OAKTON 37250-00, 08369-50, 37250-10, 37250-20, 08369-70 Instruction Manual

Page 1
OPERATING INSTRUCTIONS
OAKTON®Models 08369-50, -70
37250-00, -10, -20
Hygrothermographs
Printed in the U.S.A. 1/00
00703-07
OAKTON warrants this instrument to be free from significant deviations in material and workmanship for a period of one year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warrantied time period, please return—freight prepaid—and correction will be made without charge. OAKTON alone will determine if the product problem is due to deviations or customer misuse.
Out-of-warranty products will be repaired on a charge basis.
11. Return of items
Authorization must be obtained from our Customer Service Department before returning items for any reason. When applying for authorization, please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possi­ble damage or loss. We will not be responsible for damage resulting from careless or insufficient packing. Arestocking charge will be made on all unauthorized returns.
NOTE: We reserve the right to make improvements in design, construction, and
appearance of products without notice.
Page 2
32
Table of Contents
1. Introduction.........................................................................................4-5
2. Components............................................................................................6
3. Getting started .......................................................................................7
3.1 Unpacking .............................................................................................................................7
3.2 Attaching the cylinder drum..............................................................................................7
4. Hygrothermograph operation...........................................................8-21
4.1 Powering your hygrothermograph................................................................................8-9
4.2 Cartridge pens ....................................................................................................................10
4.3 Setting the time...................................................................................................................11
4.4 Chart paper....................................................................................................................12-17
4.5 Calibration adjustment.................................................................................................18-19
4.6 Changing drum rotation length..................................................................................20-21
5. Installing optional axial fan.................................................................22
6. Precautions ...........................................................................................23
7. Specifications...................................................................................24-25
8. Ordering information...........................................................................26
9. Appendix 1: Readings in areas with high humidity ............................27
10. Warranty .............................................................................................28
11. Return of Items ...................................................................................28
NOTES
Write down the name and information of your OAKTON distributor here.
Page 3
54
1. Introduction
Thank you for selecting an OAKTON thermohygrometer. These precision instru­ments are useful for monitoring temperature and relative humidity conditions over time. They create a permanent chart recording of these measurements for your records. Their quartz-controlled drive maintains an even drum speed even when batteries are weak (or when the spring winding runs down).
Typical applications include general weather measurements, agriculture, horticulture, monitoring the condition of test laboratories, humidity chambers, computer rooms, precision machinery rooms, food storage facilities, warehouses, museums, bank vaults, film and book storage facilities.
You have one of five available models:
Minidrum hygrothermograph
Model number: 08369-70 Special features:
• Compact size
• Records on 7-day rotation
• Protective acrylic cover locks into place
Domed minidrum hygrothermograph
Model number: 08369-50 Special features:
• Our most compact model
• Records on 7-day rotation
• Protective glass dome
Spring-wound, two-speed hygrothermograph
Model number: 37250-10 Special features:
• Spring wound— no batteries needed!
• Records on 1- or 7-day rotation
• Protective acrylic cover locks into place
Three-speed hygrothermograph
Model number: 37250-00 Special features:
• Records on 1- or 7-, or 32-day rotation
• Protective acrylic cover locks into place
Long-cycle hygrothermograph
Model number: 37250-20 Special features:
• Records on 1 or 3 month rotation
• Protective acrylic cover locks into place
Page 4
Attaching the cylinder drum
NOTE: Long-cycle model 37250-20 already has an attached cylinder drum.
1. Unscrew the hold-down nut from the
center shaft.
2. Carefully insert the cylinder drum onto
the center shaft. Make sure the cylinder drum is completely lowered over the center shaft.
See figure
3. Replace the hold-down nut on the top of
the center shaft.
76
3. Getting started
Unpacking
1. Remove all components from the packing material.
2. Carefully remove the acrylic or glass protective cover from the hygrothermo-
graph. If your model has a handle, turn the carrying handle to the left until it stops, then lift up the protective cover.
3. Carefully remove the styrofoam piece located beneath the pen arm mechanism.
4. Remove the white paper that secures the humidity pen.
5. Lift the metal pen holding fitting up from the temperature pen until it is removed
from the pen guide.
6. Move the pen lift lever towards you to pull the temperature and humidity pens to
their resting position.
NOTE: Keep all original packing in case you need to transport the instrument.
2. Components
All OAKTON hygrothermographs have these same features:
3.1
3.2
1
2
1. Pen tip for temperature
2. Cylinder drum with quartz clock
3. Pen tip for humidity
4. Paper holder
5. Pen lift lever
6. Protective cover
7. Pen arms
8. Pen guide
3
4
5
6
7
8
A
A
Page 5
98
4. Hygrothermograph operation
Powering your hygrothermograph
4.1
Minidrum models (08369-50 and -70):
These models require one AAbattery. The battery compartment is located on top of the cylinder drum. To open the battery compartment, slide the cover of the battery compartment to the side. Make sure that the + end of the battery is facing up.
Spring wound, 2 speed model (37250-10):
This model is spring wound (requires no batteries). The spring winding shaft is located on top of the cylinder drum. To wind up the unit, turn the shaft counterclockwise until it is too stiff to move any further. This hygrothermograph will make a “clicking” sound while running. The drum will maintain an even rotation speed even when the spring is winding down.
3 speed model (37250-00):
This model requires two C batteries. The battery compartment is located on top of the cylinder drum. To open the battery compartment, pull the silver knob on top of the battery compartment away from the center of the cylinder drum. Note polarity on the lid of the battery compartment.
This model features a battery indicator. To check the battery, press the red check button next to the battery compartment. Note the color of the LED indicator:
• A green LED means the batteries are in good condition.
• A red LED means the batteries need to
be replaced.
Long cycle model (37250-20):
This model requires one C battery. The battery compartment is located on top of the cylinder drum. To open the battery compartment, flip open the lid on top of the cylinder drum. Note polarity inside the battery compartment.
This model features a battery indicator. To check the battery, press the red check button next to the battery compartment. The power indicator has green, white and red zones.
• the green zone means the battery is in good condition.
• The white zone means the battery is low
and will have to be replaced soon.
• The red zone means the battery is very
low and should be replaced immediately.
Top of Drum
Close-up of battery indicator
Close-up of battery indicator
BATTERY
Page 6
1110
Setting the time
4.3
Models 08369-50, 08369-50, 37250-10, 37250-00:
1. Pull the pen lift lever towards you to
move the temperature and humidity pens to their resting position.
See figure
2. Turn the cylinder drum one full
rotation clockwise.
See figure
3. Turn the cylinder drum counterclock-
wise until the pen tips align with the correct time of day or night. Numbers indicating time are located at the top and bottom of the chart paper.
NOTE: Make sure that the numbers on the graph are right side up. Change the chart paper when the cylinder has com­pleted one rotation.
Long cycle model (37250-20) only:
1. Pull the pen lift lever slightly towards
you to move the temperature and humidity pens away from the chart.
See figure
2. Turn the chart assembly knob
(located on top of the chart assembly) clockwise.
See figure
3. Stop when the time printed on the
chart paper and pen tips line up about two hours past the present time.
4. Turn the cylinder counterclockwise to
the correct present time.
5. To take up slack in the chart paper,
turn the chart paper roll counterclock­wise until the paper is taut. Be sure not to offset the alignment of present time and pen tips.
See figure
Cartridge pens
4.2
To operate the pens:
1. Pull the pen lift lever towards you to move the
temperature and humidity pens to their resting position.
See figure
2. Remove the pen caps by rotating them counter
clockwise.
3. Push the pen lift lever away from you until the
temperature and humidity pens are touching the paper on the cylinder drum.
NOTE: Do not touch the pen tips, because skin oils will prevent the ink from flowing smoothly.
To replace the pens:
The pen tips can be used for up to a year, but we recommend that you replace them at least every six months. Conditions such as high temperature or low humidity may shorten pen life to three months.
See the “Accessories” section on page 26 to order extra pens.
1. Hold the pen arm in your right hand and pull
the pen cartridge off with your left hand. If it is difficult to pull off, use a small pliers.
See figure
2. To attach the replacement cartridge pen, insert it
over the arm until it is fully attached.
A
B
B
B
C
D
E
D
E
C
B
A
A
A
Page 7
To replace chart paper for long-cycle model 37250-20:
The long-cycle hygrothermograph comes with one roll of paper installed. As the unit records, paper moves from the Chart Roll to the Take-Up Roll. When the paper is completely used, you need to replace the chart paper.
NOTE: do not discard cardboard tube from the original chart paper roll. You will need this cardboard tube for use on the Take-Up Roll.
For best results, we recommend you use our chart paper. See the “Accessories” section on page 26 to order extra chart paper rolls.
When the chart paper is nearing the end of its cycle, a red stripe will appear on the lower edge of the chart paper. The stripe means there are only three days left, and that the paper should be changed by that time.
1312
To change the chart paper:
1. Remove the protective cover. Pull the
pen lift lever towards you so the pen tips do not touch the paper. Write the time and date on the chart paper.
2. Wind the old chart paper onto the
Take-Up Roll by turning the Timing Roll knob clockwise (in the direction
of the arrow).
See figure
3. Release the right and left chart
assembly locks located on the base of the instrument near the chart drive mechanism.
See figure
4. Pull the chart assembly back using
the handle located behind the chart assembly. To prevent the unit from
tipping over, hold the base of the instrument with your other hand. Use the handle to keep the chart assembly in a horizontal position.
See figure
Chart paper
4.4
NOTE: See pages 13-17 for directions on changing the chart paper on long-cycle model 37250-20.
Your cylinder drum comes with one sheet of chart paper already attached. See the “Accessories” section on page 26 to order extra chart paper.
To replace chart paper for models 08369-50, 08369-50, 37250-10, 37250-00:
1. Wind the new chart onto the cylinder so
that the two ends of the paper overlap.
See figure
2. Secure this overlapping section with the
paper holder.
See figure
3. Make sure the markings on the paper
coincide where the overlap occurs.
4. Follow directions on bottom of page 11
to set the correct time.
A
A
B
A
A
B
C
C
B
B
Timing roll knob
Chart assembly locks
Page 8
1514
7. Replace the Take-Up Roll with the
empty cardboard tube into the chart assembly.
• Place the empty cardboard tube (from step 5) back over the metal spindle of the Take-Up Roll. Make sure that the slot on the Take-Up Roll fits into the ridge on the lower disk.
• Replace the upper plastic disk of the
Take-Up Roll by turning it clockwise.
See figure
• Place the pointed spindle tip (protruding through the upper disk) into the receptacle at the left (on the upper part of the assembly). Push to the left, and reinsert the right end of the take-up cylinder into the bottom groove. You may have to turn the cylinder slightly until it slides into the groove.
See figure
8. Install the new chart paper on the
Chart Roll.
• Load the new chart paper onto the Chart Roll, making sure that the slot on the chart paper roll fits into the ridge on the lower disk.
• Make sure that the chart paper is loaded correctly. The side of the paper with the oval-shaped sprocket holes should be at the upper disk (top), while the side with the round sprocket holes should be at the lower disk (bottom).
• Replace the upper plastic disk of the
Chart Roll by turning it clockwise.
See figure
5. Remove the empty cardboard tube
from the Chart Roll (see sticker on top of chart assembly for position).
See figure
• Push the Chart Roll towards the top of the chart assembly, which removes it from the groove at the base of the assembly.
• Pull the Chart Roll out from the bottom.
• Remove the upper plastic disk of the Chart Roll by turning it counter-clockwise.
• Remove the cardboard tube from the metal spindle (spindle is inside the cardboard tube). Save this cardboard tube for step 7.
See figure
6. Remove the Take-Up Roll with
the old chart paper from the chart assembly (see sticker on top of chart assembly for position).
See figure
• Push the Take-Up Roll toward the top of the chart assembly, which removes it from the groove at the base of the assembly.
• Pull the Take-Up Roll out from the bottom.
• Remove the upper plastic disk of the Take-Up Roll by turning it counter-clockwise.
• Remove the used chart paper, complete with inner cardboard tube, from the metal spindle(spindle is inside the cardboard tube).
See figure
D
D
E
E
E
E
F
G
G
F
upper disk
cardboard tube
lower disk
slot
stopper
E
upper disk
cardboard tube
lower disk
slot
stopper
E
Chart roll
Take-up roll
Take-up roll
Page 9
1716
9. Install the Chart Roll with the new
chart paper back into the chart assembly.
• Place the pointed spindle tip (pro-
truding through the upper disk) into the receptacle at the left (on the upper part of the assembly). Push to the left, and reinsert the right end of the starting roll into the groove (on the lower part of the assembly). Make sure that the roll is securely installed and can pivot easily.
See figure
• Make sure paper is positioned so
it will run in a counterclockwise direction around the Chart Roll.
See figure
10. Pull out the chart paper about 6
inches. Draw the paper around the Timing Roll. Make sure the sprock­et holes in the paper are placed correctly over the sprockets on the
Timing Roll.
See figure
11. Tape the end of the chart paper to
the Take-Up Roll. Remove any other paper that might still be attached to the Take-Up Roll. Be sure the paper is straight and even so that it tracks properly.
See figure
12. Advance the chart with the timing
knob.
13. Return the chart assembly upright
to its original position.
See figure
14. Tighten the right and left assembly
locks to hold chart assembly in place.
See figure
15. To set time, see bottom of page 11.
16. Make sure to check that pen car-
tridges and batteries are still fresh.
Notes
Be sure to check at least once a month to guarantee accuracy, since the instru­ment operates over a long period of time (30 or 90 days).
Also make sure to periodically check that the pens are operating correctly.
Vibrations could loosen the chart paper in about 60 days. The paper should be tightened for accurate time readings.
H
H
I
I
K
L
M
M
K
L
J
J
Chart roll
Take-up roll
Timing roll
Page 10
1918
Calibration adjustment
4.5
Your hygrothermograph is factory calibrated. However, vibration during shipment may make a slight readjustment necessary.
To adjust calibration:
1. Compare the hygrothermograph reading to another temperature/ humidity
instrument known to be accurate.
2. Raise and lower the reading by turning the adjustment screws.See drawings on
these two pages for location of the adjustment screws.
Minidrum hygrothermograph
(model 08369-70) On the minidrum hygrothermograph
model, the calibration screws are locat­ed behind the two stickers on the back of the beige case protecting the pen mechanism.
See figure
Remove the stickers to reach the cali­bration adjustment screws.
See figure
Domed minidrum hygrothermograph
(model 08369-50)
+
-
+
-
T
E
M
P
.
H
U
M
I
D
.
A
D
J
U
S
T
.
A
D
J
U
S
T
.
All other models
(models 37250-00, -10, -20)
Calibration adjustment screw locations
Humidity calibration adjustment screw
Temperature calibration adjustment screw
Humidity calibration adjustment screw
Temperature calibration adjustment screw
A
A
B
B
Page 11
2120
Three speed hygrothermograph 37250-00
This model lets you select between a 1, 7, and 32 day drum rotation. When you receive your hygrothermograph from the factory, it is set for a 7 day rotation.
See figure
IMPORTANT: make sure you have the appropriate chart paper for the rotation speed you select. See page 26 to order more chart paper.
1. To switch from 7 day operation to
1 day operation, flip the switch on top of the cylinder drum to 1D.
See figure
2. To switch from 7 day operation to
32 day operation, flip the switch on top of the cylinder drum to 32D.
See figure
7D
1D
32D
Changing drum rotation length
NOTE: Minidrum models 08369-50 and 08369-70 operate on a 7 day rotation only. Long cycle model 37250-20 accepts either 1 month or 3 month chart rolls—it is not
necessary to adjust drum speed.
Spring wound, two speed hygrothermograph 37250-10
This model lets you select between a 1 and 7 day drum rotation. When you receive your hygrothermograph from the factory, it is set for a 7 day rotation.
IMPORTANT: make sure you have the appropriate chart paper for the rotation speed you select. See page 26 to order more chart paper.
1. Look at the bottom of the cylinder
drum. You will see notches for two gears. Only one gear will be in place.
See figure
2. The alternate gear is located on a post
at the corner of the thermohygrome­ter’s base. Remove it from the post.
See figure
3. Using two screwdrivers, lift the gear
currently installed on the bottom of the cylinder drum.
See figure
4. Insert the other gear onto the oppo-
site shaft on the bottom of the drum.
• The one-day gear slides on the shaft
marked “24 H / Z = 22”. This gear has 22 teeth and a larger diameter than the 7-day gear.
• The seven day gear slides onto the
shaft marked 168H / Z = 18. This gear has 18 teeth.
5. Place the gear not in use over the
post at the corner of the thermohy­grometer’s base for storage.
6. Install the cylinder drum onto
the cylinder shaft. See page 7 for directions.
4.6
Z
=
1
8
1
6
8
H
Z
=
2
2
2
4
H
7D
1D
32D
A
A
D
D
E
E
F
F
B
B
C
C
Page 12
2322
6. Precautions
1. Do not use your hygrothermograph:
• in direct sunlight
• in temperatures below –20°C or above 50°C
• near ovens, stoves, or other heating equipment
• near harsh chemicals such as paint thinner or ammonia
• in dusty or wet environments
• within magnetic fields
• in areas with strong vibrations, such as loudspeakers or motors
2. For best results, use only the pens and chart paper listed under the “Accessories”
section of this manual (page 26). We cannot guarantee results with other pens and other paper.
3. Do not repair the instrument yourself. Should repairs be necessary, please return
the instrument to place of purchase. See the back cover of this manual for infor­mation on instrument Warranty and Return of Items.
4. If your instrument will not be used for long periods of time, make sure to replace
the pen caps and remove the battery.
5. Installing optional axial fan
NOTE: The axial fan is not for use with minidrum models 08369-50 and 08369-70.
The external draft fan (optional) attaches to the protective cover. It improves instru-
ment response and accuracy by directly pulling surrounding air into the instrument.
Use the fan in areas where ventilation is poor.
If you want to order an axial fan, see the “Accessories” section on page 26.
To attach the fan:
1. Install the fan on the single
column of five vents located on the protective cover.
See figure
2. Position fan so that the spring clips
on the fan go into the upper vent slots on the hygrothermograph cover.
See figure
3. Let the fan slide down until it stops.
Make sure both spring clips are fully engaged in the cover.
See figure
4. Position the instrument so the
flow of air from the fan will not be blocked.
5. Plug the power cord into an 110 to
120 volt, 50/60 Hz outlet. For 220 V operation, order optional 220 V transformer (sold separately in “Accessories” section on page 26).
A
B
B
C
C
A
Page 13
24 25
7. Specifications
Model Minidrum Domed minidrum
hygrothermograph hygrothermograph
Model number 08369-70 W08369-50
Humidity range 5 to 90% RH 5 to 90% RH
Humidity accuracy ±5% from 10 to 90% RH; ±5% from 10 to 90% RH;
±7% from 5 to 10% RH ±7% from 5 to 10% RH
Humidity chart range 5 to 100% RH 5 to 100% RH
Humidity sensor human hair bundle human hair bundle
Humidity chart graduations 5% RH 5% RH
Temperature range –6 to 40°C / 22 to 104°F –6 to 40°C / 22 to 104°F
Temperature accuracy ±2°C / ±3.6°F ±2°C / ±3.6°F
Temperature sensor bimetallic strip bimetallic strip
Temp. chart graduations 2°C / 2°F 2°C / 2°F
Chart size 3.6"H x 8.1"L 3.6"H x 8.1"L
Chart rotation 7 day (172 hr) 7 day (172 hr)
Power One AA battery (included) One AA battery (included)
Dimensions 6"W x 7.4"H x 3.9"D 7.5"H x 5" dia
Shipping weight 4 lbs 4 lbs
Spring wound, two speed Three speed Long cycle
hygrothermograph hygrothermograph hygrothermograph
37250-10 37250-00 37250-20
5 to 90% RH 5 to 90% RH 5 to 90% RH
±3% from 10 to 90% RH; ±3% from 10 to 90% RH; ±3% from 10 to 90% RH;
±5% from 5 to 10% RH ±5% from 5 to 10% RH ±5% from 5 to 10% RH
0 to 100% RH 0 to 100% RH 0 to 100% RH
human hair bundle human hair bundle human hair bundle
1% RH 1% RH 1% RH
–10 to 50°C / 14 to 122°F –10 to 50°C (14 to 122°F) –20 to 50°C
±1°C / ±1.8°F ±1°C / ±1.8°F ±1°C
bimetallic strip bimetallic strip bimetallic strip
1°C / 2°F 1°C / 2°F 1°C
6.6"H x 11.5"L 6.6"H x 11.5"L 6.6"H x 114" L (1-month)
6.6"H x 300" L (3-month)
1 day (26 hr) or 1 day (26 hr), 7 day (172 hr) 1-month (38 day max) or
7 day (172 hr) or 32 day (810 hr) 3-month (100 day max)
Spring wound Two C batteries (included) One C battery (included)
13.3"W x 11.5"H x 5.1"D 13.3"W x 11.5"H x 5.1"D 13.3"W x 11.5"H x 5.1"D 9 lbs 9 lbs 9 lbs
Page 14
27
26
8. Ordering information
Additional Hygrothermographs
WD-08369-70 Minidrum hygrothermograph WD-08369-70 Domed minidrum hygrothermograph WD-37250-10 Spring wound, two speed hygrothermograph WD-37250-00 Three speed hygrothermograph WD-37250-20 Long cycle hygrothermograph
9. Appendix 1: Readings in areas with high humidity
The specially processed human hair bundle, such as the one used in this instrument, is the best sensor for ambient humidity measurements in terms of linearity, repeatability, and temperature compensation. Human hair will elongate linearly as humidity increases or decreases in the range of 5 to 90% RH.
However, one of the characteristics of human hair is that exposure to humidity levels above 90%RH causes it to undergo a change of state. This condition is called the “saturated state”, and occurs with even short exposure to high humidity levels. When the hair bundle reaches the saturated state, it contracts and causes a 7 to 9% downward shift in the humidity indications. If the humidity then stays below 80%, the hair bundle will elongate linearly for approximately two weeks. After this time period, the hair bundle is in its dry state.
The hair bundle has two curves, depending on the state of the hair bundle. These characteristics can be taken into account when using the instrument. The instrument is calibrated at the factory with the hair bundle in the dry state. Generally, the instrument is used indoors with humidity levels below 90%.
When used in air conditioned areas or other areas of moderate humidity, you do not need to adjust the instrument.
How to use the instrument in high humidity areas:
For use above 85% RH, turn the humidity fine adjustment screw counterclockwise to raise the humidity reading by 7 to 9%. Readings will be within specifications over the entire range when the instrument is exposed to high humidity conditions (85 to 100%) at least once every two weeks.
Note any calibration change either on the chart paper or on the instrument itself. This way you can change the instrument back to its original calibration if required.
Model Use with Temperature Rotation Increments Quantity/ number model type range pack
WD-08369-60 Mini-drum 22 to 104°F 7-day 2-hour 100 WD-08369-55 Mini-drum –6 to 40°C 7-day 2-hour 100 WD-08368-10 2- or 3-speed 14 to 122°F 1-day 15-minute 100 WD-08368-20 2- or 3-speed 14 to 122°F 7-day 2-hour 100 WD-08368-30 2- or 3-speed –10 to 50°C 1-day 15-minute 100 WD-08368-40 2- or 3-speed –10 to 50°C 7-day 2-hour 100 WD-08368-22 3-speed 14 to 122°F 32-day 6-hour 25 WD-08368-42 3-speed –10 to 50°C 32-day 6-hour 25 WD-37250-60 Long cycle –20 to 50°C 1-month 2-hour 1 roll WD-37250-62 Long cycle –20 to 50°C 3-month 2-hour 1 roll
Extra Chart Paper
Extra Pens
For use with all OAKTON hygrothermographs.
WD-08368-75 Blue pens, 6/pack WD-08368-70 Red pens, 6/pack
Axial Fan
WD-37250-50 Optional axial fan. Use with 2-speed, 3-speed and long cycle models
for more uniform sample measurements in fluctuating environments. 110 VAC
WD-01578-02 Transformer. Use to operate axial fan at 220 VAC, 100 W