switch.
= Switch is in the on, upward position
IRQ 7 (Default Setting)
Switch # 789
ON
OFF
IRQ 3
Switch # 789
ON
OFF
IRQ 4
Switch # 789
ON
OFF
IRQ 5
Switch # 789
ON
OFF
IRQ 9
Switch # 789
ON
OFF
IRQ 10
Switch # 789
ON
OFF
IRQ 11
Switch # 789
ON
OFF
35
Appendix B: PC2000 Pin-outs and Jumpers
It is not necessary for the average user to worry about the interconnect scheme, as the cables supplied with all
of the units need only be plugged into the matching connectors on the hardware. However, if the need arises
to design and fabricate your own cabling system, the following tables supply the necessary information.
Appendix C: Calibrating the Wav elength of
Your Spectrometer
The following describes how to calibrate the wavelength of your spectrometer. Though each spectrometer
is calibrated before it leaves Ocean Optics, the wavelength for all spectrometers will drift slightly as a
function of time and environmental conditions.
You are going to be solving the following equation, which shows that the relationship between pixel
number and wavelength is a third-order polynomial . . .
λ
= I + C
p
. . . where λ is the wavelength of pixel p, I is the wavelength of pixel 0, C1 is the first coefficient
(nm/pixel), C
calculating the values for I and the three Cs.
is the second coefficient (nm/pixel2), and C3 is the third coefficient (nm/pixel3). You will be
2
p + C
1
p2 + C
2
p
3
Setting Up
3
To re-calibrate the wavelength of your spectrometer, you will need the following:
•A light source that produces spectral lines. Ocean Optics HG-1 Mercury-Argon lamp is ideal for this
purpose. If you do not have an HG-1, you will need a spectral line source that produces several (at
least 4-6) spectral lines in the wavelength region of your spectrometer.
• Your spectrometer.
• An optical fiber (the fibers that came with your CHEM2000 and CHEM2000-UV-VIS will work fine).
• Either a spreadsheet program (Excel or Quattro Pro, for example) or a calculator that performs third-
order linear regressions. If you are using Microsoft Excel, choose
AnalysisToolPak
check
AnalysisTookPak-VBA
and
.
Tools | Add-Ins
from the menu and
Calibrating the Wavelength
of Your Spectrometer
1. After placing OOIChem into Scope Mode, take a spectrum of your light source that produces spectral
lines. Adjust the integration period until there are several peaks on the screen that are not off-scale.
2. Move the cursor to one of the peaks and carefully position it so that it is at the point of maximum
intensity. Record the pixel number that is displayed in the cursor function bar (located above right of
the graph). Repeat this step for all of the peaks in your spectrum.
3. Using your spreadsheet, create a table like the one shown on the next page. In the first column, place
the exact or true wavelength of the spectral lines that you used. Most calibration line sources come
with a wavelength calib ration sheet . If you do not have a wavelength calibration sheet for your light
source, you can probably find the wavelengths for your spectral lines (if they are being produced by
pure elements) in the first column of the worksheet. In the second column of this worksheet, place the
observed pixel number. In the third column, calculate the pixel number squared, and in the fourth
column, calculate the pixel number cubed.
4. Now you are ready to calculate the wavelength calibration coefficients. In your spreadsheet program,
find the functions to perform linear regressions.
if using Quattro Pro, look under
if using Excel, look under
Tools | Advanced Math
Analysis ToolPak
5. Select the true wavelength as the dependent variable (Y). Select the pixel number, pixel number
squared and the pixel number cubed as the independent variables (X). After you execute the
regression, an output similar to the one shown below is obtained.
Regression Statistics
Multiple R0.999999831
R Square0.999999663
Adjusted R Square0.999999607
Standard Error0.125540214
Observations22
intercept
CoefficientsStandard Error
Intercept190.4739930.369047536
first coefficient
X Variable 10.362639830.001684745
X Variable 2-1.174416E-058.35279E-07
X Variable 3-2.523787E-092.656608E-10
third coefficient
second coefficient
6. The numbers of importance are indicated in the above figure. You will need to record the Intercept as
well as the First, Second, and Third Coefficients. Also, look at the value for R squared. It should be
very close to 1. If it is not, you have probably assigned one of your wavelengths incorrectly.
7. Select
Spectrometer | Spectrometer Configuration
38
from the menu. Update the wavelength coefficients.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.