Observator ANALITE NEP5000 Quick Start Manual

2091 Exchange Court Fairborn, Ohio 45324 937-426-2703 www.NexSens.com
Observator NEP5000 Turbidity Sensor
Subconn Wire Color
Wire Outputs
Calibration
Module Connection
Black
Power GND
Slot 2
White
SDI-12 Data
N/A
Red
10 to 30V DC power
Slot 1
Green
4-20mA loop to power
GND
N/A
Orange
Voltage Out
N/A
Blue
Calibration
Communication Data
Slot 6
Quick Start Guide
The Observator ANALITE NEP5000 Turbidity Sensor is compatible with the Nexsens X2 data logger using SDI-12 output, and is designed for environmental monitoring and process applications.
Calibrating the NEP5000
For configuration and calibration, the Observator ANALITE NEP5000 Configuration Kit accessory must be used. This kit comes with a dedicated Subconn connection cable.
The Subconn cable to calibration module connection is detailed in the table below.
2091 Exchange Court Fairborn, Ohio 45324 937-426-2703 www.NexSens.com
Observator NEP5000 Turbidity Sensor
Observator
Sensor Pin
Number
MCIL-6-FS
Socket
Number
Observator
Pin Signal
MCIL-6-FS*
Wire Color
J Pin
1
1
Power GND
Orange
J2 2 2
SDI-12 Data
Black
J6
3
3
10 to 30 VDC
power
Red
J5
UW-FWP
MCIL-6-FS
Quick Start Guide
Connect to the NEP5000 Turbidity Sensor and use the OEM Calibration software to enable the “Wipe on PowerUp” option.
This ensures that the optics on the sensor are cleared before each new reading is requested from the X2, which allows for the most accurate measurements.
Wiring
For compatibility, sensors must be connected to a female 6-pin wet mateable cable (MCIL-6-FS).
The flying lead for the MCIL-6-FS must be connected to a UW-FWP with the pin configuration in the table below.
*If the MCIL wire color is not listed, it is unused in the connection.
For additional information, please read the Observator sensor manual located on the Fondriest Environmental website.
https://www.fondriest.com/pdf/observator_nep5000_manual.pdf