NTI MiniSPL User Manual

Warranty Terms
Contact Details
NTI guarantees the MiniSPL and its components against defects in material or workmanship for a period of one year from the date of original purchase, and agrees to repair or to replace at its discretion any defective unit at no cost for either parts or labor during this period. This warranty does not cover damages caused through accidents, misuse, lack of care, transit, the attachment or installation of any components that were not provided with the product, loss of parts, connecting the instrument to a power supply, input signal voltage or connector type other than specified, or wrongly polarized batteries. In particular, no responsibility is granted for special, incidental or consequential damages.
This warranty becomes void if servicing or repairs of the product are performed by any party other than an authorized NTI repair center or if the instrument has been opened in a manner other than specified in this manual.
In case of malfunction, take - or ship prepaid - the MiniSPL, packed in the original box, to the authorized NTI representative in your country.
Note: Be sure to include a copy of your sales invoice as prove of purchase date and a detailed description of the fault.
For contact-details of the NTI representative please see our web page: www.nt-instruments.com
Headquarter: NTI AG
Im alten Riet 102 9494 Schaan Liechtenstein, Europe Tel. +423 - 239 6060 Fax +423 - 239 6089 E-mail info@nt-instruments.com Home www.nt-instruments.com
Subsidiary: NTI North America
3520 Griffith Street St. Laurent Quebec H4T 1A7 Canada Tel. +1 - 514 - 344 5220 Toll free1800 - 661 6388 Fax +1 - 514 - 344 522 E-mail canada@nt-instruments.com
The MiniSPL-package includes:
1x Case MiniSPL 1x MiniSPL, measuring microphone 1x Windscreen 1x Test Certificate with individual frequency response chart 1x Microphone-holder 1x Adapter 5/8“ - 3/8“ 1x User Manual MiniSPL
MiniSPL
User Manual
1/2“ Capsule, Omni-directional
Status LED, indicating On-status
Sleeve
© NTI AG All rights reserved. Subject to change without notice. Release 1.2 / February 2002 Minilyzer, MiniSPL and Minstruments are registered trademarks of NTI.
2001
MP 2/01 E 100 N 09.01
How to insert the battery
MiniSPL with Minilyzer ML1
Technical Data MiniSPL
• To insert the battery, press the push-pin in and slide the sleeve off.
• Insert one 1.5 V AA size battery according to drawing below. The polarity is marked on the sticker inside.
1
2
Push-pin
Note: The MiniSPL does not operate without
the battery.
• Close the MiniSPL. Turn the sleeve so that it is locked by the push-pin.
Calibration-sticker
Notes & Warnings
To avoid any problems during the operation of the instrument, follow the rules listed below:
• Use the instrument for the intended purpose only.
• Never connect the instrument to a high voltage output such as a power amplifier, mains power plug, etc.
• Do not disassemble the instrument.
• Never use the instrument in a wet environment.
• Remove the batteries as soon as they are flat or if the instrument is not intended to be used for a longer period of time.
• The use of rechargeable NiCd- or NiMH-batteries is causing shorter battery lifetime than specified.
• Note the correct polarity of the inserted battery.
• We recommend to calibrate the MiniSPL in yearly intervals to fulfill the specified sensitivity.
• Connect the MiniSPL to the Minilyzer ML1.
• Approximately ten seconds later the Status-LED will start to flash periodically indicating that the MiniSPL has switched on.
• An empty battery will be indicated by the non-flashing Status-LED.
• Simply unplug the MiniSPL after completed operation. Thus the Status-LED will stop flashing, indicating it is switched off.
MiniSPL as stand-alone mic
The MiniSPL may be operated as a stand-alone microphone e.g. connected to a balanced audio-line, a microphone­input or a sound-card. Note: Follow the specified load condition to achieve the flat frequency response. Phantom power causes a lower load condition, so make sure to switch it off.
At inputs with high DC-input resistance you may find the MiniSPL not switching on or the delay-time until switching on may be too long. In this case you have to build up an adapter according to the below circuit diagram. Please note that both resistance ’’R’’ should be of equal values within 1% tolerance for good signal balancing. At unbalanced inputs use pin 1 and 2. Pin 3 shall not be connected. In this case the sensitivity will be halved.
2
1
3
R
R = 10 kOhm .. 100 kOhm
connect audio-input
connect MiniSPL
Life
Ground
Return
Microphone Type 1/2”, omni-directional,
pre-polarized condenser, free field transducer
Sensitivity (20 ±2) mV/Pa,
(-34 ±1) dBV/Pa @ 1 kHz, 20°C +0.05 dB balanced out
Frequency Response acc. to IEC60651, Type 2,
tolerance curve is shown on individual calibration certificate
Peak Acoustic Input 130 dB Noise 32 dB
/ °C,
SPL
@ 1 kHz
SPL
, A-weighted
SPL
Load Conditions 20 - 200 kOhm balanced,
to achieve this always switch PHANTOM POWER off.
Power Supply 1x AA battery 1.5 V, battery
lifetime typical 300 hrs, no phantom power supply, phantom power resistant
Dimensions O 22 x 180 mm, O 0.87” x 7” Weight 100 g ( 3.5 oz ) incl. battery Temperature 0° to +45° C (32° to 113° F) Humidity < 90 % R.H., non condensing Conforms to the
Normative Standards EMC-Directives:
89/336, 92/31, 93/68 Harmonized Standards: EN 61326-1
//