This manual was originally drafted
and approved in the English language.
Printed in USA
All rights are reserved. No
part of this document may be
photocopied, reproduced or
translated to another language
without prior written consent
of OAKWORKS®, Inc.
OAKWORKS® is a registered
trademark of OAKWORKS®, Inc.
Notice
The information contained in this
document is subject to change
without notice and should not
be construed as a commitment by
OAKWORKS®, Inc.
OAKWORKS®, Inc. assumes no
responsibility for any errors that
may appear in this document nor
does it make expressed or implied
warranty of any kind with regard
to this material, including, but not
limited to, the implied warranties
of merchantability and fitness for
a particular purpose.
OAKWORKS®, Inc. shall not be
liable for incidental or consequential damages in connection with or
arising out of the furnishing, performance, or use of this document
and the program material which it
describes.
TABLE OF CONTENTS
Product Use Description .....................................................................................................1
Important Safety Instructions
Symbol Identification .................................................................................................... 1
Contact Information ......................................................................................... back cover
PRODUCT USE DESCRIPTION / IMPORTANT SAFETY INSTRUCTIONS
PRODUCT USE DESCRIPTION
The PG and Ultrasound Tables are used to support the patient during examinations and procedures.
It is intended to be operated by a healthcare professional in a medical environment. No special training is required but a review of the following Safety Instructions is important for the safety of the
operator and patient. The healthcare professional should read and understand this entire manual
before use with a patient. There are no known contradictions to the use of this equipment.
SYMBOL IDENTIFICATION
This symbol, when used in this manual and on product labels, represents a caution
warning. Be sure to read and comply with all precautions and warnings.
This symbol, when used in this manual and on product labels, warns against an
electrical shock hazard. Be sure to observe and comply with all warnings.
This symbol, when used in this manual or on product labels, indicates a Protective Earth
(Ground) Terminal.
This symbol, when used in this manual or on product labels, indicates that the
product should be protected from moisture. The humidity specifications for Transport &
Storage are listed in the environmental conditions section of this manual.
This symbol is used to indicate that the operator should consult the user manual.
This symbol, when used in this manual and on product labels, indicates that the table
and components are a Type B Applied Part pursuant to IEC 60601-1: 2005.
This symbol, when used in this manual or on product labels, indicates alternating
~
----
current (AC).
This symbol, when used in this manual or on product labels, indicates direct current
(DC).
IMPORTANT SAFETY INSTRUCTIONS
CAUTION
The Fowler (Backrest) and Leg Rest sections are not designed to support the entire weight of the patient. Do not sit on the
Fowler or Leg rest sections.
READ AND SAVE THESE INSTRUCTIONS
The table is not to be used as a gurney or to transport patient from room to room.
When the footboards are in the down position, it is possible for them to hit the ground in low height and high tilt conditions. Operate the table with the footboard up if possible. If not, then check ground clearance to prevent damage to the
footboard.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information provided in this manual.
1
IMPORTANT SAFETY INSTRUCTIONS
The use of accessories, transducers, and cables other than those specified by the manufacturer, may result in increased emissions or
decreased immunity of the table.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the table should
be observed to verify normal operation in the configuration in which it will be used.
The table is designed to be a stand alone table used with hand held ultrasound probes. This table must not be modified or
incorporated into any other equipment.
As with any moving mechanism there are potential pinch points around and underneath the table top. It is the responsibility of the operator of this equipment to insure that bystanders are not in the area below or around this equipment during
operation.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in the
area of this equipment. Directions for use of this equipment are described in this manual. The operator should read these sections carefully.
Weight Limit: (patient and accessories) 550 lbs./250 kg. Do not exceed.
Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory patient.
The patient may lose balance and fall.
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
Do not put excessive weight on the extended end (beyond the main top) of the cardio panel. Maximum capacity is 150
lbs. / 68 kg.
When lowering the table or using the Trendelenburg functions, make sure there is nothing underneath the table top that
can impede motion (like stools, cabinets, accessory parts, cleaners, etc.).
Use this furnishing only for its intended use as described in these instructions. Do not use attachments not recommended by
the manufacturer.
Close supervision is necessary when this furnishing is used by, or near children, invalids, or disabled persons.
All motor actuators & lifting columns have a duty cycle of approximately 10%, meaning 2 minutes of operation within a
20 minute period. Exceeding this duty cycle can damage this equipment.
WARNING
To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or
damaged, or dropped into water. Contact Oakworks Customer Service before use.
4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.
DANGER
For procedures where accidental table motion is a safety concern, the handset control must be locked out.
Risk of electric shock - Connect this furnishing to a properly grounded outlet only. See Grounding Instructions in this
manual.
Electrical Shock Hazard. The power supply/control module is located under the table seat. No user serviceable parts are inside
the control box. Refer servicing to qualified personnel. Unplug wall plug prior to contact with any cables connected to the
power supply.
2
PRODUCT DESCRIPTION
PG350
Paper roll
holder
Manual 4
position leg
foot control
optional hand control
rest
Electric
Lift Towers
STANDARD SPECIFICATIONS
Motion 1Powered 22”-38” (55.9-96.5 cm.) Height Range
Motion 2Powered ±15º Trendelenburg/ Reverse
Trendelenburg
Motion 3Powered 0-80º Fowler
Motion 4Manual 4 Position Leg Rest
Foot ControlAll powered Motions
Paper Roll HolderAccommodates up to 21” (53.3 cm.) rolls
(roll not included)
Table Voltage Options 120V/60Hz or 230V/50Hz
Table Capacity550 lbs. (250 kg.) (patient and accessories)