Novex EI8600, EI8700, EI8675, PowerEase 500 EI8600, PowerEase 500 EI8700 User Manual

...
USER GUIDE
PowerEase® 500 Power Supply
A programmable power supply for electrophoresis
Document Part Number A12162 Publication Number MAN0000738 Revision 9.0
For Research Use Only. Not for use in diagnostic procedures.
Table of Contents
Warning ................................................................................................................................................................... 4
Kit Contents ............................................................................................................................................................ 5
Product Specifications ........................................................................................................................................... 6
Front and Rear View of PowerEase® 500 ............................................................................................................ 7
Introduction .......................................................................................................................................................... 10
Description of PowerEase® 500 .......................................................................................................................... 10
Methods ....................................................................................................................... 12
Getting Started ..................................................................................................................................................... 12
Operational Modes .............................................................................................................................................. 13
Using PowerEase® 500 ......................................................................................................................................... 14
Printing .................................................................................................................................................................. 21
Custom Methods .................................................................................................................................................. 22
Troubleshooting ................................................................................................................................................... 24
Appendix ...................................................................................................................... 25
Instrument Symbols and Safety ......................................................................................................................... 25
Consignes De Securite ......................................................................................................................................... 26
Sicherheits Anweisungen ................................................................................................................................... 27
Run Conditions for Gels ..................................................................................................................................... 28
Repair, Maintenance, and Cleaning .................................................................................................................. 29
Background Information on Voltage, Current, and Resistance ..................................................................... 31
Accessory Products .............................................................................................................................................. 32
Technical Support ................................................................................................................................................ 33
3
Warning
Federal Communications Commission Advisory
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
4
Kit Contents
Types of Kits
This manual is supplied with the following kits:
Kit Catalog No.
PowerEase® 500 Power Supply (100/120 Vac 50/60 Hz) EI8600
PowerEase® 500 Power Supply (220/240 Vac 50/60 Hz) EI8700
Kit Components
Note
PowerEase® 500 Pre-Cast Gel System includes PowerEase
XCell
SureLock
®
500, Temperature Monitoring Probe,
Mini-Cell and XCell II™ Blot Module
EI8675
PowerEase® 500 Power Supply contains:
®
PowerEase
500 Power Supply 1 each
Instruction Manual 1 each
Extra Fuses 2 each
Power Cord (U.S., Canada, Europe, Taiwan and Japan) 1 each
Warranty Card 1 each
Temperature Probe (Optional) 1 each
®
See page 6 for specifications and detailed description of PowerEase
500 Power
Supply.
To ensure safe, reliable operation, always operate the PowerEase® 500 Power Supply in accordance with the manufacturer’s instructions. Always wear protective gloves and safety glasses when working in a laboratory environment. See safety information on pages 25–27.
Warranty information is provided on page 33.
5
Product Specifications
Specifications
Input Power
100–120 VAC: 100–120V, 50/60 Hz, 1.2 A at maximum load
Fuse: One 1A/250V/AGC fast blow fuse
220–240 VAC: 220–240VAC, 50/60 Hz, 0.5 A at maximum load
Fuses: Two 1A/250V/5 × 20 mm fast blow fuses
Output Range 1–500 V (minimum step size 1 V)
1–500 mA (minimum step size 1 mA)
0.1–50 W (minimum step size 0.1 W)
Automatic crossover on reaching set limits.
Accuracy Voltage: ± 2% or ± 2 volts
Current: ± 2% or ± 2 mA
Wattage ± 2% or ± 0.2 watt
Drift < 1% in 8 hours after 30 minute warm-up with constant supply voltage
Circuit Protection Open Circuit; Short Circuit; Thermal Protection Over Voltage, Current, Wattage
Safety Load Detection
Thermal Shutoff (with optional temperature probe)
Isolated Floating Ground
Memory Control Non-volatile, up to 24 hours
Custom Methods Stored to RAM
Computer Interface RS232C
Display LCD, view area 114 x 64 mm
240 × 128 dots
Dimensions 16.5 cm (w) × 18 cm (h) × 20.5 cm (d)
Weight 2.9 kg (6.4 lbs)
Operating Ambient temperature 4–30°C, ≤80% relative Environmental humidity, altitude < 2000 meters, indoor use only, Conditions pollution degree 2
6
Front and Rear View of PowerEase® 500
Front View of PowerEase
®
500
1
8
POWER
5
+
2
S
3
R
A
T
T
S
O
P
T
1. LCD Screen
2. Power Output Jacks, Positive (Red)
3. Power Output Jacks, Negative (Black)
4. Temperature Probe Jack
5. Functional Keys (Power, Start, Stop)
6. Directional Keys (Up, Down, Left, Right)
7. High Voltage LED
8. Power LED
P
SELECT
o
w
r
e
HIGH VOLTAGE
5
e
s
a
E
TEMP
7
0
0
6
4
Continued on next page
7
Front and Rear View of PowerEase® 500, Continued
Rear View of PowerEase
®
500
Fan
Printer Port
PRINTER
Power Entry Module with On/Off main power switch
RS232
RS232C Communi­cation Port
Continued on next page
8
Front and Rear View of PowerEase® 500, Continued
Key Pad
Power button secondary On/Off switch for the high voltage components, fan, and LCD
SELECT
screen.
Start button used to start a selected method program.
Stop button used to end the selected method and to backup to a previous screen.
Select button used to select a menu choice and to move forward to the next applicable screen.
Left directional button for moving left around the menu.
Right directional button for moving right around the menu.
Up directional button for moving up around the menu and to adjust numerical and
character values (higher).
Down direction button for moving down around the menu and to adjust numerical and character values (lower).
9
Introduction
Description of PowerEase
Product Description
Features of PowerEase
®
500
The PowerEase®500 Power Supply is a microprocessor-controlled power supply for electrophoresis of pre-cast and hand-poured mini-gels. The power supply is designed to simplify electrophoresis by combining high performance and programming flexibility with ease of use when running mini-gel and blotting applications.
®
The PowerEase you need to think only about the electrical parameters and limits for a single gel.
This manual describes the setup and operation of PowerEase including important information on safety and maintaining the unit.
The important features of PowerEase®500 Power Supply are:
Large LCD display allows electrical parameters to be displayed in either
graphical or numerical formats
Eight pre-programmed methods for running and blotting Novex
mini-gels
Four custom methods for running your custom applications
Simple three step operating procedure to start running or blotting Novex
pre-cast gels
Microprocessor control provides highly accurate electrical outputs resulting
in exceptional blotting capability
Parallel port for printing experimental results to most printers
Optional Temperature Probe to continuously monitor the buffer temperature
and shut the unit down in the event of overheating
Two sets of output jacks allows electrophoresis of multiple mini-cell units in
constant voltage, current, or watts
500 Power Supply is designed around the single gel concept:
®
500
®
500 Power Supply
®
pre-cast
®
10
Continued on next page
Description of PowerEase® 500, Continued
Flowchart
The following flowchart describes the various screens displayed on the
®
PowerEase
500 Power Supply and the keypad buttons (Start, Stop or Select)
used to navigate through the screens.
Power ON
Start-Up Screen
Start/Select
Gel Method Screen
Select
Gel Quantity Screen
Method Edit Screen
Method Run Large Display
Select
Method Run Large Display
StopStart
Previous Screen
Method Run Graphic Display
Start
Select, once Select, twice
Method Run Graphic Display
Stop
Previous Screen
Select
Method Run/Edit Screen
Select, onceSelect, twice
Method Run Large Display
11