Nova-Tech Temp 360 User Manual

Instruction Manual
RTD Datalogging Thermometer
Temp-360
Part of Thermo Fisher Scientific
68X526601 Rev 1 09/0
T
A
BLE OF CONTENTS
1. INTRODUCTION ............................... 1
2. SAFETY PRECAUTIONS.................. 2
3. SPECIFICATIONS ............................ 3
4. BATTERY INSTALLATION AND
REPLACEMENT................................ 6
5. INSERTING AND REMOVING
OPTIONAL RUBBER ARMOUR........ 7
6. ASSEMBLING OPTIONAL HANDSFREE ACCESORIES
7. CONNECTING A RTD....................... 9
8. KEY FUNCTIONS ........................... 10
9. DISPLAY OVERVIEW ..................... 11
10. MEASUREMENT MODE................. 13
11. HOLD FUNCTON ............................ 14
12. MIN, MAX, and AVE FUNCTION..... 14
13. DATA LOGGING ............................. 14
14. SETUP MODE................................. 15
15. GENERAL SETUP SCREEN .......... 16
16. CALIBRATION SCREEN
17. ALARMS SCREEN.......................... 20
18. DATA LOGGING SCREEN............. 25
19. CALIBRATION REPORT SCREEN
20. CLEAR / RESET SCREEN ............. 27
21. MAINTENANCE
22. CLEANING
23. BATTERIES
24. TROUBLE SHOOTING.................... 29
25. ACCESSORIES
26. WARRANTY
27. PRODUCT RETURN
28. INNOCAL® CALIBRATION AND
REPAIR SERVICES ........................ 32
............................... 28
...................................... 28
..................................... 28
............................... 30
.................................... 31
....................... 31
............. 8
................. 18
.26
1. INTRODUCTION
his versatile hand-held
T in
strument accu measurements instrument for
easy oper includes features:
provides highly
rate te
mperature
is designed
the follo
. The
ation
and
wing
• Menu driven setup and operation
• Datalogging for up to 2000 points
• USB output
• Operator selection of Celsius or Fahrenheit scale
• Resolution of up to 0.01°C/ °F
• Large backlit LCD and dot-matrix graphic LCD display
• Hold feature for temporarily retain a reading
• Displays minimum, maximum and average readings
• Field calibration capability
• Disabling of Auto-Off function
• Low battery warning
• Three-pin Din connector input
• Operates with a wide selection of probes
- 1 -
2. SAFETY PRECAUTIONS
WARNING:
1. This instrument is designed to accept low level signals supplied by standard RTDs. Under NO circumstances should the input voltage exceed the specified 50V RMS.
2. To prevent ignition of a hazardous atmosphere, batteries must only be changed in an area known to be non-hazardous.
CAUTION:
1. Do not use or store this instrument in microwave ovens or any abnormally hot or cold areas.
2. Weak batteries should not be left in the instrument. Dead batteries can leak and cause damage to unit.
DANGER:
1. Voltages present at the RTD may
also be present at the battery terminals. Always disconnect the RTD when changing batteries.
- 2 -
3. SPECIFICATIONS
RTD Thermometers
Temperature Range:
-201 to 1210°C (-330 to 2210°F)
Out of range display:
-
- -
-
Reso
lution
0.1 °C/°F:From -330.0 to -100 °C/°F
0.01 °C/°F:From -99.99 to 99.99 °C/°F
0.1 °C/°F:From 100.0 to 999.9 °C/°F
1 °C/°F:Above 1000 °C/°F
Accuracy
±0.1 °C/ ±0.2 °F: From -330.0 to -100 °C/°F ±0.03 °C/±0.06 °F: From -99.99 to 99.99 °C/°F ±0.1 °C/ ±0.2 °F: From 100.0 to 999.9 °C/°F ±1 °C/ ±2 °F: Above 1000 °C/°F
Display
Backlit Dot-matrix 50mm X 37.2mm
Data Logging
2000 points
Logging Interval
2 sec to 60 min
Min/Max/Avg Function
Yes
- 3 -
Auto Off (adjustable time)
Enable/Disable option available
Stability Criteria
Yes, upon stability of 5 seconds
Display update
0.5
sec per up
rate
date.
Input
One three-pin DIN connector
.
Input Protection 50V
rms
Storage
– 40°C to 65°C (– 40°F to 149°F)
Humidity
10% to 90% (non-condensing)
Battery Life Size: Three AA, 1.5V; Alkaline Life: 400 hours continuous, typical,
(without backlighting and buzzer on)
Dimensions Without Armor:
175mm (L) X 97mm (W) X 42mm (H)
With Armor:
180mm (L) X 102mm (W) X 52mm (H)
Weight with batteries Without Armor: 267g With Armor: 362g
- 4 -
Ingress protection:
Meets IEC-529 IP-54 for dust and water resistant enclosures (probe attached)
CE Compliance
EN61326-1/A1: 1998 (EU EMC Directive)
- 5 -
4. BATTERY INSTALLATION AND REPLACEMENT
The total battery life without backlighting is about 400 hours. Remaining battery power is indicated by the battery life indicator.
Indicator Voltage
Cell + 3 bars
More than 4.1 V
Cell + 2 bars
Cell + 1 bars Cell + empty bars
flashing Adaptor power supply
USB power supply USB power
More than 3.6 V
More than 3.0 V More than 2.85 V
Main power supply connected
Selected settings and
will remain in memory even power is turned are being replac
are
stored in
off, or
ed.
while
1. Before
2.
3.
4. Insert three
5.
instrume
changing
nt off
battery, tur
and disconnect RTD Loosen screw and back
of case.
Remove
pol Install
arity.
cover and
the t
hree
new
- 6 -
batteries
tighten
lift battery
AA batteries.
memory
batteries
n
cover
observing
screw
.
after
.
off
5. INSERTING AND REMOVING OPTIONAL RUBBER ARMOUR
1. To insert thermometer into the optional rubber armor, slide in from the top of meter before pushing the bottom edges of meter down to set it into position. Lift up the stand at the back of meter for bench top applications if necessary.
2. To remove thermometer from armor, push out from the bottom edges of meter until it is completely out of boot.
- 7 -
6. ASSEMBLING OPTIONAL HANDSFREE ACCESORIES
You can use the optional magnets and strap in the Handsfree Kit accessories for hands free operations.
- 8 -
7. CONNECTING A RTD
Use the correct 100 RTD (alpha =
0.003850) for your instrument. Using an incorrect probe type will result in erroneous readings. Insert the 3-pin plug into the mating connector on the top of the instrument.
If no probe is connected the display will read “open”.
s are sensitive
RTD element allow Mu probe by 5 will give you time
. W
time for the
ltip
lying
the time constant of the
required
hen
.
at the tip or
taki
ng measurements
reading to
the
stabilize.
approx
sensing
imate
,
- 9 -
8. KEY FUNCTIONS
F1
F2
F3
hold Freeze display
on/off
light
recall
log
Note: Function keys change in setup mode to provide advanced operation flexibility. Text above key will indicate function.
Step through Min, Max and Avg readings.
Toggle between F and C display
Toggle between menu and measure mode
Turns meter on and off (press and hold for 3 seconds to turn off)
Press momentarily to turn on backlight
Recalls and steps through stored readings
Stores current measured value to memory
- 10 -
9. DISPLAY OVERVIEW
The dot matrix display features a large primary display, smaller secondary displays for channel info or min/max/ave, and helpful annuciators for added measurement data
- 11 -
1 Power supply indicator
(Battery/Main adaptor/USB) 2 Date (format of mmm – dd) 3 Time (hour:min 4 Time format (am/pm/hrs) 5 Measurement mode 6 Hold function indicator 7 Stable indicator 8 Data logging indicator 9 PC data logging indicator 10 Main reading display 11 Current reading unit indicator 12 Min/Max/Avg elapsed time 13 Current Min/Max/Avg reading
14 Meter logging memory location 15 Measurement alarm active
16 Countdown time indicator with
17 Min/Max/Avg Display 18 Min/Max/Avg unit indicator 19 Function keys
indicator
indicator
countdown time
- 12 -
10. MEASUREMENT MODE
On initial start-up the meter will display the measured value for in the primary display.
F2
Pressing the
key will toggle reading
between F and C display.
F1
Pressing the
key initiates and toggles through Minimum, Maximum, and Average reading modes.
F3
Pressing
enter accesses Setup mode.
- 13 -
11. HOLD FUNCTON
Press the hold key to retain the reading on the display. Press the hold key again for normal operation.
12. MIN, MAX, and AVE FUNCTION
Press the F1 key to toggle between the minimum, maximum, and average readings. The minimum and maximum reading function is ideal for monitoring unattended operations while continually displaying every temperature change that occurs. The minimum and maximum values are sensed and automatically stored.
To exit and clear this function, press the F3 to access the Menu functions. See the Clear Reset menu section for more details.
13. DATA LOGGING
Press the
log
key to store the current reading to memory. The memory indicator M = 1234 shows the memory location for the next stored reading.
Press the
recall
key to review stored
readings.
See section on Data Logging for timed logging, and logging to a computer.
See section on Clear/Reset for information on clearing stored readings.
- 14 -
14. SETUP MODE
To access the setup mode from
measurement mode press the F3 key
(Menu).
Press ▲▼ keys on the meter key pad to scroll through options.
To enter a setup screen press Select key.
To return to measurement mode press
F3
Meas are listed
key. Following menu options
1. General Setup
2. User field calibration
3. Alarm settings
4. Data logging settings
5. View user calibration report
6. Clear/Reset options
- 15 -
F1
15. GENERAL SETUP SCREEN
The first page of the General Setup screens let you set Resolution, auto-off time, and password.
Press F1 to indicate you want to change
the setting of the current parameter or recall or log to move to the next parameter.
Press recall or log to change the options.
Press F2 to choose the next setting. Whenever set the options, press F1 for accepting the choice.
- 16 -
This screen below is used to reset/change password. In the event if uses forget his/her password, 5586 can be used to reset to a new value
On the second page you can set time and date.
- 17 -
16. CALIBRATION SCREEN
The thermometer is factory calibrated and does not require calibration before use. The Calibration function allows single point calibration of the thermometer, at any temperature point to compensate for RTD off-set error. It is NOT necessary to perform a field calibration to obtain specified meter accuracy. Use the field calibration feature to improve thermometer/probe accuracy or to compensate for RTD drift.
..
Before go into the calibration mode, must enter the password. Press F2 to change to the next digit. (Default Password is 9900)
- 18 -
There are two calibration options:
Offset – adjusts at a single point. Offset calibration can be performed at any temperature in the offset range of ±10.00 °C or 18.00 °F
Slope – adjusts at two points. The two calibration points must be at least 40.00 °C (72.00 °F) apart. The 2nd calibration point should be at a higher value than the 1st calibration point.
Use the or keys to adjust the value to match known temperature standard. Press
F1 to accept.
- 19 -
17. ALARMS SCREEN
There are two kind of alarm setting is available under the alarm setting options
Measurement alarm
Disable or enable the alarm by pressing recall or logand F1 to accept. Increase or decrease individual limit by pressing recall or log.
Meter in alarm mode
- 20 -
Countdown alarm
You can enable/disable the countdown alarm and set the countdown time from 5 sec to 1 hour.
After setting (enabling) the countdown alarm, the measurement screen should look like this:
- 21 -
Press F2 key to start the countdown:
Press F2 key in the middle of a countdown will stop the process:
You will need to restart the countdown by selecting the “start” key (F2).
Alternatively, you can also choose “Menu” to go into the timer menu and disabling the alarm.
- 22 -
The buzzer will sound for 30 seconds, or until the reset (F2) is pressed, at the end of the countdown.
Flashing announciator indicates that the countdown has expired
To reset the timer, select “restart” (F2) it. You display should look like this:
To repeat the functions, select F2
Note: The °C/ °F function is disabled when the count down function is enabled. To activate the
o
C/ oF function on
measurement screen, please go into the
- 23 -
“Alarm Setting page 2” and disable the count down alarm function.
When activated, the countdown timer temporarily over-rides the Auto-shutoff until the countdown is completed or manually stopped. If the meter is manually or auto shutoff,
The Countdown Alarm is automatically reset to “Disable” each time the meter is shut off. The 'Timer' key will revert back to the '°C/°F' key.
- 24 -
DATA LOGGING SCREEN
Press recall or logto choose the logging methods as auto or manual. If it is auto logging, using recall or logto set time interval. Its range is from 2min to 60min.
Data Transfer from Meter to Computer
Once the USB connection is establish with PC, press the Select button to download data from Meter to PC using HyperTerminal.
- 25 -
18. CALIBRATION REPORT SCREEN
The Calibration report will show the time and date along with results of the last user calibration.
- 26 -
19. CLEAR / RESET SCREEN
Press F1 to choose which data you want
to clear or reset. For calibration, logged data and reset all, you will have to enter the password to proceed. (Default Password is 9900)
- 27 -
20. MAINTENANCE
Properly us maintain calibr require service cleaning
ed, the th
ermome ation indefinitely other than
of the
housing and changing
ter
and
occasional
the batteries.
21. CLEANING
WARNING:
TO PREVENT IGNITION OF A HAZARDOUS ATMOSPHERE BY E
LEC
TROSTATIC DISCHARGE,
CLEAN WITH DAMP CLOTH.
Do
not
clean wi Use mild use excessive
22. BATTERIES
If there is no is turned AA batteries. Also
minals are clean and
ter properly install necessary INSTALLATION AND REPLA se
ction for
th
abrasives or solven
detergents,
never immerse nor
fluid.
display when
on,
check cond
the
thermometer
ition of the three
check
that the battery
batteries
ed. If
replacement is
, refer to the BATTERY
C
replacement proced
EMENT
ure.
should
not
are
of
ts.
- 28 -
23. TROUBLE SHOOTING
The following chart lists the most probable faults. There are no internal adjustments or user-replaceable parts.
FAULT ACTION
Check condition of batteries.
NO
Display
Check that batteries are inserted properly.
Display
shows
OVER /
Out of range indication
UNDER
Display
Shows
OPEN
Display
Shows
Err
No RTD connected in the Connector
If display shows this message other than during the field calibration mode, please return the instrument for servicing
- 29 -
24. ACCESSORIES
Replacem Accessories
ent
Meters and Meter
Item Oakton
Temp 360 thermometer 35426-40
Rubber Armour with Stand
Handsfree Kit (Two Magnets and a Strap)
General purpose probe (immersion Into liquids)
Penetration probe (meat, semi-soft Materials)
Surface probe (direct contact on Hot surfaces)
Air/gas probe
35427-80
35427-85
08117-70
08117-85
08117-78
08117-90
- 30 -
25. WARRANTY
The
Manufac
turer
warra
nts this product to be free from significant deviations from published period
of
adjus
tment warranty period corrected due
to
as
deter
Repair
costs period, or misuse or abus you
.
specifications
three years. If
is necessary within
, the
at
problem will be
no charge
for a
repair or
the
if it is not
misuse or abuse on your
mined by
those
the
outside
Manufacture
the
resulting
e,
may be invoiced
warr
from
anty
prod
part
r.
uct
to
26. PRODUCT RETURN
To limit seller or Manu and shipping in returning outsi retur reason protectio it Manufacturer will not be responsible for damage resulting from careless or insufficient packing.
charges and dela
the
de
of the
ning
the
for the
n, pack
agai
nst
possible damage or
fact
urer
for
structions
produ
before
ct,
either within or warranty peri produc
t,
ys, contact the
return
please
. For
the
carefully and insure
author
od. W
state the
your
loss. The
ization
hen
- 31 -
27. INNOCAL® CALIBRATION AND REPAIR SERVICES
Optimum performance of your temperature-measuring instrument is not a timeless condition. To ensure quality measurements, have your instrument calibrated regularly. Trust InnoCal® to satisfy your calibration and equipment repair needs. With over a decade of service, we've helped thousands of customers meet ISO, FDA, EPA, GLPs/cGMPs and other quality standards.
Conformity*
ISO/IEC 17025:2005 accredited NIST Handbook 150, 2000 Edition ANSI/NCSL Z540-2-1997 NIST Technical Note 1297 ISO 9000:2000
Fast Service
Our substantial inventory of replacement parts ensures a fast turnaround and prevents costly downtime. Most instruments serviced in five business days!
Excellent Value
Get quality at a fair price. Our InnoCal® NIST-traceable certificates offer extensive test data on a broad range of measurement parameters without breaking the bank!
- 32 -
Reliable Support
Trust in our free diagnostic support and troubleshooting advice. Our factory­trained metrologists and technicians are armed with years of experience and extensive technical data.
Convenient Reminders
It’s so easy to keep your instruments functioning properly. Based on your requirements, InnoCal will send you a reminder when it’s time to re-certify or service your instrument.
We provide you with the documentation you need to meet your most stringent quality requirements for the control of inspection, measuring, and test equipment.
Certification includes certificate of calibration with test data, including:
description and identification of the
item certified
condition of the item
issue date
identification of calibration procedure
calibration date
as found/as left test data (where
applicable)
signature of technician
statement of estimated uncertainty
list of equipment used to perform
calibration (including their calibration dates)
- 33 -
With today's high quality standards such as ISO 9000, certification is becoming increasingly important. Traceability is not a timeless condition. It must be verified and maintained over the life of the calibration to ensure the highest accuracy possible. When you have your calibration done by InnoCal, we will send you an automatic reminder when it is time to recalibrate your instrument.
Are your calibration certificates good enough?
InnoCal surpasses the competition by providing the most complete certificates as required by NIST. All of our certificates include measured data and point-by-point measurement uncertainty, and by request, we’ll provide test accuracy and test uncertainty ratios at no extra cost. Call us today and see why InnoCal is The Choice of Quality.
*See our Scope of Accreditation for any limitations.
- 34 -
Calibration test points against NIST­traceable standards
Four test points
across range of
instrument.
0, 100, 165,
230 C
(-4, 32, 446,
770F)
Meter only
MM­17000-04
Probe only
MM­17001-04
InnoCal—The Choice of Quality
866-InnoCal (866-466-6225)
InnoCalSolutions.com
System (meter + probe)*
MM­17002-04
- 35 -
- 36 -
TECHNICAL ASSISTANCE
sales@novatech-usa.com www.novatech-usa.com Tel: (866) 433-6682 Fax: (866) 433-6684 Tel: (281) 359-8538 Fax: (281) 359-0084
If
you have any questions abo
prod
of this
uct, contact the
Manufacturer or authorized se
For
more information on
ut the
ller.
OAKTON
use
Instruments Products, please contact your nearest distributor or visit our web site listed below:
OAKTON Instruments
P.O. Box 5136 Vernon Hills, IL 60061, USA Tel
(in U.S.): 888-462-5866 Tel (outside U.S.) 1-847-549-7600 Fax:
(1) 847-247-2984 Website: E-mail: info@oakton.com
Distributed by
www.4oakton.com
:
Loading...