Nutricia F-800 User manual

Your healthcare practitioner has advised you to administer your tube feeding using a Flocare® 800 (F800) enteral feeding pump. The F800 is a stationary pump that helps you in administering your feeding regime in an easy and safe manner.
Daily Care of the F800 pump
> Make sure that your pump, including the sensors and the rotor, is kept clean.
> Always unplug the pump from the electrical circuit prior to cleaning. > The battery should be maintained at full charge if possible. Keep the pump
connected to the mains whenever possible. If the battery is fully charged, the pump can operate for 3 hours (at an average speed of 125 ml/h). The “LOW BATTERY” alarm will sound when approximately 30 minutes of operation time remain. The battery will be fully reloaded after ca.12 hours of charging.
F800 Pump
Preparation of the F800 pump
HOLD
DOSE
DEC
RUN
HOLD
DOSE
VOL FED
DEC
INC
RUN
HOLD
RUN
HOLD
DOSE
VOL FED
DEC
INC
HOLD
DOSE
DEC
RUN
HOLD
RUN
HOLD
HOLD
DOSE
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
800
RUN
HOLD
DOSE
VOL FED
DEC
INC
Have your Flocare® feeding set ready for use.
Open the feeding set and connect it to the feed.
1
Insert the drip chamber of the feeding set into the drip detector slot of the pump.
2
Slowly open the roller clamp on the feeding set and allow the total set to fill with feed. Make sure the drip chamber is filled up about one-third.
3
Stretch the tubing around the pump wheel and insert the plastic retainer into the bracket on the pump.
4
Pull the tubing through the tubing guide. Connect the feeding set to your tube and continue to operate the pump.
41 2 3
Operating the F800 Pump
Programming continuous administration
> Turn on the pump by pressing the key. > Set the desired flow rate (ml/h) by pressing the key. > Press the key to start the pump.
Programming dose administration
> After programming the flow rate do not immediately press the key. > Press the key to set the desired dose. > Press or to program the correct dose (in ml). > Press the key to start the pump.
Once the desired dose has been delivered, the pump automatically stops, the alarm will sound and the “DOSE COMPLETE” indicator will be illuminated.
Loading...
+ 2 hidden pages