Nova-Tech pH-Dissolved Oxygen Meter User Manual

f
r
pH/Dissolved Oxygen/°C/°F Meter
pH/DO 300 Series
WATERPROOF
MODE
HOLD
ENTER
ON
OFF
CAL
MEAS
SETUP
ATC
READY
MEAS
C
%
I
P
67
Notes
OPERATING INSTRUCTIONS
Write down the name and information of your OAKTON distributor here.
OAKTON®35632-Series
Portable Waterproo
pH/Dissolved Oxyge
Mete
Table of Contents
1. Introduction ............................................................................................4
2. Display and keypad functions .............................................................5-6
2.1 Display...................................................................................................................................5
2.2 Keypad...................................................................................................................................6
3. Preparation..........................................................................................7-9
3.1 Inserting the batteries..........................................................................................................7
3.2 Probe information — Connecting the probes...............................................................8-9
4. Calibration.......................................................................................10-19
4.1 Important information on meter calibration .................................................................10
4.2 Preparing the meter for pH calibration..........................................................................11
4.3 pH calibration................................................................................................................12-13
4.4 Preparing the meter for DO calibration..........................................................................14
4.5 Dissolved oxygen calibration in % saturation..........................................................15-16
4.6 Dissolved oxygen calibration in mg/l (ppm) mode ...............................................17-18
4.7 Temperature calibration....................................................................................................19
5. Measurement...................................................................................20-24
5.1 Taking DO readings......................................................................................................20-21
5.2 Taking pressure/salinity compensated DO measurements...................................21-22
5.3 Taking pH readings......................................................................................................23-24
5.4 Manual temperature compensation................................................................................25
6. Hold function........................................................................................26
7. Backlit LCD ............................................................................................27
8. Advanced SETUP functions..............................................................28-55
8.1 Advanced setup mode overview................................................................................29-31
pH measurement mode overview ................................................................................29
% saturation measurement mode overview ...............................................................30
mg/l (ppm) measurement mode overview.................................................................31
8.2 pH mode—Program P1.0: Viewing previous calibration data....................................32
8.3 pH mode—Program P2.0: Viewing electrode data..................................................33-34
P2.1 Electrode offset .........................................................................................................33
P2.2 Electrode slope .........................................................................................................34
8.4 pH mode—Program P3.0: pH measurement configuration...................................35-37
P3.1 READY indicator and auto endpoint function....................................................35
P3.2 Selecting number of pH calibration points..........................................................36
P3.3 Selecting °C or °F .....................................................................................................37
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
8.5 pH mode—Program P4.0: Resetting to factory default settings .................................38
8.6 % saturation mode—Program P1.0: Dissolved oxygen parameters
Pressure adjustment mode..............................................................................................39
8.7 % saturation mode—Program P2.0: Offset adjustment................................................40
8.8 % saturation mode—Program P3.0: Previous calibration information......................41
8.9 % saturation mode—Program P4.0: Electrode properties...........................................42
8.10 % saturation mode—Program P5.0: Unit configuration .......................................43-44
P5.1 READY indicator and auto endpoint function....................................................43
P5.2 Selecting °C or °F temperature readout................................................................44
8.11 % saturation mode—Program P6.0: Adjusting LCD brightness ...............................45
8.12 % saturation mode—Program P7.0: Resetting to factory default .............................46
8.13 mg/l (ppm) mode—Program P1.0: Dissolved oxygen parameters....................47-48
P1.1 Pressure adjustment mode.....................................................................................47
P1.2 Salinity adjustment mode .......................................................................................48
8.14 mg/l (ppm) mode—Program P2.0: Previous calibration information.....................49
8.15 mg/l (ppm) mode—Program P3.0: Electrode properties..........................................50
8.16 mg/l (ppm) mode—Program P4.0: Unit configuration ........................................51-53
P4.1 READY indicator and auto endpoint function....................................................51
P4.2 Selecting mg/l or ppm units ..................................................................................52
P4.3 Selecting °C or °F temperature readout................................................................53
8.17 mg/l (ppm) mode—Program P5.0: Adjusting LCD brightness................................54
8.18 mg/l (ppm) mode—Program P6.0: Resetting to factory default ..............................55
9. Probe care and maintenance...........................................................56-60
9.1 pH electrode care..........................................................................................................56-57
9.2 DO probe care .....................................................................................................................57
9.3 Membrane replacement ...............................................................................................58-59
9.4 Electrolyte solution ............................................................................................................60
10. Troubleshooting..................................................................................61
11. Error Messages....................................................................................62
12. Specifications......................................................................................63
13. Accessories.....................................................................................64-65
14. Appendix 1: Meter theory ..................................................................66
15. Appendix 2: Pressure vs. Altitude table .............................................67
16. Appendix 3: Factory default settings.................................................68
17. Warranty .............................................................................................69
18. Return of Items ...................................................................................69
32
1. Introduction
pH/Dissolved Oxygen/°C/°F Meter
pH/DO 300 Series
WATERPROOF
MODE
HOLD
ENTER
ON
OFF
CAL
MEAS
SETUP
ATC
READY
MEAS
C
%
Thank you for selecting an OAKTON meter. This OAKTON portable meter is a microprocessor-based instrument that measures pH, dissolved oxygen and temperature. It’s completely waterproof—and it floats! Your meter has many user-friendly features, all of which are accessible through the membrane keypad.
Your meter includes batteries. Please read this manual thoroughly before operating your meter.
2. Display and Keypad Functions
2.1
Display
The LCD has a primary and secondary display.
• The primary display shows the measured pH or dissolved oxygen reading.
• The secondary display shows the temperature of the reading. The display also shows error messages, keypad functions and program functions.
Primary display
1.
2.
3.
4.
5.
SETUP
16.
15.
READY HOLD
ON
14.
13.
1. SETUP mode indicator
2. MEASurement mode
indicator
3. CALibration mode
indicator
4. pH indicator
5. % saturation indicator
6. Salinity input indicator
7. Oxygen solubility
indicators
OFF
12.
MEAS
CAL
ERR
10.
11.
8. Temperatur e
indicators
9.Automatic Temperature
Compensation indicator
10. ERRor indicator
11. Low battery indicator
12. Probe indicator
13. Calibration data
indicator
C
ATC
Secondary display
14. Function ON/OFF
indicator
15. HOLD indicator
16. READY indicator
pH%
ppt ppm mg/l
F
9.
6.
7.
8.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
54
2.2
Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the LCD.
ON/OFF............Powers and shuts off the meter.
HOLD ..............Freezes the measured reading. To activate, press HOLD while in
measurement mode. To release, press HOLD again.
NOTE:
When auto endpoint feature is switched on, it automatically holds reading after 5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release.
MODE..............Selects the measurement parameter. Press MODE to toggle between
pH; DO % saturation; DO mg/l (ppm). In pH and DO mg/l (ppm) calibration modes press MODE to access temperature calibration. NOTE: To switch between mg/l and ppm modes, see page 52.
CAL/MEAS......Toggles user between Calibration and Measurement mode.
• If you were in pH mode, press CAL/MEAS to enter
pH calibration mode.
• If you were in DO % saturation mode, press CAL/MEAS to enter
DO % saturation calibration mode.
• If you were in DO mg/l (ppm) mode, press CAL/MEAS
to enter DO mg/l (ppm) calibration mode.
NOTE: Temperature calibration is available from both pH and DO mg/l (ppm) calibration modes; see page 19 for directions. In advanced set-up mode: Press CAL/MEAS to return to main menu from sub menus. Press CAL/MEAS again to return to measurement mode from main menu.
ENTER ............Press to confirm values in Calibration mode and to confirm
selections in SETUP mode.
▲▼ ..........Press in SETUP mode to
scroll through subgroups. Also lets you increment/ decrement the values in the mg/l and temperature calibration modes.
ON
OFF
HOLD
MODE
SETUP..............Press to enter SETUP
mode. SETUP mode lets you customize meter
CAL
MEAS
ENTER
preferences and defaults, and view calibration and probe data.
..........................LIGHT Press to activate
backlit display.
SETUP
3. Preparation
3.1
Inserting the Batteries
Four AAA batteries are included with your meter.
1. Use a Phillips screwdriver to remove the two screws holding the battery cover.
See figure below.
2. Lift off battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position. Screw cover back into place.
Remove these two screws to access battery compartment
Battery compartment
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
76
3.2
Probe information
Your meter uses two probes:
pH electrode with BNC connector
dissolved oxygen/temperature probe with a notched 6-pin connector
The temperature sensing element built into the dissolved oxygen probe will also compensate for pH readings as long as both probes are in your solution at the same time.
You can use any standard pH electrode with a BNC connector with this meter. If you do not have the DO probe connected you will need temperature probe 35618-05 to provide temperature compensation for pH readings (not included order separately). For the dissolved oxygen/temperature probe, you must use an OAKTON dissolved oxygen/temperature probe with the notched 6-pin connector. For replacement probes, see the “Accessories” section, pages 64-65.
NOTE: Keep connectors dry and clean. Do not touch connectors with soiled hands.
To connect the pH electrode:
1. Slide the BNC connector of the probe over the BNC connector socket on the
meter. Make sure the slots of the connector are in line with the posts of the socket. Rotate and push the connector clockwise until it locks.
See figure below.
2. To remove probe, push and rotate the connector counterclockwise. While
holding onto the metal part of the connector, pull probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
To connect the dissolved oxygen probe:
NOTE: Do not substitute other DO probes. For a replacement DO probe, see the
“Accessories” section, page 64.
1. Line up the notch and 6 pins on the meter with the holes in the 6-pin connector.
Push down and turn the locking ring clockwise to lock into place.
See figure below.
2. To remove probe, turn the locking ring counterclockwise on the probe connector.
Pull probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
pH Electrode (with BNC connector)
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
DO probe (with 6-pin connector)
locking ring
top view of 6-pin
connector of
DO probe
98
4. Calibration
4.1
Important Information on Meter Calibration
When you recalibrate your meter, old calibration points are replaced on a “point by point” basis in pH and completely for % saturation and mg/l (ppm).
For example:
pH:
if you previously calibrated your meter at pH 4.01, 7.00, and 10.01, and you recalibrate at pH 7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01.
% saturation and mg/l (ppm):
calibration data.
To view current calibration points:
pH: Program P1.0 in the SETUP section, page 32.
% saturation: Program P3.0 in the SETUP section, page 41.
mg/l (ppm): Program P2.0 in the SETUP section, page 49. To completely recalibrate your meter, or when you use replacement probes,
it is best to clear old calibration data by resetting the meter.
To reset the meter to its factory defaults:
pH: Program P4.0 in the SETUP section, page 38.
% saturation: Program P7.0 in the SETUP section, page 46.
mg/l (ppm): Program P6.0 in the SETUP section, page 55. NOTE: Resetting the meter will set meter to factory defaults.
pH, % saturation and mg/l (ppm) must be reset separately.
All new calibrations automatically override existing
4.2
Preparing the Meter for pH Calibration
Before starting pH calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used. For example, if you shut the meter off in "pH" units, the meter will read "pH" units when you switch the meter on.
Do not reuse pH calibration solutions after calibration. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. See pages 64-65 for information on our high-quality pH calibration solutions.
Be sure to remove the protective rubber cap from the pH electrode before calibration or measurement. If the electrode has been stored dry, wet the probe in
tap water for 10 minutes before calibrating or taking readings to saturate the pH electrode surface and minimize drift.
Wash your probe in deionized water after use, and store in electrode storage solution. If storage solution is not available, use pH 4.0 or 7.0 buffer.
The pH/DO 300 meter is capable of up to 5-point pH calibration at pH 1.68, 4.01, 7.00,
10.01, and 12.45 to ensure accuracy across the entire pH range of the meter . The meter automatically recognizes and calibrates to these standard buffer values, which makes pH calibration faster and easier.
To calibrate at 5 points you must select a 5-pt calibration in SETUP mode program P3.2 on page 36. You can select 2, 3, 4 or 5 calibration points for pH calibration.
For directions on how to calibrate your meter:
• See sections 4.2-4.3 on pages 11-13 for pH calibration
• See section 4.4-4.5 on page 14-16 for % saturation calibration
• See section 4.4 and 4.6 on page 14, 17-18 for mg/l (ppm) calibration
• See section 4.7 on pages 19 for Temperature Calibration
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
1110
4.3
pH calibration
NOTE: We recommend that you perform at least a 2-point calibration using standard
buffers that bracket (one above and one below) the expected sample range. You can perform a 1-point calibration, but make sure that the buffer value is close to the sample value you are measuring.
The temperature sensing element is in the DO probe. For temperature compensated readings and calibration, immerse the DO probe into the calibration buffer while calibrating pH.
pH Calibration
1. If necessary, press the MODE key to
select pH mode. The pH indicator appears in the upper right hand corner of the display.
See figure
A
2. Rinse the pH electrode and DO probe
thoroughly with de-ionized water or a rinse solution. Do not wipe the probe;
this causes a build-up of electrostatic charge on the glass surface.
3. Dip both probes into the calibration
buffer. The end of both probes must be completely immersed into the sample. Stir the probe gently to create a homogeneous sample.
4. Press CAL/MEAS to enter pH calibration
mode. The CAL indicator lights. The primary display will show the measured reading while the smaller secondary display will indicate the pH standard buffer solution.
See figure
B
5. Wait for the measured pH value to
stabilize.
See figure
C
A
B
READY
C
MEAS
pH
C
ATC
CAL
pH
pH
CAL
pH
pH
6. Press ENTER to confirm calibration.
The meter is now calibrated to the current buffer.
See figure
D
• If you are performing multipoint calibration, go to step 7.
• If you are performing one-point calibration, go to step 9.
7. Rinse the probe with de-ionized water
or a rinse solution, and place it in the next pH buffer.
8. Follow steps 5 to 8 for additional
calibration points.
9. When calibration is complete,
press CAL/MEAS to return to pH measurement mode.
Notes
To exit from pH Calibration mode without confirming calibration, DO NOT press ENTER in step 6. Press CAL/MEAS instead.
If the selected buffer value is not within ±1.0 pH from the measured pH value: the electrode and buffer icon blink and the ERR annunciator appears in the lower left corner of the display.
To limit the number of pH buffer values available during calibration, see section P3.2 on page 36.
CAL
READY
pH
pH
D
If the READY indicator has been activated (set up program P3.1—see page 35), the READY annunciator lights when the reading is stable.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
1312
4.4
Preparing the meter for DO calibration
4.5
Dissolved Oxygen calibration in percent saturation
Before calibrating your meter, make sure to rinse the DO probe well with de-ionized water or rinse solution.
NOTE: Do not let membrane surface of the DO
probe touch any other surface. The DO probe guard (the piece with holes fitted over the end of the probe) protects the membrane; make sure this is always attached to the probe while it is in use.
You can calibrate for dissolved oxygen in either % saturation or mg/l (ppm). All new DO
calibrations automatically override existing DO calibration data.
Before Dissolved Oxygen calibration
Before starting DO calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used (either pH, mg/l, ppm, % saturation). For example, if you shut the meter off in “mg/l” units, the meter will read “mg/l” units when you switch the meter on.
NOTE: Most users will calibrate to 100% saturation
even when working in mg/l. When calibrat­ing the meter in mg/l mode, you can make fine adjustments, typically to a midrange value between 10 and 100%. If you are calibrating to a midrange value, you need to set the 100% saturation value first.
DO Probe guard
You can calibrate this meter quickly and easily in air. The exact calibration value depends on barometric pressure. The meter is set to a factory default of 760 mm Hg, which results in a calibration value of 100% saturation in air.
NOTE: If the barometric pressure setting has been changed from 760 mm Hg, the
calibration value in air will automatically adjust to a value other than 100%. The adjusted value will be correct for the new barometric pressure setting.
• See page 39 to change the pressure setting.
• See Appendix 2 on page 67 for a chart of adjusted % saturation values.
To calibrate 100 % saturation:
1. Rinse the probe well with deionized
rinse water or rinse solution. For best accuracy, wrap the end of the probe in a damp cloth. Do not touch the membrane.
2. Press the MODE key to select the
% saturation mode.
See figure
A
3. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show 100.0 (see NOTE above).
See figure
B
4. Hold the probe in the air (or in damp
cloth). Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup).
5. Press the ENTER key. The meter auto-
matically calibrates to 100.0% air satura­tion and returns to Measurement mode.
See figure
C
A
B
READY
C
READY
READY
MEAS
MEAS
CAL
%
C
ATC
%
%
C
ATC
Notes
The reading in the primary display in step 3 must read at 50% or above for the calibration to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
You can offset your % DO calibration: see page 40 for directions.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
1514
To calibrate 0 % saturation:
1. Rinse the probe well with deionized
rinse water or rinse solution. Do not touch the membrane.
2. Press the MODE key to select the
% saturation mode.
See figure
D
2. Place the probe in zero oxygen
solution. Stir the probe gently to homogenize the sample. Make sure that the sample is continuously flowing past the membrane sensor.
3. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show 0.0.
See figure
E
4. Wait for the reading to stabilize. If the
Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup).
5. Press the ENTER key. The meter auto-
matically calibrates to 0.0% saturation and returns to Measurement mode.
See figure
F
Notes
The reading in the primary display in step 3 must read at 10% or under for the calibra­tion to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
You can offset your % DO calibration: see page 40 for directions.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
4.6
Dissolved Oxygen calibration in mg/l (ppm) mode
Calibrating the meter to 100 % in saturation mode will also calibrate the mg/l mode at the value in mg/l corresponding to 100% saturation. This should produce acceptable results in most applications.
MEAS
READY
D
%
C
ATC
This meter also lets you make a calibration adjustment in mg/l (ppm) mode without affecting your % saturation calibration.
To select between mg/l and ppm units, see page 52.
To calibrate in mg/l (ppm) mode:
1. Calibrate 100% saturation per section 4.5
on page 15.
2. Rinse the probe well with deionized
rinse water or rinse solution. Wipe the
READY
CAL
%
outside of the probe carefully (do not touch the membrane).
3. Dip the probe into a sample of known
E
MEAS
READY
%
C
ATC
F
oxygen concentration. Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup).
NOTE: The sample must continuously flow
past the membrane for accurate readings. Astirrer can keep the solution agitated; order a magnetic stirrer on page 65.
MEAS
READY
mg/l
C
ATC
A
4. Press the MODE key to select the
mg/l (ppm) mode.
See figure
A
5. Press the CAL key. The CAL indicator
will appear above the primary display. The primary display shows the current value of the measurement and the sec­ondary display shows the temperature.
See figure
B
6. Press the and keys to adjust
the reading.
7. Press the ENTER key. The meter
automatically calibrates to the value you entered and returns to Measurement mode.
See figure
C
READY
B
READY
C
CAL
mg/l
C
ATC
CAL
mg/l
C
ATC
1716
Notes on mg/l (ppm) mode calibration
During mg/l (ppm) calibration, the meter adjusts to the barometric pressure value that is programmed into the meter. It also adjusts to the salinity value that is programmed into the meter.
You can change the barometric pressure value and salinity value in the mg/l (ppm) SETUP mode (see pages 47-48 for directions).
Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display.
4.7
Temperature Calibration
The built-in temperature sensor included in the DO probe is factory calibrated. Calibrate your sensor only if you suspect sensor drift that may have occurred over a long period of time or if you have a replacement DO probe. Make sure DO probe is attached before you perform the temperature calibration.
1. Switch the meter on. Press MODE to
select either pH or mg/l (ppm) Measurement mode.
2. Press the CAL/MEAS key to enter
pH or mg/l (ppm) calibration mode.
The CAL indicator will appear above the primary display.
See figure
A
READY
A
CAL
mg/l
C
ATC
3. While in pH or mg/l (ppm) calibration
mode, press the MODE key to enter temperature calibration mode.
The primary display shows the last set temperature reading and the secondary display shows you the factory default temperature value.
See figure
B
4. Compare the primary display reading
to a NIST-traceable thermometer or another thermometer known to be accurate.
5. Press the or keys to adjust
the primary display reading to agree with your temperature standard.
See figure
C
6. Press the ENTER key to confirm
temperature calibration and return to Measurement mode.
READY
B
READY
C
CAL
C
ATC
CAL
C
ATC
Notes
To exit from Temperature Calibration mode without confirming calibration, DO NOT
press ENTER in step 6. Press CAL/MEAS instead.
Temperature calibration is restricted to ±5°C from the factory default value displayed during calibration (shown in the
secondary display).
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
1918
5. Measurement
5.1
Taking DO readings
During measurement, the probe can be:
• fully immersed in the solution
• partially immersed in the solution Do not allow the probe’s membrane surface to touch
anything! The probe guard (the piece with holes fitted over the end of the probe) protects the membrane; you should leave this piece attached to the probe at all times.
IMPORTANT: since the DO probe consumes oxygen from the sample, the sample must constantly flow past the membrane to achieve more accurate readings. You can use a stirrer to accomplish this (see “Accessories” on page 65 for a magnetic stirrer).
To take measurements:
1. Rinse the probe well with deionized
rinse water or rinse solution.
2. Select the appropriate measurement
mode. Press the MODE key to toggle between modes:
• % saturation.
mg/l (ppm)**,
• pH
3. Dip the probe into the sample. Stir the
probe gently to homogenize the sample. Make sure that the sample is continu­ously flowing past the membrane sensor.
A
4. Note the reading on the display. If the
READY indicator is switched on, it will appear when the reading is stable.
See figure
**To select between mg/l and ppm units, see page 52. The ATC indicator should appear in the lower right hand corner of the display.
If it does not, this indicates an error.
CONTINUED ON NEXT PAGE
A
MEAS
READY
Probe guard
mg/l
C
ATC
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is stable. Switch the READY indicator on or off in the SETUP program P5.1 for % saturation measurements (page 43) or program P4.1 (page 51) for mg/l (ppm) measurements.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically “hold” the reading. The “hold” indicator appears on the left side of the display. Press the HOLD key to release the reading. Switch the Auto endpoint feature on or off in the SETUP program P5.1 for % saturation measurements (page
43) or program P4.1 (page 51) for mg/l (ppm) measurements.
5.2
Taking pressure/salinity compensated DO measurements
If necessary, you can adjust the pressure and salinity values of your measurements in the SETUP mode. The DO meter will automatically compensate for salinity and pressure based on the values entered in the setup functions. The meter is factory set at 760 mm Hg (101.3 Pascals) pressure adjustment and a factor of 0.0 salinity adjust­ment. See Appendix 2 on page 67 for a “Pressure vs Altitude” table.
Pressure adjustment
NOTE: This adjustment appears in both
the % saturation and mg/l (ppm) measurement modes.
1. Press the MODE key to select % satura-
tion or mg/l (ppm) mode. Press the SETUP key to enter Set Up mode.
See figure
A
2. Press the ENTER key until the upper
display shows a number and the lower display shows “Hg” or “PA”.
3. Use the and keys to enter the
barometric pressure. The upper display will show the value you have entered.
See figure
B
4. Press ENTER to confirm pressure value.
SETUP
A
SETUP
B
5.Press CAL/MEAS to return to measure-
ment mode, or continue with step 4 on page 22 to make a salinity adjustment [available from mg/l (ppm) mode only].
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
2120
Salinity adjustment
NOTE: this mode appears in mg/l (ppm)
measurement mode only.
1. Press the Mode key to select mg/l (ppm)
mode.
2. Press the SETUP key to enter Set Up
mode.
See figure
4. Press the ENTER key until the upper
display shows a number and the lower display shows “SAL”.
5. Use the and keys to enter the
salinity of your solution in ppt. The upper display will show the value you have entered.
See figure
6. Press the ENTER key to confirm value.
7. Press CAL/MEAS to return to
Measurement mode.
A
B
5.3
SETUP
Taking pH Measurements
Be sure to remove the protective rubber cap on the electrode before measurement. To ensure Automatic Temperature Compensation (ATC) for pH measurements, the DO probe must be plugged in and must be immersed into the sample you wish to take pH readings.
A
To take readings:
1. Rinse the pH electrode with deionized or
SETUP
ppt
B
distilled water before use to remove any impurities adhering to the probe body. If the pH electrode has dehydrated, soak it for 30 minutes in OAKTON electrode storage solution or a 2M–4M KCl solution.
2. Press ON to switch on meter. Press
MODE key to select pH measurement mode. The MEAS annunciator appears on the top center of the LCD. The ATC indicator appears in the lower right hand corner to indicate Automatic Temperature Compensation. If the ATC indicator does not appear, make sure DO probe is plugged into the meter.
See figure
A
MEAS
pH
C
ATC
A
3. Dip both the pH electrode and the DO
probe into the sample.
When dipping the pH electrode into the sample, the sensor or the glass bulb of the electrode must be completely immersed into the sample. Stir the probe gently in the sam­ple to create a homogenous sample.
4. Allow time for the reading to stabilize.
Note the reading on the display.
5. Press the MODE key to toggle between
modes:
• pH
• % saturation.
mg/l (ppm)**
**To select between mg/l and ppm units, see page 52. The ATC indicator should appear in the lower right hand corner of the display.
If it does not, this indicates an error.
CONTINUED ON NEXT PAGE
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
2322
Taking measurements with READY indicator selected on
If the READY indicator has been activated, the READY annunciator lights when the reading is stable. Switch the READY indicator on or off in Set up program P3.1— see page 35 for directions.
Taking measurements with the auto endpoint feature selected on
When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically “HOLD” the reading. The “HOLD” indicator appears on the left side of the display. Press the HOLD key to release the reading. Switch the Auto endpoint feature on or off in Set up program P3.1—see page 35 for directions.
5.4
Manual Temperature Compensation
IMPORTANT: For manual compensation, you must disconnect the DO or tempera­ture probe (see page 9).
1. Switch the meter on. Press the MODE
key to select pH mode.
2. Press the CAL/MEAS key to enter pH
calibration mode. The CAL indicator will appear above the primary display.
3. While in pH calibration mode, press
the MODE key to enter temperature
calibration mode. The primary display shows the current temperature setting and the secondary display shows the default value 25°C.
See figure
A
4. Check the temperature of your sample
using an accurate thermometer.
5. Press the or keys to offset the
temperature to the measured value from step 4.
See figure
B
6. Press ENTER to confirm the selected
temperature and to return to the pH measurement mode.
The meter will now compensate pH readings for the manually set temperature.
CAL
C
A
CAL
C
B
Notes
To exit this program without confirming the manual temperature compensation value, DO NOT press ENTER in step 6. Press CAL/MEAS instead.
24 25
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
ENT
MEAS
SETUP
6. HOLD function
7. Backlit LCD
This feature lets you freeze the value of the pH, % saturation or mg/l (ppm) and temperature readings for a delayed observation. HOLD can be used any time when in MEAS mode.
1. To hold a measurement, press the
HOLD key while in measurement mode. The “HOLD” annunciator will appear on the display.
See figures and
A
B
2. To release the held value, press HOLD
again. Continue to take measurements.
NOTE: This meter shuts off automatically
after 20 minutes of nonuse. If the meter is shut off either automatically or manually, the HOLD value will be lost.
NOTE: This meter has an auto endpoint
feature. When this feature is switched on, the display will automatically “HOLD” a reading that has been stable for more than 5 seconds. The “HOLD” indicator appears. Press the HOLD key to release the reading. To switch on or off the auto endpoint feature, see the SETUP program P3.1 on page 35 for pH, P5.1 on page 43 for % saturation and P4.1 on page 51 for mg/l (ppm).
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
Using the backlit LCD
MEAS
HOLD
A
MEAS
HOLD
B
pH
C
ATC
%
C
ATC
This feature lets you view readings in low light situations.
1. To activate the backlit LCD feature,
press the key to activate.
See figure
To keep the LCD backlit, keep the key pressed. If the key is not pressed, the backlight automatically shuts off automatically after 20 seconds after pressing the key.
A
Adjusting LCD brightness
This mode lets you adjust the brightness of the backlit LCD. Selecting a dimmer backlighting level helps conserve batteries. This adjustment is only available in the % saturation or mg/l (ppm) SETUP modes.
From measurement mode:
1. Press the MODE key to switch to
% saturation or mg/l (ppm) mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter “LCd” in the upper display.
See figure
4. Press the ENTER key. A number (0-8)
appears in the upper display and “LCd” appears in the lower display.
See figure
5. Press the and keys to select
from level 0 (dimmest light) to level 8 (brightest light).
6. Press the ENTER key to confirm
selection and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
A
B
A
SETUP
A
SETUP
B
2726
8. Advanced SETUP functions
pH
SETUP
pH
SETUP
pH
SETUP
pH
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
pH
SETUP
pH
SETUP
pH
SETUP
pH
SETUP
8.1
Advanced SETUP mode detailed overview
The advanced set up mode lets you customize your meter’s preferences and defaults. Your OAKTON waterproof meter features different sub groups that organize all SETUP parameters.
This meter displays different subgroups depending on which mode the meter is in:
pH % Saturation mg/l (ppm)
CAL: Calibration data
ELE: Probe data
COF: Configuration
rSt: Reset meter to factory default
dPr: Dissolved oxygen parameters
% saturation offset adjustment
CAL: Calibration data
ELE: Probe data
dPr: Dissolved oxygen parameters
CAL: Calibration data
ELE: Probe data
COF: Configuration
Press the SETUP key to enter SETUP mode. Press the ▲ and ▼ keys to scroll through sub groups.
From the pH measurement mode
CAL: Viewing previous calibration data
• View previous calibration data
Instructions on page 32
ELE: Viewing electrode data
• View pH electrode offset
• View pH electrode slope
Instructions on page 33-34
COF: Unit configuration
• Ready indicator on or off / auto endpoint on or off
• number of calibration points: 2, 3, 4, or 5
Instructions on page 35-37
Instructions on page 38
• Select °F or °C
rSt: Reset to factory default
• Reset to factory default settings
COF: Configuration
LCd: Backlit display
rSt: Reset meter to factory default
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
LCd: Backlit display
rSt: Reset meter to factory default
2928
From the % saturation measurement mode
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
SETUP
From the mg/l (ppm) measurement mode
Instructions on page 39
Instructions on page 40
Instructions on page 41
Instructions on page 42
Instructions on page 43-44
dPr: Dissolved oxygen parameters
• Select Hg or Pa barometric pressure units
• Select barometric pressure
OFS: % saturation offset adjustment
• Set % saturation offset adjustment
CAL: Viewing previous calibration data
• View previous calibration data
ELE: Viewing probe data
• View probe slope
• View % saturation offset
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
COF: Unit configuration
• Ready indicator on or off / auto endpoint on or off
• Select °F or °C
Instructions on page 47-48
Instructions on page 49
Instructions on page 50
Instructions on page 51-53
Instructions on page 54
dPr: Dissolved oxygen parameters
• Select Hg or Pa barometric pressure units
• Select barometric pressure
• Select salinity adjustment factor
CAL: Viewing previous calibration data
• View previous calibration data, including date/time
ELE: Viewing probe data
• View probe slope
• View mV value equivalent to 100% saturation
• View mV value equivalent to 0% saturation
COF: Unit configuration
• Ready indicator on or off / auto endpoint on or off
• Select mg/l or ppm units
• Select °F or °C
LCd: Backlit display
• Adjust brightness of backlit LCD
Instructions on page 45
Instructions on page 46
LCd: Backlit display
• Adjust brightness of backlit LCD
rSt: Reset to factory default
• Reset to factory default settings
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
rSt: Reset to factory default
• Reset to factory default settings
Instructions on page 55
3130
8.2
pH mode—P1.0: Viewing previous calibration data
8.3
pH mode—P2.0: Viewing electrode data
Program P1.0 lets you recall previous calibration data, which helps you know when to recalibrate your meter. This is a “view only” mode.
From measurement mode:
1. Press the MODE key to select pH
measurement mode.
2. Press the SETUP key to enter SETUP
mode.
See figure
A
SETUP
pH
3. Press the and keys to scroll
through subgroups until you view parameter P1.0.
A
4. Press the ENTER key repeatedly to view
previous calibration data.
P1.1 = pH 1.68 P1.2 = pH 4.01 P1.3 = pH 7.00 P1.4 = pH 10.01 P1.5 = pH 12.45
See figure
B
SETUP
pH
B
5. When you have scrolled through all
calibration data, you will automatically return to the subgroup menu. Press the CAL/MEAS key to return to measure­ment mode.
NOTES:
If there is no previous calibration data at a particular point, the primary display will show “----”.
See figure
C
SETUP
pH
C
Program P2.0 has two “view only”options that lets you check the electrode parame­ters for diagnostic purposes.
P2.1: Electrode offset
From measurement mode:
1. Press the MODE key to select pH
measurement mode.
2. Press the SETUP key to enter Set Up
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P2.0.
See figure
A
4. Press the ENTER key to select
parameter P2.1.
5. The display shows the electrode offset
value.
• the display shows the mV offset at pH 7.00. If you have not calibrated at any buffer, the primary display shows
0.00 mV.
See figure
B
6. Press the ENTER key to to proceed to
step 4 of P2.2. Or, press the CAL/MEAS key to return to measurement mode.
SETUP
pH
A
SETUP
mV
pH
B
32 33
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
P2.2: Electrode slope
NOTE: this parameter is available in pH measurement mode only.
From measurement mode:
1. If necessary, press the MODE key to
select pH measurement mode. Press the SETUP key to enter SETUP mode.
2. Press the and keys to scroll
through subgroups until you view parameter P2.0.
See figure
3. Press the ENTER key twice to select
4. The display shows electrode slope
in percentage of the range you are currently measuring. Default setting is 100.0.
See figure
5. Press the ENTER key to return to the
subgroup menu. Press the CAL/MEAS key to return to measurement mode.
D
parameter P2.2.
E
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
8.4
SETUP
pH
pH mode—P3.0: pH measurement configuration
P3.1: READY indicator and auto endpoint function
This program lets you select:
D
“READY
“READY
Auto endpoint function on.
SETUP
%
it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions.
From measurement mode:
pH
E
1. Press the MODE key to select pH
measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll
through subgroups until you view parameter P3.0.
See figure
4. Press the ENTER key to select
parameter P3.1.
See figure
5. Press the and keys to select the
configuration you require.
•OFF switches the READY indicator off.
•ON switches the READY indicator on.
•ON and HOLD together switches the auto endpoint feature on.
6. Press the ENTER key to confirm selec-
tion and to proceed to step 4 of P3.2. Press the CAL/MEAS key to return to measurement mode.
NOTES: Meter default is set for Ready indi­cator on, and auto endpoint function off.
You can also change the Ready indicator and auto endpoint function in SETUP Program P5.1 on page 43 in % saturation mode or Program P4.1 on page 51 in mg/l (ppm) mode. Any changes you make to the Ready indicator/ auto endpoint function in pH mode will also change in % saturation and mg/l (ppm) modes.
indicator on” to indicate when the reading is stable. indicator off” for faster meter response.
Select auto endpoint on to “hold” the reading when
SETUP
A
A
SETUP
READY
B
ON
SETUP
READY
OFF
SETUP
READY HOLD ON
B
pH
3534
P3.2: Selecting number of pH calibration points
P3.3 Selecting °C or °F
Program P3.2 lets you select the number of pH calibration points the meter will use in calibration mode: 2, 3, 4, or 5. The meter will automatically exit calibration mode after you have calibrated to your selected number of points.
From measurement mode:
1. Press the MODE key to select pH
measurement mode.
SETUP
pH
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P3.0.
See figure
C
C
4. Press the ENTER key twice to select
parameter P3.2.
See figure
D
SETUP
5. Press the and keys to select 2, 3, 4,
or 5 point pH calibration.
6. Press the ENTER key to confirm selection
and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
D
This meter lets you select between °C and °F units for temperature readings.
From measurement mode:
1. Press the MODE key to select pH
SETUP
measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P3.0.
See figure
E
E
4. Press the ENTER key three times to
select parameter P3.3.
See figure
F
SETUP
5. Press the and keys to toggle
between °C and °F.
6. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
SETUP
F
pH
C
F
NOTES:
You can also switch between °C and °F in SETUP Program P5.2 on page 44 in % saturation mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch between °C and °F in pH mode, the meter will also switch in % saturation and mg/l (ppm) modes.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
3736
8.5
pH mode—P4.0: Resetting to factory default settings
8.6
% saturation mode—P1.0: Dissolved Oxygen Parameters
This program lets you reset all pH parameters to factory default settings. This clears all calibration data any other pH setup functions you might have changed. The following settings will remain as you have set them:
• Temperature unit of measure (°C or °F)
• The temperature offset calibration value
• All dissolved oxygen calibration data and parameters including the LCD contrast setting
From measurement mode:
1. Press the MODE key to select pH
SETUP
pH
measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P4.0 in the lower display.
See figure
A
4. Press the ENTER key.
See figure
B
A
MEAS
SETUP
pH
5. Press the and keys to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
6. Press the ENTER key to confirm
selection and to return to measurement mode.
B
Notes:
To reset all % saturation data, see page 46. To reset all mg/l (ppm) data, see page 55. See page 68 for a table of factory default settings.
SETUP
MEAS
pH
This sub group lets you adjust the barometric pressure
Pressure adjustment mode
Barometric pressure is vital to correct dissolved oxygen measurements. You need to enter the correct barometric pressure of the area you are measuring. This mode lets you perform two functions:
• Select either mm Hg or Pascal barometric pressure units
• Adjust the barometric pressure. See Appendix 2 on page 67 for a “Pressure vs Altitude” table.
From measurement mode:
1. Press the MODE key to select %
saturation mode.
2. Press the SETUP key to enter SETUP
mode.
3. Press the and keys to scroll through
subgroups until you view parameter P1.0 in the lower display.
See figure
A
4. Press the ENTER key twice. The upper
display shows either Hg or PA and the lower display shows bAr.
5. Press the and keys to toggle between
mm Hg and Pascal units.
See figure
B
6. Press the ENTER key to confirm selection
and move to the next screen. The upper display shows the barometric pressure and the lower display shows the units selected in step 5.
See figure
C
7. Press the and keys to adjust
the barometric pressure. The pressure
adjustment range is 500 to 1499 mm Hg
(66.6 to 199.9 kPa).
8. Press the ENTER key to confirm
selection and move to step 3 on page 40. If instead you want to return to measure­ment mode, press CAL/MEAS twice.
SETUP
A
SETUP
SETUP
B
SETUP
C
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
3938
8.7
% saturation mode—P2.0: Offset Adjustment
8.8
% saturation mode—P3.0: Previous calibration information
Use the offset adjustment to act as an offset at values between 0 and 100%. It is useful to match against end-user standards (i.e. a test kit value).
From measurement mode:
1. Press the MODE key to select %
saturation mode.
SETUP
2. Press the SETUP key to enter SETUP
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P2.0 in the lower display.
See figure
A
4. Press the ENTER key. The upper
display shows the current measurement in % saturation and the lower shows OFS.
See figure
B
5. Press the and keys to offset the
% saturation measurement.
A
MEAS
SETUP
READY
%
B
6. Press the ENTER key to confirm
selection and to move back to subgroup P2.0. If you want to return to measure­ment mode, press CAL/MEAS.
This sub group shows you the previous calibration data, along with date and time of calibration. This is a “view only” parameter.
In % saturation mode: calibration information is shown in % saturation units.
From measurement mode:
1. Press the MODE key to select %
SETUP
saturation mode.
2. Press the SETUP key to enter SETUP
mode.
3. Press the and keys to scroll
through subgroups until you view
A
parameter P3.0 in the lower display.
See figure
A
SETUP
CAL
%
4. Press the ENTER key. The upper
display shows the calibration data.
See figure
6. Press the ENTER key to move back to
subgroup P3.0. If you want to return to measurement mode, press CAL/MEAS.
B
B
SETUP
CAL
%
NOTE: If you did not calibrate this meter in a particular mode, the screen will
C
show “– – –”.
See figure
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
C
4140
8.9
% saturation mode—P4.0: Electrode properties
8.10
% saturation mode—P5.0: Unit configuration
These “view only” parameters show the electrode properties for diagnostic purposes:
probe slope:
shows the calibration slope of the probe. This mode displays slope
from 0.5 to 1.999 % of slope (1.0 = 100%).
% saturation of
fset: shows the value of the % saturation offset entered in parameter
P2.0 (see page 40 for instructions).
100% saturation mV value:
sensor’s mV output corresponding to 100% saturation.
0% saturation mV value:
shows the
SETUP
shows the sensor’s millivolt output corresponding to 0% saturation.
A
From measurement mode:
1. Press the MODE key to select the
% saturation measurement mode.
2. Press the SETUP key to enter SETUP
SETUP
%
K=
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P4.0 in the lower display.
See figure
A
4. Press the ENTER key. The upper
display shows the probe slope.
See figure
B
5. Press the ENTER key. The upper
display shows the % saturation offset. See directions for setting this offset on page 40.
See figure
C
B
%
C
SETUP
mV
6. Press the ENTER key. The upper
display shows the 100 % saturation mV value.
See figure
D
7. Press the ENTER key. The upper display
shows the 0 % saturation mV value.
See figure
E
D
SETUP
mV
8. Press the ENTER key to move back to
subgroup P4.0. If you want to return to measurement mode, press CAL/MEAS.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
E
Unit configuration mode lets you select the following parameters:
A. Ready indicator and auto endpoint function B. Temperature in °C or °F
P5.1: READY indicator and auto endpoint function
The first program lets you select:
“READY
“READY
Auto endpoint function on.
it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions.
From measurement mode:
1. Press the MODE key to select the
% saturation measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P5.0 in the lower display.
See figure
4. Press the ENTER key to select
parameter P5.1 “rdY”(Ready).
See figure
5. Press the and keys to select the
configuration you require.
•ON switches the READY indicator on.
•OFF switches the READY indicator off.
• ON and HOLD together switches the
6. Press the ENTER key to confirm selection
and to move to step 3 on page 44 (P5.2). Or, press the CAL/MEAS key twice to return to measurement mode.
NOTES: Meter default is set for Ready indi­cator on, and auto endpoint function off.
You can also change the Ready indicator and auto endpoint function in SETUP Program P3.1 on page 35 in pH mode or P4.1 on page 51 in mg/l (ppm) mode. Any changes you make to the Ready indicator/ auto endpoint function in % saturation mode will also change in pH and mg/l (ppm) modes.
indicator on” to indicate when the reading is stable. indicator off” for faster meter response.
Select auto endpoint on to “hold” the reading when
A
B
auto endpoint feature on.
SETUP
A
SETUP
READY
ON
SETUP
READY
OFF
SETUP
READY HOLD
ON
B
4342
P5.2: Selecting °C or °F temperature readout
This meter lets you select between °C and °F units for temperature readings.
From measurement mode:
1. Press the MODE key to select the
% saturation measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P5.0 in the lower display.
See figure
E
4. Press the ENTER key until “C” or “F’
appears in the upper display.
See figure
F
5. Press the and keys to toggle
between °C and °F.
6. Press the ENTER key to confirm
selection and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
SETUP
E
SETUP
SETUP
F
8.11
% saturation mode—P6.0: Adjusting LCD brightness
This mode lets you adjust the brightness of the backlit LCD. Selecting a dimmer backlighting level helps conserve batteries.
From measurement mode:
1. Press the MODE key to select the
SETUP
% saturation measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P6.0 in the lower display.
See figure
A
4. Press the ENTER key. A number (0-8)
appears in the upper display and “LCd”
C
appears in the lower display.
See figure
B
A
SETUP
5. Press the and keys to select
from level 0 (dimmest light) to level 8 (brightest light).
F
6. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
B
NOTES:
You can also switch between °C and °F in SETUP Program P3.3 on page 37 in pH mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch between °C and °F in % saturation mode, the meter will also switch in pH and mg/l (ppm) modes.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
4544
8.12
% saturation mode—P7.0: Resetting to factory default settings
This program lets you reset all parameters to factory default settings. This clears all calibration data, memory, and any other setup functions you might have changed. The LCD contrast setting will be reset to 8 (the maximum setting). However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• the temperature offset calibration value
• pH calibration data and set parameters
From measurement mode:
1. Press the MODE key to select the
% saturation measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P7.0 in the lower display.
See figure
A
4. Press the ENTER key.
See figure
B
5. Press the and keys to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
6. Press the ENTER key to confirm
selection and to return to measurement mode.
Notes
To reset all pH data, see page 38. To reset all mg/l (ppm) data, see page 55. See page 68 for a table of factory default
settings.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
SETUP
A
SETUP
SETUP
B
8.13
mg/l (ppm) mode—P1.0: Dissolved Oxygen Parameters
This sub group lets you adjust the barometric pressure and salinity.
P1.1 Pressure adjustment mode
Barometric pressure is vital to correct dissolved oxygen measurements. You need to enter the correct barometric pressure of the area you are measuring. This mode lets you perform two functions:
• Select either mm Hg or Pascal barometric pressure units
• Adjust the barometric pressure. See Appendix 2 on page 67 for a “Pressure vs Altitude” table.
From measurement mode:
1. Press the MODE key to select mg/l (ppm)
mode.
2. Press the SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P1.0 in the lower display.
See figure
A
4. Press the ENTER key twice. The upper
display shows either Hg or PA and the lower display shows bAr.
5. Press the and keys to toggle between
mm Hg and Pascal units.
See figure
B
6. Press the ENTER key to confirm selection
and move to the next screen. The upper display shows the barometric pressure and the lower display shows the units selected in step 5.
See figure
C
7. Press the and keys to adjust
the barometric pressure. The pressure
adjustment range is 500 to 1499 mm Hg
(66.6 to 199.9 kPa).
8. Press the ENTER key to confirm
selection and move to P1.2 on page 48 (salinity adjustment). If instead you want to return to measurement mode, press CAL/MEAS twice.
NOTE: Salinity adjustment appears in mg/l (ppm) measurement mode only.
SETUP
A
SETUP
SETUP
B
SETUP
C
4746
P1.2 Salinity adjustment mode
Salinity correction mode lets you correct for the variations in oxygen solubility due to salt concentration in the sample.
From measurement mode:
1. Press the MODE key to select mg/l
(ppm) mode.
2. Press the SETUP key to enter SETUP
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P1.0 in the lower display.
See figure
4. Press the ENTER key five times. The
See figure
5. Press the and keys to enter the
correct salinity adjustment factor. The salinity adjustment factor range is
0.0 to 50.0 ppt.
6. Press the ENTER key to confirm
selection and to move back to subgroup P1.0. If you want to return to measure­ment mode, press CAL/MEAS.
D
upper display shows the salinity value and the lower shows SAL.
E
8.14
mg/l (ppm) mode—P2.0: Previous calibration information
This sub group shows you the previous calibration data, along with date and time of calibration. This is a “view only” parameter.
In mg/l (ppm) mode: calibration information is shown in mg/l (ppm) units.
SETUP
From measurement mode:
1. Press the MODE key to select mg/l
SETUP
(ppm) mode.
2. Press the SETUP key to enter SETUP
D
SETUP
ppt
E
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P2.0 in the lower display.
See figure
A
4. Press the ENTER key. The upper
display shows the calibration data.
See figure
B
6. Press the ENTER key to move back to
subgroup P2.0. If you want to return to measurement mode, press CAL/MEAS.
A
SETUP
CAL
mg/l
B
SETUP
CAL
mg/l
NOTE: If you did not calibrate this meter in a particular mode, the screen will
C
show “– – –”.
See figure
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
C
4948
8.15
mg/l (ppm) mode—P3.0: Electrode properties
These “view only” parameters show the electrode properties for diagnostic purposes:
probe slope:
shows the calibration slope of the probe. This mode displays slope
from 0.5 to 1.999 % of slope (1.0 = 100%).
100% saturation mV value:
shows the sensor’s mV output corresponding to
100% saturation.
0% saturation mV value:
shows the sensor’s millivolt output corresponding to
0% saturation.
SETUP
From measurement mode:
1. Press the MODE key to select the
mg/l (ppm) measurement mode.
A
2. Press the SETUP key to enter SETUP
mode.
3. Press the and keys to scroll
through subgroups until you view parameter P3.0 in the lower display.
See figure
A
SETUP
K=
mg/l
4. Press the ENTER key. The upper
display shows the probe slope.
See figure
B
B
6. Press the ENTER key. The upper
display shows the 100 % saturation mV value.
See figure
C
SETUP
mV
7. Press the ENTER key. The upper display
shows the 0 % saturation mV value.
See figure
D
C
8. Press the ENTER key to move back to
subgroup P3.0. If you want to return to measurement mode, press CAL/MEAS.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
SETUP
mV
D
8.16
mg/l (ppm) mode—P4.0: Unit configuration
Unit configuration mode lets you select the following parameters:
A. Ready indicator and auto endpoint function B. mg/l or ppm units C. Temperature in °C or °F
P4.1: READY indicator and auto endpoint function
The first program lets you select:
“READY
“READY
Auto endpoint function on.
it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions.
From measurement mode:
1. Press the MODE key to select the
mg/l (ppm) measurement mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P4.0 in the lower display.
See figure
4. Press the ENTER key to select
parameter P4.0 “rdy” (Ready).
See figure
5. Press the and keys to select the
configuration you require.
•ON switches the READY indicator on.
•OFF switches the READY indicator off.
• ON and HOLD together switches the
6. Press the ENTER key to confirm selection
and to proceed to P4.2 on page 52. Or, press the CAL/MEAS key twice to return to measurement mode.
Notes: Meter default is set for Ready indica­tor on, and auto endpoint function off.
You can also change the Ready indicator and auto endpoint function in SETUP Program P3.1 on page 35 in pH mode or P5.1 on page 43 in % saturation mode. Any changes you make to the Ready indicator/ auto endpoint function in % saturation mode will also change in pH and % saturation modes.
indicator on” to indicate when the reading is stable. indicator off” for faster meter response.
Select auto endpoint on to “hold” the reading when
A
B
auto endpoint feature on.
SETUP
A
SETUP
READY
ON
SETUP
READY
OFF
SETUP
READY HOLD ON
B
5150
P4.2: Selecting mg/l or ppm units
P4.3: Selecting °C or °F temperature readout
This mode lets you select between mg/l or ppm dissolved oxygen units.
From measurement mode:
1. Press the MODE key to select mg/l
SETUP
(ppm) mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P4.0 in the lower display.
See figure
C
4. Press the ENTER key until “dO”
appears in the upper display, P4.2 in the lower display.
See figure
D
C
SETUP
5. Press the and keys to toggle
between mg/l or ppm units.
6. Press the ENTER key to confirm
selection and move to P4.3 on page
53. Press the CAL/MEAS key twice to return to measurement mode.
SETUP
D
mg/l
ppm
This meter lets you select between °C and °F units for temperature readings.
From measurement mode:
1. Press the MODE key to select mg/l
SETUP
(ppm) mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P4.0 in the lower display.
See figure
E
4. Press the ENTER key until “C” or “F’
appears in the upper display.
See figure
F
E
SETUP
5. Press the and keys to toggle
between °C and °F.
6. Press the ENTER key to confirm
selection and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
SETUP
F
C
F
NOTES:
You can also switch between °C and °F in SETUP Program P3.3 on page 37 in pH mode or Program P5.2 on page 44 in mg/l (ppm) mode. If you switch between °C and °F in mg/l (ppm) mode, the meter will also switch in pH and % saturation modes.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
5352
8.17
mg/l (ppm) mode——P5.0: Adjusting LCD brightness
This mode lets you adjust the brightness of the backlit LCD. Selecting a dimmer backlighting level helps conserve batteries.
From measurement mode:
1. Press the MODE key to select mg/l
(ppm) mode.
SETUP
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P5.0 in the lower display.
See figure
A
4. Press the ENTER key. A number (0-8)
appears in the upper display and “LCd” appears in the lower display.
See figure
B
A
SETUP
6. Press the and keys to select
from level 0 (dimmest light) to level 8 (brightest light).
B
6. Press the ENTER key to confirm selec-
tion and to return to the subgroup menu. Press the CAL/MEAS key to return to measurement mode.
8.18
mg/l (ppm) mode—P6.0: Resetting to factory default settings
This program lets you reset all parameters to factory default settings. This clears all calibration data, memory, and any other setup functions you might have changed. The LCD contrast setting will be reset to 8 (the maximum setting). However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• the temperature offset calibration value
• pH calibration data and set parameters
From measurement mode:
1. Press the MODE key to select mg/l
(ppm) mode.
2. Press SETUP key to enter SETUP mode.
3. Press the and keys to scroll through
subgroups until you view parameter P6.0 in the lower display.
See figure
A
4. Press the ENTER key.
See figure
B
5. Press the and keys to toggle
between NO and YES.
• NO retains current settings
• YES resets to factory default settings
6. Press the ENTER key to confirm
selection and to return to measurement mode.
SETUP
A
SETUP
SETUP
B
NOTES:
To reset all pH data, see page 38. To reset all % saturation data, see page 46. See page 68 for a table of factory default settings.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
5554
9. Probe Care and Maintenance
9.1
pH Electrode Care
Since your pH electrode is susceptible to dirt and contamination, clean it every one to three months depending on the extent and condition of use.
NOTE: for specialty electrode care, consult the instruction manual included with
your electrode.
pH electrode storage
For best results, always keep the pH bulb wet. Use the rubber cap filled with OAKTON electrode storage solution to store your electrode (see page 64-65 for ordering information). Also, you can store in a pH 4 buffer with 1/100 part of saturated KCl. Other pH buffers are OK for storage, but NEVER use distilled water for storage.
After measuring
1. Rinse the pH electrode and reference junction in de-ionized water.
2. Store the electrode as recommended above in “pH electrode storage,” or as
recommended by the manufacturer.
3. Prior to next use, rinse the liquid junction with de-ionized water and tap dry—
never wipe electrode.
NOTE: If this does not restore electrode to normal response, see “Reactivating the
pH electrode” section on page 57.
pH electrode cleaning
Salt deposits: dissolve the deposits by immersing the electrode in tap water for ten
to fifteen minutes. Ten thoroughly rinse with distilled water. Oil/grease film: wash electrode pH bulb gently in some detergent and water. Rinse
electrode tip with distilled water or use a general purpose electrode cleaner (see page 65 for ordering information).
Clogged reference junction: heat a diluted KCl solution to 60 to 80°C. Place the sensing part of the electrode into the heated solution for about 10 minutes. Allow the electrode to cool in some unheated KCl solution.
Protein deposits: prepare a 1% pepsin solution in 0.1 M of HCl. Set the electrode in the solution for five to ten minutes. Rinse the electrode with distilled water.
Reactivating the pH electrode
If stored and cleaned properly, your pH electrode should be ready for immediate use. However, a dehydrated bulb may cause sluggish response. To rehydrate the bulb, immerse the electrode in a pH 4 buffer solution for 10 to 30 minutes. If this fails, you probably need to replace the pH electrode. Never touch or rub glass bulb.
Contact builds up an electrostatic charge.
9.2
DO Probe Care
The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measure­ments immediately—without the typical 15 minute wait of other dissolved oxygen probes.
The probe consists of two parts:
• an upper part consisting of an anode, a cathode, and cable
• a lower part consisting of a membrane cap, membrane, and electrolyte solution.
Under typical operating conditions, the probe should last for several years. Proper care and
maintenance will help you receive the maximum probe life and ensure more accurate readings.
Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the mem­brane, the membrane must be cleaned at regular intervals to assure maximum reliability.
After using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical to clean.
NOTE: Although the membrane is strong and
not easily damaged, wipe it gently while cleaning it. If the membrane is damaged or torn, the probe will no longer function.
There are no special probe storage requirements.
Anode
Cathode
Membrane cap
(holds membrane
and electrolyte
solution)
Probe guard
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
5756
9.3
DO Membrane Replacement
Replacement of the membrane is required only when you cannot calibrate the probe, or if the membrane is damaged.
Typical membrane damages are punctures or wrinkles caused during measurements or cleaning.
To order replacement probe components or a replacement probe, see the “Accessories” section on page 64.
To replace the probe membrane:
1. Pull off the probe guard.
2. Unscrew the membrane cap from the probe.
3. Hold the probe under a water tap and brush
away the white oxide on the cylindrical anode with a stiff plastic brush—do not use metal cleaning material.
4. It the cathode has any deposits, remove them
with a light scouring powder. Do not polish the cathode.
NOTE: If you have purchased a replacement
membrane module with preinstalled membrane, skip to step 12.
CONTINUED ON NEXT PAGE
Membrane
Membrane
A
Temperature
sensor
Anode
Cathode
lock
O-ring
Membrane
cap
Probe guard
5. Using the installation tool, unscrew and
remove the membrane lock from the membrane cap.
See figure
B
6. Remove the membrane and O-ring.
Discard both.
7. Rinse the membrane cap and membrane lock
in tap water.
8. Install a new O-ring inside the
membrane cap.
9. Install a new membrane. Make sure the
membrane covers the O-ring all around its circumference.
See figure
C
10. Using the installation tool, screw the
membrane lock back into the cap. Tighten the lock firmly over the membrane and O-ring, but do not overtighten.
11. Inspect the membrane for wrinkles.
If wrinkles exist, remove the membrane and repeat steps 8-11.
12. Fill the membrane cap with water and
inspect the bottom for leaks. If water drops are leaking from the membrane, re-seat the membrane on the O-ring (repeat steps 8-11).
13. If the assembly is leak-free, empty the water
and fill the membrane cap with electrolyte to the brim.
14. Screw the cap onto the probe.
Excess electrolyte will drain out.
15. Replace probe guard.
16. Calibrate the probe (see section 4.4-4.6 on
pages 14-18) after the % saturation readings have stabilized.
Installation tool
Insert notches on installation tool into slots on membrane lock. Then, unscrew membrane lock from membrane cap.
B
Carefully place O-ring, then membrane, into membrane cap.
C
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
5958
9.4
DO Electrolyte solution
The electrolyte solution in your probe’s cap will periodically evaporate and need to be replaced. 35640-71 replacement electrolyte solution comes premixed and ready to use. However, the electrolyte mixture package 35640-70 has a longer shelf life in its unmixed form.
If you purchase the electrolyte mixture package to make your own replacement electrolyte solution, use the following steps to prepare the solution:
1. Fill a beaker to the 400 ml mark with
deionized water.
2. Pour the entire contents of the
58.5 grams electrolyte package into the beaker.
3. Stir the solution until all of the chemical
is dissolved—until the solution is clear.
4. Pour the solution into a clean
container with a cap and keep sealed between use.
10. Troubleshooting
Problem Cause Solution Power on but a) Batteries not in place. a) Check that batteries are in
no display place and making good
contact.
b) Batteries not in correct b) Reinsert batteries with
polarity (+ and –). correct polarity.
c) Weak batteries. c) Replace batteries.
Unstable a) Insufficient electrolyte a) Fill DO probe with reference readings in DO probe. electrolyte. See page 58-59.
b) Air bubbles in probe. b) Tap probe to remove bubbles. c) Dirty DO probe or c) Clean the probe or electrode
pH electrode. and recalibrate.
d) Probe not deep d) Make sure sample entirely
enough in sample. covers the probe sensors.
e) External noise pickup e) Move or switch off
or induction caused by interfering motor. nearby electric motor.
f) Broken DO probe or f) Replace probe or electrode.
pH electrode See page 64.
Slow response a) Dirty membrane/ a) Clean probe. See "Probe
Dirty electrode Care & Maintenance",
page 56-59.
Not responding a) HOLD mode in operation. a) Cancel HOLD mode. to key press
b) Damaged key pad. b) Return to dealer. c) Internal program error. c) Reset all internal programs
by reinserting batteries.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
6160
11. Error Messages
12. Specifications
LCD Display
Err Unrecognized input Wrong input in Release key. Select annunciator from keypad selected mode. valid operations
CAL & Err Calibration error Wrong value input Check your input value, annunciators at calibration. clean probe. blink Dirty probe. See Calibration sections or
Battery Low battery level Need new batteries Clean battery contacts. indicator blinks or battery Replace batteries with fresh
* See “Warranty” and “Return of Items” on page 67
Indicates Cause Solution
depending on mode.
Probe Maintenance section.
connection is bad. ones, noting polarity.
If an error message appears in the primary display (the upper row of larger digits), switching off the meter and switching it on again may eliminate the error message.
If error persists, or the meter shows incorrect values, return the meter.
For a complete diagram of the display, see page 5.
Mode pH mg/l (ppm) % saturation Temperature Range -2 to 16 pH 0.00 to 0.0 to 200.0% 0.0 to 50.0°C /
20.00 mg/l (ppm) 32.0 to 122.0°F
Resolution 0.01 pH 0.01 mg/l (ppm) 0.1% 0.1°C / 0.1°F Accuracy ±0.01 pH ±1.5% full scale ±1.5% full scale ±0.3°C / ±0.5°F Calibration up to 5 points: one or two points (100% and 0% saturation); offset up to
pH 1.68; 4.01; plus separate mg/l (ppm) slope ±5°C or ±10°F
7.00; 10.01; 12.45
Temperature compensation: automatic from 0 to 50°C Salinity correction:
Range: 0.0 to 50.0 ppt Resolution: 0.1 ppt Method: Key in manually and automatic correction
Barometric pressure correction:
Range: 500 to 1499 mm Hg / 66.6 to 199.9 kPa Resolution: 1 mm Hg / 0.1 kPa Method: Key in manually and automatic correction
DO Probe type: galvanic probe Display: dual LCD Backlit display: select from eight levels of brightness Operating temperature: 0 to 50°C Power: four 1.5 V AAA batteries (included) Battery life: > 100 hours continuous use
use of backlit display decreases battery life
Dimensions:
Meter: 7.5"L x 3.75"W x 2.25"H (19 x 9.5 x 5.7 cm) Boxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm) DO Probe: 6.8"L x 1.3" dia (173 mm L x 32 mm dia) with 10-ft cable pH Electrode: 6.0"L x 1.0" dia (152 mm L x 25 mm dia) with 10-ft cable
Weight:
Meter: 1.0 lb (0.5 kg) Boxed: 2.0 lbs (0.9 kg)
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
6362
13. Accessories
Meters
35632-02 pH/DO 300 meter Meter only 35632-00 pH/DO 300 meter Includes DO probe with 10-ft cable (35640-50), pH elec-
trode with 10-ft cable (35805-23), and electrolyte solution
35632-60 pH/DO meter kit. Includes pH/DO meter plus DO probe with
10-ft cable, pH electrode with 10-ft cable, 20 pH ”Singles“ pouches, 500 ml of electrolyte solution, 5 replacement membranes, membrane replacement tool, rinse bottle and foam-lined hard carrying case.
OAKTON “Singles” calibration solution pouches
pH solutions have ±0.01 pH accuracy at 25°C. Shpg wt 1.1 lb (510 g) per box.
35653-00 Deionized rinse water solution pouches, 20/box. 35653-01 pH 4.01 calibration buffer solution pouches, 20/box. 35653-02 pH 7.00 Calibration buffer solution pouches, 20/box. 35653-03 pH 10.00 Calibration buffer solution pouches, 20/box. 35653-04 Assortment pack, 5 each deionized water,
pH 4.01, pH 7.00, and pH 10.00 solution pouches.
Other accessories
Probes and probe replacement parts
35640-50 Replacement DO probe with 10-ft cable 35640-52 DO probe with 25-ft cable 35805-23 Replacement submersible pH electrode with 10-ft cable,
single junction, ABS plastic housing
35640-71 Replacement DO electrolyte solution, 500 ml 35640-70 Electrolyte DO powder, 58.5 g. Mix your own electrolyte solution—
makes 400 ml. Has longer shelf life in unmixed form.
WD-35640-72 Replacement DO membrane module. This item consists of a preassembled membrane, membrane lock, O-ring and cap. Simply screw replacement membrane module onto the probe you already have.
35640-74 Replacement DO membranes, 5/pack 35640-75 Replacement DO membranes, 25/pack 35640-79 DO membrane installation tool. Required for membrane installation.
Not required if you purchase replacement membrane module 35640-72.
OAKTON calibration solutions
pH solutions have ±0.01 pH accuracy at 25°C. Shpg wt 1.1 lb (510 g).
00654-01 pH 1.68 calibration buffer, 1 pint. 00654-00 pH 4.01 calibration buffer, 1 pint. 00654-04 pH 7.01 calibration buffer, 1 pint. 00654-08 pH 10.01 calibration buffer, 1 pint. 00654-12 pH 12.45 calibration buffer, 1 pint.
00653-04 pH electrode storage solution, 1 pint 00653-06 pH electrode cleaning solution, 1 pint 00653-00 Zero oxygen solution 04804-00 Battery-powered magnetic stirrer
To order OAKTON accessories, contact your OAKTON distributor.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
6564
14. Appendix 1: Meter Theory
15. Appendix 2: Pressure vs. Altitude table
Dissolved oxygen levels in natural and waste waters depend on the physical, chemical, and biochemical activities in the water body.
This meter uses a galvanic probe. It consists of a cell that contains electrolyte and that is enclosed by a selective membrane, and two metallic electrodes. The mem­brane is practically impermeable to water and ionic dissolved matter, but is perme­able to oxygen and a few other gases.
The cathode consumes the oxygen passed through the membrane, and produces an electric current in the probe. This current is proportional to the partial pressure of oxygen in the sample.
Since the cathode consumes the oxygen in the sample, it is essential that the fluid must flow past the sensor to maintain accurate readings.
The solubility of oxygen in water varies with barometric pressure and temperature, and decreases as salinity increases. For the most accurate DO readings, you need to compensate for these factors. This meter automatically compensates for tempera­ture readings. It also allows you to enter a salinity correction factor and the baromet­ric pressure to correct for this variability.
Barometric pressure affects DO readings, therefore this meter lets you enter the cor­rect barometric pressure at your altitude. If you do not have equipment that lets you measure the exact barometric pressure at your altitude, you can estimate it using the chart below.
If you change the barometric pressure setting from its factory setting (760 mm Hg), the % saturation calibration value in air will automatically adjust to a value other than 100% (see “corrected % saturation value” column below). The adjusted value is correct for the new barometric pressure setting.
See page 39 or 47 for information on how to adjust the barometric pressure.
Pressure vs Altitude
Altitude (ft) Pressure (mm Hg) Corrected % saturation value
0 (Sea level) 760 100
500 746 98.1 1000 732 96.3 1500 720 94.7 2000 707 93.0 2500 694 91.3 3000 681 89.6 3500 668 87.8 4000 656 86.2 4500 644 84.6 5000 632 83.0 5500 621 81.6 6000 609 80.0
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
6766
16. Appendix 3: Factory Default Settings
17. Warranty
Resetting the meter to factory default settings clears all calibration data and returns other setup functions to the default settings shown in the table below. However, the following settings will remain as you have set them:
• temperature unit of measure (°C or °F)
• The temperature offset calibration value
• the LCD contrast setting See pages 38, 46 and 55 for directions on setting all modes of this meter to factory
default settings.
Program Function Options/settings Default
Clr Memory clear yes/no no dPr Pressure units Hg or Pa Hg
Barometric pressure adjust from 500-1499 mm Hg 760 mm Hg
Salinity adjustment adjust from 0-50 ppt 0 ppt OFS Set % saturation offset up to ±10% no offset CAL View calibration data – – – ELE View pH electrode offset 0.00 mV
View pH electrode slope 100.0%
View DO probe slope 1.000
View % saturation offset 0 % offset
View mV value of DO probe
@ 100% saturation 37.0 mV
View mV value of DO probe
@ 0% saturation 0.3 mV
COF Ready indicator/auto endpoint on/off Ready on only
Select mg/l or ppm units mg/l or ppm mg/l
Select temperature units °C or °F retains settings LCd Adjust backlit display levels 0 to 8 (brightest) 8 rST Reset to factory default yes/no no
OAKTON warrants this meter to be free from significant deviations in material and workmanship for a period of three years from date of purchase. OAKTON warrants this probe to be free from significant deviations in material and workmanship for a period of six months from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warrantied time period, please return—freight prepaid—and correction will be made without charge. OAKTON alone will determine if the product problem is due to deviations or customer misuse.
Out-of-warranty products will be repaired on a charge basis.
18. Return of items
Authorization must be obtained from our Customer Service Department before returning items for any reason. When applying for authorization, please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possi­ble damage or loss. We will not be responsible for damage resulting from careless or insufficient packing. Arestocking charge will be made on all unauthorized returns.
NOTE: We reserve the right to make improvements in design, construction, and
appearance of products without notice.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
6968
Loading...