NORTH AMERICAN DRÄGER Drager KIT Air Diss User manual

Page 1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
KIT,AIRDISS&CO2YOKE
INSTALLATIONPROCEDURE
NOTE: Thisprocedureoutlinesanextremelycomplexmodificationoftheanesthesia
machine.InstallationofthiskitshallbeperformedonlybyaNorthAmerican DrägerqualifiedTechnicalServiceRepresentativewiththenecessaryknowledge andexperience.
flowknobguards,andflowcontrolknobslocatedbelowthe100to1000ml/min. (fine)flowmeters.
1. TurntheSystemPowerswitchto ONanddisconnectallpipeline hoses.
2. Closeallcylindervalvesexceptthe O
cylindervalve.
2
3. Settheoxygenflowrateto5l/min.
4. OpentheN
Oflowcontrolvalveto
2
drainpressurefromthesystem.
5. ClosetheO
cylindervalve,and
2
closetheflowcontrolvalves.Press theO
FLUSHbuttontodrain
2
pressurefromthesystem.
6. TurntheSystemPowerswitchto STANDBYandremoveACpower fromthemachine.Disableallcircuit breakers.
7. Removethescrewssecuringthe tabletop,andremovethetabletop.
CAUTION: Use ESD control when
handling any electronic componentsorassemblies.
9. Removethevaporboxfrontcover. FormachineswithaMultispec
analyzer,removeallconnectionsto thePCBassemblyonthevaporbox frontplate,andremovethePCB fromtheplate.
10. Removethevaporboxbackcover.
11. Removethescrewsholdingthe angledfrontplateatthetopofthe flowmetershield,andremovethe plate.
12. Removeallflowcontrolknobs.
13. Removethetwoscrewssecuringthe knobguard;removetheguardand theflowmetershield.
8. Removetheflowmeterhousingback cover.
1
Page 2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CAUTION: Use ESD control when
handling any electronic components or assemblies.
14. Remove the flowmeter lights PCBs and channels from the studs on the O
and N2O flow channels.
2
15. Remove the stop pin hex nuts from each flow control valve.
16. Remove the flow control valves: hold each at the wrench flats and turn it counter-clockwise.
17. Remove each flowmeter tube by turning its retaining insert at the top of the assembly counter­clockwise until there is enough clearance to remove the tube from the channel.
18. Disconnect theflextubing connected to the N remove the O
O restrictor housing, and
2
and N2O restrictor
2
assemblies.
19. Remove the cross-over tubes connecting the O
and N2O fine and
2
coarse flowmeter ports.
24. Remove the three socket head cap screws securing the bottom flowmeter block to the channel assembly.
25. Pull the bottom block forward, disconnect the small diameter flex tubing from the block, and remove the block.
26. Disconnect the flex tubing from the O
and N2O pipeline gauges.
2
27. Remove the three socket head cap screws securing the flowmeter assembly, and remove the assembly.
28. Remove the kep nuts securing the cylinder and pipeline gauges to the flowmeter channel and remove the gauges.
29. Remove the socket head cap screws securing thetop block to the channel assembly.
30. Apply masking tape to the raised surfaces of and the entire inside area ofthe center flowmeter channel to prevent scratching.
20. Remove the manifold assembly connected to the O
and N2O outlets
2
at the top of the flowmeters.
21. Disconnect the 4-way fitting and restrictor assembly’s lower copper tube from the O
flowmeter inlet
2
port.
22. Disconnect the copper tube attached to the N
O flowmeter valve port.
2
23. Disconnect the copper tubes from the O
and N2O cylinder gauges.
2
31. Temporarilyattach the 4-gas bottom block to the channel with the screws that were previously removed from the original block.
32. Using the bottom block as a template, mark the circumference of the 4th gas flow control housing onto the channel: (Insert a flowmeter gasket (P/N 4102725) into the housing. This will provide a smaller diameter hole to aid in locating the center point.) The hole location is shown in Figure 1.
2
Page 3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
SP10501
7/64 IN. DIA. HOLE (DRILL)
11/16 IN. DIA.
CLEARANCE HOLE
(DRILL)
(LOCATE CENTER
USING 4-GAS BLOCK
AND GASKET AS
TEMPLATE)
1 7/8 IN.
Figure 1: DRILLING LOCATIONS: 4-GAS CONVERSION
3
Page 4
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
33. Remove the 4-gas bottom block and centering gasket.
CAUTION:
Perform the next four steps in a suitable location away from the anesthesia machine to prevent drilling chips from entering the pneumatic system.
34. Center punch the marked hole and drill an 11/16 in. dia. hole into the flowmeter channel. A hole saw is recommended. De-burr any sharp edges.
35. Measure and mark a location 1 in. up from the lower center spacer hole, and centered in the channel as shown in Figure 1.
36. Carefully drill a 7/64 in. dia. hole at this location, and de-burr any sharp edges.
37. Remove all masking tape previously applied, and carefully clean up any metal chips from the flowmeter assembly and surrounding area.
38. Using a small amount of Loctite #271 (red), thread a 3/16 in. tube straight fitting (P/N 4109402) into the back of a high pressure gauge (P/N 4110575-002).
39. Apply a small amount of Loctite #271 (red) to the threads of a 1/16 in. hose barb (P/N 4111771) and pipeline pressure gauge (P/N 4110575-001). Assemble the fittings with coupling (P/N 4103668) as shown in Figure 2.
4
Page 5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
40. Place the pipeline gauge assemblies into the upper holes of the gauge channel and secure each gauge with two 10-32 kep nuts (P/N HW55002 x2). See Figure 2. Reuse the original hardware to secure the O
and N2O
2
gauges.
FLOWMETER CHANNEL
PIPELINE PRESSURE GAUGE
41. Place the cylinder gauge assemblies
SIDE VIEW
into the lower holes of the gauge channel and secure each gauge with two 10-32 kep nuts (P/N HW55002 x2). Reuse the original hardware to secure the O
and N2O gauges.
2
HOSE BARB
SP10502
FITTING
CYLINDER PRESSURE GAUGE
GAUGE MOUNTING NUTS
Figure 2: PIPELINE AND CYLINDER GAUGE ASSEMBLY AND MOUNTING
5
COUPLING
STRAIGHT
FITTING
(2X EACH GAUGE)
Page 6
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
42. Apply a small amount of Loctite #271 (red) to the threads of two 2-56 x ¼ in. pan head screws (P/N HW02000 x2). Place a #2 lock washer (P/N HW67012 x2) on each screw, and secure the two shoulder spacers (P/N4110843 x2) to the inside of the center flowmeter channel.
43. Reinstall the top block assembly to the flowmeter channel using the hardware previously removed.
44. Threadtwo inserts (P/N4102506 x2) into the center locations of the top flowmeter block until snug.
45. Secure the 4-gas bottom block (P/N
4110572) to the flowmeter channel using the hardware previously removed from the original block.
47. Install a #113 O-ring (P/N 4102792 x4) in the grove of each flowmeter insert on all four inserts. Apply a small amount of Parker Super-O­Lube to these O-rings.
48. Thread the chromed flowmeter insert into the upper hole of the bottom block, and the remaining brass inserts into the lower holes until snug. DO NOT overtighten the inserts.
49. Install the flow control valves (P/N 4103352 x2) into the flowmeter inserts. Reuse the original O N
O flow control valves. CAUTION:
2
2
and
Before tightening the cartridge, rotate the valve shaft several turns counter-clockwise to prevent bottomingthe valve element into the seat when the cartridge is tightened.
46. Install a #018 O-ring (P/N 4102336 x4) at the shoulder of each of the four bottom flowmeter inserts (consisting of P/N 4110576 x3 and
4110573) as shown in Figure 3.
50. Install the stop pin nuts (P/N 4103382 x2) onto the flow control assemblies. Reuse the original stop pin nuts for the O
and N2O
2
assemblies.
51. Thread two 10-32 x 1/16 in. hose barb fittings w/seal (P/N 4112707­001 x2) into the back of the bottom flowmeter block.
52. Remove the nuts and ferrules from eight ¼ in. tube straight fittings (P/N 4109408 x8). Apply a small amount of Loctite #271 (red) to the threads of each fitting, and thread the fittings into the top and bottom blocks at the locations marked A Figure 3.
6
Page 7
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
REAR VIEW OF FLOWMETER CHANEL
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
SP10503
AA
HOSE BARB FITTING W/SEAL
BOTTOM BLOCK MOUNTING SCREWS (3X)
STOP PIN NUT
FLOW CONTROL VALVE
A
AA AAA
DETAIL OF FLOW CONTROL
INSERT ASSEMBLY
CROSS-OVER
TUBES (2X)
HOSE BARB
FITTING W/SEAL
(SIDE VIEW)
O-RING, #113
INSERT
O-RING, #018
Figure 3: INSTALLATION OF FITTINGS AND FLOW CONTROL VALVE INSERT ASSEMBLY
7
Page 8
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
53. Assemble the compression fittings onto the open ended tubes on the 4­gas manifold assembly (consisting of P/Ns 4108483, 4109408 x4,4109410, 4110834-009 x3, and 4110837-013 x2) as shown in Figure 5. Ensure that the Swagelok fittings with the two-piece ferrules are assembled as shown in Figure 4.
TWO-PIECE FERRULE
Figure 4: ASSEMBLY OF SWAGELOK FITTINGS
Insert the copper tubes into the manifold assembly as shown in Figure 5. Before tightening the compression nuts, ensure that the tubing is inserted into the fittings as far as possible.
COPPER
TUBING
COMPRESSION
NUT
SP10504
54. Attach the copper tubes on the manifold assembly on the ports at the top of the flowmeter bank as shown in Figure 5. Ensure that the tubing is inserted into the fittings as far as possible, and tighten the compression nuts.
O2
MIX
55. Attach the correct gas-identifying label (P/N 4109871, 4109872, 4109873, 4109874, and 4109901) to each tube on the manifold assembly as shown in Figure 5.
SP10505
N20
AIR
MANIFOLD
TOP VIEW
Figure 5: FOUR GAS MANIFOLD ASSEMBLY WITH LABELS
8
CO2
Page 9
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
56. Reinstall the flowmeter channel assembly using the hardware previously removed.
57. Reinstall the cross-over tubes connecting the oxygen and nitrous oxide fine and coarse flowmeters. Verify that the correct identification labels are attached to each tube.
58. Mount the cylinder support (P/N
4106690) on the bottom frame rail at the back of the machine (see Figure
6) using two ¼-20 x 2½ in. hex head screws (P/N HW08009 x2), lock washers (P/N HW65010 x2) and flat washers (P/N HW66004 x2).
REAR VIEW
59. Mount the CO (consisting of P/Ns 4111792, HW06006, 1101624, 4104716, 4105929 x2, and 4112755-001), (and spacer block (P/N 1101593) on earlier models) on the upper frame rail of the machine as shown in Figure 6. Use two 5/16-24 x 1¾ in. socket head screws (P/N HW01058 x2) andlock washers (P/N HW65005 x2). Apply a CO
1101639) on this yoke.
YOKE ASSEMBLY CO2
SPACER
yoke assembly
2
label (P/N
2
YOKE
MOUNTING
LOCATION
SP10506
CYLINDER SUPPORT
MOUNTING LOCATION
Figure 6: CO2YOKE AND CYLINDER SUPPORT MOUNTING LOCATIONS
9
CYLINDER
SUPPORT
Page 10
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
60. Position the CO
cylinder pressure
2
regulator assembly (consisting of P/Ns 4103591, 4109415, 4102906x2, 4109409, and 4109401) as shown in Figure 7, with the Inlet marking toward the front of the machine.
61. Install two 10-32 x ½ in. set screws (P/N HW10001 x2) in the regulator mounting bracket and secure the regulator.
62. Connect a 3/16 in. dia. (P/N
4104215) pre-bent copper tube between the CO the inlet fitting on the CO
cylinder yoke and
2
cylinder
2
regulator.
TOP VIEW OF MACHINE
Ensure that the tubing is inserted into the ferrules correctly, and tighten the fittings securely. Install aCO
label (P/N 4109873 x2) on
2
each end of this tube.
63. Connect a 3/16 in. dia. (P/N
4104214) pre-bent copper tube between the CO
cylinder gauge and
2
the port marked "HP" on the CO cylinder regulator. Ensure that the tubing is inserted correctly, and tighten the fittings securely. Install aCO
(P/N 4109873 x2) label on
2
each end of this tube.
2
CO2 CYLINDER YOKE
3/16 IN. DIA. COPPER TUBE
SP10507
3/16 IN. DIA. COPPER TUBE
TO CYLINDER PRESSURE GAUGE
"HP"
PORT
SET
SCREWS (2X)
CO2 CYLINDER
PRESSURE
REGULATOR
"INLET" PORT
Figure 7: CO2CYLINDER REGULATOR MOUNTING AND H.P. CONNECTIONS
10
Page 11
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
64. Reattach the copper tubespreviously removed from the O
and N2O
2
cylinder gauges, and tighten the fittings securely. Verify that the gas identification labels are correct.
65. Reattach the ¼ in. dia. copper tubes previously removed from the O N
O flow control valve ports and the
2
2
and
top flowmeter blocks, and tighten the fittings securely.
66. Remove the ¾ in. dia. plastic plug from the flowmeter housing directly abovethe O
pipelineinlet assembly.
2
67. Install a DISS air inlet housing assembly (consisting of P/Ns 4113363 and 4102886) in the hole where the plug was removed, using a -18 hex nut (P/N HW52002) and lock washer (P/N HW67001). Orient the DISS connector at a 70° angle as shown in Figure 8. Apply an AIR label (P/N 4102742) to the inlet housing.
68. Apply a small amount of Loctite #271 (red) to the threads of the filter assembly (consisting of P/Ns 4102532 and 4106897) and install it into the air inlet housing.
69. Apply a small amount of Loctite #271 (red) to the threads of tee fitting (P/N 4102760). Install it into the filter assembly as shown in Figure 8, with its side port facing downward.
70. Apply a small amount of Loctite #271 (red) to the threads of a NPT x 1/16 in. hose barb fitting (P/N
4111771). Install it into the end port of the tee as shown in Figure 8.
71. Remove the nut and ferrules from ¼ in. tube to NPT elbow fitting (P/N
4109410). Using a small amount of Loctite #271 (red) install it in the remaining port as shown in Figure
8. Reattach the nut and ferrules on the fitting (ref. Figure 4).
72. Install an 8 in.length of flex tubing (P/N ML08003 x8) between the air pipeline gauge and the hose barb fitting installed in the previous step. Secure each end of the tubing with a press-on clamp (P/N 4104161 x2), and install an AIR label (P/N 4109872 x2) on each end of this tube.
73. Reattach the flex tubing previously removed from the O
and N2O
2
pipeline gauges. Secure the connections with the press-on clamps previously removed (use tie straps P/N 4106068 x2 for early style nylon fitting), and verify that the gas identification labels are correct.
74. Remove the ¼ in. x 10-32 plug from the left port of the 4-way fitting and restrictor assembly, and insert a 10­32 x 1/16 in. hose barb fitting w/seal (P/N 4112707-001) into this port.
11
Page 12
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
SP10508
SIDE VIEW
FILTER
AIR INLET HOUSING
O
70
ELBOW
FITTING
REAR VIEW
TEE FITTING
HOSE BARB FITTING
PRESS-ON
HOSE CLAMP (2X)
FLEX TUBING
AIR LABEL (2X)
Figure 8: AIR PIPE LINE INLET ASSEMBLY
12
Page 13
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
75. Connect the short end of a ¼ in. dia. "L" tube (P/N 4110837-015) to the CO2cylinder pressure regulator as shown in Figure 9. Install a CO label (P/N 4109873 x2) on each end of this tube.
76. Connect the other end of the "L" tube to a ¼ in. T fitting (P/N
4108636), and install a ¼ in. plug (P/N 4103072) in the side port of the T fitting, and tighten the fittings securely.
77. Connect the short end of a ¼ in.dia. pre-bent tube (P/N 4110279) to the remaining port on the tee fitting as
2
shown in Figure 9. Install a CO
2
label (P/N 4109873 x2) on each end of this tube.
SP10509
1/4 IN. DIA. TUBE TO CO2 OFPD
PLUG
Figure 9: CO2CYLINDER REGULATOR L.P. CONNECTIONS
1/4 IN. DIA.
"L" TUBE
TEE FITTING
13
Page 14
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
78. Connect either end of a ¼ in.dia. S tube (P/N 4110836-006) to the CO flow control valve port. See Figure
10. Install a CO
label (P/N 4109873
2
x2) on each end of this tube.
79. Position the CO
OFPD (consisting
2
of P/Ns 4111771, 4103681, 4102713, 4109408, and 4109410) in the flowmeter housing and connect the side fitting on the OFPD to the previously installed S tube from the CO
flow control valve. Tighten the
2
fittings securely
80. Connect the bottom fitting on the CO
OFPD to the copper tube from
2
the CO
regulator that was
2
previously installed. Tighten the fittings securely.
81. Position the Air OFPD (consisting of
2
P/Ns 4111771, 4103681, 4102713, 4102784, 4015815, and 4109410 x2) in the flowmeter housing and connectan Stube (P/N4110836-004) between the Air flow control valve and the side fitting on the OFPD. Tighten the fittings securely. See Figure 10. Install an AIR label (P/N 4109872 x2) on each end of this tube.
82. Connect a copper tube (P/N
4108803) between the air pipeline inlet assembly elbow fitting and the elbow fitting at the lower end of the check valve on the Air OFPD, and tighten the fittings securely. Install an AIR label (P/N 4109872 x2) on each end of this tube.
83. Attach the minimum flow cutoff valve assembly (consisting of P/Ns 4111771, 4102055, 4110792-013, 4103549,and 4112707-001 x2) to the N
O flowmeter crossover tube with
2
two tie straps (P/N 1101732 x2) as shown in Figure 10.
14
Page 15
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
TIE STRAP (2X)
MINIMUM FLOW
CUTOFF VALVE
SP10510
FROM CO2
FLOW CONTROL
VALVE
FROM AIR
FLOW CONTROL
VALVE
AIR OFPD
CO2 OFPD
FROM CO2
REGULATOR
FROM AIR PIPELINE INLET
Figure 10: OFPD AND MINIMUM FLOW VALVE INSTALLATION
15
Page 16
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
84. Mount the gas selector valve (consisting of P/Ns 4112707-001 x4, HW02028 x2, 4108749 x2, 4103621 x2, and 4110518) to the floor of the vapor box, oriented as shown in Figure 11, with two 10-32 x 5/16 in. socket head screws (P/N HW01022 x2), lock washers (P/N HW67006 x2) and flatwashers (P/N HW66003 x2).
NOTE:
Skip the next nine steps if the ORMC does not have a wire harness and switch.
85. Securethe pressureswitch assembly (consisting of P/Ns 4103444, 4106367, and 4112707-001 x2 to the copper tubing in the vapor box with two tie straps (P/N 1101732 x2) as shown in Figure 11.
86. Disconnect all wires from the ORMC wire harness adapter.
87. Attach the female connector of the ORMC wire harness adapter to the terminal on the right side of the pressure switch.
88. Attach the female connectors of the AIR OPT/PRESS SW wire harness (P/N 4106324) to the male connectors on the wire harness adapter.
89. Connect the WHT/YEL wire from the ORMC to the AIR OPT/PRESS SW wire harness.
90. Connect the WHT wire from the alarm channel to the remaining terminal on the AIR OPT/PRESS SW wire harness.
SP10511
TO 4-WAY FITTING PORT "A"
TO AIR OFPD
TO MINIMUM FLOW VALVE
FROM ALARM CHANNEL
FROM ORMC
WH/YEL
WH
WH/YEL
VAPOR BOX
TOP VIEW
NYLON TEE
FITTING
GAS SELECTOR
VALVE ASSEMBLY
PRESSURE SWITCH
ASSEMBLY
Figure 11: GAS SELECT VALVE AND PRESSURE SWITCH CONNECTIONS
16
Page 17
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
91. Connect the WHT/YEL wire from the alarm channel harness to the remaining terminal on the pressure switch assembly.
92. Install a 16 in. length of flex tubing (P/N ML08003 x16) between the right side port of the pressure switch and the front port on the gas selector switch assembly. Secure each end of the tubing with a press­on clamp (P/N 4104161 x2). Install an O
label (P/N 4109871 x2) on
2
each end of this tube.
93. Connect a 14 in. length of flex tubing (P/N ML08003 x14) to the left side port of the pressure switch in the vapor box, and secure it with a press-on clamp. Connect the other end of the tubing to the top hose barb on the minimum flow valve and secure it with a press-on clamp (P/N 4104161 x2). Install an O
label
2
(P/N 4109871 x2) at each end of this tube.
NOTE: Skip the next step if the
ORMC has a wire harness and switch.
94. Connect a 33 in. length of flex tubing (P/N ML08003 x33) between the front port on the gas selector switch and secure it with a press-on clamp. Connect the other end of the tubing to the top hose barb on the minimum O
flow valve and secure
2
it with a press-on clamp (P/N 4104161 x2). Install an O
label (P/N
2
4109871 x2) at each end of this tube.
95. Connect an 8 in. length of flex tubing (P/N ML08003 x8) from the side port of the rear valve on the gas selector and secure it with a press­on clamp (P/N 4104161). Connect the other end of the tubing to the center port of the nylon tee fitting (P/N 4102337), and secure it with a tie strap (P/N 4106068). Install an O
label (P/N 4109871 x2) on each
2
end of this tube.
96. Connect an 11 in. length of flex tubing (P/N ML08003 x11) to the side port of the front valve on the gas selector, and secure it with a press-on clamp (P/N 4104161). Connect the other end of the tubing to the nylon tee fitting installed in the previous step, and secure the connection with a tie strap (P/N
4106068). Install an O
label (P/N
2
4109871 x2) on each end of this tube.
97. Connect a 24 in. length of flex tubing (P/N ML08003 x24) from the rear port on the gas selector assembly to the hose barb on the top of the air OFPD in the flowmeter housing. See Figures 11 and 12. Secure each end with a press-on clamp (P/N 4104161 x2). Install an O
label (P/N 4109871 x2) on each
2
end of this tube.
98. Cut the flex tubing two inches above the hose barb on the top of the air OFPD and install a nylon tee fitting (P/N 4102337) as shown in Figure
12. Install an O
label (P/N 4109871
2
x2) on each end of this junction.
17
Page 18
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
FROM PRESSURE SWITCH ASSEMBLY
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
FROM TEE FITTING IN VAPOR BOX
TEE FITTING "C" (EXISTING)
4-WAY FITTING
FROM GAS
SELECTOR
TEE FITTING
(INSTALL)
2 IN. LENGTH
6 IN. LENGTH
N2O OFPD
AIR OFPD
SP10512
Figure 12: FLEX TUBING CONNECTIONS
CO2 OFPD
18
Page 19
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
99. Connect a 6 in. length of flex tubing (P/N ML08003 x6) to the side port of the nylon tee installed in the previous step.Connect the other end of this tube to hose barb on the top of the CO
OFPD, and secure this
2
connection with a press-on clamp (P/N 4104161). Secure all nylon tee fitting junctions with tie straps (P/N 4106068 x3). Install an O
label (P/N
2
4109871 x2) on each end of the tubing.
100. Connect a 17 in. length of flex tubing (P/N ML08003 x17) from the remaining port of the nylon tee fitting in the vapor box and secure it with a tie strap (P/N 4106068). Connect the other end of the tubing to the side port "A" on the 4-way fitting and secure it with a press-on clamp (P/N 4104161). Install an O label (P/N 4109871 x2) on each end of this tube. See Figure 12.
101. Remove the existing tubing from side port "B" on the 4-way fitting. Connect an 8 in. length of flex tubing (P/N ML08003 x8) from side port "B" on the 4-way fitting to the bottom port of the minimum flow valve, and secure each connection with a press-on clamp (P/N 4104161 x2). Install an O
label (P/N
2
4109871 x2) on each end of this tube. See Figure 12.
102. Cut the tie strap securing the flex tubing on the side port of tee fitting "C" in the flowmeter housing and remove the tubing.
103. Connect a 10 in. length of flex tubing (P/N ML08003 x10) to the side port of nylon tee fitting "C", and secure it with a tie strap (P/N
4106068). Connect the other end of the tubing to the side port of the minimum flow valve, and secure it with a press-on clamp (P/N
4104161). Install an O
label (P/N
2
4109871 x2) on each end of this tube.
104. Reconnectthe flex tubing previously removed from the hose barb located below the O
flow control valve, and
2
secure the connection with the press-on clamp previously removed. Verify that an O
label is attached to
2
each end of this hose.
105. Reconnectthe flex tubing previously
2
removed from the hose barb located below the N
O flow control valve,
2
and secure the connection with the press-on clamp previously removed. Verify that a N
O label is attached
2
to each end of this hose.
106. ConnecttheAir pipeline supplyhose to the machine.
107. Install a CO
cylinder in the new
2
yoke. Verify the presence of two index pins on the yoke assembly and the correct engagement with the cylinder index holes. Open one cylinder valve each of Oxygen, Nitrous Oxide and CO
108. Close the N
O flow control valve;
2
.
2
ensure that all other valves are fully open.
19
Page 20
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
109. Turn the System Power switch ON, and depress the rear gas selector valve to flush any debris from the new piping, and return the switch to the STANDBY position.
110. Install the CO
restrictor housing
2
(P/N 4103440) with a yellow restrictor (P/N 4110738-006) and O­ring (P/N 4101872) as shown in Figure 14.
111. Place gaskets (P/N 4102724 x2) on top of the CO
restrictor housing and
2
in the insert at the top flowmeter block for CO
112. Carefully install the CO
.
2
flowtube
2
(P/N 4112557-001) with the CO marking facing forward as shown in Figure 14.
113. Install the AIR restrictor housing (P/N 4103440) with an O-ring (P/N
4101872) as shown in Figure 14.
114. Place gaskets (P/N 4102724 x2) on top of the AIR restrictor housing and in the insert at the top flowmeter block for AIR.
115. Carefully install the AIR flowtube (P/N 4112558-001) with its AIR marking facing forward as shown in Figure 14.
116. Close the AIR and CO
flow control
2
valves.
117. Reinstall the O
restrictor assembly
2
that was previously removed from the bottom flowmeter block.
119. Turn the System Power switch to ON to flush any debris from the new piping, and return the switch to the STANDBY position.
120. Carefully reinstall the O flowtube with the O
marking facing
2
2
coarse
forward as shown in Figure 14.
121. Turn the System Power switch ON to flush any debris from the new piping, and return the switch to the STANDBY position.
122. Reconnectthe flex tubing previously removed from the N
O restrictor
2
assembly and verify that the connection is secured with a press-
2
123. Reinstall the N
on clamp.
O restrictor
2
assembly that was previously removed from the bottom flowmeter block.
124. Turn the System Power switch ON. Set the O N
O flow control valve fully to its
2
flow to 4 l/min. Open the
2
stop to flush any debris from the new piping. Return the System Power switch to STANDBY.
125. Carefully reinstall the N flowtube with the N
O fine
2
O marking
2
facing forward as shown in Figure
14.
126. Turn the System Power switch ON to flush any debris from the new piping, and return the switch to the STANDBY position. Close the N
O
2
flow control valve.
118. Carefully reinstall the O flowtube with the O
marking facing
2
forward.
fine
2
20
Page 21
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
SP10513
GASKET
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
LABEL "CAUTION-
HYPOXIC MIXTURE"
LABEL "O2 & N2O"
GASKET
LABEL "ALL GASES"
CO2 FLOWMETER TUBE
4-GAS CO2 FLOWMETER SHIELD
AIR FLOWMETER TUBE
GASKET
RESTRICTOR (YELLOW)
P/N 4110738-006
RESTRICTOR HOUSING
O-RING, #010
FLOW CONTROL KNOB
AND CO2 LABEL
GASKET
RESTRICTOR HOUSING
O-RING, #010
FLOW CONTROL KNOB
AND O2 LABEL
Figure 13: AIR AND CO2FLOWTUBE INSTALLATION
21
FLOW CONTROL KNOB
AND AIR LABEL
FLOW CONTROL KNOB
AND N2O LABEL
Page 22
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
127. Carefully reinstall the N flowtube with the N
O coarse
2
O marking
2
facing forward as shown in Figure
14.
128. Slidethe4-gas flowmeter lights PCB (P/N 4110857) into the flow lights channel (P/N 4110582) and attach the wire harness to the assembly. (The black wire is connected to the terminal at the back of the PCB.) Insert the assembly into the shoulder studs at the center flowmeter. Reinstall the remaining lights in the same manner.
129. Install the 4-gas flowmeter shield (P/N 4110587-004).
130. Reattach the 3-gas flowmeter knob guard to the flowmeter assembly reusing the hardware previously removed.
131. Use the appropriate vapor box front plate (P/N 4111510 or 4111520) on machines with a multigas analyzer. Assemble the gas selector switch cam assembly (consisting of P/Ns 4111693, 4103169, 4103423, 4108461, 4108462, 4109866, and
4109867) to the new vapor box front panel with two 8-32 x in. socket head screws (P/N HW01012 x2) and lock washers (P/N HW67000 x2). See Figure 13.
132. If applicable, attach the vapor indicator PCB assembly to the front panel with the hardware that was previously removed. Restore all wire connections to the vapor indicator PCB.
133. Installthe vapor box front panel and ensure that the selector switch cams operate the gas select valves correctly. Set the selector to the ALL GAS position. Attach a HYPOXIC MIXTURE label (P/N 4110953) to the left of the gas selector switch. If applicable, apply arrow indicator labels (P/N 4112055 x3) at the three locations shown in Figure 13.
134. Reinstall the angled front plate at the top of the flowmeter shield reusing the hardware previously removed. On earlier machines this plate is retained by two screws inside the flowmeter housing. (On later machines with a one-piece plate, this plate was installed in the previous step as part of the vapor box front cover.)
135. Install the AIR and CO
flow knobs
2
(P/N 4103736 x2) and reinstall the N
O and O2flow knobs.
2
136. Turn the System Power switch ON. Set the Gas Selector switch to ALL GAS. Verify that the O and CO
flowmeters operate
2
2,N2
O, AIR,
properly over their full range. Close all flow control valves.
137. Connect the 15 mm connector to a Capnomed flowmeter test stand and verify that each "off stop" is properly set. Readjust the stops as necessary. Removethe testflowmeter. Turnthe System Power switch ON. Verify that all flow meters operate properly over their full range.
22
Page 23
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
VAPOR BOX TOP VIEW
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
GAS SELECTOR SWITCH ASSEMBLY
HYPOXIC MIXTURE CAUTION LABEL
SP10514
VAPOR BOX FRONT VIEW
Figure 14: GAS SELECTOR SWITCH ASSEMBLY
SWITCH ASSEMBLY
MOUNTING SCREWS AND
LOCK WASHERS (2X)
VAPOR BOX
FRONT PANEL
ARROW
LABELS (3X)
23
Page 24
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
INSTALLATION PROCEDURE (continued)
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
138. Enable the circuit breakers and connect AC power to the machine.
139. Removeany existing labels from the flowmeter knobs. Install the correct label (P/Ns 4103905, 4103908, 4103904, and 4103178) on each flowmeter knob.
TOP VIEW OF MACHINE
WITH DRAWER OPEN
140. Attach a new instruction label (P/N
4104818) to the inside of the top drawer as shown in Figure 15.
INSTRUCTION LABEL
System Power
Figure 15: NEW INSTRUCTION LABEL LOCATION
24
TOP
DRAWER
SP10515
Page 25
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CO2AND AIR CIRCUIT ADJUSTMENT AND TEST
1.0 CYLINDER PRESSURE REGULATOR ADJUSTMENT
1.1 Open the CO
flow control valve
2
to the full open position.
1.2 Loosen or remove the acorn nut
from the CO
regulator.
2
1.3 Turn the System Power switch
to ON.
1.4 Set the Gas Select switch to the
ALL GAS position.
1.5 Open the O
and CO2cylinder
2
valves.
1.6 Turn the CO
regulator
2
adjustment screw until the CO flowmeter indicates a flow rate of 550 ml/min.
2.0 HIGH PRESSURE CYLINDER LEAK TEST
2.1 Turn the System Power switch
to STANDBY.
2.2 Open all cylinder valves and
allow the gauge pressures to stabilize.
2.3 Close all cylinder valves and
remove all cylinders from their yokes. If applicable, close the Auxiliary O
flowmeter valve.
2
Observe the cylinder pressure gauges. The pressure should not drop more than 50 psi over the next two minutes.
2
2.4 Reattach the cylinders to the
yokes.
Replace the acorn nut on the regulator.
2.5 Turn the System Power switch
to ON and repeat the high pressure leak test.
25
Page 26
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CO2AND AIR CIRCUIT ADJUSTMENT AND TEST (continued)
3.0 LOW PRESSURE LEAK TEST
3.1 Ensure that all flow control valves are closed.
3.2 Turn the System Power switch to STANDBY.
3.3 Remove the 15mm connector from the FRESH GAS OUTLET. Connect a test gauge and B.P. bulb to the fresh gas outlet, and pressurize the system to 50 cm H
O.
2
3.4 The pressure should not drop more than 10 cm H
O in thirty
2
seconds.
3.5 Disconnect the test gauge. Reconnect the 15mm connector to the FRESH GAS OUTLET.
5.0 FLOWMETER TEST
5.1 Open the O
cylinder valve.
2
5.2 Adjust the flow of AIR over the full range of the flowmeter. The float should move freely over the entire range.
5.3 Adjust the flow of CO maximum position. The flow should be between 500 and 600 ml/min. If the flow is not correct, repeat test procedure
1.0 (cylinder pressure regulator adjustment).
5.4 Adjust the O
flowmeter to 3
2
l/min.
5.5 Adjust the flow of N
O over the
2
full range of the flowmeter. The float should move freely over the entire range.
to the
2
4.0 OFPD TESTS
4.1 Turn the System Power switch to ON.
4.2 Open the O
,N2O, and CO
2
cylinder valves.
4.3 Set the O
,N2O, and AIR flow
2
rates to 1.0 l/min.; set the CO flow rate to 500 ml/min.
4.4 Close the O
cylinder valve.
2
When the oxygen flow stops, the N
O, AIR and the CO2flows
2
must also drop to zero.
6.0 GAS SELECTOR SWITCH TEST
6.1 With the switch in the ALL GAS position, open all of the
2
flow control valves and verify that all gases are able to flow.
6.2 Turn the Gas Select switch to O
2
+N2O. The flow of AIR and
2
CO
should stop.
2
6.3 Close all of the flow control valves. There should be a minimum oxygen flow.
6.4 Turn the Gas Select switch to ALL GAS. The minimum flow should stop.
26
Page 27
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CO2AND AIR CIRCUIT ADJUSTMENT AND TEST (continued)
7.0 OXYGEN CONCENTRATION VERIFICATION TEST
7.1 Connect a 22mm hose between
the bellows VENTILATOR HOSE andthe inspiratory valve terminals.
7.2 Activate the waste gas
scavenger.
7.3 Set theManual/Automaticvalve
to BAG.
7.4 Occlude the bag mount.
7.5 Insert the sensor from a
calibrated O
Med into the valve
2
dome on the inspiratory valve.
7.6 Set the O
7.7 Verify that the O
flow to 4 l/min.
2
concentration
2
is between 97% and 100% within 3 minutes.
7.8 Set the N
7.9 Verify that the O
O flow to 2 l/min.
2
concentration
2
is between 64% and 70% within 3 minutes.
7.12 Verify that the O
concentration
2
is between 71% and 77% within 3 minutes.
7.13 Close the AIR flow control valve.
7.14 Set the O
7.15 Set the CO
7.16 Verify that the O
flow to 1 l/min.
2
flow to 500 ml/min.
2
concentration
2
is between 64% and 70% within 3 minutes.
7.17 Close the CO
and O2flow
2
control valves.
7.18 Remove the occlusion from the bag mount.
7.19 Remove the 22mm hose from the bellows VENTILATOR HOSE and inspiratory valve terminals.
7.10 Close the N
O flow control
2
valve.
7.11 Set the AIR flow to 2 l/min.
27
Page 28
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
CO2AND AIR CIRCUIT ADJUSTMENT AND TEST (continued)
8.0 ORMC TEST NOTE:
If the alarm channel on the machine does not have an O
2/N2
lamp, disregard the alarm functions in the following tests.
8.1 Fully open the N
O flow control
2
valve. Slowly open and close the O
flow control valve, and
2
observe that the ORMC is controlling the flow of nitrous oxide. There should be no ORMC alarm with the Gas Select switch in the ALL GAS position.
8.2 Turn the Gas Select switch to
O
+N2O. Slowly open and close
2
the O
flow control valve, and
2
observe that the ORMC is controlling the flow of nitrous oxide. The ORMC alarm should function correctly with the Gas Select switch in the O position.
O alarm
+N2O
2
9.0 RE-ASSEMBLY & FINAL TESTING
9.1 Close all cylinder valves except the O
9.2 Set the O
9.3 Open the N
cylinder valve.
2
flow rate to 5 l/min.
2
O flow control valve
2
to drain pressure from the system.
9.4 Close the O open the O and press the O
cylinder valve,
2
flow control valve,
2
FLUSH button
2
to drain pressure from the system.
9.5 Turn the System Power switch to STANDBY.
9.6 Reinstall the table top on the machine.
9.7 Reinstall the vapor box back coverandthe flowmeter housing back cover.
9.8 Perform a complete PMS procedure on the machine.
28
Page 29
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTSRETURN TO CD-ROM TABLE OF CONTENTS
Technical Service Department 3122 Commerce Drive Telford, PA 18969 (215) 721-5402 (800) 543-5047 (215) 723-5935 Fax
Quality Service for Life
®
Part Number: SP00105 Rev: B
Date: June 21, 1996 © 1996 N.A.D., Inc.
Loading...