Nova Stat Profile PRIME CCS Analyzer User Manual

52920 EN
Only
Stat Prole PRIME
®
CCS Analyzer
Instructions for Use
Manual
Nova Prime CCS Analyzer Quick Start Guide
Conrm the analyzer is
1
Ready for analysis.
Displays Ready, desired test menu shows no Orange
Icons.
Login if necessary. Press
2
the Login icon then enter or scan your User ID and Password.
Select the Sample Container and the Panel.
3
Prepare the sample for analysis then press
4
to extend the sample probe.
Position the sample
5
over the sample probe then press to aspirate the sample.
Enter any additional
6
patient or sample information needed while the analysis is completed.
Review results.
7
LEVEL
REP
EC
SN
NOVA BIOMEDICAL SYMBOL DIRECTORY
In vitro diagnostic medical device
Device for near-patient testing
Caution
Consult Instructions for Use
Biological risk
Catalog number
Manufactured by
Control
Level
YYYY - MM - DD
Only
Batch code
Serial number
Temperature limitation
Use by date
Electronic Waste
Authorized representative in the European Commission
Laser radiation: Do not stare into beam
Prescription use only
Prime CCS Instructions for Use Manual
Revision History
PN52920 Prime CCS Instructions for Use Manual
Rev Release Description
H 05 – 2022 Update to meet IVDR requirements
J 01 – 2023 Hazardous Waste Disposal Advisory
Preface
Ordering Information
The Stat Prole PRIME CCS Analyzer Instructions for Use Manual can be ordered from Nova Biomedical Order Services. Call, fax, or write to:
Nova Biomedical Corporation Phone: 1-800-458-5813
200 Prospect Street FAX: 1-781-893-6998 Waltham, MA 02454 (in the USA) USA
The Stat Prole Prime CCS analyzer is manufactured in the USA by Nova Biomedical Corporation.
REP
EC
Nova Biomedical GmbH Phone: + 49 (0) 61054505-0 Hessenring 13A, Geb. G FAX: + 49 (0) 61054505-37
64546 Mörfelden-Walldorf
Germany
Authorized Representative
+1-781-891-9718
(outside the USA)
Web: www.novabiomedical.com
Trademark
Stat Prole is a registered Trademark of Nova Biomedical Corporation.
Stat Prole Prime is a registered Trademark of Nova Biomedical Corporation.
®
®
Copyright
Printed in the U.S.A. Copyright 2023, Nova Biomedical Corporation, Waltham, MA 02454.
i
Prime CCS Instructions for Use Manual
T echnical Assistance
For technical assistance inside the United States, call Nova Biomedical Technical Services:
1-800-545-NOVA (6682)
1-781-894-0800
FAX: 1-781-894-0585
For technical assistance outside the United States, call your local Nova subsidiary or authorized distributor.
Nova Biomedical recommends users report any serious incidents/ adverse events to Nova Biomedical or Nova Biomedical’s Authorized Representative, as well as to local Competent Authorities as required.
Nova Biomedical Canada Ltd.
17-2900 Argentia Road
Mississauga, Ontario L5N 7X9
Canada
Tel: 1 905 567 7700
Toll Free: 800 263 5999
Fax: 1 905 567 5496
ca-info@novabio.com
Nova Biomedical France
Parc Technopolis
Bât. Sigma 3 Avenue du Canada
91940 Les Ulis Courtaboeuf
France
Tel: (33) 1-64 86 11 74
Fax: (33) 1-64 46 24 03
Nova Biomedical UK
Innovation House
Aston Lane South
Runcorn, Cheshire,
WA7 3FY, UK
Tel: + 44 1928 704040
Fax: + 44 1928 796792
uk-info@novabio.com
Nova Biomedical Iberia, S.L.
c/Vic 17, Planta 3A
08173 Sant Cugat del Vallès
Barcelona
Spain
Tel: +34 935531173
es-info@novabio.com
or pt-info@novabio.com
Nova Biomedical Italia Srl
Via Como 19
20045 Lainate Milano
Italy
Tel: +39 02 87070041
Fax: +39 02 87071482
it-info@novabio.com
Nova Biomedical GmbH
Hessenring 13A, Geb. G
64546 Mörfelden-Walldorf
Germany
Tel: + 49 6105 4505-0
Fax: + 49 6105 4505-37
de-info@novabio.com
Nova Biomedical Benelux BV
Europalaan 4
5232 BC Den Bosch
The Netherlands
Tel: +31 (0) 733032701
NL-info@novabio.com
or BE-info@novabio.com
Nova Biomedical Switzerland
GmbH
Herostrasse 7
8048 Zürich Switzerland
Tel: +41-41-521-6655
Fax: +41-41-521-6656
e-mail: ch-info@novabio.com
Nova Biomedical Brasil
Rua Massena, 107
Jardim Canadá
Nova Lima – MG Cep: 34007-746
Brasil
Tel: (55) 31 3360-2500
E-mail: sac@novabiomedical.com.br
Nova Biomedical KK
Harumi Island
Triton Square Ofce Tower X 7F
1-8-10 Harumi, Chuo-ku
Tokyo 104-6007
Japan
Tel: 03-5144-4144
Fax: 03-5144-4177
jp-info@novabio.com
ii
Contents
1 Introduction 1-1
1.1 About This Manual .....................................................................1-1
1.2 Safety .........................................................................................1-2
1.3 Installation and Use ...................................................................1-5
1.4 Requirements ............................................................................1-5
1.5 Cleaning the Analyzer ................................................................1-6
1.6 Intended Use, Tests Performed, and Clinical Utility ...................1-6
1.7 The Sample ...............................................................................1-8
1.7.1 Handling Requirements..................................................1-8
1.7.2 Acceptable Anticoagulants .............................................1-9
2 Getting Started 2-1
2.1 Power Up Procedure .................................................................2-3
2.2 The Home Screen: Ready ........................................................2-4
2.2.1 Header Bar .....................................................................2-4
2.2.2 Selection Area ................................................................2-5
2.2.3 Sensor Status Screen ....................................................2-6
2.2.4 Menu Bar .......................................................................2-7
2.3 Login to the Analyzer .................................................................2-8
2.4 Automatic Calibrations ...............................................................2-8
2.4.1 Manual Calibrations
2.4.2 Air Detector Calibration
3 Sample Analysis 3-1
..........................................................2-9
.....................................................2-9
3.1 Analyzing Patient Samples ........................................................3-1
3.1.1 Analyzing Syringe Samples ...........................................3-1
3.1.2 Utilizing the Safety Sample Port ....................................3-3
3.1.3
Analyzing Sample from a Blood Collection Tube
3.1.4 Analyzing Sample from a Capillary Tube .......................3-6
3.2 The Sample Results Display ......................................................3-7
3.2.1 Sample Result Printout ..................................................3-8
3.3 Analyzing QC and Prociency Samples ...................................3-9
3.3.1 Analyzing Internal QC Samples ...................................3-10
3.3.2 Analyzing External QC Samples ..................................3-11
3.3.3 Analyzing Prociency Samples ....................................3-12
52920J EN 2023 – 01
.................3-5
iii
Prime CCS Instructions for Use Manual
4 Reviewing Patient and QC Data 4-1
4.1 Reviewing Patient Data .............................................................4-1
4.2 Reviewing QC Data ...................................................................4-3
4.3 QC Statistics ..............................................................................4-4
4.4 Levey Jennings Graphs .............................................................4-4
5 Consumables Replacement 5-1
5.1
Calibrator Cartridge/Auto QC Cartridge Replacement
5.1.1 Replace the Calibrator Cartridge....................................5-3
5.1.2 Replace the Auto QC Cartridge......................................5-5
5.2 Replace the MicroSensor Card ..................................................5-6
5.3 Prime MicroSensor Card Warranty Process ..............................5-8
6 Periodic Replacements 6-1
6.1 Pump Tubing Harness Replacement .........................................6-1
6.1.1 Install the Pump Tubing .................................................6-3
6.2 Probe Replacement ...................................................................6-4
6.3 Reference Cartridge Replacement ............................................6-6
6.3.1 Install the Reference Cartridge .........................................6-7
6.4 Printer Paper Replacement .......................................................6-8
6.5 Safety Sample Port Replacement ..............................................6-9
7 Troubleshooting 7-1
.....................5-2
7.1 Event Log ...................................................................................7-2
7.2 Resolving Event Codes .............................................................7-2
7.2.1 Flow Event Codes .............................................................7-3
7.2.2 Printer Event Codes ..........................................................7-7
7.2.3 MicroSensor Card Event Codes .......................................7-8
7.3 Troubleshooting Flow Problems ..............................................7-14
7.3.1 The Flow Path Flush Tool ...............................................7-14
7.3.2 Flushing the Flow Path ...................................................7-15
7.3.3 Flushing the Sample Probe/S-Line .................................7-17
7.3.4 Flow T est ......................................................................7-18
7.4 Long-Term Shutdown ............................................................... 7-19
iv
Contents
A Appendix A-1
A.1 Specications .............................................................................A-1
A.2 Quality Control and Calibration ..................................................A-3
A.2.1 Traceability of Calibrators, Controls, and Standards .......A-3
A.2.2 Quality Control ................................................................. A-3
A.2.3 Calibrator Cartridge ......................................................... A-4
A.3 Analytical Specicity, Human Whole Blood....................................A-4
A.4 Healthcare Professional Analytical Performance Studies ...........................
A.4.1 Method Comparison Study .............................................. A-8
A.4.2 Analytical Precision or Repeatability .............................A-16
A.4.3 Run-to-Run or Total Imprecision .................................... A-22
A.5 Point-of-Care/Near Patient Testing Performance Studies .......A-28
A.5.1 Within-Run Precision or Reproducibility ........................ A-29
A.5.2 Quality Control Total Imprecision ................................... A-34
A.5.3 Syringe Mode Method Comparison ............................... A-35
A.5.4 Capillary Mode Method Comparison ............................. A-36
A.6 Reference Values ....................................................................A-38
A.7 Cybersecurity ...........................................................................A-39
A.7.1 Cybersecurity Protection Overview .............................A-39
A.7.2 Software Updates .......................................................A-39
A.7.3 Operating System Patches .........................................A-39
A.7.4 Malware Control ..........................................................A-39
A.7.5 Creation of Software for Release ................................A-40
A.7.6 Security Related to the USB Port ................................A-40
A.8 Ordering Information ................................................................A-41
A.9 Warranty ..................................................................................A-42
A-8
B Principles of Measurement B-1
B.1 Measured Values .......................................................................B-1
B.1.1 Sodium, Potassium, Chloride, and Ionized Calcium .....B-1
B.1.2 pH Sensor .....................................................................B-2
B.1.3 Partial Pressure of Carbon Dioxide ...............................B-3
B.1.4 Partial Pressure of Oxygen ...........................................B-4
B.1.5 Hematocrit ..................................................................... B-4
B.1.6 Glucose ......................................................................... B-4
B.1.7 Lactate .......................................................................... B-5
B.2 Calculated Values ......................................................................B-6
B.2.1 Temperature Correction for Measured Values ..............B-6
B.2.2 Bicarbonate Concentration ...........................................B-7
B.2.3 Total Carbon Dioxide Content ....................................... B-7
B.2.4 Hemoglobin ................................................................... B-7
v
Prime CCS Instructions for Use Manual
B.2.5 Base Excess of Blood ...................................................B-8
B.2.6 Standard Bicarbonate Concentration ............................B-8
B.2.7 Base Excess Extracellular Fluid ....................................B-9
B.2.8 Oxygen Content ............................................................B-9
B.2.9 Oxygen Saturation ......................................................B-10
B.2.10 Alveolar Oxygen .......................................................... B-10
B.2.11 Arterial Alveolar Oxygen Tension Gradient
B.2.12 Arterial Alveolar Oxygen Tension Ratio ........................B-11
B.2.13 P50 or PO2 (0.5) ......................................................... B-12
B.2.14 Ionized Calcium Normalized to pH 7.4 ........................ B-13
B.2.15 Anion Gap ................................................................... B-14
B.2.16 Oxygen Capacity of Hemoglobin (O2Cap) .................. B-14
B.2.17 PO2/FIO2 Oxygenation Index (PO2/FI) ....................... B-14
B.2.18 Respiratory Index (RI) ................................................. B-15
References
........................................................................................B-17
...................B-11
vi
1 Introduction
1 Introduction
This manual provides all necessary instructions for the routine operation and upkeep of the Stat Prole Prime CCS Analyzer. Please read this manual carefully. It has been prepared to help you attain optimum performance from your Analyzer.
Throughout this manual: NOTE indicates especially important information; CAUTION indicates information
that is critical to avoid instrument damage or incorrect
results; WARNING indicates possible hazard to the
operator.
WARNING: Blood samples and blood products
are potential sources of infectious agents.
Handle all blood products, ow path components, and accessories (waste-line,
capillary adaptor, probe, MicroSensor Card,
clot catchers, ush tools/xtures, safety sample port, etc.) with care. Gloves and protective clothing are recommended. When performing maintenance and troubleshooting procedures, also use protective eyewear.
1.1 About This Manual
This manual is for the Stat Prole Prime CCS Analyzer.
This section introduces the Prime CCS Analyzer and covers requirements, tests performed, procedural limitations, clinical utility, and sample handling.
1. Intro.
1-1
Prime CCS Instructions for Use Manual
1.2 Safety
Personnel operating this analyzer must be procient in the operating and replacement procedures of the analyzer. The following safety procedures must be followed.
General Safety
1. Read the safety and operating instructions before operating the analyzer.
2. Retain the safety and operating instructions for future reference.
3. Observe all warnings on the analyzer and in the operating instructions.
4. Follow all operating and use instructions.
5. Do not use the analyzer near water, for example near a sink, etc.
6. Use only with a cart or stand that is recommended by the manufacturer.
The analyzer and cart combination should be
used with care. Quick stops, excessive force, and uneven surfaces may cause the analyzer and cart combination to overturn.
7. Place the analyzer so that its location or position does not interfere with its proper ventilation.
8. Place the analyzer away from heat sources.
9. Connect the analyzer to a power supply only of the type described in the operating instructions or marked on the analyzer.
10. Do not defeat the safety purpose of the polarized or grounding type plug.
11. Route power cords so that they are not likely to be walked on or pinched by items placed upon or against them, paying particular attention to cords at plugs, power sockets, and at the point where they exit from the analyzer.
1-2
1 Introduction
12. The analyzer should be cleaned only as recommended by the manufacturer.
13. Take care not to let objects or liquids fall into the analyzer.
14. The analyzer should be serviced by qualied service personnel.
15. Do not attempt to service the analyzer beyond that described in the operating instructions. All other servicing should be referred to qualied service personnel.
Electrical Safety
1. To reduce risk of electric shock, do not remove the cover.
2. There are no user serviceable parts inside the analyzer.
3. Servicing must be done by qualied service personnel.
4. To reduce the risk of re or electric shock, do not expose the analyzer to water.
5. Use Nova Part Number 52413 external power supply to power up the analyzer.
6. Ensure that the wall outlet receptacle is properly wired and earth grounded.
7. DO NOT use a 3-to-2 wire plug adaptor.
8. DO NOT use a 2-wire extension cord or a 2-wire multiple-outlet power strip.
Chemical and Biological Safety
1. Observe all precautionary information printed on the original solution containers.
2. Operate the analyzer in the appropriate environment.
1. Intro.
1-3
Prime CCS Instructions for Use Manual
3. Take all necessary precautions when using pathologic or toxic materials to prevent the generation of aerosols.
4. Wear appropriate laboratory attire, e.g., safety glasses, gloves, lab coat, and breathing apparatus, when working with hazardous materials.
5. Dispose of all waste solutions according to standard hospital procedures.
Barcode Scanner Safety
A Class 2 laser is incorporated into the bar code scanner.
WARNING: Do not stare into the laser beam.
Laser Specications:
• Wavelength: 650 nm
• Max Output: 1.9 mW
EN 60825-1: 2014
The laser complies with 21 CFR 1040.10 and
1040.11, except for deviations pursuant to Laser
Notice No. 50, dated June 24, 2007.
Disposal of Electronic Waste
This symbol ( ) on the product label indicates that the product should not be treated as household waste.
Devices/Accessories: To ensure the product is disposed
properly, decontaminate the product according to the instructions provided in section 1.5 of this manual and hand over the product to the applicable collection point for the recycling of electrical and electronic equipment.
1-4
1 Introduction
1.3 Installation and Use
This section covers the installation requirements and assembly procedures for the Stat Prole Prime CCS Analyzer.
Prior to use of the analyzer, operators should be familiar with Chapter 2 Operation and Chapter 3 Operating Procedures.
NOTE: Under the Warranty, a Nova service
representative will install this equipment for you.
1.4 Requirements
Working Area Requirements (Environmental):
Keep the working area around the system free of dirt, corrosive fumes, vibration, and excessive temperature changes.
Table 1-1 Prime CCS Requirements
Electrical Requirements
Operating Voltage Range 100 – 240 VAC Operating Frequency 50 – 60 Hz Power Consumption Less than 100 W
Ambient Operating Temperature 15 °C – 32 °C (59°F – 89.6°F) Operate at Humidity 20 to 85% without condensation Operate at Altitude 12,000 feet (3650 meters)
Dimensions
Height 15.4 in (39.1 cm) Width 12.0 in (30.5 cm) Depth 14.4 in (36.2 cm)
Weight
17.5 lb (8.164 kg) without reagent pack 23 lb (10.45 kg) with full reagent pack
1. Intro.
1-5
Prime CCS Instructions for Use Manual
Lifting the Analyzer:
1. One person is needed to lift the analyzer.
CAUTION: Never use the door (open or closed) to
assist you in lifting the analyzer. The door cannot support the weight of the analyzer.
2. From the front of the analyzer, place your hands under each side of the analyzer.
3. Lift the analyzer. Remember to bend your knees and lift with your legs and not your back.
4. Place the analyzer onto a clean and at surface.
1.5 Cleaning the Analyzer
Nova Biomedical recommends using 70% Reagent Alcohol (V/V) or Isopropyl Alcohol (IPA) for cleaning the various analyzer surfaces or components when required. Use a lint-free cloth lightly dampened with the cleaning reagent to wipe down analyzer surfac es. Never spray or pour reagent directly onto or into the analyzer. Once wiped down, all residual uid should be dried with a lint­free cloth.
1.6 Intended Use, Tests Performed, and Clinical Utility
Intended Use
The Stat Prole Prime CCS Analyzer System is intended for in vitro diagnostic use by health care professionals in clinical laboratory settings and in Point-of-Care/Near­Patient Testing settings for the quantitative determination
of pH, PCO2, PO
and Lac (Lactate) in heparinized whole blood.
Measured Parameters
Stat Prole Prime CCS Analyzer: pH, PCO2, PO
Na+, K+, Cl-, iCa, Glu (Glucose), and Lac (Lactate).
, Hct, Na+, K+, Cl-, iCa, Glu (Glucose),
2
, Hct,
2
1-6
1 Introduction
Calculated Parameters
From the directly measured results, the calculated results are shown in Table 1-2.
Table 1-2 Calculated Parameters
pH, PCO2, PO2 (corrected to patient temperature)
Bicarbonate level (HCO Total Carbon Dioxide (TCO2)
Base Excess of the blood (BE-b)
Base Excess of extracellular uid
(BE-ecf) Standard Bicarbonate Concentration
(SBC) Oxygen Content (O2Ct) Oxygen Capacity (O2Cap) Alveolar Oxygen (A)
Clinical Utility
3
1
-
)
Arterial Alveolar Oxygen Tension Gradient (AaDO
Arterial Alveolar Oxygen Tension Ratio (a/A)
Respiratory Index (RI) P50
PO2/FIO2 ratio
Oxygen Saturation (SO2%) Hemoglobin (Hb)
Anion Gap Normalized Calcium (nCa)
)
2
Table 1-3 provides the clinical utility information for each of the analytes measured on the Stat Prole Prime CCS Analyzer.
Table 1-3 Prime CCS Analyzer Clinical Utility
1. Intro.
pH
PCO
PO
Whole blood measurement of certain gases in Whole blood, or pH of whole blood, is used in the Diagnosis and treatment of
2
life-threatening acid-base disturbances.
2
Whole blood measurements of the packed red cell volume of
Hct
a blood sample are used to distinguish normal from abnormal states, such as anemia and erythrocytosis (an increase in the number of red cells).
Na
+
Sodium measurement is used in the diagnosis and treatment of electrolyte imbalance.
1-7
Prime CCS Instructions for Use Manual
Table 1-3 Prime CCS Analyzer Clinical Utility
Potassium measurement is used to monitor electrolyte balance in the diagnosis and treatment of disease conditions character­ized by low or high potassium levels.
Chloride measurement is used in the diagnosis and treatment of electrolyte and metabolic disorders.
Calcium measurements are used in the diagnosis and treat­ment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
Glucose measurement is used in the diagnosis and treatment of carbohydrate metabolism disturbances including diabetes mellitus and hypoglycemia.
Lactate (lactic acid) measurement is used to evaluate the acid­base status of animals suspected of having lactic acidosis.
K
Cl
iCa
Glu
Lac
+
-
1.7 The Sample
Lithium heparin whole blood samples from syringes, open tubes, small cups, and capillary tubes can be used on the Stat Prole Prime CCS Analyzer. The minimum sample size for analysis is 100 µL.
1.7.1 Handling Requirements
pH, PCO2, PO
Correct sample handling is critical to ensure that the blood gas values obtained accurately reect the in vivo state. Ensure that all samples have been obtained and stored following consistent, clinically accepted protocols. It is particularly important to ensure that samples are well mixed before introduction into the analyzer. Nova Biomedical recommends that you analyze the sample within 15 minutes for blood gases. Storing samples on ice is not recommended. Using iced samples may elevate
the PO
2
2
result.2
1-8
1 Introduction
Potassium
Correct sample handling is critical to ensure whole blood potassium values obtained accurately reect the in vivo state. For example, a hemolyzed specimen of 50 mg/dL hemoglobin will increase the potassium blood concentration by 3%.3
1.7.2 Acceptable Anticoagulants
Lithium heparin is the acceptable anticoagulant for use with the analyzer.
EDTA, citrate, oxalate, sodium heparin, and sodium uoride ARE NOT acceptable for use.
Depending on the amount of heparin used in the collection syringe and whether it is lled to capacity with blood, heparin concentrations of 20 I.U. per mL to over 100 I.U. per mL may result.
Liquid or dry heparin when present in excess may cause errors. Ensure blood collection devices are lled per manufacturer instructions.
Ensure blood collection devices are lled per manufacturer instructions.
Our experience suggests that lyophilized lithium heparin giving a nal concentration in blood of not more than 20 I.U. per mL is acceptable.
1. Intro.
CAUTION:
1-9
Stat Prole Prime CCS Analyzer users should take careful note of these considerations when establishing reference intervals and interpreting results.
2 Getting Started
2 Getting Started
The Stat Prole Prime CCS Analyzer is pictured below.
1
4
1. Touch-screen Display
2. Printer
3. Sampler
4. Door/Front Panel
Figure 2.1 Nova Stat Prole Prime CCS
2
3
2. Started
2-1
Prime CCS Instructions for Use Manual
1
3
8
9
7
4 5
1. Waste Line
2. Reference Line
3. Pump and Pump Tubing Cartridge
4. Calibrator Cartridge Opening
5. Control Cartridge Opening
6. Sampler
7. Air Detector
8. MicroSensor Card (under cover)
9. Reference Cartridge (under cover)
2
6
2-2
Figure 2.2 Analytical Compartment
2 Getting Started
2.1 Power Up Procedure
When the analyzer is powered on, it displays the Stat Prole Prime CCS logo. During this time, an internal Power On Self Test (POST) is run. Any errors encountered during the POST will display on the analyzer's screen.
After successfully completing the POST, the Home
screen displays with Initializing. During initialization,
an internal diagnostic sequence is run: the MicroSensor Card use life; the calibrator cartridge uid level; and the internal auto QC cartridge uid level are checked.
Figure 2.3 Initializing Screen
2. Started
The Prime CCS performs a prime cycle. After completion,
the screen displays Not Ready.
Figure 2.4 Not Ready Screen
2-3
Prime CCS Instructions for Use Manual
2.2 The Home Screen: Ready
The screen of the Prime CCS Analyzer is a Touch-screen. The touch-screen display provides prompts, menus, status information, sensor status, sample container and panel selection, and date and time.
2.2.1 Header Bar
The Header Bar is the top section of the display. This is where Ready or Not Ready, Date and Time, Login,
and MicroSensor Card, Calibrator Cartridge, and Control Cartridge status are displayed.
Header Bar
Selection Area
Menu Bar
Figure 2.5 Home Screen: Ready
2-4
The current Date and Time is displayed.
When a timed operation is in process, the Date
and Time is replaced by a countdown timer.
Login using the Lock icon displayed in the
Header Bar. Press the Lock and proceed to login with your Operator ID and password.
Only one person can be logged into the
analyzer at a time. When logged onto the analyzer, an open lock is shown with the logged in operator ID displayed under it.
2 Getting Started
The analyzer can also be run with the login featured turned off.
The upper right corner of the Home screen
(Header Bar) has the Status Graph which when touched will display the status of the MicroSensor Card, calibrator cartridge, and QC cartridge.
2.2.2 Selection Area
The Selection Area is the middle of the display. Panel
selection, Sample container selection, and information about sensor availability are found here.
Figure 2.6 Selection Area
Analytes displayed in Orange are unavailable
for analysis. If you press the icon of the Orange Analyte, the Sensor Status Screen with additional information will display.
Analytes displayed in Blue are available and
selected for analysis. If you press the icon of
a Blue Analyte, it turns Grey indicating it is not
selected for analysis.
Analytes displayed in Grey are available but
not selected for analysis. If you press the icon
of a Grey Analyte, it turns blue, indicating it is
selected for analysis.
Use the Container button to select thetype of conatiner and sample to be analyzed.
Use the Panel button to select from a predened list of test panels.
2. Started
2-5
Prime CCS Instructions for Use Manual
2.2.3 Sensor Status Screen
Analytes displayed in Orange are unavailable for
analysis. Press an orange analyte button to see the Sensor Status Screen, which displays the status of that analyte. Detected sensor errors or QC Lockout conditions display on the screen.
Touching the Calibrate button starts a Calibration
sequence then returns to the Home Screen.
Touching the QC button displays the Analyze QC Screen if more than one Internal QC is locked out or an External QC is locked out for all the sensors.
Touching the QC button starts the QC Level sequence for the QC Lockout, then returns to the Home Screen if there is only 1 Internal QC Locked out for all the sensors.
Touching the Fix button starts a Calibration sequence, then returns to the Home Screen.
If all sensors pass Calibration from a Fix, all Internal QCs that failed QC Lockout will be executed.
Touching the Right Arrow button displays the status of the next sensor that is unavailable.
Touching the Left Arrow button displays the status of the prior sensor that is unavailable.
2-6
Figure 2.7 Sensor Status Screen
2 Getting Started
2.2.4 Menu Bar
The Menu Bar is the bottom section of the screen. The
Tool Box icon (System Menu screens), Find Results icon, QC icon (to run QC and QC Menu Screens), and the Start (Run Test) or Calibrate icon.
The Home icon returns the analyzer to the Home
screen by touching this icon. This icon does not display on the Home screen.
The Tool Box icon is located at the Menu Bar.
Press this icon to display Screen one of the System Menus. The up/down arrow key is pressed to display screen two. From the System Menu, you can also navigate to the Setup Menu.
The recall results icon of the Menu Bar will
display all the patient results stored on the analyzer.
The QC icon will display the QC Menu screen:
Run QC, Setup QC Levels, View QC Data, and Setup QC Operations.
The Calibrate icon is displayed when all analytes
are not calibrated. Press Calibrate to initiate a
system calibration.
If one or more analytes are calibrated, the Start
icon displays. Press Start to begin an analysis.
The Prime icon is displayed in the footer after an
Auto QC Cartridge, a Control Cartridge, or tubing
is replaced. Press Prime to initiate a system
prime.
Pressing the Enter button moves the analyzer to
the next screen in the procedure.
Screens may contain other navigational icons including:
Press the Back icon to return to the previous
screen.
The Page Up and Page Down icons scroll
through the menus that have multiple pages.
2. Started
2-7
Prime CCS Instructions for Use Manual
2.3 Login to the Analyzer
From the Home screen, login if you are prompted to login.
1. Press the Login icon to log into the analyzer.
2. Enter or scan your Operator ID then press the
Enter button.
3. If required, enter or scan your Password, then
press the Enter button.
2.4 Automatic Calibrations
The Stat Prole Prime CCS analyzer performs a 2-point calibration 30 minutes after being powered on and regularly thereafter to maintain optimal MicroSensor Card and air detector performance.
Figure 2.8 Operator ID Screen and Operator Password Screen
A 1-point calibration is performed at regular intervals to monitor MicroSensor Card performance between each 2-point calibration. If a calibration error occurs, an alert is shown to notify the operator and the test button of the
affected analyte is displayed with an orange background
to indicate it is unavailable for testing.
Scheduled 2-point calibrations can be delayed once for 10 minutes by pressing the Cancel button. After 10 minutes the rescheduled calibration begins and cannot be cancelled.
2-8
2 Getting Started
2.4.1 Manual Calibrations
A manually initiated 2-point calibration can be performed whenever the analyzer displays Ready or Not Ready on the header bar.
A Not Ready (Not Calibrated) status is displayed after powering the analyzer on, after replacing some consumable items or as a result of a system error. When the analyzer displays Not Ready, samples cannot be run until a 2-point calibration is performed that successfully calibrates the air detectors and at least one analyte. To initiate a calibration from the Not Ready state, press the Calibrate button
A Ready status indicates the air detectors and one or more analytes are calibrated and ready for analysis. To manually calibrate the analyzer from the Ready state select the Toolbox icon and then select the Calibrate button .
Analytes that display an orange button may be
uncalibrated. Press the orange button and then select the Calibrate button, if displayed, to initiate a 2-point calibration.
CalibrateCalibrate
on the Menu Bar.
2. Started
Figure 2.9 Not Ready Screen and Ready Screen
2.4.2 Air Detector Calibration
The Air Detectors are automatically calibrated once a day. An Air Detector calibration can be initiated manually, if needed, as follows:
> > > .
2-9
3 Sample Analysis
3 Sample Analysis
WARNING: While the probe is extended, do not open
or close the door.
When Ready is displayed on the Home screen the analyzer
is ready to analyze samples for any analyte not displaying an
Orange test button. The analyzer can measure whole blood
samples from capillary tubes, syringes, test tubes, and open containers as well as external Quality Control material from ampules and internal Quality Control material from an internal auto QC cartridge.
3.1 Analyzing Patient Samples
Before running a patient sample verify the analyzer is Ready to perform the analysis and that all the desired analytes are available for selection. If necessary, refer to Chapter 2 for additional information.
INTERFERENCE WARNING: Do not perform glucose
and lactate testing on patients taking the drug hydroxyurea. Refer to section A.3 for
additional interference information.
3.1.1 Analyzing Syringe Samples
From the Home screen, log in if necessary.
1. Select the syringe button from the container drop-down list.
2.
Select the desired Test Panel from the drop-down list or select analytes to create a Custom Panel.
3. Analysis
Figure 3.1 Ready Screen: Container and Panel Drop-down Lists
3-1
Prime CCS Instructions for Use Manual
3. Press the Start button to begin the analysis.
4. If prompted, enter all Required Information and press Start again to begin the analysis.
Figure 3.2
Sample Information Screen
5. Prepare the sample for analysis (mix well) then position the sample over the probe and press the
Aspirate
The sample probe retracts automatically once sufcient sample has been aspirated into the analyzer.
6. Enter any Required or Optional information while the analysis is running.
button.
NOTE:
The Sample Information remains disabled until all Required elds have been entered. The analysis can be cancelled by pressing the X icon but results will not be printed or transmitted.
Figure 3.3 Syringe Sample
3-2
Figure 3.4
Sample Information Screen
3 Sample Analysis
3.1.2 Using the Safety Sample Port
The Safety Sample Port provides a means of attaching a syringe to the analyzer instead of manually positioning the sample probe in the sample.
When using the Safety Sample Port, Nova recommends using the Nova Syringe Clot Catcher to ensure that the sample is positioned correctly for aspiration and to prevent clots from entering the ow path. If a clot catcher is not used, syringes must be lled with sufcient sample for the probe to travel approximately 1-inch (26 mm) into the syringe.
Figure 3.5 Safety Sample Port
From the Home screen:
1. Select the appropriate syringe button
from the container drop-down list.
2. Select the desired Test Panel from the drop-down list or select one or more analytes to create a Custom Panel.
3. Analysis
Figure 3.6 Ready Screen: Container and Panel Drop-down List
3-3
Prime CCS Instructions for Use Manual
3. Prepare the sample for analysis (mix well) then attach the syringe to the Safety Sample Port.
4. Press the Start button to begin the analysis.
5. If prompted, enter all Required Information and press the Start button once more to begin the analysis.
6. Press the Aspirate button to aspirate sample into the analyzer. The sample probe retracts automatically once sufcient sample has been aspirated into the analyzer.
7.
Remove the syringe from the Safety Sample Port.
8. Enter any Required or Optional information while the analysis is running.
NOTE: The
Required elds have been entered. The analysis can be cancelled by pressing the X icon but results will not be printed or transmitted.
Figure 3.7
Sample Information Screen
3-4
Sample Information screen will remain until all
Figure 3.8 Position Sample Screen:
Safety Sample Port
3 Sample Analysis
3.1.3 Analyzing Sample from a Blood Collection Tube
From the Home screen:
1. Select the blood collection tube icon from the container drop-down list.
2. Select the desired Test Panel from the drop­down list or select one or more analytes to create a Custom Panel.
Figure 3.9 Ready Screen: Container and Panel Drop-down List
3. Press the button to begin the analysis.
4. If prompted, enter all Required Information and press Start once more to begin the analysis.
5. Prepare the sample for analysis (mix well) then position the sample over the probe and press Aspirate . The sample probe will retract automatically once sufcient sample has been aspirated into the analyzer.
6. Enter any Required or Optional information while the analysis is running.
3. Analysis
Figure 3.10 Sample Information Screen
3-5
Figure 3.11 Position Sample Screen:
Prime CCS Instructions for Use Manual
NOTE:
The Sample Information screen will remain until all Required elds have been entered. The analysis can be cancelled by pressing the X icon but results will not be printed or transmitted.
3.1.4 Analyzing Sample from a Capillary Tube
From the Home screen:
1. Select the capillary icon from the container drop-down list.
2. Select the desired Test Panel from the drop-down list or select one or more analytes to create a Custom Panel.
3. Press Start to begin the analysis.
4. If prompted, enter all Required Information and press the Start button again to begin the analysis.
5. Prepare the sample for analysis (mix well). Then position the capillary tube into the capillary adaptor and press Aspirate .
6. When prompted, remove the capillary tube and press .
7. Enter any Required or Optional information while the analysis is running.
NOTE: The
3-6
Sample Information screen will remain until all Required elds have been entered. The analysis can be canceled by pressing the X icon but results will not be printed or transmitted.
Figure 3.12 Ready Screen: Container and Panel Drop-down Lists
3 Sample Analysis
Figure 3.13 Sample Information Screen
Figure 3.14 Position Sample Screen:
3.2 The Sample Results Display
Once the sample analysis is complete, results for the selected and calculated analytes display (Figure 3.16).
Each analyte is shown with its measured value, the unit of measure, and a 3-segment bar that provides a
visual indication of the sample concentration: Green for normal results, Orange exceeds normal limits, and
Red exceeds panic limits. Sample results displayed as
-- were determined to be outside the analyzer's analytical
measurement range.
NOTE: Results can be displayed 2 different ways
(setup congurable).
The color bar consists of 3 segments:
1. The rst (left hand) segment indicates the
sample result is lower than the entered normal range.
The segment is displayed with a Orange
background if a sample result is between the low Normal and low Alert range.
The segment is displayed with a Red back-
ground when a sample exceeds the low Alert range.
2. The middle segment indicates the sample result
is within the entered normal range.
The segment is displayed with a Green
background when the sample result is within the entered normal range.
3. Analysis
3-7
Prime CCS Instructions for Use Manual
Figure 3.15
Blood Results Screens: Displayed in 1 Column (left) or 2 Columns
3. The last (right hand) segment indicates the sample result is higher than the entered normal range.
• The segment is Orange when a sample result is
between the high Normal and high Alert range.
• The segment is Red when a sample exceeds the
high Alert range.
Use the Page Up and Page Down buttons to scroll through additional pages of result screens. The number of pages is shown in the upper left corner of the display, e.g., 1 of 3.
Press the Print button to print the results on the analyzer’s thermal printer.
Press the Transmit button to transmit the results to the LIS/HIS system.
Press the Home Button to return to Home screen.
3.2.1 Sample Results Printout
The Sample Results printout contains a customizable header followed by the measured and calculated test
results. Each test result contains the test name, the measured value and the unit of measure. A result
ag consisting of one or more up () or down () arrows may also be displayed. Some event codes prevent a test
3-8
3 Sample Analysis
result from being printed. Should this occur the event code is printed in place of the test result. Results printed
as - - were determined to be outside the analyzer's
measurement range.
Flags
A single up arrow  is printed if the measured value is above the upper end of the normal range.
A double up arrow  is printed if the measured value is above than the upper end of the critical range.
A triple up arrow value is above than the upper end of the measurement range.
A single down arrow  is printed if the measured value is below the lower end of the normal range.
A double down arrow  is printed if the measured value is below the lower end of the critical range.
A triple down arrow measured value is below the lower end of the measurement range.

is printed if the measured

is printed if the
3. Analysis
3.3 AnalyzingQCandProciencySamples
Before running a QC sample, verify the analyzer is Ready to perform the analysis and that all the desired analytes are available for selection. If necessary, refer to Chapter 2 for additional information.
3-9
Prime CCS Instructions for Use Manual
3.3.1 Analyzing Internal QC Samples
From the Home screen:
1.
From the Home Screen, press the QC button .
2. Press the Analyze QC button .
3. Press the Select QC Level button.
Figure 3.16 Quality Control Screens
4. From the drop-down list, select the Internal Control Level to be analyzed.
5. Enter a QC comment if desired.
6. Press Start to begin the analysis.
7. Wait until probe is fully extended.
8. Position the well-mixed sample over the probe. Select the Aspirate button. The sample probe retracts automatically when sufcient sample has been aspirated. Tests are performed and the QC results are displayed on the screen.
9. Once the analysis is complete press the Save button to keep the QC results or press Delete to discard the QC results.
3-10
Figure 3.17 Quality Control Results Screen
3 Sample Analysis
3.3.2 Analyzing External QC Samples
From the Home screen:
1.
From the Home Screen, press the QC button
2. Press the Analyze QC button .
Figure 3.18 Quality Control Screens
3. From the drop-down list, select the External Control Level to be analyzed.
4. Select the lot number to be analyzed.
5. Enter a QC Comment if desired.
6. Press Start to begin the analysis.
7. Wait for the Sample Probe to fully extend.
8. Prepare the sample for analysis (mix well), then position the sample over the probe and press
the Aspirate button. The sample probe will
retract automatically once sufcient sample has been aspirated into the analyzer.
9. Once the analysis is complete, press Save to
keep the QC results or press Delete to discard the QC results.
3. Analysis
Figure 3.19 Position External QC Screen
3-11
Prime CCS Instructions for Use Manual
3.3.3 AnalyzingProciencySamples
1.
From the Home Screen, press the QC button
2. Press the Analyze QC button .
3. From the drop-down list, press Select QC Level to Analyze.
4. From the drop-down list select Prociency.
5. Press Start to begin the analysis.
6. Wait for the Sample Probe to fully extend.
Figure 3.20 Quality Control Screens
7. Prepare the sample for analysis (mix well) then position the sample over the probe and press Aspirate . The sample probe will retract automatically once sufcient sample has been aspirated into the analyzer.
8. Once the analysis is complete press Save to keep the QC results or press Delete to discard them.
3-12
Figure 3.21 Prociency Sample Positioned on Probe
4 Reviewing Patient and QC Data
4 Reviewing Patient and QC Data
Patient and QC Data are stored on the analyzer and can be reviewed at any time. The following section will demonstrate how to nd your data.
4.1 Reviewing Patient Data
To recall patient data, proceed as follows:
1. Press the Recall Results icon on the Menu Bar to display the current date's patient results
2.
Select the patient result to review.(Figure 4.1)
3.
Then press to display results (Figure 4.2).
the selected sample
4. Data
Figure 4.1 Results Screen
Figure 4.2 Patient Result Selected Data
4-1
Prime CCS Instructions for Use Manual
4. If there is more than one page of data, up and down arrow buttons will appear on the footer to display all screens.
5. To view additional sample results, press the "Start Date End Date" button.
6. Press the drop down menu for "Today, Yesterday, Week, Month, Year, or All" for patient data; or select the start and end dates on the screen.
7. Press the back button to display patient results for the selected date range.
8. Select the desired patient result, then press to display
9. To print the data, press the print icon on the footer of the screen.
results.
4-2
Drop Down
Menu
Start and End Date Selection
Figure 4.3 Results Screens
4 Reviewing Patient and QC Data
4.2 Reviewing QC Data
To recall any QC data on the analyzer, proceed as follows:
1. Press the QC button on the Home screen located in the footer of the screen.
2. The Quality Control screen displays: press the "View QC Data" button.
3. The View QC Data screen displays. From this screen, press the Start Date End Date button.
4. Press the drop down menu for "Today, Yesterday, Week, Month, Year, or All" for QC data; or select the start and end dates on the screen.
5. Press the back button to display QC results.
6. A Level button is also available to select a QC level from a drop down list.
7. A Lot button is available for choosing a QC lot.
8. Select a QC data date and press the to display the data screen.
9. To print the data, press the print icon on the footer of the screen.
4. Data
Figure 4.5 QC Results Screen
4-3
Figure 4.4 View QC Data Screen
Prime CCS Instructions for Use Manual
4.3 QC Statistics
To view QC Statistics on the analyzer, proceed as follows:
1. Press the QC button on the Home screen.
The Quality Control Screen displays.
2. Press the QC Statistics button to display the QC Statistics screen.
Figure 4.6 Quality Control Screen
3. On this screen, press the Start Date End Date button.
4. Press the drop down menu for "Today, Yesterday, Week, Month, Year, or All" for patient data; or select the start and end dates on the screen.
5. Press the back button to display all the QC Statistics for these selected dates.
6. To print the QC Statistics, press the print icon on the footer of the screen.
4.4 Levey Jennings Graphs
To generate a Levey Jennings graph on the analyzer, proceed as follows:
1. Press the QC button on the Home screen.
The Quality Control Screen displays.
3. Press the Levey Jennings button to display the Levey Jennings Graph screen.
4-4
4 Reviewing Patient and QC Data
Figure 4.7 Quality Control Screen
4. Select the QC Level with the Level button for the QC data. This displays a list of the levels that have QC data.
5. Select the Test with the Test button. This displays a list of the tests congured for the analyzer.
6. Select the Lot Number with the Lot Number button. This displays a list of the available lots for the QC Level.
7. A graph dispays showing a Y-axis for the test result and an X-axis for each day of a selected level,lot, test, and month.
8. Touch the Right Arrow button to graph the QC data of the next month that is available. Touch the Left Arrow button to graph the QC data of the prior month that is available.
9. To print the Levey Jennings Graph, press the print icon on the footer of the screen.
4. Data
Figure 4.8 Levey Jennings Graph
4-5
5 Consumables Replacement
5 Consumables Replacement
The following sections provide directions to operate and maintain the Stat Prole Prime CCS Analyzer at peak efciency. From the Home screen, press the Tool box icon to display the System Menu screens (Figure 5.1). From these screens, press the appropriate button to replace the following consumables:
MicroSensor Card
Calibrator Cartridge
Auto QC Cartridge
WARNING: Blood samples and blood products are
potential sources of infectious agents. Handle
all blood products, ow path components, and accessories (waste-line, probe, MicroSensor Card, clot catchers, ush tools/xtures, safety sample port, etc.) with care. Gloves and protective clothing are recommended. When performing replacement and troubleshooting procedures, also use protective eyewear.
5. Con Repl
Figure 5.1 System Menu Screens
The Cartridge Required screen (Figure 5.2) displays when a new MicroSensor Card, Calibrator Cartridge, Auto QC Cartridge, or Reference Cartridge has been removed or requires replacement.
The Sensor Card button displays if the
MicroSensor Card is not present, has zero remaining life, or has not completed hydration.
5-1
Prime CCS Instructions for Use Manual
Figure 5.2 Cartridge Required Screen
The Calibrator Cartridge and/or Auto QC Cartridge buttons display if the cartridge is not present or has zero remaining life.
The Reference Sensor button displays if the Reference Sensor is not present.
Touching the button will bring you to the screen for replacing this item.
5.1 Calibrator Cartridge and Auto QC Cartridge Replacement
The Calibrator Cartridge and/or Auto QC Cartridge should be changed when the system indicates the cartridge is empty. From the Home screen, press the Tool Box icon . Then press Replace Calibrator Cartridge or Replace Auto QC, as needed.
Mix the cartridge thoroughly by gentle inversions.
Then follow the directions on the screen to replace the cartridge and the Capillary Adaptor.
WARNING: When the Calibrator Cartridge or Auto QC
Cartridge is removed, keep your ngers and hands away from the back of the cartridge compartment. Sharp needles can cause
injury, and the waste needle is a biohazard.
5-2
5 Consumables Replacement
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
NOTE: The Calibrator Cartridge and the Auto QC
Cartridge must be replaced through the Tool Box screens. If you remove and replace a cartridge (even if it is the same one) outside of these screens, you will not be able to prime the analyzer, and you will not be able to calibrate or to analyze samples (Calibrator Cartridge) or to analyze internal controls (Auto QC Cartridge). If you have removed and replaced a cartridge outside of these screens, go to the appropriate screen and press Prime.
NOTE: The Capillary Adaptor comes in the Calibrator
Cartridge box. It is very important for the proper operation of the analyzer that the Capillary Adaptor be changed with every Calibrator Cartridge change.
5.1.1 Replace the Calibrator Cartridge
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
5. Con Repl
1. Press the Tool Box icon.
2. From the System Menu, press the Replace Calibrator Cartridge button.
Figure 5.3 Remove Calibrator Cartridge and Capillary Adaptor
5-3
Prime CCS Instructions for Use Manual
3. Open the door and remove the cartridge.
4. Slide in a new cartridge past the front retaining lip.
CAUTION: Probe will move when Enter is pressed.
5. To replace the Capillary Adaptor, press the Enter
button. Slide off the used capillary adaptor. Slide on the adaptor provided with Calibrator Cartridge.
6. Close the analyzer door. Press the Prime button.
Calibrator Cartridge status can be viewed at any time by pressing the Cartridge Status indicator icon at upper right of the header bar.
5-4
Figure 5.4 MicroSensor Card Status
The Calibrator status bar is empty (white) when no Calibrator Cartridge is installed.
The status bar is green and indicates the with the remaining use life of the Calibrator Cartridge when the use life remaining percentage is > 5%.
The Calibrator status bar is orange when the remaining use life is ≤ 5%.
The screen displays the lot expiration date, lot number, use life expiration date and time, number of samples remaining, and percentage use life remaining.
5 Consumables Replacement
When no cartridge is installed, all of the above are blank.
Press as appropriate the Install button or Replace button.
5.1.2 Replace the Auto QC Cartridge
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
1. Press the Tool Box icon.
2. From the System Menu, press the Replace Auto QC button.
3. Open the analyzer door. Remove the Auto QC Cartridge, if present.
4 Slide in a new cartridge, past the front
retaining lip.
5. Close the door. Press the Prime button.
5. Con Repl
Figure 5.5 Remove Auto QC Cartridge
Auto QC Cartridge status can be viewed at any time by pressing the Status icon at upper right of the screen.
The Auto QC status bar is empty (white) when the Auto QC Cartridge is not installed.
The Auto QC status bar is green when the percentage remaining use life is > 5%.
5-5
Prime CCS Instructions for Use Manual
Figure 5.6 Auto QC Cartridge Status
The Auto QC bar is orange with the percentage remaining use life of the Auto QC Cartridge if the use life remaining percentage is ≤ 5%.
The screen displays the lot expiration date, lot number, use life expiration date, use life expiration time, number of samples remaining, and percentage use life remaining.
If no pack is installed, all above is blank.
There is either an install button to Install an Auto QC Cartridge or a Replace button to replace an Auto QC Cartridge.
5.2 Replace the MicroSensor Card
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
1. Press the Tool Box icon.
2. From the System Menu, press the Replace Sensor Card button. Wait for the pump to stop.
3. Open the analyzer door. Then open the MicroSensor Card door. Remove the card.
NOTE: Hold the MicroSensor Card as indicated in
Figure 5.7.
5-6
5 Consumables Replacement
Hold by
Edges
Figure 5.7 MicroSensor Card
4 Insert the new card. Close the MicroSensor Card
door.
5. Close the analyzer door. Press the Prime button.
MicroSensor Card status can be viewed at any time by pressing the Status icon on the upper right corner of the screen.
The sensor status bar is empty (white) when no MicroSensor Card is installed.
The bar is green and indicates the remaining use life of the MicroSensor Card if the remaining use life is greater than 5%.
The status remaining use life use life
bar is orange
and indicates the
of the MicroSensor Card if the
remaining is less than or equal to 5%.
5. Con Repl
Figure 5.8 MicroSensor Card Status
5-7
Prime CCS Instructions for Use Manual
The screen displays the lot expiration date, the lot number, the use life expiration date, the use life expiration time, the number of samples remaining, the percentage use life remaining.
If no MicroSensor Card is installed, all above is blank.
There is either an Install button to install a MicroSensor Card or a Replace button to replace the Card.
5.3 Prime MicroSensor Card Warranty Process
If one or more analytes are unable to calibrate within the MicroSensor Card warrantied use life, the analyzer will generate a 16-digit warranty code that can be submitted for credit towards a new MicroSensor card.
The scenarios described in Table 5-1 generate a warranty code.
Warranty code generation depends upon the answers to a series of questions the user is prompted to answer. Following is a description of the screens and choices presented to the user.
1. If a MicroSensor Card fails prior to its stated warranty, the user is presented with the following screen.
5-8
Figure 5.9 Warranty Screen
2. The user must select one of three options:
Option 1: “Claim Sensor Card”. This option
provides the user with a warranty code which will reimburse them for all of the available parameters remaining on the MicroSensor Card to the end of
5 Consumables Replacement
Table 5-1 Failures that Generate Warranty Codes
Failure Mode Action
Hydration Failure – PCO
channel
2
Hydration Failure – All other parameters
PCO
channel failure
2
during normal use
Any other channel in normal use
the warranty period. If this option is selected, the software will permanently deactivate the card.
Option 2: “Claim Sensor”. This option
provides the user with a credit for only the failed parameter but does not disable the remaining parameters. The user is allowed to continue using the remaining parameters on the card for patient testing.
Option 3: Choose to defer the choice
by selecting the cancel button . This only defers the choice until the next time the failure is detected.
The entire card is disabled. A single warranty code is generated. The customer must install a new card.
The sensor is automatically disabled and a warranty code is generated and printed automatically. The customer can continue using the card or disable the remaining channels. To disable the remaining channels and generate a code for the rest of the card, the user must select the disabled sensor – and follow the prompts.
The PCO a warranty code is generated and printed automatically. The customer can continue using the card or disable the remaining channels. To disable the remaining channels and generate a code for the rest of the card, the user must select the white PCO prompts.
The customer is presented with a popup that allows disabling of the sensor or card.
is automatically disabled and
2
icon and follow the
2
5. Con Repl
5-9
Prime CCS Instructions for Use Manual
3. Once the user has made a selection, the appropriate conrmation dialogues are presented.
Figure 5.10 Conrmation Dialogs
4. After conrmation, the Stat Prole Prime analyzer will automatically print and save the 16-digit code. The code is stored in the analyzer for future retrieval and is not lost.
5. The warranty codes can be retrieved at a later time by pressing Menu > Service > Credits. The codes are searchable by date range or printed status. Note that the “Printed” status (right hand column) refers to a reprint initiated from this screen. It does not refer to the automatic printing that occurs. This allows for easy identication of any unclaimed codes.
6. Once the code is generated, the user must provide it to customer support for a warranty credit. Customers in the USA should contact Nova Technical Support at 800-545–6682, In Canada contact Nova Technical Support at 1-800-263-5999. Outside the USA or Canada contact your local distributor.
5-10
Figure 5.11 Warranty Credits Screen
6 Periodic Replacements
6 Periodic Replacements
Periodically the Pump Tubing Harness, Reference Cartridge replaced. This section gives procedures on replacements of these consumable Items.
6.1 Pump Tubing Harness Replacement
The Pump Tubing Harness should be replaced at intervals prescribed in the maintenance log. Replace the tubing that goes around the pump as follows.
WARNING: Exposure to Blood Borne Pathogens.
, Sample Probe, or printer paper may need to be
Follow laboratory procedures.
Waste (W)
Line
Pressure
Plate
Tab of Tubing
Clip
Pump Tubing
Manifold; to
Reference
Cartridge
Pressure Plate
Release Button
Figure 6.1 Pump Tubing
Reference (R)
Line
Tubing Clip
6. Periodic
6-1
Prime CCS Instructions for Use Manual
Figure 6.2 Replace and Install Pump Tubing Screens
1. From the Home screen, press the Tool Box icon.
2. From the System Menu, select Replace Pump Tubing. Wait for the pump to stop.
3. Open the analyzer door. Then open the MicroSensor Card door. Remove the MicroSensor Card, if present.
4. Push the white button to release the Pump Tubing Pressure Plate.
6-2
Figure 6.3 Release Pump Tubing Pressure Plate
5. Disconnect the W and R tubes from the analyzer.
6. Disconnect the tubing connector from the Reference Cartridge.
7. Remove the tubing harness and discard.
8. Press the Enter button to continue.
Figure 6.4
Disconnect Pump Manifold
6 Periodic Replacements
Figure 6.5 Disonnect Tubing Connector from Reference Cartridge
6.1.1 Install the Pump Tubing
1. Install the tubing over the pump rollers.
2. Slide the tubing bracket under the locating tabs.
3. Connect the tubing connector to the Reference Cartridge.
4. Connect the W and R tubes to the analyzer.
5. Close and latch the Pump Tubing Pressure Plate.
6. Insert the MicroSensor Card.
7. Close the MicroSensor Card door.
8. Close the analyzer door.
9. Press the Prime button to continue.
6. Periodic
6-3
Prime CCS Instructions for Use Manual
6.2 Probe Replacement
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
If the Probe or Sample/Air Detector line become damaged, replace the assembly. Use the following procedure:
1.
From the Home screen, press the Tool Box icon. From the System Menu select Replace S-Line Probe and wait for the pump to stop.
2. Remove the capillary adaptor from the front of the probe by gently pulling (Figure 6.6).
3. Disconnect the Air Detector Tube (Figure 6.7).
4. Disconnect the Air Detector Sample Line from the Reference Cartridge module using the removal
Air Detector Tube
Connector
Figure 6.6
Remove Capillary Adaptor
6-4
Figure 6.7 Disconnect
Air Detector Tube
6 Periodic Replacements
Removal Tool
Sample (S)-Line (connects to
Reference Cartridge)
Figure 6.8 Remove the S-line from the Reference Cartridge
tool (Figure 6.8).
5. Squeeze the white pinch clamp (Figure 6.9). Remove the Probe, S-line, and Air Detector tube
Pinch
Clamp
S-Line (clear)
with Connection
to Reference
Cartridge
Air Detector
Line (gray) to
Analyzer
Figure 6.9 Remove Probe, S-line, and Air Detector Line
and discard.
6. Insert the new Probe assembly until it clicks into place.
7. Install the Capillary Adaptor onto the probe.
8. Reconnect the S-line to the Reference Cartridge
9. Reconnect the Air Detector line into the analyzer.
10. Close the door.
11. Press the Calibrate button.
6-5
6. Periodic
Prime CCS Instructions for Use Manual
6.3 Reference Cartridge Replacement
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
1. From the Home screen, press the Tool Box icon.
2. From the System Menu, select the Replace Reference button.
3. Open the door. Open the MicroSensor Card door.
4 Remove the MicroSensor Card, if present.
5. Disconnect the Sample (S)-line from the bottom of the Reference Cartridge (Figure 6.10).
Pump Tubing
Manifold
Reference
Cartridge
MicroSensor
Card
6-6
S-Line
Figure 6.10 Reference Cartridge
6. Disconnect the Pump Tubing from the top of the Reference Cartridge.
7. Slide the Reference Cartridge right to remove (Figure
6.11)..
8. Press the Enter button to continue.
6 Periodic Replacements
Figure 6.11 Slide the Reference Cartridge Right to Remove
6.3.1 Install the Reference Cartridge
1. Position the new Reference Cartridge to the right of its installed position.
2. Slide the Reference Cartridge left and into place.
3. Attach the S-Line to the bottom and Pump Tubing to the top.
4. Insert the MicroSensor Card.
5. Close the MicroSensor Card door.
6. Close the analyzer door.
7. Press the Prime button to continue.
6. Periodic
6-7
Prime CCS Instructions for Use Manual
6.4 Printer Paper Replacement
1. Open the printer cover.
2. Remove the depleted roll of paper.
3. Insert a new roll of paper. The loose end of the paper should feed from the bottom of the roll.
4. Feed paper past the cover. Then close the printer cover.
6-8
Figure 6.12 Replace the Printer Paper
6 Periodic Replacements
6.5 Safety Sample Port Replacement
WARNING: Exposure to Blood Borne Pathogens.
Follow laboratory procedures.
1. Open the door.
2. Slide out the old Safety Sample Port.
3. Slide in a new Safety Sample Port.
4. Close the door.
Squeeze Pinch Clamp to Remove Safety Sample
Port
Figure 6.13 Replace the Safety Sample Port
6-9
6. Periodic
Prime CCS Instructions for Use Manual
6-10
7 Troubleshooting
7 Troubleshooting
This section describes the recommended troubleshooting procedures for use with the Stat Prole Prime CCS analyzer. The procedures use the most logical and direct steps to resolve each problem and are written to minimize the replacement of any unnecessary parts. If the recommended solutions do not resolve the problem please contact Nova Biomedical Technical Support for troubleshooting assistance.
FOR TECHNICAL ASSISTANCE, CALL TOLL FREE:
USA 1-800-545-NOVA Canada 1-800-263-5999 Other Countries Contact the local Nova
Biomedical Sales Ofce or Authorized Nova Biomedical Distributor
WARNING: Blood samples and blood products
are potential sources of infectious agents. Handle all blood products, accessories,
and ow path components (waste-line,
capillary adaptor, probe, sensor cartridge,
etc.) with care. Gloves and protective clothing are recommended. When performing maintenance and troubleshooting procedures, also use protective eyewear.
7. Trshoot.
7-1
Prime CCS Instructions for Use Manual
7.1 Event Log
The Event Log displays a list of events that have occurred during a selected time frame. To access the Event Log: from the Home Screen, press:
> > >
The Event Log initially displays events that occurred on the current date but may be changed to show events that occurred during a specied time frame or that contain a specic Event ID. Events are displayed chronologically, with the most recent event at the top of the page.
Each event is shown with the date and time the event occurred, the event ID and a description of the event. To view additional details of a specic event, select (highlight) the event of interest then press the Details button . To print the Event Log press .
Figure 7.1 Event Screen
7.2 Resolving Event Codes
Event Codes are grouped into one of 5 categories, Cartridge errors, Flow errors, Printer errors, MicroSensor Card errors, and Hardware/Software errors. Use the following troubleshooting steps to resolve the listed codes.
If a displayed code is not listed, please contact Nova Biomedical Technical Support for assistance.
7-2
7 Troubleshooting
7.2.1 Flow Event Codes
Event Code
Description and Corrective Action
602 Insufcient Sample – During the last sample
analysis, the leading edge of the sample was not detected at the reference air detector when expected.
Recommended Solution
1. Rerun the sample and insure the sample probe is not being obstructed by the sample container.
2. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
3. Replace the W/R Pump Harness.
4. Contact Nova Biomedical Technical Support.
604 No Flush When Required – During the last
calibration or sample analysis, Flush solution was not detected when expected.
Recommended Solution
1. Verify the % remaining in the Calibrator Cartridge. If the pack indicates less than 10% remaining, replace the Calibrator Cartridge.
2. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
3. Calibrate the Air Detector.
4. Contact Nova Biomedical Technical Support.
7. Trshoot.
7-3
Prime CCS Instructions for Use Manual
605 No Air When Required - During the last
calibration or sample analysis, Air was not detected when expected.
Recommended Solution
1. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Prime Flow Path.
2. Calibrate the Air Detector.
3. Contact Nova Biomedical Technical Support.
606 No Standard A When Required - During the
last calibration or sample analysis, calibrator standard A solution was not detected when expected.
Recommended Solution
1. Verify the % remaining in the Calibrator Cartridge. If the pack indicates less than 10% remaining, replace the Calibrator Cartridge.
2. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
3. Calibrate the Air Detector.
4. Contact Nova Biomedical Technical Support.
607 No Standard B When Required - During the
last calibration, calibrator standard B solution was not detected when expected.
Recommended Solution
1. Verify the % remaining in the Calibrator Cartridge. If the pack indicates less than 10% remaining, replace the Calibrator Cartridge.
2. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
3. Calibrate the Air Detector.
4. Contact Nova Biomedical Technical Support.
7-4
7 Troubleshooting
608 No Internal QC When Required – During the
last Internal QC analysis, QC solution was not detected when expected.
Recommended Solution
1. Verify the % remaining in the Internal QC Cartridge. If the pack indicates less than 10% remaining, replace the QC Cartridge.
2. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
3. Calibrate the Air Detector.
4. Contact Nova Biomedical Technical Support.
609 No Sample When Required – During the last
sample analysis, no sample was detected by the analyzer.
Recommended Solutions
1. If analyzing a whole blood sample, verify that the sample is not clotted. If the sample is clotted, Nova Biomedical recommends that the sample be redrawn or that a clot catcher be utilized prior to repeating the analysis.
2. If analyzing an external QC sample, repeat the analysis. If the problem recurs, proceed to step
3.
3. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
4. Calibrate the Air Detector.
5. Contact Nova Biomedical Technical Support.
7. Trshoot.
7-5
Prime CCS Instructions for Use Manual
611 Sample Position Not Conrmed – During the
last analysis, the sample was not detected at one of the air detectors.
Recommended Solutions
1. If analyzing a whole blood sample, verify that the sample is not clotted. If the sample is clotted, Nova Biomedical recommends that the sample be redrawn or that a clot catcher be utilized prior to repeating the analysis.
2. If analyzing an external QC sample, repeat the analysis. If the problem recurs, proceed to step
3.
3. Flush the ow path with deionized water and recalibrate the analyzer. Refer to Section 7.3, Flushing the Flow Path.
4. Contact Nova Biomedical Technical Support.
7-6
7 Troubleshooting
7.2.2 Printer Event Codes
Event Code
Description and Corrective Action
904 Printer Paper Out – No paper was detected in
the thermal printer.
Recommended Solution
1. Check and or replace printer paper supply.
2. Contact Nova Biomedical Technical Support if
printer paper is not recognized.
905 Printer Cover Open – The printer cover is not
completely closed.
Recommended Solution
1. Insure the printer cover is fully closed.
2. Contact Nova Biomedical Technical Support if
unable to resolve.
906 Printer – A printer error has occurred.
Recommended Solution
1. Check the printer paper supply. Clear any paper
jam that may have occurred.
2. Contact Nova Biomedical Technical Support if
unable to resolve.
7. Trshoot.
7-7
Prime CCS Instructions for Use Manual
7.2.3 MicroSensor Card Event Codes
Event Code
1201 1209 1217 1225 1233 1241 1249 1257 1265 1289 1297
Description and Corrective Action
Slope pH Slope
PO2 Slope PCO2 Slope
K Slope Na Slope
The measured difference between the indicated analytes’ calibration standards did not meet the minimum specications for a properly performing sensor during the last 2-point calibration.
Cl Slope Ca Slope Hct Slope Glucose Slope Lactate Slope
Recommended Solution
1. Recalibrate the analyzer.
2. Flush the ow path and recalibrate the analyzer. Refer to Section 7.3.2, Flushing the Flow Path.
3. Replace the MicroSensor Card.
4. Replace the Calibrator Cartridge.
5. Replace the Reference Cartridge.
6. Call Nova Biomedical Technical Support.
7-8
7 Troubleshooting
1202 1210 1218 1226 1234 1242 1250 1258 1266 1290 1298
Overload pH Overload
PO2 Overload PCO2 Overload
During the last calibration or analysis sequence, the indicated analytes’ sensor reading exceeded the software limits.
K Overload Na Overload Cl Overload Ca Overload Hct Overload Glu Overload Lac Overload
Recommended Solution
1. Recalibrate the analyzer.
2. Flush the ow path and recalibrate the analyzer. Refer to Section 7.3.2 Flushing the Flow Path.
3. Replace the MicroSensor Card.
4. Call Nova Biomedical Technical Support.
7-9
7. Trshoot.
Prime CCS Instructions for Use Manual
1205 1213 1221 1229 1237 1245 1253 1261 1269 1293 1301
Stability pH Stability
PO2 Stability PCO2 Stability
During the last calibration or analysis sequence, the indicated analytes’ sensor did not reach a stable endpoint.
K Stability Na Stability Cl Stability Ca Stability Hct Stability Glu Stability Lac Stability
Recommended Solution
1. Recalibrate the analyzer.
2. Flush the ow path and recalibrate the analyzer. Refer to Section 7.3.2 Flushing the Flow Path.
3. Replace the MicroSensor Card.
4. Replace the Reference Cartridge.
5. Call Nova Biomedical Technical Support.
7-10
7 Troubleshooting
1206 1214 1222 1230 1238 1246 1254 1262 1270 1294 1302
E-Zero Drift pH E-Zero Drift
PO2 E-Zero Drift P
CO2 E-Zero Drift
K E-Zero Drift
During the last analysis sequence, the indicated analytes’ performance changed signicantly since the last successful 2-point calibration.
Na E-Zero Drift Cl E-Zero Drift Ca E-Zero Drift Hct E-Zero Drift Glu E-Zero Drift Lac E-Zero Drift
Recommended Solution
1. Recalibrate the analyzer.
2. If the problem persists, ush the ow path and recalibrate the analyzer. Refer to Section 7.3.2, Flushing the Flow Path.
3. Replace the MicroSensor Card.
4. Call Nova Biomedical Technical Support.
7-11
7. Trshoot.
Prime CCS Instructions for Use Manual
1207 1215 1223 1231 1239 1247 1255 1263 1271 1295 1303
A to A Drift pH A to A Drift
During the last analysis sequence, the indicated analytes’ performance
PO2 A to A Drift PCO2 A to A
changed signicantly from the previous analysis.
Drift K A to A Drift Na A to A Drift Cl A to A Drift Ca A to A Drift Hct A to A Drift Glu A to A Drift Lac A to A Drift
Recommended Solution
1. Recalibrate the analyzer.
2. If the problem persists, ush the ow path and
recalibrate the analyzer. Refer to Section 7.3.2
Flushing the Flow Path.
3. Replace the MicroSensor Card.
4. Call Nova Biomedical Technical Support.
7-12
7 Troubleshooting
1208 1216 1224 1232 1240 1248 1256 1264 1272 1296 1304
Slope Drift pH Slope Drift
PO2 Slope Drift P
CO2 Slope
Drift K Slope Drift Na Slope Drift Cl Slope Drift Ca Slope Drift Hct Slope Drift Glu Slope Drift Lac Slope Drift
Recommended Solution
1. Recalibrate the analyzer.
2. If the problem persists, ush the ow path
and recalibrate the analyzer.
3. Replace the MicroSensor Card.
4. Call Nova Biomedical Technical Support.
During the last calibration, the indicated analytes’ performance changed signicantly from the previous calibration.
7-13
7. Trshoot.
Prime CCS Instructions for Use Manual
7.3 Troubleshooting Flow Problems
The analyzer may experience ow problems as a result of aspirating clots from poorly heparinized whole blood samples. If this should occur operators can use the following procedures to clear the analyzer ow path and verify the analyzer is capable of aspirating from the sample probe.
NOTE: Nova Biomedical recommends but does not
require the use of clot catchers as an effective means of preventing the aspiration of clots into the analyzer’s ow path.
7.3.1 The Flow Path Flush Tool
The Flow Path Flush Tool (Figure 7.2) consists of a 30 mL syringe and a special adaptor to safely clear the ow path in the event it becomes obstructed.
The tool can be used with or without the adaptor to ush individual ow path components. The adaptor is keyed to connect to the Reference Cartridge in only one direction (Figure 7.3). The tubing connected to the adaptor should point up to ensure it is not pinched shut when the MicroSensor Card door is closed.
7-14
Figure 7.2 Flow Path Flush Tool
7 Troubleshooting
Figure 7.3 Tubing Connected to the Adaptor Points Up
7.3.2 Flushing the Flow Path
The analyzer’s ow path can be ushed to remove clots or other debris that may have been aspirated into the system. Use of a device other than the Flush Tool may damage the MicroSensor Card and is not recommended.
Following is the recommended procedure for clearing the ow path when needed.
1. From the Home screen, press >
2. When the pump stops turning, open the analyzer door then open the MicroSensor Card door.
3. Aspirate a small amount of deionized water into the Flow Path Flush Tool.
4. Disconnect the W/R Pump Harness from the Reference Cartridge.
Figure 7.4 Flush Flow Path Screens
7-15
7. Trshoot.
Prime CCS Instructions for Use Manual
5. Connect the Flush Tool to the Reference Cartridge and close the MicroSensor Card door.
6. Using moderate pressure, carefully ush the ow path with 1-2 ml of deionized water into a container placed in front of the sample probe.
7. Rell the Flush Tool with air and ush the ow path to remove any remaining liquid.
8. Disconnect the Flushing Tool and reattach the W/R tubing harness to the Reference Cartridge.
9. Close the MicroSensor Card door; press Done.
10. Recalibrate the analyzer 3 times.
7-16
Figure 7.5
Flushing the Flow Path into Gauze
7 Troubleshooting
7.3.3 Flushing the Sample Probe/S-Line
The Sample Probe is designed to prevent large clots and other debris from advancing into the analyzer’s ow path if possible. As a result, the sample probe may become obstructed and require manual ushing to remove an obstruction. Following is the recommended procedure for clearing obstructions from the sample probe and S-line.
1. From the Home screen, press >
2. When the pump stops turning, open the analyzer door.
3. Draw a small amount of deionized water into the Flow Path Flush Tool.
4. Carefully disconnect the Sample probe/S­line tubing from the bottom of the Reference Cartridge using the removal tool.
5. Slide the ush tool tubing over the extended sample probe.
6. Using moderate pressure, ush water through the sample probe and S-line into a container or gauze placed in front of the S-line.
7. Rell the Flush Tool with air and ush the sample probe to remove any remaining liquid.
8. Press Done when nished.
9. Recalibrate the analyzer as needed.
Figure 7.6 Flushing the S-Line
7-17
7. Trshoot.
Prime CCS Instructions for Use Manual
7.3.4 Flow Test
The following procedure can be used to verify the analyzer is able to aspirate from the sample probe. Use this procedure in tandem with the ushing procedures to clear obstructions and verify ow.
1. From the Home Screen, press
2. Once the pump stops and the sample probe has been extended, press to scroll to page 2.
3. Fill a small container with de-ionized water and immerse the sample probe into the water.
4. Press on the display to start the peristaltic pump.
The water level in the container should drop quickly if the analyzer is aspirating correctly.
5. Press Done when nished.
6. Recalibrate the analyzer as needed.
>
7-18
Figure 7.7 Flow Test
7 Troubleshooting
7.4 Long Term Shutdown
The Prime ABG analyzer can be safely shut down for several hours without taking any special precautions. However, if the analyzer is to be shut down for a longer period, the following shutdown procedure is strongly recommended.
1. Remove the Calibrator Cartridge and Internal QC Cartridge from the analyzer.
2. Insert the Calibrator Flush Fixture into the Calibrator Cartridge bay. Insert the Control Flush Fixture into the Internal QC Cartridge bay.
3. Place the Waste (W) line into an empty container. Place the remaining tubing lines into a container of deionized water.
4. From the Home screen, press Toolbox > Service > Diagnostics.
5. Press the Pump Speed button. Select Very Fast.
6. Press the Rotary Valve and select Flush.
7. Press Run Pump. Allow the pump to run long enough to ensure that water has passed through the entire owpath and into the waste container.
8. Repeat Steps 6 and 7, selecting Standard A, Standard B, and OBC (On Board Control) 1, 2 and 3.
9. Remove the tubing lines from the container of deionized water. Repeat Steps 6, 7, and 8 until all water has been ushed from the system.
10. Remove the Flush Fixtures. Remove the pump tubing harness from the pump. Power off the analyzer.
7. Trshoot.
7-19
Appendix A
A Appendix
Appendix A includes analyzer specications, performance data, solutions and reagents, consumables lists, reference information, and warranty terms for the Stat Prole Prime CCS Analyzer.
A.1Specications
Analyte Measurement Ranges
Analyte
pH
PCO2
PO2 5.0 - 765.0 mmHg 0.66 - 102.00 kPa
Hct 12% - 70%
Na+ 80.0 - 200.0 mmol/L
K+ 1.0 - 20.0 mmol/L
Cl- 50.0 - 200.0 mmol/L iCa 0.20 - 2.70 mmol/L 0.8 - 10.8 mg/dL Glu 15 - 500 mg/dL 0.8 - 28.0 mmol/L
Lac 0.4 - 20.0 mmol/L 4 - 178 mg/dL
Default Alternate
6.500 - 8.000 (pH units) 316.2 - 10.0 nmol/L (H+ units)
3.0 - 200.0 mmHg 0.40 - 26.70 kPa
Units of Measure
Appendix A
A-1
Prime CCS Instructions for Use Manual
Calculated Result Resolution
Parameter
Units of Measure
Default Alternate
-
HCO
TCO
3
2
0.1 mmol/L
0.1 mmol/L
nCa 0.1 mmol/L
BE-ecf 0.1 mmol/L
BE-B 0.1 mmol/L
SBC 0.1 mmol/L
O2Ct 0.1 mL/dL 1 mL/L 0.1 mmol/L
P50 0.1 mmHg 0.1 kPa
O2Cap 0.1 mL/dL 1 mL/L 0.1 mmol/L
SO2% 0.1
Hb 0.1 g/dL 0.1 mmol/L 1 g/L
Anion Gap 0.1 mmol/L
With Entered FiO
2
A 0.1 mmHg 0.01 kPa
AaDO
2
0.1 mmHg 0.01 kPa
a/A 0.1
PO2/FiO
2
0.1 mmHg 0.01 kPa
RI 0.1
Acceptable Sample Sample Volume,
Capillary or Syringe
Barometer
A-2
Sample
Whole Blood (heparinized) 100 µL
400 – 800 ±1 mmHg, accurate to 1.5%
Appendix A
A.2 Quality Control and Calibration A.2.1 Traceability of Calibrators, Controls, and Standards
Chemistry analytes are traceable to the Standard
Reference Materials of the National Institute of Standards
and Technology (NIST). SO
A.2.2 Quality Control
Healthcare facilities should follow federal, state, and
local guidelines for testing quality control materials.
At a minimum, Nova Biomedical recommends that
each laboratory performs the following minimum QC
procedures (Auto-Cartridge QC or External Ampule QC)
on each analyzer:
During each 8 hours of testing, analyze one level of Control.
Analyze all 3 levels during each day of operation.
After performing system maintenance, follow good laboratory practice guidelines for performing quality control analysis.
CAUTION: Sensor performance may be affected
by use of controls other than Stat Prole Prime Controls. Contact Nova Biomedical for
additional information.
When a new lot number of Auto-Cartridge QC is installed, the previous lot number becomes inactive. Thus, you are unable to run lots in parallel to validate the new lot to the old by alternating packs on the same unit.
Nova Recommendation: All Nova controls ship with a product
insert sheet. This product insert sheet contains the target value ranges for each level of QC contained in the pack. Nova’s recommendation for conversion to a new lot number is to use the product insert sheet range levels for the rst 30 days or until sufcient data is collected to establish the new target values. After sufcient data is collected, the established values and ranges can be entered into the analyzer.
is traceable to tonometry.
2
Appendix A
A-3
Prime CCS Instructions for Use Manual
Alternate Method: If this method is inadequate, Nova
recommends the use of the external controls run in parallel and overlapping with the on-board product change over. This method offers continuity in monitoring performance during the change over period. The external QC monitoring can be done using the QC program on the analyzer.
A.2.3 Calibrator Cartridge
In addition to the calibrators and solutions, the Calibrator Cartridge has a self-contained waste bag for safe disposal of waste. For information on Automatic and Manual Calibrations, see Section 2.4
A.3AnalyticalSpecicity,HumanWholeBlood
An interference study was performed according to CLSI guideline EP7-A2. The study used spiked and diluted human specimens containing potential interfering
substances for pH, PO2, PCO
and lactate at normal physiological levels. Each sample containing the interfering substance was evaluated against a reference specimen without the interfering substance.
Potential interfering substances were selected for test based upon a known potential to interfere with the test methodology. The following table represents substances that were tested without demonstrating a clinically signicant effect on test results:
, Na, K, iCa, Cl, glucose
2
A-4
Appendix A
Interfering
Substance
Highest
Concentration
Tested
Analytes Tested
Acetaminophen 20.0 mg/dL Glu, Lac Acetoacetate 2.0 mmol/L
pH, Na, K, iCa, Cl, Glu, Lac
Acetylsalicylic acid 3.62 mmol/L Na, K, Cl, Glu, Lac Albumin 15 g/dL Hct
Ammonium Chloride 107.0 µmol/L Na, K, Cl, iCa, Glu, Lac Ascorbic Acid 50 mg/dL Cl, Glu, Lac
Benzylkonium Chloride BetaHydroxybutyrate Bilirubin
10.0 mg/L
2.0 mmol.L Glu, Lac
20.0 mg/dL
Calcium Chloride 2.0 mmol/L pH,PCO
pH, Na, K, Cl, iCa, Glu, Lac
Hct, pH, PCO
, PO2,
2
Na,K,Cl,iCa,Glu,Lac
,PO2,Na,K
2
D-Galactose 1.0 mmol/L Glu, Lac Dobutamine 2.0 mg/dL pH,Na,K,iCa,Glu, Lac Dopamine Hydrochloride
5.87 µmol/L Glu, Lac
EDTA 3.4 umol/L Glu, Lac Ethanol 86.8 mmol/L Glu,Lac,pH,PCO Fluorescein 1.0 mmol/L PCO2, PO
, PO
2
2
Fluoride 105 µmol/L Glu, Lac Glucose 1,000 mg/dL Lac Glycolic Acid 1 mmol/L Glu, Lac Glucosamine 30.0 µmol/L Glu, Lac
Hemoglobin 2.0 g/L
Hct, pH, PCO Na,K,Cl iCa, Glu, Lac
, PO2,
2
Heparin 100 IU/mL Glu, Lac, Hct Ibuprofen 2.4 mmol/L Na, K, iCa, Cl, Glu, Lac
2
Appendix A
Continued on next page
A-5
Prime CCS Instructions for Use Manual
Continued from previous page
Interfering
Substance
Intralipid 10.0 mg/mL Lithium Lactate 6.6 mmol/L Na, K, iCa, Glu
Magnesium Chloride 15.0 mmol/L Na, Cl Maltose 13.0 mmol/L Glu, Lac Mannose 1.0 mmol/L Glu, Lac Perchlorate 1.0 mmol/L iCa Potassium Chloride 5.0 mmol/L pH, PCO Pyruvate 309 µmol/L Glu, Lac Salicylic Acid 4.34 mmol/L Na, K, Cl, Glu, Lac Sodium Bromide 37.5 mmol/L pH, K, iCa, Lac Sodium Chloride 10.0 mmol/L pH, PCO2, PO2, iCa Sodium Citrate 12.0 mmol/L Cl, Glu, Lac
Sodium Oxalate 500 mg/dL Cl, Glu, Lac Thiocyanate 6.8 mmol/L Cl, Glu, Lac
Urea 40.0 mg/dL Lac Uric Acid 1.4 mmol/L Lac Xylose 25.0 mg/dL Glu, Lac Zinc Chloride 1.3 mg/dL Na, K, iCa
Highest
Concentration
Tested
Analytes Tested
Hct, pH, PCO Na,K,Cl iCa, Glu, Lac
, PO2,
2
, PO2, iCa
2
A-6
Appendix A
The following table represents substances that were tested that demonstrated a clinically signicant effect on test results.
Analyte
Chloride
Ionized Calcium
Glucose
Lactate
Hct
Interfering Substance
Bromide
Thiocyanate
MgCl
2
Hydroxyurea 0.2 mg/dL No interference observed
Oxalate
Thiocyanate
Glycolic acid 0.0 mmol/L No interference observed
Hydroxyurea 0.0 mg/dL No interference observed
Albumin
Hemolysis
Triglycerides
White Blood Count (WBC)
Concentration Interference
2.5 mmol/L No interference observed
5.0 mmol/L Bias of 12.7%
3.4 mmol/L No interference observed
5.1 mmol/L Bias of 15.2%
3.75 mmol/L No interference observed
7.50 mmol/L Bias of 13.5%
0.4 mg/dL Bias of 19.2%
125 mg/dL No interference observed
250 mg/dL Bias of -10.9%
1.7 mmol/L No interference observed
3.4 mmol/L Bias of 10.0%
0.25 mmol/L Bias of 11.7%
0.2 mg/dL Bias of 20.1%
2.8 g/dL No interference observed
5.7 g/dL Bias of 12.7%
5% No interference observed
10% Bias of -10.7%
986.4 mg/dL No interference observed
1233 mg/dL Bias of 12.9%
>50,000 WBC/µL May increase the Hct Value
Appendix A
NOTE: Elevated White Blood Count (WBC) >50,000
WBC/μL may increase the Hct Value.
A-7
18
Prime CCS Instructions for Use Manual
A.4 Healthcare Professional Analytical Performance
Studies
Three Stat Prole Prime CCS analyzers were compared to 2 Stat Prole pHOx Ultra Analyzers in a laboratory setting by healthcare professionals. The protocol consisted of within run precision runs, day­to-day precision runs, linearity validation, and method comparison studies comparing the performance of the Stat Prole Prime CCS Analyzers to the Stat Prole pHOx Ultra Analyzers.
A.4.1 Method Comparison Study
Lithium Heparin arterial whole blood discarded specimens from hospital patients were analyzed in duplicate on the 3 Stat Prole Prime CCS Analyzers and 2 Stat Prole pHOx Ultra reference analyzers. The number of samples per run and the total number of runs each day depended upon the availability of blood specimens on any given test day. Some additional whole blood specimens from consenting donors were tonometered, spiked, or diluted with saline to cover the analytical measurement range for all analytes. The number of data points (N) varies for each parameter due to error, instrument calibration status, or insufcient sample volume.
A minimum of 150 whole blood specimens were analyzed for each parameter in syringe collection devices. The samples were analyzed on each of the Stat Prole Prime CCS analyzers and on each of the pHOx Ultra analyzers. The Stat Prole Prime CCS results for each analyzer were compared to the average of the 2 results from the pHOx Ultra comparative method.
A minimum of 100 whole blood specimens were analyzed for each parameter in capillary collection tubes. Each specimen was analyzed one time from a capillary container on each Stat Prole Prime CCS analyzer and then immediately run as a syringe specimen on the same Stat Prole Prime CCS analyzer. The capillary test result was compared to the syringe result from each test system.
A-8
Appendix A
Bias Chart Results
The method comparison bias estimate was analyzed using CLSI Standard EP09-A2 as a reference document. The bias plots for each parameter are summarized and include boundary lines that represent the 95% condence interval across the measurement range based upon each parameter’s between analyzer day-to-day (+/-2SD) performance specication or CV% (whichever is greater). Each bias plot represents 3 Stat Prole Prime CCS analyzers compared to the average result from 2 Stat Prole pHOx Ultra analyzers. Medically relevant low and high concentrations are annotated.
A-9
Appendix A
Prime CCS Instructions for Use Manual
A-10
Appendix A
A-11
Appendix A
Prime CCS Instructions for Use Manual
A-12
Appendix A
iCa Bias Plot
A-13
Appendix A
Prime CCS Instructions for Use Manual
Syringe Method Comparison Study Results
vs. Stat Prole pHOx Ultra
Test
Parameter
pH
PCO
2
PO
2
Hct
Na
K
iCa
Cl
Glu
Lac
Analyzer
#1 172 40 #2 170 38 #3 168 41 #1 179 34 3.4 - 200.0 0.9854 0.9344 0.9977 #2 181 29 3.1 - 192.6 1.0091 0.1547 0.9920 #3 176 32 3.3 - 199.10 1.0019 1.1679 0.9980 #1 177 43 26.9 - 586.3 1.0046 -1.2710 0.9986 #2 167 43 29.5 - 593.2 0.9897 1.4508 0.9988 #3 180 42 31.3 - 587.6 1.0035 0.5961 0.9990 #1 174 22 12 - 70 1.0445 -1.9271 0.9889 #2 170 24 12 - 68 1.0007 -0.6236 0.9871 #3 164 19 13 - 70 1.0207 -0.9936 0.9895 #1 180 30 85.5 - 195.7 1.0189 -2.2841 0.9955 #2 186 29 85.1 - 196.7 1.0109 -2.0438 0.9960 #3 181 31 85.0 - 198.2 1.0278 -3.5873 0.9961 #1 179 26 1.11 - 19.75 1.0163 -0.0371 0.9996 #2 182 25 1.11 - 19.54 1.0138 -0.0619 0.9995 #3 183 25 1.12 - 19.79 1.0272 -0.0769 0.9995 #1 181 25 0.25 - 2.48 0.9880 0.0457 0.9974 #2 180 25 0.25 - 2.42 0.9752 0.0432 0.9958 #3 179 25 0.25 - 2.52 1.0059 0.0345 0.9962 #1 186 39 52.8 - 189.3 1.0003 1.0158 0.9955 #2 183 40 54.1 - 190.6 0.9952 0.6980 0.9787 #3 180 37 51.0 - 179.5 0.9569 4.4537 0.9944 #1 181 22 35 - 432 0.9987 1.1978 0.9960 #2 184 22 39 - 466 1.0005 0.6501 0.9940 #3 185 24 39 - 474 1.0007 -2.6844 0.9892 #1 182 25 0.4 - 17.8 0.9841 0.0937 0.9974 #2 182 26 0.5 - 20.0 1.0463 -0.0577 0.9959 #3 182 26 0.4 - 18.7 1.0101 -0.0342 0.9946
Total
specimens
Altered
specimens
Specimen
range
6.523 - 7.862
6.519 - 7.875
6.520 - 7.953
Slope
Intercept
0.9976 0.0099 0.9985
0.9977 0.0106 0.9990
1.0018 -0.0225 0.9989
r
A-14
Appendix A
Syringe Method Comparison Study Results
vs. Capillary Results
Individual Analyzer Performance Data
Capillary vs. Syringe Comparison
Parameter Analyzer
pH
pH units
PCO
mmHg
PO
mmHg
Hct
%
Na
mmol/L
K
mmol/L
iCa
mmol/L
Cl
mmol/L
Glu
mg/dL
Lac
mmol/L
#1 100 6.787 - 7.683 1.0094 -0.0721 0.9988 #2 100 6.820 - 7.669 1.0157 -0.1176 0.9986 #3 100 6.806 - 7.668 1.0097 -0.0714 0.9989 #1 100 17.7 - 111.0 1.0026 -0.4347 0.9989
2
#2 100 19.6 - 103.7 0.9939 -0.1404 0.9981 #3 100 18.0 - 123.2 0.9897 -0.1897 0.9991 #1 100 25.5 - 435.2 0.9942 2.1791 0.9996
2
#2 100 25.1 - 399.1 1.0082 0.3311 0.9994 #3 100 25.6 - 442.7 0.9944 2.2551 0.9994 #1 100 14 - 69 1.0013 0.0485 0.9963 #2 100 14 - 66 0.9863 0.6676 0.9960 #3 100 13 - 67 1.0161 -0.4917 0.9950 #1 100 85.0 - 198.1 0.9995 -0.1711 0.9978 #2 100 85.0 - 192.0 1.0016 -0.4681 0.9988 #3 100 85.0 - 194.7 0.9926 0.9061 0.9987 #1 100 2.70 - 19.37 0.9966 0.0934 0.9996 #2 100 2.63 - 19.36 0.9933 0.0872 0.9996 #3 100 2.64 - 19.48 1.0042 #1 98 0.33 - 2.76 1.0228 -0.0603 0.9855 #2 98 0.32 - 2.70 0.9995 -0.0140 0.9826 #3 99 0.34 - 2.52 1.0308 -0.0542 0.9803 #1 100 55.8 - 197.1 0.9897 0.1776 0.9997 #2 100 51.0 - 184.1 0.9921 -0.0870 0.9988 #3 100 54.1 - 199.3 0.9905 0.9342 0.9978 #1 100 17 - 488 0.9855 -0.4734 0.9998 #2 100 19 - 491 0.9919 -0.5176 0.9998 #3 100 21 - 489 0.9813 0.2346 0.9999 #1 100 1.1 - 18.1 1.0034 0.0120 0.9994 #2 100 1.2 - 19.9 1.0030 -0.0057 0.9995 #3 100 1.2 - 19.5 0.9911 -0.0010 0.9994
Total #
specimens
Specimen
range
Slope Intercept r
0.0375 0.9995
Appendix A
A-15
Loading...