Nova-Tech pH-CON 300 User Manual

f
pH/Conductivity/°C/°F Meter
pH/CON 300 Series
WATERPROOF
MODE
HOLD
ENTER
RANGE
ON
OFF
CAL
MEAS
SETUP
ATC
READY
MEAS
C
µS
I
P
67
16. Warranty
OPERATING INSTRUCTIONS
OAKTON warrants this meter to be free from significant deviations in material and workmanship for a period of three years from date of purchase. OAKTON warrants this probe to be free from significant deviations in material and workmanship for a period of six months from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warrantied time period, please return—freight prepaid—and correction will be made without charge. OAKTON alone will determine if the product problem is due to deviations or customer misuse.
Out-of-warranty products will be repaired on a charge basis.
17. Return of items
Authorization must be obtained from our Customer Service Department before returning items for any reason. When applying for authorization, please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possi­ble damage or loss. We will not be responsible for damage resulting from careless or insufficient packing. Arestocking charge will be made on all unauthorized returns.
NOTE: We reserve the right to make improvements in design, construction, and
appearance of products without notice.
OAKTON®35631-00
Portable Waterproo
pH/CON 300 Mete
00702-89
R1 Printed in the U.S.A. 10/00
Table of Contents
1. Introduction ............................................................................................4
2. Display and keypad functions .............................................................5-6
2.1 Display...................................................................................................................................5
2.2 Keypad...................................................................................................................................6
3. Preparation..........................................................................................7-9
3.1 Inserting the batteries..........................................................................................................7
3.2 Probe information.............................................................................................................8-9
4. Calibration.......................................................................................10-16
4.1 Important information on meter calibration .................................................................10
4.2 Preparing the meter for calibration .................................................................................10
4.3 pH calibration................................................................................................................11-12
4.4 Conductivity and TDS calibration .............................................................................13-15
4.5 Temperature calibration....................................................................................................16
5. Measurement...................................................................................17-22
5.1 Taking pH measurements .................................................................................................17
5.2 Taking conductivity or TDS measurements...................................................................18
5.3 HOLD function / Ready indicator / Auto endpoint...................................................19
5.4 Using manual ranging function: conductivity/TDS....................................................20
5.5 Selecting manual temperature compensation: conductivity/TDS.............................21
5.6 Selecting the manual temperature compensation value..............................................22
ogram 4.0: Resetting to factory default settings (pH) ...............................................31
6.5 Pr
ogram 5.0: Viewing previous conductivity/TDS calibration data..........................32
6.6 Pr
ogram 6.0: Viewing conductivity/TDS probe data (effective cell constant) .........33
6.7 Pr
ogram 7.0: Conductivity/TDS measurement configuration ..............................34-37
6.8 Pr
P7.1 Selecting Ready indicator/auto endpoint function.............................................34
P7.2 Selecting °C or °F......................................................................................................35
P7.3 Selecting automatic or manual temperature compensation...............................36
P7.4 Setting the TDS factor ..............................................................................................37
ogram 8.0: Temperature settings.............................................................................38-39
6.9 Pr
P8.1 Setting the temperature coefficient........................................................................38
P8.2 Selecting the normalization temperature ..............................................................39
ogram 9.0: Resetting to factory default settings (conductivity/TDS)..................40
6.10 Pr
7. Probe care and maintenance...........................................................41-42
7.1 pH electrode care..........................................................................................................40-41
7.2 Conductivity probe care....................................................................................................41
8. Troubleshooting....................................................................................43
9. Error Messages......................................................................................44
10. Specifications......................................................................................45
11. Accessories.....................................................................................46-47
6. Advanced setup functions...............................................................23-40
6.1 Advanced setup mode overview................................................................................24-25
pH .......................................................................................................................................24
Conductivity/TDS ............................................................................................................25
ogram 1.0: Viewing previous pH calibration data....................................................26
6.2 Pr
ogram 2.0: Viewing pH electrode data (slope and offset)........................................27
6.3 Pr
ogram 3.0: pH measurement configuration..........................................................28-30
6.4 Pr
P3.1 Selecting Ready indicator/auto endpoint function.............................................28
P3.2 Selecting number of pH calibration points ...........................................................29
P3.3 Selecting °C or °F......................................................................................................30
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
12. Appendix 1: conductivity to TDS conversion factors.........................48
13. Appendix 2: calculating TDS conversion factors................................49
14. Appendix 3: calculating temperature coefficients ............................50
15. Appendix 4: meter factory default settings ......................................51
16. Warranty .............................................................................................52
17. Return of Items ...................................................................................52
32
1. Introduction
pH/Conductivity/°C/°F Meter
pH/CON 300 Series
WATERPROOF
MODE
HOLD
ENTER
RANGE
ON
OFF
CAL
MEAS
SETUP
ATC
READY
MEAS
C
µS
Thank you for selecting an OAKTON meter. This OAKTON portable meter is a microprocessor-based instrument that measures pH, conductivity, TDS and temperature! Your meter has many user-friendly features, all of which are accessible through the membrane keypad.
Your meter includes a single-junction pH electrode, a combination conductivity/ temperature probe, and batteries. Please read this manual thoroughly before operating your meter.
2. Display and Keypad Functions
2.1
Display
The LCD has a primary and secondary display.
• The primary display shows the measured pH or conductivity reading.
• The secondary display shows the temperature of the reading. The display also shows error messages, keypad functions and program functions.
Primary display
1.
2.
3.
CAL
21
20.
SETUP
READY HOLD
MEAS
K=
ON
19.
18.
17.
1. SETup mode indicator
2. MEASurement mode
indicator
3. CALibration indicator
4. mV indicator
5. % indicator
6. pH measurement
indicator
7. Millisiemens indicator
8. Microsiemens indicator
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
OFF
ERR
15.
16.
14.
9. Parts per thousand
indicator
10.
Parts per million indicator
11.
Temperature indicators
12.
pH setup indicator
13.
Automatic Temperature Compensation indicator
Secondary display
14.
ERRor indicator
15.
Low battery indicator
16.
Probe indicator
17.
Calibration indicator
18.
Cell constant indicator
19.
ON/OFF indicator
20.
HOLD indicator
21.
READY indicator
mV
pH mS µS ppt ppm
F
C
pH
ATC
13.
%
4.
5.
6.
7.
8.
9.
10.
11.
12.
54
2.2
Keypad
The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the LCD.
ON/OFF............Powers and shuts off the meter.
HOLD ..............Freezes the measured reading. To activate, press HOLD while in
measurement mode. To release, press HOLD again.
NOTE:
When auto endpoint feature is switched on, meter automati­cally holds reading after 5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release auto endpoint feature.
MODE..............Selects the measurement parameter. Press MODE to toggle between
pH, conductivity and TDS readings. In calibration mode, press MODE to access temperature calibration.
CAL/MEAS......Toggles user between Calibration and Measurement mode. For
example, if you are in pH measurement mode, press CAL/MEAS to enter pH calibration mode. NOTE: Temperature calibration is available from calibration mode; see page 16 for directions. In advanced set-up mode: Press CAL/MEAS to return to main menu from sub menus. Press CAL/MEAS again to return to measurement mode from main menu.
ENTER ............Press to confirm values in Calibration mode and to confirm
selections in Setup mode.
RANGE............Press to switch to manual ranging in Conductivity (or TDS) mode.
▲▼ ..........Press in Setup mode to scroll through subgroups. Also lets
you increment/decrement the values in the conductivity and temperature calibration modes.
SETUP..............Press to enter SETUP mode. SETUP mode lets you
customize meter preferences and defaults, and view calibration and probe data.
3. Preparation
3.1
Inserting the Batteries
Four AAA batteries are included with your meter.
1. Use a Phillips screwdriver to remove the two screws holding the battery cover.
See figure below.
2. Lift off battery cover to expose batteries.
3. Insert batteries. Follow the diagram inside the cover for correct polarity.
4. Replace the battery cover into its original position. Screw cover back into place.
Remove these two screws to access battery compartment
Battery compartment
ON
OFF
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
CAL
MEAS
HOLD
ENTER
RANGE
SETUP
MODE
76
3.2
Probe information
Your meter includes two probes:
pH electrode with BNC connector
conductivity/TDS/temperature probe with a notched 6-pin connector
The temperature sensing element built into the conductivity probe will also compensate for pH readings as long as both probes are in your solution at the same time.
If you want to use an “All in One” pH probe with a built-in temperature element, or if you want to use a separate temperature probe, you will need to disconnect the conductivity/TDS probe to allow for connection of the separate temperature sensor.
You can use any standard pH electrode with a BNC connector with this meter. Conductivity/TDS probes, “All in One” pH electrodes with a built-in temperature element, and temperature probes require a probe with a notched 6-pin connector. For replacement probes, see the “Accessories” section, pages 46-47.
NOTE: Keep connector dry and clean. Do not touch connector with soiled hands.
To connect the pH electrode:
1. Slide the BNC connector of the probe over the BNC connector socket on the
meter. Make sure the slots of the connector are in line with the posts of the socket. Rotate and push the connector clockwise until it locks.
See figure below.
2. To remove probe, push and rotate the connector counterclockwise. While
holding onto the metal part of the connector, pull probe away from the meter.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
To connect the conductivity/TDS/temperature probe:
1. Line up the notch and 6 pins on the meter with the holes in the 6-pin connector.
Push down and turn the locking ring clockwise to lock into place.
See figure below.
2. To remove probe, turn the locking ring counterclockwise on the probe connector.
Pull probe away from the meter.
NOTE: follow the same directions to connect an optional separate temperature
element.
CAUTION: Do not pull on the probe cord or the probe wires might disconnect.
pH electrode (with BNC connector)
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
conductivity/TDS?temperature probe (with 6-pin connector)
locking ring
top view of 6-pin
connector—
conductivity/TDS
temperature probe
98
4. Calibration
4.1
Important Information on Meter Calibration
When you recalibrate your meter, old calibration points are replaced on a “point by point” basis in pH, and on a “range by range” basis in conductivity or TDS.
For example:
pH:
if you previously calibrated your meter at pH 4.01, 7.00, and 10.01, and you recalibrate at pH 7.00, the meter retains the old calibration data at pH 4.01 and pH 10.01.
Conductivity/TDS:
µS in the 0 to 1999 µS range and you recalibrate at 1500 µS (also in the 0 to 1999 µS range), the meter will replace the old calibration data (1413 µS) in that range. The meter will retain all calibration data in other ranges.
TDS values are proportional to conductivity values. Note that if you calibrate a TDS value in an equivalent conductivity range, the TDS value will replace the previous conductivity value, and vice versa.
To view current calibration points:
pH: Program P1.0 in the SETUP section, page 26.
Conductivity/TDS: Program P5.0 in the SETUP section, page 32. To completely recalibrate your meter, or when you use a replacement probe,
it is best to clear old calibration data by resetting the meter.
To reset the meter to its factory defaults:
pH: Program P4.0 in the SETUP section, page 31.
Conductivity/TDS: Program P9.0 in the SETUP section, page 40. NOTE: Resetting the meter will set meter to factory defaults.
Conductivity/TDS and pH must be reset separately.
For directions on how to calibrate your meter:
• See section 4.3 on pages 11-12 for pH calibration
• See section 4.4 on page 13-15 for conductivity and TDS calibration
• See section 4.5 on page 16 for Temperature Calibration
4.2
Preparing the Meter for Calibration
Before starting calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used. For example, if you shut the meter off in "pH" units, the meter will read "pH" units when you switch the meter on.
Do not reuse calibration solutions after calibration. Contaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. See pages 46-47 for information on our high-quality calibration solutions.
if you previously calibrated your conductivity meter at 1413
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
4.3
pH calibration
We recommend that you perform at least a 2-point calibration using standard buffers that bracket (one above and one below) the expected sample range.
Preparing for pH calibration
This meter is capable of up to 5-point pH calibration to ensure accuracy across the entire pH range of the meter. Select from the following buffer options:
pH 1.68, 4.01, 7.00, 10.01, and 12.45. The meter automatically recognizes and calibrates to these standard buffer values,
which makes pH calibration faster and easier.
Be sure to remove the protective electrode storage bottle or rubber cap of the probe before calibration or measurement. If the electrode has been stored dry,
wet the probe in tap water for 10 minutes before calibrating or taking readings to saturate the pH electrode surface and minimize drift.
Wash your probe in deionized water after use, and store in electrode storage solution. If storage solution is not available, use pH 4.0 or 7.0 buffer.
To calibrate pH:
1. If necessary, press the MODE key to
select pH measurement mode. The pH indicator appears in the upper right hand corner of the display.
See figure
A
2. Rinse the pH electrode thoroughly with
de-ionized water or a rinse solution. Do not wipe the probe; this causes a build-up of electrostatic charge on the glass surface.
3. Dip the pH electrode into the calibration
buffer. The end of the probe must be completely immersed into the sample. Stir the probe gently to create a homogeneous sample.
NOTE: The temperature element is in the conductivity cell. For temperature compen­sated readings, dip the conductivity cell into the calibration buffer as well.
CONTINUED ON NEXT PAGE
A
READY
MEAS
pH
C
ATC
1110
4. Press CAL/MEAS to enter pH calibration
mode. The CAL indicator lights. The primary display will show the measured reading while the smaller secondary display will indicate the pH standard
READY
CAL
pH
buffer solution.
See figure
5. Wait for the measured pH value to
B
pH
B
stabilize. If the READY indicator has been activated (set up program P3.1— see page 28), the READY annunciator lights when the reading is stable.
6. Press ENTER to confirm calibration.
The meter is now calibrated to the current buffer. The lower display scrolls through the remaining buffer options.
• If you are performing multipoint calibration, go to step 7.
• If you are performing one-point calibration, go to step 9.
7. Rinse the electrode with de-ionized
READY
MEAS
pH
water or a rinse solution, and place it in the next pH buffer.
8. Follow steps 5 to 8 for additional
calibration points.
9. When calibration is complete,
press CAL/MEAS to return to pH measurement mode.
See figure
C
C
C
ATC
Notes
To exit from pH Calibration mode without confirming calibration, DO NOT press ENTER in step 6. Press CAL/MEAS instead.
If the selected buffer value is not within ±1.0 pH from the measured pH value: the electrode and buffer icon blink and the ERR annunciator appears in the lower left corner of the display.
To limit the number of pH buffer values available during calibration, see Set-up program P3.2 on page 29.
4.3
Conductivity and TDS calibration
Calibrate up to 5-point conductivity or 5 point TDS calibration at one point per range: Range
Conductivity: TDS: R1 0.00-19.99 µS 0.00-9.99 ppm R2 0.0-199.9 µS 10.0-99.9 ppm R3 0-1999 µS 100-999 ppm R4 0.00-19.99 mS 1.00-9.99 ppt R5 0.0-199.9 mS 10.0-199.9 ppt
If you are measuring values in more than one range, make sure to calibrate each of the ranges you are measuring. All new calibration data will over-ride existing stored calibration data for each measuring range you calibrate.
• If you are measuring in ranges near to or greater than 20 mS (10 ppt), or near to
or lower than 100 µS (50 ppm), calibrate the meter at least once a week to get specified ±1% F.S. accuracy.
• If you are measuring in the mid ranges and you washed the probe in deionized
water and stored it dry, calibrate the meter at least once a month.
• If you take measurements at extreme temperatures, calibrate the meter at least
once a week.
Preparing for conductivity/TDS calibration
For best results, select a standard value close to the sample value you are measuring. Alternatively, use a calibration solution value that is approximately
2
3 the full scale value of the measurement range you plan to use. For example,
in the 0 to 1999 µS conductivity range, use a 1413 µS solution for calibration. See the table below for recommended calibration solution ranges:
Conductivity Recommended TDS Recommended Range Cal. Solution Range Range Cal. Solution Range
0.00-19.99 µS 6.00 to 17.00 µS 0.00-9.99 ppm 3.00 to 8.50 ppm
0.0-199.9 µS 60.0 to 170.0 µS 10.0-99.9 ppm 30.0 to 85.0 ppm 0-1999 µS 600 to 1700 µS 100-999 ppm 300 to 850 ppm
0.00-19.99 mS 6.00 to 17.00 mS 1.00-9.99 ppt 3.00 to 8.50 ppt
0.0-199.9 mS 60.0 to 170.0 mS 10.0-199.9 ppt 30.0 to 170 ppt
Temperature coefficient: These meters are factory set to a temperature coefficient of 2.1% per °C. For most applications this will provide good results. See Program P8.1 on page 38 to set the temperature coefficient to a different value. See Appendix 31, “Calculating Temperature Coefficients” on page 50 to determine the appropriate temperature coefficient for your solution.
Normalization temperature: The factory default value for normalization tempera­ture is 25°C. If you need to normalize to a value other than 25°C, see Program P8.2 on page 39.
CONTINUED ON NEXT PAGE
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
1312
To calibrate conductivity:
1. If necessary, press the MODE key to
select conductivity mode.
See figure
A
2. Rinse the probe thoroughly with
de-ionized water or a rinse solution,
then rinse with a small amount of calibration standard.
3. Dip the probe into the calibration
standard. Immerse the probe tip beyond the upper steel band. Stir the probe gently to create a homogeneous sample.
See figure
B
4. Wait for the measured conductivity
value to stabilize. If the READY indica­tor has been activated (set up program P7.1—see page 34), the READY annunci­ator lights when the reading is stable.
5. Press CAL/MEAS to enter conductivity
calibration mode. The CAL indicator will appear above the display.
See figure
C
6. Press the or to change the value on
the primary display to match the value of the calibration standard.
See figure
D
7. Press ENTER to confirm the calibration
value. The meter returns to the MEAS (measurement) mode.
8. Repeat steps 1-7 for other ranges.
A
Immerse probe beyond upper steel band
B
C
D
READY
READY
READY
MEAS
µS
C
ATC
CAL
µS
C
ATC
CAL
µS
C
ATC
Notes
When entering calibration mode, the meter will display the factory default value. If the meter was previously calibrated, the display may “jump” to the factory default value when switching from measurement to calibration mode.
To exit from Conductivity Calibration mode without confirming calibration, DO NOT press ENTER in step 6. Press CAL/MEAS instead. This will retain the meter’s old calibra­tion data in the measuring range of the calibration.
You can offset the conductivity reading up to ±20% from the default setting. If your measured value differs by more than ±20%, clean or replace probe as needed. See page 47 for a wide selection of high-quality calibration standards.
http://www.novatech-usa.com/Products/Laboratory-Equipment-Supply-Products Tel: (281) 359-8538 Toll Free:(866) 433-6682
Calibrating for TDS directly
The factory default setting for TDS conversion factor is 0.5. If your solution has a dif­ferent TDS factor, you can improve calibration accuracy by setting the TDS factor prior to calibration. See SetUp Program P7.4 on page 37 for directions. See Appendixes 1 and 2 on pages 48-49 to determine your exact TDS factor.
1. If necessary, press the MODE key to
select TDS mode.
See figure
A
2. Rinse the probe with de-ionized water or
a rinse solution, then rinse with a small amount of calibration standard.
A
MEAS
READY
ppm
C
ATC
3. Dip the probe into the calibration
standard. Immerse the probe tip beyond the upper steel band. Stir the probe gently to create a homogeneous sample.
See figure
B
Immerse probe beyond upper steel band
4. Wait for the measured TDS value to
stabilize. If the READY indicator has been activated (set up program P7.1— see page 34), the READY annunciator lights when the reading is stable.
5. Press CAL/MEAS to enter TDS calibra-
tion mode. The CAL indicator will appear above the display.
See figure
C
6. Press the or keys to change the
value on the primary display to match the value of the calibration standard.
See figure
D
7. Press ENTER to confirm the calibration
value. The meter returns to the MEAS (measurement) mode.
8. Repeat steps 1-7 for other ranges.
B
CAL
READY
ppm
C
ATC
C
CAL
READY
ppm
C
ATC
D
Notes
When entering calibration mode, the meter will display the factory default value. If the meter was previously calibrated, the display may “jump” to the factory default value when switching from measurement to calibration mode.
To exit from TDS Calibration mode without confirming calibration, DO NOT press ENTER in step 6. Press CAL/MEAS instead. This will retain the meter’s old calibration
data in the measuring range of the calibration. You can offset the TDS reading up to ±20% from the default setting. If your meas-
ured value differs by more than ±20%, clean or replace probe as needed.
1514
Loading...
+ 18 hidden pages