Equipment Description:Model - Stat Profile® pHOx® Analyzer
Laboratory Equipment
100 - 120 / 220 - 240 V, 130 W, 50/60 Hz
Protection Class I
Year of Manufacture:See serial number showing date.
Applicable Directives:73/23/EEC, Low Voltage Directive
Laws for electrical equipment within certain voltage limits
89/336/EEC, EMC Directive
Laws relating to electrical magnetic compatibility
Applicable Standards:EN 61010-1/A2:1995, EN 61010-2-010/A1:1996
Equipment for Measurement, Control and Laboratory use
EN 50081-1:1992
Electromagnetic Compatibility. Generic Emission Standard
EN 50082-1:1992
Electromagnetic Compatibility. Generic Immunity Standard
Type Examination Certificate:GS-mark License Number AL 01 06 20747 011
TUV Product Service, Munich/Germany
Authorized by : ________________________ 3/15/99
Francis C. ManganaroDate
President
Nova Biomedical Corporation
PrefaceStat Profile pHOx Reference Manual
Nova Stat Profile
®
pHOx® Reference Manual
Part Number and Ordering Information
The Stat Profile® pHOx® Reference Manual (PN 22363) can be ordered from Nova Biomedical
Order Services. Write or call:
Nova Biomedical
200 Prospect Street
Waltham, MA 02454-9141
U.S.A.
Telephone:1-800-458-5813
FAX:(781) 893-6998 (in the U.S.A.) or
(781) 899-0417 (outside the U.S.A.)
Technical Assistance
For technical assistance, call Nova Biomedical Technical Services at:
Telephone:1-800-545-NOVA or
FAX:(781) 894-0585
Trademarks and Patents
Stat Profile® and pHOx® are registered trademarks of Nova Biomedical.
The Stat Profile pHOx Analyzer is covered by the following patents:
U.S. Patent No. 4,686,479, U.S. Patent No. 5,578,194.
Copyright
(781) 894-0800
Printed in the U.S.A. Copyright 2001, Nova Biomedical, Waltham, MA 02454-9141.
i
PrefaceStat Profile pHOx Reference Manual
Note: U.S. Legislative Requirements — Nova Analyzers
The new legislative requirements for clinical laboratories have established a new regulatory
environment and have changed the responsibility that Nova Biomedical has regarding the
proper operation and control of Nova analyzers. With the advent of the requirements that the
end user specifically follows all directions for use from the manufacturer, Nova becomes a
partner with the user and becomes responsible for providing instructions that will virtually
guarantee that the analyzer is in control. This requires Nova to provide instructions on
maintenance and control which are virtually foolproof in regards to assuring that an analyzer
is in control.
In addition to asking that all operators perform the maintenance and calibration procedures
described in this manual, we also recommend that only Nova reagents, calibrators, cleaning
agents, and controls be run through the analyzers. These are the only materials that we have fully
characterized and tested. We have neither tested the performance nor determined the long term
effects of products provided by others for use on our instruments. Confirming that an analyzer
is in control by using reagents and controls not certified by Nova Biomedical can not be
guaranteed by Nova, and Nova will not be responsible for the consequences. Users who elect
to use reagents, standards and/or controls from other producers should confirm by the
appropriate testing protocols from those other manufacturers that Nova’s instrumentation is in
control and in compliance with all legislative requirements. This testing verification should
also include a means of troubleshooting when the instrument is not in control, since Nova
Biomedical cannot, in the present environment, provide that service. Since Nova Biomedical
does not know the formulations of the reagents and controls used by other manufacturers,
including the levels of surfactants, preservatives, viscosity adjusters, and other additives that
may be used, there is no way for Nova to be certain that the results obtained on controls or
patient samples will be correct when these products are used. There is also no way to tell
whether there will be long term or short term adverse effects on sensor performance, sensor
life, tubing or flow cell degradation, or on data quality/accuracy or reliability.
Nova Biomedical manufactures totally integrated systems to assure proper analyzer performance. We cannot characterize all products that might be used on Nova analyzers, nor can we
control changes that other manufacturers might choose to make to their products. Therefore,
use of these products with a Nova analyzer must be the responsibility of the individual end user
and of that manufacturer rather than Nova.
Under these new legislative requirements, user modification of an in vitro diagnostic device,
is also regulated. Modification of the product includes use of untested products or products
which have not been tested to determine substantial equivalency and safety and effectiveness
in use with Nova analyzers.
Nova Biomedical’s Reference Manuals include the specific information needed for the proper
maintenance, qualification, and operation of Nova Biomedical Analyzers. If the user follows
the directions in those manuals, they should have very few problems. In the case where they
need assistance from Nova Biomedical in troubleshooting their systems, we will stand ready
to assist them in any and all ways necessary.
This manual complies with the NCCLS guidelines Vol. 4, Section 2.1. As a user, you
can copy and compile these above sections into one compact NCCLS Nova pHOx Manual or
use the cross-reference to find the appropriate section in the Nova Stat Profile pHOx
Reference Manual.
This manual provides all necessary instructions for the routine operation and maintenance of
the Stat Profile pHOx Analyzer. Please read this manual carefully. It has been prepared to help
you attain optimum performance from your Stat Profile pHOx Analyzer.
1 Introduction
1. Intro.
WARNING: Blood samples and blood products are potential sources of hepatitis and
other infectious agents. Handle all blood products and flow path components (wasteline, capillary adapter, probe, sensor module, etc.) with care. Gloves and protective
clothing are recommended.
!
This section introduces the Stat Profile pHOx Analyzer and covers requirements, tests
performed, procedural limitations, clinical utility, and sample handling.
1.1Installation
This section covers the installation requirements and assembly procedures for the Stat Profile
pHOx Analyzer.
1.1.1 Requirements
Working Area Requirements:
Keep the working area around the system free of dirt, corrosive fumes, vibration, and excessive
temperature changes. Ambient operating temperature is 15 °C to 30 °C (59°F to 86 °F). Operate
at humidity of 0 to 95% without condensation.
NOTE:
Analyzer and means refer to the manual.
NOTE:
equipment for you.
This International Caution Label appears on the rear of the pHOx
Under the Warranty, a Nova service representative will install this
Electrical Requirements:
A grounded, 3-wire receptacle within 5 feet of the system is required for operation. The U.S.
models require a 120 Volt AC line at 50/60 Hz frequency. The analyzer can be operated at
100 - 120; 220 - 240 Volt AC 50/60 Hz.
Fuse requirements:
•2 Amp Time Delay (SB 2A or T2A) at 100 - 120 Volt AC line.
•1 Amp Time Delay (T1A) at 220 - 240 Volt AC line.
1-1
Stat Profile pHOx Reference Manual
1.2Intended Use, Tests Performed, and Clinical Utility
1. Intro.
Intended Use
The Stat Profile pHOx Analyzer is intended for in vitro diagnostic use by health care
professionals in the quantitative determination of pH,
hematocrit (Hct), hemoglobin (Hb) in heparinized whole blood.
Measured Parameters
pH, PCO2, PO2, SO2%, Hct, (Na+ required for Hct), Hb, and barometric pressure
Calculated Parameters
From the directly measured results, the calculated results are
•Base Excess of the blood (BE-b)
•Base Excess of extracellular fluid (BE-ecf)
•Bicarbonate level (HCO
•Standard Bicarbonate Concentration (SBC)
•Total Carbon Dioxide (TCO2)
•Oxygen Content (O2Ct)
•Oxygen Saturation (SO2%) - (If measured result not available)
•Hemoglobin (Hbc) - (If selected or measured result not available)
•P50 (measure single point)
•pH, PCO2, PO2 (corrected to patient temperature)
˙
•
˙
/
Qsp Qt
(physiological shunt - requires 2 samples: mixed venous and arterial)
•Respiratory Index (RI - uses entered %FIO2 or default value of 20.9)
•PO2/FIO2 ratio
-
)
3
PCO
, PO2, oxygen saturation (SO2%),
2
1-2
With the Nova CO-OXIMETER inputs of Oxygen Capacity of hemoglobin (O2Cap) and
Oxygen Content of hemoglobin (O2Ct), the additional combined CO-Oximeter/Stat Profile
pHOx calculated results are
•CcO
•CaO
•Cv- O
•a-v- DO
2
2
2
2
•P50
1 Introduction
Test Dependencies
•If O2Cap from the Nova CO-OXIMETER is available, that value is displayed.
•If O2Ct from the Nova CO-OXIMETER is available, that value is displayed instead
of the calculated O2Ct value.
•If SO2% from the Nova CO-OXIMETER is available, that value is displayed instead
of the SO2% measured by the reflectance method.
•If Hb from the Nova CO-OXIMETER is available, that value is displayed instead of
the Hb measured by the conductivity and photometric methods.
1. Intro.
Clinical Utility
1
The following list includes the clinical utility information for each of the analytes measured
on the Stat Profile pHOx Analyzer.
Blood Gases:Whole blood measurement of blood gases is used in the diagnosis
(PCO2, PO2,and treatment of life-threatening acid-base disturbances in critically ill
and pH)patients with numerous metabolic and pulmonary diseases.
OxygenUsed to assess the oxygenation of hemoglobin and the adequacy of tissue
Saturationoxygenation in the evaluation of pulmonary function. Also used in the
diagnosis and treatment of cyanosis.
HematocritWhole blood measurement of hematocrit is used to estimate that red blood
cells are present in sufficient quantity to carry oxygen and carbon dioxide.
HemoglobinOxygen is carried from the lungs throughout the body by hemoglobin present
in red blood cells. Measurement of hemoglobin provides the clinician with
information regarding the evaluation of chronic and acute anemias and also
with information pertaining to the potential oxygen transport capability of the
hemoglobin.
PhysiologicThe physiologic shunt or AV Shunt calculation relates to the small fraction
Shuntof the cardiac output which goes to the lungs that does not come into contact
with oxygen exchange units (alveoli). In health, this shunt fraction is less than
5%. Increased shunt fraction is seen in cases where there is a pathophysiologic process that reduces the number of functional gas exchange alveoli or
blood is directed away from functional alveoli.
Ref.1.Tietz, N.W. ed. 1986. Textbook of Clinical Chemistry. W. B. Saunders Co.
1-3
Stat Profile pHOx Reference Manual
1.3The Sample
1. Intro.
1.3.1 Handling Requirements
Sodium or lithium heparin whole blood sample from syringes, open tubes, small cups, and
capillary tubes can be used on the Stat Profile pHOx Analyzer. The sample size is 70 µL for
normal mode and 45 µL for micro mode (blood gases only).
Correct sample handling is critical to ensure that the blood gas values obtained accurately
reflect the in vivo state. Ensure that all samples have been obtained and stored following
consistent, clinically accepted protocols. It is particularly important to ensure that samples are
well mixed before introduction into the analyzer. Nova Biomedical recommends that you
analyze the sample within 15 minutes for blood gases. Storing samples on ice is not
recommended. Using iced samples may elevate the PO2 result.
1.National Committee for Clinical Laboratory Standards. Considerations in the Simultaneous Mea-
surement of Blood Gases, Electrolytes, and Related Analytes in Whole Blood; Proposed Guideline.
NCCLS Document C32-P. Vol. 13, No. 17.
1
1.3.2 Acceptable Anticoagulants
Sodium and lithium heparin are the recommended anticoagulants for use with the Stat Profile
pHOx Analyzer. EDTA, citrate, oxalate, or sodium fluoride are not recommended for use.
Depending on the amount of heparin used in the collection syringe and whether it is filled to
capacity with blood, heparin concentrations of 20 I.U. per mL to over 100 I.U. per mL may be
obtained. Liquid heparin when present in excess may cause errors by dilution in pH, PCO2, and
PO
.
2
Our experience suggests that lyophilized lithium heparin giving a final concentration in blood
of not more than 20 I.U. per mL is acceptable in the critical care laboratory. Stat Profile pHOx
Analyzer users should take careful note of these considerations when establishing reference
intervals and interpreting results.
1-4
Update to PN 22363 Rev. D 3/2002
1.3.3 Interfering Substances
S
O2% Interferences:
High levels of COHb and MetHb will increase the reported SO2% value.
MetHb values above 15% will interfere with the SO2% value.
10% Intralipid solutions will interfere with the
Intralipid are >500 mg/dL.
Hemolyzed samples will interfere with the SO2% value.
Hematocrit (Hct) Interference:
White Blood Count (WBC) greater than 50,000 WBC/µL may increase the hematocrit value.
1.3.4 Matrix Effects
Nova instruments are designed for clinical environments to analyze actual patient specimens,
not modified blood samples. Specimens removed from the patient, anticoagulated appropriately, and promptly analyzed are the only type of sample where the measurement results will
be reliable. Matrix effects/interferences can occur when patient specimens are removed from
the body, modified and then measured on a Nova instrument. For example, matrix effects have
been seen on Nova analyzers when attempting to analyze samples collected from cell savers
used in various surgical procedures. Also, evaluation laboratories run specimens from patients
with a wide variety of pathologies and from patients who are being treated with a broad
spectrum of therapeutic and pharmacological agents. Despite extensive clinical trials, it is not
possible to anticipate every possible combination of transfused blood products, crystalloids,
and drugs (or their metabolites) that may be present in a blood sample. As a result, some users
have found that their particular patient mix has necessitated making adjustments to maintenance. For example, a high number of cardiopulmonary bypass pump or ECMO (extracorporeal membrane oxygenation) samples result in a need for increased analyzer maintenance. If
you are experiencing excessive downtime, you may need to modify your own maintenance
schedules. Nova’s Clinical Applications Group will assist you in tailoring a maintenance
program to meet these needs.
1 Introduction
SO
% when the blood concentrations of
2
1. Intro.
1.4About This Reference Manual
This manual is for Stat Profile pHOx Analyzer.
Throughout this manual,
information that is critical to avoid instrument damage or incorrect results, and
indicates possible hazard to the operator.
NOTE:
indicates especially important information,
CAUTION:
indicates
WARNING:
1-5
Stat Profile pHOx Reference Manual
1. Intro.
1-6
2Setup
This section describes how to setup the Stat Profile pHOx Analyzer.
2.1Installing the Stat Profile pHOx
The analyzer is initially installed by a factory authorized representative.
2 Setup
2. Setup
CAUTION:
at all times. This is necessary to prevent crystallization of salts in the fluid
lines and cuvette. If it is to be shut down indefinitely, purge all fluid lines
with distilled water and then with air. The purge sequence is selected from
the Operational Menu.
2.2Power Up Procedure
After power up, the analyzer checks the error condition status of the instrument by the Power
On Self Test (POST). The bicolor LED on the front panel will be a solid yellow if the analyzer
has passed this test.
NOTE:
than 1 minute, the Not Ready screen will appear with the LED at solid
yellow. The analyzer will need to be calibrated.
The pHOx analyzer is designed to be left on with adequate fluids
After power up, the analyzer displays the pHOx logo screen. In less
2.3Using the Keypad and Display
OverviewThe keypad, display, and status lights are located on the front panel.
DisplayThe display provides prompts, menus, status information, error messages,
patient results, etc. The top line gives the screen's name (i.e., Setup Menu) and
in some screens the date and the time. The second line displays directions for
the screen or additional information about the displayed data. The middle of the
screen is for the menu items that you can select, detailed direction for
procedures, patient information, or electrode statuses. The bottom line defines
the soft keys.
2-1
Stat Profile pHOx Reference Manual
Soft KeysThe 4 soft keys are defined by the labels currently shown on the lower line
of the display. The keys cause system action, such as selection of menus,
initiation of maintenance procedures, and other displays.
Some common soft keys are
Home returns you back to the READY screen.
Next Screen moves you to the next screen in the sequence.
Cancel returns to the previous screen or cancels a sequence.
2. Setup
Status LightsThe 2 status lights on the front panel reveal the system status as follows:
Steady Green - All air detectors and one or more channels are calibrated. The
analyzer is ready for analysis or input.
FlashingGreen - All air detectors and one or more channels are calibrated, but
the analyzer is busy analyzing, priming, accepting external data, etc. The
analyzer will not allow any analytical or any other sequence to be started until
it becomes not busy.
Steady Yellow - One or more air detectors or all channels are not calibrated.
Flashing Yellow - One or more air detectors or all channels are uncalibrated and
the analyzer is busy.
Status SymbolsThere are a number of symbols that can appear after the results. The symbols
have the following meanings:
↑ (single up arrow), ↓ (single down arrow) - The result is higher or lower that the
defined reference range for the parameter.
↑↑ (double up arrow), ↓↓ (double down arrow) - The result is higher or lower that
the defined alert range for the parameter.
↑↑↑ (triple up arrow), ↓↓↓ (triple down arrow) - The result is out of the
analyzer's operating range.
X (an A through an analyte prefix) - The channel is uncalibrated.
? (question mark) - Insufficient sample is detected during sample reading.
* (asterisk) - The result is calculated using a default sodium concentration.
— (a line through an analyte prefix) - The channel did not pass QC and QC
lockout is enabled; or the results have been suppressed.
KeypadThe keypad allows you to enter information into memory. It consists of 12
number keys (including a decimal point key and a dash key), up (↑), down
(↓), left (←), and right (→) arrow keys, and an ENTER (↵) key. The arrow
keys move the cursor on the display. Also, the left arrow key can be used as
a backspace when entering numerical information. The ENTER (↵) key
places data on the display into memory.
2-2
Analyze KeysThere are 2 keys to initiate an analysis: one key has an icon of a capillary tube
and the other key has an icon of a syringe. To perform an analysis with a
capillary sample, press the capillary key. To perform an analysis with a
syringe sample, press the syringe key. These 2 keys are only active when the
Home screen or the Result screen are displaying.
2.3.1 General Keypad Entry
Entering numbers:
1. In general, press the arrow keys to select a character entry field or option. Then follow
the screen instructions.
2. For numeric ID entry, press a key to add a character, and the cursor moves to the right.
Up to 20 characters are accepted. If editing an entry, press the left arrow (←) to erase
the last character.
3. After data has been entered correctly, press ENTER (
information into memory.
Soft Keys
2 Setup
↵
) as a final step to store the
Analyze Capillary
2. Setup
Number Keys
Green LED
Yellow LED
1
4
7
2
5
8
0
Figure 2-1. Stat Profile pHOx Keypad
3
6
9
ENTER
Analyze Syringe
Up, Down, Left,
Right Keys
Enter Key
2-3
Stat Profile pHOx Reference Manual
2.4Overview of the Displays (User Interface)
The format that allows the operator to change displays, enter data, and perform functions is
generally referred to as the user interface (controlled by software).
Here are some general rules for the user interface:
2. Setup
•To perform any operation, use the screen instructions to guide you.
•To go back to the Ready screen or the menu for a selected section (i.e., QC, Setup,
etc.), press Home.
•To move back to the previous display, press Previous Page (soft key), if applicable.
•Press Cancel (soft key) to terminate the current operation or sequence or to return to
previous display.
•Press Exit (soft key) to return to the previous display.
•Analyzing a sample and performing calculations take precedence over the user
interface. Therefore, you are temporarily locked out of accessing displays that can
interfere with an ongoing sequence or operation.
•A status message indicates an error condition.
2.5Adapting the Program to Your Clinical Requirements with the Setup Menu
Use the Setup Menu to adapt the analyzer to your requirements.
2-4
2.6Setup Options
The Setup Menu has the following setup options:
•Results Configuration Menu
•Operation Configuration Menu
2 Setup
-Reference and Alert Limits Setup
PCO
pH,
-Electrode Offsets Setup
pH,
PCO
-Results Units Setup
Temperature: °C or °F
Blood Gas:mmHg or kPa
pH/H+:pH or H
Hb:g/dL or mmol/L or g/L
-Results Suppression:Suppress or not suppress
-Remote Review:ON or OFF
-Patient Name:ON or OFF
-Mandatory Patient ID:ON or OFF
-Analysis Configuration
Analysis Transmit ModeManual/Auto
Transmit Diagnostic DataOff/ON
Analysis Print ModeManual/Auto
Print Diagnostic DataOFF/ON
CO-Ox Data MergeAccession nbr/Patient ID
Set Analyzer ID Number:
-Calibration Configuration
Set 1 Point Calibration Frequency 30, 45 min.
Set 2 Point Calibration Frequency 2, 4, 6 hr.
Transmit Diagnostic DataOFF/ON
Print Diagnostic DataOFF/ON
Transmit Drift DataOFF/ON
Print Drift DataOFF/ON
-System Configuration
Set Date FormatDD/MM/YY, MM/DD/YY, YY/MM/DD
Set Time Mode24/12 hr. Mode
Date and TimeSet
Measured Barometric Pressure760 mmHg
Corrected Barometric Pressure0.0 mmHg
Set Default Hb Used in Calculations:14.3 g/dL
Set Tone Frequency (100-500Hz):3500 Hz
-Analysis Mode:A or B
-Hb Type:Measured or Calculated
-STAT Mode:ON or OFF
, PO2, SO2%, Hct, Hb
2
, PO2, SO2%, Hct, Hb
2
+
2. Setup
2-5
Stat Profile pHOx Reference Manual
•Communications Menu
-Baud (4800, 9600, 19200)
-Data Bits (7, 8)
-Stop (1,2)
-Parity (None, Even, Odd)
-External Keyboard: Yes or No
-Co-oximeter attached: Yes or No
2. Setup
2.6.1 Password
•PrinterInstalled or Not Installed
•Passwords:System or Operator
•Operator Passwords: ON or OFF
•Language:English, Chinese, French, German, Italian, Japanese, and Spanish
The first option you should call up is Set Password: either System or Operator.
The System Password, which safeguards the setup parameters, is required to enter the Setup
Menu. Set up System Passwords as follows:
1. Press Menu on the Ready screen to display the Operational Menu screen.
2. Press Setup to display the Setup Menu screen.
3. A pop-up Password screen appears. Enter the default password, 0. This is the
password until one is officially entered.
4. Scroll down with the arrow keys to the Password option and press Enter. A pop-up
screen appears: select System Password.
5Key in a password number between 1 and 9999 (up to 4 numeric characters) then
press ENTER.
6. Proceed to the next setup option or return to the Operational Menu.
2-6
The Operator Passwords allows you the ability to enable and to enter 200 unique passwords
with privilege levels. When enabled, password entry will be required whenever any of the
home screen soft keys, syringe key, or capillary key are pressed. Entry into any one of the
protected areas depends on the privilege level assigned to the operator of the analyzer.
There are 3 privilege levels: Privilege Level 1 is the most privileged operator and Privilege
level 3 is the default level.
•Privilege Level 1 operators have access to all areas of the analyzer except those
protected by the existing System Password. Level 1 operators do not require PDM
review and may override this feature. Level 1 operators can override QC lockout.
•Privilege Level 2 operators have access to all areas of the analyzer except those
protected by the existing System Password. Level 2 operators will require PDM
review, but may override this feature. Level 2 operators cannot override QC lockout.
•Privilege Level 3 operators have access to analysis only. Level 3 operators will
require PDM review before the results are released. Level 3 operators cannot override
QC lockout.
Set up Operator Passwords as follows:
1. Press Menu on the Ready screen to display the Operational Menu screen.
2. Press Setup to display the Setup Menu screen.
3. A pop-up Password screen appears. Enter the default password, 0. This is the
password until one is officially entered.
3. Scroll down with the arrow keys to the Password option and press Enter. A pop-up
screen appears: select Operator Password.
4. Key in a password number, up to 4 numeric characters, and press ENTER.
5. To enable Operator Passwords, scroll down to Operator Passwords then press the
Enter key: Off will turn to On.
6. Proceed to the next setup option or return to the Operational Menu.
2.6.2 Results Configuration Menu
The Results Configuration Menu screen allows you to set reference and alert limits (Low and
High) for all analytes; to adjust electrode offsets for slope and intercept for all analytes; to select
units for temperature, pH, blood gas, and hemoglobin; to turn Remote Review On or Off
(operator Password is enabled when On); Mandatory Patient ID (On or Off); Patient Name (On
or Off); Global Suppression (Test Results).
2 Setup
2. Setup
2.6.2.1 Remote Review
The operator passwords are enabled whenever Remote Review is enabled. To enable Remote
Review (password required), go to the Results Configuration screen, select Remote Review,
then press the Enter key to enable (On) - disable is Off.
Features of Remote Review are as follows:
•Send Results Data to PDM for Review: When all results and calculated parameters are
available, the pHOx Analyzer transmits all this data to the PDM. The analysis cycle
is suspended indefinitely until the PDM returns record or the review is bypassed or
cancelled.
•Suppress Results Data: The PDM will return a reviewed data record to the pHOx
Analyzer. The analyzer will suppress test results based on the information received
from PDM.
•Bypass Review: A pHOx operator with password level 1 or level 2 can bypass a remote
review session. When the session is bypassed, the results are saved, displayed, and
printed (if automatic printing is on). A reviewed data record sent by PDM is ignored
if the review session is bypassed.
2-7
Stat Profile pHOx Reference Manual
•Cancel Review: A pHOx operator can cancel a remote review session. When the session
is cancelled, the results are saved, but the results will not be displayed or printed. A
reviewed data record sent by PDM is ignored if the review session is cancelled.
•Review Patient Name: PDM can return a patient name with the suppressed test results.
The patient name will be displayed and printed with the result data if the patient name
entry feature is ON.
•Review Accession Number: PDM can return an accession number with the suppressed
2. Setup
2.6.2.2 Results Suppression
In the Results Configuration screen, test results can be suppressed: will not be displayed,
printed, or transmitted. The suppressed results will only affect analysis results: it will still
calibrate.
Measured tests that depend on the results from a suppressed test will not be displayed. A
dependency error is displayed.
Calculated parameters that depend on suppressed results are not calculated.
Suppress results as follows:
test results. The accession number will overwrite the existing accession number. The
accession number will be displayed and printed with the result data.
1. In the Results Configuration screen, scroll down to Results Suppression.
2. Press the Enter key to get a pop-up screen of the tests.
3. Scroll to the test that you want to suppress then press the Enter key.
4. Scroll to the next test or exit from screen.
5. On the Home screen, the suppressed test will have a strike through the test;
no results will be reported on this test.
2.6.2.3 Mandatory Patient ID
In the Results Configuration screen, the Mandatory Patient ID is enabled or disabled. If
enabled, the Sample Information screen will prompt for Patient ID before the sample analysis
can continue.
A patient ID can be sent from PDM as part of a remote review session.
A patient ID can also be received from a bar code device or external keyboard through the bar
code port.
2-8
2.6.3 Operation Configuration Menu
The Operation Configuration Menu screen allows you to setup the analysis output options, to
configure calibration frequency, and to set date, time, barometric pressure, analyzer ID
number, and to change the analysis mode (A or B); to select Hb Measured or Calculated; to
select STAT Mode.
2.6.4 Communications
The Communication screen allows you to setup the communication ports: Baud, Stop, Data, and
Parity. Also, the External Keyboard and CO-Oximeter attached (Yes or No) is selected here.
2.7QC Setup
2 Setup
2. Setup
To access the QC Setup screen, press QC (soft key) on the Ready screen. Then use the down
arrow key to select QC Setup and press the ENTER key to display the QC Setup screen.
For External Controls, enter the Lot Number, the Expiration Date, the Daily Analysis Times
(up to 3 times per control per day), and the Ranges of each control.
For Internal Controls, the Lot Number, the Expiration Date, and the Control Ranges are
automatically read from the control pack when the QC Auto Cartridge is installed. The Daily
Analysis Times (up to 3 times per control per day) must be manually entered.
2.7.1 QC Lockout
QC Lockout can be enabled (ON) in one of 2 modes or disabled (OFF) from the QC Setup
screen (a password protected screen). When QC Lockout is selected from this screen, a popup screen appears. Lockout OFF, Mode A, and Mode B are the choices. When QC Lockout is
ON (Mode A or Mode B) and QC is due, all test that are scheduled for QC will be lock out. A
QC analysis must be done for a locked out test before an analysis can be performed.
Channel Lockout
A channel will be locked out when it does not pass a QC level. The channel appears with a
strike-through it. This channel will not report results unless a password overrides it. When all
channels become locked out, the analyzer displays the Not Ready screen.
NOTE:
track (Mode A) of the channels that would be locked out if the QC Lockout
feature was ON. If the QC Lockout feature is now turned ON, all those
channels will now become locked out.
If the QC Lockout feature is OFF, the analyzer will internally keep
2-9
Stat Profile pHOx Reference Manual
Level Lockout
QC level lockout is enforced for both Mode A and Mode B. If a QC level is not run and QC
Lockout is enabled, all channels in that QC level will become locked out. These channels
remain locked out until the QC analysis cycle of that level is run. A message is displayed on
the Ready/Not Ready screen that a QC analysis is due for a particular level.
2. Setup
NOTE:
track of all channels in this level that would be locked out if the QC Lockout
feature was ON. If the QC Lockout feature is now turned ON, all those
channels will now become locked out.
QC Lockout is activated by selecting either Mode A or Mode B.
Mode A QC lockout locks or unlocks channels based on the last QC level run. A lockout
channel is displayed with a strike-through the channel name. Results for a locked out channel
are not calculated. If more than one QC level is run for the QC cycle, only the last QC level
that a channel is in determines whether the channel will be locked or unlocked. If 3 levels were
run and the channel passed the first 2 levels but failed the last, the channel will be locked out.
To be unlocked, this channel must pass that last level: passing the first 2 does not count. If the
reverse happened and the channel failed the first 2 levels but passed only the last level, the
channel will not be locked out.
Mode B QC lockout locks or unlocks channels based on the last QC run for all levels. A
lockout channel is displayed with a strike-through the channel name. Results for a locked out
channel are not calculated. If a channel fails any level run in a QC cycle, the channel will be
locked. To be unlocked, the channel must pass all levels that it failed.
If the QC Lockout feature is OFF, the analyzer will internally keep
2-10
QC Lockout OFF
When the QC lockout mode is set to OFF, the analyzer internally tracks the lockout state for
each channel. The tracking uses the Mode A lockout. If QC lockout is enabled by selecting
either Mode A or Mode B, any number of the tracked channel can now become locked out.
QC Lockout is password protected, and a password override (Level 1 operators only) is
provided with a pop-up screen. Analysis can be initiated for all channels in a lockout state, but
results will be displayed with the message "Scheduled QC Not Run" and printed with this error
message.
For Automatic QC mode to perform, turn Automatic QC Analysis ON in the Quality Control
screen.
Messages displayed on the Ready/Not Ready screen detail the reason for the QC lockout. These
message are only displayed if QC lockout is enabled (Mode A or Mode B). The following are
the messages and their meanings.
Loading...
+ 115 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.