NORTH AMERICAN DRÄGER Drager KIT Air Diss User manual

RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
KIT,AIRDISS&CO2YOKE
INSTALLATIONPROCEDURE
NOTE: Thisprocedureoutlinesanextremelycomplexmodificationoftheanesthesia
machine.InstallationofthiskitshallbeperformedonlybyaNorthAmerican DrägerqualifiedTechnicalServiceRepresentativewiththenecessaryknowledge andexperience.
flowknobguards,andflowcontrolknobslocatedbelowthe100to1000ml/min. (fine)flowmeters.
1. TurntheSystemPowerswitchto ONanddisconnectallpipeline hoses.
2. Closeallcylindervalvesexceptthe O
cylindervalve.
2
3. Settheoxygenflowrateto5l/min.
4. OpentheN
Oflowcontrolvalveto
2
drainpressurefromthesystem.
5. ClosetheO
cylindervalve,and
2
closetheflowcontrolvalves.Press theO
FLUSHbuttontodrain
2
pressurefromthesystem.
6. TurntheSystemPowerswitchto STANDBYandremoveACpower fromthemachine.Disableallcircuit breakers.
7. Removethescrewssecuringthe tabletop,andremovethetabletop.
CAUTION: Use ESD control when
handling any electronic componentsorassemblies.
9. Removethevaporboxfrontcover. FormachineswithaMultispec
analyzer,removeallconnectionsto thePCBassemblyonthevaporbox frontplate,andremovethePCB fromtheplate.
10. Removethevaporboxbackcover.
11. Removethescrewsholdingthe angledfrontplateatthetopofthe flowmetershield,andremovethe plate.
12. Removeallflowcontrolknobs.
13. Removethetwoscrewssecuringthe knobguard;removetheguardand theflowmetershield.
8. Removetheflowmeterhousingback cover.
1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CAUTION: Use ESD control when
handling any electronic components or assemblies.
14. Remove the flowmeter lights PCBs and channels from the studs on the O
and N2O flow channels.
2
15. Remove the stop pin hex nuts from each flow control valve.
16. Remove the flow control valves: hold each at the wrench flats and turn it counter-clockwise.
17. Remove each flowmeter tube by turning its retaining insert at the top of the assembly counter­clockwise until there is enough clearance to remove the tube from the channel.
18. Disconnect theflextubing connected to the N remove the O
O restrictor housing, and
2
and N2O restrictor
2
assemblies.
19. Remove the cross-over tubes connecting the O
and N2O fine and
2
coarse flowmeter ports.
24. Remove the three socket head cap screws securing the bottom flowmeter block to the channel assembly.
25. Pull the bottom block forward, disconnect the small diameter flex tubing from the block, and remove the block.
26. Disconnect the flex tubing from the O
and N2O pipeline gauges.
2
27. Remove the three socket head cap screws securing the flowmeter assembly, and remove the assembly.
28. Remove the kep nuts securing the cylinder and pipeline gauges to the flowmeter channel and remove the gauges.
29. Remove the socket head cap screws securing thetop block to the channel assembly.
30. Apply masking tape to the raised surfaces of and the entire inside area ofthe center flowmeter channel to prevent scratching.
20. Remove the manifold assembly connected to the O
and N2O outlets
2
at the top of the flowmeters.
21. Disconnect the 4-way fitting and restrictor assembly’s lower copper tube from the O
flowmeter inlet
2
port.
22. Disconnect the copper tube attached to the N
O flowmeter valve port.
2
23. Disconnect the copper tubes from the O
and N2O cylinder gauges.
2
31. Temporarilyattach the 4-gas bottom block to the channel with the screws that were previously removed from the original block.
32. Using the bottom block as a template, mark the circumference of the 4th gas flow control housing onto the channel: (Insert a flowmeter gasket (P/N 4102725) into the housing. This will provide a smaller diameter hole to aid in locating the center point.) The hole location is shown in Figure 1.
2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
SP10501
7/64 IN. DIA. HOLE (DRILL)
11/16 IN. DIA.
CLEARANCE HOLE
(DRILL)
(LOCATE CENTER
USING 4-GAS BLOCK
AND GASKET AS
TEMPLATE)
1 7/8 IN.
Figure 1: DRILLING LOCATIONS: 4-GAS CONVERSION
3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
33. Remove the 4-gas bottom block and centering gasket.
CAUTION:
Perform the next four steps in a suitable location away from the anesthesia machine to prevent drilling chips from entering the pneumatic system.
34. Center punch the marked hole and drill an 11/16 in. dia. hole into the flowmeter channel. A hole saw is recommended. De-burr any sharp edges.
35. Measure and mark a location 1 in. up from the lower center spacer hole, and centered in the channel as shown in Figure 1.
36. Carefully drill a 7/64 in. dia. hole at this location, and de-burr any sharp edges.
37. Remove all masking tape previously applied, and carefully clean up any metal chips from the flowmeter assembly and surrounding area.
38. Using a small amount of Loctite #271 (red), thread a 3/16 in. tube straight fitting (P/N 4109402) into the back of a high pressure gauge (P/N 4110575-002).
39. Apply a small amount of Loctite #271 (red) to the threads of a 1/16 in. hose barb (P/N 4111771) and pipeline pressure gauge (P/N 4110575-001). Assemble the fittings with coupling (P/N 4103668) as shown in Figure 2.
4
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
40. Place the pipeline gauge assemblies into the upper holes of the gauge channel and secure each gauge with two 10-32 kep nuts (P/N HW55002 x2). See Figure 2. Reuse the original hardware to secure the O
and N2O
2
gauges.
FLOWMETER CHANNEL
PIPELINE PRESSURE GAUGE
41. Place the cylinder gauge assemblies
SIDE VIEW
into the lower holes of the gauge channel and secure each gauge with two 10-32 kep nuts (P/N HW55002 x2). Reuse the original hardware to secure the O
and N2O gauges.
2
HOSE BARB
SP10502
FITTING
CYLINDER PRESSURE GAUGE
GAUGE MOUNTING NUTS
Figure 2: PIPELINE AND CYLINDER GAUGE ASSEMBLY AND MOUNTING
5
COUPLING
STRAIGHT
FITTING
(2X EACH GAUGE)
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
42. Apply a small amount of Loctite #271 (red) to the threads of two 2-56 x ¼ in. pan head screws (P/N HW02000 x2). Place a #2 lock washer (P/N HW67012 x2) on each screw, and secure the two shoulder spacers (P/N4110843 x2) to the inside of the center flowmeter channel.
43. Reinstall the top block assembly to the flowmeter channel using the hardware previously removed.
44. Threadtwo inserts (P/N4102506 x2) into the center locations of the top flowmeter block until snug.
45. Secure the 4-gas bottom block (P/N
4110572) to the flowmeter channel using the hardware previously removed from the original block.
47. Install a #113 O-ring (P/N 4102792 x4) in the grove of each flowmeter insert on all four inserts. Apply a small amount of Parker Super-O­Lube to these O-rings.
48. Thread the chromed flowmeter insert into the upper hole of the bottom block, and the remaining brass inserts into the lower holes until snug. DO NOT overtighten the inserts.
49. Install the flow control valves (P/N 4103352 x2) into the flowmeter inserts. Reuse the original O N
O flow control valves. CAUTION:
2
2
and
Before tightening the cartridge, rotate the valve shaft several turns counter-clockwise to prevent bottomingthe valve element into the seat when the cartridge is tightened.
46. Install a #018 O-ring (P/N 4102336 x4) at the shoulder of each of the four bottom flowmeter inserts (consisting of P/N 4110576 x3 and
4110573) as shown in Figure 3.
50. Install the stop pin nuts (P/N 4103382 x2) onto the flow control assemblies. Reuse the original stop pin nuts for the O
and N2O
2
assemblies.
51. Thread two 10-32 x 1/16 in. hose barb fittings w/seal (P/N 4112707­001 x2) into the back of the bottom flowmeter block.
52. Remove the nuts and ferrules from eight ¼ in. tube straight fittings (P/N 4109408 x8). Apply a small amount of Loctite #271 (red) to the threads of each fitting, and thread the fittings into the top and bottom blocks at the locations marked A Figure 3.
6
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
REAR VIEW OF FLOWMETER CHANEL
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
SP10503
AA
HOSE BARB FITTING W/SEAL
BOTTOM BLOCK MOUNTING SCREWS (3X)
STOP PIN NUT
FLOW CONTROL VALVE
A
AA AAA
DETAIL OF FLOW CONTROL
INSERT ASSEMBLY
CROSS-OVER
TUBES (2X)
HOSE BARB
FITTING W/SEAL
(SIDE VIEW)
O-RING, #113
INSERT
O-RING, #018
Figure 3: INSTALLATION OF FITTINGS AND FLOW CONTROL VALVE INSERT ASSEMBLY
7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
53. Assemble the compression fittings onto the open ended tubes on the 4­gas manifold assembly (consisting of P/Ns 4108483, 4109408 x4,4109410, 4110834-009 x3, and 4110837-013 x2) as shown in Figure 5. Ensure that the Swagelok fittings with the two-piece ferrules are assembled as shown in Figure 4.
TWO-PIECE FERRULE
Figure 4: ASSEMBLY OF SWAGELOK FITTINGS
Insert the copper tubes into the manifold assembly as shown in Figure 5. Before tightening the compression nuts, ensure that the tubing is inserted into the fittings as far as possible.
COPPER
TUBING
COMPRESSION
NUT
SP10504
54. Attach the copper tubes on the manifold assembly on the ports at the top of the flowmeter bank as shown in Figure 5. Ensure that the tubing is inserted into the fittings as far as possible, and tighten the compression nuts.
O2
MIX
55. Attach the correct gas-identifying label (P/N 4109871, 4109872, 4109873, 4109874, and 4109901) to each tube on the manifold assembly as shown in Figure 5.
SP10505
N20
AIR
MANIFOLD
TOP VIEW
Figure 5: FOUR GAS MANIFOLD ASSEMBLY WITH LABELS
8
CO2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
56. Reinstall the flowmeter channel assembly using the hardware previously removed.
57. Reinstall the cross-over tubes connecting the oxygen and nitrous oxide fine and coarse flowmeters. Verify that the correct identification labels are attached to each tube.
58. Mount the cylinder support (P/N
4106690) on the bottom frame rail at the back of the machine (see Figure
6) using two ¼-20 x 2½ in. hex head screws (P/N HW08009 x2), lock washers (P/N HW65010 x2) and flat washers (P/N HW66004 x2).
REAR VIEW
59. Mount the CO (consisting of P/Ns 4111792, HW06006, 1101624, 4104716, 4105929 x2, and 4112755-001), (and spacer block (P/N 1101593) on earlier models) on the upper frame rail of the machine as shown in Figure 6. Use two 5/16-24 x 1¾ in. socket head screws (P/N HW01058 x2) andlock washers (P/N HW65005 x2). Apply a CO
1101639) on this yoke.
YOKE ASSEMBLY CO2
SPACER
yoke assembly
2
label (P/N
2
YOKE
MOUNTING
LOCATION
SP10506
CYLINDER SUPPORT
MOUNTING LOCATION
Figure 6: CO2YOKE AND CYLINDER SUPPORT MOUNTING LOCATIONS
9
CYLINDER
SUPPORT
Loading...
+ 20 hidden pages