NORTH AMERICAN DRÄGER D-Tec User manual

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Field
Service
Procedure
Part Number: SP00189
Date: 12 June 2002
© 2002 Draeger Medical, Inc.
Rev: C
D-TEC Desflurane Vaporizer Installation
Removable Vaporizer Mounting Systems
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
D-TEC Desflurane Vaporizer Installation on
Machines with Removable Vaporizer Mounting Systems
Installation Procedure
NOTE: This procedure only applies to installation of D-Tec Desflurane vaporizer on Narkomed 6000, Narkomed M, Narkomed GS Euro, Julian, Fabius GS and Fabius machines with removable vaporizer mounting systems. Refer to Field Service Procedure SP00091 for other desflurane vaporizer installations.
¡¡¡¡WARNING: Do not install a D-Tec vaporizer in the right hand position of a
Narkomed 6000 as this will result in an incorrect fit and may cause leaks in the fresh gas delivery system.
¡¡¡¡WARNING: Do not install a D-Tec vaporizer on a Narkomed Mobile without the
correct vaporizer support block as this will result in an incorrect fit and may cause leaks in the fresh gas delivery system. Refer to this procedure to verify correct vaporizer support block prior to D-Tec vaporizer mounting.
¡¡¡¡WARNING: The O-rings supplied with this vaporizer are incompatible for use on
the Narkomed 6000, Narkomed GS Euro, Narkomed M, Fabius GS or the USA manufactured Fabius. These systems shall only use O-RING #113 (NITRILE) P/N
4115864.
NOTE:Proceed to Section 2.0 if you are not
removing an existing vaporizer.
1.0 Vaporizer Draining, Drying and Removal Procedure:
¡¡¡¡WARNING: Do Not install a D-Tec
vaporizer in the right hand position of a Narkomed 6000 as this will result in an incorrect fit and may cause leaks in the fresh gas delivery system.
Before removing an existing vaporizer from the machine perform a fresh gas leak test at the fresh gas outlet to verify the initial gas circuit integrity. Refer to the device service manual for specific instructions.
¡¡¡¡Caution: The steps in this section
must be performed in the following sequence:
1.1 Activate the waste gas scavenger.
1.2 Attach a breathing bag to the bag mount port.
1.3 Open the APL valve to the minimum position.
1.4 Connect the inspiratory and expiratory valves using a 22 mm hose.
1.5 Ensure the fresh gas hose is connected between the absorber system and the machine.
If there is an existing vaporizer in place, it must be drained and dried as follows:
1.6 Turn the System Power switch to ON.
1
Installation Procedure (continued)
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1.7 Set the ventilation control to Manual or MAN/Spont mode.
1.8 Set all vaporizer handwheels to their “0” position.
¡¡¡¡WARNING: Do Not inhale anesthetic
vapors as this could result in personal injury.
1.9 Open the filler and drain plugs and drain the vaporizer into a suitable container. Dispose of the residual agent in an approved manner.
1.10 Close the filler and drain plugs.
1.11 Move the exclusion lever away from the vaporizer being removed. Depress the Zero release button and rotate the vaporizer handwheel to the maximum concentration setting.
1.12 Set the oxygen flow to 10 L/min. for at least 20 minutes.
1.13 Turn the vaporizer handwheel to the Zero position, set the oxygen flow to minimum, and turn the System Power switch to STANDBY.
1.14 Locate the vaporizer lock down lever at the rear of the vaporizer, and turn the lever 90° counter-clockwise.
NOTE: Earlier units are not configured as
“User Removable” and therefore require a special wrench (P/N
4114522) to be inserted into the keyed spline which is covered by a “DO NOT REMOVE” label. Remove this label for access to the keyed spline through the rocker plate at the rear of the vaporizer, and turn the tool 90° counter-clockwise. See Figure 1.
1.15 Verify that the exclusion slider bar is disengaged from the vaporizer being removed.
1.16 Hold the vaporizer with two hands and lift it up until it clears the mounting block, then bring it forward to clear the machine.
2.0 Machine Preparation:
2.1 Before installing a vaporizer, perform a fresh gas leak test at the fresh gas outlet to verify gas circuit integrity. Refer to the device service manual for specific instructions.
2.2 Examine the vapor mount O-rings and replace them if needed.
¡¡¡¡WARNING: The O-rings supplied with
this vaporizer are incompatible for use on the Narkomed 6000, Narkomed GS Euro, Narkomed M, Fabius GS or the USA manufactured Fabius. These systems shall use only O-ring #113 (NITRILE) P/N 4115864.
Note: The following two steps apply to
the Narkomed Mobile and Fabius machines.
2.3 Verify the correct vaporizer support block is installed on the vaporizer mount. See Figure 2.
2
Installation Procedure (continued)
LOCK DOWN TOOL
P/N 4114522
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
FIGURE 1. Vapor 19.3 Vaporizer Installation
VAPOR 19.3
MOUNT
VAPORIZER
SP18901
3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
Installation Procedure (continued)
¡¡¡¡WARNING: Do not install a D-Tec
vaporizer on a Narkomed Mobile or Fabius without the correct vaporizer support block as this will result in an incorrect fit and may cause leaks in the fresh gas delivery system. Refer to this procedure to verify correct vaporizer support block prior to D-Tec vaporizer mounting.
RETURN TO CD-ROM TABLE OF CONTENTS
2.4 If applicable, remove the original vaporizer support block and install the appropriate vaporizer support block using the original hardware as shown in Figure 2.
SP18902
INCORRECT
VAPORIZER
SUPPORT BLOCK
CORRECT
VAPORIZER
SUPPORT BLOCK
FIGURE 2. Vaporizer Support Block
(NM Mobile and Fabius only)
VAPORIZER
SUPPORT BLOCK
P/N 4116999
USE EXISTING
SUPPORT BLOCK
HARDWARE (2X)
4
Loading...
+ 12 hidden pages