NTI ICP Adapter ASD Data Sheet

PRODUCT DATA
ICP Adapter ASD
for XL2 Analyzer
The ICP Adapter connects to the XL2 and generates ICP pow­er supply for accelerometers and other custom sensors. The adapter offers an electronic data sheet, which stores the sen­sitivity and individual serial number of the connected sensor.
Signal
Block Diagram
ICP Adapter ASD
Application
Vibration measurements•
Electronic Data Sheet
The Automated Sensor Detection (ASD) of the XL2 Analyzer au­tomatically reads the electronic data sheet of this adapter, i.e. the adapter model, the sensor serial number and calibration data. This promotes faster setup and ensures accurate measurements. The information of the electronic data sheet is documented in every XL2 measurement report.
XLRBNC
2
3 1 ScreenASD
Specifications
Output / Input XLR / BNC
Power Supply for Sensors ICP•
Power Supply from XL2 48 V Phantom Power
Maximum Input Level 8.84 V (146.9 dB @ Sensitivity = 20 mV/Pa)
Residual Noise typical LA = 2.5 µV
®
4 mA +/- 0.02 mA•
LZ = 3.5 µV
THD @ 1 kHz, 1Vrms < 80 dB
Frequency Range 2.1 Hz - 300 kHz
NTi Audio # 600 010 223
ICP® is a registered trademark of PCB Piezotronics.
www.nti-audio.com Page 1 / 2
XL2 Projector
Serial Number Setting
PRODUCT DATA
Prior First Use
Serial Number Setting
Store the serial number of your connected sensor in the electronic data sheet of the adapter.
Connect the XL2 to the computer and start the XL2 Projector • software V2.6 or higher. You will see the XL2 screen live on the PC monitor. The latest XL2 Projector software is available for download at http://my.nti-audio.com/support/xl2.
Press F4 on the keyboard to open the ASD Adapter S/N Writer.•
Click • Read from ASD Adapter.
Enter the sensor serial number.•
Click • Write S/N to Adapter.
In case you use multiple custom sensors, then we suggest uti­lizing an individual adapter for each sensor. This saves time and simplifies your operation.
21
User Calibration
Set the sensitivity according to the sensor specifications:
Connect the adapter to the XL2.•
Power on the XL2 and select • Calibrte in the main menu. The ASD indication
2
will be shown. In case ASD is not shown, then select the measurement function SLMeter shortly and afterwards Calibrte again. The XL2 reads the sensor data at start up and change of measurement function.
Select • Sensitivity
1
with the rotary wheel .
Press enter • and adjust the sensitivity with the rotary wheel
in 0.1 mV/Pa steps. The setting range 1 µV/Pa - 9.99 V/Pa is
supported.
43
Confirm with enter • .
The configuration is completed. The serial number
sitivity
4
are stored in the electronic data sheet.
3
and sen-
Keep User Calibration
The XL2 deletes the user calibration at the serial num­ber setting. Thus always set the serial number first and the user calibration afterwards.
Dispose the instrument in accordance with the legal environmental regulations in the country.
www.nti-audio.com Page 2 / 2
Loading...