Introduction
The Nox W7 LINK KITs come in three different versions and can be used with the Nox recording devices when configured for an ambulatory study. Note there is a slight difference in the configuration dependent on if the Nox T3 recorder or Nox A1 recorder is used (Refer to section Instuctions for Use).
W7 LINK KIT - A
To be used with:
• Resmed AirSe
7 LI
• Nox W
To be used w
• Radiometer TCM4, TCMCombiM, TCM40, TCM TOSCA
• Nonin RespSense
• Resmed S9™
• Nox W7 L
To be used with:
• SenTec Digital Monitor (Software SMB07.0x and SMB08.00)
The Nox W7 LINK KIT
Nox W7 LINK (
• Power Adapter (FW7721M/05)
• USA Plug Adapter
• European Plug Adapter
• UK Plug Adapter
• Nox USB Mini Cable
• DB9 Male to Male Mini Null Modem
• Nox User Instruction Leaflet
Intended Use
The Nox W7 LINK is a Medical Device Data System (MDDS) and allows for electronic conversion and wireless transfer of medical device data between an auxiliary medical device and a Nox recording device. The Nox W7 LINK is connected to a serial port of an auxiliary medical device. Before the data is transmitted over Bluetooth® link to the applicable Nox recording device it is converted from serial to Bluetooth format.
NK KIT - R
INK KIT - S
nse™10
ith:
packages include:
Uncord Serial Port Plug
F2M01SXA)
Scope
This leaflet is designed to instruc
LINK with supported auxiliary devices.
t users on how to use the Nox W7
Warnings and Cautions
• Warning: Do not use damaged equipment, sensors or accessories. This may result in bad performance of the Nox W7 LINK or patient/ operator injury.
• Warning: The Nox W7 LINK is not certified to be used for continuous monitoring where failure to operate can cause injuries or death of the patient.
• Warning: The RF function of the Nox W7 LINK might influence the performance of other devices in close proximity and lead to potential serious harm.
• Warning: Only use the medical grade power adapter provided by Nox Medical as part of the Nox W7 LINK KIT package to power the Nox W7 LINK. The use of other power adapters may result in electric shock or overheating of the power adapter/burns.
• Warning: Use of the Nox W7 LINK with auxiliary devices that have not been validated by Nox Medical may result in electric shock. Always use the Nox W7 LINK with validated auxiliary devices. Auxiliary devices need to comply with applicable IEC 60601-1 safety standards for medical electrical equipment.
• Warning: Do not use the Nox W7 LINK KITs in a Magne Imaging (MRI) environment. That might lead to serious patient harm.
tic Resonance
Instructions for Use
1. Preparing a Study with the Nox W7 LINK
When configuring a No
Nox W7 LINK in the Noxturnal software, the applicable Nox W7 LINK is added by entering its Bluetooth address in the relevant field. Two different addresses are used, depending on if the Nox A1 recorder is being used or the Nox T3 recorder:
Nox A1 Recorder: The Nox W7 LINK’s Bluetooth address (BDA) is used and has 12 digits. The BDA can be found on the Nox W7 LINK’s label.
Nox T3 Recorder: The Nox W7 LINK’s serial number (SN) is used and has 9 digits. The SN can be found on the Nox W7 LINK’s label.
2. Connecting to a Resmed S9™/Ai
The Nox W7 LINK is conne the Resmed Embletta Adapter. The Nox W7 LINK has thumbscrews which can be used for more secure placement. For thumbscrews use a 4mm flat screwdriver. Note: The supported modes are AutoSet and CPAP.
3. Connecting to a Radiometer Device
The Nox W7 LINK KIT - R includes a DB9 male to male mini null modem. Connect the null modem to the Nox W7 LINK, use thumbscrews for secure placement. For thumbscrews use a 4mm flat screwdriver. Connect the applicable serial port extension cable (available from Radiometer) to the Radiometer device and connect the Nox W7 LINK to the serial interface located on the extension cable.
4. Connecting to a Nonin RespSense Device
The Nox W7 LINK KIT - R includes a DB9 male to male mini null modem. Connect the null modem to the Nox W7 LINK. Connect the Nox W7 Link
to the serial interface output located on the side of the Nonin RespSense.
Use the thumbscrews for secure placement. For thumbscrews use a 4mm flat screwdriver.
5. Connecting to SenTec Digital Monitors
The Nox W7 LINK KIT - S includes a DB9 male to male mini null modem.
The null modem is connected to the serial interface output located at the
back of the SenTec device. Connect the Nox W7 LINK to the null modem. Use thumbscrews for more secure placement. For thumbscrews use a 4mm flat screwdriver.
6. Starting a Study
a. Connect the Nox W7 LINK to the applicable auxiliary device as instructed above (Step 2, 3, 4 or 5) and make sure the Nox Mini USB cable is connected to the NOX W7 LINK and the Power Adapter (FW7721M/05). b. Turn on the auxiliary device. The blue LED on the link starts blinking. c. Turn on the Nox recording device. d. The green LED on the NOX W7 LINK starts blinking, indicating Bluetooth connection with the Nox recording device. Note: Establishing Bluetooth connection may take up to 2 minutes. e. Now a Bluetooth connection has been established between the Nox recording device and the auxiliary device. The auxiliary device’s signals are being recorded.
States of the
The
NOX W7 LINK
status of the transmission. It blinks when the device is communicating with the Nox recording device. It is steady when the device is on standby. The blue LED indicates that the auxiliary device is streaming data to the device.
x recording device for an ambulatory study with the
cted to the serial interfa
Nox W7 LINK
is equipped with two LEDs. The green LED indicates the
’s LEDs
rSense™10 Device
ce output located on
Maintenance
When the Nox W7 LINK is not in use, disconnect it from auxiliary devices and store in a dry place.
Environmental Conditions
Operation:
Temperature: 0 °C to +40 °C (+32 °F to +104 °F) Relative Humidity: 15-95% (non-condensing) Atmospheric Pressure: 700 hPa - 1060 hPa
Storage: Temperature: -20 °C to +70 °C (-4 °F to +158 °F) Relative Humidity: 10-95% (non-condensing) Atmospheric Pressure: 700 hPa - 1060 hPa
Disposal
According to the regulation in Europe on Waste of Electrical and Electronic Equipment (WEEE) the equipment may not be disposed of as unsorted municipal waste. The equipment shall be collected separately and returned to the appropriate collection system available. Please contact your sales representative regarding take-back or recycling of the equipment.
Cleaning
The Nox W7 LINK tolerates cleaning with a soft cloth moistened with a mild soap solution.
Technical Specifications
Nox W7 LINK
• Physical Size: 36x17x60 (WxHxL) mm
Voltage Range: 5-30V DC
• Current Co
• Ba
• Operating Frequency: 2402-2480 MHz
• Tra
• Sensitivit
• Range: 1
Mini Null Modem
• DB9 Male to Male Mini Null Modem Adapter
Power Adapter (FW7721M/05)
• Input voltage 100 to 240 V AC (± 10%)
• Input curr
• Frequency 50 to 60 Hz
• Efficiency 80% typ. at full load
• Output 5V DC/1.5A
• Exchangeable plug system: European, U
Nox USB Mini Cable
• Mini-B plug
• Standard-A plug
nsumption (max): 300mA
ud Rate: Configured up to 230400 baud/s
nsmit Power: -8 to 16 dBm
y at 0.1% BER: -90 dBm 0m in line of sight when used with Nox recording devices
ent 200 to 100 mA
K, USA
Certifications
The Uncord Serial Port Plug F2M01SXA is certified according to CE and
FCC regulations and is fully Bluetooth® certified as an end product (QDID: B012540).
Uncord Serial Port Plug
R&TTE Directive 1999/5/CE, the European Community Directive 73/23/EEC and 93/68/EEC
• EN 300 328
• EN 301
• E
The power ad standards:
• EN 55011, EN 55022/ B
• FCC 47 Part
• EN 61000­ EN 61000-4-5, EN 61000-4-6, EN 61000-4-11
489-1/-17
N 60950
apter (FW7721M/05) conforms to the following EMC
IEC 60601-1-2
15
3-2, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4,
The power adapter (FW7721M/05) ful
following standards:
• ES 60601
• IEC 60601-1
• UL 2601
• Fulfills medi
cal application class B / BF / CF
F2M01SXA complies with the requirements of
fils class II, SELV according to the
FCC Information
This
Uncord Serial Port Plug
comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separatio
F2M01SXA has been tested and found to
n between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help
FCC Caution: Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
IMPORTANT NOTE:
Radiation Exposure Statement: The complies with FCC radiation exposure limits set forth for an uncontrolled environment. End users must follow the specific operating instructions for satisfying RF exposure compliance. To maintain compliance with FCC RF exposure compliance requirements, please follow operation instruction as documented in this manual.
This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. Highest reported body SAR is:
0.581W/kg.
FCC ID: R47F2M01SX
Uncord Serial Port Plug
F2M01SXA
Descriptions of Symbols and Labels
Operating instuctions /Consult instructions for use
Manufacturer information
Date of manufacture
Serial number
Catalouge number / Reference number
+50°C
-20°C
95%
10%
1060 hPa
700hPa
(01)1569431111XXXX (11)YYMMDD (21)WWWWWWWWW
(01)1569431111XXXX (11)YYDDMM (10)ZZZZZZ
Copyright Notice
No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language or computer language, in any form, or by any means: electronic, mechanical, magnetic, optical, chemical, manual, or otherwise, without the prior written authorization from Nox Medical.
Disclaimer
This document may contain typographical errors or technical inaccuracies. Nox Medical does not accept any liability for the use or misuse whether direct or indirect of the products, or for damages arising out of the use of or inability to use the products. Users must accept all responsibility for any results obtained by or concluded from data obtained by the products including software from Nox Medical. All clinical conclusions and decisions that are based on the use of this product are the responsibility of the user.
For distributor information go to: www.noxmedical.com
Reference number for Nox W7 LINK KIT - A is 54 40 12 Reference number for Nox W7 LINK KIT - R is 54 40 11 Reference number for Nox W7 LINK KIT - S is 54 40 10
Batch code / Lot number
Temperature limitation
Humidity limitation
Atmospheric pressure limitation
Keep dry
Fragile, handle with care
In compliance with the European Directive on Waste of Electrical and Electronical Equipment (WEEE) 2002/96/EC, do not dispose of this product as unsorted municipal waste
Federal Communication Commission (FCC) logo
CE marking indicating conformance to EC directives 93/42/EEC and 2007/47/EC concerning medical devices
®
Bluetooth
Unique Device Identifier (UDI): Application Identifier (01) represents the device identifier (DI) (“1569431111XXXX”), the Application Identifier (11) the production date/date of manufacture (“YYMMDD”, with “YY” the last two digits of the production year, “MM” the production month and “DD” the production day), the Application Identifier (21) the serial number of the device (“WWWWWWWWW”) if applicable, and the Application Identifier (10)ZZZZZZ the lot number of the device (“ZZZZZZ”) if applicable
wireless technology
NOX-BT1LEA-0101-REV2
Copyright ©2017 Nox Medical - All rights reserved