Overview
Measurement Microphones
M2230M2211M2215M4261MA220
Overview
PreAmplifier
Certified Class 1
measurement
microphone
in accordance with
IEC 61672,
metal diaphragm
4
General purpose
measurement
microphone
class 1 frequency
response,
metal diaphragm
Measurement
microphone
for high sound levels
(up to 153 dB),
class 1 frequency
response,
metal diaphragm
Cost-effective class
2 measurement mi-
crophone for general
sound level testing,
commissioning and
service of audio-
acoustic installations
Microphone
preamplifier
compatible with
1/2” pre-polarized
capsules
Outdoor Measurement Microphones
Overview
M2230-WP
Outdoor Microphone
Outdoor measurement microphone,
class 1 in accordance with IEC 61672
consists of
M2230 Measurement Microphone
+ WP30 Weather Protection
M4261-WP
Outdoor Microphone
Outdoor measurement microphone,
class 2 in accordance with IEC 61672
consists of
M4261 Measurement Microphone
+ WP61 Weather Protection
5
Scope of Delivery
M2230• Measurement Microphone
consisting of
- MA220 Microphone PreAmplifier
- Microphone Capsule MC230 or MC230A
• Dust Cap
• 50 mm Windscreen
• Microphone-holder MH01
with Adapter 5/8“ - 3/8“
• Operating Manual
• Individual Frequency Response Chart
M2211• Measurement Microphone
consisting of
- Microphone PreAmplifier MA220
- Microphone Capsule 7052
• Dust Cap
• 33 mm Windscreen
• Microphone-holder
with Adapter 5/8“ - 3/8“
• Operating Manual
Overview
M2215• M2215 Measurement Microphone
consisting of
- Microphone PreAmplifier MA220
- Microphone Capsule 7056
• Dust Cap
• 33 mm Windscreen
• Microphone-holder
with Adapter 5/8“ - 3/8“
• Operating Manual
M4261• Measurement Microphone
• 33 mm Windscreen
• Microphone-holder
with Adapter 5/8“ - 3/8“
• Operating Manual
MA220• PreAmplifier
• Dust Cap
• Microphone-holder
with Adapter 5/8“ - 3/8“
• Operating Manual
6
WP30• WP30 Weather Protection for M2230
- Bird spike
- 90mm Wind screen
- Protection cage
- Upper body tube with allen key mount
- Lower body tube
- Footer plate with tripod mounting thread
(incl. 3 allen screws)
- Allen key
WP61• WP61 Weather Protection for M4261
- Bird spike
- 90mm Wind screen
- Protection cage
- Upper body tube with allen key mount
- Lower body tube
- Footer plate with tripod mounting thread
(incl. 3 allen screws)
- Allen key
Overview
7
Microphone Description
Description
The plug-on measurement microphones combined with the
XL2 Analyzer form a comprehensive sound level meter and
acoustic analyzer. The microphones are 48 VDC phantom-powered and include an electronic data sheet.
Integrated Preamplifier
The microphone bodies contain a preamplifier and require
48 VDC phantom power supply for operation. They combine
high dynamic range and wide frequency range with low noise.
The measurement microphones can also be connected with
an ASD Cable to the XL2 Audio and Acoustic Analyzer for measurements at remote locations or for reduction of acoustic reflections.
Electronic Data Sheet
The microphones include an electronic data sheet. The Automated Sensor Detection (ASD) of the XL2 Analyzer automatically reads this data, i.e. the microphone model and calibration
data. This promotes faster setup and ensures accurate measurements.
Screen connection used for
ASD technology
XLR M
Pin 2 +
Pin 3 Pin 1 Shield
2
2
3
3
1
1
Connection diagram of measurement microphones
with electronic data sheet
Microphone
2
3
1
ASD
Microphone
Capsule
8
Microphone Description
Connecting to XL2
Microphone plugs directly into the XL2
The XL2 automatically reads the electronic data sheet of the
connected microphone as follows:
• Connect the measurement microphone to the XL2.
• Switch on the XL2.
The XL2 reads the electronic data sheet of the connected
microphone during a brief initialization process prior to the
first measurement.
Microphone Connection via the ASD Cable
The NTi Audio measurement microphones can be connected
with an ASD Cable to the XL2 Analyzer for measurements at
remote locations or for reducing acoustic reflections. The electronic data sheet is transmitted via the XLR connector’s housing. Do not touch this during the brief initialization period to
ensure the complete data sheet is recognized by the XL2. The
automated sensor detection does not disturb any measurements. You may join 5- or 10-meter ASD Cables together in
series. The ASD technology supports accurate data communication up to a combined cable length of 20 meters (= 65 feet).
Microphone Connection via a professional Audio Cable
For distances longer than 20 meter (= 65 feet) use a high quality, low capacitance standard professional audio cable. The microphone sensitivity has to be entered manually into the XL2
Analyzer.
Alternatively connect the microphone first directly to the Analyzer. The XL2 reads the sensitivity and remembers this value.
Afterwards connect the audio cable.
9
Microphone Description
Capsule Replacement Instruction
The microphones for the XL2 Analyzer include an electronic
data sheet. The Automated Sensor Detection (ASD) of the XL2
Analyzer automatically reads this data, i.e. the microphone
model and calibration data. This promotes faster setup and ensures accurate measurements. In case of a capsule replacement, the electronic data sheet needs to be updated with the
data of the new capsule.
Step-by-Step-Instruction
• Install the new capsule on the microphone preamplifier.
• Plug the measurement microphone directly into the XL2.
• Install the latest firmware in the XL2, available at
https://my.nti-audio.com/support/xl2.
• Start the XL2 Projector PRO Software. The computer requires online connection to the web.
• Connect the XL2 with the USB cable to the Projector PRO
software, thus you see the XL2 display live on the computer
monitor. (if prompted select COM-Port on the XL2)
• Press the computer keyboard keys “Ctrl + Shift + F5” at the
same time (alternatively “Ctrl + Alt + F5”)
10
• Select Original NTi Audio Microphone or Custom micro-phone.
• Case: Original NTi Audio Microphone
- Enter the serial number of the new capsule
- Confirm by clicking Write to MA220.
- Now XL2 reads the factory sensitivity of the new capsule
from the NTi Audio server and stores the new data into
the electronic data sheet of the preamplifier. You will be
prompted if all is in good order.
• Case: Custom microphone
- Enter the microphone sensitivity
- Confirm by clicking Write to MA220.
- Now XL2 stores the microphone sensitivity as factory sensitivity in the electronic data sheet of the MA220 PreAmplifier.
• Verify the setting in the “CALIBRATE” screen of the XL2 and
perform a user calibration to verify if the new capsule works
in good order.
Microphone Description
11
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.