Oakworks Medical PF, PF400, PF200, PF250 User Manual

USER MANUAL
OAKWORKS® PF Series Tables
www.oakworksmed.com · 717.235.6807
made in the USA
with US & imported parts
© Copyright 2018 OAKWORKS® , Inc.
Printed in U.S.A.
All rights are reserved. No part of this document may be photocopied, reproduced or translated to another language without prior written consent of OAKWORKS®, Inc.
OAKWORKS® is a registered trademark of OAKWORKS®, Inc.
Notice
The information contained within this document is subject to change without notice and should not be construed as a commitment by OAKWORKS®, Inc.
OAKWORKS®, Inc. encourages requests for technical specifications and the like documentation to ensure accu­racy. The appropriate documentation is available upon request.
OAKWORKS®, Inc. shall not be liable for incidental or consequential damages in connection with or arising out of the furnishing, performance, or use of this document and the program material which it describes.
TABLE OF CONTENTS
Product Use Description ..............................................................................................................................1
Important Safety Instructions
Symbol Identification .........................................................................................................................1
Safety Instructions ........................................................................................................................... 1-2
Product Description & Photo
PF400 ......................................................................................................................................................3
PF250 ......................................................................................................................................................4
PF200 ......................................................................................................................................................5
Installation
Grounding .............................................................................................................................................6
Directions for Use
Foot Control Operations
PF400 Tables Only ..................................................................................................................... 7
PF250 & PF200 Tables Only ....................................................................................................8
Optional Hand Control Operations
PF400 Tables Only ..................................................................................................................... 9
PF250 & PF200 Tables Only ..................................................................................................10
Moving the Table ...............................................................................................................................11
Leveling the Table .............................................................................................................................11
Lift-Assist Backrest Adjustment .....................................................................................................12
Lowered Arm Rest Use .....................................................................................................................12
Locking Casters Use ..........................................................................................................................13
Table Extender ...................................................................................................................................13
Adjustable Arm Rest Installation and Adjustment ...................................................................14
QuickLock Face Rest .........................................................................................................................15
Cleaning & Disinfection
Recommended Cleaners/Disinfectants .......................................................................................16
Cleaning Process ................................................................................................................................17
Inspections & Maintenance .......................................................................................................................17
Warranty Information ..................................................................................................................................17
Model Number & Serial Number ..............................................................................................................18
Specifications
Product Specifications .....................................................................................................................19
Environmental Conditions ..............................................................................................................19
Electrical Specifications ...................................................................................................................19
Guidance and manufacturer’s declaration -
Electromagnetic emissions .............................................................................................................20
Recommended separation distances ...........................................................................................20
Guidance and manufacturer’s declaration -
Electromagnetic immunity....................................................................................................... 21-22
Contact Information ....................................................................................................................back cover
PRODUCT USE DESCRIPTION / IMPORTANT SAFETY INSTRUCTIONS
PRODUCT USE DESCRIPTION
The PF Exam Table is a powered positioning table used for routine treatment and examination procedures. It is intended to be operated by a healthcare professional in a medical environment. No special training is required but a review of the following User Manual is important for the safety of the operator and patient. The healthcare professional should read and understand this entire manual before use with a patient. There are no known contraindications with the use of this equipment.
SYMBOL IDENTIFICATION
This symbol, when used in this manual and on product labels, represents a caution warning. Be sure to read and comply with all precautions and warnings.
This symbol, when used in this manual and on product labels, warns against an electrical shock hazard. Be sure to observe and comply with all warnings.
This symbol, when used in this manual or on product labels, indicates a Protective Earth (Ground) Terminal.
This symbol, when used in this manual or on product labels, indicates that the product should be protected from moisture. The humidity specifications for Transport & Storage are listed in the environmental conditions section of this manual.
This symbol is used to indicate that the operator should consult the user manual.
This symbol, when used in this manual and on product labels, indicates that the table and components are a Type B Applied Part pursuant to IEC 60601-1: 2005.
This symbol, when used in this manual or on product labels, indicates alternating
~
----
current (AC).
This symbol, when used in this manual or on product labels, indicates direct current (DC).
IMPORTANT SAFETY INSTRUCTIONS
CAUTION
The Backrest and Leg Rest sections are not designed to support the entire weight of the patient. Do not sit on the Backrest or Leg rest sections.
READ AND SAVE THESE INSTRUCTIONS
The table is not to be used as a gurney or to transport patient from room to room.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
1
IMPORTANT SAFETY INSTRUCTIONS
The use of accessories, transducers, and cables other than those specified by the manufacturer, may result in increased emissions or decreased immunity of the table.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the table should be observed to verify normal operation in the configuration in which it will be used.
The table is designed to be a stand alone table. This table must not be modified or incorporated into any other equipment.
As with any moving mechanism there are potential pinch points around and underneath the table top. It is the respon­sibility of the operator of this equipment to insure that bystanders are not in the area below or around this equipment during operation.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in the area of this equipment. Directions for use of this equipment are described in this manual. The operator should read these sec­tions carefully.
Weight Limit: (patient and accessories) 550 lbs./250 kg. Do not exceed.
Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory patient. The patient may lose balance and fall.
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
Do not lift the table by the table top. This can damage the table.
When lowering the table or using the Trendelenburg functions, make sure there is nothing underneath the table top that can impede motion (like stools, cabinets, accessory parts, cleaners, etc.).
Use this furnishing only for its intended use as described in these instructions. Do not use attachments not recommended by the manufacturer.
Close supervision is necessary when this furnishing is used by, or near children, invalids, or disabled persons.
WARNING
To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Contact Oakworks Customer Service before use.
4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.
DANGER
Risk of electric shock - Connect this furnishing to a properly grounded outlet only. See Grounding Instructions in this manual.
Electrical Shock Hazard. The power supply/control module is located under the table seat. No user serviceable parts are inside the control box. Refer servicing to qualified personnel. Unplug wall plug prior to contact with any cables connected to the power supply.
2
PRODUCT DESCRIPTION
PF400
Electric Lift Towers (2)
Universal Table Extender (accessory)
Adjustable Arm rests (2) (accessory)
Accessory outlets (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 21”-37” (54-94 cm.) Height Range
Motion 2
Motion 3 Powered 0-70º Backrest Motion 4 Powered 0-60º Foot Rest Foot Control All powered motions
Table Voltage Options
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths
Length - 4 section 72” (183 cm.) Padding 4” (10 cm.) Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight
*
Warranty 3 year parts, 1 year labor Safety Listings
Powered ±15º Trendelenburg/ Reverse Trendelenburg
100V/60Hz - 230V/50Hz Switch Mode Power Supply
27” (69 cm.), 29” (74 cm.), 31” (79 cm.) or 33” (84 cm.)
234 lbs. (106 kg.)
FDA, cETLus
Shown with optional wood finish
OPTIONS
Wood Finishes Locking Casters Hand Control
ACCESSORIES
Adjustable Side Arm Rests (2) QuickLock Face Pad Table Extender - 12” (30.5 cm.) Long Lowered Arm Rest
For electrical specs, see Specifications section.
3
PRODUCT DESCRIPTION
PF250
Electric Lift Towers (2)
Manual Lift-Assist Backrest Control
Accessory outlets (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 21”-37” (54-94 cm.) Height Range Motion 2 Powered ±15º Trendelenburg/ Reverse
Trendelenburg Motion 3 Manual 0-80º Backrest Foot Control All powered motions
Table Voltage Options
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths
Length - 2 section 74” (188 cm.) Padding 3” (8 cm.) Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight Warranty 3 year parts, 1 year labor Safety Listings
*
100V/60Hz - 230V/50Hz Switch Mode Power
Supply
27” (69 cm.), 29” (74 cm.), 31” (79 cm.) or 33”
(84 cm.)
212 lbs. (96 kg.)
FDA, cETLus
Shown with optional wood finish
OPTIONS
4” (10 cm.) Comfort Foam Wood Finishes Locking Casters Hand Control
ACCESSORIES
QuickLock Face Pad Table Extender - 12” (30.5 cm.) Long Lowered Arm Rest
For electrical specs, see Specifications section.
4
PRODUCT DESCRIPTION
PF200
Electric Lift Towers (2)
QuickLock™ Face Pad (option)
Accessory outlets (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 19.5”-35.5” (50-90 cm.) Height Range
Motion 2
Foot Control All powered motions
Table Voltage Options
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths
Length - 2 section 73” (185 cm.) Padding 3” (8 cm.) Comfort Foam Upholstery Terra Touch Wood Stain Natural Maple Finish Table Weight Warranty 3 year parts, 1 year labor Safety Listings
*
Powered ±15º Trendelenburg/ Reverse
Trendelenburg
100V/60Hz - 230V/50Hz Switch Mode Power
Supply
27” (69 cm.), 29” (74 cm.), 31” (79 cm.) or 33”
(84 cm.)
212 lbs. (96 kg.)
FDA, cETLus
Shown with optional wood finish
OPTIONS
4” (10 cm.) Comfort Foam Wood nishes Locking Casters Hand Control
ACCESSORIES
QuickLock Face Pad Table Extender - 12” (30.5 cm.) Long Lowered Arm Rest
For electrical specs, see Specifications section.
5
INSTALLATION
All tables come completely assembled and ready to use. Plug the cord into a functioning outlet that is rated for the table. (see Grounding below)
Arrange the power cord and control cords so that they will not create a tripping hazard and where the controls are located to your liking and are conveniently accessible.
Be sure access to plug is not blocked for disconnecting the table from power.
GROUNDING
DANGER
This product must be grounded. If it should malfunction or break down, grounding provides a path of least resistance for electrical current to reduce the risk of electric shock. This product is equipped with a cord having an equipment-grounding conductor and a grounding pin. The pin must be plugged into an appropriate outlet that is prop­erly installed and grounded in accordance with all local codes and ordinances. See U.S. sample below.
Improper connection of the equipment-grounding conductor can result in a risk of elec­tric shock. Check with a qualified electrician or service person if you are in doubt as to whether the product is properly grounded. Do not modify the plug provided with the product - if it will not fit the outlet; have a proper outlet installed by a qualified elec­trician.
Risk of Electric Shock - Connect this furnishing to a properly grounded outlet only.
Grounding Methods
Correct Implementation
Incorrect Implementation
6