Oakton Instruments PH 150 User manual

Page 1
Output (USB or
RS232 – PC or Printer
Communication)
150 & 450 Series Waterproof Handheld Meters
pH/mV Operation Instructions
Models:
pH 150
pH/mV
pH 450
pH/mV/ISE
PC 450
pH/mV/Con
PD 450
pH/mV/DO
Getting Started/Connections
After installing (2) AA batteries and/or connecting the optional 110/220 VAC power supply, connect the desired sensors to the corresponding ports.
DO: 9-Pin RDO/ATC (PD 450 only)
A/C Power Adapter
12 mm and 16 mm probes can utilize the Grip-Clip or more sensors to a beaker and to the instrument as needed. The stand can be extended as shown above or used for wall-mounting.
BNC: pH, mV/ORP or Ion Selective
ATC/Con:
6-Pin ATC or ATC/Conductivity (PC 450 only)
Output (USB or RS232 – PC or Printer Communication)
to attach one
Keypad Functions
Press once to power ON in the mode that was previously used. Press again to turn backlight on for one minute or off (450 series only). Hold for 3 seconds to power OFF.
Toggle between measurement and calibration modes. In SETUP mode, BACK serves to return to the previous menu option or setting.
Confi rm calibration values in CAL mode. Confi rm selections in SETUP mode. Freeze or release the measured reading.
Customize instrument settings and preferences. (See also Setup Programs)
Toggle between available measurement types.
Save measurement into memory. Increase value or scroll up in SETUP or manual calibration.
Recall saved values from memory. Decrease value or scroll down in SETUP or manual calibration.
Send output data to printer or computer. (450 series only).
Setup Programs
To access the settings below, press SETUP. Up/down arrows will display the available options. Press ENTER to accept the desired setting, or BACK to return to the previous option and/or exit.
Confi guration Options
• Ready indicator ON / OFF / or Automatic HOLD when stable
• Choose ºCelsius or ºFahrenheit
Find Quality Products Online at: sales@GlobalTestSupply.com
pH Buffer Options
• Select the desired pH Buffer Calibration Group:
USA (1.68, 4.01, 7.00, 10.01, 12.45) or NIST (1.68, 4.01, 6.86, 9.18, 12.45) or DIN (1.09, 3.06, 4.65, 6.79, 9.23, 12.75) or MAN (manual adjustment of any custom pH values that are ≥1 pH unit
apart. 450 series only)
• Select number of calibration points
Select Calibration Due Reminder
• Set number of days from 0-60 for desired parameter
View Calibration Data
• Press ENTER to view each point that is calibrated.
View Electrode Data
• Press ENTER to view mV Offset and Slope % of the measured reading.
System Settings
Data Logging: MANUAL upon key press only TIMED interval. Choose (SEC / MIN / HOUR) interval.
• Automatic shut off after 10 minutes. Choose ON or OFF.
Clock Settings: Date: Choose USA (MM/DD/YYYY) or Euro (DD/MM/YYYY). Time: Choose (24HR or 12HR). If 12HR, choose AM or PM.
Set Printer Type:
CSV (Comma Separated Values) – best format for computer Printer (Text) – best format for printer.
Choose Manual (MAN) upon key press or TIMED interval. If timed, choose (SEC / MIN / HOUR).
Reset
NO. Exits from reset menu options without action.
FACTORY RESET. Returns all settings except date/time and ATC calibration to factory default values after ENTER is pressed then restarts meter.
DATA RESET. Erases data stored in memory while retaining other settings after ENTER is pressed.
CALIBRATION RESET. Erases non-ATC calibration data while retaining other settings after ENTER is pressed.
pH Calibration
For best results, periodic calibration with known, accurate standards is recommended. Calibrate with standards that bracket your intended measuring range while including a neutral standard (pH 7.00 or 6.86). For example, if you expect to measure samples from pH 6.2 to pH 9.5, calibration with 4.01, 7.00, and 10.01 standards will work well. Provide stirring for best results. After calibration with two or more points, the active slope segment of the measurement will be visible on the bottom display during measurement. 100 % slope will be shown if only one calibration point is performed and “- - -“ if no calibration is performed. The meter will automatically return to measurement mode upon successful completion of the number of specifi ed calibration points. To specify a different number of pH calibration points, see pH Buffer Options.
Using Automatic Buffer Recognition
1. While in pH measurement mode, dip the pH and ATC sensor(s) into
your fi rst standard, then press CAL. The primary display will search for the nearest standard value, while the secondary display will show the un-adjusted value.
2. When the “READY” indicator appears, press ENTER to accept.
The primary reading will fl ash “DONE”.
3. Rinse your electrode(s) then dip into the next pH standard. The primary
display will search for the nearest standard value that has not yet been calibrated, while the secondary display will show the unadjusted value. When the “READY” indicator appears, press ENTER to accept.
4. To calibrate another pH standard repeat Step 3 or press MEAS to return
to pH measurement mode.
Using Manual Recognition / Custom Buffers (450 Series Only)
1. While in pH measurement mode, dip the pH and ATC sensor(s) into your
fi rst standard then press CAL.
2. When the READY indicator appears, use up/down arrows to adjust
the primary reading to match the standard value at the measured temperature, then press ENTER.
3. Rinse your electrode(s) then repeat Step 2 with a standard that is ≥1 pH
unit from the previous standard value.
4. To calibrate another pH standard, repeat Step 3 or press MEAS to return
to pH measurement mode.
mV Offset Adjustment
1. While in mV measurement mode, dip the ORP and ATC sensors into
a solution with a known mV value (i.e. Zobel, Light’s, quinhydrone, or iodide/triiodide) and stir.
2. When the “READY” indicator appears, use up/down arrows to adjust the
primary reading to match the mV value at the measured temperature, then press ENTER. The meter allows an adjustable maximum value of ±200 mV from the factory default mV value. When an offset has been stored successfully, R.mV replaces mV.
www.GlobalTestSupply.com
Temperature Calibration/Manual ATC
1. Press CAL from any measurement, then press MODE.
2. Skip to step 3 for manual ATC, otherwise, dip the temperature sensor into a solution with a known accurate temperature. The upper display shows the active temperature while the lower display shows the factory default temperature without adjustment.
3. Use up/down arrows to adjust the upper display. Press ENTER to accept the calibration temperature. The maximum adjustable value is ±10 ºC (or ±18 ºF) from factory default.
Error Messages
“ERR” will appear when an error condition exists or the incorrect key is pressed. Common examples include:
• Pressing ENTER during calibration before the “READY” indicator
appears. Wait for the “READY” indicator before pressing ENTER.
UR (Under Range) • OR (Over Range)
Intended Use, Maintenance & Precautions
These handheld meters use sensors to detect various parameters for water-based measurements. For routine maintenance disconnect the power cord or battery, then dust or wipe the display using a damp cloth. If necessary, warm water or a mild water based detergent can be used. Immediately remove any spilled substance from contact with the meter using the proper cleaning procedure for the type of spill.
• Do not use this equipment in potentially explosive atmospheres.
• Refer to the electrode instructions for use, storage and cleaning.
• Ensure that no liquid enters the instrument.
• Do not use any aggressive cleaning chemicals (solvents or similar
agents).
• There are no user serviceable parts inside. Attempts to service internal
parts may void the warranty.
• Not intended for medical applications or patient use.
WARNING: No modifi cation of this equipment is allowed.
Instrument Operating Conditions
Operating Ambient Temp. 5 to 45 ºC Operating Relative Humidity 5 to 85 %, non-condensing Storage Temp. -20 to +60 ºC Storage Relative Humidity 5 to 85 %, non-condensing Pollution Degree 2 Overvoltage Category II Weight 500 g Size (L x W x H) 21.15 x 9.87 x 5.85 cm Regulatory & Safety CE, TUV 3-1, FCC Class A Power Rating DC Input: 9 VDC 1 A
Battery Requirement
Vibration Shipping/handling per ISTA #1A Shock Drop test in packaging per ISTA #1A Enclosure (Designed To Meet) IP67 (using rubber covers)
Universal Power Adapter Operating Conditions
Operating Ambient Temp. 0 to 50 ºC Operating Relative Humidity 0 to 90 %, non-condensing Storage Temp. -20 to +75 ºC Storage Relative Humidity 0 to 90 %, non-condensing Pollution Degree 2 Overvoltage Category II
Power Rating
2 x AA (LR6) 1.5 V batteries (replace batteries when battery sign blinks)
I/P: 100 - 240 V, 50/60 Hz, 0.3A O/P: 9 VDC 1 A