Installation:
NOSHOK pressure transmitters/transducers may
be mounted in any plane with negligible effect on
performance. Although these units are designed
and manufactured to withstand substantial shock
and vibration, it is recommended that they be
mounted in an area of minimal vibration. Always
use a wrench on the wrench flats when installing.
NEVER use a pipe wrench on the housing or in the
area of the electrical connection.
Maintenance/Calibration:
NOSHOK pressure transmitters/transducers
require no maintenance. Recalibration is
dependent on the users Quality Assurance
Program. If no program is in place, NOSHOK
recommends a 1 year cycle.
Alignment Procedure (applies only to 100,
200, 615/616, and 640 series):
Using a pressure source and meter with adequate
accuracy, perform the following steps:
• Open sensor
• With no pressure applied,
adjust the “Z”
potentiometer for the
correct Zero output
• Apply the correct full scale
pressure to the unit
• Adjust the “S”
potentiometer for the
correct Span output
NWD 08-3
Span
Zero
NOSHOK
TRANSMITTERS
TRANSDUCERS
Wiring Diagrams & Electrical
Connections for:
100, 200, 300, 612, 613, 615/616,
640, 660, and 800 Series
CORPORATE HEADQUARTERS
1010 West Bagley Rd. • Berea, OH 44017
440.243.0888 • FAX 440.243.3472
E-mail: noshok@noshok.com
Web: www.noshok.com
SERIES 100 SERIES 200
Current output, 2 wire
Power
Supply
Output
+ Supply
+ Output
M12 x 1 4-pin
round connector
Red
Power
Supply
Black
Output
2 WIRE
WIRING
DIAGRAM
EXAMPLE
Wiring Mini-Hirschmann
1
2
connector
Power
Supply
+ Supply
3
Wiring -
Output
+ Output
Load Limitations
4 mA to 20 mA Output Only
Vmin = 10V + (.020 x R
R
L = RS + RW
RL = Loop Resistance (ohms)
R
S = Sense Resistance (ohms)
R
W = Wire Resistance (ohms)
Series 100 4 mA to 20 mA
CONNECTION TYPE
(CODE)
Mini-Hirschmann
(7)
Mini-Hirschmann
w/ Cable (1)
4 or 6 Pin Bendix
(2 or 3)
1/2" NPT Conduit
w/ Cable (6)
M12 x 1, 4-Pin (25) 1 3
Integral Cable (36) Red Black
+ SUPPLY + OUTPUT
1 2
Red Black
A B
Red Black
Voltage output, 3 wire
Power
Supply
Output
L)
Series 200
CONNECTION TYPE
Mini-Hirschmann
Mini-Hirschmann
w/ Cable (1)
4 or 6 Pin Bendix
(2 or 3)
1/2" NPT Conduit
w/ Cable (6)
M12 x 1,
4-Pin (25)
Integral Cable
+ Supply
1
3
2
+ Output
Common
Wiring -
M12 x 1 4-pin
round connector
3 WIRE WIRING
DIAGRAM EXAMPLE
0-5 Vdc, 1-5 Vdc,
1-6 Vdc,
0-10 Vdc, 1-11 Vdc
(7)
+ SUPPLY COMMON + OUTPUT
1 2 3
Red Black White
A B C
Red Black White
1 3 4
Red Black White
(CODE)
(36)
Wiring Mini-Hirschmann
connector
Power
Supply
Output
Red
Power
Supply
Black
+ Supply
+ Output
Common
White
Output