Minilyzer ML1
Technical Specification, Software V2.00
updated Jul 2001
Technical Data General Functions
Measurements - Frequency, Level-RMS, Level-Relative
- THD+N, vu+PPM
- Polarity Test, Signal Balance Error
- Sweep, Frequency Sweep, Time Sweep
- 1/3rd Octave Spectrum, Scope
Frequency
Range 10 Hz to 20 kHz
Resolution 4 digits
Accuracy < ± 0.1 %
Level
Units dBu, dBV, V
Resolution 3 digits (dB-scale) or 4 digits (V-scale)
Accuracy ± 0.5 % @ 1 kHz
Bandwidth 20 Hz to 20 kHz
Flatness ± 0.1 dB
THD+N (Total Harmonic Distortion + Noise)
Meas. Bandwidth 10 Hz to 20 kHz
Resolution 3 digits (dB-scale) or 4 digits (%-scale)
Residual THD+N balanced < -85 dB @ -10 dBu to +20 dBu
unbalanced < -74 dB @ 0 dBu to +14 dBu
vu+PPM (vu-Indicator and Peak Program Meter)
according to IEC 60268 and DIN 45406.
PPM Type I, IIa and Nordic.
Both meters with adjustable reference and with
analog & numerical peak-hold readout.
Polarity Test (with MR1 test signal)
Positive / Negative detection through internal
microphone or XLR/RCA connector. Checks polarity
of tweeters, midrange-speakers, woofers, subwoofers and cables. Down to 10 dB S/N ratio of
input signal.
RMS
Signal Balance Error Indication range 0.0 % to 100 %
Deviation from perfect balance in % or *1
Sweep Frequency Sweep: Level as function of frequency.
Time Sweep: Measurement of level, THD+N and
frequency as function of time.
1/3rdOctave Spectrum acc. IEC 1260, class II and ANSI S1.11-
1976, class II from 50 Hz to 20 kHz, Bargraph
for Level
Scope Auto triggering, auto ranging, auto scaling
Filters Linear, A-weighting, C-weigting, C-message,
Highpass 22 Hz / 60 Hz / 400 Hz, X-Curve-1, Voice
bandpass
Input Connectors XLR balanced, RCA unbalanced
Input Impedance 40 kOhm balanced, 20 kOhm unbalanced
Input RMS1 ( upper meas. limit )
balanced +20 dBu (7.75 V
+14 dBu (3.8 V
Max. DC Input ± 50 V
Residual Noise < 12 µV, XLR-input shorted
Microphone Input (for Polarity measurement only)
Omnidirectional
Monitor Output Jack 3.5 mm (1/8“), suitable for all common
headsets
Display Graphic LCD 64 x 100 pixel, with backlight
Batteries 3x AA package dry batteries (alkaline) Typical
battery lifetime > 16 hrs
Dimensions (L x W x H)
163 x 86 x 42 mm (6.4“ x 3.38“ x 1.63“)
Weight 300 g (10.5 oz) incl. batteries
Temperature 0° to +45° C (32° to 113° F)
Humidity < 90 % R.H., non condensing
20 Hz to 20 kHz
RMS
RMS
DC
) unbalanced
)
RMS
page 1 of 2
1
for input levels > 20 dBu (balanced) the ML1 Adapter -20 dB is
available
Minilyzer ML1
Technical Data Acoustic Functions
(firmware 2.00 or higher, applicable with MiniSPL only)
Measurements - Sound Pressure Level
- 1/3rd Octave Spectrum
Acoustical Functions
in accordance with IEC60651 and IEC 60804
• Instantaneous Sound Pressure Level (Lp)
• Maximum/Minimum Sound Pressure Level (Lmax/Lmin)
• Time Response selectable
• Weighting Filters
• Equivalent Continuous Sound Pressure Level (Leq)
• Pause- and Continue Function
Sound Pressure Level
Units dB
Resolution 3 digits
Display Ranges 20 - 140 dB
Bandwidth 20 Hz to 20 kHz
Flatness According to class 1
SPL
, dB
, dB
Leq
, dB
LAeq
LCeq
in 3 bands 20 - 100 dB
SPL
40 - 120 dB
60 - 140 dB
SPL
SPL
SPL
Time Response Selectable fast, slow, impuls
Weighting Filters Selectable A, C, linear,
X-Curve-1 (for 1/3rd octave spectrum only)
Integration Pause- and Continue Function
1/3rdOctave Spectrum
31 octave band display 20 Hz to 20 kHz
Sensitivity Selectable default value (MiniSPL),
calibration to external source,
editable sensitivity value
page 2 of 2