Nu-Tek Levator Elite LE9011 User Manual

Levator Elite LE9011
Nu-Tek® Levator Elite
Single Channel EMG, EMS & ETS
• EMG Triggered Stimulation (Passive +Active), Muscle Stimulation and EMG Modes
• One channel ETS with stimulation on one channel
• Nu-Tek scale scoring for Pelvic Floor Muscle and used as an assessment tool
• Programs for Incontinence treatment, Pelvic floor muscle development and neuromuscular Rehabilitation
• User programs for EMG/ETS (1 phase) and STIM (up to 5 stimulation phases) mode
Application:
• Promote continence
• Improve Pelvic floor Exercise
• Re-educate the Pelvic muscles
Standard Accessories:
NULEADPATRD NULEADEARTH
ACPROBEV ACF35050
Nu-Tek Patient Lead- Red Nu-Tek Earth Lead - Black
Incontinence Probe - Vaginal AllCare Electrodes - 5cm x 5cm - Square-
Self Adhesive
Specifications:
Power Supply: 1.5V AA baeries, 6V Low Voltage Warning: ≤4V±0.2V Dimensions: 139mm×68mm×33mm (L*W*H) Weight: 156g (without baeries)
EMG: One Channel: EMG Range: 0.2 to 2000μV Sensitivity: 0.1μV Accuracy: 4% of μ V reading, +/- 0.3μV at 200 Hz Work/Rest Periods: 2-99 seconds
STIM: One Channel: Stimulation Intensity: 90V±10V (V max), Adjustable from 0 to
Pulse Width: Pulse Rate: Work/Rest periods: Ramp Time (up and
down): Treatment Time: Programs
90mA (on 1000Ω) 50-450μS (2% accuracy) 2-100Hz (2% accuracy) 1-99 seconds
0.1-9.9 seconds 1-99 minutes
Pre-set: 11 Customs:3
Optional Accessories:
NUSOFTWARE ACPROBEA ACF35090
ACF350R
Nu-Tek Biofeedback Soware Incontinence Probe - Anal AllCare Electrodes - 5cm x 9cm - Rectangle
- Self Adhesive AllCare Electrodes - 5cm - Round
- Self Adhesive
Code: Description: NULEVATOR Nu-Tek® Levator Elite
physiosupplies.co.uk
REDEFINING | PHYSIOTHERAPY | FITNESS | MEDICAL
The Warehouse Beck Bank West, Pinchbeck, Spalding, Lincolnshire, PE11 3QN Phone: 01775 640972 Email: sales@physiosupplies.co.uk
Nu-Tek Levator Elite
User Manual
www.nutekmedical.com
- Page 2 -
Nu-Tek Levator Elite User Manual
Table of Contents
Contents
Warnings-------------------------------------------------------------------------- Introduction---------------------------------------------------------------------- Contra-Indication and Precautions----------------------------------------- Keypad Layout------------------------------------------------------------------- Lead / Electrode connection Assembly----------------------------------- Quick start instructions-------------------------------------------------------- Clinical Mode - Program - Phase--------------------------------------------
EMG mode operation-------------------------------------------------- Examples of EMG treatment----------------------------------------- EMG Parameters-------------------------------------------------------- EMG and ETS Threshold---------------------------------------------- EMG in patient mode ------------------------------------------------- EMG measurement----------------------------------------------------- Conditions respond to EMG----------------------------------------- STIM mode operation------------------------------------------------- STIM Parameters-------------------------------------------------------- ETS mode operation--------------------------------------------------- ETS Parameters---------------------------------------------------------
Data Report for progress----------------------------------------------------- System setting---------------------------------------------------------------- Statistics-------------------------------------------------------------------------- More about EMG statistics---------------------------------------------------- Electrodes Types and Tips---------------------------------------------------- Care, Maintenance, Accessories and Disposal-------------------------- Specications-------------------------------------------------------------------- Information regarding Electromagnetic compatibility and interference (EMC)------------------------------------ Trouble shooting--------------------------------------------------------------- Warranty-------------------------------------------------------------------------- Appendix-program------------------------------------------------------------
Page
3 4 6 7 8 9
13
14 15 17 18 19 20 21 22 23 24 25
26 27 29 31 32 33 35
37 42 45
Issue Date: 06/05/2016 Document Number: LE9011-Rev. V1.1 2016
c
46
- Page 3 -
Nu-Tek Levator Elite User Manual
Warnings
This device must be used with the guidance of a Physiotherapist or Doctor. Type BF equipment, Continuous Operation. Do not immerse device into water or any other substance. Do not use the device in the presence of a ammable anaesthetic gas
mixture and air or with Oxygen or Nitrous Oxide. This device uses 4 x AA Batteries. If using rechargeable Nickel Metal Hydride batteries, be sure to use a CE approved battery charger. Never connect it directly to a battery charger or to any other mains powered equipment. To avoid the eects of electromagnetic interference, never use the device in the EMG mode, within 4 metres of a mobile telephone or near any other powerful radio interference producing equipment that causes electrical sparks etc. In the EMG mode, the device may be susceptible to strong interfering radio type emissions that may lead to temporally increased EMG microvolt readings. The reading will immediately return to the correct value when the interference ceases. (Remember that a relaxed muscle should read below 4μV). Patient electrodes including all skin surface electrodes, vaginal and rectal probe are for single patient use only! Do not use stimulation on your facial area unless you are under strict guidance from a qualied clinician. Application of electrodes near the thorax may increase the risk of cardiac brillation. Operation in close proximity (e.g. 1m) to a shortwave or microwave therapy equipment may produce instability in the stimulator output. Simultaneous connection of a patient to a high frequency surgical equipment may result in burns at the site of the stimulator electrodes and possible damage to the stimulator.
No modication of this equipment is allowed! Keep the device out of reach of children Skin irritation from the electrode gel and electrode burns are potential
adverse reactions. If skin irritation occurs, discontinue use and consult your physician. Do not apply stimulation over the patient's neck because this could cause severe muscle spasms resulting in closure of the airway, diculty in breathing, or adverse eects on heart rhythm or blood pressure. Do not apply stimulation in the presence of electronic monitoring equipment (e.g., cardiac monitors, ECG alarms), which may not operate properly when the electrical stimulation device is in use.
Device name: Nu-Tek Levator Elite Model number: LE9011
- Page 4 -
Nu-Tek Levator Elite User Manual
Introduction
The Nu-Tek Levator Elite combined EMG [electromyography] and Neuromuscular Stimulation is a simple to use advanced product for reducing incontinence in females with urinary incontinence. The device has been developed to enhance and support the Clinician, to assist the end user at home to learn more about their Pelvic muscle and above all to improve their Pelvic Muscle strength and their incontinence therapy. The Nu-Tek Levator Elite: EMG Biofeedback measures the Pelvic contraction, and assists users to maximise Pelvic oor exercises, after periodic training, it also helps to assess the pelvic oor muscle condition, quantify the need for further treatment and evaluate report on the patients progress. Muscular Stimulation improves blood circulation, capillary bed density and strengthens the Pelvic oor muscles. EMG triggered stimulation facilitate those with accid muscles; ETS helps reduce the symptoms in Genuine Stress incontinence in some patients and is being used more frequently by clinicians for this condition
For the end user, in patient mode, a simple and easy to understand EMG Biofeedback bar graph displays the Nu-Tek pelvic muscle strength grading on a scale of 1 to 6, and assists the user to meet their pre-set targets. In the therapy mode, the EMG graph can be changed to a waveform format, as viewed on the LCD screen of the device or using the Nu-Tek software link on a desk top or laptop computer . The waveform can be used to help analyse the patients condition. Periodically the doctor or therapist will be able to download newly developed programs for Muscle Stimulation, EMG Biofeedback and EMG- TRIGGERED STIMULATION. This helps this unique product to keep abreast of any clinical trials where the conclusion is that a specic program or program set may assist in enhancing the treatment of continence. The Nu-Tek Levator Elite has 3 clinical modes: EMG, ETS, and STIM. Each of these modes has custom programs. The device also includes pre-set EMG and STIM programs: there are 20 pre-set Pelvic muscle stimulation programs: Genuine Stress, Urge, Frequency, weekly maintenance. Lack of sensation, rectal stimulation and several other Patient conditions. One pre-set EMG program for assessment. The custom program has a diverse range of parameters that can be programmed by the patient or therapist to meet individual user conditions. Any one of the pre-set or custom programs can also be selected freely by the user. The essential performance of the device are free from the production of unwanted or excessive stimulation output and free from the display of incorrect numerical values associated with measure to be performed.
- Page 5 -
Nu-Tek Levator Elite User Manual
Customer Care
We welcome constructive comments regarding our equipment particularly those that might help us to improve existing features, add new ones and / or develop new products for the future.
- Page 6 -
Nu-Tek Levator Elite User Manual
Contra-Indications and Precautions
STIM: Neuromuscular Stimulation (NMS)
Before using this device you must rst seek the advice of your doctor or therapist.
Neuromuscular Stimulation should not be used by:
Patients tted with demand style cardiac pacemakers During pregnancy (unless medically advised) Patients with undiagnosed pain conditions Do not place electrodes:
- Over carotid sinus nerves
- Over larynx or trachea
- Inside mouth
- On anaesthetised or desensitised skin
- Do not drive a vehicle while the device is stimulating and attached to your body Skin irritation from the treatment of NMS or EMG itself does not generally occur. However, rubber electrodes may irritate some skin types, therefore; in this case we recommend using hypoallergenic self adhesive electrodes.
The patient should only use the device for what it was prescribed for Do not immerse the device in water or any other liquid substance Do not use stimulation on your facial area unless you are under strict
guidance from a qualied clinician
EMG
There are no precautions when using EMG unless used for pelvic oor exercising or assessment. In this case EMG should not be used:
During menstrual period Inammation or infection in the vaginal area or urinary tract. With patients who have diminished mental capacity or physical
competence who cannot handle the device properly.
The absence of sensation due to denervation of the pelvic oor. In children - when using internal probe.
- Page 7 -
Nu-Tek Levator Elite User Manual
Keypad Layout
1
3
2
4
6
5
1. ON/OFF: press to power on or switch o the device.
2. ESC: press to nish the session (program) or the settings, or return to previous menu.
3. THRS+ and THRS-: adjust the EMG threshold level (ETS target) and adjust the other parameters and settings.
4. mA+ and mA -: start the STIM or ETS phase, increase or decrease the stimulation intensity.
5. OK: press to complete selection
6. , , , : select mode in main menu, or switch parameter and settings.
- Page 8 -
Nu-Tek Levator Elite User Manual
Lead / Electrode Connection Assembly
USB cable: Connection to the PC REF: Reference wire (REF) for precise EMG measurement CH: dual conductor lead wire for STIM or EMG
Note: The REF EMG lead / electrode is only required for EMG and ETS and NOT for Neuromuscular stimulation. The lead wires, electrodes, vaginal probe and REF wires are supplied as part of the kit.
E
V
- Page 9 -
Nu-Tek Levator Elite User Manual
Quick Start Instructions
1. Insert four AA batteries
Remove battery cover. Insert batteries as labelled inside the battery compartment and then replace the back cover.
2. Insert lead wire
Insert the lead wire/s into the sockets of the device. The round black EMG reference lead wire connects to the round black socket in the top of the unit; the red stimulation lead wire connects to the red socket. IMPORTANT! If you dont use the round EMG reference wire (REF), your EMG and ETS results will be inaccurate.
3. Skin electrodes/probe placement
When using the EMG or ETS phase, connect the EMG reference lead wire to a surface skin electrode and place it appropriately on the body, making sure the skin is free from grease and dirt; repeat the above procedure with the other two skin electrodes. If using a probe, place the single surface skin electrode on the thigh area and then insert the probe.
4. Turn on Levator Elite by pressing the ON/OFF button once for 3 seconds. Note: This product has a Patient Mode which we suggest the general
public use and a Therapy Mode which is more practical for the Clinician. Selecting the Therapy Mode requires a password. When selected the Patient Mode, the product would be locked. The details setting method refer to Change using mode of page 27. In Patient Mode, you can not select the therapy program and set the parameters of the custom program.
5. The EMG, the Patient Mode uses the Nu Tek bar graph; the Therapy Mode uses a line graph.
6. There are pre-set (ready to use) programs and custom programs (which can be adjusted by the patient or therapist in Therapy Mode). The parameters of pre-set programs can not be changed. Note: Individual custom programs set for the patient should ideally be under the guidance of a doctor or therapist in Therapy Mode. Using the circular button black arrow keys navigate to select EMG and press the OK button. The screen displays Training Program and Assessment Program. In Assessment Program, there is only one preset program: Threshold = 30μV, WIDE Filter, ABOVE Biofeedback sound, Work time = 5s, Rest time = 5s, and 5 Trials, AUTO threshold is advised to be used to see the treatment progress during periodic training. In the Training program, there are two custom programs, you can select all the EMG Biofeedback parameters by using the up and down black arrow keys, and change settings by using the THRS+ and THRS- buttons. In Patient Mode, the program, work time, rest time,trial and A/M threshold can not be adjust.
- Page 10 -
Nu-Tek Levator Elite User Manual
When you have completed your settings, press the OK button which will take you into the vertical Bar graph Patient Mode or the EMG line graph Therapy
Mode, then press the OK button to begin the session.
Working with EMG
Always use the Reference wire (REF) for more accurate EMG measurement! Place the reference electrode anywhere on your skin. When using the device with a vaginal probe, place the reference electrode on your thigh.
Place the device on the desk, in its stand or hold the device. Relax so that the microvolt reading is as low as you can manage; below
6µV is acceptable while below 4µV is ideal.
ETS (EMG TRIGGERED STIMULATION) Patient / Therapy MODE
Using the circular button select ETS, press the OK button to enter the ETS parameter settings interface. Using the circular button scroll and using the THRS+ and THRS- buttons to change the EMG Biofeedback and Stimulation settings. In Patient Mode, the program, work time, rest time,trial, A/M threshold and stimulation parameters can not be adjust. When you have completed your settings, press the OK button this will then take you into the standby mode. Press the mA+ button to increase the electrical current ( mA) and to start the treatment. As soon as the patient reaches the target level (threshold) in the work period stimulation takes place for several seconds which helps to contract the pelvic muscles. Ideally contract your pelvic muscle along with the electrical stimulation.
STIM (STIMULATION) MODE
Using the circular button select STIM, press the OK button which brings up the pre-set and custom programs, in Therapy Mode press the THRS+ or THRS- buttons to select the required program. Press the OK button, and then press the mA+ button to increase the electrical current (mA) and to start the treatment. The parameters of pre-set programs can not be changed; those of custom programs can be adjusted by the therapist in Therapy Mode. Using the THRS+ and THRS- the phase parameters of the program can be viewed. At the end of the session the screen will display the Date and Time, Average Current used (mA), Program, and the Frequency (Hz) used. Only the mA (current) will be saved.
COMMUNICATE
Connect the device to the PC via a USB cable, then using the circular black arrow keys select COMMUNICATE, press the OK button, the data can be transferred between PC and the device. (This function is an optional extra
- Page 11 -
Nu-Tek Levator Elite User Manual
and the user will need to purchase the Nu-Tek System, the USB connection cable to the PC).
DATA MANAGE
Using the circular button to select DATA MANAGE, press the OK button, then using or select Data query , Delete data or Send data to pc, then press the OK button. The information is stored and displayed on the two linear graphs on the LCD screen. The two vertical graphs are combined by up to 90 records respectively, and the history of the users sessions can be viewed. For training data, the device saves the data on a daily basis to form statistics. In the Assessment data, the device stores the weekly statistics by carrying out an assessment of the programs once or twice per week. The device can store more than one patients statistics, but it can not distinguish between dierent patients. Only when the device connected via the Nu- Tek PC software, can the recorded data in PC be distinguished between dierent patients. The user can delete the treatment data when selected Delete Data and press OK button.
SYSTEM SETTING
Using the circular button select SYSTEM SETTING, press the OK button which will bring up the various options, including Date and Time, Backlight Setting,
Sound Setting, Set Language (English, German, French, Italian, Spanish), Change Using Mode, Factory Data Reset. Using the circular button scroll
down to the settings you want to change, press the OK button, then using the THRS+ and THRS- buttons to set the various options. Use or to select Save or Cancel, press the OK button. If you select Factory Data Reset, all the values are restored to the original factory setting. (Note: If the factory setting is restored all the patients statistics will be lost)
PROGRAM MANAGE
Using the circular button select PROGRAM MANAGE, press the OK button. Then using or select Downloaded Programs (There are three modes: EMG, ETS, and STIM. For either mode, there are up to a maximum of 3 programs) or To Download Programs, then press the OK button. For example, select Downloaded Programs, to view the downloaded program, press the OK button, to use the program. Only the last selected downloaded program in Therapy Mode is available for display and for use in the Patient Mode. Select to download Programs by connecting the device to the PC via the USB cable. If three programs have been downloaded the fourth program will not be possible, at this point the screen displays a prompt for the user to choose which one program to select to make up the three programs.
- Page 12 -
Nu-Tek Levator Elite User Manual
T
PROGRAM
EMG THERAPY MODE
PARAMETER MANAGE
Using the circular button select PARAMETER MANAGE, press the OK button to bring up the details.
OPERATING MANAGE
Using the circular button select OPERATING MANAGE, press OK button to bring up the details.
7. After completing the settings, Press the ESC button to exit.
8. When you have nished, press and hold the ON/OFF button for 3 seconds to turn o the device. Remove and replace the skin electrodes onto the clear plastic lm, reseal them in the plastic zip bag and store them in a cool place. If using a vaginal or rectal probe thoroughly clean the probe and seal it in plastic zip bag.
- Page 13 -
Nu-Tek Levator Elite User Manual
Clinical Mode - Program - Phase
The Nu-Tek Levator Elite enables the patient or therapist to select a pre-set program or a custom program to make the appropriate treatment. Clinical Mode: Using the circular button select either EMG, STIM or ETS and press the OK button to enter the mode. In EMG mode, there are two functions: assessment and training. For assessment, only one pre-set program is used, the user can view the treatment progress.
The device has two custom EMG/ETS programs. You can set up one EMG/ ETS phase with your chosen parameters.
In STIM mode, Levator Elite has 20 pre-set Pelvic floor muscle stimulation programs covering Genuine Stress, Urge, Frequency, weekly maintenance, Lack of sensation, Rectal stimulation and several other patient conditions. There are 3 custom programs that can be programmed to meet individual user conditions and have a diverse range of parameters to meet any current stimulation requirements. The programs can be divided into a maximum of 5 sequential phases. Phase Time is the time remaining for the Current Phase. The overall time is a combined time of all the program phases. Overall time always indicates the remaining time of the program session.
A
e
T
- Page 14 -
Nu-Tek Levator Elite User Manual
EMG mode operation
1. Using the circular button black arrow keys navigate to the EMG and press the OK button, the screen will then display two modes: Assessment Program or Training Program. Select mode and press the OK button to enter the parameter mode.
2. After completing the settings, press the OK button to enter standby mode of the vertical Bar graph (Patient Mode) or the EMG line graph
(Therapy Mode).
3. Press the OK button again to begin the session.
4. At the end of the session the screen will display the work/rest microvolt, onset/relax time
5. There are two EMG custom programs for training purposes, all the parameters can be adjusted by the patient or therapist in Therapy Mode.
6. If Assessment Program is selected, there is only one pre-set program used, the user can view their progress on the LCD screen
Patient mode Therapy mode
Time:
Work Time: 04/05 Trial: 05/05
Mode:
:
T
Loading...
+ 33 hidden pages