NOSHOK 100-300-1-1-2-7, 100-150-1-1-2-7, 300-150-1-1-2-8, 100-100-1-1-2-7, 100-200-1-1-2-7 User manual

...
Current output, 2 wire
Power Supply
Output
+ Supply
+ Output
M12 x 1 4-pin
round connector
Red
Power
Supply
Black
2 WIRE
WIRING DIAGRAM EXAMPLE
SERIES 100 SERIES 200
Wiring ­Mini-Hirschmann
1
2
connector
Power Supply
+ Supply
3
Wiring -
Output
Load Limitations
4 mA to 20 mA Output Only
Output
R
L = RS + RW
RL = Loop Resistance (ohms)
R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
Series 100 4 mA to 20 mA
CONNECTION TYPE
(CODE)
Mini-Hirschmann
(7)
Mini-Hirschmann
w/ Cable (1)
4 or 6 Pin Bendix
(2 or 3)
1/2" NPT Conduit
w/ Cable (6)
M12 x 1, 4-Pin (25) 1 3
Integral Cable (36) Red Black
+ SUPPLY + OUTPUT
1 2
Red Black
A B
Red Black
+ Output
Voltage output, 3 wire
Power Supply
Output
L)
Series 200
CONNECTION TYPE
Mini-Hirschmann
Mini-Hirschmann
w/ Cable (1)
4 or 6 Pin Bendix
1/2" NPT Conduit
w/ Cable (6)
4-Pin (25)
Integral Cable
+ Supply
1
3
+ Output
Common
M12 x 1 4-pin
round connector
3 WIRE WIRING
DIAGRAM EXAMPLE
0-5 Vdc, 1-5 Vdc,
0-10 Vdc, 1-11 Vdc
(7)
+ SUPPLY COMMON + OUTPUT
1 2 3
Red Black White
A B C
Red Black White
1 3 4
Red Black White
(CODE)
(2 or 3)
M12 x 1,
(36)
Wiring ­Mini-Hirschmann connector
2
Wiring -
1-6 Vdc,
Power Supply
Power Supply
Output
Red
Black
+ Supply
White
+ Output
Common
Output
SERIES 300 SERIES 612 & 613
2 WIRE WIRING
DIAGRAM EXAMPLE
Series 300 4 mA to 20 mA
CONNECTION TYPE (CODE) + SUPPLY + OUTPUT
Hirschmann
(7, 8 or 14)
Hirschmann w/ Cable (1)
M12 x 1, 4-Pin (25) 1 3
Integral Cable (36) Brown Blue
Load Limitations 4 mA to 20 mA Output Only
L = RS + RW
RL = Loop Resistance (ohms) R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
3 WIRE WIRING
DIAGRAM EXAMPLE
Series 300
CONNECTION TYPE
Hirschmann
(7, 8 or 14)
Hirschmann w/ Cable (1)
M12 x 1,
4-Pin (25)
Integral Cable
0-5 Vdc, 1-5 Vdc, 1-6 Vdc,
(CODE)
+ SUPPLY COMMON + OUTPUT
(36)
1 2
Red Black
L)
Power Supply
0-10 Vdc, 1-11 Vdc
1 2 3
Red Black White
1 3 4
Brown Blue Black
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
Power Supply
Red
Output
Black
Red
Black
White
Output
www.GlobalTestSupply.com
2 WIRE WIRING
DIAGRAM EXAMPLE
Power
Supply
Series 612/613 4 mA to 20 mA
CONNECTION TYPE + SUPPLY + OUTPUT CASE GROUND
Cable Red Black Shield
Load Limitations 4 mA to 20 mA Output Only
Vmin = [10V + (.020 x R
RL = RS + RW
L)] - O.04354 x
cable length
RL = Loop Resistance (ohms) R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
3 WIRE WIRING
DIAGRAM EXAMPLE
Series 612/613
CONNECTION
TYPE
Cable Red Black White Shield
0-5 Vdc, 0-10 Vdc,
0.5 to 2.5 Vdc
+ SUPPLY COMMON + OUTPUT
Power Supply
GROUND
Red
CASE
Ft.
Red
Black
Shield
Black
White
Shield
Red
Power
Supply
Output
Load Limitations 4 mA to 20 mA Output Only
L = RS + RW
RL = Loop Resistance (ohms) R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
Series 615/616
CONNECTION TYPE
Output
(CODE)
Hirschmann
(8 or 14)
Hirschmann w/ Cable (1)
6 Pin Bendix (3) A B C
1/2" NPT Conduit
w/ Cable (6)
M12 x 1, 4-Pin (25) 1 3 4
Integral Cable (36) Red Black White
SERIES 615/616
Black
CONNECTION TYPE (CODE) + SUPPLY + OUTPUT
L)
0-5 Vdc, 1-5 Vdc,
1-6 Vdc,
0-10 Vdc, 1-11 Vdc
+ SUPPLY COMMON + OUTPUT
1 2 3
Red Black White
Red Black White
2 WIRE WIRING
Output
DIAGRAM EXAMPLE
Series 615/616 4 mA to 20 mA
Hirschmann
(8 or 14)
Hirschmann w/ Cable (1)
6 Pin Bendix (3) A B
1/2" NPT Conduit
w/ Cable (6)
M12 x 1, 4-Pin (25) 1 3
Integral Cable (36) Red Black
3 WIRE WIRING
DIAGRAM EXAMPLE
Power Supply
Red
Black
1 2
Red Black
Red Black
White
SERIES 640
Current output, 2 wire
+ Supply Power Supply
Output
+ Output
Voltage output, 3 wire
+ Supply
Power Supply
Output
+ Output
Common
RS 232 Interface
Output
Series 640 0-5 Vdc, 0-10 Vdc, 0-20 mA
CONNECTION TYPE (CODE) + SUPPLY COMMON + OUTPUT
Hirschmann (8 or 14) 1 2 3
Hirschmann w/ Cable (1) Red Black White
6 Pin Bendix (3) A B C
M12 x 1, 4-Pin (25) 1 3 4
Integral Cable (36) Brown Black Blue
Wiring - M12 x 1 4-pin
round connector
Series 640 4 mA to 20 mA
CONNECTION TYPE (CODE)
Hirschmann
(8 or 14)
Hirschmann w/ Cable (1)
6 Pin Bendix (3) A B
M12 x 1,
4-Pin (25)
Integral Cable (36) Brown Blue
Load Limitations 4 mA to 20 mA Output Only
L = RS + RW
RL = Loop Resistance (ohms) R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
Pressure Sensor
+ SUPPLY + OUTPUT
1 2
Red Black
1 3
L)
Wiring - Mini-Hirschmann connector
Current output, 2 wire
+ Supply Power Supply
Output
+ Output
1
3
2
Load Limitations 4 mA to 20 mA Output Only
L = RS + RW
RL = Loop Resistance (ohms) R
Voltage output, 3 wire
Power Supply
Output
Series 660 4 mA to 20 mA
CONNECTION TYPE (CODE) + SUPPLY + OUTPUT
Mini-Hirschmann (7) 1 2
Mini-Hirschmann w/ Cable (1) Red Black
M12 x 1, 4-Pin (25) 1 3
Integral Cable (36) Brown Green
+ Supply
+ Output
Common
1
3
2
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
SERIES 660
Mini-Hirschmann w/ Cable (1) Red Black White
Wiring - M12 x 1 4-pin round connector
Current output, 2 wire
+ Supply
Power Supply
L)
Output
Voltage output, 3 wire
Power
Supply
Output
Series 660 1-5 Vdc, 0.1-10 Vdc
CONNECTION TYPE (CODE) + SUPPLY COMMON + OUTPUT
Mini-Hirschmann (7) 1 2 3
M12 x 1, 4-Pin (25) 1 3 4
Integral Cable (36) Brown Green White
+ Output
+ Supply
+ Output
Common
SERIES 800
4 mA to 20 mA, 2 wire
Power Supply
+ Supply
Load Limitations
Output
+ Output
4 mA to 20 mA Output Only
L = RS + RW
L)
RL = Loop Resistance (ohms) R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
Series 800 4 mA to 20 mA
CONNECTION TYPE (CODE) + SUPPLY + OUTPUT
Hirschmann (8 or 14) 1 2
Hirschmann w/ Cable (1) Red Black
M12 x 1, 4-Pin (25) 1 3
Find Q uality Pro ducts Onli ne at: sales@ GlobalTestSupp ly.com
www.GlobalTestSupply.com
Series 800 0-10 Vdc
CONNECTION TYPE (CODE) + SUPPLY COMMON + OUTPUT
Hirschmann (8 or 14) 1 2 3
Hirschmann w/ Cable (1) Red Black White
M12 x 1, 4-Pin (25) 1 3 4
0 Vdc to 10 Vdc, 3 wire
Power Supply
Output
+ Supply
+ Output
Common
Installation:
NOSHOK pressure transmitters/transducers may be mounted in any plane with negligible effect on performance. Although these units are designed and manufactured to withstand substantial shock and vibration, it is recommended that they be mounted in an area of minimal vibration. Always use a wrench on the wrench flats when installing. NEVER use a pipe wrench on the housing or in the area of the electrical connection.
Maintenance/Calibration:
NOSHOK pressure transmitters/transducers require no maintenance. Recalibration is dependent on the users Quality Assurance Program. If no program is in place, NOSHOK recommends a 1 year cycle.
Alignment Procedure (applies only to 100, 200, 615/616, and 640 series):
Using a pressure source and meter with adequate accuracy, perform the following steps:
• Open sensor
• With no pressure applied, adjust the “Z” potentiometer for the correct Zero output
• Apply the correct full scale pressure to the unit
• Adjust the “S” potentiometer for the correct Span output
NWD 08-6
Span
Zero
NOSHOK
TRANSMITTERS
TRANSDUCERS
Wiring Diagrams & Electrical
Connections for:
100, 200, 300, 612, 613, 615/616,
640, 660, and 800 Series
CORPORATE HEADQUARTERS
1010 West Bagley Rd. • Berea, OH 44017
440.243.0888 • FAX 440.243.3472 E-mail: noshok@noshok.com Web: www.noshok.com