Oakworks PF250 User Manual

USER MANUAL
OAKWORKS® PF Series Tables
Shown with optional QuickLock™ Face rest platform and Boiance™ crescent
Shown with optional Adjustable Side Arm Rests and Universal Table Extender
www.oakworksmed.com · 717.235.6807
made in the USA
with US & imported parts
© Copyright 2013 OAKWORKS® , Inc.
English, Printed in USA
All rights are reserved. No part of this document may be photocopied, reproduced or translated to another language without prior written consent of OAKWORKS®, Inc.
OAKWORKS® is a registered trademark of OAKWORKS®, Inc.
Notice
The information contained in this document is subject to change without notice and should not be construed as a commitment by OAKWORKS®, Inc.
OAKWORKS®, Inc. assumes no responsibility for any errors that may appear in this document nor does it make expressed or implied warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
OAKWORKS®, Inc. shall not be liable for incidental or consequen­tial damages in connection with or arising out of the furnishing, per­formance, or use of this document and the program material which it describes.
TABLE OF CONTENTS
Product Use Description ...................................................................................................... 1
Important Safety Instructions
Symbol Identification .................................................................................................... 1
Safety Instructions ......................................................................................................1-2
Product Description & Photo
PF400 ..............................................................................................................................3
PF250 ..............................................................................................................................4
PF200 ..............................................................................................................................5
Installation
Grounding ...................................................................................................................... 6
Directions for Use
Control Operations .....................................................................................................7-8
Moving the Table ...........................................................................................................9
Leveling the Table .........................................................................................................9
Lift-Assist Backrest Adjustment .................................................................................10
Lowered Arm Rest Use ............................................................................................... 10
Locking Casters Use .................................................................................................... 11
Retractable Foot Rests (Stirrup) Use .......................................................................... 11
Adjustable Arm Rest Installation and Adjustment .................................................. 12
Table Extender ............................................................................................................. 13
QuickLock Face Rest ...................................................................................................13
Cleaning & Disinfection ....................................................................................................14
Inspections & Maintenance ............................................................................................. 14
Warranty Information ......................................................................................................... 14
Model Number & Serial Number ..................................................................................... 15
Specifications
Product Specifications ................................................................................................ 16
Environmental Conditions ........................................................................................ 16
Electrical Specifications ............................................................................................. 16
Guidance and manufacturer’s declaration -
Electromagnetic emissions..........................................................................................17
Recommended separation distances .......................................................................... 17
Guidance and manufacturer’s declaration -
Electromagnetic immunity ....................................................................................18-19
Contact Information ............................................................................................back cover
PRODUCT USE DESCRIPTION / IMPORTANT SAFETY INSTRUCTIONS
PRODUCT USE DESCRIPTION
The PF Exam Table is a powered positioning table used for routine treatment and examination proce­dures. It is intended to be operated by a healthcare professional in a medical environment. No special training is required but a review of the following Safety Instructions is important for the safety of the operator and patient. The healthcare professional should read and understand this entire manual before use with a patient. There are no known contradictions to the use of this equipment.
SYMBOL IDENTIFICATION
This symbol, when used in this manual and on product labels, represents a caution warning. Be sure to read and comply with all precautions and warnings.
This symbol, when used in this manual and on product labels, warns against an electrical shock hazard. Be sure to observe and comply with all warnings.
This symbol, when used in this manual or on product labels, indicates a Protective Earth (Ground) Terminal.
This symbol, when used in this manual or on product labels, indicates that the product should be protected from moisture. The humidity specifications for Transport & Storage are listed in the environmental conditions section of this manual.
This symbol is used to indicate that the operator should consult the user manual.
This symbol, when used in this manual and on product labels, indicates that the table and components are a Type B Applied Part pursuant to IEC 60601-1: 2005.
This symbol, when used in this manual or on product labels, indicates alternating
~
----
current (AC).
This symbol, when used in this manual or on product labels, indicates direct current (DC).
IMPORTANT SAFETY INSTRUCTIONS
CAUTION
The Backrest and Leg Rest sections are not designed to support the entire weight of the patient. Do not sit on the Backrest or Leg rest sections.
READ AND SAVE THESE INSTRUCTIONS
The table is not to be used as a gurney or to transport patient from room to room.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
1
IMPORTANT SAFETY INSTRUCTIONS
The use of accessories, transducers, and cables other than those specified by the manufacturer, may result in increased emissions or decreased immunity of the table.
The table should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the table should be observed to verify normal operation in the configuration in which it will be used.
The table is designed to be a stand alone table. This table must not be modified or incorporated into any other equipment.
As with any moving mechanism there are potential pinch points around and underneath the table top. It is the respon­sibility of the operator of this equipment to insure that bystanders are not in the area below or around this equipment during operation.
Proper operation of this equipment is very important for the safety of the operator, patient, and any other individuals in the area of this equipment. Directions for use of this equipment are described in this manual. The operator should read these sec­tions carefully.
Weight Limit: (patient and accessories) 550 lbs./250 kg. Do not exceed.
Be certain that the table is completely lowered without any tilt being present prior to discharging an ambulatory patient. The patient may lose balance and fall.
This table is not suitable for use in the presence of flammable anesthetic mixture with air or oxygen or nitrous oxide.
Do not lift the table by the table top. This can damage the table.
When lowering the table or using the Trendelenburg functions, make sure there is nothing underneath the table top that can impede motion (like stools, cabinets, accessory parts, cleaners, etc.).
Use this furnishing only for its intended use as described in these instructions. Do not use attachments not recommended by the manufacturer.
Close supervision is necessary when this furnishing is used by, or near children, invalids, or disabled persons.
WARNING
To reduce the risk of burns, fire, electric shock or injury to persons:
1. Unplug this furnishing from the electrical outlet before cleaning.
2. Unplug from outlet before adding or removing parts.
3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Contact Oakworks Customer Service before use.
4. Keep the cord away from heated surfaces.
5. Never drop or insert any object into any opening.
6. Do not use outdoors.
7. Do not operate where aerosol (spray) products are being used or where oxygen is being administered.
DANGER
Risk of electric shock - Connect this furnishing to a properly grounded outlet only. See Grounding Instructions in this manual.
Electrical Shock Hazard. The power supply/control module is located under the table seat. No user serviceable parts are inside the control box. Refer servicing to qualified personnel. Unplug wall plug prior to contact with any cables connected to the power supply.
2
PRODUCT DESCRIPTION
PF400
Universal Table Extender (accessory)
Adjustable Arm rests (2) (accessory)
Electric Lift Towers (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 22”-38” (56-97 cm.) Height Range
Motion 2 Powered ±15º Trendelenburg/ Reverse
Trendelenburg
Motion 3 Powered 0-53º Backrest
Motion 4 Powered 0-51º Foot Rest
Foot Control All powered motions
Table Voltage Options 120V/60Hz or 230V/50Hz
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths 27” (69 cm.), 29” (74 cm.) or 31” (79 cm.)
Length - 4 section 73” (185 cm.)
Padding 4” (10 cm.) Comfort Foam
Upholstery Terra Touch
Wood Stain Natural Maple Finish
Table Weight
*
234 lbs. (106 kg.)
Warranty 3 year parts, 1 year labor
Safety Listings FDA, cETLus, CE marked
Retractable Foot Rests (Stirrups) (2) (option)
Accessory outlets (2)
Shown with optional wood finish
OPTIONS
Wood Finishes
Locking Casters
Hand Control
Retractable Foot Rests (Stirrups) (2)
ACCESSORIES
Adjustable Side Arm Rests (2)
QuickLock Face Pad
Table Extender - 12” (30.5 cm.) Long
Lowered Arm Rest
For electrical specs, see Specifications section.
3
PRODUCT DESCRIPTION
PF250
Electric Lift Towers (2)
Manual Lift-Assist Backrest Control
Accessory outlets (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 22”-38” (56-97 cm.) Height Range
Motion 2 Powered ±15º Trendelenburg/ Reverse
Trendelenburg
Motion 3 Manual 0-80º Backrest
Foot Control All powered motions
Table Voltage Options 120V/60Hz or 230V/50Hz
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths 27” (69 cm.), 29” (74 cm.) or 31” (79 cm.)
Length - 2 section 73” (185 cm.)
Padding 3” (8 cm.) Comfort Foam
Upholstery Terra Touch
Wood Stain Natural Maple Finish
Table Weight
Warranty 3 year parts, 1 year labor
Safety Listings FDA, cETLus, CE marked
*
212 lbs. (96 kg.)
Shown with optional wood finish
OPTIONS
4” (10 cm.) Comfort Foam
Wood Finishes
Locking Casters
Hand Control
ACCESSORIES
QuickLock Face Pad
Table Extender - 12” (30.5 cm.) Long
Lowered Arm Rest
For electrical specs, see Specifications section.
4
PRODUCT DESCRIPTION
PF200
Electric Lift Towers (2)
QuickLock™ Face Pad (option)
Accessory outlets (2)
Foot Control
STANDARD SPECIFICATIONS
Motion 1 Powered 21”-37” (56-94 cm.) Height Range
Motion 2 Powered ±15º Trendelenburg/ Reverse
Trendelenburg
Foot Control All powered motions
Table Voltage Options 120V/60Hz or 230V/50Hz
Table Capacity 550 lbs. (250 kg.) (patient and accessories)
Widths 27” (69 cm.), 29” (74 cm.) or 31” (79 cm.)
Length - 2 section 73” (185 cm.)
Padding 3” (8 cm.) Comfort Foam
Upholstery Terra Touch
Wood Stain Natural Maple Finish
Table Weight
Warranty 3 year parts, 1 year labor
Safety Listings FDA, cETLus, CE marked
*
212 lbs. (96 kg.)
Shown with optional wood finish
OPTIONS
4” (10 cm.) Comfort Foam
Wood nishes
Locking Casters
Hand Control
ACCESSORIES
QuickLock Face Pad
Table Extender - 12” (30.5 cm.) Long
Lowered Arm Rest
For electrical specs, see Specifications section.
5