Conversion Instructions 84 13 790 for
kit/option “NeoFlow”
1 Information
1.1 Intended use
This conversion instructions is applicable for:
• Option NeoFlow 84 13 785
• Kit NeoFlow 84 13 563
These options and conversion kits can be used for the following devices:
Description P/N Evita 4 Evita 2 dura EvitaXL Evita 4 / 2 dura
with option EvitaXL
Option NeoFlow 84 13 785 Yes Yes Yes * Yes *
Kit NeoFlow 84 13 563 Yes Yes Yes * Yes *
* Approved only if language specific Instructions for Use / Operating Instructions
labeled with “Supplement to the Instructions for Use / Operating Instructions of
I
Lübeck
EvitaXL” is available!
Schutz nach DIN 34 beansprucht.
Use of this document / contents is forbidden
without express authority. All rights reserved.
1.2 Bill of material of option / kit
Dräger Medical AG & Co. KGaA, D-23542
1.2.1 Option NeoFlow P/N 84 13 785
- 1 Conversion instructions 84 13 790
- 1 Assembly instructions PCBs 84 15 604
- 2 Countersunk flat head screw 13 43 246
- 1 PCB Pediatric-Flow 83 06 631
- 2 Flow-Sensor (ISO) 84 11 130
- 2 Kit water trap 84 13 125
- 1 Flow sensor cable 84 09 626
- 1 Label NeoFlow 84 13 283
- 2 Fold hose flex 0.13m 84 09 634
1.2.2 Kit NeoFlow P/N 84 13 563:
- 1 Option NeoFlow 84 13 785
- optional 1 Instructions for Use / Operating Instructions depending on device configuration
and region:
For Evita 4 and Evita 2 dura:
- 1 Instructions for Use NeoFlow de 90 29 556
- 1 Instructions for Use NeoFlow en 90 29 557
8413790 - 07 , en
RELEASED
Seite 1 von 4

- 1 Instructions for Use NeoFlow fr 90 29 558
- 1 Instructions for Use NeoFlow es 90 29 559
- 1 Instructions for Use NeoFlow nl 90 29 561
- 1 Instructions for Use NeoFlow it 90 29 560
- 1 Instructions for Use NeoFlow sv 90 29 562
- 1 Operating Instructions NeoFlow enUS 90 29 563
- 1 Instruction for Use record 90 37 157
For EvitaXL and Evita 4 / 2 dura with option EvitaXL:
- 1 Instructions for Use NeoFlow EvitaXL de 90 37 566
- 1 Instructions for Use NeoFlow EvitaXL en 90 37 682
- 1 Instructions for Use NeoFlow EvitaXL fr 90 37 719
- 1 Instructions for Use NeoFlow EvitaXL es 90 37 718
- 1 Instructions for Use NeoFlow EvitaXL nl 90 37 721
- 1 Instructions for Use NeoFlow EvitaXL it 90 37 720
- 1 Instructions for Use NeoFlow EvitaXL sv 90 37 722
- 1 Operating Instructions NeoFlow EvitaXL enUS 90 37 683
- 1 Instruction for Use record 90 37 157
1.3 Conversion information
Required software version installed in device for conversion:
- Evita 4: SW 2.20 or higher
- Evita 2 dura: SW 3.10 or higher
I
- EvitaXL: SW 5.00 or higher
Lübeck
- Evita 4 / 2 dura with OptionXL: SW 5.00 or higher
Schutz nach DIN 34 beansprucht.
Dräger Medical AG & Co. KGaA, D-23542
Use of this document / contents is forbidden
without express authority. All rights reserved.
2 Installation and configuration of PCB
Electrostatic discharge may damage electrostatic sensitive devices. Use an
antistatic mat and a wrist strap when handling electrostatic sensitive devices.
• Switch off EvitaXL / Evita 4 / Evita 2 dura and disconnect device from mains.
• Create ESD conditions.
• Install Pediatric-Flow PCB (NeoFlow) of the kit / option according to assembly
instructions PCBs 84 15 604 in electronics on rear side of device:
- If PCBs have been installed differently during previous conversions reinstall them
according to assembly instructions PCBs 84 15 604.
- Ensure that PCBs are firmly in place on both sides in the lower groove of PCB slots.
- Fix PCBs with the two screws 13 43 246 each.
- Cover free slots with cover plates. Fix cover plates with two screws 13 43 246 each.
8413790 - 07 , en
Seite 2 von 4
RELEASED