Nova-Tech PR-a User Manual

Liquid Crystal Display (LCD)
ZERO Setting Switch
SCALE Switch
START/OFF Switch
Prism
The continuing evolution of the Palette –– The Debut of the Palette
1
2
3
Control of dilution ratio (multiples in %)
When water soluble cutting oils, aqueous cleaning agents, surfactants, liquid chemicals. etc. are diluted for use, their dilution ratio (multiples in %) can be managed by the Palette series.
Control of water content (%)
The water content in certain kinds of metal working oils, liquid cleansers, and liquid medicine, can evaporate while being used. To maintain the level of highest efficiency for the liquid, water needs to be replenished for optimal efficiency. In such cases, change in the water content can be managed by the Palette series.
Control the blend ratio of two liquids (%)
When a liquid additive is mixed with a primary liquid (for example, when flux or working oil is dissolved into a liquid cleanser), the blend ratio (%) of the foreign liquid can be managed by the Palette series.
Measures concentration of solutions with a light beam
Names of Components
Main
Applications
Metalworking Oils
Water soluble cutting oils, lubricating oils, working oils, rolling oils, quenching oils, stretching oils, mold lubricant (except insoluble agents), etc.
Detergents
Aqueous cleaning agents, quasi-aqueous detergents, hydrocarbon detergents, petrochemical detergents, alkaline detergents, isopropyl alcohol (IPA), etc.
Other liquids
Alkaline solutions, ammonia solutions, surfactants, caustic sodas, hydrogen peroxide water, glycerin, caulking agents, sealants, sealing agents, liquid resin, fire extinguishing solutions, salt water, adhesive agents, gelatin, protein water, DMF, freeze protective agents, thermal catalytic solutions (for freezer, solar apparatus), adhesives (all kinds), etching regents, anti-freezing agents, flux, corrosion protective agents, formalin, water glass, plating agents, lining agents, etc.
Examples of
Measurement
NEW!NEW
!
NEW
!
NEW!NEW
!
This instrument measures the refractive index of liquids and converts the measured value into the sugar concentration value, ‘Brix.’ The relationship between the refractive index and Brix (%) concentration is based on the type of liquid. With the user function, direct concentration measurements can now be displayed by establishing and programming the respective coefficient into the user scale.
The Palette series is the ultimate compact instrument which is the result of years of proven reliable performance, ease of operation and the functionality of design.
Sample size
is Just
1 Drop!
Measurement
time is Just
3 seconds!
About
displayed values
PR-32
Cat.No.3405
Four models of the Palette are available for your different measurement samples.
Main Features
Accuracy improved to ±0.1%
User scale feature
The user scale function has been added. Concentrations of various samples can be displayed by setting up a coefficient. Three separate user scales can be entered.
Convenient temperature display
A temperature display function has been added. Temperature is displayed by pressing a switch.
External-light-Interference (ELI)*
Conventionally, digital optical instruments have the potential to display incorrect measurement values if sampling is performed near a window or outdoors where lighting may be too intense. The new ELI technology provides security under these conditions to ensure accurate measurements.
Two-year warranty
ATAGO has extended the warranty term for the Palette series to 2 years. For the terms and conditions, contact ATAGO.
Automatic Temperature Compensation (ATC) function
With ATC, there is no need to worry about the measurement temperature.
Large, easy to read display
The large-sized liquid crystal panel makes it easy to read the measured result.
Apply 1 drop of sample solution onto the prism surface.
Press the START/OFF Switch.
Brix (%) value displayed in 3 seconds!
1
2 3
Measurement Method
NEW!NEW
!
NEW
!
NEW!NEW
!
Accuracy assured!!
The Palette is manufactured to ISO 9001 standards with the highest quality materials to ensure that the sample liquids such as water-soluble cutting oils, washing liquids, industrial liquids and various chemicals, will be measured accurately.
PR-101
Cat.No.3442
PR-201
Cat.No.3452
PR-301
Cat.No.3462
*
*
Calibration with the Standard Liquid LG.
Cat.No. 3405 3442 3452 3462
Measurement Range Brix0.0~32.0
%
Brix0.0~45.0
%
Brix0.0~60.0
%
Brix45.0~90.0
%
Minimum Indication Brix 0.1
%
Measurement Accuracy Brix ±0.1% (5~40°C)
Brix ±0.1% (10~40°C)
Brix ±0.1% (10~30°C)
Measurement 5~40°C 10~40°C
Temperature Automatic Temperature Compensation Automatic Temperature Compensation
Ambient Temperature 5~40°C 10~40°C
Sample Volume 0.1ml or more
Measurement Time 3 seconds (approximately)
Power Supply 006P dry battery (9V)
International
IP64 Water Resistant
Protection Class
Dimensions and Weight
17 (W) 9 (D) 4 (H) cm, 300g
PR-32 PR-101 PR-201 PR-301
* External Light Interference (ELI)
- Patent Pending
When intense light penetrates the prism of a digital refractometer, the light waves may interfere with the sensor resulting in inaccurate readings. To ensure accurate measurement results each time, the Palette is programmed with the ELI function to display the warning message “nnn” when intense direct light is detected. Forming a habit of shading the sample stage with your hand and re-pressing the START/OFF Switch (when the warning message from the ELI function is displayed) will ensure accurate measurement results each time.
sales@novatech-usa.com www.novatech-usa.com Tel: (866) 433-6682 Fax: (866) 433-6684 Tel: (281) 359-8538 Fax: (281) 359-0084