171102 / BP&MC 11090802
> The pump has been left in “hold” mode
for 3.5 minutes.
> Reset the alarm by pressing the “HOLD” key, or
> Continue feeding by pressing the “RUN” key.
> The feeding set is incorrectly placed. > Press the “HOLD” key, insert the plastic
retainer into the bracket on the pump, and
then “RUN”.
> The pump has finished administering
the programmed dose.
> Switch the pump off by pressing “OFF” key, or
> Program the pump to administer a repeat dose
by pressing “HOLD” and then “RUN”.
> The feeding container is empty.
> The liquid level in the drip chamber is too
high.
> The sensor area is dirty.
> There is an obstruction to flow,
either in the feeding set or in the tube.
> Change the feeding container or switch off
the pump.
> Remove the set from the pump; hold the set
upside-down so that feed runs back in to the
container, until the chamber is ca. one-third
full. Reload the set.
> Remove the drip chamber from the slot and
clean the drip detector sensors with a damp
cotton swab.
> Check the feeding set, ensure that the roller
clamp is fully opened and that there are no
kinks in the tubing, and/or
> Remove the set from the pump and check if
feed is flowing from top to bottom (with the
roller clamp open), and/or
> Check if your tube is clogged by flushing the
tube.
> Ca. 30 minutes of battery time remain. > Connect the pump to the mains and charge
the pump. During charging the pump can be
used.
> The pump needs repair. > Contact your healthcare practitioner.
HOLD
SET
DOSE COMPLETE
OCCLUSION
NO FLOW
LOW BATTERY
MALFUNCTION
(E-1 to E-11)
Problem/Alarm Likely cause Instructions
Flocare 800
Problem solving
instructions
Error messages and instructions
Please refer to detailed Instructions for Use for additional information