Nutricia Flocare 800 Short Manual

Preparation and start of the pump
Operating instructions
Turning on the pump
Programming Continuous Feeding
Programming Dose Feeding
> Turn on the pump by pressing the “POWER/ON” key. The pump is in “HOLD” mode.
> Set the desired flow rate (ml/h) by pressing the “INC” or “DEC” keys. > Flow rates can be set from 1 to 300 ml/h, in 1 ml increments. > Press the “RUN” key to start the pump.
> After programming the flow rate do not immediately press the “RUN” key. > Press the “DOSE” key to set the desired dose. > Press “INC” or “DEC” to program the correct dose (in ml). > Press the “RUN” key to start the pump.
> Press “HOLD” key, pump will stop, all memory and programming is retained.
Use this to: change flow rate, alter dose setting, change feed, administer medication. > Press “RUN” to restart the pump.
Flocare 800
Short Manual
1 2 3 4
> Insert the drip chamber of the feeding set into the drip detector slot of the pump.
> Slowly open the roller clamp on the feeding set and allow the total set to fill with feed. Make sure the drip chamber
is filled up about one-third.
> Stretch the tubing around the pump wheel and insert the plastic retainer into the bracket on the pump.
> Pull the tubing through the tubing guide.
Connect the feeding set to your tube and continue to operate the pump.
1
2
3
4
HOLD
DOSE
DEC
RUN
HOLD
DOSE
VOL FED
DEC
INC
HOLD
DOSE
DEC
RUN
HOLD
RUN
HOLD
DOSE
VOL FED
DEC
INC
HOLD
DOSE
DEC
RUN
HOLD
RUN
HOLD
HOLD
HOLD
DOSE
Viewing total volume delivered
Turning off the pump
> Press “VOL FED” key, display will indicate total volume delivered.
> Turn on the pump by pressing the “OFF” key for ca 3 seconds.
RUN
HOLD
DOSE
VOL FED
DEC
INC
RUN
HOLD
DOSE
VOL FED
ENTERAL FEEDING PUMP
RUN
RATE ML/HR
DOSE COMPLETE
OCCLUSION NO FLOW
LOW BATTERY
MALFUNCTION
AC POWER
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
171102 / BP&MC 11090802
> The pump has been left in “hold” mode
for 3.5 minutes.
> Reset the alarm by pressing the “HOLD” key, or > Continue feeding by pressing the “RUN” key.
> The feeding set is incorrectly placed. > Press the “HOLD” key, insert the plastic
retainer into the bracket on the pump, and then “RUN”.
> The pump has finished administering
the programmed dose.
> Switch the pump off by pressing “OFF” key, or > Program the pump to administer a repeat dose
by pressing “HOLD” and then “RUN”.
> The feeding container is empty.
> The liquid level in the drip chamber is too
high.
> The sensor area is dirty.
> There is an obstruction to flow,
either in the feeding set or in the tube.
> Change the feeding container or switch off
the pump.
> Remove the set from the pump; hold the set
upside-down so that feed runs back in to the container, until the chamber is ca. one-third full. Reload the set.
> Remove the drip chamber from the slot and
clean the drip detector sensors with a damp cotton swab.
> Check the feeding set, ensure that the roller
clamp is fully opened and that there are no kinks in the tubing, and/or
> Remove the set from the pump and check if
feed is flowing from top to bottom (with the roller clamp open), and/or
> Check if your tube is clogged by flushing the
tube.
> Ca. 30 minutes of battery time remain. > Connect the pump to the mains and charge
the pump. During charging the pump can be used.
> The pump needs repair. > Contact your healthcare practitioner.
HOLD
SET
DOSE COMPLETE
OCCLUSION
NO FLOW
LOW BATTERY
MALFUNCTION
(E-1 to E-11)
Problem/Alarm Likely cause Instructions
Flocare 800
Problem solving instructions
Error messages and instructions
Please refer to detailed Instructions for Use for additional information