Thank you for purchasing NSK Perio Mate.
This product is a handpiece for cleaning sub- and supragingival tooth surfaces.
Please read this Operation Manual carefully before use to become familiar with operation instructions and
care & maintenance. Keep this Operation Manual for future reference.
English
English
1. Features
2. User and Indications for Use
3. Precautions for handling and operation
4. Correspondence Coupling
5. Package Contents
6. Operation
7. Care After Use
8. Cleaning and Sterilization
9. Periodical Maintenance Checks
10. Troubleshooting
11. Specifications
12. Symbols
13. Warranty
14. Option Parts List
15. Spare Parts List
16. Disposing Product
1
Ř The 360°twist free swivel coupling enables comfortable and fatigueless operation even during prolonged
procedures, allowing effortless access to difficult areas.
<Powder for cleaning the tooth surface>
Ř NSK Perio Mate Powder used in conjunction with Perio Mate can remove plaque and eliminate sub- and
supragingival biofilm quickly and efficiently.
Ř The main ingredient of Perio Mate Powder is glycine, a type of amino acid.
User: Dentist, Dental hygienist
Indications for Use: Perio Mate is for sub- and supragingival plaque removal. When applied subgingivally, it
may be used to support periodontitis and periimplantitis therapy.
1
Precautions for handling and operation
3
Please read these precautions carefully and use only as intended or instructed.
Safety instructions are intended to avoid potential hazards that could result in personal injury or damage
to the device. Safety instructions are classified as follows in accordance with the seriousness of the risk.
Class
WARNING
CAUTION
NOTICE
Ř If any hypersensitivity such as eruption and dermatitis develops due to treatment using this product,
stop using the product and seek medical attention.
Ř The Perio Mate system can be used in periodontal pockets. Therefore, attention is required for patients
with emphysema. If emphysema develops due to treatment using this product, immediately stop using
the product and seek medical attention.
Ř This instrument is for use by dentists or hygienists only.Ř Do not use on the following patients
1) Those with respiratory diseases.
2) Those with pathologically deep periodontal pockets (6 mm or more) or mucosal lesion. Prolonged
application of the spray to soft tissue or salivary glands may cause powder air to enter, which may
cause emphysema.
3) Those with severe ulcers in digestive organs.
4) Those with hepatic dysfunction.
5) Those with lung or cardiac disturbance or ventilatory impairment.
6) Those with damage or abnormality in the oral cavity.
7) Those with hyperemia, hemorrhage, or inflammation in the oral cavity.
8) Those that tend to develop an inflammation or sores in the oral mucous membrane.
9) Those with allergies.
10) Depending on the patient’s constitution, allergy symptoms or stomatitis may be caused in very rare
cases. If such symptoms are detected, stop using the product immediately.
Ř Do not use Perio Mate with the patients listed below. When used in periodontal pockets for these
patients, this product may develop bacillemia.
-Patients with endocardial inflammation
-Patients pregnant or lactating
-Patients with a contagious illness or disease
-Patients with an immune deficiency disease (neutropenia, granulocytopenia, agranulocytosis,
diabetes, hemophilia)
-Patients under irradiation treatment, chemotherapy, or administration of an antibiotic
Ř Use this product only for patients with healthy periodontal pockets for maintenance after treatment of
a gum disease. When used for any patients other than those, this product may cause emphysema.
Ř Use this product only in periodontal pockets that meet the following conditions:
-Periodontal pockets caused by a gum disease
-Periodontal pockets of 6 mm or less
Ř Check the depth of periodontal pockets with a measurement probe or roentgenography before use.
Hazard that could result in serious injury or damage to the device if the
safety instructions are not correctly followed.
Hazard that could result in light or moderate injury or damage to the device
if the safety instructions are not correctly followed.
General product specification information highlighted to avoid product
malfunction and performance reduction.
WARNING
Degree of Risk
2
Ř Use Perio Mate Powder as cleaning powder. If other powder is used, the Perio Mate Nozzle Tip will
be damaged and it will increase the risk of damaging substantia ossea and dentine or causing
emphysema. We shall assume no liability for any failures caused by the use of other powder.
Ř During use, both the operator and the assistant should always wear protective glasses and protective
face masks. Also, continuously vacuum and collect ejected powder during use. In case any powder
gets in the eyes wash out immediately with copious amount of water and consult an ophthalmologist.
Ř Set supply air pressure at hand to 0.3 - 0.4 MPa (3.0 - 4.0 kgf/cm
2
). If this product is used with
incorrect pressure, the operator, assistant or patient may be injured due to a damaged Powder Case
Cover.
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize it since it may be damaged or
contaminated.
Ř Make sure that the Nozzle Tip package is not open or damaged before use. If the package is open or
damaged, do not use that Nozzle Tip as it may be damaged or contaminated.
Ř Follow the use-by date on the Nozzle Tip package. If the use-by date has lapsed, do not use that
Nozzle Tip as it may be damaged or contaminated.
CAUTION
Ř Read this Operation Manual before use to fully understand the product functions, and file for future
reference.
Ř When operating the product always consider the safety of the patient.
Ř The end user shall be responsible for any judgment that relates to the application of this product to a
patient.
Ř Users are responsible for the operational control, maintenance and continual inspection of this product.
Ř Do not attempt to disassemble the product nor tamper with the mechanism except as recommend by
NSK in this Operation Manual.
Ř This product is a precision instrument. Do not allow any impact on to the product. Do not drop the
product. Even the slightest impact may damage and deform the product.
Ř Should the product function abnormally, cease operation immediately and contact your Authorized NSK
Dealer.
Ř Be very careful to avoid cleaning powder entering a patient’s mucosal areas (eyes, nose etc) other
than the oral cavity. Also, protect the face with a towel, etc., or protective glasses, etc., to prevent entry
of cleaning powder into eyes.
*If the patient wears contact lenses, have the patient wear safety glasses or remove his/her contact
lenses before treatment.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the air may cause
internal solidification of the cleaning powder.
Ř Securely attach the Powder Case Cap. Do not open the Powder Case Cap during use. The Powder Cap
and internal powder may be ejected since pressure is high in the Powder Case.
Ř Before connecting the Powder Case to the Handpiece Hose, be sure to blow off water remaining
on the Hose connection with dry air. Residual moisture may enter the supply air, leading to internal
solidification of the cleaning powder.
Ř If too much cleaning powder is put into the Powder Case, the proper amount of powder may not be
sprayed immediately after the start of use.
Ř The Perio Mate Powder has the property of easily scattering because of its fine particles. Fill the
chamber slowly and be careful to prevent the powder from scattering.
Ř Powder stored for a long while in opened packages can attract moisture and the powder may clog in
the Nozzle, etc. Make sure to use new dry powder. Securely close the lid of the Perio Mate Powder
container after filling the powder.
Ř Apply Vaseline, etc., to the patient’s lips to prevent drying or chapping at the corners of the mouth.
Ř Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
3
English
Ř Perform regular function and maintenance checks.
Ř If the device has not been used for a long time, check that it functions correctly and safely before
using it on a patient.
Ř U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
Correspondence Coupling
4
Joint TypeModel
NSK PHATELUS Coupling, MACH CouplingPER-PTL-P
®
KaVo
MULTIflex®, MULTIflex® LUX Coupling TypePER-KV-P
®
W&H
Roto Quick® Coupling TypePER-WH-P
®
Bien-Air
Sirona
Unifix®, Unifix® L Coupling TypePER-BA-P
®
Quick Coupling TypePER-SR-P
ISO 9168 Type2 (Midwest 4-hole) JointPER-M4-P
ISO 9168 Type1 (Borden 2-hole) JointPER-B2-P
KaVo® and MULTIflex® are registered trademarks of Kaltenbach & Voigt GmbH & Co., Germany.
Sirona® is a registered trademark of Sirona Dental System GmbH, Germany.
W&H® and Roto Quick® are registered trademarks of W&H dentalwerk Burmoos GmbH, Austria.
Bien-Air® and Unifix® are registered trademarks of Bien-Air Dental S.A., Switzerland.
Package Contents
5
Handpiece 1 Set
Powder Case (With Powder Case Cap) 1 Set
Perio Mate Nozzle Tip 40 pcs.
Perio Mate Powder 1
Maintenance Kit
Cleaning File (For Nozzle tip, For Internal Nozzle) 1
Cleaning Wire (S) (For the root of the Nozzle) 1
Cleaning Wire (L) (for other uses) 1
Cleaning Brush 1
Perio Mate Tip Remover 1
Ring Wrench 1
Turbine Oil 1
Blower Nozzle 1
(The shape varies depending on the Joint Type.)
Spare parts
Powder Case Cover 1
O-Ring (Handpiece End) 1 Set
O-Ring (Coupling End) (NSK PHATELUS coupling ONLY) 1 Set
Ř The Nozzle, Handpiece, Powder Case, and Blower Nozzle cannot be used for NSK
CAUTION
Prophy-Mate neo.
4
Component Names
Nozzle Tip
Operation
6
WARNING
Cover Holder
Powder Case Cap
Hose Connection
Internal Nozzle
Flow Volume
Adjustment Ring
Nozzle
Powder Case Cover
Rubber Packing
Handpiece Joint
O-Ring
Handpiece
Handpiece Release Collar
Powder Case
Ř Avoid prolonged continuous use, and allow the patient to rinse at appropriate times.
Ř Never impact the Powder Case and Powder Case Cap as this may cause the Powder
Case or Powder Case Cap to scratch or crack.
Ř Never take the Perio Mate apart during operation. Never pull the Handpiece Release
Collar or the coupling release during operation. Attempting to do this may cause
components of the Perio Mate to unexpectedly detach and the Perio Mate Powder to
scatter.
Ř Do not open the Powder Case Cap while using the Perio Mate. Powder may scatter.
Ř Use this product by setting the pressure to 0.3 - 0.4 MPa (3.0 - 4.0 kgf/cm
2
). Using the
product at an excessive pressure increases the risk of causing emphysema.
Ř Thoroughly dry the Perio Mate and all its components inside and out before use.
Insufficient drying may cause clogging of powder.
English
CAUTION
Ř Be very careful to avoid Perio Mate Powder entering patient’s mucosal areas (eyes,
nose etc) other than the oral cavity. Also, protect the face with a towel, etc., or protective
glasses, etc., to prevent entry of powder into eyes.
Ř If chemicals, oils, etc, adhere to the Powder Case Cover, wipe them off immediately as
they could cause damage.
5
6-1 Air Pressure setting
Measure the supply pressure at the handpiece
point and set the pressure to the value specified
on the specification table. (Fig. 1)
*For Multi Gauge information refer to Option Parts
List.
Ř Do not exceed the optimum pressure specified on the specification table.
WARNING
Ř Excess air pressure could cause the Powder Case and Powder Case Cap to crack or
fracture, which could lead to injury.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the
CAUTION
air may cause internal solidification of the cleaning powder.
Exhaust Air
Pressure
Handpiece
6-2 Connecting the Powder Case to the Handpiece Hose
Ř Before connecting each part, thoroughly dry the Coupling, Handpiece, Hose Joint section
CAUTION
of the Powder Case, inside of the Powder Case, and Nozzle. Also, perform idle running
with air alone to expel any internal water.
Chip Air Pressure
Water Pressure
Drive Air
Pressure
Fig. 1
AIR
Coupling
AIR
AIR
Powder Case Cap
AIR
Handpiece connection
Powder Case
AIR
AIR
Internal Nozzle
6
Connect the hose to the Powder Case.
PER-PTL-P
-
Designed for when NSK FM-CL (B2/B3), PTL-CL-LED (6 pins), or PTL-CL-FV (5 Hole) coupling is used.
Insert each coupling into the Hose Connection of the Powder Case.
-PER-M4-P, PER-B2-P
Used when the hose of Midwest 4 Hole (M4) or Borden 2 Hole (B2) is used. Align each pipe at the Hose
Connection of the Powder Case with the ports in the hose. Carefully insert the pipe straight into the hose
until it stops and securely tighten the Hose Nut.
-PER-KV-P, PER-WH-P, PER-BA-P, PER-SR-P
Used when KaVo
manual of the coupling you have and connect Perio Mate accordingly.
®
, W&H®, Bien-Air® or Sirona® coupling etc., is used. Check with the instruction
6-3 Connecting the Handpiece to the Powder Case
Pull back the Handpiece Release Collar on the
Powder Case and push the handpiece straight on
to the connection. Release the Handpiece Release
Collar once the handpiece is in place. (Fig. 2)
Once connected gently pull handpiece to ensure it
is securely connected.
Handpiece Release Collar
6-4 Check before treatment
Before use, check by operating the device away from a patient. If any abnormality is detected such
as vibration, sound or other during checking, or when using, stop use immediately and contact the
Authorized NSK Dealer.
English
Fig. 2
6-5 Powder filling
Remove the Powder Case Cap and fill the Perio Mate Powder to the position marked ‘Max’ in the
Powder Case. (Approx. 10 g of powder can be filled.) (Fig. 3)
Securely close the Perio Mate Powder container lid.
Securely close the Powder Case Cap. (Fig. 4)
Replacement dose of Perio Mate Powder
CLOSE
Fig. 3Fig. 4
*For Perio Mate Powder information refer to Spare Parts List.
7
Ř If excess cleaning powder is put into the Powder Chamber, the correct amount of
CAUTION
powder may not be sprayed immediately after the start of use.
Ř The Perio Mate Powder has the property of easily scattering because of its fine particles.
Perform filling slowly and be careful to prevent the powder from scattering.
Ř Do not use any powder stored for a long while in opened packages. Do not use leftover
powder. Using such powder may cause clogging of the nozzle.
Ř Use only Perio Mate Powder. Other powders have different particle sizes and will clog
and not perform as expected. Using other powders may also cause significant damage
to the Nozzle Tip.
Ř Always securely close the container lid of the Perio Mate Powder. Failure to do so will
attract moisture which may cause clogging.
Ř Use Perio Mate Powder as cleaning powder. If other powder is used, the Perio Mate
Nozzle Tip will be damaged and it will increase the risk of damaging substantia ossea
and dentine or causing emphysema. NSK shall assume no liability for any failures
caused by the use of other powders.
Ř If powder adheres to the Powder Case and/or the threaded portion of Powder Case Cap,
clean it immediately or it will become difficult to close the Cap.
6-6 Operation
1) Attach the Nozzle Tip to the Nozzle of the Perio
Mate Handpiece.
2) After filling the powder, spray it into a spittoon,
etc., for one second and confirm that powder
is mixed with water and air, and spraying is
appropriately performed.
Nozzle Tip
3) Insert the Nozzle Tip into a periodontal pocket
by 3 mm or more. (Fig. 6)
If the Nozzle Tip is inserted less than 3 mm,
powder may not be sprayed into the periodontal
pocket efficiently.
4) Start operation. Spray powder in the same
periodontal pocket for up to 5 seconds or on
the same tooth for up to 20 seconds. Do not
spray longer than this specified time.
5) After treatment, remove the Nozzle Tip from
the Perio Mate Handpiece using the Nozzle Tip
Remover and dispose of it. (Fig. 7)
Fig. 5
Fig. 6
Nozzle Tip Remover
Fig. 7
8
NOTICE
WARNING
Ř The powder spray volume can be adjusted with the Flow Volume Adjustment Ring (Fig. 8).
Start by setting it to Min. then adjust according to the application and patient condition.
Ř After adjusting the powder spray amount with the Flow Volume Adjustment Ring,
clogging may occur due to the powder condition. (for example, when moisture is
contained) In such a case, the problem may be resolved by again setting the Flow
Volume Adjustment Ring to Max, then adjusting down.
Ř Securely insert the Tip all the way by referring to the Fig. 9. If the Tip is inserted
incorrectly, the Nozzle Tip will come off and accidental ingestion may result.
Water Flow
Fig. 8
Fig. 9
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize the tip since it may be
damaged or contaminated. If the Nozzle Tip is reused, sufficient performance may not
be exerted due to nozzle deformation during removal, or accidental ingestion may result
due to detachment of the Nozzle Tip during use.
Ř Do not use this product in the same area immediately after scaling treatment of a tooth
surface in a periodontal pocket.
Ř Do not insert the Nozzle Tip into the same periodontal pocket during the same treatment
more than once.
Ř Powder should be sprayed for 5 seconds at maximum in the same periodontal pocket (20
seconds at maximum for one tooth). Never spray powder longer than this specified time.
Ř Do not strongly force the Nozzle Tip into a periodontal pocket.
Ř Make sure to use Perio Mate Powder.
Ř Do not spray powder while the Nozzle Tip end is located at the bottom of a periodontal
pocket. Emphysema may be caused.
Ř Never spray powder toward a human body. Spray only toward a tooth surface.
Ř Do not use the Nozzle Tip in any places other than periodontal pockets.
Ř If the hose swells or air or powder leaks from connections between the Powder Case
and Handpiece, during use, immediately stop use.
Ř Do not use a deformed or damaged Nozzle Tip. It may detach during use.
Ř Excessive air pressure supply may cause emphysema. Make sure the air pressure is
correctly set, as specified. Set the spray air pressure to the extent that the spray slightly
flows out from the periodontal pocket, even at maximum. Adjust the spray air pressure
within an adequate supply air pressure while checking the patient’s condition.
Ř Use the product while spraying water.
Ř Make sure to use sufficient evacuation vacuum to prevent a patient from ingesting a
large amount of powder. Avoid continuous use for a long period of time and allow the
patient to rinse out the mouth as necessary.
English
9
Ř Even after the foot control pedal is released, powder may still come out until the air
CAUTION
Operation without Nozzle Tip
Perio Mate can also perform cleaning of
supragingival tooth surfaces or around marginal
gingiva. Remove the Nozzle Tip from the Perio
Mate Handpiece and perform cleaning. During
operation, use the Nozzle around 10mm away
from the tooth surface at angles of 30 to 60
degrees. (Fig. 10) Do not spray powder directly to
the oral mucosa and do not insert the Nozzle into
periodontal pockets.
WARNING
7
After each patient maintain the product as follows.
pressure in the Powder Case is expelled. After each operation, vacuum excess powder
until the powder ejection stops in order to prevent it from entering the oral cavity of the
patient.
Ř Make sure the compressed air supply is dry and clean. Water or oil content mixed in the
air may cause internal solidification of the cleaning powder.
Ř Use only drinkable water for supply water.
Ř When using Perio Mate without the Nozzle Tip, avoid use on the oral mucosa.
Care After Use
30 - 60 degrees
10mm
Fig. 10
WARNING
Ř If organic solvents, etc., adhere to the Powder Case Cap, this may result in damage.
Ř The Powder Case Cover will deteriorate over time under the influence of ultra violet
rays and other environmental factors and is therefore considered a consumable item.
Replace the powder case cover every six months.
Ř Make sure to perform 7-1 Discard of Perio Mate Powder - 7-4 Lubrication after each
use. If powder remains in the Powder Case or Handpiece, it will solidify which may
cause clogging.
Ř Remove the Powder Case Cover and Rubber Packing from the Cover Holder once a
week to remove powder caught in the clearance. Check that there are no cracks or
scratches on the Powder Case Cover at that time.
Ř Never use the Powder Case Cover with cracks or scratches. If used in such a condition
damage and injury may result from any broken pieces.
10
Ř In case of failure, make sure to contact your Authorized NSK Dealer for repairs.
CAUTION
Ř After use always thoroughly remove powder from the product then perform cleaning,
washing and sterilization. Ensure the instrument is totally dry for storage.
Ř Make sure to perform Ultrasonic Cleaning once a week to prevent clogging of powder.
(See 8-1 Ultrasonic Cleaning.)
Ř Never lubricate Perio Mate. (However, lubricate the O-rings and Flow Volume Adjustment
Ring with the supplied turbine oil. (See 7-4 Lubrication.)
7-1 Discard of Perio Mate Powder
1) Open the powder case cap, and remove the
remaining cleaning powder. (Fig. 11)
2) Perform idle running with air alone to discharge
water and powder in the Nozzle. (Fig. 12)
English
Fig. 11
AIR
3) Blow off the powder remaining in the powder
case with air. (Fig. 13)
4) Wipe the Nozzle and handpiece with a soft cloth
saturated in alcohol to remove saliva and other
physical contamination.
Fig. 12
Fig. 13
11
7-2 Removal of powder from the Powder Case, Handpiece and Nozzle
1) Remove the Powder Case from the hose.
2) Pull back the Handpiece Release Collar and
remove the handpiece.
3) Insert the Nozzle into the Ring Wrench and
loosen the Nozzle. Remove the Nozzle from the
Handpiece. (Fig. 14)
4) Fit the supplied Blower Nozzle to the coupling
and attach to the handpiece. Run idle with no
water more than 5 sec. to clear lines. (Fig. 15)
Blower Nozzle
(PHATELUS, MACH)
LOOSEN
Ring Wrench
Fig. 14
CouplingHandpiece
*The Blower Nozzle in the product package depends on the joint type you purchase. Do not use a blower nozzle
other than the supplied blower nozzle.
5) Connect the Blower Nozzle to the Nozzle end
(Fig. 16) and idle with air alone. Residual
Blower Nozzle
®
MULTIflex®,
(KaVo
MULTIflex® LUX Coupling)
Blower Nozzle
Nozzle
powder in the Nozzle will be discharged.
Ř When idling, hold the Nozzle and Handpiece tight and be careful not to allow the Nozzle
CAUTION
to be ejected by air pressure.
Ř When idling, stop water supply and idle with air alone.
7-3 Cleaning and washing
1) Remove powder from the Handpiece and
Powder Case connections and the Adjustment
Ring section.
Clean the inside of the Handpiece connection
with the supplied cleaning brush. (Fig. 17: Left)
Clean the Powder Case connections with the
supplied cleaning brush in the same way.
(Fig. 17: Right)
Handpiece
Powder Case
Fig. 15
Fig. 16
Fig. 17
12
*If powder adheres to the O-ring section, remove
the O-rings, and carefully clean the O-rings and
grooves. (Fig. 18)
English
Fig. 18
2) Remove the Rubber Packing and disassemble
the Powder Case Cap. (Fig. 19)
3) Wash away the Perio Mate Powder adhering
to the inside of the Powder Case/Powder Case
Cap under running water. (Fig. 20)
4) Blow water off the Powder Case Cover, Rubber
Packing, and the Cover Holder by air and dry.
Then, assemble.
Rubber Packing
Fig. 19
Fig. 20
CAUTION
Ř If cleaning powder remains in the Powder Case, powder will solidify, and clogging may
occur.
Ř Do not immerse the Powder Case or the Handpiece, with residual powder, into water.
The powder will solidify, and clogging may occur.
13
7-4 Lubrication
1) Dry the O-ring section of the Powder Case and the Adjustment Ring section, apply a drop of the
supplied turbine oil and rub it in with your finger, etc. (Fig. 21, 22)
OILOIL
OIL
Fig. 21Fig. 22
2) Perform autoclave sterilization.
Ř If you fail to lubricate the O-ring section on the Handpiece side, the Joint section will be
CAUTION
8
Only the handpiece can be washed via Thermo Disinfector.
damaged and this may cause the Handpiece connection to be difficult to rotate. It may
also cause water leakage.
Ř After autoclave sterilization, the Flow Volume Adjustment Ring can be difficult to rotate
in some rare cases. However, it may return to the normal state once you rotate it. If the
Adjustment Ring cannot rotate properly, contact your Authorized NSK Dealer.
Cleaning and Sterilization
Ř Do not use the Thermo Disinfector for the Powder Case, Powder Case Cap and Nozzle.
CAUTION
8-1 Ultrasonic Cleaning
Ř Make sure to perform Ultrasonic Cleaning once a week or after the remedy described in
CAUTION
1) After disassembling, perform cleaning. (See 7-1 Discard of Perio Mate Powder - 7-3 Cleaning and
Washing.)
2) Pass through the supplied Ring Wrench from the Nozzle end, loosen the Nozzle, and remove the
Nozzle from the Handpiece.
3) Perform Ultrasonic Cleaning with hot water for the Handpiece, Powder Case and Nozzle.
4) After cleaning, blow waterdrops from all parts by air to dry them.
5) Reassemble the parts following the disassembly procedures in reverse.
6) Connect the Powder Case and Hose with the cover not attached and discharge air. This allows water
in the Internal Nozzle to be discharged. Then thoroughly dry the inside.
10. Troubleshooting to prevent clogging of powder.
14
Ř When attaching the Nozzle, lightly thread the Nozzle screw into the Handpiece by hand
CAUTION
at first and then securely tighten it with the supplied Ring Wrench.
Ř After Ultrasonic Cleaning, blow remaining water from the inside with a syringe and
sufficiently dry the parts before use.
8-2 Sterilization
Sterilize the product by autoclave sterilization. Remove the Nozzle Tip and dispose of powder after each
patient and sterilize as noted below.
All parts of this product can be autoclaved.
Autoclave Procedure
1) Disassemble the product and clean it. (See 7-1 Discard of Perio Mate Powder - 7-4 Lubrication.)
2) Insert into an autoclave pouch. Seal the pouch.
3) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
4) Keep the product in the autoclave pouch to maintain sterility until it is used. To use, perform
reassembly in reverse order to disassembly.
Ř The Nozzle Tip is a single use product. Do not reuse or re-sterilize it since it may be
CAUTION
damaged or contaminated.
Ř Before use, thoroughly dry all parts.
Ř Remove the Powder Case Cap from the Powder Case, and then sterilize by autoclave.
Sterilization with the Powder Case Cap attached may cause breakage of the Powder
Case Cap.
Ř Do not autoclave the product with other instruments even when it is in a pouch. This is
to prevent possible discoloration and damage to the product from chemical residue from
other instruments.
Ř If the sterilizer chamber temperature may exceed 135°C during the dry cycle then
delete the dry cycle. Even in this case, sufficient drying of the product is required.
Ř Do not wipe with, or clean, or immerse in, high acid water or sterilizing solutions.
Ř Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation,
and sunlight. The air should be free from dust, salt and sulphur.
Ř Do not heat or cool the product too quickly. Rapid change in temperature could cause
damage to the product.
Ř The product can be sterilized only by autoclave sterilization.
Ř Do not touch the product immediately after autoclaving as it will be very hot and must
remain in a sterile condition.
English
NOTICE
Ř NSK recommends Class B sterilizers as stated in EN13060.
15
Periodical Maintenance Checks
9
Perform periodical maintenance checks every three months, referring to the check sheet below. If any
abnormalities are found, contact your Authorized NSK Dealer.
Points to checkDetails
Powder Case CapCheck that there are no large/deep scratches.
Powder SprayAfter filling the powder chamber, spray powder into a spittoon, etc., and
confirm that powder has not leaked from connections other than the Nozzle
Tip connection.
Water SprayAfter filling the powder chamber, spray powder into a spittoon, etc., and
confirm that water has not leaked from connections other than those of the
Nozzle Tip connection.
Adjustment RingAfter filling the powder chamber, spray powder into a spittoon, etc., and
confirm that powder spray amount changes when the Flow Volume Adjustment
Ring is switched from Min. to Max.
16
Cleaning wire (small)
Cleaning file
English
Cleaning wire (large)
Replace the Nozzle Tip.
The inside of the Nozzle
Tip is clogged with
powder.
When the Nozzle Tip
is removed and air is
supplied, air and powder
come out from the
TroubleCheck ItemCauseRemedy
Troubleshooting
Air and powder do not
come out.
Or air and powder come
10
out but the pressure is
Loosen the Nozzle screw with the supplied Ring Wrench and remove the Nozzle from
the Handpiece. Clean the Nozzle base with the supplied Cleaning wire (small) and
Nozzle end with the cleaning file. In addition, clean the inside of the Handpiece with
the Cleaning wire (large). If powder and air do not flow after performing the above
cleaning, perform Ultrasonic Cleaning with hot water. (See 8-1 Ultrasonic Cleaning.)
The inside of the
Handpiece is clogged
with powder.
Nozzle section.
When the Handpiece
is removed and air is
supplied, air and powder
come out from the
Handpiece connection.
not strong enough and
cleaning power is also
weak.
17
Insert the Cleaning wire (large) into the center hole on the end of the Handpiece
section connection of the Powder Case while turning the Wire and clean the inside. In
addition, clean the Internal Nozzle in the Powder Case with the cleaning file. If powder
and air do not flow after performing the above cleaning, perform Ultrasonic Cleaning
with hot water. (See 8-1 Ultrasonic Cleaning.)
The inside of the Powder
Case is clogged with
powder.
When the Handpiece
is removed and air is
supplied, air and powder
do not come out of the
Powder Case.
Cleaning wire (large)
Cleaning file
Clean with the supplied Cleaning wire or cleaning file (see the above figure) and blow
air to remove the remaining powder inside. If the problem is not resolved, perform
Ultrasonic Cleaning with hot water. Then, thoroughly dry the inside before use. (See
8-1 Ultrasonic Cleaning.)
Powder has partially
solidified with water, etc.,
in the middle of the air
passage.
Powder has partially
solidified in the
Handpiece or in some
part in the Powder Case,
reducing the passage
Replace powder with new powder.
Powder has long been
width.
Powder has partially
opened and exposed to
air.
solidified in the Powder
Case.
18
O-ring grooves
O-ring
English
Loosen
If you use the model for PHATELUS coupling or MACH coupling, confirm that the
O-ring is not damaged in the coupling joint section. If the O-ring is damaged, replace
it. If your model is other than those for PHATELUS and MACH couplings, confirm the
coupling for use.
Water is mixed in supply
air.
Taper Ring
Ring is not sufficient, the Powder Case may be ejected during use.
After O-ring replacement, securely tighten the Taper Ring. If tightening of the Taper
Open the filter in the air supply circuit, and the drain in the compressor, and check that
water has not accumulated.
of the supplied turbine oil and rub it in with your finger, etc. (See 7-3 Cleaning and
Washing - 7-4 Lubrication.)
Remove O-rings from the Handpiece joint section of the Powder Case and carefully
remove the powder adhered to the grooves and O-rings with the supplied cleaning
brush. Then, thoroughly wipe off the Handpiece joint section with a cloth soaked with
alcohol. Install the O-rings, apply a drop of the supplied turbine oil and rub it in with
your finger, etc. (See 7-3 Cleaning and Washing - 7-4 Lubrication.)
The O-ring is scratched. Replace the O-ring in the Handpiece joint section of the Powder Case, apply a drop
Check the O-ring in the
Powder is caught in
the O-ring grooves
and connection of the
Handpiece joint section
of the Powder Case.
Check O-rings in the
Powder Case and
Handpiece joint section.
coupling section has
become tight.
19
Water leaks from
between the Handpiece
and Powder Case.
It is difficult to rotate
the connection between
the Powder Case and
Handpiece connections.
*If the trouble is not resolved by any of the above remedies, contact your Authorized NSK Dealer for service repair.
Specifications
11
ModelRefer to 4. Correspondence Coupling
Hose connection typeRefer to 4. Correspondence Coupling
Air Pressure0.3 - 0.4MPa (3.0 - 4.0 kgf/cm
Max. Consumption Air17 NL/min
Supplied Water pressure0.1 - 0.2MPa (1.0 - 2.0 kgf/cm
Supply Water Flow Rate≥40ml/min
Use EnvironmentTemperature: 0 - 40°C, Humidity: 30 - 75%
Transportation and Store Environment Temperature: -10 - 50°C, Humidity: 10 - 85%,
Atmospheric Pressure: 500 - 1,060hPa
PerioMate Powder
Use EnvironmentTemperature: 15 - 25°C, Humidity: 30 - 75%
Transportation and Store Environment Temperature: 15 - 25°C, Humidity: 30 - 75%
Symbols
12
This product is Autoclavable up to Max.135ºC.
This product can be washed via Thermo Disinfector.
Conforms to CE European Directive of “Medical device directive 93/42/EEC.”
Manufacturer.
Authorized representative in the European community.
2
)
2
)
Do not reuse.
Use by
Batch Code
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
Warranty
13
NSK products are warranted against manufacturing errors and defects in materials. NSK reserves the right
to analyze and determine the cause of any problem. Warranty is voided should the product be not used
correctly or for the intended purpose or has been tampered with by unqualified personnel or has had non
NSK parts installed. Replacement parts are available for seven years beyond discontinuation of the model.
20
Option Parts List
14
ModelOrder CodeNote
MG-4H Multi GaugeZ109400PER-M4-P
MG-2/3H Multi GaugeZ109200PER-B2-P
Spare Parts List
15
ModelOrder CodeNote
Perio Mate PowderY9009382 Bottle Set (160g each)
Perio Mate Nozzle TipY100274140 pcs/package
PTL O-Ring setY900580PER-PTL-P
PM O-Ring setY900054Handpiece joint section of the Powder Case
Disposing Product
16
In order to avoid the health risks of operators handling the disposal of medical equipment, as well as the
risks of environmental contamination caused thereof, a surgeon or a dentist is required to confirm the
equipment is sterile. Ask specialist firms who are licensed to dispose of specially controlled industrial
wastes, to dispose the product for you.
English
21
Vielen Dank für den Kauf des Perio Mate von NSK.
Dieses Produkt ist ein Handstück für die Reinigung von sub- und supragingivalen Zahnoberflächen.
Bitten lesen Sie sich diese Betriebsanleitung vor der Verwendung gründlich durch, um sich mit Betrieb und
Wartungsmaßnahmen vertraut zu machen. Bewahren Sie diese Betriebsanleitung für spätere Verwendung
auf.
Deutsch
1. Funktionen
2. Benutzer und Verwendungszweck
3. Sicherheitsanweisungen für die Handhabung und Bedienung
4. Verfügbare Modelle
5. Packungsinhalt
6. Betrieb
7. Pflege nach dem Gebrauch
8. Reinigung und Sterilisation
9. Regelmäßige Wartungsprüfungen
10. Problembehandlung
11. Technische Daten
12. Symbol
13. Garantie
14. Stückliste
15. Ersatzteilliste
16. Entsorgung des Produkts
1
Ř Das um 360° frei drehbare Handstück sowie die große Beweglichkeit des Instruments auf der
Schnellkupplung ermöglichen den reibungslosen und komfortablen Betrieb des Perio Mate auch bei
hohem Luftdruck. Das leicht zu drehende Handstück ermöglicht den einfachen Zugang zu schwer
erreichbaren Mundregionen.
<Pulver zur Reinigung der Zahnoberfläche>
Ř NSK bietet zur Verwendung mit Perio Mate das Perio Mate Powder zur Reinigung der Zahnoberflächen
an. Die Kombination aus diesem Pulver und Perio Mate entfernt Zahnbelag und sub- bzw.
supragingivalen Biofilm schnell und effizient.
Ř Der Hauptbestandteil des Pulvers Perio Mate ist Glycin, ein Aminosäuretyp.
Benutzer: Zahnarzt/ärztin, Dentalhygieniker/in
Verwendungszweck: Perio Mate ist für sub- und supragingivale Plaqueentfernung konzipiert. Bei
subgingivaler Anwendung kann es die Parodontitistherapie und Periimplantitistherapie
unterstützen.
23
Sicherheitsanweisungen für die Handhabung und Bedienung
3
Lesen Sie bitte diese Sicherheitsanweisungen aufmerksam durch und benutzen Sie das Gerät
ausschließlich gemäß seines bestimmungsgemäßen Gebrauchs und entsprechend der Anweisungen.
Die Sicherheitsanweisungen sind dafür vorgesehen, mögliche Gefahren zu verhindern, die
Körperverletzungen oder Beschädigungen am Gerät verursachen können. Die Sicherheitsanweisungen
werden gemäß der Schwere des Risikos wie folgt klassifiziert.
KLASSE
ACHTUNG
VORSICHT
HINWEIS
Ř Stellen Sie die Verwendung des Produkts ein und suchen Sie einen Arzt auf, falls sich aufgrund der
Behandlung mit diesem Produkt eine Überempfindlichkeit wie Ausschlag oder Dermatitis entwickeln
sollte.
Ř Perio Mate kommt in parodontalen Taschen zur Anwendung. Daher ist bei Patienten mit Emphysemen
Vorsicht geboten. Stellen Sie die Verwendung des Produkts sofort ein und wenden Sie sich an einen
Arzt, falls sich aufgrund der Behandlung mit diesem Produkt ein Emphysem entwickelt.
Ř Dieses Instrument ist ausschließlich für die Verwendung durch Zahnärzte oder Dentalhygieniker
bestimmt.
Ř Verwenden Sie das Instrument nicht bei den folgenden Patienten:
1) Patienten mit Erkrankungen der Atemwege.
2) Patienten mit krankhaft tiefen Parodontaltaschen (6 mm oder mehr) mit Schleimhautläsionen.
Durch dauerhafte Applikation des Pulver-Luft-Gemischs in das Weichgewebe oder in die Nähe von
Speicheldrüsen kann Pulverluft eintreten und Emphyseme verursachen.
3) Patienten mit schweren Geschwüren an den Verdauungsorganen.
4) Patienten mit Nierenversagen.
5) Patienten mit Lungen- oder kardialen Störungen oder Beeinträchtigung der Atmung.
6) Patienten mit Schäden oder Anomalien in der Mundhöhle.
7) Patienten mit Hyperämie, Blutungen oder Entzündungen in der Mundhöhle.
8) Patienten, die zur Entwicklung von Entzündungen oder Geschwüren in der Mundschleimhaut neigen.
9) Patienten mit Allergien.
10) Je nach Konstitution des Patienten können in sehr seltenen Fällen allergische Symptome oder
Stomatitis verursacht werden. Stellen Sie die Verwendung des Produkts sofort ein, wenn derartige
Symptome festgestellt werden.
Ř Verwenden Sie Perio Mate nicht bei den nachfolgend aufgeführten Patienten. Beim Einsatz in den
Parodontaltaschen dieser Patienten kann die Verwendung des Produkts die Entwicklung einer
Bazillemie verursachen.
-Patienten mit endokardialen Entzündungen
-Schwangere und stillende Patienten
-Patienten mit ansteckenden Krankheiten
- Patienten mit Immunschwächekrankheiten (Neutropenie, Granulozytopenie, Agranulozytose,
Diabetes, Bluterkrankheit)
-Patienten, die mit Bestrahlung, Chemotherapie oder Antibiotika behandelt werden.
Risiko, das zu schweren Verletzungen oder Schäden am Gerät führen
kann, wenn die Sicherheitshinweise nicht beachtet werden.
Ein mögliches Risiko, das zu leichten bis mittleren Personen- oder
Sachschäden führen kann, wenn die Sicherheitshinweise nicht beachtet werden.
Allgemeine Produktinformationen, die besonders hervorgehoben werden,
um Störungen oder Leistungsminderungen zu vermeiden.
Ausmaß und Schwere der Gefährdung oder Schädigung
ACHTUNG
24
Ř Verwenden Sie dieses Produkt nur bei Patienten mit gesunden Parodontaltaschen wie für die Pflege
nach der Behandlung einer Zahnfleischerkrankung. Dieses Produkt kann Emphyseme verursachen,
wenn es an anderen als den hier genannten Patienten angewandt wird.
Ř Verwenden Sie dieses Produkt nur in Parodontaltaschen, die die folgenden Bedingungen erfüllen:
-Durch eine Zahnfleischerkrankung verursachte Parodontaltaschen.
-Parodontaltaschen von 6 mm oder weniger.
Ř Überprüfen Sie die Tiefe der Parodontaltaschen vor der Verwendung durch Messung (Sondierung) oder
Radiographie.
Ř Verwenden Sie als Reinigungspulver Perio Mate Powder. Wenn andere Pulver verwendet werden,
kann es zu einer Beschädigung der Düsenspitze kommen, wodurch die Substantia ossea oder Dentin
verletzt werden können oder ein Emphysem entstehen kann. NSK übernimmt keine Haftung für
etwaige Fehlfunktionen aufgrund der Verwendung anderer Pulver.
Ř Während der Verwendung sollten sowohl Anwender als auch Assistenzpersonal stets eine Schutzbrille
und Atemschutzmaske tragen. Ebenfalls sollte das Pulver während der Behandlung kontinuierlich
abgesaugt werden. Spülen Sie sofort mit reichlich Wasser und suchen Sie einen Augenarzt auf, wenn
Pulver in die Augen gelangt.
Ř Stellen Sie die Druckluftzufuhr auf einen Druck von 0,3 - 0,4 MPa (3,0 - 4,0 bar) ein. Wenn dieses
Produkt mit falschem Druck verwendet wird, können Anwender, Assistenzpersonal oder Patient durch
einen beschädigten Pulverkammerdeckel werden.
Ř Die Düsenspitze ist ein Einwegprodukt. Sie darf nicht wiederverwendet oder erneut sterilisiert werden,
da sie möglicherweise beschädigt oder verschmutzt ist.
Ř Stellen Sie vor der Verwendung sicher, dass die Düsenspitzenverpackung weder geöffnet noch
beschädigt ist. Verwenden Sie die Düsenspitze nicht, wenn die Verpackung geöffnet oder beschädigt
ist, da die Düsenspitze möglicherweise beschädigt oder verschmutzt ist.
Ř Halten Sie das Verfallsdatum auf der Düsenspitzenverpackung ein. Verwenden Sie die Düsenspitze
nicht, wenn das Verfallsdatum überschritten wurde, da die Düsenspitze möglicherweise beschädigt
oder verschmutzt ist.
VORSICHT
Ř Lesen Sie diese Bedienungsanleitung vor dem Gebrauch sorgfältig durch, um sich mit allen
Bedienungsfunktionen vertraut zu machen, und bewahren Sie sie zum späteren Nachschlagen auf.
Ř Achten Sie bei der Bedienung dieses Produkts stets auf die Sicherheit des Patienten.
Ř Es obliegt dem Anwender, über die Verwendbarkeit des Produkts am jeweiligen Patienten zu
entscheiden.
Ř Der Benutzer ist für die Funktionsprüfung, Wartung und ständige Überprüfung dieses Produkts
verantwortlich.
Ř Versuchen Sie nicht, das Produkt zu zerlegen und nehmen Sie keine Änderungen am Mechanismus
vor, sofern dies nicht von NSK in diesem Handbuch empfohlen wird.
Ř Schützen Sie das Produkt vor Stößen. Lassen Sie das Produkt nicht fallen. Da dieses Produkt ein
Präzisionsinstrument ist, treten möglicherweise Verformungen auf.
Ř Stellen Sie den Betrieb sofort ein und wenden Sie sich an Ihren autorisierten NSK-Händler, wenn das
Produkt nicht normal funktioniert.
Ř Achten Sie sehr sorgfältig darauf, dass kein Reinigungspulver in andere Schleimhäute (Augen, Nase
usw.) des Patienten als in der Mundhöhle gerät. Schützen Sie das Gesicht des Patienten außerdem mit
einem Handtuch oder einer Schutzbrille usw., um zu verhindern, dass Reinigungspulver in die Augen
gerät.
* Wenn der Patient Kontaktlinsen trägt, muss der Patient eine Schutzbrille tragen oder seine
Kontaktlinsen vor der Behandlung entfernen.
Ř Achten Sie darauf, dass Ihre Druckluft trocken und sauber ist. In der Luft vorhandenes Wasser oder Öl
kann zu einer Verfestigung/Verklumpung des Reinigungspulvers im Gerät führen.
25
Deutsch
Ř Befestigen Sie den Deckel der Pulverkammer sicher. Öffnen Sie den Pulverkammerdeckel während der
Verwendung nicht. Der Pulverkammerdeckel und das eingefüllte Pulver könnten ausgeworfen werden,
da der Druck in der Pulverkammer hoch ist.
Ř Blasen Sie vor dem Anschließen des Perio Mate im Bereich des Schlauchanschlusses vorhandene
Wassertropfen mit trockener Luft ab. Verbleibende Tropfen könnten in die Luftleitung gelangen und
somit zu einer Verfestigung/Verklumpung sowie einer daraus resultierenden Verstopfung des Geräts
führen.
Ř Befindet sich zu viel Pulver in der Pulverkammer, wird unmittelbar nach Beginn der Verwendung
möglicherweise eine nicht korrekte Pulvermenge aus dem Gerät gesprüht.
Ř Eine Eigenschaft des Perio Mate Powder besteht in der leichten Zerstäubung aufgrund seiner feinen
Partikel. Führen Sie die Befüllung langsam aus und gehen Sie vorsichtig vor, um zu verhindern, dass
Pulver verschwendet wird.
Ř Pulver in lange Zeit geöffneten Packungen zieht Feuchtigkeit an und kann sich in der Düse usw.
festsetzen. Achten Sie darauf, stets neues und trockenes Pulver zu verwenden. Verschließen Sie den
Deckel des Perio Mate Powder nach dem Füllvorgang fest und sicher.
Ř Tragen Sie auf die Lippen des Patienten Vaseline o. ä. auf, um ein Austrocknen oder Aufplatzen der
Mundwinkel zu vermeiden.
Ř Verwenden Sie zum Abwischen, Eintauchen oder Reinigen des Produkts kein Wasser mit hohem
Säuregehalt und keine Sterilisationslösungen.
Ř Führen Sie in regelmäßigen Abständen Funktionsprüfungen und Wartungen durch.
Ř Überprüfen Sie das Gerät vor der Anwendung am Patienten zunächst auf einwandfreies und sicheres
Funktionieren, wenn es längere Zeit nicht verwendet wurde.
Ř Ein U.S. Bundesgesetz schreibt vor, dass dieses Produkt nur durch einen oder im Auftrag eines
QuickPER-SR-P
Gelenk ISO 9168, Typ 2 (Midwest 4-Loch)PER-M4-P
Gelenk ISO 9168, Typ 1 (Borden 2-Loch)PER-B2-P
KaVo® und MULTIflex® sind eingetragene Handelsmarken der Kaltenbach & Voigt GmbH & Co., Deutschland.
Sirona® ist eine eingetragene Handelsmarke der Sirona Dental System GmbH, Deutschland.
W&H® und Roto Quick® sind eingetragene Handelsmarken der W&H Dentalwerk Burmoos GmbH, Österreich.
Bien-Air® und Unifix® sind eingetragene Handelsmarken der Bien-Air Dental S.A., Schweiz.
26
Packungsinhalt
5
Handstück 1 Satz
Pulverkammer (mit Pulverkammerdeckel) 1 Satz
Perio Mate Düsenspitze 40 Stück
Perio Mate Powder 1
Wartungskit
Reinigungsfeile (für die Düsenspitze der Innendüse) 1
Reinigungsdraht (S) (für die Wurzel der Düse) 1
Reinigungsdraht (L) (für sonstige Verwendungszwecke) 1
Reinigungsbürste 1
Düsenspitzenentferner Perio Mate 1
Ringschlüssel 1
Turbinenöl 1
Ausblasdüse 1
(Type abhängig vom jew. Perio Mate Kupplungstyp)
Ř Düse und Düsenspitze, Handstück, Pulverfach und Gebläsedüse können für das NSK
VORSICHT
Prophy-Mate neo nicht verwendet werden und sind nicht kompatibel.
Komponentenbezeichnungen
Deutsch
Pulverfluss-Einstellring
Düse
Düsenende
Deckelhalterung
Pulverkammerabdeckung
Gummidichtung
Handstückanschluss
O-Ring
Handstück
Pulverkammerdeckel
Entriegelungsring Handstück
Pulverkammer
27
Kupplungs- bzw.
Schlauchanschluss
Innendüse
Betrieb
6
ACHTUNG
VORSICHT
Ř Vermeiden Sie eine lange, kontinuierliche Verwendung und ermöglichen Sie es dem
Patienten, regelmäßig zu spülen.
Ř Setzen Sie Pulverkammer und Pulverkammerdeckel keinen Stößen oder Schlägen aus,
da dies an den beiden Teilen Kratzer oder Risse verursachen kann.
Ř Zerlegen Sie Perio Mate niemals während des Betriebs. Ziehen Sie während des Betriebs
niemals am Entriegelungsring des Handstücks oder an der Kupplungsentriegelung. Dies
kann dazu führen, dass sich die Komponenten des Perio Mate unerwartet lösen und das
Perio Mate Powder unkontrolliert austritt.
Ř Öffnen Sie den Pulverkammerdeckel nicht während der Verwendung des Perio Mate.
Dadurch kann Pulver unkontrolliert austreten.
Ř Stellen Sie die Druckluftzufuhr auf einen Wert von 0,3 - 0,4 MPa (3,0 - 4,0 bar)
ein. Die Verwendung des Produkts mit zu hohem Luftdruck erhöht das Risiko von
Emphysemen.
Ř Trocken Sie Perio Mate und alle Komponenten gründlich innen und außen.
Unzureichende Trocknung kann dazu führen, dass Pulver sich festsetzt.
Ř Achten Sie sehr sorgfältig darauf, dass kein Perio Mate Powder in andere Schleimhäute
(Augen, Nase usw.) als in der Mundhöhle gerät. Schützen Sie das Gesicht des Patienten
außerdem mit einem Handtuch oder einer Schutzbrille usw., um zu verhindern, dass
Pulver in die Augen gerät.
Ř Wischen Sie am Pulverfachdeckel haftende Chemikalien, Öle usw. umgehend ab. Diese
nicht zu entfernen kann Schäden verursachen.
6-1 Druckluft einstellen
Messen Sie den Versorgungsdruck am
Handstückpunkt und stellen Sie den Wert für
den Druck entsprechend den Angaben in der
Spezifikationstabelle ein. (Abb. 1)
* Für Informationen zum Mehrfachmessgerät Multi
Gauge sh. Kapitel 14. Stückliste.
Ř Überschreiten Sie nicht die in der Spezifikationstabelle genannten optimalen Drücke.
ACHTUNG
VORSICHT
Ř Ein zu hoher Luftdruck kann Risse oder Brüche an Pulverfach und Pulverfachdeckel
verursachen, was zu Verletzungen führen kann.
Ř Achten Sie darauf, dass Ihre Druckluft trocken und sauber ist. In der Luft vorhandenes
Wasser oder Öl kann zur inneren Verfestigung des Reinigungspulvers führen.
28
RückluftDruck
Handstück
Sprayluft-Druck
Wasserdruck
AntriebsluftDruck
Abb. 1
6-2 Anschluss des Pulverfachs an den Handstückschlauch
Ř Trocknen Sie vor dem Anschluss der Teile Kupplung, Handstück, Schlauchanschluss der
VORSICHT
Pulverkammer, die Pulverkammer innen und die Düse gründlich. Führen Sie außerdem
einen Leerlauf nur mit Luft durch, um eventuell in den Kanälen vorhandene Feuchtigkeit
herauszublasen.
Luft
Handstückanschluss
Luft
Luft
Kupplung
Luft
Pulverkammerdeckel
Luft
Pulverkammer
Luft
Innendüse
Schließen Sie den Schlauch an die Pulverkammer an.
-PER-PTL-P
Wird verwendet, wenn das Gerät auf einer Kupplung FM-CL (B2/B3), PTL-CL-LED (6 Stifte) oder
PTL-CL-FV (5-Loch) betrieben wird. Stecken Sie alle Kupplungen in den Schlauchanschluss der
Pulverkammer.
-PER-M4-P, PER-B2-P
Wird verwendet, wenn das Gerät auf einem Schlauch des Typs Midwest-4-Loch (M4) oder Borden-2Loch (B2) eingesetzt wird. Richten Sie jede Leitung am Schlauchanschluss der Pulverkammer mit einem
Loch im Schlauch aus. Stecken Sie die Leitung vorsichtig bis zum Anschlag gerade in den Schlauch und
ziehen Sie die Schlauchmutter fest an.
-PER-KV-P, PER-WH-P, PER-BA-P, PER-SR-P
Wird verwendet, wenn das Gerät auf einer Kupplung KaVo
®
, W&H®, Bien-Air® oder Sirona® usw.
eingesetzt wird. Ziehen Sie die Bedienungsanleitung für Ihre Kupplung heran und schließen Sie den
Perio Mate dementsprechend an.
Deutsch
29
6-3 Anschluss des Handstücks an die Pulverkammer
Ziehen Sie den Entriegelungsring des Handstücks
an der Pulverkammer zurück und drücken Sie das
Handstück gerade auf den Anschluss. Lassen Sie
den Entriegelungsring des Handstücks auslösen,
sobald sich das Handstück in Position befindet.
(Abb. 2)
Ziehen Sie nach erfolgtem Anschluss vorsichtig am
Handstück, um einen einwandfreien Anschluss zu
gewährleisten.
Entriegelungsring Handstück
6-4 Überprüfung vor der Behandlung
Überprüfen Sie das Gerät vor der Verwendung ohne Einsatz am Patienten. Stellen Sie die Verwendung
ein und wenden Sie sich an den Händler, wenn eine Anomalie wie zum Beispiel Vibration, Geräusch oder
Ähnliches festgestellt wird.
6-5 Befüllung mit Pulver
Entfernen Sie den Pulverkammerdeckel und befüllen Sie die Pulverkammer bis zu der mit‚ Max’
markierten Position mit Perio Mate Pulver (es können etwa 10 g Pulver eingefüllt werden).
Schließen Sie den Flaschendeckel des Perio Mate Pulvers nach dem Befüllen fest und sicher.
Schließen Sie den Deckel der Pulverkammer nach dem Befüllen fest und sicher. (Abb. 4)
Max. Füllstand Perio Mate Pulver
SCHLIESSEN
Abb. 2
Abb. 3Abb. 4
*Informationen zum Perio Mate Pulver finden Sie in der Ersatzteilliste.
Ř Wenn zu viel Reinigungspulver in die Pulverkammer gefüllt wird, wird unmittelbar nach
VORSICHT
dem Verwendungsstart möglicherweise keine korrekte Menge an Pulver versprüht.
Ř Eine Eigenschaft des Perio Mate Pulvers besteht in der leichten Zerstäubung aufgrund
seiner feinen Partikel. Führen Sie die Befüllung langsam aus und gehen Sie vorsichtig
vor, um zu verhindern, dass Pulver verstreut wird.
Ř Verwenden Sie kein Pulver aus bereits lange geöffneten Packungen oder übrig
gebliebenes Pulver. Die Verwendung derartigen Pulvers kann zur Verstopfung der Düse
führen.
Ř Verwenden Sie als Pulver ausschließlich Perio Mate Powder. Andere Pulver haben eine
andere Partikelgröße, daher kann es zu einer Verstopfung und unzureichender Leistung
des Geräts kommen. Die Verwendung anderer Pulver kann außerdem zu erheblichen
Schäden an der Düsenspitze führen.