Novatech 1732 Operator's Manual

Oxygen Transmitter
Model 1732
Operators Manual
February 2015
1. USING THIS MANUAL ................................................................................................................................. 3
1.1 CAUTIONS .................................................................................................................................................. 3
1.2 WARNING SYMBOLS ................................................................................................................................... 4
2. INTRODUCTION ........................................................................................................................................... 5
3. DEVICE SPECIFICATIONS .......................................................................................................................... 7
3.1 HARDWARE SPECIFICATIONS ....................................................................................................................... 7
3.2 OPERATIONAL SPECIFICATIONS ................................................................................................................... 8
3.2.1 SCALING OF ANALOG OUTPUTS CHANNEL #1 AND CHANNEL #2 ............................................................. 8
3.2.2 LOCAL DISPLAY OF PROCESS VARIABLES .............................................................................................. 8
4. DISPLAY AND KEYPAD .............................................................................................................................. 9
4.1 RUN MODE DISPLAY ................................................................................................................................... 9
4.2 OXYGEN DISPLAY ..................................................................................................................................... 10
4.3 KEYPAD ................................................................................................................................................... 11
4.3.1 KEYPAD IN RUN MODE ....................................................................................................................... 11
4.3.2 KEYPAD IN THE SETUP MENU ............................................................................................................. 12
4.3.3 INFORMATION SCREEN ....................................................................................................................... 13
5. SETUP MENU ............................................................................................................................................. 15
5.1 SETUP MENU FUNCTION SUMMARY ........................................................................................................... 15
5.2 SETUP MENU DISPLAY .............................................................................................................................. 15
5.3 SETUP MENU FUNCTIONS ......................................................................................................................... 16
5.3.1 PROBE 1 & PROBE 2 OFFSET ............................................................................................................. 16
5.3.2 LOWER LINE ITEMS ............................................................................................................................ 16
5.3.3 OXYGEN DISPLAY UNITS .................................................................................................................... 17
5.3.4 DAMPING FACTOR .............................................................................................................................. 17
5.3.5 PROCESS ALARMS ............................................................................................................................. 17
6. ALARMS ..................................................................................................................................................... 19
6.1 CHECKING AND ACCEPTING AN ALARM ...................................................................................................... 20
6.1.1 CURRENT ALARMS ............................................................................................................................. 20
6.1.2 ALARM LOG ....................................................................................................................................... 21
6.2 ALARM RELAYS ........................................................................................................................................ 21
6.3 COMMON ALARMS .................................................................................................................................... 22
6.4 SELECTABLE PROCESS ALARMS ................................................................................................................ 23
6.5 WARNING MESSAGES ............................................................................................................................... 23
7. GAS CALIBRATE AND PURGE ................................................................................................................ 25
7.1 ACTIONS THAT OCCUR WHEN THE GAS SOLENOID BUTTONS ARE PRESSED ................................................. 26
8. INDEX .......................................................................................................................................................... 27
February 2015 Operators Manual 1732 Oxygen Transmitter 1
Copyright NOVATECH CONTROLS PTY LTD — 2015
This manual describes the transmitter firmware version 1.29, February 2015
Neither the whole nor any part of the information contained in, or t he product described in, this manual may be adapted or reproduced in any material form except with the prior written approval of Novatech Controls Pty Ltd (Novatech).
The product described in this manual and products for use with it are subject to continuous developments and improvement. All information of a technical nature and particulars of the product and its use (including the information in this manual) are given by Novatech in good faith. However, it is acknowledged that there may be errors or omissions in this manual. A list of details of any amendments or revisions to this manual can be obtained upon request from Novatech Controls Technical Enquiries. Novatech Controls welcome comments and suggestions relating to the produ ct and this manual.
All correspondence should be addressed to: ­ Technical Enquiries
Novatech Controls Pty Ltd 309 Reserve Road, Tel: +61 3 9585 2833 Cheltenham Fax: +61 3 9585 2844 Victoria 3192 Email: info@novatech.com.au Australia Web site: http://www.novatech.com.au/
Novatech Controls or their authorised dealers should carry out all maintenance and service on the product. Novatech Controls can accept no liability whatsoev er for any loss or damage caused by service or maintenance by unauthorised personnel. This manual is intended only to assist the reader in the use of the product, and therefore Novatech Controls shall not be liable for any loss or damage whatsoever arising from the use of any information or particulars in, or any error or om i ssion in, this manual, or any incorrect use of the product.
Operators Manual February 2015 2 1732 Oxygen Transmitter
1. USING THIS MANUAL
This manual is a shortened version of the 1732 Oxygen Transmitter product manual to be used by the operator. It does not cover the commissioning, calibration or servicing of the 1732 Oxygen Transmitter. If more detailed information is required than what is provided in this manual please refer to the more detailed 1732 Oxygen Transmitter Technical Manual. B oth manuals are supplied with each transmitter and are also available to download at the Novatech Controls website.
http://www.novatech.com.au/1732
It is assumed in this manual that the transmitter has been installed by qualified personal and that the wiring to the main power supply, the oxygen probe and all the associated signal devices complies with the local safety codes and regulations.
1.1 Cautions
Please read the safety information below before co nnecting power to the transmitter.
CAUTION 1
The probe heater is supplied with MAINS VOLTAGE. This supply has electrical shock danger to maintenance personnel. Always isolate the transmitter before working with the probe. The EARTH wire (green) from a heated probe must ALWAY S be connected to earth.
CAUTION 2
Combustion or atmosphere control systems can be dange rous. Burners must be mechanically set up so th at in the worst case of equipment failure, the system cannot generate explosive atmospheres. This danger is normally avoided with flue gas trim systems by adju stment so that in the case of failure the appliance will not generate CO in excess of 400 ppm in the flue. The CO level in the flue should be measured with a separate CO instrument, normally an infrared or fuel cell t ype.
CAUTION 3
The oxygen probe is heated to over 700°C/1290°F and is a source of ignition. Since raw fuel leaks can occur during burner shutdown, the transmitter has an interlocking relay that removes power from the probe heater when the main fuel shut-off valve power is off. If this configuration does not suit or if it is possible for raw fuel to come into contact with a hot oxygen probe then t he M odel 1732 transmitter with a heated probe will not be safe in your application. An unheated probe can be utilised in such application s, however the oxygen readings are valid only above 650°C/1200°F.
CAUTION 4
The reducing oxygen signal from the transmitter and the associated alarm r el ay can be used as an explosive warning or trip. This measurement assumes complete combustion. If incomplete combustion is possible then this signal will read less reducing and should not be used as an alarm or trip. A true excess combustibles analyser, normally incorporating a c atalyst or thermal conductivity bridge, would be more appropriate where incomplete combustion is possible. Also read the probe electrical shock caution in the pro be heat er interlock caution in chapter 4.7 of the Model 1732 Technical Manual.
CAUTION 5
FIL-3 filter. If the optional FIL-3 has been fitted to the 1231 probe in this installation, please read the Important Notice in the Technical Manual.
February 2015 Operators Manual 1732 Oxygen Transmitter 3
1.2 Warning Symbols
Danger, high voltage. Risk of electrical shock.
Caution hot surface.
Caution, risk of danger. See additional informatio n in the manual.
Operators Manual February 2015 4 1732 Oxygen Transmitter
2. INTRODUCTION
The Novatech 1732 Dual Probe Oxygen Transmitter is designed to measure the oxygen gas con cent ration in combustion applications such as power station s and industrial boilers. The measurement is used to control the air / fuel ration in the burner to maintain efficiency and safe operation conditions.
The 1732 Oxygen Transmitter is based on the well-known model 1632 transmitter. It includes a number of hardware and software improvements such as a gra phic display, larger characters, faster microprocessor, simplified set up menu, alarm logging, faster probe heater control and more calculated values such as burner efficiency.
The 1732 Oxygen Transmitter has a variety of user-selectable functions. They are simple to use because each selection is menu driven. For options you are not sure about, read the manual on that particular item in chapter 5, Setup Menu.
Features include:-
Inputs
Two zirconia oxygen probe, heated or unheated Oxygen range from 1 x 10
-30
to 100% Furnace, kiln or flue thermocouple, field selectable as type K, J, R or S Main flame established safety interlock (for heated probes only) Purge flow switch
Outputs
Two linearised 4-20mA or 0-20mA DC isolated outputs, max. load 1000 The output function and the range are field selectabl e Common alarm relay (programmable) Three other alarm relays with selectable functions
Computer
RS 232 or RS 485 for connection of a computer terminal or printer for diagnostics of the transmitter, probe or combustion process. This connection is suitable f or network connection to a computer, DCS or PLC using MODBUS protocol.
Display
Multi font graphical display Large characters for the oxygen on the top line Multiple lower line items for the secondary functions. i e P robe temperature, % Oxygen for the second probe Alarm display mode that shows the time the alarm occurred and the acceptance time Alarm log mode that keeps the time the alarm occurred, the acceptance time and the time the alarm was cleared for the last 4,000 alarms
Power
Universal mains supply voltage, 100 to 240VAC Automatically detects the mains voltage and freque ncy and set the power control accordingly
February 2015 Operators Manual 1732 Oxygen Transmitter 5
This page has been intentionally left blank.
Operators Manual February 2015 6 1732 Oxygen Transmitter
3. DEVICE SPECIFICATIONS
3.1 Hardware Specifications
Number of Oxygen Probes:
1 or 2
Oxygen Range:
1 x 10-30 to 100%
Oxygen Accuracy:
±1% of actual measured oxy gen value with a repeatability of ±0.5% of the
measured value
Thermocouple Types:
Type K and J
Temperature Accuracy:
±2°C
Analog Outputs:
0-20mA or 4-20mA field selectable
Active Outputs
(WARNING: DO NOT LOOP POWER OUTPUTS. Use only passive receivers for commissioning and
testing. The use of loop powered receivers will damage the output)
Output Load:
1000 ohm max
Alarm Relays:
4
Alarm Relay Contacts:
2A 240VAC, 2A 30VDC
Reinforced insulation when used with mains voltage
(WARNING: Do not use both mains voltage and low voltage connections to adjacent alarm contacts)
Mains Voltage Supply:
100 to 240VAC 50/60 Hz
Reinforced insulation
Overvoltage:
Category II (IEC60364-4-443)
Power:
5 Watts for controller plus probe power
530W max., 25% duty cycle each probe on 240VAC 110W max., 100% duty cycle each probe on 110VAC 576W (2.4A) max
Environmental Rating:
Operating Temperature -25°C to 55°C
Relative Humidity 5% to 95% (non-condensing)
Altitude
2000m Maximum
Degree of Protection:
IP65
IP54 with internal reference air pump
Case Size:
315mm (12.4”) wide, 190mm (7.5”) high, 110mm (4.3”) deep
Case Weight:
3 Kg (6.6 lbs.)
WARNING: All signal level connections onto the transmitter must be treated as safety extra-low voltage (SELV) as defined in the standard IEC61140. Double insulation must be used when connecting these terminals to systems that might carry high voltage.
February 2015 Operators Manual 1732 Oxygen Transmitter 7
Loading...
+ 21 hidden pages