Nova Stat Profile pHOx Training manual

Training Manual
Stat Profile pHOx Training Record
Instructor's Name: ___________________________________
Trainee's Name: ___________________________________
3
4
5
6
7
8
9
0
Date Completed/Initials
DWG #10-1016A
Handling and Running Patient Samples .............................................. ___________________
Handling and Running Controls ......................................................... ___________________
Replacing the Reagent Pack and Paper ............................................. ___________________
Identifying Flow Path and System Components .................................. ___________________
Calibrating the Analyzer ................................................................... ___________________
Daily Maintenance ........................................................................... ___________________
Troubleshooting ............................................................................... ___________________

Table of Contents Stat Profile pHOx Training Manual

QUICK REFERENCE
Key Facts (pHOx) ............................................................................. QR-1
Maintenance Planner ......................................................................... QR-2
Resolving Results Problems.............................................................. QR-3
Key Facts (BioProfile pHOx)............................................................ QR-4
Preface
Stat Profile pHOx Customer Training Programs ...................................... i
Learning Objectives .................................................................................ii
1 System Identification ............................................................ 1-1
1.1 System Overview .....................................................................1-1
1.2 Flow Overview......................................................................... 1-3
1.2.1 Reagents ......................................................................1-3
1.3 Flow Components ....................................................................1-3
1.3.1 Reagent Pack............................................................... 1-3
1.3.2 Reagent Harness..........................................................1-4
1.3.3 Pinch Valves................................................................1-4
1.3.4 Peristaltic Pump ..........................................................1-4
1.3.5 Sampler Assembly — Sample Probe.......................... 1-4
1.3.6 Sensor Module ............................................................1-5
1.3.7 Sensors ........................................................................1-5
1.3.8 Reference Electrode ....................................................1-6
1.3.9 Sensor Blanks..............................................................1-6
1.3.10 Interconnect Tubing ....................................................1-6
1.3.11 Barometric Pressure Module.......................................1-6
1.3.12 Pump Tubing............................................................... 1-7
2 Running an Analysis ...................................................................... 2-1
2.1 Is the Instrument Ready for Analysis?.....................................2-1
2.1.1 Home Screen ...............................................................2-1
2.2 Patient Samples — Acceptable Samples/Sample Handling ....2-1
2.2.1 Acceptable Samples ....................................................2-1
2.2.2 Acceptable Anticoagulants .........................................2-2
2.2.3 Syringe Handling Tips ................................................2-2
2.2.4 Capillary Handling Tips..............................................2-3
2.2.5 Sample is Presented and Analysis Occurs ..................2-3
2.3 Sample Information..................................................................2-4
2.4 Results Screen and Printout .....................................................2-4
2.5 Running an Analysis Exercise .................................................2-4
2.6 Standby Mode ..........................................................................2-4
PN 24302 Rev. B 7/2001 TOC-1
Table of Contents Stat Profile pHOx Training Manual
3 Operational Status ............................................................... 3-1
3.1 Home Screen - Ready for Analysis..........................................3-1
3.2 Questionable Results................................................................ 3-1
3.3 System Status Screen ...............................................................3-1
4 Calibrating the Analyzer ....................................................... 4-1
4.1 Full 2-Point Calibration - all except Hb and SO
4.2 Calibration Timing ...................................................................4-1
4.3 Other Types of Calibration ......................................................4-2
4.3.1 SO2 and Hb Calibration ..............................................4-2
4.3.2 Single Point Calibration for Non-Gas Sensors ...........4-2
5 Using and Handling Controls ................................................. 5-1
5.1 Controls ....................................................................................5-1
5.1.1 Stat Profile pHOx Controls .........................................5-1
5.2 Handling Controls ....................................................................5-1
5.2.1 Storage Temperature ...................................................5-1
5.2.2 Stability of Opened Ampules......................................5-2
5.2.3 Altitude and Barometric Pressure Effects...................5-2
5.3 When to Use .............................................................................5-2
5.4 Using the Stat Profile pHOx's On-Board QC Features...........5-2
5.4.1 The QC Soft key .........................................................5-3
5.4.2 Automatic QC Analysis (Internal) ..............................5-3
5.4.3 Manual QC Analysis (External)..................................5-3
5.5 QC Data....................................................................................5-7
5.5.1 View Today's Data ......................................................5-7
5.5.2 View Daily Statistics...................................................5-7
5.5.3 View Monthly Statistics..............................................5-8
5.5.4 Printing Levey-Jennings Charts.................................. 5-9
.......................
2
4-1
6 Replacing Reagents and Paper .............................................. 6-1
6.1 When to Replace ......................................................................6-1
6.2 How to Replace ........................................................................6-1
6.3 Replacing the Reagent Pack.....................................................6-1
6.4 Replacing the Control Pack......................................................6-1
6.5 Replacing the Printer Paper......................................................6-1
7 Troubleshooting ................................................................... 7-1
7.1 Sensor Screens .........................................................................7-1
7.2 System Test .............................................................................. 7-1
7.3 Error Log ..................................................................................7-2
7.4 Flow Related Problems ............................................................ 7-2
7.5 Sensor Problems.......................................................................7-3
7.6 Results Related Problems.........................................................7-3
7.7 Miscellaneous Problems - Operational ....................................7-3
7.8 Call Nova..................................................................................7-3
TOC-2 PN 24302 Rev. B 7/2001
Table of Contents Stat Profile pHOx Training Manual
8 Flow Test and Flushing ......................................................... 8-1
8.1 Manual Flow Test Lab ............................................................. 8-1
8.1.1 Water (Flow) Test .......................................................8-4
8.1.2 Pump Test....................................................................8-5
8.1.3 Flushing Sensor Module, Air Detector, and Probe.....8-7
8.1.4 Flushing the Reference Electrode ...............................8-8
8.2 Flushing the Entire Flow Path..................................................8-9
8.3 Reference Solution Test .........................................................8-12
9 Maintenance ........................................................................ 9-1
9.1 Logging Maintenance...............................................................9-1
9.1.1 Maintenance Log Sheets .............................................9-1
10 Routine Maintenance .......................................................... 10-1
10.1 Overview ................................................................................10-1
10.2 General Maintenance Procedures...........................................10-1
10.2.1 Suggested Daily Start-up Procedure .........................10-2
10.2.2 Flowpath Cleaning/Deproteinizing...........................10-2
10.2.3 Flow Cell Conditioning.............................................10-5
11 Maintaining Sensors ........................................................... 11-1
11.1 Overview ................................................................................11-1
12 Periodic Maintenance.......................................................... 12-1
12.1 Replacement of Pump Tubing, R-Line, and W-Line.............12-1
12.1.1 Pump Tubing Replacement.......................................12-2
12.1.2 Waste Line Replacement ..........................................12-3
12.1.3 Reference Line Replacement ....................................12-4
12.2 Probe and Air Detector Replacement ....................................12-5
13 Setup ................................................................................ 13-1
13.1 Adapting the Analyzer to Meet Your Requirements ............. 13-1
13.1.1 Data Input Display Options ......................................13-1
13.1.2 General Operating Options .......................................13-2
13.2 Inputting Selections................................................................13-2
13.3 Setup Menu Checklist ............................................................ 13-3
13.3.1 Results Configuration Menu Checklist.....................13-4
13.3.1.1 Reference & Alert Limits Setup Checklist 13-5
13.3.1.2 Electrode Offsets Setup Checklist............. 13-6
13.3.1.3 Results Units Setup Checklist ...................13-7
13.3.1.4 Results Suppression Checklist...................13-8
13.3.1.5 Remote Review..........................................13-9
13.3.1.6 Patient Name..............................................13-9
13.3.1.7 Mandatory Patient ID ................................13-9
PN 24302 Rev. B 7/2001 TOC-3
Table of Contents Stat Profile pHOx Training Manual
13.3.2 Operation Configuration Checklist .........................13-10
13.3.2.1 Analysis Configuration Checklist ...........13-11
13.3.2.2 Calibration Configuration Checklist .......13-12
13.3.2.3 System Configuration Checklist..............13-13
13.3.2.4 Analysis Mode Checklist.........................13-15
13.3.2.5 Hb Type Checklist ...................................13-15
13.3.2.6 Stat Mode.................................................13-15
13.3.3 Communication Menu Checklist ............................13-16
13.3.4 Printer Checklist......................................................13-17
13.3.5 System Password Checklist ....................................13-18
13.3.6 Operator Password Checklist ..................................13-18
13.3.7 Language Checklist.................................................13-19
14 QC Setup ........................................................................... 14-1
14.1 QC Setup Checklist ................................................................14-1
14.1.1 Control Lot Checklist................................................14-3
14.1.2 Expiration Date Checklist .........................................14-3
14.1.3 Daily Analysis Time Checklist .................................14-4
14.1.4 QC Lockout Checklist...............................................14-5
14.1.5 Set Ranges Checklist.................................................14-5
15 Troubleshooting Lab ........................................................... 15-1
15.1 Troubleshooting Exercises .....................................................15-1
Problem #1 .............................................................................15-2
Problem #2 .............................................................................15-2
Problem #3 .............................................................................15-3
Problem #4 .............................................................................15-3
Problem #5 .............................................................................15-4
Problem #6 .............................................................................15-4
16 Examination/Evaluation ..................................................... 16-1
16.1 Certificate and C.E.U. Credits ...............................................16-1
Training Examination.............................................................16-3
Training Evaluation................................................................16-7
A Appendix ............................................................................. A-1
A.1 Technical Assistance............................................................... A-1
A.1.1 Telephone Technical Support (Hot Line) ..................A-1
A.1.2 Field Service Support .................................................A-2
A.2 Ordering Parts ......................................................................... A-3
A.2.1 Recommended Spare Parts......................................... A-3
A.3 Warranty Replacements .......................................................... A-4
A.4 Emergency (After Hours) Parts Shipments.............................A-4
A.5 Reaching your Nova Sales Representative ............................. A-4
TOC-4 PN 24302 Rev. B 7/2001

Quick Reference Stat Profile pHOx Training Manual

Key Facts (pHOx and pHOx Basic)
NOTE:
The pHOx Basic Analyzer measures only pH, PCO2, and PO2.
Sample Volume: 45 microliters whole blood (Micro sample)
70 microliters whole blood (Normal sample)
Slope Limits:
pH 9.1 - 11.6
PCO PO
2
SO
2
+
Na
2
(-15.0) - ( -1.6)
7.9 - 12.6
6.9 - 18.2
8.8 - 11.5
Hct 12.0 - 50.0
Measurement Range:
pH 6.50 - 8.00 pH units H+ 316.23 - 10.00 nmol/L
PCO PO
2
SO
2
2
3.0 - 200 mmHg 0.4 - 26.7 kPa 0 - 800 mmHg 0.0 - 106.7 kPa
0.0 - 100 % 0.0 - 1.0
Hb 4.0 - 24.0 g/dL 40.0 - 240.0 g/L 2.5 - 14.9 mmol/L Hct 12 - 70 %
Dependency Rule and Information:
Hct Requires Na+ calibrated for calibration/results.
PCO SO
2
2
Requires pH calibrated for results. (pH is checked on standards to assure bicarb is correct.) Requires Hct and PO2 calibrated for results (uses calculated SO2 if measured SO2 is not available).
Hgb Requires Hct and SO2 for reporting.
Priority:
1. If linked with the CO-Ox, then CO-Ox Hgb is reported.
2. If no CO-Ox or measured result, calculated result from Hct reading (Hct/3) is reported.
3. If no CO-Ox, no measured, and no Hct, analyzer reports Default Value (Hgb).
P50 Measured reported only with PO
between 30 and 75 mmHg.
2
P50c is reported when PO2 is outside above limits. Also P50 requires calibrated SO2% and Hgb/Hct for reporting.
Qsp Qt Requires 2 separate blood samples for determination: mixed venous and arterial. RI Utilizes input or default FIO2% for calculation. Calculated Results Requires calibrated and proper results by measured tests used in calculation. Suppression (Hgb) Only Hgb can be suppressed. Your option is to turn Hgb from measured to calculated (one
or the other will always appear). Option is done in the Setup Menu.
Update to PN 24302 Rev. B 3/2002
QR-1
Quick Reference Stat Profile pHOx Training Manual

Maintenance Planner

Weekly
Monthly
Quarterly
NOTE:
For the BioProfile pHOx, SO2 is not applicable.
Reposition Tubing in Pinch Valves Record Slopes
Calibrate SO2
Clean SO2 Sensor and Cuvette Replace Pump Tubing
Replace W-line Replace R-line
As Needed
Deproteinize Probe/Sensor Module Replace Probe Replace Reagent Pack Replace PO2 Cap Condition pH Sensor Replace Control Pack
Replace PCO2 Cap Replace Sensors Replace Printer Paper
QR-2
Quick Reference Stat Profile pHOx Training Manual
Resolving Results Problems
Flow problems are often accompanied by sensor related problems. Do not change sensors or mem­branes until flow problems have been resolved first.
NOTE:
For the BioProfile pHOx, use protein solution instead of whole blood;
blood gases are just gases; SO2 is not applicable.
Blood Gases SO2 PO2/PCO2
Results High Verify controls are at room temperature or Low
Check for low slope Change membrane Change sensor Results High Verify controls are at room temperature Debubble sensor or Low Check for low sensor slope (<9.5) Condition sensor module w/whole blood Condition w/pH Conditioning Solution Calibrate Check for air leak in system Clean probe and preheater Check waste line for restrictions Change reference electrode
Results High Recalibrate, clean cuvette and sensor, or Low inspect or replace washer
pH
Condition sensor module w/whole blood Verify reference solution delivery Change pH sensor
QR-3
Quick Reference Stat Profile pHOx Training Manual

Key Facts (BioProfile pHOx)

Sample Volume: 300 microliters sample size for full panel
Slope Limits:
pH 9.1 - 11.6
PCO
2
PO
(-15.0) - (-1.6)
2
Measurement Range:
pH 5.00 - 8.00 pH units
PCO
2
PO
2
BarP 400.0 - 800.0 mmHg 53.3 - 106.7 kPa 15.7 - 31.5 inHg
7.9 - 12.6
3.0 - 200 mmHg 0.4 - 26.7 kPa 0 - 800 mmHg 0.0 - 106.7 kPa
Dependency Rule and Information:
PCO
2
Requires pH calibrated for results. (pH is checked on standards to assure bicarb is correct.)
Calculated Results Requires calibrated and proper results by measured tests used in calculation.
QR-4

Preface Stat Profile pHOx Training Manual

Stat Profile
Stat Profile® and pHOx® are registered trademarks of Nova Biomedical.
Training on the Stat Profile pHOx Analyzer is intended to be carried out by a Nova training specialist utilizing this training guide, the reference manual shipped with the analyzer, training aids, and supple­mental hands-on exercises as appropriate.
Training has been designed to be carried out in several stages corresponding to the operator’s involvement level. When carried out at your location, the time you make available per training session will determine how much material can be covered.
Basic operator training requires approximately 1-2 hours, based on analyzer features and the participants' prior experience. Key or principal operator training requires an additional 2 hours of training. See the following page for topics generally covered.
®
pHOx® Customer Training Programs
NOTES:
For additional information on training programs and materials, con­tact the Nova customer training department at:
(800) 545-NOVA ext. 571
i
Preface Stat Profile pHOx Training Manual
NOTES:
Learning Objectives
Basic Operator Training is designed to provide a new (first line) operator
with skills to:
Perform analyses of patient and QC samples
Locate and use selected software screens and functions
Identify major components
Perform routine maintenance procedures
Identify error conditions
Carry out simple corrective procedures
Use the reference manual or the Nova Hot Line as necessary
Key Operator Training adds those skills and information to:
Perform all routine maintenance procedures
Locate and utilize all necessary software screens and functions
Identify performance trends and patterns
Introduction to Nova support service
Customize maintenance schedules
Apply advanced troubleshooting techniques to resolve fluidic, calibration, and results-related problems
Identify pre-analytical and other sources of sample handling error
Use troubleshooting guides to resolve results-related problems
Train additional operators in the workplace
Management Options will be discussed with the medical and/or technical director. They include the following:
Adapting the analyzer to the facility’s needs (setup options)
Guidelines for analyzer performance verification (See
Getting Your Nova Analyzer On Line
- Linearity
- Precision
- Correlation
QA Program options
CLIA compliance
ii
, PN 17361.)

1 System Identification Stat Profile pHOx Training Manual

NOTES:
1 System Identification

1.1 System Overview

The display is a backlit liquid crystal display (LCD) for menus and messages. Menus allow use of operating options and features.
The keypad is the primary input device.
The internal printer prints patient results, QC, setup infor­mation, error log, etc.
The analytical compartment is where all calibration re­agents enter and the sensor measurements take place.
Reagents enter from the reagent pack. Waste materials (samples and reagents) are expelled into the reagent pack waste bottle.
Sensors and air detectors are automatically calibrated peri­odically to insure accuracy.
1-1
1 System Identification Stat Profile pHOx Training Manual
NOTES:
11
1
10
2
9
3
4
8
7
DWG #10-1017A
5
1-2
6
Figure 1.1 Analytical Compartment
1. Waste Line 7. Sampler
2. Reference Line 8. Air Detector
3. Pinch Valve (Reference) 9. Sensor Module with Sensors
4. Pump and Pump Tubing 10. Reference Electrode
5. Reagent Pack Opening 11. Pinch Valve (Waste)
6. Control Pack Opening
1 System Identification Stat Profile pHOx Training Manual
NOTES:

1.2 Flow Overview

1.2.1 Reagents

Reagents are drawn from the reagent pack, through the reagent harness, through the rotary valve, to the fluid fountain by way of the sample tubing, where they are aspirated by the sample probe. The sample probe can aspirate different reagents depending upon the position of the rotary valve. Aspirated fluids are drawn through the probe, the air detector, the sensor module with sensors, reference electrode, around the pump and out through the waste line to the waste container in the reagent pack.

1.3 Flow Components

1.3.1 Reagent Pack

The reagent pack contains 6 flexible bags in a cardboard carton. One of these is a waste bag to collect the used reagents, controls, and samples. The other 5 bags contain standard reagents: A, B, C, D, and R. Each bag includes a polyethylene fitment with a septa seal. The waste bag has a hydrophobic air vent. Septa are arranged in a line along the rear of the pack. The septa are pierced during insertion of the pack. The lot number and expiration date are at the front of the pack. The Reagent Management System (RMS) of the reagent pack auto­matically enters the calibration values, the lot number, the fluid volumes, and the expiration date to the analyzer's computer after insertion of the reagent pack.
Instructor shows all these parts as they are described.
Instructor points out
that the reagent pack
can be inserted and
removed a few times.
1-3
1 System Identification Stat Profile pHOx Training Manual
NOTES:

1.3.2 Reagent Harness

The Reagent Harness transports reagents from the reagent pack to the rotary valve. It is replaced as part of annual maintenance by Nova Service.

1.3.3 Pinch Valves

There are 2 pinch valves: one is used to control the flow of the reference fluid and the other one is used to control the flow of fluids through the flow cell.

1.3.4 Peristaltic Pump

The pump is a 6-roller peristaltic pump driven by a stepper motor.
1.3.5 Sampler Assembly — Sample Probe
The sampler allows for the aspiration of the sample. The sampler has 2 sampling positions: horizontal for the aspiration of a sample from a capillary tube and inclined for the aspiration of the sample from a syringe. A capillary adapter automatically moves to the end of the probe when the Capillary Analyze key is pressed. No special adapters are required to aspirate a sample from a capillary tube. The capillary or syringe positions are selected by pressing the Black Capillary key or the White Syringe key.
1-4
1 System Identification Stat Profile pHOx Training Manual
NOTES:

1.3.6 Sensor Module

The sensor module includes the preheater and flow cell. The preheater preheats samples and controls to 37°C. In addition, it contains the hematocrit impedance sensor and 2 air detectors. The sensor module geometry is an interlaced configuration with the reference electrode at the top of the sensor module, 3 sensors on the left side, 2 sensors on the right side. In addition to the Hct sensor located in the preheater, there are 6 sensors: Reference Electrode, Na+, PCO2, SO2 (optic), pH, and PO2. A window in the door allows flow path visibility and is augmented by a backlight.

1.3.7 Sensors

The Sensors housed in the sensor module are the core of the Stat Profile pHOx Analyzer. The methodology used by each sensor are
Sensor Methodology
+
Na
Sodium ion-selective electrode
pH Hydrogen ion-selective glass electrode
PCO PO
2
2
Severinghaus-type electrode
Polarographic Clark-type electrode Hct Impedance electrode Hb Impedance electrode/photometry
SO
2
Reflectance photometry (fiber optics)
The sensors clip into the sensor module, and an electrical contact is automatically made.
1-5
1 System Identification Stat Profile pHOx Training Manual
NOTES:

1.3.8 Reference Electrode

The Reference Electrode is mounted above the sensor module. It is a solid-state Ag/AgCl electrode and provides the reference voltage for comparison to sample voltages. The exit port of the flow path is located on this electrode.
1.3.9 Sensor Blanks (PN 22507)
Sensor Blanks (see Figure 1.2) are used in place of sensors that are not present on the analyzer or that are removed for certain maintenance and troubleshooting procedures. Periodically (as needed) remove blanks and clean the flow cell and tip of the blank with a cotton-tipped applicator that is moistened with bleach.
DWG #7-0-039A
Figure 1.2 Sensor Blank
1.3.10Interconnect Tubing (PN 07161)
Interconnect Tubing is installed between the reference electrode and the sensor module. Required for the proper sealing of these components.

1.3.11Barometric Pressure Module

The Barometric Pressure Module, located on a printed circuit board, continuously monitors the barometric pressure. This barometer can be calibrated against an external barometer, if desired, through the software.
1-6
1 System Identification Stat Profile pHOx Training Manual
NOTES:

1.3.12Pump Tubing

The pump tubing has 2 manifolds (a top and bottom) and reference and waste tubing, labelled R and W. The top manifold can be identified by the half circle (right-sided). The R and W-pump tubing are connected to the corresponding labeled outlets below the pump.
G #10-1007A
DW
Figure 1.3 Pump Manifolds and Tubing
1-7
1 System Identification Stat Profile pHOx Training Manual
NOTES:
1-8
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:

2 Running an Analysis

2.1 Is the Instrument Ready for Analysis?

2.1.1 Home Screen

Check the Home screen for Ready to analyze. The screen displays all analytes with the uncalibrated ones x'd out. Also, the next QC and Calibration times are displayed. A bar graph for the amount of reagents and controls shows the approximate percent remaining.
For whole blood
2.2 Patient Samples — Acceptable Samples/Sample Handling
analysis only.

2.2.1 Acceptable Samples

The acceptable sample is heparinized whole blood.
Nova instruments are designed for clinical environments to analyze
actual patient specimens, not modified blood samples. Specimens removed from the patient, anticoagulated appropriately, and promptly analyzed are the only type of sample where the measurement results will be reliable. Matrix effects/interferences can occur when patient specimens are removed from the body, modified and then measured on a Nova instrument. For example, matrix effects have been seen on Nova analyzers when attempting to analyze samples collected from cell savers used in various surgical procedures. Also, evaluation laboratories run specimens from patients with a wide variety of pathologies and from patients who are being treated with a broad spectrum of therapeutic and pharmacological agents. Despite exten­sive clinical trials, it is not possible to anticipate every possible combination of transfused blood products, crystalloids, and drugs (or their metabolites) that may be present in a blood sample. As a result, some users have found that their particular patient mix has necessi­tated making adjustments to maintenance. For example, a high number of cardiopulmonary bypass pump or ECMO (extracorporeal membrane oxygenation) samples result in a need for increased ana­lyzer maintenance. If you are experiencing excessive downtime, you may need to modify your own maintenance schedules. Nova’s Clini­cal Applications Group will assist you in tailoring a maintenance program to meet these needs.
No serum or plasma
unless operator has
validated the system
for that purpose.
No urine, CSF,
cardioplegia solution,
or other body fluids.
Update to PN 24302 Rev. B 3/2002
2-1
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:
This means no EDTA (purple tops), NaF (gray tops), or uncentrifuged red tops.
A general rule of thumb for blood gas and electrolyte analysis is to run the sample within 15 minutes of the time it was collected.
P
O2.... may increase in a
blood sample collected into a plastic syringe and stored on ice. The NCCLS document C32-P discusses these increases.

2.2.2 Acceptable Anticoagulants

Sodium and lithium heparin resulting in a final concentration of not more than 20 I.U. per mL in blood are the only acceptable anticoagulants.

2.2.3 Syringe Handling Tips

1. Remove air at time of collection.
2. Mix sample thoroughly by rolling between hands.
3. Run as soon as possible after collection. Do not ice plastic syringes.*
4. Check for clots by expelling a drop of blood onto a gauze pad.
5. For repeat samples, expel any air left in syringe.
6. Position sample so that probe tip stays deep in the syringe but not touching the syringe plunger.
* NCCLS. Nov., 1993. Considerations in the simultaneous measurement of blood gases, electrolytes and related analytes in whole blood., NCCLS document C32-P. Vol. 13, No. 17.
2-2
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:

2.2.4 Capillary Handling Tips

Collect in heparinized capillary tubes only.
Place metal “flea” (stirring bar) in capillary prior to collect­ing sample.
Sample must be collected without air spaces or gaps in capillary.
Using magnet, mix well after collection.
Cover tube ends with removable caps (not clay seal) for transportation to lab.
See additional notes in Chapter 1 of the Reference Manual.

2.2.5 Sample is Presented and Analysis Occurs

Press the White Button (Syringe) or the Black Button (Capillary) to run an analysis. The Aspirate screen displays. Wait until the sample probe is in position. Present the sample to the probe.
Instructor
demonstrates
analysis.
Samples can be presented in a syringe, opened vacuum tube, sample cup, or capillary.
Probe
Syringe
G #10-1024A
DW
Figure 2.1 Syringe Sample Presentation
2-3
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:
After the sample is positioned, press Continue (soft Key). (If not pressed within 45 seconds, the analyzer returns to the Ready screen.)
In STAT Mode Analysis, the Sample Information screen is not displayed. The ABG Results­Measured screen is displayed after aspiration is completed.
If the view results soft key is not selected the analysis will not go to completion. No results will be displayed or printed out.
NOTE:
If you attach an external key board, alpha keys are available.

2.3 Sample Information

2.4 Results Screen and Printout

The pump turns, pulling the sample through the probe, air detector tubing, and sensor module. A pop-up screen appears. Remove the sample from the probe. Press Analyze. (If Analyze -soft key- is not pressed within 30 seconds, the probe will automatically move down­ward.) The probe moves downward. The analyzer displays the Sample Information screen with a bar graph until completion. Input sample information, then press View Results to see the com­pleted results.
The Sample information screen is displayed after aspiration has been completed. This screen allows you to enter Accession number, Operator ID, Patient ID number, Combine with CO-Oximeter, com­bine with last sample for A/V & Shunt results, Patient temperature, and Sample Type. Use arrow keys to move to next option; Use the number keys or the Enter key to toggle or get pop-up selection.
2-4
The ABG Results -Measured screen is displayed and results fill in the table as they become available. The results on the screen can be printed by pressing Print (soft key). Press Next Page (soft key) to go to the next results screen, etc.

2.5 Running an Analysis Exercise

See Running an Analysis Checklist to perform the lab.

2.6 Standby Mode

See Standby Mode Checklist to perform the lab.
2 Running an Analysis Stat Profile pHOx Training Manual
Running a Syringe Analysis Checklist
NOTES:
Step 1 Check the instrument for readiness.
Check the display for the Home (Ready) screen.
Step 2 Present sample and begin analysis.
(Procedure for syringes)
Press the White Syringe Button.
Probe positions for sample aspiration. Slide the syringe tip over the extended probe.
CAUTION:
the base of the syringe plunger or the bottom of the sample cup or tube.
Do not allow the sample probe to touch
Step 3 Analysis begins.
Press Aspirate Normal (soft key) to start the sampling. (Or, press Aspirate Micro (soft key) for a smaller sample: 45µL instead of 70µL.)
NOTE:
5 minutes of the time the probe extends, the probe automatically starts to move down.
When the system signals by an audible alarm, withdraw the sample.
Press Analyze. The probe moves down.
NOTE:
after the sample has been aspirated, the probe automatically starts to move down.
If one of these soft keys is not pressed within
If Analyze is not pressed within 30 seconds
The analysis begins.
The Sample information screen is displayed.
Update to PN 24302 Rev. B 3/2002
2-5
2 Running an Analysis Stat Profile pHOx Training ManualRunning a Syringe Analysis Checklist
NOTES:
Step 4 Enter patient information.
The Sample information screen is displayed. (If in STAT Mode, this screen does not display. Go to next step.)
Enter patient data.
NOTE:
Press arrow keys to move through options. Press number keys or Enter key to toggle or get pop-up screen.
If the View Results (soft key) is not pressed, the analysis will not go to completion and will not print out the results.
Step 5 Results are displayed and printed.
Press View Results (soft key) to view the results.
Measured results are displayed on the ABG Results ­Measured screen as they become available. Selected calculated results can be viewed by pressing Next Page (soft key) from the Results screen.
To print the results, press Print (soft key).
Results are transmitted to an optional external printer or computer, if enabled.
After transmission and/or printout, all patient information becomes READ ONLY. They no longer can be changed or combined.
Press Home (soft key) to return to the Home (Ready) screen.
2-6
2 Running an Analysis Stat Profile pHOx Training Manual
Running a Capillary Analysis Checklist
NOTES:
Step 1 Check the instrument for readiness.
Check the display for the Home (Ready) screen.
Step 2 Prepare capillary.
(See Section 2.2.4, Capillary Handling Tips)
Mix sample again by drawing magnet back and forth along the length of the capillary.
Keeping capillary level, remove cap (or caps) from the ends.
Remove mixing flea by pulling it out one of the free ends with the magnet.
Step 3 Present sample.
Press the Black Capillary Button and wait for probe to position.
Holding one gloved finger over the open end of the capil­lary, carefully slide the tip of the capillary into the adapter until it just touches the probe.
The capillary should be in line with probe, not tilted horizon­tally.
Demo this before
trainees attempt it.
Skip this section if
they do not run
capillaries.
CAUTION: Do not
end of the capillary yet.
Step 4 Begin analysis.
Press Aspirate Normal (soft key) to start the sampling. (Or, press Aspirate Micro (soft key) for a smaller sample: 45µL instead of 70µL.)
NOTE:
5 minutes of the time the probe extends, the probe automatically starts to move down.
As soon as the pump starts to turn, remove your finger from the open end of the capillary, allowing the sample to be drawn up.
Update to PN 24302 Rev. B 3/2002
If one of these soft keys is not pressed within
remove your finger from the open
2-7
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:
Running a Capillary Analysis Checklist
When the system signals by an audible alarm, remove the capillary from the probe and adapter.
Press Analyze. The probe moves down.
NOTE:
after the sample has been aspirated, the probe automatically starts to move down.
The analysis begins.
The Sample information screen is displayed.
If Analyze is not pressed within 30 seconds
Step 5 Enter patient information.
The Sample information screen is displayed. (If in STAT Mode, this screen does not display. Go to next step.)
Enter patient data.
NOTE:
Press arrow keys to move through options. Press number keys or Enter key to toggle or get pop-up screen.
Step 6 Results are displayed and printed.
If the View Results (soft key) is not pressed, the analysis will not go to completion and will not print out the results.
Press View Results (soft key) to view the results.
Measured results are displayed on the ABG Results ­Measured screen as they become available. Selected calculated results can be viewed by pressing Next Page (soft key) from the Results screen.
To print the results, press Print (soft key).
Results are transmitted to an optional external printer or computer, if enabled.
After transmission and/or printout, all patient information becomes READ ONLY. They no longer can be changed or combined.
Press Home (soft key) to return to the Home (Ready) screen.
2-8
2 Running an Analysis Stat Profile pHOx Training Manual
Running an AV Shunt Analysis Checklist
NOTES:
Step 1 Check the instruments for readiness.
NOTE:
This checklist can only be performed if a Nova CO-Oximeter is in communication with the pHOx Analyzer, so that the results can be com­bined. This checklist may be removed if not needed.
Check the pHOx display for the Home (Ready) screen.
Check the CO-Oximeter for readiness. (See the Nova CO-Oximeter Reference Manual.)
Step 2 Present sample and begin analysis.
Press the White Syringe Button. Probe positions for sample aspiration.
Press the AV Shunt (soft key).
Position the mixed venous sample over the extended probe.
You must press soft
CAUTION:
the base of the syringe plunger or the bottom of the sample cup or tube.
Do not allow the sample probe to touch
keys within 45
seconds or the cycle
will abort
automatically.
Step 3 Analysis begins.
Press Aspirate (soft key).
When the system signals by an audible alarm, withdraw the sample.
Press Analyze (soft key). The probe moves down.
The analysis begins.
The Sample information screen is displayed; enter patient data.
NOTE:
Press arrow keys to move through options. Press number keys or Enter key to toggle or get pop-up screen.
If Analyze is not pressed within 30 seconds after the
sample has been
aspirated, the probe
automatically starts
to move down.
2-9
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:
Running an AV Shunt Analysis Checklist
Step 4 Continue Analysis on CO-Oximeter.
Analyze the mixed venous sample on the CO-Oximeter.
Step 5 Run the Arterial Sample.
When the message, Mixed Venous sample complete, posi­tion the arterial sample over the extended probe.
Press Aspirate (soft key).
After aspiration, remove arterial sample and press Analyze (soft key).
Step 6 Continue Analysis on CO-Oximeter.
If the View Results (soft key) is not pressed, the analysis will not go to completion and will not print out the results.
Analyze the arterial sample on the CO-Oximeter.
Step 7 Results are displayed and printed.
Press View Results (soft key) to view the results.
Measured results are displayed on the ABG Results ­Measured screen as they become available. Selected calculated results can be viewed by pressing Next Page (soft key) from the Results screen.
To print the results, press Print (soft key).
Results are transmitted to an optional external printer or computer, if enabled.
After transmission and/or printout, all patient information becomes READ ONLY. They no longer can be changed or combined.
Press Home (soft key) to return to the Home (Ready) screen.
2-10
2 Running an Analysis Stat Profile pHOx Training Manual
Standby Mode Checklist
NOTES:
Step 1 Start the procedure.
NOTE:
This option allows you to place the analyzer into a standby mode and allows you to reactivate the analyzer out of the mode at a programmed time and date. The standby mode will use less reagents and controls because the analyzer will not run its automatic calibration cycles. Calibration is lost
SO
except for set time is elapsed, if you initiate a calibration, or if
you turn Standby off.
From the Operational Menu, scroll down to the Standby Mode.
Press ENTER (once if mode is OFF or twice if Mode is ON).
and Hb. Standby mode ends after the
2
Step 2 Set the date.
Press ENTER (again) - the date becomes blank.
Key in the date that you want the analyzer to come out of Standby Mode.
Press ENTER to set the date. (DO NOT press OK.)
Step 3 Set the time.
Press any Arrow key to go to the time field.
Press ENTER - the time becomes blank.
Key in the time that you want the analyzer to come out of Standby Mode.
Press ENTER to set the time. (DO NOT press OK.)
2-11
2 Running an Analysis Stat Profile pHOx Training Manual
NOTES:
Standby Mode Checklist
Step 4a Start the Standby Mode.
Now press OK (soft key) to initiate the Standby Mode.
Another pop-up screen appears to verify that you want to go into Standby Mode. Press OK (soft key) again and the Standby Mode is officially ON.
Step 4b Cancel the Standby Mode.
Press Cancel (soft key) to cancel the Standby Mode.
The date and time fields revert to the previous settings.
The Standby Mode is now OFF.
2-12

3 Operational Status Stat Profile pHOx Training Manual

NOTES:
3 Operational Status

3.1 Home Screen - Ready for Analysis

Indicates readiness of the analyzer.
All analytes are shown. Uncalibrated analytes are x'd out and a — (a line through an analyte prefix) when the channel did not pass QC and QC lockout is enable.
The Next Cal and if programmed the Next QC is displayed.
Two fluid gauges display the approximate remaining solu­tions for the reagent cartridge and control cartridge.

3.2 Questionable Results

Results with a problem display a symbol in the Alert column. There are a number of symbols that can appear after the results. The symbols have the following meanings:
(single up arrow), (single down arrow) - The result is higher or
lower than the defined reference range for the parameter.
↑↑ (double up arrow), ↓↓ (double down arrow) - The result is higher
or lower than the defined alert range for the parameter.
↑↑↑ (triple up arrow), ↓↓↓ (triple down arrow) - The result is out of
the analyzer's operating range.
X (an X through an analyte prefix) - The channel is uncalibrated.
? (question mark) - Insufficient sample is detected during sample
reading.
* (asterisk) - The result is calculated using a default sodium concentration.
— (a line through an analyte prefix) - The channel did not pass QC and QC lockout is enabled; or the results have been suppressed.

3.3 System Status Screen

This screen only appears if there are operational messages or status codes.
3-1
3 Operational Status Stat Profile pHOx Training Manual
NOTES:
3-2

4 Calibrating the Analyzer Stat Profile pHOx Training Manual

NOTES:
4 Calibrating the Analyzer
4.1 Full 2-Point Calibration - all except Hb and
Uses internal calibrators (standards)
Calculates slopes for sensors and air detectors
Occurs at regular intervals or can be manually initiated by pressing Calibrate (soft key)
Included as part of most maintenance routines
Resets calibration timers
Can generally be stopped by pressing Exit (soft key)
Can be manually run by pressing the Calibration (soft key) from the Ready (or Not Ready) screen.
1. The 2 pt. Calibration screen will display.
2. Select option 1: pH, PCO2, PO2, Hct Calibration (ABG).
3. Press Enter.

4.2 Calibration Timing

Auto calibration takes place at intervals of 2, 4, or 6 hours depending on analyzer activity. See Chapter 3 of the Reference Manual for more details.
S
O
2
4-1
4 Calibrating the Analyzer Stat Profile pHOx Training Manual
NOTES:

4.3 Other Types of Calibration

4.3.1SO2 and Hb Calibration

Uses external calibrators (2 levels)
Initiated by the operator monthly or as needed
Access from Ready screen: Calibrate (soft key)
Instructor shows Calibration Configuration screen for calibration options. Access the Setup Menu (Password), select Operation Configuration Menu (press Enter). Select the Calibration Configuration option (press Enter). Discuss the options.
CLIA regs. require 30 minutes.

4.3.2 Single Point Calibration for Non-Gas Sensors

Values for calibration (standards) are lot specific and must be verified or reentered at time of calibration.
S
O
See Calibrating
Occurs with each gas Cal I (30-45 min. intervals) if Mode A is selected
Occurs with each analysis if Mode B is selected
Uses a calibration standard (either C or A) and increases time to appearance of results
Checks for sensor drift
Printing of drift is operator selected.
and Hb Checklist.
2
Mode A: One point fluid & Gas Cal done every 30-45 minutes.
Mode B: One point fluid & Gas Cal done with EACH analysis.
4-2
4 Calibrating the Analyzer Stat Profile pHOx Training Manual
Calibrating SO2 Checklist
NOTES:
Step 1 Begin calibration.
From the Home (Ready) screen, press Calibrate (soft key).
Then select option 2: External two standard Hb, Calibration then press ENTER.
The External Standard Hb, SO2% Calibration screen ap­pears.
SO
%
2
Step 2 Verify calibrator (standard) values.
Consult insert sheet for this lot and compare to values on the screen.
Key in the assay value for Std. #1, if necessary.
Step 3 Prepare calibrator #1.
Mix gently by inversion.
Open the ampule carefully to avoid injury (glass edges).
Position Calibrator #1 so that the extended probe tip is well covered by the solution.
Press Continue (soft key).
When system signals by audible alarm, remove the ampule,
Press Analyze (soft key).
Step 4 Prepare calibrator #2.
Repeat procedure for calibrator #2.
Enter/verify assay value for Std. #2. (Press ENTER after entering the assay value.)
Mix gently by inversion.
Open the ampule carefully to avoid injury (glass edges).
Position Calibrator #2 so that the extended probe tip is well covered by the solution.
Press Continue (soft key).
When system signals by audible alarm, remove the ampule,
Press Analyze (soft key).
4-3
4 Calibrating the Analyzer Stat Profile pHOx Training Manual
NOTES:
4-4

5 Using and Handling Controls Stat Profile pHOx Training Manual

NOTES:
5 Using and Handling Controls

5.1 Controls

CAUTION:
of controls other than Nova controls and will void warranty on components that come into contact with them.
The Stat Profile pHOx Analyzer has on board QC and on board controls. Controls can be analyzed either with the internal control pack or with external Nova control ampules.
Performance of the analyzer may be affected by the use

5.1.1 Stat Profile pHOx Controls

Monitor performance of all blood gases and electrolytes
Contain no preservatives, dyes, or surfactants that can damage sensors or other flow components
Formulated at 3 clinically significant levels: L1 Acidosis, with low pH, high PCO2, low PO2, low SO2,
low normal Hct and low normal Hb values
L2 Normal, with normal pH, PCO2 and PO2 values L3 Alkalosis, with high pH, low PCO2, high PO2 high
SO
, and normal high Hct/Hb values
2
Internal in a self-containing control pack (Auto- Cartridge QC) or external in individual ampules
The Auto- Cartridge
QC has a use-life of 35
days.

5.2 Handling Controls

5.2.1 Storage Temperature

Store at 25°C (77°F). Warmer or colder controls affect gases and pH. Hold vial between thumb and fingertip when shaking to avoid changing temperature. For every 1°C 1%. The opposite effect is seen at temperatures
CAUTION:
sill or near a heating or air-conditioned duct.
Do not refrigerate. Avoid storing controls on a window
above
25°C, PO2 recovery drops
below
25°C.
5-1
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:

5.2.2 Stability of Opened Ampules

Shake and analyze within 30 seconds of opening. Contamination with room air will begin to affect PO2, PCO2, and pH results after that time.
PO
Room air has a Exposure of controls to room air causes:
PCO
to decrease.
2

5.2.3 Altitude and Barometric Pressure Effects

Aqueous controls are assayed at sea level. Changes in altitude will affect gas recovery. For every 1000 ft above sea level, PO2 recovery drops by approximately 0.5%. An incorrect barometric pressure will also affect gas recovery. System barometric pressure is displayed on the printout, barometric pressure, or Status screen. If significantly different from a known, accurate measurement, enter an offset by selecting Setup, enter password, select Operations Configuration, press Enter, select System Configuration Menu (use arrow keys to go to corrected Barometric Pressure), enter mmHg, Press Home to exit to Home (Ready) screen.
of about 150 mmHg and a PCO2 of ~0 mmHg.
2
PO
and pH to increase,
2

5.3 When to Use

As of the date of publication of this guide, a minimum of one blood gas control must be analyzed every 8 hour shift of lab operation. Also, after maintenance or troubleshooting or as required by the laboratory’s quality control program. All levels must be analyzed within a 24 hour period.

5.4 Using the Stat Profile pHOx's On-Board QC Features

The analyzer's QC features allow you to:
Set ranges and units for acceptable results.
Store statistics.
Program for internal or external control source.
Display or print QC statistics in the form of reports.
Remind or require operators to run QC at specific intervals.
Analyze other forms of QC such as proficiency surveys or linearity standards without “offsets”.
5-2
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:
These features must be enabled or setup before they can be used. See Quality Control Setup in the analyzer’s reference manual for details. If you will be using them, the installer will have done at least a partial setup so that you can learn how they work.

5.4.1 The QC Soft key

This key has more than one function:
To run a QC analysis either internal or external
To view or to print QC data
To set up QC levels
To turn automatic QC analysis on or off

5.4.2 Automatic QC Analysis (Internal)

To use the Automatic QC Analysis feature properly, the QC analysis times need to be programmed under the Daily Analysis Times screen during the QC Setup and the Automatic QC Analysis needs to be turned on. QC is automatically performed at these programed times with the controls from the Auto-Cartridge QC.
NOTE:
Internal QC can also be run manually: see the checklist.
5.4.3 Manual QC Analysis (External)
To use the Manual QC Analysis feature properly, the QC analysis times need to be programmed under the Daily Analysis Times screen during the QC Setup and the Automatic QC Analysis needs to be turned on. When QC analysis is needed, a Pop up screen will appear prompting you to run the required QC. Press OK (soft key). Then go to the Analyze QC screen and select the appropriate QC.
GO TO CHECKLIST.
5-3
5 Using and Handling Controls Stat Profile pHOx Training Manual
DWG #6010A
NOTES:
Analyzing Controls Checklist
Step 1 Preparing for Analysis
Hold QC ampule between thumb and forefinger and agitate vigor­ously for several seconds.
NOTE:
Do not warm ampule in hands.
Step 2 Initiating the QC Analysis
Press QC (soft key) on the Home screen.
With Analyze QC Highlighted, press ENTER.
Select the QC to be analyzed, i.e., QC Level #1 External.
Change if necessary by pressing Previous Control (soft key) or Next Control (soft key).
Press the White Button (Syringe) or Analyze (soft key).
Step 3 Completing the Cycle
Figure 5.1 Agitating the Ampule
5-4
Once probe has extended, open the ampule while protecting yourself from the sharp glass edges.
Fully immerse the probe into the fluid of the ampule. Do not allow the probe to touch the bottom of the ampule.
Press Continue (soft key).
When the system signals by an audible alarm, remove the ampule and press Analyze (soft key). The probe moves downward.
Discard the ampule.
NOTE:
cause the system to to the Quality Control screen.
Failure to press Continue (soft key) within 45 seconds will
time-out
canceling the cycle and returning you
5 Using and Handling Controls Stat Profile pHOx Training Manual
Analyzing Controls Checklist
NOTES:
Step 4 Accepting or Rejecting Results
When results are ready, they appear on the screen along with an indication of whether they have passed or failed the QC range check. If one or more tests are out of range, the QC Results screen flashes Exceed Limits.
To save all data, press QC (soft key).
To reject the data, press Delete (soft key). Then press Continue to discard all the data.
All data is either accepted or rejected.
Step 5 Repeat for next level
5-5
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:
Analyzing Internal Controls Checklist
Step 1 Preparing for Analysis
From the Ready screen, press QC (soft key) on the Home screen.
Step 2 Initiating the QC Analysis
With Analyze QC Highlighted, press ENTER.
Select the QC to be analyzed, i.e., QC Select Internal L1.
Change if necessary by pressing Previous Control (soft key) or Next Control (soft key).
Press Analyze (soft key).
Step 3 Accepting or Rejecting Results
When results are ready, they appear on the screen along with an indication of whether they have passed or failed the QC range check. If one or more tests are out of range, the QC Results screen flashes Exceed Limits.
To save all data, press QC (soft key).
To reject the data, press Delete (soft key). A pop-up screen appears: enter your password. Then press Continue to dis­card all the data.
All data is either accepted or rejected.
Step 4 Repeat for next level
5-6
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:

5.5 QC Data

To view QC data, select View or Print QC Data on the Quality Control (QC) screen then press Enter. The QC Data screen for the selected level is displayed. From this screen, you can view today's data, daily statistics, monthly statistics, cumulative, or Levey-Jennings.

5.5.1 View Today's Data

From the QC Data screen, select View Today's Data then press Enter. The QC Results screen is displayed. To view the next result, press Next Page (soft key). For options, press Options (soft key). A pop-up window appears. One option is for deleting this result. You can also print or view other statistics.

5.5.2 View Daily Statistics

From the QC Data screen, select View Daily Statistics then press Enter. This screen is the QC Daily Statistics for the selected level. From this screen, you can view the next level, and/or print the statistics.
5-7
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:
5.5.3 View Monthly Statistics
The Monthly Statistics is a record of all QC accepted for the past month. Print this record for your files at the end of each calendar month.
1. From the Quality Control screen, select View or Print QC Data.
2. From the QC Data screen select View monthly statistics.
3. Press Print (soft key).
=======Stat profile QC Statistics======
Date 6-23-98 Time 10:54 Analyzer ID 0 Operator ID 1 Lot#1234567890 Exp 7-31-98
----------------------------------------------------
DAILY (n=6)
Test Mean ±Range SD CV
pH 7.595 0.050 2.3 2.3 pCO223.0 10.0 2.3 2.3 pO SO Hb 7.5 10.0 3.5 3.5 Hct 18 10.0 2.3 2.3
----------------------------------------------------
Test Mean ±Range SD CV
pH 7.595 0.050 2.3 2.3 pCO223.0 10.0 2.3 2.3 pO SO Hb 7.5 10.0 3.5 3.5 Hct 18 10.0 2.3 2.3
----------------------------------------------------
QC CUMULATIVE (n=1,498 during163 days)
Test Mean ±Range SD CV
pH 7.595 0.050 2.3 2.3 pCO223.0 10.0 2.3 2.3 pO SO Hb 7.5 10.0 3.5 3.5 Hct 18 10.0 2.3 2.3
----------------------------------------------------
155 12.0 1.0 1.0
2
.... .... .... ....
2
MONTHLY (n=97 during 30 days)
155 12.0 1.0 1.0
2
.... .... .... ....
2
155 12.0 1.0 1.0
2
.... .... .... ....
2
5-8
Figure 5.2 QC Statistics Printout
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:

5.5.4 Printing Levey-Jennings Charts

The Levey-Jennings data graph is displayed on screen. When print­ing, each graph will be rotated 90° with the date appearing on the long axis of the chart.
1. From the Quality Control screen, select View or Print QC Data.
2. From the QC Data screen, select View Levey-Jennings.
3. Highlight the analyte using the arrow key.
4. Press enter to view data on the screen.
5. Press Print (soft key).
=======Stat profile QC Statistics======
LEVEY-JENNINGS CHARTS Analyzer ID 0 Operator ID ... Lot # 1234567890 Exp Date 07-31-98 Date Lower Limit Upper Limit
----------------------------------------------------
7.585 7.600 7.635
---------------------------------------------------­11-23 * 11-24 * 11-25 * 11-26 * 11-27 * 11-28 * 11-29 * 11-30 * 12-01 * 12-02 *
Figure 5.3 Levey-Jennings Graphs Printout
5-9
5 Using and Handling Controls Stat Profile pHOx Training Manual
NOTES:
5-10
6 Replacing Reagents & Paper Stat Profile pHOx Training Manual
NOTES:

6 Replacing Reagents and Paper

6.1 When to Replace

Replace the fluid pack and/or the control pack when it has reached its expiration date or when the graphic appears empty on the Home screen. Also, replace the capillary adapter found in the reagent pack.

6.2 How to Replace

To install a new reagent pack and/or control pack, you will be using one of the automated Operational Menu procedures: Change Reagent Pack or Change Control Pack. These procedures come up as screens with the steps necessary to carry out the changing of the pack. Because the reminders are brief, you should refer to the complete illustrated proce­dures located in the maintenance section of the analyzer's reference manual if you need additional details. To stop automated maintenance procedures, press Cancel (soft key).
The on-board Reagent
Cartridge has a use-
life of 45 days.
The on-board QC
Cartridge has a use-
life of 35 days.

6.3 Replacing the Reagent Pack

See Replacing the Reagent Pack Checklist to perform the lab.

6.4 Replacing the Control Pack

See Replacing the Control Pack Checklist to perform the lab.

6.5 Replacing the Printer Paper

See Replacing the Printer Paper Checklist.
6-1
6 Replacing Reagents & Paper Stat Profile pHOx Training Manual
NOTES:
Replacing the Reagent Pack/Capillary Adapter Checklist
WARNING: Blood samples and blood products are potential sources of hepatitis and other infectious agents. Handle all blood products and flow path components (waste-line, septum assembly, probe, flow cell, etc.) with care. Gloves and protective clothing are recommended. Dispose of the reagent waste bottle according to your laboratory's biohazardous waste policies.
Step 1 Replace the capillary adapter.
From the Home (Ready) screen, press Menu.
Select Change Pack.
Press Enter. A pop-up screen appears. Change Reagent Pack
Change Control Pack Exit
Highlight Reagent Pack.
Press Enter.
Press Move Probe (soft key).
Open the door.
Remove the capillary adapter from the front of the probe by
pulling it off.
Replace the adapter with a new one found in the reagent
pack. As you put on the adapter, make sure the probe goes into the center hole of the adapter.
Press Move Probe (soft key) when finished.
Press Cancel then press Cancel again to return to the Home
screen.
6-2
Step 2 Remove old cartridge.
Open the door.
Remove the old cartridge - lift the cartridge up slightly to
clear the lip.
Mix the new cartridge by gently inverting for several seconds.
DO NOT SHAKE THE PACK.
Slide in the new cartridge.
Close the door.
Update to PN 24302 Rev. B 3/2002
6 Replacing Reagents & Paper Stat Profile pHOx Training ManualReplacing the Reagent Pack/Capillary Adapter Checklist
Step 3 Prime the analyzer.
Press Prime (soft key): this primes and validates the car­tridge.
Step 4 Calibrate the analyzer.
After priming, a pop-up screen appears with a question, "Do you want to recalibrate? Yes or No." Press Yes (soft key).
Calibration takes place.
Step 5 Validate performance with QC.
Run QC.
Document the results.
NOTES:
6-3
6 Replacing Reagents & Paper Stat Profile pHOx Training Manual
NOTES:
Replacing the Control Pack Checklist
WARNING: Blood samples and blood products are potential sources of hepatitis and other infectious agents. Handle all blood products and flow path components (waste-line, septum assembly, probe, flow cell, etc.) with care. Gloves and protective clothing are recommended. Dispose of the reagent waste bottle according to your laboratory's biohazardous waste policies.
Step 1 Initiate the procedure
From the Home (Ready) screen, press Menu.
Select Change Pack.
Press enter. A pop-up screen appears. Change Reagent Pack
Change Control Pack Exit
Highlight Control Pack by using the down arrow key.
Press Enter.
Step 2 Follow the screen directions.
Open the cover.
Remove the old cartridge - lift the cartridge up slightly to
clear the lip.
Mix the new cartridge by gently inverting for several seconds.
DO NOT SHAKE THE PACK.
Slide in the new cartridge.
Close the door.
Step 3 Prime the analyzer.
Press Prime (soft key).
Step 4 Calibrate the analyzer.
After priming, a pop-up screen appears with a question, "Do
you want to recalibrate? Yes or No." Press Yes (soft key).
Calibration takes place.
6-4
Step 5 Validate performance with QC.
Run QC.
Document the results.
Update to PN 24302 Rev. B 3/2002
6 Replacing Reagents & Paper Stat Profile pHOx Training Manual
Replacement of the Printer Paper Checklist
NOTES:
Step 1 Open the printer cover.
Step 2 Lift print head.
Open the printer platen.
Gently pull the lever near the paper advance knob forward.
Remove the depleted roll of paper.
Step 3 Remove paper holder from the old roll of paper.
Step 4 Install new roll of paper.
Insert the paper holder into a new roll of paper. The loose end of the paper should feed from the bot­tom of the roll.
Install the paper holder with paper into the support col­lars.
Push the paper through the back of the roller.
Step 5 Reposition print head.
DWG #10-1002B
Platen
Paper Advance
Knob
Figure 6.1 Paper Installation
Center the paper and close the printer platen by moving the lever back to the original position.
Advance the paper by using the paper advance knob.
Feed the paper through the cover. Then close the cover.
6-5
6 Replacing Reagents & Paper Stat Profile pHOx Training Manual
Replacement of the Printer Paper Checklist
NOTES:
Step 6 Test printer (Optional).
From the Home screen, press Menu (soft key); press Service (soft key); use the down arrow key to select Printer Menu: press ENTER; use the down arrow key to select Character Set Test; finally press ENTER to obtain a character printout to verify proper operation (see Figure 6.2).
After the self test is completed, a pop-up screen appears. Press OK.
Press Home (soft key) to return to the Home screen
** TEST PRINT ** !”#$%&’()*+,-./01234567 89:;<=>?@ABCDEFGHIJKLMNO PQRSTUVWXYZ[\] hijklmnopqrstuvwxyz{|}~
^
‘abcdefg
ЗьйвдаезклипомДЕ ЙжЖфцтыщяЦЬ¢£¥ ƒбнуъсСao
1
¿
αβΓπΣσµτΦθΩδ∞φ∈ ≡±≥≤ ÷≈°.−
DWG #9-1000A
1
〈〈 〉〉
/
/
2
4
n2
§ßØø °
Figure 6.2 Sample Test Printout
¨
6-6

7 Troubleshooting Stat Profile pHOx Training Manual

NOTES:
7 Troubleshooting
Instructor: Show
these screens as they

7.1 Sensor Screens

For a full description of these screens, turn to Chapter 6 of the Reference Manual.
From the Home screen, press Menu (soft key). Then press Service (soft key). Using the down arrow key, select Sensor Subsystem. Then press ENTER. Calibration and analysis data for each sensor, the SO2 LEDs, and the air detectors are contained on the following screens.

7.2 System Test

For a full description of this screen, turn to Chapter 6 of the Reference Manual.
From the Home screen, press Menu (soft key). Then press Service (soft key). Using the down arrow key, select System Test then press ENTER. The System Test screen displays millivolt readings for all sensors. From this screen, you can manipulate the pHOx Plus to perform many actions, i.e., priming. It also allows you to check the sampler, rotary valve, pump, waste valve, reference valve, ADTs, and SO2 LEDs. Turn to Chapter 6 of the Reference Manual for procedures on how to perform these tests. The operator flow test is performed from the System Test screen by pressing Operator Flow Test (soft key). The pump turns on, the probe moves to the syringe position, and the waste valve opens. This allows the user to present a cup of water to the probe and observe the flow of water.
are discussed.
7-1
7 Troubleshooting Stat Profile pHOx Training Manual
NOTES:

7.3 Error Log

From the Home screen, press Menu (soft key). Then press Service (soft key). Using the down arrow key, select Error Log and press ENTER. This screen shows all errors displayed in chronological order fol­lowed by a short explanatory message - holds 96 errors. For more detailed information, refer to the Reference Manual (Chapter 5,
Troubleshooting).
An index lists all the status messages by code with a symptom and corrective action.

7.4 Flow Related Problems

The cause of flow problems are clots, hidden blockages, leaks, and pumping problems. The errors show up as flow status codes. Resolve by referring to solutions described in the Reference Manual, Troubleshooting chapter, for that status code.
NOTE:
Always fix flow problems first.
Flow problems are often accompanied by sensor related problems. Do not change sensors or membranes until flow problems have been resolved. Once flow problems are resolved, sensor problems often disappear. Diagnose with flow tests: Water Test, Pump Test, and Reference Solution Test.
Water Test — Checks to verify that water can be pulled through the system from the probe. If water cannot be aspirated through the probe, a clog, leak, or a mechanical pump problem exists in the system.
Pump Test — Checks to verify that the pump and waste valve are functioning properly. Also verifies if the W-line from the reference electrode to the waste bottle is free of block­ages.
Reference Solution Test — Verifies reference solution is get­ting to the reference electrode.
7-2
7 Troubleshooting Stat Profile pHOx Training Manual
NOTES:

7.5 Sensor Problems

Problems may be uncalibrated, overload, instability, slope, drift, out of range, and dependency status codes. Resolve by referring to solutions described in the Chapter 5, Trouble-
shooting, in the Reference Manual for that status code.

7.6 Results Related Problems

Problems may occur because sample results are either too high or too low. Can be combination of two or more results problems. See Resolving Results Problems in the Quick Reference section. Resolve by referring to solutions described in the Chapter 5, Trouble-
shooting, in the Reference Manual for that error code.

7.7 Miscellaneous Problems - Operational

Problems may be related to lab temperature, incorrect barometric pressure, open door, or reagent pack problems. Resolve problems by lowering/raising temperature, correcting baro­metric pressure in the analyzer, closing door, and reinstalling/install­ing a new reagent pack.

7.8 Call Nova

For problems involving hardware, electronics, printer, or software error messages. Problems that cannot be resolved by you or at your site. Resolve problems by calling Nova Technical Services at 1-800-545-NOVA. Have the serial number and the closest phone extension to the instrument ready.
7-3
7 Troubleshooting Stat Profile pHOx Training Manual
NOTES:
7-4
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:

8 Flow Test and Flushing

8.1 Manual Flow Test Lab

WARNING: Use appropriate protective face shield, gloves, etc., when attempting to flush, clean or handle any components of the analyzer which come in contact with blood specimens.
The Stat Profile pHOx flow tests are used to troubleshoot fluidic problems that might give the following status codes:
71 Flow - No Std A
72 Flow - No Std B
73 Flow - No Std C
74 Flow - No Std D
75 Flow - Flow long/short
76 Flow - No sample
77 Flow - No air
The fluid system can be isolated into 3 large sections:
Fluid Pack to fluid fountain
Sampler/sensor module
W/R-lines to waste tubing
Performing the flow test limits the search to one of the fluid sections.
What you will need to do this test:
1. A sample cup, 25 mL medicine cup, or 25 mL beaker
2. Deionized water
3. Syringe with blunt needle and silicone tubing (probe cleaning kit or equivalent)
4. Lint-free tissue or gauze
8-1
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:
From ReadyServiceSystem TestCustomer Flow Test
Operator Flow Test
1. - Lift the tubing out of the Waste solenoid.
1
- If tubing remains pinched, roll it between fingers to loosen.
- Place the sample probe in a cup of water.
1
2
Is fluid flow noticed in the waste line tubing?
A. IF YES, call your Nova Service Representative. B. IF NO, go to Step 2.
2. - Disconnect the Waste (W) from the Reference Electrode.
2
- Place this end of the W tubing into a cup of water. Is fluid flow seen in the Waste line tubing?
A. IF YES, press “PUMP to turn off the pump motor.
Go to the next page,
“Sensor Module Back Flush procedure.
B. IF NO, go to Step 3.
3. - Disconnect the Waste line tubing from the front panel port.
3
- Place this end of the tubing over a gauze to prevent spilling.
- Aspirate water from the same location as in Step 2 above.
Is fluid seen exiting the Waste tubing end you just disconnected?
A. IF YES, go to Step 4. B. IF NO, Replace the W/R Tubing Harness ( P/N 23023)
4. - Press the “PUMP key to stop the Pump Motor
4
- Remove the Reagents Cartridge from the unit.
- Install the flush adapter.
- Place the flush adapter tubes into an empty beaker.
- Inject water to the front panel Waste port.
- Water should flow through the adapter waste tube. (Adapter waste line is last to the right)
* Loss of Reference flow will change the flow rate.
Reference solution will flow at a rate of 1 -2 drops/second during a pumping cycle.
** If this tubing cannot be flushed call your Nova Service Representative. If tubing flushes easily, change the Reagents
Cartridge. In a small number of cases the original problem may have been poor alignment of the Reagents Cartridge.
8-2
4
3
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:
Sensor Module Back Flush Procedure
5. If you are starting from the Operator Flow Test
5
-Turn the Pump OFF by pressing the PUMP” key.
6. If you are starting from the Ready Screen
7
8
6
- Press Service, System Test and select the Operator Flow Test.
- Turn the pump OFF as in Step 5.
7. Connect a syringe filled with water to the Reference
7
electrode waste port.
8
4. Back flush the flow path by injecting the water. Caution: Place a gauze or towel at the sample probe tip to receive the obstruction or water.
9. If the water easily flushes through the sensor module,
9
you may have an air leak into the flowpath. This will be from an interconnect tubing, failed membrane, or poorly seated electrode. Go t o Step 15.
10
6. If water does NOT exit the sample probe.
10
- Lift the Reference electrode off the sensor module.
- Flush the Reference electrode.
11
If it flows freely, confirm that the interconnect tubing is properly positioned and reinstall the Reference Electrode.
13
14
7. Disconnect the Air Detector from the bottom of Sensor
11
Module.
8. Inject water into the W port of the Reference Electrode.
12
Does water flow through the sensor module? IF YES, go to Step 13. IF NO, go to Step 14.
9. - Connect the syringe to the Sample Probe tip.
13
- Inject water through the Probe Water should exit the Air Detector (ADT) tubing. IF NO flow is noted, flush the ADT and probe separately. IF the water flows easily, the seal between the ADT and probe may be leaking. Replace the Air Detector.
(NO from Step 12) The obstruction lies in the flowcell.
14
While continuing to inject, remove one electrode at a time. Starting with the bottom electrode. When water starts to flow either the obstruction flushed into the flowcell or the electrode just removed was the cause. Re-membrane if needed. Dry the flowcell before re-inserting the electrode.
( From Step 9) Connect the syringe with water to the
15
sample probe tip. Reconnect the Reference electrode Waste line. Apply a slight pressure to the syringe. Look for water entering a flowcell, electrode cap, or leaking from the interconnect tubing. Replace the membrane, or tubing as required.
15
If the above does not identify the problem, please contact your Nova Service Representative.
8-3
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:

8.1.1 Water (Flow) Test

The water (flow) test verifies that fluid can be pulled through the system from the probe. If water cannot be pulled through the system, a clog, leak, or mechanical pump problem exists.
1. From the Home screen, press Menu (soft key)
2. Press Service (soft key).
3. From the Service screen, select System Test. Then press Enter.
4. Move the sampler to the syringe position by using the arrow keys at the bottom right corner of the key pad.
a. Select Sampler. b. Press Enter. c. Select Syringe. d. Press Enter.
5. Open the waste valve by using the arrow keys at the bottom right corner of the keypad.
a. Select waste valve b. Press Enter. (Valve toggles to open position if
closed.)
6. Turn the pump on to fast speed by using the arrow keys at the bottom right corner of the key pad.
a. Select Pump. b. Press Enter. c. Select Fast. d. Press Enter.
7. Fill a 2 mL sample cup with water.
8. Open the door.
9. Insert the probe into the cup of water and observe for water being aspirated.
8-4
If water is NOT aspirated from the cup, then go to the Pump Test to check waste aspiration. If water is aspirated from the cup, then the sample probe, sensor module, external tubing, and waste line are functioning with no blockage. The problem exists between the reagent pack and fluid fountain. Call Nova Service.
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:

8.1.2 Pump Test

The pump test verifies that the pump is working properly and is capable of providing suction to the sensor module and probe. If already on the System Test screen, skip to Step 4., otherwise, start with Step 1 below.
1. From the Home Screen, press Menu (soft Key)
2. Press Service (soft key).
3. From the Service screen, select System Test. Then press Enter.
4. Move the sampler to the syringe position by using the arrow keys at the bottom right corner of the keypad.
a. Select Sampler. b. Press Enter. c. Select Syringe. d. Press Enter.
5. Open the waste valve by using the arrow keys at the bottom right corner of the keypad.
a. Select Waste Valve. b. Press Enter. (Valve toggles to open position if
closed.)
6. Turn pump to fast speed by using the arrow keys at the bottom right corner of the key pad.
a. Select Pump. b. Press Enter. c. Select Fast. d. Press Enter.
7. To check the W-line for aspiration, fill a 2 mL sample cup with water.
8. Remove the tubing from the waste port of the reference electrode and place the waste tubing in the cup of water.
9. Observe for water being aspirated.
If water is aspirated during the pump test and not in the water (flow) test, there is a blockage or leak between the reference electrode, the sensor module, air detector, and probe. Proceed to the Flow Path Flush Procedure. If water is not aspirated during the pump test, the problem is with the W-line, waste valve, pump, or waste bag. Proceed to Step 10 below.
8-5
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:
10. If water is not aspirated, lift the waste tubing out of the waste valve. If flow starts, the tubing may have been collapsed or pinched, or the waste valve may be dirty or not working. Correct the problem. (Clean, if necessary, or call Nova service)
11. If flow does not start, remove the waste line from the front panel connection below the pump. Be ready to catch any water that is expelled from the tubing. If water flows from this tube there is a plug in the internal waste tubing or waste bag of the reagent pack.
12. Turn off the pump by using the arrow keys at the bottom right corner of the keypad.
a. Select the Pump. b. Press Enter. c. Select Stop. d. Press Enter.
13. Connect a syringe (Probe Cleaning Syringe PN 02702) filled with water, that has a length of tubing attached, to the waste port on the front panel. Attempt to inject water through the line to clear the plug.
14. If there is not flow, remove the reagent pack to ensure the problem is not a poorly seated or damaged waste connec­tion. Should the problem continue call Nova Service.
15. If the problem is resolved, reconnect the correct ends of the W-line to the waste port on the front panel and on the waste port of the reference electrode. Also position the tubing back into the waste valve.
16. Rerun the water test to verify proper flow.
8-6
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:

8.1.3 Flushing Sensor Module, Air Detector, and Probe

If already on the system test screen, skip to Step 4, otherwise, start with Step 1 Below.
1. From the Home Screen, press menu (soft Key)
2. Press Service (soft key).
3. From the Service screen, select System Test. Then press Enter.
4. Move the sampler to the syringe position, if not already there, by using the arrow keys at the bottom right corner of the keypad.
a. Select Sampler. b. Press Enter. c. Select Syringe. d. Press Enter.
5. Remove the W-line tubing from the waste port of the reference electrode.
6. Connect a syringe (Probe Cleaning Kit PN 02702) filled with water, that has a length of tubing attached, to the waste port of the reference electrode (see Figure 8.1).
NOTE:
The probe should still be in the syringe position.
WARNING:
Water may spray out of the probe.
7. Gently push the water through the reference electrode, sensor module, air detector, and probe. If the flow path is clear, the plunger can be depressed with no resistance, otherwise, resis­tance will be felt. The blockage may be dislodged by pushing the water through the flow path. If not, flush the suspected individual components by proceeding to the Flushing the Reference Electrode procedure (Section 5.5.4).
8. When the blockage is cleared, a nice steady stream of water should be seen coming out of the probe.
9. Rerun the water test to verify proper flow. The Sampler should be in the syringe position.
a. Turn pump to fast speed. b. Open waste valve. c. Insert the probe into a cup of water and observe for
water being aspirated. If water flowed freely during the flush procedure but still does not pass the water test, then air is entering the system preventing aspiration.
8-7
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:
a. Verify interconnect tubing is properly connected.
Reseat if necessary.
b. Inspect the seal between the reference electrode
and sensor module. Replace interconnect tubing, if necessary.
c. Inspect each sensor to assure internal fill solution
is at the proper level. If not replace cap.
d. Inspect pH sensor to assure glass tip is not broken.
Replace sensor, if necessary
10. Rerun the water test.
11. Once proper flow has been established, turn the pump off. Use the arrow keys at the bottom right corner of the keypad.
a. Select pump. b. Press Enter. c. Select Stop. d. Press Enter.
12. Move the sampler back to the Home position. Use the arrow keys at the bottom right corner of the keypad.
a. Select Sampler. b. Press Enter. c. Select Home. d. Press Enter.
13. Close the waste valve by using the arrow keys at the bottom right corner of the keypad.
a. Select waste valve. b. Press Enter (valve toggles to closed position if
open).
14. Press Home (soft key).
8-8

8.1.4 Flushing the Reference Electrode

1. Remove the W-line and R-line from the reference electrode.
2. Lift the reference electrode off the sensor module by pulling up on the locking pin that is on the top of the reference electrode.
3. Connect a syringe (Probe Cleaning Syringe PN 02702) filled with water, that has a length of tubing attached to it, to the W-port of the reference electrode. While covering the R­port with your finger, push water through the electrode so that the water flows out of the flow cell port. Repeat by attaching the syringe to the R-port and covering the W-port while pushing water through the electrode.
4. Reinstall the reference electrode.
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:

8.2 Flushing the Entire Flow Path

Flush flow path when a flow problem has been detected. Typical flow problem error codes are Insufficient Sample, No Std A, B, C, or D Fluid. Flush individual components only if the problem is not cleared by a flow path flush.
WARNING:
Use appro­priate protective face shields, gloves, etc., when attempting to flush, clean, or handle any components of the analyzer which come in contact with blood specimens.
DWG #10-1025A
Figure 8.1 Flow Path Air Flush
8-9
8 Flow Test & Flushing Stat Profile pHOx Training ManualFlushing the Flow Path Checklist
NOTES:
WARNING:
when attempting to flush, clean, or handle any components of the analyzer which come in contact with blood specimens.
Step 1 Initiate Flowpath maintenance.
Press Menu (soft key) on the Home screen.
The Operational Menu screen appears.
Select Service Menu (soft key); enter password, if enabled.
Select System Test. Press Enter.
Select Sampler by moving the cursor with the arrow keys.
Select syringe by moving the cursor with the arrow keys.
Open the door.
Use appropriate protective face shields, gloves, etc.,
Then press Enter.
Then press Enter.
Step 2 Prepare syringe.
Disconnect the W-line from the side of the reference elec­trode.
Attach a length of tubing to the end of the Probe Cleaning Syringe (PN 02702).
Pull back the plunger of the syringe to fill it with air.
Attach the syringe tubing to the W-line port. (Refer to Figure
8.1, if necessary.)
Step 3 Flush air through flow path.
Press syringe plunger in slowly. If flow path is clear, the plunger can be depressed with no resistance. If a clot or other obstruction is present, resistance will be felt. If necessary, repeat the procedure using deionized or distilled water.
8-10
8 Flow Test & Flushing Stat Profile pHOx Training Manual
Flushing the Flow Path Checklist
NOTES:
Step 4 Determine results.
Continue to Step 5 if the obstruction has been flushed free.
If not, repeat flush procedure using deionized or distilled
water or contact key operator of lab for assistance.
Step 5 End flush procedure if no other components need
flushing.
Remove syringe.
Reattach the W-line to the reference electrode.
Press Exit (soft key) to return to the home position.
Close the door.
Press Home (soft key) twice to return to Home screen.
Step 6 Prime the flow path.
From the Home screen, press Menu (soft key).
Select Flowpath/Probe maintenance. Then press Enter.
Press Continue (soft key) to prime the flow path.
Step 7 Calibrate.
A pop-up screen appears to question - Do you want to recalibrate? Yes or No.
Press Yes; calibration takes place.
NOTE:
screen, press Calibrate (soft key) to recalibrate analyzer.
If No is pressed, you can still recalibrate. From the Home
Contact key operator (or lab) for assistance if problem is not resolved.
8-11
8 Flow Test & Flushing Stat Profile pHOx Training Manual
NOTES:
8.3 Reference Solution Test
1. Press Menu (soft key) on the Home screen.
2. The Operational Menu screen appears.
3. Select Service Menu (soft key); enter password, if enabled.
4. Select System Test. Press Enter.
5. Using the arrow keys, highlight Ref Valve.
6. Press Enter to toggle from Closed to Open.
7. Using the arrow keys, highlight Pump.
8. Press Enter and select Fast using the arrow keys.
9. Press Enter.
10. Open the door and disconnect the R-tubing from the refer­ence electrode.
11. Hold the R-tubing over a piece of gauze or sample cup to catch any solution that may flow out of the tubing.
12. Reference solution should flow at a minimum of 1 drop per second.
13. If there is flow, turn off the pump. If there is no reference solution dripping out of the tubing, verify there is reference solution or a blockage. Check for blockage in the reference tubing or the pump harness tubing. If necessary, replace the reagent pack.
8-12

9 Maintenance Stat Profile pHOx Training Manual

NOTES:
9 Maintenance

9.1 Logging Maintenance

The Stat Profile pHOx Maintenance Log (PN 22362) contains each of the following types of log sheets — maintenance, quality control, and service record.

9.1.1 Maintenance Log Sheets

Allows the operator to record the daily and preventive maintenance procedures. A sample is shown.
9-1
9 Maintenance Stat Profile pHOx Training Manual
GOLECNANETNIAM
_____________________#rezylanA____________raeY/htnoM
123456789 01112131415161
dedeensa)40721NP(retaeherPelpmaSezinietorpeD
ylkeeWsevlaVhcniPnignibuTnoitisopeR
rehtO
GOLATADECNAMROFREP
123456789 01112131415161
:ylkeeWsepolSedortcelEdroceR
Hp
P
OC
2
P
O
2
aN
+
S
O
2
tcH
MAINTENANCE LOG
Month/Year____________ Analyzer #_______________________
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Deproteinize Probe/Sensor Module (Use PN 12704) as needed
Reposition Tubing in Pinch Valves Weekly
Clean the
S
O
2
Sensor Every 3 Months
Initial and Date When Done. Tech's Initial's_______ Date_______
Change the Pump Tubing Every 3 Months
Initial and Date When Done. Tech's Initial's_______ Date_______
Change the R-Line and W-Line Every 3 Months
Initial and Date When Done. Tech's Initial's_______ Date_______
Other
PERFORMANCE DATA LOG
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Record Electrode Slopes Weekly:
pH
P
CO
2
P
O
2
Na
+
S
O
2
(LED1)
Hct
NOTES:
Figure 9.1 Maintenance Schedule and Performance Data Log
9-2
9 Maintenance Stat Profile pHOx Training Manual
GOLLORTNOCYTILAUQ
slortnoC
egnaRyassAyaD 123456789 01112131415161
woLnaeMhgiH
Hp
06722NP1leveL
P
OC
2
_____________#toL
P
O
2
_________etaD.pxE
S
O
2
tcH
bH
16722NP2leveL
Hp
_____________#toL
P
OC
2
_________etaD.pxE
P
O
2
Hp
26722NP3leveL
P
OC
2
_____________#toL
P
O
2
_________etaD.pxE
S
O
2
tcH
bH
SLAITINIS'NAICINHCET
NOTES:
Figure 9.2 Quality Control Log
9-3
9 Maintenance Stat Profile pHOx Training Manual
NOTES:
SERVICE RECORD
NOTE: Record serial/lot number of components
Date Nature of Problem Resolution/Comments
9-4
Figure 9.3 Service Record

10 Routine Maintenance Stat Profile pHOx Training Manual

NOTES:
10 Routine Maintenance

10.1 Overview

How often maintenance procedures are done is in part dependent on how much the instrument is in use. Each schedule includes daily, weekly, monthly, every 3 months, and annual procedures. The Stat Profile pHOx Maintenance Log helps you remember when to do each procedure. It contains the daily and preventative maintenance log sheets, quality control log sheets, and service record log sheets.
NOTE:
Make sure to use the correct maintenance log sheet. The
maintenance log sheets list procedures due for each month.

10.2 General Maintenance Procedures

See the Maintenance Planner (Quick Reference) for daily, weekly, monthly, quarterly, and annual procedures. Changes in the timing of these procedures should be made according to the number of samples per day, the quality of the samples, and the lab circumstances encountered. Consult with the Nova Hot Line at 1-800-545-NOVA if you have any questions.
10-1
10 Routine Maintenance Stat Profile pHOx Training Manual
NOTES:

10.2.1 Suggested Daily Start-up Procedure

“Housekeeping” Procedures
Discuss modifying this to meet lab situation.
Verify reagent and control levels. Review Maintenance Logs for procedures due.
Analyze Controls (Level 1, 2, and 3).
Document QC.
10.2.2Flowpath Cleaning/Deproteinizing
1. Deproteinizing Solution (PN 12704) - used for routine cleaning) is packaged in ampules. Use this solution when routine clean­ing is required.
The cleaning cycle:
Removes protein buildup
Frequency depends on the number of samples or the condi­tion of those samples
Should be performed when the hematocrit sensor is uncalibrated
Must be followed by a full calibration
Takes 5 minutes
Can be run at the end of a shift or when most convenient
Remove the sodium sensor and install a blank.
10-2
See Flowpath Cleaning Procedure Checklists.
10 Routine Maintenance Stat Profile pHOx Training Manual
Flowpath Cleaning Procedure Checklist
NOTES:
Step 1 Remove the Na+ sensor.
Select Flowpath/Probe Maintenance; wait until system is finished removing fluid from the flow path.
Open the door and remove the Na+ sensor. (Squeeze the front and rear of the black body {sensor clip} and gently pull it out of the sensor module.)
Replace it with a blank sensor.
Press Cancel.
At pop-up screen, press Cancel again.
Step 2 Begin sample preheater cleaning procedure.
Press Menu (soft key on the Home screen) to display Operational Menu.
Select Flowpath Cleaning.
Probe extends for aspiration.
Step 3 Prepare solution.
Open an ampule of Deproteinizing Solution (PN 12704).
Step 4 Present sample and begin cycle.
Position the ampule over the extended probe.
Press Continue (soft key) to aspirate the solution.
CAUTION:
the ampule.
Do not allow the sample probe to touch the bottom of
When the system signals by an audible alarm, remove the ampule of deproteinizing solution.
A pop-up screen appears when the aspiration is completed.
10-3
10 Routine Maintenance Stat Profile pHOx Training Manual
NOTES:
Flowpath Cleaning Procedure Checklist
Step 5 Sequence begins.
Press Analyze.
Discard any unused solution. Do not save it for use later.
At the completion of the cycle, a pop-up screen appears questioning to recalibrate. Press NO.
Step 6 Replace the Na+ sensor.
Press Menu.
Select Flowpath/Probe Maintenance.
Remove blank sensor.
Install the Na+ sensor.
Press Continue.
Step 7 Calibrate.
At the completion of the Flowpath Priming Cycle, a pop-up screen appears questioning to Recalibrate. Press Yes.
NOTE:
Menu screen. Press Home (soft key). To calibrate on this screen, press Calibrate (soft key).
If no is pressed, the analyzer returns to the Operational
10-4
10 Routine Maintenance Stat Profile pHOx Training Manual
NOTES:

10.2.3Flow Cell Conditioning

Decreases bubble hang-up in the flow cell and at the sensor tips by wetting the flow cell.
Coats the sensor membranes with protein to improve their performance.
Perform during periods of infrequent sampling and after flow cell replacement, PO2 or PCO2 sensor maintenance, and troubleshooting. Takes 5.5 minutes.
See Flow Cell Conditioning Cycle Checklist.
10-5
10 Routine Maintenance Stat Profile pHOx Training Manual
NOTES:
Flow Cell Conditioning Checklist
Step 1 Prepare sample.
Fill a 2 mL sample cup 1/2-full with heparinized whole blood (or serum) or select a syringe containing at least 1 mL of heparinized whole blood.
Step 2 Begin flow cell conditioning cycle.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Select Sensor Conditioning.
Press Enter
Step 3 Present sample and begin sequence.
Pour whole blood (or serum) into a cup.
Position the cup over the extended probe.
CAUTION:
cup.
Do not allow the sample probe to touch the bottom of the
Press Continue (soft key).
After the audible alarm, remove the cup.
Press Analyze. The probe moves downward.
The sequence begins. Whole blood fills the entire flow path.
Conditions for 5 minutes. Then flushes the flow cell.
Step 5 Calibrate.
At the completion of the cycle, a pop-up screen appears questioning to recalibrate. Press Yes to recalibrate.
NOTE:
Menu screen. Press Home (soft key). To calibrate on this screen, press Calibrate (soft key).
If no is pressed, the analyzer returns to the Operational
10-6

11 Maintaining Sensors Stat Profile pHOx Training Manual

NOTES:
11 Maintaining Sensors

11.1 Overview

Correctly functioning sensors are key to obtaining accurate results.
Maintenance is essential for sensors to function correctly.
Frequency of conditioning or replacement is dependent on the number of samples per day and will vary from lab to lab.
Monitoring trends and shifts or both in control values and sensor slope values using your logbook is the best way to determine an appropriate maintenance interval for your laboratory.
Use the maintenance checklists to do preventive mainte­nance or to correct problems.
Before replacing sensors, membranes, etc.:
Press Menu (soft key) from the Home screen. Then select Flowpath/Probe Maintenance. Press Enter.
Remove sensor and shake down.
Dry sensor and flow cell chamber.
Reseat sensor.
Perform calibration.
Table of Contents This section contains maintenance procedures for the Manual Stat Profile pHOx Analyzer sensors. These procedures
are done as part of a preventive maintenance program or when a problem occurs. The procedures can be used during training or for reviewing on your own.
Maintaining Sodium (Na+).................................... 11-3
pH Sensors............................................................ 11-5
Maintaining PCO2 Sensor ....................................11-8
Maintaining PO2 Sensor..................................... 11-12
Maintaining SO2................................................11-17
WARNING
Use appropriate protective face shields, gloves, etc., when attempting to flush, clean, or handle any compo­nents of the analyzer which come in contact with blood specimens.
11-1
11 Maintaining Sensors Stat Profile pHOx Training Manual
NOTES:
Na+ Maintenance Checklist
When Change the Sodium (Na+ ) sensor when it has problems
with slope, instability, overload, or drift.
What You Need Sodium sensor (PN 21523)
Stat Profile pHOx Controls
Stat Profile pHOx Maintenance Log (PN 22362) Cotton swabs
Step 1 Record slope for comparison purposes.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the Na Calibration & Analysis Data is displayed.
Press Print (soft key) to print the slope information. Save for comparison in Step 7.
Press Exit (soft key) twice to return to the Operational Menu screen.
11-2
Step 2 Begin Na+ maintenance.
Select Flowpath/Probe Maintenance.
Wait until system is finished removing fluid from the flow path.
Open the door.
Remove the Na+ sensor by squeezing the front and rear of the black body (sensor clip) and gently pulling it out of the sensor module.
CAUTION: Do Not condition the Na
+
sensor; you will destroy the sensor.
11 Maintaining Sensors Stat Profile pHOx Training Manual
Na+ Maintenance Checklist
NOTES:
Step 3 Clean and dry the sensor module cuvette with a cotton
swab.
Step 4 Replace with a new sensor.
Insert a new Sodium Sensor (PN 21523).
STOP
TRAINING CLASS ONLY — Continue with the next sensor.
ø
Step 5 Prime the flow path.
Close the door.
Press Continue (soft key) to prime the flow path.
Step 6 Calibrate.
At the completion of the cycle, a pop-up screen appears questioning to recalibrate. Press Yes to recalibrate.
Calibration begins.
NOTE:
Menu screen. Press Home (soft key). To calibrate on this screen, press Calibrate (soft key).
If no is pressed, the analyzer returns to the Operational
11-3
11 Maintaining Sensors Stat Profile pHOx Training ManualNa+ Maintenance Checklist
NOTES:
Step 7 Review calibration status and slope performance
number.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the Na+ Calibration & Analysis Data is displayed. Compare current versus previous slope performance num­bers saved from Step 1.
If higher and within range, continue with Step 8.
or
If not, replace sensor.
Step 8 Verify sensor performance by running controls.
Run QC samples, 3 levels.
CAUTION:
Controls should be stored at approximately 25°C.
Verify controls are in range by comparing with data insert sheet or within lab criteria.
Record procedure and document quality control results.
11-4
11 Maintaining Sensors Stat Profile pHOx Training Manual
When Condition the pH sensor when the sensor fails to cali-
brate because of low slope or when pH control results are high.
What You Need pH Sensor Conditioning Solution (PN 23397)
Sensor Conditioning Holder (PN 09458) Stat Profile pHOx Controls Multipack (PN 22763)
Stat Profile pHOx Maintenance Log (PN 22362) Lint-free tissues Cotton swabs Deionized or distilled water
pH Maintenance Checklist
NOTES:
Step 1 Record slope for comparison purposes.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the pH Calibration & Analysis Data is displayed.
Press Print (soft key) to print the slope information. Save for comparison in Step 7.
Press Exit (soft key) twice to return to the Operational Menu screen.
Step 2 Begin pH maintenance.
Select Flowpath/Probe Maintenance.
Wait until system is finished removing fluid from the flow path.
Open the door.
Remove the pH sensor by squeezing the front and rear of the black body (sensor clip) and gently pulling it out of the sensor module.
11-5
11 Maintaining Sensors Stat Profile pHOx Training ManualpH Maintenance Checklist
NOTES:
Step 3 Condition glass tip, replace, and reconnect sensor.
(Troubleshooting Only)
Fill bottom chamber of sensor conditioning holder (PN
09458) with pH conditioning solution (PN 23397). Immerse sensor tip in conditioning solution to soak. The pH sensor is conditioned for 15 minutes.
Remove sensor. Rinse tip with deionized water.
Dry tip with lint-free tissue. Dry flow cell with cotton swab covered with lint-free tissue.
Reinsert sensor.
STOP
TRAINING CLASS ONLY — Continue with the next sensor.
ø
Step 4 Prime the flow path.
Close the door.
Press Continue (soft key) to prime the flow path.
When the priming is completed, a pop-up screen appears to question for recalibration. Press No.
Step 5 Condition the flow cell.
From the Operational Menu, select Sensor Conditioning.
NOTE:
Conditioning cycle.
All sensors must be in the flow cell during a Flow Cell
Fill a 2 mL sample cup 1/2-full with whole blood.
Insert the probe into the blood.
Press Continue (soft key) to aspirate the blood.
After the audible alarm, remove the cup.
Press Analyze on the pop-up screen. The probe moves downward. The sequence begins. Whole blood fills the entire flow path. Conditioning is a 5.5 minute cycle. Then the system flushes the flow cell.
11-6
11 Maintaining Sensors Stat Profile pHOx Training Manual
pH Maintenance Checklist
NOTES:
Step 6 Calibrate.
At the completion of the cycle, a pop-up screen appears questioning to recalibrate. Press Yes to recalibrate.
Calibration begins.
NOTE:
Menu screen. Press Home (soft key). To calibrate on this screen, press Calibrate (soft key).
If no is pressed, the analyzer returns to the Operational
Step 7 Review calibration status and slope performance
number.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the pH Calibration & Analysis Data is displayed. Compare current versus previous slope performance num­bers saved from Step 1.
If higher and within range, continue with Step 8.
or
If not, replace sensor.
Step 8 Verify sensor performance by running controls.
Run QC samples, 3 levels.
CAUTION:
Controls should be stored at approximately 25°C.
Verify controls are in range by comparing with data insert sheet or within lab criteria.
Record procedure and document quality control results.
11-7
11 Maintaining Sensors Stat Profile pHOx Training Manual
NOTES:
P
CO2 Maintenance Checklist
When Change the membrane for low slope, controls out of
range, and PCO2 instability, drift, or overload error codes.
What You Need PCO
Membraned Cap Kit (PN 25048)
2
Stat Profile pHOx Controls Multipack (PN 22763)
Stat Profile pHOx Maintenance Log (PN 22362) Lint-free tissues Cotton swabs Deionized or distilled water Heparinized whole blood
Step 1 Record slope for comparison purposes.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the PCO2 Calibration & Analysis Data is displayed.
Press Print (soft key) to print the slope information. Save for comparison in Step 12.
Press Exit (soft key) twice to return to the Operational Menu screen.
11-8
Step 2 Begin PCO2 Sensor Membraning and Replacement
Select Flowpath/Probe Maintenance.
Wait until system is finished removing fluid from the flow path.
Open the door.
Remove the PCO2 Sensor by squeezing the front and rear of the black body (sensor clip) and gently pulling it out of the sensor module.
11 Maintaining Sensors Stat Profile pHOx Training Manual
P
CO2 Maintenance Checklist
NOTES:
Step 3 Separate cap and membrane from sensor.
Unscrew the used membrane cap from the sensor body and discard it. Dispose of the plastic cap as a biohazard.
CAUTION:
If installing a new sensor, carefully remove the shipping cap before installing a new membrane cap. Do not touch the electrical contacts of the sensor.
Step 4 Replace membrane cap on sensor.
NOTE:
filling solution. Do not add or remove any Internal Filling Solu­tion. The prefilled level of solution is all that is needed to operate the analyzer successfully. Adding or subtracting from the prefilled level will effect the biosensor's performance.
The prefilled PCO2 cap does not require any additional
Open foil pouch by tearing at score mark on the bottom of the package.
Take a new PCO2 Membrane Cap (PN 25048) that is prefilled with internal filling solution. Hold the shipping plug end of the membrane cap and shake it gently, as if shaking down a thermometer, to ensure that the air bubble is at the threaded end of the membrane cap.
Unscrew the black shipping plug from the cap. Screw the cap onto the sensor. Discard the black shipping plug. Do not tilt the membrane cap when installing onto the sensor; the internal filling solution may drip out.
DWG #10-1066A
Figure 11.1 Shaking Down
Membrane Cap
Shipping Plug
See air gap
after shaking
Internal filling solution
DWG #10-1067A
CAUTION:
Do not touch the surface of the membrane.
Step 5 Check for air bubbles at the tip of sensor.
If new sensor, remove from the foil pack.
Hold the sensor with the cap downward. With a wrist­snapping motion, shake the sensor down to move air bubbles to the back of the sensor.
With the sensor tip still downward, observe the tip for bubbles. If bubbles are present, tap the sensor with a finger to loosen the bubbles and again shake the sensor down. Repeat if necessary.
CAUTION:
Do not point sensor tip upward.
Figure 11.2 Membrane Cap
11-9
11 Maintaining Sensors Stat Profile pHOx Training Manual
NOTES:
123456
P
CO2 Maintenance Checklist
Step 6 Dry the sensor and membrane.
Sensor
Vent Hole
Vent Cover
Figure 11.3 PCO2 sensor
with vent hole covered
Wipe the sensor body and cap dry.
DWG #10-1068A
Remove a single vent cover from the blue backing and place over the vent hole.
Press firmly to ensure a good seal.
Step 7 Clean the sensor module cuvette with a cotton swab.
Step 8 Replace and reconnect sensor.
Insert the sensor into the sensor module by sliding the sensor body into the sensor module until the sensor clips into place.
TRAINING CLASS ONLY —Continue with the next sensor.
STOP
ø
Step 9 Prime the flow path.
Close the door.
Press Continue (soft key) to prime the flow path.
When the priming is completed, a pop-up screen appears to question for recalibration. Press No. The analyzer returns to the Operational Menu.
Step 10 Condition the sensor module.
NOTE:
All sensors must be in the sensor module during a Sensor
Conditioning cycle.
Fill a 2 mL sample cup 1/2-full of whole blood.
Select Sensor Conditioning from the Operation Menu and press Enter.
Immerse the probe in the whole blood.
Press Continue (soft key) to aspirate the blood.
After the audible alarm, remove the cup.
11-10
11 Maintaining Sensors Stat Profile pHOx Training Manual
Press Analyze on the pop-up screen. The probe moves downward. The sequence begins. Whole blood fills the entire flow path. Conditioning is a 5.5 minute cycle. Then the system flushes the flow cell.
P
CO2 Maintenance Checklist
NOTES:
Step 11 Calibrate.
At the completion of the cycle, a pop-up screen appears questioning to recalibrate. Press Yes to recalibrate.
Calibration begins.
NOTE:
Menu screen. Press Home (soft key). To calibrate on this screen, press Calibrate (soft key).
If no is pressed, the analyzer returns to the Operational
Step 12 Review calibration status and slope performance number.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the PCO2 Calibration & Analysis Data is displayed. Compare current versus previous slope performance num­bers saved from Step 1.
If higher and within range, continue with Step 13.
or
If not, replace sensor.
Step 13 Verify sensor performance by running controls.
Run QC samples, 3 levels.
CAUTION:
Controls should be stored at approximately 25°C.
Verify controls are in range by comparing with data insert sheet or within lab criteria.
Record procedure and document quality control results.
11-11
11 Maintaining Sensors Stat Profile pHOx Training Manual
NOTES:
P
O2 Maintenance Checklist
When You will need to polish and remembrane the oxygen
(PO2) sensor if you are experiencing a problem such as slope out of range, controls out of range, or drift error codes on analysis.
What You Need PO2 Premembraned Cap Kit (PN 21795)
Stat Profile pHOx Controls Multipack (PN 22763)
Stat Profile pHOx Maintenance Log (PN 22362) Polishing paper Lint-free tissues Cotton swabs Deionized or distilled water Heparinized whole blood
Step 1 Record slope for comparison purposes.
Press Menu (soft key on the Home screen) to display the Operational Menu screen.
Press Service (soft key) and enter your password to display the Service Menu screen.
Select Sensor Subsystem, press Enter, and press Next Page (soft key) until the PO2 Calibration & Analysis Data is displayed.
Press Print (soft key) to print the slope information. Save for comparison in Step 13.
Press Exit (soft key) twice to return to the Operational Menu screen.
Step 2 Begin PO2 Sensor Polishing, Membraning, and Replace-
ment
Select Flowpath/Probe Maintenance.
Wait until system is finished removing fluid from the flow path.
Open the door.
Remove the PO2 Sensor by squeezing the front and rear of the black body (sensor clip) and gently pulling it out of the sensor module.
11-12
11 Maintaining Sensors Stat Profile pHOx Training Manual
P
O2 Maintenance Checklist
NOTES:
Step 3 Separate cap and membrane from sensor.
Unscrew the used membrane cap from the sensor body and discard it. Dispose of the plastic cap as a biohazard.
Step 4 Polish the sensor (Troubleshooting only).
If troubleshooting, polish the sensor: if not, skip to Step 5.
Take a polishing paper and place a couple of drops of deionized water onto it.
Hold the PO2 polishing paper so that the tip of your index finger provides light pressure against the back of the paper.
Gently polish the sensor tip on the paper, move the tip in a circular motion for about 10 seconds.
CAUTION:
bench top.
Discard the used polishing paper.
Wipe the sensor tip with a lint-free tissue soaked in deion-
Never polish the sensor on a hard surface such as a
ized water.
Step 5 Replace membrane cap on sensor.
Open foil pouch by tearing at score mark on the bottom of the package.
Take a new PO2 Premembraned Cap (PN 21795) that is prefilled with internal filling solution, and shake it gently to ensure that the solution is away from the threaded end and at the membrane end. Unscrew and discard the shipping plug from the cap.
Insert the sensor body straight down (vertical position only to ensure no loss of internal filling solution) into the filled cap and screw the cap onto the sensor body.
CAUTION:
Do not touch the surface of the membrane.
The prefilled PO2 cap
does not require any
additional filling
solution. Do not add
or remove any Internal
Filling Solution. The
prefilled level of
solution is all that is
needed to operate
the analyzer
successfully. Adding
or subtracting from
the prefilled level will
effect the biosensor's
performance.
11-13
11 Maintaining Sensors Stat Profile pHOx Training Manual
NOTES:
P
O2 Maintenance Checklist
Step 6 Check for air bubbles at the tip of sensor.
Hold the sensor with the cap downward. With a wrist­snapping motion, shake the sensor down to move air bubbles to the back of the sensor.
With the sensor tip still downward, observe the tip for bubbles. If bubbles are present, tap the sensor with a finger to loosen the bubbles and again shake the sensor down. Repeat if necessary.
CAUTION:
Do not point sensor tip upward.
Step 7 Dry the sensor and membrane.
Dry the sensor with a lint-free tissue.
Do not touch the membrane.
Step 8 Clean the sensor module cuvette with a cotton swab.
Step 9 Check for air bubbles, replace and reconnect sensor.
Remove air bubbles by repeating Step 6.
Insert the sensor into the sensor module by sliding the sensor body into the sensor module until the sensor clips into place.
11-14
STOP
ø
TRAINING CLASS ONLY —Continue with the next sensor.
Loading...