SPECIFICATION:
1. Bettery: DC 1.55/1.5v (SR41/LR41)
2. Power Consumption: 0.2mW
3. Battery Life: More than 200 hours of continuous operation.
(Approx. 2 years if used 10 minutes per day)
4. Display Unit: 0.1°C/ 0.1°F
6. Accuracy: ±0.1°C(±0.2°F) between 35°C~38°C (95°F~100.4°F), ±0.2°C(±0.4°F) for
remainder of range at room temperature of 22°C (71.6°F)
7. Dimensions: 40mm x 43mm x 53mm
8. Weight: Approx. 14g (including battery)
9. Nipple Size: Medium
CAUTION:
1. Any pacifier wears out. To PREVENT POSSIBLE CHOKING HAZARD, test the strength
of the nipple by pulling on the bulb portion before each use.
2. Examine the pacifier bulb before each use for cracks, tears, tackiness or other signs of
wear. Discard the worn unit immediately.
3. THE UNIT IS A MEDICAL DEVICE, do not substitute or use as a regular pacifier. Store
out of reach of children.
4. Do not contact the unit with any thinner or chemical solvent.
5. Do not store the unit in direct sunlight or at high temperature.
6. Do not attempt to disassemble the unit.
7. Do not let children play with the unit. TEMPERATURE SHOULD BE TAKEN UNDER
ADULT SUPERVISION.
8. When the battery is almost exhausted, the ( ) symbol will appear in the display window,
meaning it's time to discard the unit. Do not use as a regular pacifier.
9. ORAL TEMPERATURE READINGS ARE OFTEN (0.5°C) 1.0 °F LOWER THAN ACTUAL
BODY TEMPERATURE READINGS.
For use up to age 5 years.
A BABY’S THERMOMETER.
MP9100
Digital Pacifier Thermometer
32.0°C~42.9°C (90.0°F ~109.9°F).
Less than 32°C/ 90°F, Lo°C / Lo°F Display
More than 42.9°C/ 109.9°F, Hi°C / Hi°F Display
Power button
display window
10. Do not use this thermometer for measurements other than for oral use.
11. Components may be harmful if swallowed. Contact your physician immediately should
this occur.
12. The Thermometer nipple comes in one size intended to fit all mouth sizes. Please gauge
your child's mouth size and the nipple size prior to thermometer use. A child using a
nipple that is too large has an increase risk of choking.
HOW TO OPERATE:
1. Press the on-off button to turn on the thermometer, the window will display 188.8M°E and
a beep sound will be heard. This indicates that the unit is in working condition.
2. Upon releasing the on-off button, last measured temperature will be showed for about two
seconds with letter “M” under the degree sign, then 37.0°C/98.6°F, and then Lo°C/Lo°F
displayed (if the room temperature is greater than 32°C/ 90°F, it will be displayed rather
than Lo°C/ Lo°F). The flash °C/°F indicates the thermometer is ready to take a tempera-
ture.
3. Position the pacifier bulb into babies' mouth. Note: Drinking or eating prior to use may
affect reading. Measurement should be taken one half hour after eating or drinking.
4. Once the degree sign °C/°F on the display has stopped flashing (approx. 3 minutes) and
sound an alarm (Bi—Bi—Bi, if temperature exceeds 37.8°C/100°F it sounds Bi-Bi-Bi
---Bi-Bi-Bi), representing the temperature is ready to read. The reading is reflecting the
oral temperature of the baby.
5. It is recommended to leave the thermometer in baby's mouth for 6~8 minutes to allow the
nipple to warm up and get a more accurate reading.
6. The unit will automatically turn off in 10 minutes. However, to prolong battery life, it is best
to press the power button to turn it off once the temperature has been taken.
7. Repeat operation procedures if interrupted during use.
CARE INSTRUCTIONS:
1. Before and after each use, clean the unit with warm water and sterilize the pacifier bulb
with ethyl alcohol.
2. Do not boil or wash in dishwasher.
3. Wipe clean with dry soft cloth before storage.
4. Use the original cap to store the unit. Store out of reach of children.
WARNING:
DO NOT TIE PACIFIER AROUND CHILD'S NECK AS PRESENTS A STRANGULATION
DANGER.
Loading...