DHD Direct Heat
Invitrocell CO2 Incubator
Models: NU-5810, NU-5820, NU-5831, NU-5841
NU-5810E, NU-5820E, NU-5831E, NU-5841E
Operation and Maintenance Manual
1.0 General Description
The NuAire Direct Heat Decon (DHD) Automatic CO2 Direct Heat Incubator has been designed to provide a reliable
controlled
In-vitro environment for optimum tissue cell culture growth. The chamber also provides an environment for the storage
and preservation of gametes and animal tissue cell cultures intended for research at or near body temperature.
There are five parameters that contribute to optimum growth conditions, they are:
Humidity
Precise temperature control
Precise CO
Sterility
Reliability
Like all NuAire equipment, this Incubator has been designed to provide the highest quality standards of performance with
matching computer technology, precise temperature control, and CO
technology with years of design, quality, and manufacturing experience.
1.1 Incubator Chamber
The design and size of the DHD inner chamber provides a large capacity (5.7 cu. ft.), and ease of use. The chamber walls are
directly heated by physically attached foil heating elements on the sides, bottom, top, and back of the chamber, providing
a temperature uniformity of +0.3°C. A high-density insulation that has a high "R" rating covers the complete outer surfaces
of the Incubator inner chamber.
A continuous operating air pump draws environmental sample from the chamber and circulates it through a HEPA filter
capsule then through sensors that measure CO
control
2
gas control system combining state-of-the-art
2
volume (as a percentage) in the chamber for monitoring and control.
2
1.3 Incubator Control Electronics
The NuAire Incubator Control Electronics is a state-of-the-art microcomputer based control system specifically designed to
service the precise control requirements of the chambers environment, providing optimum programmable conditions for
culture growth. The microcomputer is “user-friendly” and provides a 5x7 LCD screen which provides operating control
parameters, status indicators, additional key operational parameters, and an imbedded touch panel to permit efficient
operator entry of data.
The microcomputer is supported with Read Only Memory (ROM) containing executable software, Random Access Memory
(RAM) for temporary storage, and Electronically Erasable Programmable Read Only Memory (EEPROM) for control set
points and parameters. The EEPROM provides for indefinite storage of these values during periods of power off or power
interruption (power fault tolerant).
1.4 Incubator CO
The NuAire direct heat Incubator incorporates a microprocessor-based, non-dispersive infrared CO
The amount of energy received at the detector is an approximate logarithmic function of the CO
between source and detector. The wavelengths used are absorbed only by CO
Control
2
induction sensor.
2
concentration in the gas
2
making the measurement insensitive to
2
other components, such as water vapor. Detector linearization is performed with 32 bit digital accuracy. Advanced design
provides a very stable output minimizing drift and requiring less frequent calibration. The output is digital, alleviating errors
brought about by analog signals. Calibration of this control is accomplished by a zero & span calibration done on the sensor
and a display off- set calibration. These calibrations are accessible through the NuTouch LCD located on the front of the
unit.
OM0246 Page 3 of 64
Rev 5 June/2020
Page 4
1.5 Incubator Construction
The outer shell of the air-jacketed Incubator is cold-rolled steel with a powder coat paint finish. The front frame surface of
the outer shell is heated with a foil type heater directly attached behind the front perimeter opening.
The front frame perimeter heater, as well as the outer front door heater is duty cycle controlled (manually adjusted for
specific ambient conditions) to balance the heat that reaches the chamber and thus reduces the possibility of condensation
forming on the inner glass door and the inner chamber walls.
The inner chamber is 16 gauge, type 304L, polished stainless steel using crevice-free construction, which provides an easily
cleanable inert surface that does not in itself promote biological growth. In addition, all shelves, shelf supports, and guide
rails are easily removable and can be autoclaved separately if so desired.
Remember: The chamber environment is not selective. The growth environment is applied equally to all microorganisms
(specimens and contaminates) within the chamber.
1.6 Incubator Humidity
Models NU-5810 and NU-5831
A relative humidity level of up to 95% is achieved in the Incubator by the use of a stainless steel pan filled with distilled
water no purer than 1 mega ohm, and placed on the bottom of the chamber. It is necessary to set the duty cycle of the
door and front perimeter heater in proper proportions to reduce the possibility of condensation forming on the glass inner
door and the chamber walls. It is also important to thoroughly wipe the walls and the glass door clean before adding the
humidity water pan. Condensation will occur more readily at contamination points. There is no electronic sensing and thus
no automatic control of the humidity level in the air jacketed Incubator. An air pump injects fresh air into the chamber at a
preset duty cycle to reduce condensation forming on the chamber walls and front glass door. The air injections are settable
by the user to help control condensation. (See section 8.8 for this option.)
Models NU-5820 and NU-5841 use a humidity monitoring/control system for chamber humidity; the system is external to
the chamber. The chamber air is routed through the water vapor saturated air in an evaporator box to add humidity when
the RH sensor indicates that more humidity is required maintain a set point.
1.7 Cabinet Ventilator Fan
Two axial fans are mounted to the bottom cover panel of the Incubator and runs continuously when the unit is switched on.
These fans pull air into the cabinet shell thru the top panel louver openings and discharges it out of the bottom of the unit.
The operation of this fan is necessary to assure accurate chamber temperature control over the 60°F to 85°F (15°C to 30°C)
ambient temperature range that the equipment may see.
Safety Instructions
1.8
These safety instructions describe the safety features of the INVITROCELL Incubator.
The incubator has been manufactured using the latest technological developments and has been thoroughly tested before
delivery. It may, however, present potential hazards if it is not used according to the intended purpose or outside of
operating parameters. Therefore, the following procedures must always be observed:
•
•
The incubator must be operated only by trained and authorized personnel.
For any operation of this unit, the operator must prepare clear and concise written instructions for operating and
cleaning, utilizing applicable safety data sheets, plant hygiene guidelines, and technical regulations, in particular.
o Which decontamination measures are to be applied for the incubator and accessories?
Which protective measures apply while specific agents are used?
o
Which measures are to be taken in the case of an accident?
o
•
Repairs to the device must be carried out only by trained and authorized expert personnel.
•
Keep these operating instructions close to the unit so that safety instructions and important information are
always accessible.
•
Should you encounter problems that are not detailed adequately in the operating instructions, please contact your
NuAire Representative or NuAire technical Services.
OM0246 Page 4 of 64
Rev 5 June/2020
Page 5
1.9 Explanation of symbols
Safety alert symbol indicates a potentially hazardous
CAUTION
!
Safety alert symbol indicates a potentially hazardous
WARNING
!
Note:
Biohazard
Hazardous Gases!
Chemical Hazard
Flammable Hazard
Lead Free
CAUTION used without the safety alert symbol
avoided, may result in property damage.
Potential electrical hazard,
only qualified person to access.
Ground, Earth
Symbol (in text box) Description (in text box)
situation which, if not avoided, could result in death of
serious injury.
situation which, if not avoided, may result in minor or
moderate injury.
CAUTION
indicates a potentially hazardous situation which, if not
Personal Protection Equipment Required.
OM0246 Page 5 of 64
Rev 5 June/2020
Hot Surface Burn Potential
Page 6
2.0 Performance Parameters and Features
• Both the interior and exterior of the DHD are constructed of 18 to 16-gauge steel.
The interior is highly polished type 304 L 16 gage stainless steel, using crevice-free construction.
All exposed edges are deburred to insure no sharp edges.
The exterior is cold rolled 16 & 18 gauge steel finished in a powder coated polyurethane finish, which is resistant to
chemicals and easily cleaned using mild household detergents.
• The DHD's microcomputer temperature control system has two temperature sensors located inside the chamber at the
back wall. The temperature sensors compare the values to a set point and execute a differential control algorithm that
energizes a solid-state switch, supplying power to the heaters.
• Easily removable inner shelves and rails for sterilization.
• Up to 16 shelves can be placed inside the chamber (4 shelves are shipped with each DHD).
• Space-age high temperature material is used to insulate the inner chamber walls.
• Foil heaters are directly attached to the chamber walls and the top and bottom surfaces of the inner chamber.
• Most electronics, motors, pumps, and valves are fully accessible from the back of the unit.
• Access to the filters and CO₂ sensor through the front of the incubator.
• A CO
• The CO
• Automatic recovery of the CO
• The outer door includes an internal radiant heater in order to minimize condensation on the inner glass door.
• The inner glass door is 3/16” tempered with smooth-ground edges and seals are tight against a silicone rubber gasket.
• All control electronics are protected with a circuit breaker that may trip at 110% of loading rating but will trip at 145%
• The Incubator has factory-installed adjustable leveling legs to compensate for uneven laboratory surfaces.
sample port is provided on the front panel just below the door to check the concentration of CO2 in the chamber.
2
percentage is controlled by a solid-state gas infrared sensor using a single light source with dual wave length
2
detection, providing accurate monitoring of CO
level, after a 1 minute door opening, to 5.0 +.02/- 0.5% is in 5 minutes as average.
2
A magnetic outer door gasket helps to insure a tight seal against the cabinet.
The door latch is cam action.
A momentary switch monitors the door position, the position being opened or closed.
of load rating in less than 2 seconds.
Should the circuit breaker open (pop-out button will appear), merely depress to reset.
.
2
OM0246 Page 6 of 64
Rev 5 June/2020
Page 7
3.0 Models & Features
3.1 Invitrocell DHD (heated decontamination) all Models.
3.2 Dimensions (see also Specification Drawing BCD-18314)
Overall Dimensions - inches (mm): Model NU-5810/E thru NU-5841/E
Height: Exterior: 39.688 in (1008 mm)
Width: 26.750 in (680 mm)
Depth: 27.188 in (691 mm)
Foot print: 23.5 in (597 mm) wide x 20.75 in (527 mm) deep
Height: Interior: 28.500 in (723.900mm)
Width: 21.257 in (539.928mm)
Depth: 20.053 in (509.346mm)
Shelf Capacity:
Size:19.75 Inches (502mm) x 18.75 Inches (476mm)
Supplied: 4 Shelves
Maximum Capacity: 20 Shelves
Maximum Weight Capacity 25 lbs. per DHD shelf
125 lbs. per DHD Incubator
(Make sure self is empty before changing position or removing.)
Water Pan:
Dimensions: Mean Length 12.0 inches (305mm)
Mean Width 10.0 inches (254mm)
Depth 1.5 inches (38mm)
Capacity: Maximum 2.0 Liters
Recommended Fill 1.5 Liters
3.3 Performance Parameters :
Temperature Control: Available on All Models Performance Determined with O₂ Set Point @ 21.0%
• Temperature Range: 5°C to 55°C (37°C default) (5°C above ambient to 30°C maximum ambient)
• Temperature Uniformity: +0.3°C @ 37°C.
• Temperature Accuracy: +0.1°C.
• Temperature Recovery: 0.12°C/minute Average
• Temperature Display Resolution: 0.1°C
• Temperature Sensor Type: Precision Integrated Circuit
•Door and Perimeter: Proportional base duty cycle based on Temperature set point and -20 to +20% manually adjustable to adapt to ambient conditions.
Control: Available on All Models Performance Determined with O₂ Set Point @ 21.0%
CO
2
• CO
• CO2 Accuracy: +0.1%
• CO2 Recovery: Up to 5% +0.2/-0.5 in 5 minutes average.
• CO
• CO
• CO
• Sensor Bay Temp Chamber Set Point +3.5 Degree C.
Range: 0.1 to 20% (default 5%). (0.0 set point idles system)
2
Display Resolution: 0.1%
2
Control Logic: Fixed Algorithm/Manual Environmental Adaptable.
2
Sensor Type: Infrared single source dual wave length
2
OM0246 Page 7 of 64
Rev 5 June/2020
Page 8
Relative Humidity Control: Available on NU-5820 and NU-5841
Performance Determined with O₂ Set Point @ 21.0%
• RH Range: Low is ambient dependent to 95% maximum (default 90%)
(0.0 set point idles the system)
• RH Accuracy: +5% / -3%
• RH Recovery: 90% ±5% : 25 minutes on Average
• RH Water tank capacity: 3 Liters (approximate)
• RH Evaporator tank temperature: Chamber temperature Set point +10⁰ C
O
control and monitoring system w/ the Fujikura Zirconia sensor Available on the NU-5831 & NU-5841
2
•O
measuring range: 0 to 25 %
2
• Default Set-Point: 21% (system is idle at this set point)
• O
set- point Range: .5 to 21% (2 to 21% when RH setpoint is above 90%)
2
• O2 Accuracy: +.25%
• O2 Recovery 5% +2%: 15 minutes on average
Note: Performance ratings apply ONLY when the Incubator is installed and set up properly per the
instructions in the manual, has stabilized at all set points and is in a stable environment.
See sections 8 and 9 for proper set up.
While idle a system will not inject gas or RH and will not alarm.
3.4 Standard Items Packed With Unit
• Four (4) stainless steel shelves
• Eight (8) stainless steel shelf brackets
• One (1) water pan
(Note: Water pan is included with units having RH Control System for calibration purposes only. See section 9.5.1)
• Gas supply tube(s)
• Access port plug with breather hole
• One (1), 2 meter (6.5 ft.) electrical cord 10 Amp 18awg with 105⁰C rated jacket *
• Operation and Maintenance Manual
• Operating Instructions
*Do not replace electrical cord with an inadequately rated cord.
A factory supplied replacement is recommended.
3.5 Consult factory for additional Accessories (Ordered Separately)
OM0246 Page 8 of 64
Rev 5 June/2020
Page 9
OM0246 Page 9 of 64
Rev 5 June/2020
Page 10
4.0 Test Performance & Procedures
All equipment is thoroughly inspected at the NuAire Factory at the time of shipment. Quality control is maintained by
constant surveillance over the product, beginning at the receipt of purchased material and concluding with a final
inspection, which certifies the Incubator performance to the specifications. In all instances where product quality cannot
be easily assessed on the end item, the product is inspected during sub-assembly fabrication. The following test procedures
are conducted on every DHD Incubator and a copy of the test report is included with each Incubator.
4.1 Visual Inspection
•Each Incubator is visually inspected to insure that the interior is clean and free from scratches, nicks, and burrs
and that all welds, both interior and exterior are finished.
• Painted surfaces are inspected to be free of scratches, nicks, insufficient covering, and noticeable paint runs.
• The doors open and close freely without binding of the hinges.
• The gasket seals the inner glass door to the Incubator tightly across its entire opening.
• The glass door is free of scratches.
4.2 Electrical Tests
•Electrical Leakage Test
All Incubators are required to have a primary-circuit filtering to meet EMC (electromagnetic compatibility)
regulations. Electrical leakage does not exceed 1.0 milliamperes.
•Dielectric Voltage – Withstand Test
1200 VDC (115 volt units) or 2000 VDC (230 volt units) is applied between dead metal parts and the hot/neutral
power source lead with no electrical breakdown using an Associated Research Model 7564SA Quadchek II, 8106
Omnia 6, or 8006 Omnia 6, or other models.
•Grounding Continuity
The resistance between the green bonding conductor of the supply cord and any dead metal part of the cabinet
shall not exceed 0.10 ohms.
4.3 Functional Tests
These functional tests are performed on every Incubator over a minimum of a 48-hour /burn-in period. The results are
recorded on the final Inspection report and a copy is included with each incubator that is shipped. All failures are noted,
corrected, and any failed test is successfully repeated. Changes from any default value of the optional configuration
parameters meant to improve performance are recorded on this report for your reference.
• Control Systems
All performance and diagnostic functions are exercised to insure proper operation of control systems,
components, and alarms.
• Temperature
Each Incubator is monitored for stable temperature control over the 48-hour burn-in period at (37°C).
• CO
• CO
• Humidity
Control
2
Each Incubator is calibrated to function at a 5% CO
The concentration is checked with an independent calibrated instrument.
Each unit is monitored during the 48-hour burn-in period and only accepted with zero failures.
Recovery
2
Each Incubator is exercised for CO
The door is opened for 1 minute to deplete the CO
After the door is closed, the unit shall recover to 5% +0.2/- 0.5% within a 5 minute period.
Each unit has a default RH level intended to avoid condensation.
Door and perimeter duty cycles are determined by a Base duty cycle calculation based on the chamber
Temperature set point. Automatically adjusted in response to environmental conditions but can be manually
adjusted if needed to control condensation and tune in the chamber temperature uniformity for the environment
that the incubator is installed in.
If unit is master reset these settings will be changed to a default value of 0.0.
The factory default door and perimeter settings are: Door: 0.0, Perimeter: 0.0.
level.
2
recovery time at the end of the 48-hour burn in period.
2
.
2
OM0246 Page 10 of 64
Rev 5 June/2020
Page 11
5.0 Warranty
Details regarding product warranties can be found in the published warranty data separate from this manual and included
within the data packet send with the unit.
6.0 Shipments
NuAire, Inc. takes every reasonable precaution to assure that your Incubator arrives without damage. Motor carriers are
carefully selected and shipping cartons have been specifically designed to insure your purchase arrives safely and
undamaged. However, damage can occur in any shipment and the following paragraph outlines those steps you should
take on receipt of a NuAire Incubator to be sure that if damage has occurred, the proper claims and actions are taken
immediately.
6.1 Damaged Shipments
6.1.1 Terms are F.O.B. factory, unless stated otherwise.
Therefore, it is important to check each shipment before acceptance.
6.1.2 If there is visible damage, the material can be accepted after the driver makes a notation on the consignee's copy of
the freight bill. Then an inspection must be made to verify the claim against the carrier. This inspection is the basis
of your filing the claim against the carrier.
6.1.3 If concealed damage is found, it is absolutely necessary to NOTIFY THE FREIGHT AGENT AT ONCE, and request an
inspection. Without this inspection, the transportation company may not accept a claim for loss or damage. If the
carrier will not perform the inspection, an affidavit must be prepared stating that he was contacted on a certain date
and that he failed to comply with the request. This, along with other papers in the customer's possession will
support the claim.
OM0246 Page 11 of 64
Rev 5 June/2020
Page 12
7.0 Installation
CAUTION
!
The Incubator is fastened to the base skid and it is usually the best procedure to leave the skid attached until the Incubator
is located in its approximate position to facilitate ease in handling. The base skid can then be removed by removing the two
bolts holding the cabinet to the skid. See BCD-18314 for dimensional and feature information about the incubator.
It is recommended that no less than two people
are present when the incubator needs to be moved.
Examine the Incubator carefully.
INSPECT both the exterior and the interior for any transit damage before discarding the shipping crate.
7.1 Location
In locating the Incubator, consider all possible conditions that might affect its performance as well as laboratory procedures
for its intended purpose. Do not locate near heating or cooling ducts, or next to equipment that generates heat (steam
radiators, stoves, ovens, autoclaves, etc.). Avoid direct sunrays and rapidly moving air currents. These conditions adversely
affect the even heat dissipation required from the exterior surfaces of the Incubator and may cause the temperature
variation in the chamber to exceed specifications as stated in section 3.2. Since the Incubator needs even heat dissipation
on all surfaces in order to maintain an internal temperature variation of less than ± 0.2°C, a minimum of 2 inches (50mm)
must be allowed between the rear and sides of the Incubator and any walls, partitions or obstructions to facilitate adequate
convection of air around the Incubator. Confirm clearance with a tape measure if needed. Adjust the Incubator location
accordingly. In addition, the Incubator location should address access to its power cord for connection/disconnection if
necessary. For maintenance and service purposes, the control center electronics should remain accessible. Leave at least 3
ft of clearance in front for the swinging door.
Note: When locating Incubators side-by-side, they should be spaced at least 10” apart to permit one
Incubator to be in a heated Decon Cycle with a minimal heating disturbance on adjacent Incubators
running in their normal “run” mode.
7.2 Leveling
Prior to use the incubator should be leveled using a bubble level on a middle shelf in the chamber. The Incubator should
have all 4 leveling feet firmly on the bench or floor. Level from side to side and front to back. By turning the adjustable
leveling feet counter-clockwise, raises the Incubator.
7.3 Shelf & Water Pan Preparation.
Shelves
Incubator is shipped with the shelves installed. The shelf mounted water pan is also in place that option is ordered. NuAire
recommends decontaminating all surfaces within the interior chamber, glass door, and outer door with gasket. They can be
wiped down with a disinfectant of 70 percent alcohol or similar non-corrosive antimicrobial agent. Use only disinfectants
that are compatible with the vinyl gasket, the powder coat paint on the outer door and the 304L stainless steel in the
chamber. See BCD-15169
Note: Absolutely no chlorinated or halogen materials are to be used in the chamber.
NOTE: NuAire does not offer any product warranty with respect to cleaning material compatibility. USE AT YOUR
OWN RISK! The information provided above is from raw material suppliers and known general source documents
for use to develop application cleaning SOP’s.
Four (4) shelves are installed in each Incubator. The shelf positions are easily changed by removing the shelf and moving
the horizontal guides to the position desired. If the shelf does not bind or disengage from the horizontal bracket it is
installed correctly. Additional shelves and shelf guides are available.
Water Pan
Place water pan in the center on the bottom of the chamber and fill with Single distilled water no purer than 1 Mega Ohm.
It is recommended to fill the pan to a maximum of about 1/2 inch below the top rim. See Sections 3.1 and 8.2 for more
specifications and operational details regarding proper maintenance instruction regarding the water.
OM0246 Page 12 of 64
Rev 5 June/2020
Page 13
OM0246 Page 13 of 64
Rev 5 June/2020
Page 14
7.4Electrical
CAUTION
!
The electrical supply circuit to the Incubator must conform to all national and local electrical codes.
Consult the serial-data plate, located at the front of the right side of the Incubator, for voltage, cycle, phase, and ampere
requirements before making connection. Plug the power cord securely into a grounded power source. VOLTAGE SHOULD
NOT VARY MORE THAN 10% FROM SERIAL PLATE RATINGS. Have a qualified technician check with the power source with a
properly rated volt meter if needed. A separate branch circuit is recommended to prevent possible loss of product due to
overloading or failure of other equipment on the same circuit. A SURGE PROTECTOR IS STRONGLY RECOMMENDED to
avoid power-related faults.
7.5 Air Inlet Connection
The air inlet and filter are installed at the Factory. The filter has flexible hose that attaches to bulkhead fittings in the
bottom of the front panel and can be found under the valance that is beneath the door. One 50mm polypropylene .3micron HEPA filter is connected to the hoses. The filter is labeled “Air filter”. To remove the valance the outer door must
be open. Reach in under the valance and lift it to remove. See BCD-18316. The air inlet on the back of the incubator is
connected to the filter by a hose at the factory and does not require installation.
CAUTION: This is a free air supply. DO NOT connect to a pressurized air source.
7.6 CO2 / N2 Gas Supply Setup
7.6.1 Gas and Air supply HEPA filtering is installed at the factory
on the front of the incubator as illustrated in the picture.
Each incubator is equipped with the filters required for
the control systems available on that model.
• CO
Pressure to the Invitrocell is rated at 20 PSIG or 1.4 BAR.
2/N2
Do not exceed 25 PSIG or 1.8 BAR to avoid damaging the incubator.
• CO
of medical grade is recommended.
2/N2
• A two-stage pressure regulator, Linde #19590 (NU-1564)
or equal is recommended.
• DO NOT USE a single stage regulator.
7.6.2 CO
/ N2 Gas Supply Tubing Connections
2
Included with this unit and for all models of the 5800 series is (1) six foot (2 m) of vinyl tube.
It connects between the low pressure side of the 2 stage regulator at the CO
INLET (as shown in the following Figure). Next, secure each connection with the clamps supplied.
CO
2
supply and incubator fitting labeled
2
On Models NU-5831 and NU-5841 there is a second six foot (2 m) of vinyl tube.
It connects between the low pressure side of the 2 stage regulator at the N
supply gas and incubator fitting labeled
2
INLET. Next, secure each connection with the clamps supplied.
N
2
Figure illustrating CO
2
OM0246 Page 14 of 64
Rev 5 June/2020
and N2 gas supply connections
Page 15
7.6.3 CO2/ N2 Supply
CAUTION
!
CAUTION
!
•CO
/ N2 of medical grade is recommended.
2
• A two-stage pressure regulator, Linde # 19590, or equal, is recommended.
• DO NOT USE a single stage regulator.
It will not give a stable output at 20 psi and exposes the Incubator to the gas cylinder pressure.
• High concentrations of CO
• Install Incubator in a well-ventilated area.
• Gas Supply pressure to the Invitrocell is rated at 20psi (1.4 bar).
• Do not exceed 25psi (1.8 bar) or damage to the incubator may occur.
This Incubator is designed to be operated with CO
Connecting a flammable or toxic gas can result in a hazardous condition.
Gases other than CO
CO
gas cylinders have a UN1013 label on the cylinder and are equipped with a CGA 320 outlet valve.
2
Check the gas cylinder for the proper identification labels.
Do not use CO
A siphon tube is used to extract liquid CO
Consult with your gas supplier to ensure that the CO
7.6.4 CO
/ N2 Pressure Regulators
2
The regulator’s high-pressure stage direct from the supply cylinder must have a range of from 0 to 2000 PSI or
0 to 140 BAR. This gauge indicates actual tank pressure.
The low-pressure stage should have a range of 0 to 30 PSI or 0 to 2 BAR (100 PSI or 6 BARS maximum). This gauge
will indicate the actual CO
Some single stage CO
All NuAire Incubators use CO
maximum performance.
To connect the regulator:
First: Open the CO
or dirt that may have collected in the valve outlet.
BE SURE to keep your face away from the valve outlet to protect your eyes from dust or dirt.
Second: Make sure the regulator pressure-adjusting screw is released by turning it counterclockwise until it turns
freely.
Third: Attach the regulator to the cylinder valve and tighten the connection nut with a wrench.
BE SURE DISC SEAL IS IN PLACE BEFORE MAKING CONNECTION.
7.6.5 CO
/ N2 Supply Adjustment
2
With the regulator OFF (i.e. fully counterclockwise), open the cylinder valve slowly usually 1 to 2 turns is sufficient.
NEVER STAND IN FRONT OR BEHIND THE REGULATOR WHEN
OPENING THE VALVE. ALWAYS STAND TO ONE SIDE.
The cylinder tank pressure should read: 700 to 800 PSI (48 to 55 BAR) for CO
more or less depending on the temperature of the cylinder.
for N
2
Next turn each regulator's pressure adjusting knob clockwise until the low-pressure gauge reads 20 PSI or 1.4 BAR.
Both gas connections are now complete.
Note: OSHA requires the CO
to prevent accidental damage to cylinder.
7.6.6 Checking the Connections
The connections can be checked for leakage by brushing a small amount of soapy water on each one.
Observe to see if any bubbles are blown. If not the connection is secured properly.
If it is, the cause of the leaking needs to be determined, (i.e. loose clamp or damaged hose) and corrected.
OM0246 Page 15 of 64
Rev 5 June/2020
/ N2 should not be connected to this equipment.
2
gas cylinders equipped with siphon tubes.
2
2
pressure to the Incubator.
2/N2
pressure regulators have two gauges. USE A TWO-STAGE REGULATOR.
2/N2
in such quantities that precise metering of CO2/N2input pressure is important for
2/N2
cylinder slightly, for an instant (this is termed “cracking the valve.”) This will blow out dust
2
tanks to be physically restrained (i.e. via chained to wall)
2
gas can cause asphyxiation!
2
gas only.
2
from the cylinder which can damage the pressure regulator.
cylinder does not contain a siphon tube.
2
or 2550 to 2650 PSI (176 to 183 BAR)
2
Page 16
7.7 Removing the cable tie used to secure the air pump for shipping:
7.8 Setting up the access port:
When utilizing the access port do not block or restrict the relief holes in the center of the port plug. Blocking this venting
can cause performance issues for the incubator. Cut slits or new holes in the vented plug to make passage for cables or
other items requiring access to the chamber if needed.
7.9 Correct Installation
When the Incubator is installed correctly connected to the power source it is rated for gas connections made properly, the
water pan filled, the shelves in place, and the unit is leveled. Read section 8 and follow all instructions for setting set points
at the desired value. Then calibrate the Incubator control systems as explained in section 9. If the Incubator is installed and
calibrated correctly it will meet the performance specs listed in section 3.2.
7.10 Reversing Incubator Hinges
Reversing the incubator hinges from the right to the left side SEE PTB0236 for instructions.
This requires disassembly of the incubator and should be done by a qualified service technician.
OM0246 Page 16 of 64
Rev 5 June/2020
Page 17
8.0 DHD Operation
564
191
ATTENTION ACCOMPANY'S
SYMBOL
POTENTIAL ELECTRICAL
PERSON TO ACCESS
188
HOT SURFACE
!
CAUTION
!
Safety alert symbol indicates potentially hazardous
Situation which, if not avoided, may result in minor
Or Moderate injury.
CAUTION: A qualified technician who is familiar with the proper maintenance procedures required for this equipment,
as well as repair must perform all maintenance actions on this equipment.
The Incubator is designed to provide a sterile, constant temperature, constant CO
atmosphere for optimum growth of tissue cell cultures and other organisms requiring this precise environment. To operate
the Incubator properly, the following parameters must be reviewed, carefully set, and/or prepared.
8.1 Sterility
The chamber environment is not selective. The growth environment is applied equally to all microorganisms (specimens
and contaminates) within the chamber. Therefore, before placing any cultures in the Incubator, the shelves and sidewall
top plenums should be sterilized. The interior sidewalls, top, bottom, door, as well as the gasket should be wiped clean
with a 70% solution of isopropyl alcohol or other disinfectant compatible with the chamber construction, to remove any
contamination. Use mild detergent to clean the exterior of the Incubator. This Incubator provides the lab professional with
a choice of 2 heated decontamination cycles. (See section 10.1 for further information.)
INFORMATION OR IMPORTANT
HAZARD ONLY QUALIFIED
BURN POTENTIAL
level and naturally humidified
2
NOTE: NuAire does not offer any product warranty with respect to cleaning material compatibility. USE AT YOUR OWN RISK! The information provided above is from raw material suppliers and known general source documents
for use to develop application cleaning SOP’s.
8.2 Humidity
Water Pan available on all Models
Humidification of the Incubator chamber is achieved through the process of water evaporation (vapor water pressure) from
a stainless steel water pan placed on the bottom of the chamber. Materials of different thermal resistance (e.g., glass,
plastic) do not offer sufficient thermal recovery and are not recommended for water pan usage. Although some metals
offer better thermal coefficients than stainless steel, dissimilar metals cause electrolysis in the acid atmosphere (carbonic
acid) and should never be used, or placed within the chamber.
Use only distilled or mineral-free water, no purer than 1 mega ohm, in the stainless steel pan. The water should be
changed at least once a week. FLOODING THE BOTTOM OF THE INCUBATOR IS NOT RECOMMENDED since it is difficult to
change the water weekly and almost always necessitates the use of chemicals, which are not recommended and may
damage the stainless steel. Also, it promotes condensation on the inner walls because it steals the natural convection, heat
flow through the inner chamber and condensation points occur.
NOTE: ABSOLUTELY NO CHLORINATED OR HALOGEN MATERIALS ARE TO BE USED IN THE CHAMBER.
Humidity Recovery: Using the water pan supplied with the incubator and with a 5.0% CO₂ set point.
NU-5810: After a 15 second door opening RH recovers to 90% of what the RH level was prior to opening the door for
15 seconds in 20-40 minutes on average.
NU-5831: Recovery after a door opening is directly dependent on the O₂ set point and ambient RH.
The lower each of these factors is the longer the recovery time.
Contamination in the water pan may be avoided by adding a small amount of copper sulfate to the water pan after each
Incubator decontamination.
NOTE: Please contact the copper sulfate manufacturer to determine the proper amount of copper sulfate that should be
added to the water pan.
OM0246 Page 17 of 64
Rev 5 June/2020
Page 18
Condensation on the glass door, walls, top, or bottom of the chamber indicates an incorrect balance of door and front
perimeter heat. Both the door and front perimeter heaters operate on a duty cycle. A good starting point for these duty
cycles is the default setting for the door & perimeter heaters in a room ambient temp of 22°C (72°F) at a temperature set
point of 37.0°C. The fresh airflow that the air pump delivers to the chamber has been preset at the factory. If condensation
starts forming on the sides or back wall in the chamber, the number and length of air injections into the chamber can be
increased. (See section 8.8.1.2 for instructions.) To alleviate the condensation, increase the airflow. To increase RH in the
chamber reduce the airflow. If airflow is decreased, some condensation may be unavoidable. The port plug with breather
hole must be used on the side port.
Humidity Monitoring and control system are available on Models NU-5820 and NU-5841.
These models humidify the chamber by circulating the chamber air through heated water vapor saturated air in a tank
located outside of the chamber. This gives the operator control over the amount of humidity in the chamber
ambient dependent low value to 95%. After a door opening, the recovery time to 90% +5% / -3% is 25 minutes on average.
The control system has a +3% accuracy from any given set-point 90% and below. Much of the information regarding the
water pan can be applied with the exception of the water pan.
RH setpoints exceeding 90% will limit the O2 setpoint input (on 5841 models) to a minimum of 2%.
The control system uses a solid-state capacitance humidity sensor to monitor the relative humidity within the chamber.
If humidity is required, the control system will activate a solenoid valve routing the chamber air through a humidified
heated enclosure just before being returned to the chamber. The control will continue rout the chamber air through the
water vapor saturated air until the set-point is achieved. Use only single distilled water, NO PURER THAN 1 MEGA OHM to
fill the reservoir. The relative humidity display and control option offers an accurate and reliable method to control
humidity when required. The alarm set-point defaults at 8% over the set-point and 3% below.
NOTE: DO NOT SEAL OR RESTRICT THE OPENING IN THE PORT PLUG. THIS IS A PRESSURE RELIEF FOR THE CHAMBER.
8.3 Control System Introduction
The NuAire Incubator Control Electronics system is designed to serve the control requirements of the Incubator chamber.
Temperature, CO
optimum conditions for culture growth within a chamber. Operator input is coordinated through the NuTouch LCD keypad
and status displays.
Figure 1 shows the various inputs and outputs of the system.
levels in all models and RH & O2 in available models are controlled by preset values to provide the
2
from an
OM0246 Page 18 of 64
Rev 5 June/2020
Page 19
Button Description
Action when touched
Control system description
(i.e.….TEMP or CO2)
Brings a pop up screen that describes the control system
that was touched
Goes to the Environmental Set Point Screen
the password function is enabled)
Goes to “System Settings” Screen for
access to more menus
Date & time display
Goes to screen for adjusting time and date
8.3.1 Chamber control in a single electronic package.
8.3.2 Enhanced information presentation on a NuTouch LCD
Main Screen Continuously displays:
• Measured Chamber Temperature
• Measured CO
2
%
• Measured RH % in Available models
• Measured O
% in available models
2
• Current Screen ID “MAIN SCREEN”
• A mode status indicator (RUN, STANDBY or Power loss Recovery)
• Door ajar status displays when door is ajar
• Door Closed status displays when door is closed and flashes during the post door close delay time.
• NuAire Logo button Leads to the System settings
menu when pressed
• Date and Time
• Air, CO
• PIDT: CO
RH, & O2 inject indicators show when gas or RH is injecting.
2,
RH, & O2 indicate that the system showing is in Post Inject Delay Time.
2,
• A heated decon Cycle display
• An Alarm Status Menu
• Maintenance Required displayed
8.3.3 Active buttons on the MAIN SCREEN:
Measured value of
each control system
NuAire Logo
8.3.4 Accessing the Incubator Control Screens
OM0246 Page 19 of 64
Rev 5 June/2020
(Key pad to enter password will pop up if
Page 20
Accessing the many Control screens, performing functions and diagnostic tests, and maneuvering through the
Save
-
This command saves the changes you have made and returns you to the calling screen or, in some
cases, back to the main menu.
Exit
-
This command exits you from the current screen and returns you to the calling screen.
Close
-
The command causes the current screen to be removed and returns back to the calling screen.
Back
-
This command returns you to the previous screen.
Start
-
This command causes a main event to begin (e.g. Decon Cycle).
Next
-
This command causes another step (screen), in a list of steps (screens) to be displayed
(e.g., Decon cycles and some CO2 calibration screens).
Decon
-
This command begins the Decon sequence.
Cancel
-
This command will cancel the changes you have made, and in some circumstances, cancels the
(e.g., cancel in the middle of the Decon screen setup returns you to the calling screen)
NuTouch LCD menu system is accomplished through the following methods:
•Touching a “graphic” Icon like the NuAire Logo in the Main Screen which access the System Settings menu,
or lead you to an Enter Password screen when a password has been established.
•Touching a “Button function” button command that is visible on the screen.
The button will have text in it indication the function you are about to perform.
The following buttons are used on the NU-5810:
An explanation of these buttons is as follows:
“set of requests” that you have entered and returns you back to the calling screen.
•Touching a “Text” button command. The text or numerical value will be in blue indicating that is a link
similar to the computer or on the internet. The system settings menu screen shown below is a good
example of both the Text commands and Button commands:
Note: The Main Screen is the one exception for the above described buttons. The green clock text is and is a link to adjusting the time.
The large white Control system measured values are also a “second” link to the “Environmental
Set-Points” menu and the control system descriptions are a link to system description pop up
screens when touched.
8.3.5 Functioning accessed from the System Settings screen:
The “System Settings” menu screen can be accessed from the “Main Screen” by touching the NUAIRE logo:
Environmental Set-Points: Allows adjustment of Temperature, CO
settings.
Note: You can also access the Environmental set points by touching any
control system numerical value in the MAIN SCREEN.
Performance Monitor: Views a function’s Set-Point and Measured values over the last seven days.
Service Settings: Protected by the password 9876, it facilitates Calibrations, Option settings,
and access to Diagnostics.
System Information:Views the units Model Number, Serial Number, Control board assembly (CCBA) Software Revision and the Display board assembly (LCBA) Software Revision.
Display Settings:Brightness: Allows adjustment of the screen brightness. Slide right to increase / left to decrease
Save: Remember to press the “Save” button after making changes or
calibrating the screen. They will be lost if “Cancel” is pressed to exit
the screen.
Set Password: A 4 digit password can be set and will protect access to Environmental Set points,
running Decon cycles, and changing the password. See section 8.3.7 for instructions
, Door, and Perimeter heater power level
2
Rev 5 June/2020
OM0246 Page 20 of 64
Page 21
to set the User Password.
9 8 7 6
Decon: Begins Decon setup instructions and starts the heated Decon Cycle when pressed.
Also, it advances through the phases of the Decon Cycle.
System Settings Screen
8.3.6 Automatic notification of abnormal situations (See Section 11 for details)
The red “Alarm Status” text message (button) will appear on the Main
Screen to indicate a fault within the Incubator system. An audible
alarm is also heard. The audible alarm is on for 10 seconds at the beginning
of an alarm condition and thereafter, it will be on 1 second every 30
seconds.
By depressing (touching) the Alarm Status text on the Main Screen, you
now reach the Incubator Alarm Status screen which will identify which
Incubator sub-systems are in an alarm state (Temperature, CO
General). You may exit the screen or touch any of the sub-system buttons
which will give you more specific information on alarm conditions. See Section 11.0 for more specifics on alarm
conditions.
8.3.7 Password Protection
There are 2 levels of password protection.
1. The permanent numerical password required for accessing the “Service Settings” menu screen.
This screen is accessed by pressing the “Service Settings” text Icon in the “System Settings” menu screen.
The password is and it cannot be changed or deleted.
, or
2
The functions in the “Service Settings” screen are intended to be used by qualified technicians to calibrate
the control system sensors like Temperature and CO
incubator performance, perform system diagnostics and to Factory or Master reset for returning the
incubator to Default settings.
2. A selectable 4 digit numerical password that is entered by the Incubator user to prevent others that are not
authorized to change the Incubator set points (CO
Decon Cycle or changing passwords. The password protection may be initiated by accessing the “Set
Password” menu in the System Settings menu screen. Once initiated, a four digit number must be entered,
then re-entered for verification.
OM0246 Page 21 of 64
Rev 5 June/2020
, adjust optional parameter settings to customize
2
%, Temp), running conditions of the Incubator, initiate a
2
Page 22
8.3.8 Standard remote communication capability
The NU-5810 Incubator comes equipped with the following remote communication tools:
• USB Output of daily, weekly & monthly performance logs and an event log that indicates when alarms,
• 4-20 milliamp analog Output
• RS-485 2 way communications
• A USB method of updating the software
8.3.9 Diagnostic and calibration assists.
To access the Service Settings Menu, perform the following steps: NuAire
Logo → Service Settings → 9876 Enter. You are now in the Service Settings
Menu (screen). This screen provides you the capability to review and
analyze numerous Incubator functions including the following:
• Individual analog inputs may be displayed to assist calibration.
• Individual outputs may be forced to an on or off condition.
• Individual digital inputs may be displayed.
• Calibrated and Uncalibrated values are displayed.
• All options may be individually tested.
8.4 NuTouch Display Screen
The system LCD touch screen has the following functions.
8.4.1 Front LCD touch screen Descriptions
The NuTouch LCD allows the user to access a wide variety of unit
configurations and diagnostic menus by touching the NuAire logo icon.
The Main Screen displays:
• Power Interrupted message displayed when power is lost and
• current Calibrated values of enabled control systems
• Current operating mode (run, standby or Power loss recovery)
• Door Ajar, Alarms, and Maintenance notices
8.4.2 Inject CO
On the bottom right of ‘Main’ screen, “CO
is flowing into the chamber maintain set point for each of the control systems.
Each text icon appears independently when that System is actively injecting.
8.4.3 Post Inject Delay Time (PIDT)
On the bottom right of the ‘Main” screen just below the inject status icons from 8.4.2, “CO
when these systems are in post inject time after the injection is done.
Each text icon appears independently when that system is in the delay time.
8.4.4 Inject Air Status On the bottom right of the Main screen “Air” text indicates that fresh air is currently being injected into the chamber.
8.4.5 Door Ajar Status
The door ajar white text message indicates when the inner glass door is not closed and latched; the inner glass door
is open. The door ajar status is a result of the pressure switch (open/closed) located along the top middle of the
front panel. After the door is closed, the text will display “Door Closed Delay” indicating the delay time (60 sec
adjustable).
8.4.6 Power interruption notice
door openings and power interruptions occur.
then re-connected
/ O2 / RH Status
2
/ O2 / RH” text indicates when the inject valve is open and the gas or RH
2
/ O2 / RH” text indicates
2
OM0246 Page 22 of 64
Rev 5 June/2020
Page 23
Indicates that the power has been interrupted. Touch text to acknowledge and Nuaire Logo will reappear.
8.4.7 USB status
The USB status icon appears in the lower right side of the Main screen when a memory stick is placed in the USB port
at the back of the unit while down loading performance and event history. This icon disappears when the memory
stick is removed from the port.
CAUTION: When a memory stick is placed in the USB port it can interfere with the RS-485 communication output.
The RS-485 communication may have to be restarted after the memory stick is removed from the port.
8.4.8 Maintenance Required Notice
The Maintenance required notice appears in the middle of the right side of the screen indicating that either the
RH water tank needs to be filled or that it is time to check the capsule filter in the Sensor bay.
Press the notification button to bring up the screen that indicates the maintenance that is required.
8.4.9 Alarm Status Notice
The “Alarm Status” button indicates an abnormal status condition.
This text appears just above the AIR CO
status messages (paragraph 8.4.2 and 8.4.3). The “Alarm Status” button and
2
audible alarm indicates the abnormality. A catastrophic temperature control condition will de-energize the safety
relay and cause the chamber to cool below the set point. The unit will resume heating when it has recovered from
the abnormality. See section 8.3.5 and 11.0 for more details on Alarms.
8.5 Rear Component Panel Detail
The rear control panel contains the following functions (see also BDC-15170)
8.5.1 Power Cord
The power cord is 6.5-foot (2 m) in length, type “SVT” molded plug, allowing for long life and easy cleaning.
8.5.2 Circuit Breaker
All control electronics are protected with a circuit breaker that may trip at 110% of load rating, but will trip at 145%
of load rating in less than 2 seconds.
Should the circuit breaker open (pop-out button will appear), merely depress to reset.
8.5.3 N2 Inlet on NU-5831 and NU-5841
inlet provides a fitting for vinyl tubing for oxygen control.
The N
2
Be sure to follow the recommended inlet pressure to insure proper flow rates and consistent O
readings.
8.5.4 CO
The CO
Inlet All models
2
inlet provides a fitting for vinyl tubing.
2
Be sure to follow the recommended inlet pressure to insure proper flow rates and consistent CO
readings.
8.5.5 Power Switch
The power switch, located at the top of the rear panel, controls all power to the Incubator.
8.5.6 Standard Communication Interface’s / Alarm
Three receptacle connectors are provided for direct field connection of a computer for 2 way communications and
incubator monitoring/or output to a printer or other device only capable of receiving RS-485 digital information,
dry contact remote alarm indicator, also 4 to 20 mA output to an analog monitoring device or system.
All come standard and ready to use. See section 10 for more detail.
8.5.7 Air Inlet (All Models)
This is a free air supply.
DO NOT CONNECT TO A PRESSURIZED SOURCE.
Air filter is installed at the factory on the front of the unit.
mak
percentage
2
percentage
2
OM0246 Page 23 of 64
Rev 5 June/2020
Page 24
8.6 Front Component Panel
The front component panel contains the following functions. BCD-18316
OM0246 Page 24 of 64
Rev 5 June/2020
Page 25
8.6.1 RH Fill Port (Available on models NU-5820 and NU-5841)
The RH evaporator tank is filled through the port that has the red plug (w O-ring) screwed in it. Because of the
pressure in the evaporator tank during an RH injection the O-Ring must seal against the top of the fill port.
8.6.2 N
Supply filter (Available on models NU-5831 and NU-5841)
2
supply filter is installed at the factory and should be changed following the instructions found in Maintenance
The N
2
Section 10.
8.6.3 CO
Supply Filter
2
The CO
supply filter is installed at the factory and should be changed following the instructions found in
2
Maintenance Section 10.
8.6.4 Air Inlet Filter - All models The air inlet filter is installed at the factory and should be changed following the instructions found in Maintenance
Section 10. This is a free air supply.
Note: DO NOT CONNECT TO A PRESSURIZED SOURCE.
8.6.5 Chamber sample port
The Sample port is provided for checking the CO
level in the chamber to monitor the calibration the Sensor and
2
determine if it need adjustment.
8.6.6 Sensor bay access panel
Provides access to the CO2 sensor, Capsule Filter and the Hydrophonic filter for maintenance.
OM0246 Page 25 of 64
Rev 5 June/2020
Page 26
8.7 Run Mode Operator Interactions
OM0246 Page 26 of 64
Rev 5 June/2020
Page 27
8.7.1 THE FIRST TIME THE INCUBATOR IS TURNED ON
•The NuTouch boot up screen appears then shows a screen instructing you to “Read the manual and follow
all instructions to properly install the incubator” Both screens will automatically advance after 5 minutes.
To manually advance:
o Press the exit button then a thank you screen will appear.
o Press the “Continue” button then the unit will go to the “MAIN SCREEN”.
These screens will not show again unless you “MASTER RESET” the incubator control board.
•Setting the Clock
8.7.1.1 Clock display organization:
YYYY-MM-DD Hr: Min and is displayed in green text in the lower left of the display screen.
8.7.1.2 Press the green text and the Time and date adjustment screen appears.
After a power interruption the date and time will flash as follows:
•Green text indicates there was a power interruption but the real time clock remained powered.
Press the flashing text to bring up the “set the clock” screen. If no change is needed press cancel
to go back to the main screen. Otherwise make the required changes then, press the save
button to save the changes and return to the main screen.
•Yellow text indicates the real time clock lost the backup power and the date and time need to be
reset. Press the flashing text to bring up the date/time setting screen
8.7.1.3 Use the up/down arrows to increase or decrease the highlighted value.
Use the left/right arrows to move between date and time values that require changing.
8.7.1.4 Press “Save” to save the changes you just made or
Press “Cancel” to abort the changes and return to the main screen.
The mode display, on the ‘Main’ screen, is used to identify the mode of chamber operation: either standby or run.
In general, there is no need for operator interaction in "run" mode. However, operator interaction is required when
the incubator is first turned on then to perform calibration functions and address abnormal conditions status. If an
abnormal condition has, or does, exist for a particular parameter, an alarm becomes active and the ‘Alarm Status’
button will appear. This could be a catastrophic alarm condition, which could harm the tissue culture cells. It does
indicate an operational abnormality and should be checked. To acknowledge the abnormality, simply press the
“Alarm Status’ button to view the alarm, to dismiss it press the corresponding alarm’s button. The audio and visual
alarms will then extinguish. If the abnormality still exists, the return after waiting the specified alarm time.
Let the Incubator run normally, and if the alarm doesn't come back on, then everything is normal. If the alarm does
come back on, use the troubleshooting guide to correct the abnormality (see Section 11.0).
8.8 Standby Mode Operator Interactions
The NuAire incubator automatically enters the standby mode when necessary.
There is no action required by the user to enter this mode.
All functions that put the incubator in standby mode automatically start a 10 minute timer.
If the function is not interfaced through the screen on the NuTouch display, the incubator then returns to the MAIN SCREEN
8.8.1.1 Accessing the Environmental Set point screen used to manage set
points is done one of two ways:
• Press any control system value displayed in the main screen.
• Press the NuAire logo then press the Environmental Set
Point text in the Systems Settings screen
8.8.1.2 Pressing the blue numerical value next to the control system that
you intend to change will bring up a number key pad. Punch in the
new value and hit the
“ent” key to leave the key pad and return to the Set-point screen.
You will be asked if you want to save the new set point(s) when you
press Exit to leave the set point screen and return to the Main screen.
Press “Save” to retain the changes.
OM0246 Page 27 of 64
Rev 5 June/2020
Page 28
Temperature
37.0 C
All Models
Note 1: A set point of 0 (zero) will put the system into an Idle
level of gas/RH in the chamber.
Note 1: A set point of 0 (zero) will put the system into an Idle
level of gas/RH in the chamber.
Note 2: A set point of 21.0% will put the system into an idle
lower O₂ set points.
Automatic maximum
O₂ Adjustment
0.1% to 1.0%
20.7%
1.1% to 3.0%
20.0%
3.1% to 6.0%
19.0%
6.1% to 10.0%
18.0%
10.1% to 15.0%
17.0%
15.1% to 20.0%
16.0%
8.8.1.3 Default Set points
CO₂5.0%
RH
(Humidity)
90.0%
state where it does not control and all alarms are
inhibited. The sensor remains active and reports the
state where it does not control and all alarms are
inhibited. The sensor remains active and reports the
All Models
state where it does not control and all alarms are
inhibited.
O₂21.0%
Note 3:O₂ set points below 10% will automatically reduce
the Air Injection Time value to “0”. This stops air
injections that will interfere with achieving these
8.8.1.4 Automatic Adjustment of Post Inject Delay Times when O₂ control system is active
Post inject delays are automatically adjusted according to the O₂ set point to facilitate gas and RH recovery
after a door opening. The value displayed in the option menu will not change but is the basis for a formula in
the software that calculates the new delay time depending on the O₂ set point. As the set point is reduced
the delay times are shortened automatically through the software.
The value of the post inject delay time used is the value that shows in the Options display for that control
system whether it is the default or user selected.
8.8.1.5 When the CO₂ system is active the level of CO₂ inversely affects the O₂ present in the chamber.
The maximum achievable O₂ set point is indicated by the following table.
NU-5820
and
NU-5841
NU-5831
and
NU-5841
CO₂ Set Point
OM0246 Page 28 of 64
Rev 5 June/2020
Page 29
8.9 Diagnostic and Checkout Procedures
The Incubator controller provides general diagnostic facilities:
• Diagnostic mode is intended for factory and field technicians.
• It allows them to turn the controller's output signals (heaters, valves, safety devices etc.) on and off.
• Diagnostics are located in the “Service Settings” menu categorized by the control systems (i.e. Temperature, CO
RH and O
) plus a general category for diagnostics not directly related to the control systems.
2
Press the “Diag” button to access them.
•Each screen and the function buttons in them initiate a 5 minute timer. When the timer runs out the display will
exit that screen and go back to the main screen. Manually exiting the diagnostic screen will automatically stop that
timer and return to the Service Settings screen. The function button that was activated will return the function to
the default diagnostic condition at the end of the 5 minute timer. Manually putting the button back to the default
condition will reset and stop the timer
8.9.1 Diagnostic Mode
To initiate the diagnostic mode, perform the following from the ‘Main’ screen:
• Select the NuAire Logo to enter “System Settings”
• Select “Service Settings” and enter 9876, and then hit enter (ENT)
• In the Service Settings menu, select the “Diagnostics” button for the desired function
1. Power Supply Regulated +12 VDC (current voltage)
2. Power Supply Regulated -12 VDC (current voltage)
3. Air Inject Solenoid (on/off)
4. Alarm dry contact Relay (NO/C/NC) switch states
5. Air Pump Triac (on/off)
6. Spare AC Triac
(Note: Default values are in bold)
The paragraphs on the following page provide a more complete description of the diagnostics listed above.
,
2
OM0246 Page 29 of 64
Rev 5 June/2020
Page 30
The Diagnostic screen is shown to the right of each of the control systems Temperature, CO2, RH & O2 and the General
category for your reference. Once entered, the Diagnostics screen will display the calibrated and uncalibrated values of the
given function and all related components.
Temperature Control Diagnostics
1. Safety Relay
This function shows the current state of the safety relay.
LCD display will show “On” or “Off” corresponding to the
relay condition. On for closed or set or yes.
2. Chamber Temperature Sensor
This function shows on LCD the current values, calibrated (displayed) and
not calibrated of the temperature sensor controlling the chamber
temperature.
3. Safety Temperature Sensor
This function shows on LCD the current values, calibrated
and not calibrated of the Safety temperature sensor located in the chamber.
4.Chamber Heater Triac Control (0%, 50%, and 100%)
This function allows the jacket heater to be turned on at different percentages 0, 50,100)
5. Door Heater
This function shows on LCD the current state of the door heater. This function also allows the door heater to be
turned on at different percentages (0, 50, and 100 percent). The LCD display will show phase firing percentage
chosen default is 0%.
6. Perimeter Heater
This function shows the current state of the perimeter heater. This function also allows the perimeter heater to be
turned on at different percentages (0, 50, and 100 percent). The LCD display will indicate percentage of phase
firing chosen default is 0%.
Control Diagnostics
CO
2
1. CO
Sensor
2
This function shows the current CO
sensor reading both the
2
“Un-Calibrated sensor signal and the Calibrated value shown in the
display screen.
2. CO
Inject Valve / on/off
2
This function shows on LCD the current state of the CO
inject valve. LCD
2
will show “ON” or “OFF” corresponding to the valve condition. On for
open or set or yes.
3.Sensor Bay Heater Triac Condition / on/off
This function shows on LCD the current state of the sensor bay Heater
triac. LCD will show 0, 50, and 100 percent corresponding to the relay
position. The triac can be toggled to each percent of phase firing manually.
4. CO
Sensor Bay Temperature Sensor
2
This function shows on LCD the sensor bay temperature in degrees C.
The Sensor bay is heated to temperature set point +3.5 degrees
5.Air inject Solenoid / on/off
This function shows on LCD the current state of the Air Inject Solenoid.
LCD will show “ON” or “OFF” corresponding to the solenoid position.
Default is off-solenoid activates when the button is touched
6.Air Pump / on/off
OM0246 Page 30 of 64
Rev 5 June/2020
Page 31
RH Control Diagnostics functioning
1. RH Sensor
This function displays the current value of the chambers RH sensor from
0 to 100%. It shows both the claibrated value shown in the display
screen and un-calibrated sensor signal.
2.RH Inject Solenoid
This function indicates the on/off (default off) state of the RH Inject
solenoid valve and allows the activation of it.
3.RH Water Level Float Switch
This function shows when the float switch is:
Open=water level OK or Closed=water level Low.
4.RH Water Heater
This function reports back on whether the triac/heater circuit is on or off (default off) and allows the activation of
it each time the RH heater button is pushed.
5.RH Water Temperature Sensor
This function shows the current temperature (in C) for the water in the water reservoir. The water temperature
should be 10 deg. greater than the chamber temperature set point. It can reduce some during long RH injections
after a door opening or when the system is first turned on but should recover within a few minutes.
Control Diagnostics Functioning
O
2
Sensor
1. O
2
This value displays the current value of the chambers oxygen sensor both
the Calibrated value shown in the display screen and
Un-Calibrated sensor signal.
Inject Valve /on/off :
2. N
2
This function indicates the on/off (default off) state of the N
Inject
2
solenoid valve and allows the activation of it.
Bypass valve on/off: THIS FEATURE IS NOT ACTIVE.
3. O
2
4. Blocking Solenoid /on/off
This function indicates the on/off (default off) state of the Blocking
solenoid valve and allows the activation of it.
General Control Diagnostics
1. Power Supply Regulated +12 VDC
This function will display on the LCD screen the current value (+12.xx) of
the regulated +12 VDC power supply.
2. Power Supply Regulated – 12VDC
This function will display on the LED screen the current value (-12xx)
of the regulated -12VDC power supply.
3. Power Supply Regulated +5 VDC
This function will display on LCD screen the current value (+5.00) of the
regulated +5 VDC power supply.
4. Chamber Air Inject Solenoid
This function shows on LCD the current state of the Air Inject Solenoid. LCD will show “ON” or “OFF”
corresponding to the solenoid position.
Default is off-solenoid activates when the button is touched
5. Remote Alarm Relay
This function shows on LCD the current state of the remote alarm relay. The LCD display will show “ON” or “OFF”
corresponding to the relay condition and its Normally Open (NO0 pair of contacts
6. Air Pump Triac
This function shows on LCD the current state of the Air Pump Triac. LCD will show “ON” or “OFF” corresponding to
the relay position. The triac can be manually turned On and Off.
OM0246 Page 31 of 64
Rev 5 June/2020
Page 32
Option Configuration Parameters
Format: Option (Default, Minimum/Maximum)
All the options for the NU-5810 Incubator use this same format. The default is set at the factory and can be restored in
the field with a Master or a Factory reset command. On the following pages, there is a short description of each Option
with its Default/Minimum/Maximum setting, followed by a more detailed description.
Temperature
1. Maximum Above Set Point (1.0°, .1/10.0)
2. Temp Low Time Out Timer (240 minutes, 0/300)
3. Minimum Below Set Point (0.5°, .1/10.0)
4. Temp Sensor Differential (4.0°, 0/10.0)
5. Recovery Delay Time (60 seconds, 0/120)
CO
2
1. System Enable (On, On/Off) (in Settings Screen)
2. Maximum Above Set Point (1.0%, .1/10.0)
3. CO
Low Time Out Timer (10 minutes, 0/240)
2
4. Minimum Below Set Point (0.5%, .1/10.0)
5. Post Inject Delay Time (PIDT) (60 seconds, 0/250)
Relative Humidity
1. System enable (On, On/Off) (in Settings Screen)
2. Maximum above Set Point (8.0 %, 0.0/30.0)
3. RH Low Time Out timer (60 minutes, 0/300)
4. Minimum below Set Point (3.0 %, 0.1/15.0)
5. Post Inject Delay time (PIDT) (90 seconds, 0/300)
O
2
1. System Enable (On, On/Off) (in Settings Screen)
2. Maximum above Set Point (1.0 %, 0.0/10.0)
3. O₂ Time out timer (30 minutes, 0/300)
4. Minimum below Set Point (1.0 %, .1/10)
5. N₂ Post Inject Delay time (PIDT) (30 seconds, 0/300)
General
1. Door Close Delay Time (60 seconds, 0/300)
2. Air Inject Time (10 seconds, 0/100)
3. Air Inject Cycle Time (1 minute, 0/120)
4. Digital Output Interval (0 minutes, 0/120)
5. Door Ajar Alarm Delay (5 minutes, 1/60)
6. Settings (Text in top left side of screen)
Note: Selecting any of the Option title texts will open a brief description of that option’s functions.
Default values are in bold
The following paragraphs provide a more complete description of the Optional configuration parameters listed above.
OM0246 Page 32 of 64
Rev 5 June/2020
Page 33
Temperature
1. Maximum Above Set point
This value determines the maximum deviation, measured in (°C) that the chamber is permitted to rise above set
point, once it is reached before an alarm condition is declared. An alarm condition will cause the chamber to enter
a safe condition where no power is enabled to any of the system output controls until the situation is rectified.
2. Temp Time Out
This value determines the time, in minutes, for the temperature to achieve set point. If the temperature doesn't
get to within 0.2° of set point within this time period, an alarm condition is declared.
3. Minimum Below Set point
This value determines the maximum deviation, measured in (°C) that the chamber is permitted to fall below set
point, once it is reached before an alarm condition is declared.
4. Temp Sensor Differential
This value specifies a maximum differential, measured in temperature (°C) that the two temperature sensors may
deviate from one another, or from the last read value. If this differential is exceeded, a warning LED is shown on
the running chamber. If multiple sensors fail to read within the specified limits, an alarm condition is declared. An
alarm condition will cause the chamber to enter a safe condition where no power is enabled to any of the system
output controls until the situation is rectified.
5. Recovery Delay Time
This value determines the time, in seconds, to turn off the chamber heaters in half-degree increments during a
temperature recovery cycle. The delay time is required to prevent temperature overshoot of the control set point.
CO
2
1. CO₂ System Enable (Enabled/disabled) Is located in the bottom right of the screen
This function will enable or disable the CO₂ system. If the system is Enabled it will be displayed in the MAIN
screen. If it is disabled it will not be displayed In the Main or Option/Settings screen for the CO₂ system:
System Enabled (with check mark) means the System is enabled
System Enabled (without check mark) indicates that the system is disabled
2. Maximum Above Set point
This value determines the maximum deviation, measured in (%) that the chamber is permitted to rise above set-
point, once it is reached before an alarm condition is declared.
3. CO
This value determines the time, in minutes, for the CO
Time Out Timer
2
percentage to achieve set point. If the CO2 percentage
2
doesn't get to within its lower limit of set point within this time period an alarm condition is declared. CO2
injections are stopped, and air is injected for 15 minutes incase CO2 was not being detected by the sensor. This
alarm can be cleared through the alarm screens or by opening the inner door.
4. Minimum Below Set Point
This Function gives the ability to set the value in percent CO₂ below set point that starts the CO₂ time out timer
described in item 2.
5. Post Inject Delay Time
This value specifies the time, in seconds, for an injection of CO
to be measurable at the sensor. When CO2 is
2
injected into the chamber, the system delays until this time period has elapsed before making a new control
decision. In this manner, diffusion induced delays will not cause the CO
system to overshoot the control set point.
2
OM0246 Page 33 of 64
Rev 5 June/2020
Page 34
Relative Humidity
1. System Enable (Enabled/disabled) on models NU-5820 and NU-5841
This function will enable or disable the RH system. If the system is Enabled it will be displayed in the MAIN screen.
If it is disabled it will not be displayed in the Main or Option/Settings screen for the RH system:
System Enabled (with check mark) means the System is enabled
System Enabled (without check mark) indicates that the system is disabled
2. Maximum above set-point
This value determines the amount, in percent, that the measured value can be above the set-point prior to setting
an alarm condition.
3. RH Time Out Timer (Default 30 minutes)
This value determines the amount of time before declaring a RH Low alarm because RH level is not reaching
Set-point.
4. RH Minimum below Set-Point (Default 3.0%)
This function gives the ability to set the value in percent RH below set-point that starts the “Not achieving set-
point timer” described in item 2.
5. Post Inject Delay Time
This value specifies the time, in seconds, that the Rh system delays control to allow detection of the level RH that
was just injected into the chamber.
O₂
1. System Enable (Enabled/disabled) on models NU-5831 and NU-5841
This function will enable or disable the O₂ system. If the system is Enabled it will be displayed in the MAIN screen.
If it is disabled it will not be displayed in the Main orthe Option/Settings screen for the O₂ system:
System Enabled (with check mark) means the System is enabled
System Enabled (without check mark) indicates that the system is disabled
2. Maximum Above Set-Point
This value determines the maximum deviation, measured in (%) that the chamber is permitted to rise above set point, once it is reached before an alarm condition is declared.
3. O₂ Time Out Timer
This value determines the time, in minutes, for the O₂ percentage to achieve set-point. If the O₂ percentage doesn’t get to within its upper limit of set-point within this time period an alarm condition is declared.
4. Minimum below set-point to Start Alarm Timer
This value determines the limit in percent that starts the Alarm Timer for declaring the alarm.
5. N₂ Post Inject Delay Time
This value specifies the time, in seconds, for an injection of N₂ to be measurable at the sensor. When N₂ is injected into the chamber, the system delays until this time period has elapsed before making a new control decision. In
this manner, diffusion induced delays do not cause the O₂ system to overshoot the control set-point.
(To Access this function: press Settings Text in Service Control Screen)
Note: System is disabled on all other models
(To Access this function press Settings Text in Service Control Screen)
Note: System is disabled on all other models
OM0246 Page 34 of 64
Rev 5 June/2020
Page 35
General Options
1. Door Close Delay Time
This value determines the time, in seconds, to increase the door and parameter heater to duty cycle by 10% after
the inner door is closed. Also inhibits gas and RH control for the same period.
2. Air Inject Time
This value specifies the time, in seconds, for an injection of air into the chamber
3. Air Inject Cycle Time
This value specifies the amount of time in minutes between each injection of air into the chamber
4. Digital Output Interval
This parameter specifies the frequency, in minutes that performance and set point data is output to a device such
as a printer that is only capable of receiving data. If the frequency is specified as zero, no data will be output.
5. Door Ajar Alarm Delay
Determines how long in minutes the glass door can be open before an alarm is declared.
7. Settings (text located at the top of screen)
Pressing the next button will go to the second page containing the following:
•The Filter Maintenance button to access:
Resetting the change filter notice
Setting time to reminder (24 months/1/48) 0 disables the feature.
View time remaining until current notice is triggered.
•The Clear Password button that clears the 4 digit password that was input through the Set Password
screen.
•Change language Button: Pressing this button brings up a screen with the language choices of English
(default), Spanish, French & German. Pressing the button in that screen of the language desired will
change the user screens to that language.
•Screen Lock button (default Main Screen Unlocked). Pressing this button will change it to “Main screen
Locked” (Lockout = 1). This puts a “Press to unlock” button on the Nuaire Logo in the Main screen. This
feature stops accidental screen changes by brushing the touch screen. You must press and hold (2
seconds) until the message disappears which unlocks the main screen so that you can navigate into; the
System settings, Alarms, Environmental Set Point and Clock Adjustment screens. Pressing the button
again will change the text to “Main Screen Unlocked (Lockout = 0). This removes the “Press to Unlock”
button from the Nuaire Logo and the Main screen goes back to the default condition.
•Power Loss Mode cancel button. Cancels Power loss mode. Only use this if it is certain that there is no
condensation in the circulation system. See Section 10.2 for further details about the Power Loss Mode.
OM0246 Page 35 of 64
Rev 5 June/2020
Page 36
8.10 Resetting the Electronics Software
8.10.1 Factory Reset
The Factory Reset is similar to the Mast Reset. However, it only changes the Options (returned to default settings)
and the environmental set-points (returned to their defaults).
To perform a Factory Reset, you will need to do the following:
NuAire Logo → Service Settings → 9876 Enter → Reset → Factory Reset → Confirm
The Incubator will perform the Factory Reset and return to the Main screen.
8.10.2 Master Reset
The master reset diagnostic function is the last effort to correct operational faults which otherwise cannot be
solved. By reloading the default configuration, the entire memory will be reset and ALL CALIBRATION OFFSETS,
USER PASSWORD, CONFIGURATION OPTIONS, MODEL NUMBER OF THE INCUBATOR, AND THE SERIAL NUMBER
WILL BE LOST. All calibrations will need to be performed following a master reset.
To perform a Master Reset, you will need to do the following:
NuAire Logo →
Service Settings →
9876 Enter →
Reset →
Master Reset →
Confirm
The Incubator will perform the Master Reset then the “Factory Setup” Screen will appear
The Factory Setup screen shows the Model Number, Serial
Number and Language that was chosen for the user screens.
As long as this information is correct you can press save which
will save the information displayed in the screen then, take you
to the date and time set screen. If the date and time are correct
you can press exit and go to the MAIN screen. If you press Exit
you will go to the main screen however if there is a power cycle
the incubator will reboot to the Factory set up screen again. All
setup information will show correctly but, you must press save to
boot up into the MAIN screen after a power cycle. The model
number and the serial number can be verified by comparing them
to the serial number label found down behind the valence under the door. The language defaults to English. It can
be changed any time through the Service Settings/General/Option/settings menu. See section 8.10 for further
instructions.
Note: After performing either Factory or Master Reset remember to turn the incubator off
then on again to ensure that all intended parameters were reset.
8.10.3 Inputting the Factory settings:
The “Factory Set-up” screen will appear after the master reset. If the Model Number or the Serial Number need to
be re-input for any reason the following steps should be taken:
Setting the model number sets the following function:
•All Nuaire customers will press Nuaire in the “Choose Company” Screen. This will bring up the
Nuaire logo in the main screen. All other customers will press the appropriate company
button to bring up the identifiers specific to that company.
•Model number NU-5800 and NU-5810 and E models
Enables Temperature and CO
Chooses the algorithm suited for low watt heating of the chamber
Supresses the decon button in the “Systems Settings” screen
OM0246 Page 36 of 64
Rev 5 June/2020
the only control systems available on this model
2
Page 37
Loads the model number in the System information screen
•Model numbers NU-5820 through NU-5841 and E or D models
Enables Temperature and CO
plus RH and O2 control systems as available on the various models
2
Chooses the algorithm suited for high watt heating of the chamber suited for heated Decon cycles.
Activates the heated decon cycle choices in the “Systems Settings” screen
•Touch the “Model Number” text then follow the screen prompts as shown below.
Pressing continue on the choose Region screen will bring you back to the Factory Set-Up screen showing
the Model Number you just entered. Default is NU-5800 if no model number is entered
•Touch the Serial Number text then enter the 12 digit serial number which can be found on the serial
number label located behind the valance under the sample port. Press the ENT button on the key pad
screen and you will return to the Factory Set-Up screen showing the serial number you just entered.
Default is blank if no number is entered.
•Touch the Set Language text then touch the save button. This sets the language on all of the user screens.
The language in the Service Settings menu /screen will remain in English. English is default language if
none is chosen. Touch the green save button to exit the factory screen.
•The final change that is requested is the clock. It should not have been reset. As long as the date and
time are correct you can press cancel or save in that screen to exit to the Main screen.
See section 7.8.1 if for instructions to reset date and/or time.
•If the incubator loses power the date/time will flash in green text until it is manually interfaced with. If
the date and time are correct you can cancel and the display will return to the main screen.
•When power to the incubator is lost long enough the real time clock will lose it’s backup power and the
clock will revert to it’s default time. When power is restored to the incubator the date/time will flash in
yellow text and continue until it is reset. Press save to return to the main screen and continue normal
operation.
OM0246 Page 37 of 64
Rev 5 June/2020
Page 38
9.0 Calibration
Proper calibration of the incubator involves following parameters. On all models of the Invitrocell incubator the chamber
temperature, door temperature, perimeter temperature, and CO
should be checked on a regular schedule. The first three, chamber, door, and perimeter temperature should be completed and
stabilized before any CO
detail. For the best results, follow the procedure carefully, and if the desired result is not achieved, try procedure again from
the start.
9.1 Chamber Temperature Calibration
The DH’s TEMPERATURE CALIBRATION MUST BE PERFORMED WITHIN 1°C OF THE PLANNED OPERATING TEMPERATURE.
Normally, 37.0°C is the most common set point. To initiate the procedure, turn on the incubator via the power switch on
the back panel. The default set point is 37.0°C; enter “Environmental Set Points” and adjust if desired Let the unit stabilize
for 8 to 12 hours. Use an independent calibrated instrument to check and calibrate the temperature.
Below is a description of calibration with a glass thermometer as one example for temperature calibration.
At the beginning of this procedure, set a calibrated glass thermometer in a glass beaker filled with water resting on a shelf
in the middle of the chamber. Do not place the glass beaker on the bottom of the chamber because it will result in a slightly
higher temperature reading due to the heater located on the chamber bottom. Placing the thermometer in glass beaker on
the middle shelf will give the most accurate results for calibration. The chamber should be humidified to avoid false low
readings due to evaporation of water from the flask. An accurate digital thermometer with a type K thermal couple could
also be used. No matter what instrument is used the measurement for this calibration should be taken in the middle of the
incubator chamber and in a stabilized condition.
sensor calibration is performed. On Invitrocell Below, each calibration procedure is described in
2
sensor require calibration at installation and the calibration
2
When the unit has stabilized at the operating temperature, perform the following calibration procedure.
• Allow Incubator to stabilize at its given temperature set point in run mode.
• Enter “Service Settings” by means of your provided password.
• Press the Temperature Calibration button.
• Determine actual temperature within chamber by reading temperature measurement instrument.
• Select the number digits below the “Current Calibrated Sensor Reading” blue text button
• Input the current value displayed on your temperature measurement instrument and exit calibration.
Note: THE DISPLAY DOES NOT UPDATE IMMEDIATELY; IT TAKES 2-3 SECONDS
The chamber temperature calibration is complete. Let unit stabilize for 8 to 12 hours. If the chamber temperature (actual
thermometer) still does not match the display, perform the above procedure again. In some cases it might be necessary to
calibrate several times to achieve a stable condition due to ambient conditions of temperature and humidity within the
laboratory.
OM0246 Page 38 of 64
Rev 5 June/2020
Page 39
9.2 Door and Perimeter Duty Cycle adjustments for Temperature Calibration
Location of condensation
Adjustment
On the inner glass door
Increase the door heater duty cycle in increments no
greater than 5% at a time.
On the inner door gasket
Increase the perimeter heater duty cycle in increments no
greater than 5% at a time.
In the back of the chamber
First decrease the door heater in increments of no greater
heater in the same manor until condensation stops form.
The Door and Perimeter heaters are run on “Base Duty Cycle” calculations that are automatically increased or decreased to
run in accordance with the chamber temperature at a chosen set point. These duty cycles work in the controlled lab
environment but due to the many variables involved in temperature control there is also a manual adjustment. The heater
duty cycles can be adjusted to control condensation or to improve the temperature uniformity in the chamber.
The inner glass door and outer shell perimeter temperature calibration is best accomplished by running the Incubator for
a minimum of 24 hours with the water pan in place and perform the following calibration sequence, if required. Open the
outer Incubator door and look for general condensation. Some condensation on the glass door can be desirable as an
indication of adequate humidity in the chamber. Typically, one to two inches of condensation in the corners of the glass
door and/or gasket indicates a properly calibrated door and perimeter heater. The perimeter heater is located on the outer
shell next to the glass door gasket. If calibration is required, simply perform the procedure as stated below. As stated
earlier the door and perimeter heaters operate on a base duty cycle calculated for the chamber temperature set point
being used. In the door and perimeter heater (adjustment) screen “0.0” = the calculated base duty cycle. Adjusting the
value up or down will increase or decrease the percent of power to each heater in 0.1% increments. Small adjustments are
recommended to a maximum of 5.0% at a time.
The following steps should be taken for setting these duty cycle percentages:
• Allow Incubator to stabilize at its given temperature and humidity level in run mode.
• Enter the “Environmental Set Points” screen either from the MAIN SCREEN or the SYSTEM SETTINGS screen.
• Select the numerical digits following the “Door” or “Perimeter” text. Go to the adjustment screen.
• Press UP or DOWN keys to increase or decrease the numerical value to the desired level. The limits the settings
can be changed to are +50 and – 20.
• Press “Done” to set current value and exit calibration.
• The reset button for both the door and the perimeter heaters resets the number displayed to 0.0 removing
any adjustment previously made.
Adjusting Door and perimeter heater duty cycles to control condensation:
If condensation persists see section 9.3 Setting air injections.
OM0246 Page 39 of 64
Rev 5 June/2020
than 5%. If condensation persists and condensation starts
forming on the door then move to reducing the perimeter
Page 40
9.3 Setting Air InjectionsIf there is still some undesired condensation in the chamber when the door and perimeter heaters are set for the desired
result, the air injections can be adjusted. There is a control for length of the air injection labeled, Air Inject Time, and the
frequency that air is injected called, Air Inject Cycle. These controls are described in more detail in the "General Options"
section of the Service Settings menu. The default is 10-second injections every 1 minute. Start by increasing the length of
the injection by a few seconds at a time then increase the frequency if needed. If the humidity in the chamber is less than
desired, reduce the Air Inject Time.
Air and OWhen the O
interfere with achieving the lower O
Levels
2
set point is set below 10% the “air inj time” is automatically reduced to 0(zero) to stop air injects which will
2
set points.
2
9.4 CO
The DHD infrared CO
Calibration
2
and CO
sensor may be calibrated using one of four techniques: Open Door CO2, Closed Door CO2, Off-Set,
2
injection calibrations. CO2 calibrations can be performed in approximately ten minutes, by accessing the CO2
2
calibrations menu in “Service Settings” and following the on screen instructions. To perform Off-Set, Open Door, and Closed
Door calibrations attach an independent CO
measurement device to the Sample Port in the front of the unit before the
2
calibration begins.
9.4.1 CO
Off Set Calibration
2
CO2 Off Set Calibration can be performed anytime an independent measurement doesn't correlate to the front panel
display. However, this calibration SHOULD NOT BE PERFORMED MORE THAN ONCE PER WEEK. Sensor calibration
should be performed if an independent measurement, if it doesn't match the display within +0.3 percent within one
week after a sensor calibration. Before doing the following calibration, check and change, if necessary, the Incubator
in-line filter found within the Sensor bay.
When unit has stabilized at the operational temperature and CO
percentage, take an independent measurement
2
and, if necessary, perform the following:
• In “Service Settings” enter the CO
• Press the “CO
Offset Calibration” text button
2
•Use an independent instrument to determine actual CO
Calibrations menu.
2
percentage (compare the display CO2 to the
2
independent measurement). If these two readings have a difference of equal to or less than 0.5 percent,
proceed to enter the CO
below) If the difference is greater than 0.5 percent, it is recommended to perform a CO
Fyrite value. This calibration only allows a maximum of 0.5% offset. (See * Note
2
Open Door
2
Calibration.
• Enter the value measured by the independent device
• Exit and Save to retain this value.
*Note: When the display value is more that 0.3% different from the measured value, offset display
1/2 the difference measured. Allow the Incubator to stabilize back to set point, then measure
the CO
in the chamber again. Offset the display again if necessary.
2
9.4.2 CO
Sensor Calibration (Zero/Span/CAL. INJ.)
2
There are 2 sensor (zero/span) calibration routines available to the lab professional. The first option is the “open
door” routine involving opening the outer and inner door to zero the sensor. This routine also automatically
calibrates to the CO
injection rate during the injection for the span portion of the sensor calibration. It is
2
recommended that this routine be used during the initial setup of the Incubator, if the set point of the system is
changed or if other changes are made on the Incubator affecting the CO
system.
2
The second option is a “closed door” routine. This routine allows calibration of the sensor without opening the door
avoiding undue exposure to the cultures that may be in process. This routine injects “fresh air” into the detector
head of the sensor to calibrate zero. The chamber air is then allowed back into the detector head to calibrate the gas
span that is detected.
OM0246 Page 40 of 64
Rev 5 June/2020
Page 41
9.4.3 Open Door CO
• Attach independent CO
• Enter CO
• Select the “CO
Sensor Calibration Routine
2
monitoring device to the sample port on the front of the unit
2
calibrations by means of Service Settings
2
Open Door Zero Span” text
2
• Follow on screen instructions
The CO
unnecessary to run the separate “Injection Calibration”. The injection is dependent on the proper gas pressure and
factory set flow rate. Any changes in either will result in a change to the value reached during this injection.
9.4.4 Closed Door CO
inject rate is automatically calculated from the CO2 injection made during the span calibration making it
2
Sensor Calibration Routine
2
• Attach independent CO
• Enter CO
• Select the “CO
calibrations by means of Service Settings
2
Closed Door Zero Span” text
2
monitoring device to the sample port on the front of the unit
2
• Follow on screen instructions
Enter measurement on independent device when prompted
Allow unit to run and stabilize for a minimum of 2 hours then, check calibration with an independent instrument.
Compare the display CO
greater than 0.3%, repeat above procedure. If these two readings have a difference of less than 0.3%, perform the
Offset calibration procedure in Section 9.4.1.
CO
2
9.4.5 CO
Injection Calibration
2
The CO
injection calibration can be performed separately from zero/span calibration to optimize the gas injection
2
time required to recover the CO
possible with virtually no overshoot. Verify the CO
The injection calibration is not required after an “Open Door” sensor calibration since it is performed automatically
during this routine. The injection calibration should be performed after a “Closed Door” sensor calibration when
possible, if CO
supply pressure of the Incubator is changed, or is the CO2 flow control valve is disturbed.
2
The following steps should be taken for the CO
• Enter CO
Calibrations by means of the Service Settings menu
2
• Select the “CO
• Follow the on screen instructions
9.5 Relative Humidity sensor calibration - System available on Models NU-5820 and NU-5841
Relative humidity calibration can be performed anytime if the relative humidity system has been installed. The relative
humidity sensor can be calibrated from a known source of humidity within the Incubator chamber. Typically, the water pan
is used because it has a known minimum humidity level of 96 percent after 12 hours. A hygrometer can also be used to
calibrate the chamber Relative Humidity. See the following instructions for both methods of calibration.
9.5.1 Calibrating with a water pan (included with the Incubator)
As stated in the introduction the relative humidity sensor can be calibrated from a known source of humidity within
the Incubator chamber. Typically, the water pan is used because it has a known minimum humidity level of 96
percent after 12 hours when the Air injections are deactivated. This calibration is done with the O
to disable N
injections during the procedure.
2
To use the water pan
• Access the Environmental Set points screen to set the RH set point to zero if not set there already.
• Go to the Service Settings/General/Options button and change the Air Inject time from 5 (sec) to 0 (sec).
This will stop air injections which is necessary to reach RH saturation of the chamber.
•After 12 hours refer to 9.5.3 Display Offset Calibration section for instructions on how to adjust the display
displayed value for RH.
• Set the calibration to 96% RH
• Remember to reset the air inject time to 5(sec.) after you complete the calibration and choose your RH set
point above zero.
Note: Some condensation can be expected in the chamber at this humidity level with the air injections turned off.
OM0246 Page 41 of 64
Rev 5 June/2020
percent to your independent measurement. If these two readings have a difference
2
level to set point after a door opening. The recovery time should be as minimal as
2
sensor calibration prior to performing an injection calibration.
2
injection calibration:
2
Injection Time” text
2
set point at 21%
2
Page 42
9.5.2 Calibrating with a Hygrometer
When unit chamber has stabilized at the operational temperature, CO
water in the tank the heater has been powered for at least 60 minutes ensuring it is up to temperature. Allow the
RH system to achieve set point and stabilize there for at least 30 minutes. Then a measurement with an independent
hygrometer probe can be made It is usually inserted into the chamber through the chamber access port located on
the right side of the incubator. The Hygrometer must be up left undisturbed until it indicates that it is up to the
chamber temperature or until the temperature on the hygrometer stops increasing for at least 5 minutes. Take a
reading then move to the Display Offset Calibration described in the next section to adjust the RH value for the
incubator system.
ACCURACY OF CALIBRATION W/ THE HYGROMETER:
The hygrometer used to calibrate the Incubator sensor will have a tolerance that is specified by the manufacturer of
the instrument. The tolerance of accurate hygrometers can be as tight as + 2% others can have tolerances of + 3% or
greater at this humidity level. This means that when 90% is read on the instrument the actual relative humidity
percentage can be at the high or low limit of the hygrometer’s tolerance. If condensation is encountered the value
that the incubator displays can be adjusted accordingly.
9.5.3Display Offset Calibration:
Pressing the Calibrate button on the Service settings screen produces the
Calibrate the RH offset screen as show to the right. Press the Calibrated
sensor reading value (blue text) to access the key pad screen then, to
input the independently measured value and press the ENT button on the
key pad. This returns you to the “Calibrate RH offset” screen. I might take
a couple of seconds for the new value to appear in this screen. Then,
press the SAVE button to return to the “Service settings screen”. Pressing
the exit button in that screen will return you to the MAIN screen.
NOTES:
1. When you use the Water pan method of calibration remember to set the “Air Inject Time” back to 5 (sec).
Otherwise Condensation is likely to form in the chamber. That Option is accessed through Service
settings/General/Options button.
2. Choose the desired set-point that you want to run the system at in the “Environmental Set Point screen
remembering that it was set to 0 (zero) for the calibration.
Calibration – System available on Models NU-5831 and NU-5841
9.6 O
2
When unit chamber has stabilized at the operational temperature, CO
least 10 minutes ensuring the sensor is up to temperature - take an O
through the chamber sample port used for taking CO
readings. It is located behind the cover under the incubator outer
2
door and, if necessary, perform the following:
NOTE: If a Fyrite instrument is used,
BE SURE TO READ THE O
FOR PROPER O
MEASUREMENTS AND SAFETY CONSIDERATIONS.
2
FYRITE OPERATING AND SAFETY INSTRUCTIONS
2
Display Offset Calibration:
Press the Calibrate button on the Service settings screen produces the Calibrate the O
sensor reading value to access the key pad to input the independently measured value then press the ENT button on the
key pad. When you return the Calibrate O
offset screen press the save button to return to the System settings screen.
2
Press the exit button in this screen to go back to the main screen.
%, the RH set point has been chosen and with
2
%, and RH%, O2 system has been powered for at
2
measurement with an independent instrument
2
offset screen. Press the Calibrated
2
OM0246 Page 42 of 64
Rev 5 June/2020
Page 43
10.0 Maintaining Your DHD
DHD Chamber
The chamber maintenance is up to the discretion of the owner and the extent of cleanliness and sterility desired.
The shelves and bracket supports are all removable and autoclavable.
The interior should be wiped down with an appropriate disinfectant such as 70% ISOPROPYL ALCOHOL or equivalent.
Note: DO NOT USE ANY CHLORINATED OR HALOGEN MATERIALS IN THE CHAMBER.
SUCH MATERIAL IS HARMFUL TO THE POLISHED STAINLESS STEEL.
NOTE: NuAire does not offer any product warranty with respect to cleaning material compatibility. USE AT YOUR
OWN RISK! The information provided above is from raw material suppliers and known general source documents for use
to develop application cleaning SOP’s.
The humidity pan should also be sterilized and the water changed regularly to assure sterility.
A small amount of copper sulfate may be added to the humidity pan to inhibit bacterial growth.
/ N2 Supply Filter
CO
2
P/N X-980385 (50 mm Disk, Uni-directional In-Line, Dry)*
The CO
Note: Direction of flow (IN is labeled on one side of filter) when replacing filter.
Sensor Bay Capsule Filter P/N RF566 (Capsule, Uni-Directional In-Line, Wet)
This filter should be replaced EVERY TWO YEARS to assure optimum performance.
A visual check should be performed when maintenance reminder appears in the main screen to assure filter integrity.
Remove sensor bay cover to perform visual check.
Air Supply Filter P/N X-980385 (50 mm Disk, Uni-directional In-Line, Dry)*
The Air Supply Filter should be replaced every fifth empty CO
Note: direction of flow (IN is labeled on one side of filter) when replacing filter.
CO
P/N X-980366 (50mm Disk, Uni-Directional In-Line, Wet)*
The filter should be changed when discolored (yellow brown).
The filter is plumbed in the chamber sample hose and can be inspected when the cover is removed to check the air pump filter.
This filter has a green dot to distinguish it from the "dry" filter.
OM0246 Page 43 of 64
Rev 5 June/2020
/ N2 Supply Filter should be replaced every fifth empty CO2 tank or when the filter is visibly discolored (yellow-brown).
2
tank or when the filter is visibly discolored (yellow-brown).
2
Sensor Filter
2
* Note: The word "In" on the outer rink of the body indicates the inlet side of the filter
and should be installed toward the gas supply.
Page 44
Back to
System Settings
Back to
Main
Tough Decon Button
10.1 Heated Decon Cycles
Heated Decontamination Cycle Preparation and use Procedures for Incubator models:
NU-5810, NU-5820, NU-5831and NU-5841
IMPORTANT
READ AND UNDERSTAND ALL OF THE ATTACHED INSTRUCTIONS BEFORE USING
THE HEATED DECONTAMINATION CYCLES
NuAire gives the lab professional the choice of the 2 most commonly accepted, heated decontamination cycles available.
They are a 95°C humidified cycle and a 145°C dry cycle. To choose, the user goes to the System Settings menu and
pushes the Decon button in the lower right hand of the display. From there it will ask which Decon (95 or 145), and
proceed to instruct the user in preparation of the chamber.
Choosing the 145° DECON CYCLE
The 145
°C decontamination cycle is high temperature dry cycle to ensure that the contaminating agents are eradicated.
The complete cycle takes less than 8 hours to run. (See graph for cycle phases). There is no need to recalibrate or
remove the CO
IR sensor since it is not located in the chamber. See the instructions above for choosing this cycle in the
2
System Settings Menu.
OUTER DOOR MUST REMAIN CLOSED DURING THE DOCONTAMINATION CYCLE OR IF CHAMBER
TEMPERATURE IS OVER 55
°.
THE DECONTAMINATION CYCLE WILL HEAT THE INCUBATOR INTERIOR SURFACES TO 145° C +.
CONTACT WITH ANY SURFACE INSIDE OF THE OUTER DOOR DURING THIS CYCLE CAN RESULT IN BURNS.
PREPARATION
Following the prompts in the incubator display screen you will:
• Remove culture cells, samples, dishes, instruments, or other user-introduced equipment from the chamber.
• Wiping out all spills and materials from chamber walls, shelves, plenums, & inner door may be required to
avoid stains, odor and material baked on these surfaces during the cycle.
• IF a cleaner is used make sure it is compatible with the construction of the chamber (stainless steel).
• Ensure the hole in the access port plug at the side of the Incubator is open.
This is a relief for the chamber during the Decon Cycle.
•Models NU-5820 and NU-5841 remove the RH sensor from upper left side of the back wall of the chamber. The
sensor will be damaged if it is not removed. The sensor cable is made to withstand the heat of a decon cycle
and is left in place.
OM0246 Page 44 of 64
Rev 5 June/2020
Page 45
145
10:00:00
TEMPERATURE DEG C
37
FOLLOW ALL FEDERAL, STATE AND LOCAL REGULATIONS THAT APPLY WHEN USING CLEANER. ANY
RESIDUE LEFT OF THE CLEANER WILL BE EXPOSED TO THE DECON. TEMPERATURES CHECK WITH
MANUFACTURER ABOUT EXPOSURE OF THE CLEANER TO THESE TEMPERATURES.
DRY 145°C DECON CYCLE
Note:
• CO
• Protection of the CO
• RH sensor must be removed from all models equipped with RH control the as stated in the Preparation section and
• The Decon Cycle will not advance to the heat up phase unless the door is opened.
• Follow the on screen instructions, pressing the Next button when each step is completed
Start the decon cycle by following the prompts on the NuTouch display screen.
The decon cycle will automatically go through the following phases:
HEAT UP 2 hours on average
either schedule a start time or press the
DECONTAMINATION 3 hours
temperature is reached.
blowing HEPA filtered air to cool the chamber to the user chosen temperature set point.
TIME
RH and O₂ control systems will be idle during the cycle.
2,
1:00:00
2:00:00
and O₂ sensors during the heated cycle is automatic.
2
3:00:00
4:00:00
TIME (HRS:MIN:SEC)
5:00:00
6:00:00
7:00:00
8:00:00
9:00:00
in the NuTouch screen prompts that show during the startup of the cycle. The sensor will be damaged if it is not
removed.
Complete the chamber preparation and
Start button to start the heat up portion of
the cycle.
•Temperature display shows chamber
temperature.
•Text displays that the unit is currently in the
Heating Up cycle
•A timer tracks how long the unit has been
heating thus far
The unit automatically starts a countdown timer for the correct length of cycle when the decontamination
• Temperature display shows chamber temperature.
• Text displays that the unit is currently in the Decontamination cycle.
COOL DOWN & STABILIZING 4 hours
When the decontamination period is complete the heaters are shut off and the air pump is turned on
OM0246 Page 45 of 64
Rev 5 June/2020
Page 46
• Temperature display shows chamber temperature during cycle.
!
• Text displays that the unit is currently in the Cooling Down portion of the DECON cycle
• A count up timer (incrementing time) tracks how long the unit has been cooling down
CYCLE COMPLETE
Indicates the user should:
• Refill the water pan with single distilled water no purer than 1 Mega OHM in preparation to return to normal
running.
Note: If Decon Cycle is run as routine maintenance, it is ok to reuse the HEPA filters
for the service life indicated in Section 10.
If contamination is an issue due to the room environment, HEPA filters can be replaced as necessary.
RESUME NORMAL OPERATION
Press the Exit button after fulfilling the on screen instructions
• This puts the Incubator into normal run mode
• CO
• Checking the calibration of the temperature sensor is recommended.
Follow the instructions in section 9.1 of the manual
DECON CYCLE NOTES:
control resumes
2
• Aborting the cycle:
To abort any phase of either the 95 or 145 decontamination cycles press the cancel button to step through the
different phases of the cycle that are described above. The cool down phase cannot be bypassed until
temperature reduces to the temperature set point. See cautions below.
• In case of a power interruption, the heat up, drying & cool down phases will resume when power is restored.
The decon phase will reset to the beginning of the 3-hour cycle or if the temperature drops more than 10° it will
go back to the heat up phase then resume the cycle when decon temperature is reached.
• It is considered normal for some odor to occur during the Decon Cycles.
This odor reduces with use of the Decon Cycle.
• Some discoloration due to heat of the chamber and its components can be expected.
• If decon temperature is not reached within the time allotted (set by the temp. Time Out option).
The Incubator will alarm and indicate an Alarm Status condition in the display.
• If the inner glass door is opened during the Decon Cycle an alarm sounds and the Decon Door Open Alarm will
overtake the display telling the user to close the door.
• Place Incubators 10" apart when placed side-by-side to enable running the heated Decon Cycle while
surrounding Incubators are in normal run mode.
USING THE 95° DECON CYCLE
The 95° C decontamination cycle is humidified to ensure that the contaminating agents are eradicated. The complete
cycle takes about 14 hours to run. (See graph for cycle phases). There is no need to recalibrate or remove the CO
sensor since it is remote from the chamber.
OUTER DOOR MUST REMAIN CLOSED DURING THE DOCONTAMINATION CYCLE OR
IF TEMPERATURE IS OVER 55°.
THE DECONTAMINATION CYCLE WILL HEAT THE INCUBATOR INTERIOR SURFACES TO 95° C +.
CONTACT WITH ANY SURFACE INSIDE OF THE OUTER DOOR DURING THIS CYCLE CAN RESULT IN BURNS.
OM0246 Page 46 of 64
Rev 5 June/2020
2
IR
Page 47
HUMIDIFIED 95 DEG. C DECON CYCLE
37
95
0:00:00
1:00:00
2:00:00
3:00:00
4:00:00
5:00:00
6:00:00
7:00:00
8:00:00
9:00:00
10:00:00
11:00:00
12:00:00
13:00:00
14:00:00
TIME (HR:MN:SEC)
TEMPERATURE DEG. C / RH %
TEMPERATURE
RELA TIVE HUMIDITY
Decon Phase
Drying Phase
!
PREPARATION
Following the prompts in the incubator display screen you will:
• Remove culture cells, samples, dishes, instruments, or other user-introduced equipment from the chamber.
• Wipe out all spills and materials from chamber walls, shelves, and plenums using a disinfectant of choice that is
compatible with construction of the Incubator chamber and instrumentation.
•Ensure the hole in the access port plug at the side of the Incubator is open.
This is a relief for the chamber during the Decon Cycle.
•Ensure that the sample port in the front panel of the Incubator is capped.
Condensation can form in tubing if it is left open.
•Models NU-5820 and NU-5841 remove the RH sensor from upper left side of the back wall of the chamber. The
sensor will be damaged if it is not removed. The sensor cable is made to withstand the heat of a decon cycle
and is left in place.
•Empty water pan then clean it with a disinfectant then, fill with 300ml of water. Place it back in the chamber on
the highest shelf for the decon cycle.
FOLLOW ALL FEDERAL, STATE AND LOCAL REGULATIONS THAT APPLY
TO THE DISINFECTANT USED TO CLEAN THE CHAMBER.
Note:
• CO
• Protection of the CO
RH and O₂ systems will be idle during the cycle.
2,
and O₂ sensors during the heated cycle is automatic.
2
• RH sensor must be removed from all models equipped with RH control the as stated in the Preparation section
and in the NuTouch screen prompts that show during the startup of the cycle. The sensor will be damaged if it is
not removed.
• The Decon Cycle will not advance to the heat up phase unless the door is opened so that the NuTouch screen
prompts to empty the water pan, clean the chamber and remove the RH sensor are followed.
OM0246 Page 47 of 64
Rev 5 June/2020
Page 48
•Follow the on screen instructions, pressing the Next button when each step is completed
Start the decon cycle by following the prompts on the NuTouch display screen.
The decon cycle will automatically go through the following phases:
HEAT UP 1 hour
Complete the chamber preparation and either schedule a start time or press the Start button to start the heat up portion of the cycle.
• Temperature display shows chamber temperature.
• Text displays that the unit is currently in the Heating Up cycle
• A timer tracks how long the unit has been heating thus far
DECONTAMINATION 9 hours
The unit automatically starts a timer for the correct length of cycle when the decontamination temperature is
reached.
• Temperature display shows chamber temperature.
• The Text displays that the Decontamination cycle is in progress
• A timer tracks the remaining time in Decontamination
HUMIDITY REDUCTION 1 hour
The unit remains at the decontamination temperature and the air pump is turned on to reduce the humidity left
from the decontamination phase.
• Temperature display shows chamber temperature during cycle.
• The Text displays that the Decontamination cycle is in progress
• A timer tracks the remaining time in Decontamination.
COOL DOWN 3 hours
When the humidity reduction period is complete the heaters are shut off to cool the chamber to the user chosen
temperature set point.
• Temperature display shows chamber temperature during cycle.
• The Text displays that the unit is now in the Cooling Down cycle
• A timer tracks how long the unit has been cooling
CYCLE COMPLETE
Indicates the user should:
•Refill the water pan with single distilled water no purer than 1 Mega OHM in preparation to return to normal
running.
Note: If Decon Cycle is run as routine maintenance, it is ok to reuse the HEPA filters for the service life indicated in
Section 10. If contamination is an issue due to the room environment, HEPA filters can be replaced as necessary.
RESUME NORMAL OPERATION
Press the Exit button after fulfilling the on screen instructions.
• This puts the Incubator into normal run mode
• CO
control resumes
2
• Checking the calibration of the temperature sensor is recommended.
• Follow the instructions in section 9.1 of the manual
OM0246 Page 48 of 64
Rev 5 June/2020
Page 49
10.2 Chemical Decontamination of the Incubator Chamber
To chemically decontaminate NuAire Incubator inner chambers, users may use the traditional formaldehyde or Chlorine
Dioxide in gas form, or Vapor Phased Hydrogen Peroxide. All three of the chemicals are compatible to all parts within
NuAire Incubator chambers. The incubator should be turned off during these procedures
Note: As stated previously, the chamber and components can also be wiped down with a
70% solution of Isopropyl Alcohol for cleaning and decontamination.
10.3 Power Loss Recovery Mode
When an incubator experiences a sudden loss of power there is an increased chance of condensation building up in the
circulation system tubing. This mode helps to dissipate that buildup before it causes the sensors to fail
- Conditions for entering mode: - Incubator must be without power for 45 minutes or longer. This includes first time
setup. Mode is automatically triggered after power is restored (after 45+ min outage). If less than 45 minutes have
passed incubator will enter normal Run mode.
- When in Power Loss Recovery Mode:
o Blocking solenoid will be turned “ON” which will bypass the CO2and O2 sensors in the circulation system.
o Both systems will be put in standby mode meaning they will not control or alarm while in this mode.
o “Power Loss Recovery” shows in place of “RUN” in the main screen.
o Heaters and Air Pump run as normal.
- Once entered, this mode will run for 1 hour. A count down timer screen (seen below) can be found when pressing
the CO2 or O2 Diagnostics button in the Service Settings screen. It is also recorded in the event log.
10.4 Shutting down the Incubator
Prior to shutting down the Incubator open the inner and outer doors and remove the water pan. Leave doors open for at
least 5 minutes prior to shutting it off. This will purge the chamber, circulating system and the sensors of humidity that
could condense and cause faulty readings when the Incubator is turned back on. Be sure to empty the water pan prior to
putting it back into the chamber if the Incubator is going to be shut off for any length of time.
10.5 Automatic Filter Change Notification
The Capsule filter located in the Sensor bay (see BCD-18319 for location) should
be checked average every 2 years to see if it should be changed. The Filter
material is normally a white material. Discoloration of the filter material is an
indicator that it should be changed. Retro Fit Kit RF566 should be ordered for the
replacement filter. A timed notice triggers a Maintenance Notification button in
the Main Screen when this timer counts down to zero. Pressing the Maintenance
Required button produces a screen indicating that the timer for filter life has
expired.
OM0246 Page 49 of 64
Rev 5 June/2020
Page 50
Responding to the notification:
From the main screen press the NuAire Logo / Service settings / General Option /
Settings / Filter Maintenance to access the Filter Maintenance Adjustments screen.
In this screen you can:
•Extend the alarm by 1 month by pushing the Extend. Press the Exit button
to retain the extension and go the General Settings screen. Pressing Save
will not retain the time extension.
•Reset the notification timer to the default of 24 months by pressing save.
This is recommended.
•Change the timer by sliding the green button right (to decrease) or left (to
increase) the amount of time until the next notification. The number of months showing will change as the
button is moved.
•Press the Save button to save the timer setting and go the General Setting screen. Pressing the Exit button will
not save the new timer setting.
•Setting the timer to 0 months deactivates the notification.
After extending or resetting the timer you can return to the main screen by following the screen prompts to save your
choice and exit the menu.
OM0246 Page 50 of 64
Rev 5 June/2020
Page 51
11.0 Error Indicators & Troubleshooting
Step 1 NOTE ALL ERROR INDICATORS
When the Incubator is running, the Alarm Status button indicates an error.
Pressing the Alarm Status button, entering the correlating Alarms Menu and pressing the Silence button in the
Alarms Menu will silence the audible alarm until Diagnostics is entered or another alarm becomes active.
Step 2 CLEAR ERROR INDICATORS.
Error indicators can be cleared by pressing the Alarm Status button and entering the Alarms Menu and pressing the
Clear All button; if the error is still present the indicator will return when the alarm time reaches 0.
Step 3 MONITOR REOCCURRENCE OF ERROR INDICATORS.
If reoccurrence of the error indicator is immediate or daily, use the Error Indicator Troubleshooting Guide on the
following pages to address the issue indicated by the reoccurring alarm.
OM0246 Page 51 of 64
Rev 5 June/2020
Page 52
Temperature System
Displayed Error Code
Code Description
Checks and Corrections
Temp Over Set Point,
Decon Temp High
-Temperature over set point
Normal mode and Decon Cycle
Check temperature sensor calibration.
Faulty TRIAC, replace control board.
-Temperature time out error
for 240 minutes
Check temperature sensor calibration
Door/Perimeter heater needs to be increased with a high
temperature set-point in a low ambient temperature
-Sensor temperature (differential)
sensors exceeds 4°C.
Check temperature sensor calibration
-Sensor temperature (differential)
sensors exceeds 10° C.
Check temperature sensor calibration
CO2 Sensor Shroud Heater
-Defective shroud temperature
sensor
Confirm it is plugged in at board connector P6.
Replace CO2 shroud temperature sensor
Check control board output to relay, replace board if
Check sensor bay heater for function, replace if faulty.
Check control board output to relay, replace board if
bay.
CO2 System
Displayed Error Code
Code Description
Checks and Corrections
Perform CO2 sensor calibration
Check sensor and disk filter for condensation
Check CO2 gas supply - inline gas filters, CO2 gas tank
Check connection on control board
One or both temp sensors faulty, replace
Check connection on control board
One or both temp sensors faulty, replace
Displayed Error Code Code Description Checks and Corrections
Temp Sensor Failure
Sensor Bay Temp Low
Sensor bay is below temp set point
+2⁰ - See CO
diagnostics screen
2
faulty.
Check that all connections are secure
Sensor Bay Temp High Sensor bay temp is above 65 deg C
CO2 Over Set Point - CO2 over set point
time out error. The measured
- CO
2
CO2 level has not reached with in the
CO2 Under Set Point
allowed percentage in the allowed
time. CO2 stops injecting and air is
injected for 15 minutes in case the
sensor is not detecting the CO2.
Cal Inject Failed
OM0246 Page 52 of 64
Rev 5 June/2020
Not enough increase in the CO2
faulty.
Check that all connections are secure
Check sensor bay heater for function, replace if faulty.
Check for uncontrolled sources of heat influencing the
Check injection solenoid for leaking valve
pressure, CO2 sensor function
Run Cal Inj. Calibration (see Section 9.3.3)
Check/replace CO
gas supply tanks
2
Check for leaks in chamber
- inner door gasket, chamber recirculating tubes
Call NuAire Technical Services if error persists.
Page 53
RH System
Displayed Error Code
Code Description
Checks and Corrections
Perform RH sensor calibration
moisture
Perform RH sensor calibration
Check sensor and inject solenoid for loose connections
Check for loose connections on the tank heater
Check the tank heater for function replace if faulty.
O2 System
Perform O2 sensor calibration
Check sensor and disk filter for condensation
Check O2 gas supply - inline gas filters, CO2 gas tank
Check for leaks in air pump and hosing
RH Level High
- RH over set point beyond the alarm
limit
Check injection solenoid for leaking valve
Check sensor & moisture barrier in sensor cap for
Check RH tank for water level if low check Float switch in
RH Level Low
- RH Level is below the set point
beyond the alarm limit and the timer
has counted down to zero
tank for function if Maintenance reminder does not
appear in the MAIN screen
Check for leaks in chamber
Check for leaks in air pump and hosing
Evaporator Temp low
-Water temperature in the water
tank is below the alarm limit for 120
minutes
Check for loose connections on the tank temperature
sensor
Displayed Error Code Code Description Checks and Corrections
O2 Under Set Point - O2 under set point
Check injection solenoid for leaking valve
pressure, O2 sensor function
O2 Over Set Point - O2 time out error
Check/replace O
Check for leaks in chamber
gas supply tanks
2
- inner door gasket, chamber blower fan shaft seal
OM0246 Page 53 of 64
Rev 5 June/2020
Page 54
General Indicators
- Inner glass door is not closed or
the control board.
Close and latch inner glass door.
Check door switch, if faulty, replace.
“Power Interuption” Text
instead of the NuAire Logo
Flashing Green:
return the date and time to green text.
Shows when incubator has lost power for more than 45
dissipate. Gas control is then resumed.
Corrupted memory was detected
during boot up of the incubator
Replace Control Board
A change in parameters failed to write
to memory
Replace Control board
Check output @ Power Supply-If good could be Control Bd
DC power supply failed - replace
Check output @ Power Supply-If good could be Control Bd
DC power supply failed - replace
Check output @ Power Supply-If good could be Control Bd
DC power supply failed - replace
Check for loose line voltage connection to control board
Replace Control Board
Memory stick no compatible try another one. May take
USB port not functioning contact Nuaire technical service
Increase Door heater duty cycle in 5% increments or less
at a time
On inner door gasket and/or front of
the chamber
Increase Perimeter heater duty cycle in 5% increments or
less.
Decrease door then perimeter heater duty cycles in 5%
increments or less.
Note: Depending on the room ambient that the Incubator is operating, the operator may have to readjust the door and
perimeter
Displayed Error Code Code Description Checks and Corrections
Door Ajar
switch not making connection with
- Interruption of power Press the power interruption text to dismiss the notice.
date/time flashing - Interruption of power
Power Loss Recovery Mode
Displayed in main screen in place
of RUN or STANDBY.
Corrupted Memory
Data Write Failure
Confirm connection on the Control Board
Indicates that the real time clock did not lose power. See
section 8.7.1.1 to address this. Canceling or saving in the
clock set screen will return to the main screen and stop the
flashing.
Flashing Yellow
Indicates the real time clock lost backup power and needs to
be set again. See section 8.7.1.1 to address this. Save must
be pressed in the clock set screen to stop the flashing and
minutes long enough to get condensation in the circulation
system and cause the CO2 and O2 sensors to malfunction.
This mode blocks circulation through the sensors and does
not control the gas for 1 hour to allow the condensation to
+ 12 VDC Power Failure
- 12 VDC Power Failure
5 VDC Power Failure
Zero Cross Failure
Universal Serial Bus Error
Condensation Excessive
Low voltage detected on the 12 VDC
power circuit
Low voltage detected on the 12 VDC
power circuit
Low voltage detected on the 5 VDC
power circuit
The Power Triacs control circuit not
working
Stops trying to write to USB memory
stick after several tries
On Glass Door
(Using Water Pan)
See section 9 for more
detailed instructions
On any wall toward back of chamber
Check power supply connection to Control board
Check power supply connection to Control board
Check power supply connection to Control board
Replace fuse with slow blow fuse of the correct amp
rating as specified on the wiring diagram
several tries
OM0246 Page 54 of 64
Rev 5 June/2020
Page 55
heater duty cycle are too high and likewise if the door and perimeter heater duty cycles are too low the glass door and
front
wall of the chamber will show excessive condensation.
Condensation Persists After
Cycles are Adjusted
Increase air injections, increasing duration first,
& 1 minute (Air Cyc)
Check side access port stopper pressure relief hole(s) for
the chamber to be less than flow into the chamber.
Check for and remove the block of shipping foam from
the room where the incubator is installed
Message says it is time to change
Check and Change filters in sensor bay and reset timer
responses
Fill with maximum 3 liters of distilled single water no
Than 1 Mega Ohm – removes message on main screen
Door and Perimeter Duty
Condensation on the face
of the incubator outside of
the chamber
Pressure in chamber high enough to
cause blow by around the inner door
gasket
recommended initial change to 6 sec (Air Inj)
blockage.
Check circulation system for blockage causing flow out of
under the air pump.
Excess Vibration
Check jacket fan mounted in bottom on Incubator
Turn the incubator off. If the vibration persists it is from
Maintenance Required
Press button
Filter
Check Filter and extend timer if filter doesn’t need
changing
Extend Timer – See section 10.4 for more detail on all
Button in Main Screen
Message is RH Watertank
purer
For further assistance, call NuAire Customer Service at 1-800-328-3352 or (763) 553-1270 USA.
12.0 Communication Outputs
12.1 Remote Alarm Contacts
The NuAire DHD Incubator contains a set of contact points to connect to a remote alarm system. The contacts are
located on the rear panel (see page 14). The contacts are housed in a modular (RJ-11) telephone jack and rated for (30V
at 1 Amp). The contacts provided are normally open (NO), normally closed (NC) and common (COM) as shown below.
The alarm contacts do not distinguish between a CO
contacts upon an alarm condition. Power interruption will also change the state of the contacts.
temperature and any other alarm. Each will open or close the
2
Note: The relay is powered when the incubator is turned on in order to detect a power interruption.
The relay contacts are in their normal state (NO, NC) when there is an alarm or during the loss of power to the
incubator control board.
OM0246 Page 55 of 64
Rev 5 June/2020
Page 56
12.2 4 to 20mA Analog System Performance Output: Refer to PTB0235 Revision 3 or higher.
CAUTION
!
WARNING
!
12.3 RS-485 2-way Communications: Refer to PTB0234
12.4 On-Board USB Port:
Uploading Service Performance/Event Data: Refer to STB0348
Uploading Incubator Programming: Refer to PTB0347
When connecting external accessories or components to the system,
attach only components that have been tested and certified for
compliance to UL/IEC 60601-1 or UL/IEC 60950
13.0 Electrical/Environmental Requirements
13.1 Electrical
Models: NU-5810, NU-5820, NU-5831 and NU-5841
Domestic Units: 115V, 50/60Hz, 1 Phase, 8.5 Amps
“E” Units: 230V, 50/60Hz, 1 Phase, 4.3 Amps
Start Up Power 625 Watts
Running Power 250 Watts
Decon Cycle 995 Watts
13.2 Operational Performance (for indoor use only)
Environment Temperature Range: 60°F-85°F (15.5°C – 29.4°C)
Environment Humidity: Maximum relative humidity 80% for temperatures up to
31°C decreasing linearly to 50% relative humidity at 40°C
Environment Altitude: 6562 ft (2000 m) over sea level maximum
13.3 Light Exposure
Standard Fluorescent Lighting @ 150 ft. candles (1614 LUX) maximum intensity.
13.4 Installation Category: 2.0
Installation category (overvoltage category) defines the level of transient overvoltage, which the instrument is designed
to withstand safely. It depends on the nature of the electricity supply and its overvoltage protection means. For
example, in CAT II, which is the category used for instruments in installations supplied from a supply comparable to public
mains such as hospital and research laboratories and most industrial laboratories, the expected transient overvoltage is
2500V for a 230V supply and 1500V for a 120V supply.
13.5 Pollution Degree: 2.0
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree
2.0 assumes that normally only non-conductive pollution such as dust occurs with the exception of occasional
conductivity caused by condensation.
13.6 Chemical Exposure
Chemical exposure should be limited to antibacterial materials used for cleaning and disinfecting. Chlorinated and
Halogen materials are not recommended for use on stainless steel surfaces. Chamber decontamination can be
accomplished by paraformaldehyde, vapor phased Hydrogen Peroxide or Ethylene Oxide without degradation of cabinet
materials.
13.7 EMC Performance (classified for light industrial)
Emissions: EN61326
Immunity: EN61326
Class A equipment is intended for use in an industrial environment.
In the documentation for the user, a statement shall be included drawing
attention to the fact that there may be potential difficulties in ensuring
electromagnetic compatibility in other environments, due to conducted
as well as radiated disturbances.
OM0246 Page 56 of 64
Rev 5 June/2020
Page 57
Component
Material
Incubator Cabinet
Painted Steel
Outer Door
Painted Steel
Chamber
Stainless Steel
Outer Door Gasket
Vinyl Clad Magnet
Inner Door Gasket
Silicon Rubber
Front Service Panel
Painted Steel
Electronics Panel
Painted Steel
Sensor Bay
Aluminum
Exhaust Filter
Aluminum
Hosing
Silicon Rubber
Tubing
Stainless Steel
Air Pump
Various Steel/Copper/Nylon
Motor
Various Steel/Copper/PVC
Printed Wiring Assembly
Lead Free Electronic
Wire
PVC Coated Copper
Solenoid Valves
Various Steel/Copper
Connectors
PVC
Hardware
Stainless Steel and Steel
BIOHAZARD
RECYCLE
LEAD FREE
CAUTION
!
13.8 Heat Rejection: 14 BTU/Minutes.
14.0 Disposal and Recycle
Incubators that are no longer in use and are ready for disposal contain reusable materials. ALL components with the
exception of the HEPA filters may be disposed and/or recycled after they are known to be properly disinfected.
Note: Follow all local, state and federal guidelines for disposal of HEPA filter solid waste.
Prior to any disassembly for disposal the incubator must be decontaminated
Note: Material type can be verified with use of a magnet with stainless and aluminum being non-magnetic.
OM0246 Page 57 of 64
Rev 5 June/2020
Page 58
OM0246 Page 58 of 64
Rev 5 June/2020
Page 59
OM0246 Page 59 of 64
Rev 5 June/2020
Page 60
OM0246 Page 60 of 64
Rev 5 June/2020
Page 61
OM0246 Page 61 of 64
Rev 5 June/2020
Page 62
OM0246 Page 62 of 64
Rev 5 June/2020
Page 63
OM0246 Page 63 of 64
Rev 5 June/2020
Page 64
OM0246 Page 64 of 64
Rev 5 June/2020