NORTH AMERICAN DRÄGER Babylog 8000, Babylog 8000SC, Babylog 8000 Plus User manual

Babylog 8000 plus
Infant Care Ventilator
Operating Instructions Software 5.n
D
MEDICAL
MT-549-97
NOTICE
This document contains information in which Draeger Medical, Inc. claimed proprietary rights. The information may not be reproduced in whole or in part except as authorized in writing by Dräger. This information is the property of Draeger Medical, Inc., it is provided solely for the use intended.
Repairs/Modifications Repairs on this device shall be performed only by DrägerService or
Factory Authorized Service Centers. Information about repairs can be obtained from Dräger or Authorized Dealers. Draeger Medical, Inc. will not be responsible for injury to persons or damage to property arising directly or indirectly out of unauthorized repairs or modifications to this device. Furthermore, any unauthorized repairs or modifications void any warranty extended by Dräger.
This document is provided for your information only. It will not be exchanged or updated without request.
Trademarks
The Dräger name and logo are registered trademarks of Dräger.
Babylog
is a registered trademark of Dräger.
ThinkJet is a registered trademark of Hewlett Packard Corporation Dräger Medical AG & Co. KGaA,1997-2001
All rights reserved, subject to modifications
MT 550-97
Working with these Operating Instructions
3
Working with these Operating Instructions
Header line – the title... of the main chapter
The title of the respective sub-section is printed underneath the main header – to help you find your way quickly from subject to subject.
Page body... the Operating Instructions
combine text and illustrations. The information is presented in the form of required steps of action, giving the user hands-on experience in learning how to use the ventilator.
Left-hand column... the text
provides explanations and instructs users step-by-step in the practical use of the product, with short, clear instructions in easy to follow sequences. Bullet points indicate separate actions. Where several actions are described, numbers are used to refer both to relevant details in the illustrations and to specify the sequence of actions.
Calibrating the O2-Sensor Manually
After sensor has been replaced, but possible at all times.
1 Press »Cal Config.« menu key.
2 Press »O2-Cal« menu key.
3 After about 5 minutes, O2-Cal disappears from
the display, calibration is now completed.
1
D Babylog 8000 plus
2
Preparation
Calibrations
Right-hand column... the illustrations
provide the visual reference for the text and make it easier to locate the various parts of the device. Elements mentioned in the text are highlighted. Unnecessary details are omitted. Rendering of screen displays guide the user and allow to reconfirm actions performed.
Typing conventions... Controls are designated as »Control Name« , e.g:
»PEEP/CPAP«
Screen messages are printed in bold, e.g:
Calibrate flow sensor!
The respective text message may now be cleared: 1 Press »Confirm« key.
1
D Babylog 8000 plus
WARNING !
Strictly follow this Operator's Instruction Manual Any use of the product requires full under-
standing and strict observation of all portions of these instructions. The equipment is only to be used for the purpose specified under "Intended Use" (see page 18) and in conjunction with appropriate airway monitoring (see page 19). Observe all WARNINGS and CAUTIONS as rendered throughout this manual and on labels on the equipment.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
4
Contents
Contents
Important Safety Information READ THIS FIRST ! 7
Intended Use 17
Operating Concept 21
Preparation 27
Operation 51
Care 97
Troubleshooting 105
What's What 111
Technical Data 119
Theory of Operation 129
Ordering Information 143
Index 144
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
5
This page intentionally left blank
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
6
Important Safety Information
Contents
Important Safety Information
Contents
Operator's Responsibility for Patient Safety.................................................8
Limitation of Liability.................................................................................... 8
Warranty.....................................................................................................9
Definitions................................................................................................. 10
Summary of WARNINGS and CAUTIONS................................................ 10
General Precautions..................................................................................10
Precautions During Preparation................................................................. 12
Precautions During Operation....................................................................13
Precautions During Care........................................................................... 15
Precautions During Maintenance............................................................... 16
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
7
Important Safety Information
Operator's Responsibility for Patient Safety Limitation of Liability
Operator's Responsibility for Patient Safety
WARNING !
Strictly follow this Operator's Instruction Manual.
Any use of the product requires full under­standing and strict observation of all portions of these instructions. The equipment is only to be used for the purpose specified under "Intended Use" (page 18) and in conjunction with appropriate airway monitoring (see page
19). Observe all WARNINGS and CAUTIONS as rendered throughout this manual and on labels on the equipment.
The design of the equipment, the accompanying literature, and the labeling on the equipment take into consideration that the purchase and use of the equipment are restricted to trained professionals, and that certain inherent characteristics of the equipment are known to the trained operator. Instructions, warnings, and caution statements are limited, therefore, largely to the specifics of the Dräger design. This publication excludes refe­rences to various hazards which are obvious to a medical professional and operator of this equipment, to the consequences of product misuse, and to potentially adverse effects in patients with abnormal conditions. Product modification or misuse can be dangerous. Dräger disclaims all liability for the consequences of product alterations or modifications, as well as for the consequences which might result from the combination of this product with other products whether supplied by Dräger or by other manufacturers if such a combination is not endorsed by Dräger.
Limitation of Liability
Dräger's liability, whether arising out of or related to manufacture and sale of the goods, their installation, demonstration, sales representation, use, performance, or otherwise, including any liability based upon Dräger's Product Warranty, is subject to and limited to the exclu­sive terms and conditions as set forth, whether based upon breach of warranty or any other cause of action whatsoever, regardless of any fault attributable to Dräger and regardless of the form of action (including, without limitation, breach of warranty, negligence, strict liability, or otherwise).
THE STATED EXPRESSED WARRANTlES ARE IN LlEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING, WITHOUT LIMITATION, WARRANTlES OF MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE, OR NONINFRINGEMENT.
Dräger shall not be liable for, nor shall buyer be entitled to recover any special incidental, or consequential damages or for any liability incurred by buyer to any third party in any way arising out of or relating to the goods.
The operators of the ventilator system must recognize their responsibility for choosing appropriate safety moni­toring that supplies adequate information on equipment performance and patient condition. Patient safety may be achieved through a wide variety of different means ranging from electronic surveillance of equipment perfor­mance and patient condition to simple, direct observation of clinical signs. The responsibility for the selection of the best level of patient monitoring lies solely with the equipment operator (see also page 19, "Mandatory Ventilation Monitoring").
8
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
Warranty
Important Safety Information
Warranty
All Dräger products are guaranteed to be free of defects for a period of one year from date of delivery. The following are exceptions to this warranty:
1. The defect shall be a result of workmanship or material. Defects caused by misuse, mishandling, tampering, or by modifications not authorized by Dräger or its representatives are not covered.
2. Rubber and plastic components and materials are warranted to be free of defects at time of delivery.
3. Oxygen sensors capsules have a six-month limited warranty from the date of delivery.
Any product which proves to be defective in workman­ship or material will be replaced, credited, or repaired with Dräger holding the option. Dräger is not respon-sib­le for deterioration, wear, or abuse. In any case, Dräger will not be liable beyond the original selling price.
Application of this warranty is subject to the following conditions:
1. Dräger or its authorized representative must be promptly notified, in writing, upon detection of the defective material or equipment.
2. Defective material or equipment must be returned, shipping prepaid, to Dräger or its authorized representative.
3. Examination by Dräger or its authorized represen­tative must confirm that the defect is covered by the terms of this warranty.
4. Notification in writing, of defective material or equipment must be received by Dräger or its authorized representative no later than two (2) weeks following expiration of this warranty.
The above is the sole warranty provided by Dräger. No other warranty expressed or implied is intended. Rep­resentatives of Dräger are not authorized to modify the terms of this warranty.
Draeger Medical, Inc., Telford, PA
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
9
Important Safety Information
Definitions Summary of WARNINGS and CAUTIONS
Definitions
WARNING !
A WARNING statement refers to conditions with a possibility of personal injury if disregarded.
CAUTION !
A CAUTION statement designates the possibility of damage to equipment if disregarded.
NOTE: A NOTE provides additional information intended to avoid inconveniences during operation.
Inspection = examination of actual condition Service = measures to maintain specified
condition
Summary of WARNINGS and CAUTIONS
General Precautions
WARNING !
Strictly follow this Operator's Instruction Manual
Any use of the product requires full under­standing and strict observation of all portions of these instructions. The equipment is only to be used for the purpose specified under "Intended Use" (page 18) and in conjunction with appropriate airway monitoring (see page
19). Observe all WARNINGS and CAUTIONS as rendered throughout this manual and on labels on the equipment.
WARNING !
Babylog 8000 plus is solely intended for use by physicians or by allied health care profession­als on the order of a physician. All users must be properly trained and completely familiar with these Operating Instructions.
Repair = measures to restore specified
condition
Maintenance = inspection, service, and repair,
where necessary
Preventive = Maintenance measures at regular Maintenance intervals
Typing conventions in this manual
Controls (knobs, hard keys and menu keys) are designated as »Control Name«, e.g.:
»PEEP/CPAP« On-screen messages are printed in bold, e.g.:
Calibrate flow sensor!
WARNING !
This device is to be used only in rooms with line power installations complying with national safety standards for hospital patient rooms. (e.g., IEC 601.1, "Safety of Medical Equipment).
To maintain grounding integrity, connect only to a"hospital grade" receptacle.
Always disconnect supply before servicing.
WARNING !
DANGER, risk of explosion if used in the presence of flammable anesthetics.
This device is neither approved nor certified for use in areas where combustible or explo­sive gas mixtures with air or with nitrous oxide are likely.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
10
Important Safety Information
Summary of WARNINGS and CAUTIONS
WARNING !
Mobile telephones must not be used within 10 meters (33 feet) of the equipment. Mobile telephones can interfere with the function of electromedical equipment and therefore endanger the patient!
WARNING !
Do not use ventilator in conjunction with nuclear spin tomography (MRT, NMR, or NMI)!
Equipment malfunction may result.
WARNING !
Sidestream monitors may cause negative pressures in the event of a blocked inspiratory circuit. Always connect monitoring lines of a suctioning sidestream monitor via adapter 84 12 448 with safety valve.
WARNING !
Whenever a patient is connected to the venti­lator, constant attention by qualified medical staff is required in order to provide immediate corrective action in case of a malfunction.
WARNING !
If a fault is detected in the ventilator and its life-support functions are in doubt, ventilation must be started without delay with an inde­pendent ventilation device (resuscitation bag)
- using PEEP and/or increased inspiratory O2 concentration where necessary and appro­priate. The unit should then be removed from use and serviced by an authorized service technician.
CAUTION !
Restriction of Distribution
Federal Law and Regulations in the United States and Canada restrict this device to sale by or on the order of a physician.
WARNING !
Always use independent apnea monitor when the built-in apnea monitoring has been switched off. The operator of the ventilator must still assume full responsibility for proper ventilation and patient safety in all situations.
WARNING !
In case of malfunction of any of the built-in monitoring a substitute must be provided in order to maintain an adequate level of moni­toring. The operator of the ventilator must still assume full responsibility for proper venti­lation and patient safety in all situations.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
11
Important Safety Information
Summary of WARNINGS and CAUTIONS
Precautions During Preparation
WARNING !
Always install expiratory valve that has been cleaned and disinfected.
WARNING !
Treatment of batteries and O2-sensor capsules:
Do not throw into fire! Risk of explosion. Do not force open! Danger of bodily injury. Follow all local, state, and federal regulations
with respect to environmental protection when disposing of batteries and O2-sensor capsules.
WARNING !
Always use medical grade oxygen and air that is dry and free from dust and oil. Contami­nated gas may cause ventilator malfunction.
WARNING !
Always provide adequate ventilation! When the equipment is being operated with
O2, ensure that the ambient concentration does not exceed 24% O2. Otherwise, the risk of fire may be increased.
WARNING !
To maintain grounding integrity, connect only to a "hospital grade" receptacle. Always disconnect supply before servicing.
WARNING !
Dräger cannot warrant or endorse the safe performance of third party humidifiers for use with the Babylog 8000 plus infant ventilator. Specifically, combinations of humidifiers not intended for use with infant patients bear risks of delivering breathing gas not maintained at a proper temperature.
Excessive rainout in the inspiratory line of the breathing circuit may disable the flow and volume measuring capabilities of the Babylog 8000 plus ventilator.
Increased pneumatic resistance and compli­ance in the inspiratory line caused by a humidifier may result in inaccurate airway pressure readings.
We recommend contacting the manufacturers /distributors of heated humidifying devices about compliance of their products with the requested performance characteristics.
WARNING !
Dräger cannot warrant or endorse the safe performance of third party ventilator circuits for use with the Babylog 8000 plus infant ventilator. Increased pneumatic resistance caused by some circuit systems may result in inaccurate airway pressure readings.
Only use circuits with an inner diameter of at least 10 mm (0.4 inch)
We recommend contacting the manufacturers /distributors of infant patient circuits regarding compliance of their products with the requested performance characteristics (see p. 32).
WARNING !
In order to avoid any risk of electric shock in the event of faulty grounding of patient moni­toring equipment, do not use antistatic or electrically conductive patient circuits
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
12
WARNING !
Note arrows for inspiratory and expiratory connector on ventilator ports. If circuit ends are confused, humidification will not be effective.
Important Safety Information
Summary of WARNINGS and CAUTIONS
Precautions During Operation
WARNING !
To maintain grounding integrity, connect only to a "hospital grade" receptacle. Always disconnect supply before servicing.
WARNING !
Always use safety valve when connecting sampling CO2 or NO monitors. Otherwise negative airway pressures may result in the event of a blocked inspiratory circuit.
WARNING !
The ventilator is ready for operation only when:
- it is completely assembled with all required auxiliary equipment in place,
- all sensors are calibrated (O2, Flow),
- all checks have been completed successfully.
CAUTION !
Do not place containers of liquids on top of the Babylog 8000 plus ventilator. Liquids penetrating the ventilator can cause equipment malfunction and damage.
CAUTION !
If Air and O2 supply lines are contaminated with particles or moisture, as is likely during the use with a medical air compressor, use high pressure line filters that retain at least 0.1 micron particles and stop water droplets (e.g. Dräger part no. D 800 130, D 800 132).
WARNING !
Always use ventilator that has been cleaned and disinfected and has been successfully tested to be ready for operation.
WARNING !
Whenever a patient is connected to the venti­lator, constant attention by qualified medical staff is required in order to provide immediate corrective action in case of a malfunction. Not any and every critical condition can be expected to cause a ventilator alarm.
WARNING !
If a fault is detected in the ventilator and its life support functions are in doubt, ventilation must be started without delay with an inde­pendent ventilation device (resuscitation bag)
- using PEEP and/or increased inspiratory O2 concentration where necessary and appro­priate. The unit should then be removed from use and serviced by an authorized service technician.
WARNING !
In case of malfunction of any of the built-in monitoring, a substitute must be provided in order to maintain an adequate level of moni­toring. The operator of the ventilator must still assume full responsibility for proper ventila­tion and patient safety in all situations. This is particularly important when, during a nebulizer treatment, the Babylog flow sensor has been removed to protect it against contamination.
CAUTION !
Always hold end cuffs when pulling a circuit from a port. Otherwise damage to the patient circuit will result.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
WARNING !
The operator of the ventilator must still assume full responsibility for proper ventilation and patient safety when switching off flow measurement during nasal CPAP.
Use appropriate external monitoring to assure patient's oxygenation status and to detect apnea.
13
Important Safety Information
Summary of WARNINGS and CAUTIONS
WARNING !
Always heed all precautions and follow all hospital protocols with respect to the admin­istration of oxygen. It is absolutely essential that the ventilation regime and the oxygen concentration administered is selected on the basis of arterially measured oxygen partial pressure in the blood of the infant. This is the only way of minimizing the risk of both hyper­oxemia, which might cause, above all, retro­lental fibroplasia, and hypoxemia, which might contribute to intraventricular hemor­rhage and damage to the baby's brain.
WARNING !
The accuracy of the concentration of oxygen delivered to the patient must be monitored at all times. If the oxygen analyzer integrated into the Babylog 8000 plus ventilator is not operable, an independent oxygen monitor must be connected to the inspiratory limb of the patient circuit.
WARNING !
• Always use extreme caution when using oxygen!
• Oxygen intensely supports any burning! No smoking, no open fire in areas where oxygen is in use!
• Always provide adequate ventilation in order to maintain ambient O2 concentrations < 24 %.
• Always secure O2 cylinders against tipping, do not expose to extreme heat.
• Do not use oil or grease on O2 equipment such as tank valves or pressure regulators. Do not touch with oily hands. Risk of fire!
• Open and close valves slowly, with smooth turns. Do not use any tools.
WARNING !
Do not block air intake. Ventilator malfunction will result.
Be careful not to cause back pressure when using scavenging systems.
WARNING !
Aerosol medications may clog expiratory valves which can impair ventilation.
Therefore, exchange expiratory valve immediately after nebulizing aerosols.
WARNING !
Aerosol medications may clog bacteria filters which can impair ventilation.
Therefore, do not use bacteria filters at the nebulizer output or in the expiratory side of the patient circuit.
WARNING !
The wires of the Babylog 8000 plus flow sensor operate at a high temperature. If the sensor is left in the patient circuit for an extended period of time while nebulizing without cleaning, residue buildup might result from the pharma­ceutical aerosols that would impair flow measurement.
As a worst case, these residues might ignite! To prevent this, it is not sufficient to remove the
electrical connector from the flow sensor. Therefore, it is mandatory to physically remove
the flow sensor from the circuit, (resp. the flow sensor element from the wye) before beginning any nebulizer treatment.
WARNING !
When the flow and volume measuring system of the Babylog 8000 plus ventilator is not operable, detection of spontaneous breathing activity is not possible. Therefore no apnea alarm is pro­vided at that time.
Tidal volumes for ventilated infants are signifi­cantly influenced by the compressible volume of the breathing circuit (in addition to parameters of the patients lungs). Adjustment of inspiratory time, expiratory time, pressure limits, and inspir­atory flow rate all affect the tidal volume deliv­ered to the patient. Assessment of ventilator set­tings must be made with special care during pe­riods where no patient flow/volume information is available. Otherwise, lung damage due to ex­cessive tidal volumes and/or pressures may re­sult.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
14
Important Safety Information
Summary of WARNINGS and CAUTIONS
Precautions During Care
WARNING !
Water traps should be used in appropriate loca­tions of the breathing circuit in order to prevent water accumulation in the tubing from being drained toward the patient's airway.
WARNING !
Warning or Caution level audible alarms require immediate operator attention to avert or to prevent development of situations with the possibility of patient injury.
WARNING !
The alarm silence button is intended to provide a way of muting audible alarms while corrective action is taken. The operator of the ventilator must still assume responsibility for proper venti­lator function and patient safety in the event of an alarm. Failure to identify and correct alarm situations may result in patient injury.
WARNING !
Therapeutic decisions should not be made solely on the basis of the data transmitted via the com­munications Interface. Always use other diag­nostic means in addition.
WARNING !
Always follow accepted hospital procedures for handling equipment contaminated with body fluids.
WARNING !
Follow all accepted hospital procedures for disinfecting parts contaminated by body fluids (protective clothing, eyewear, etc.).
WARNING !
The ventilator is only ready for operation when it has been completely assembled and the flow sensor been calibrated after switching the ventilator on.
CAUTION !
Certain components of the ventilator consist of materials that are sensitive to certain organic solvents sometimes used for cleaning and disinfecting (e.g., phenols, halogen releasing compounds, oxygen releasing compounds, strong organic acids, etc.). Exposure to such substances may cause damage that is not always immediately apparent. Sterilization with ethylene oxide (EtO) is also not recommended.
CAUTION !
Use only dry and clean compressed air and oxy­gen. Water and/or oil in the air or oxygen supply will cause equipment malfunction and damage.
CAUTION !
Do not place containers of liquids on top of the Babylog 8000 plus ventilator. Liquids penetrating the ventilator can cause equipment malfunction and damage.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
CAUTION !
Do not process flow sensor element in cleaning and disinfection equipment.
Do not use compressed air, a brush or similar tools to clean the flow sensor element, as this would possibly damage the thin wires in the flow sensor.
CAUTION !
Ensure that no liquid remains in the pressure measuring canal of the expiratory valve, as it might cause malfunction.
15
Important Safety Information
Summary of WARNINGS and CAUTIONS
Precautions During Maintenance
WARNING !
To avoid any risk of infection, clean and disin­fect ventilator and accessories before any maintenance according to established hospital procedures - this applies also when returning ventilators or parts for repair.
WARNING !
Preventive Maintenance work on the Babylog 8000 plus ventilator shall be performed by properly trained and factory authorized staff only.
WARNING !
When servicing the ventilator, always use replacement parts that are qualified to Dräger standards.
Dräger cannot warrant or endorse the safe performance of third party replacement parts for use with the Babylog 8000 plus ventilator.
WARNING !
Treatment of batteries and O2-sensor capsules:
Do not throw into fire! Risk of explosion. Do not force open! Danger of bodily injury. Follow all local, state, and federal regulations
with respect to environmental protection when disposing of batteries and O2-sensor capsules.
CAUTION !
The device must be inspected and serviced at regular 6 month intervals. A record must be kept on this preventive maintenance. We recommend obtaining a service contract with DrägerService through your vendor.
For repairs of the Babylog 8000 plus infant venti­lators, we recommend that you contact DrägerService.
WARNING !
Disconnect power! Before opening the unit, always disconnect
plug at the wall outlet first.
WARNING !
Never operate the ventilator if it has suffered physical damage or does not seem to operate properly. In this case always refer servicing to properly trained and factory authorized service personnel.
For ventilators with liquid crystal displays (LCD):
WARNING !
If, in case of damage to the ventilator, the glass of the LCD screen is broken, a liquid chemical may escape that should not come into contact with the skin. In case of contamination, wash off with soap immediately.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
16
Intended Use
Contents
Intended Use
Contents
Intended Medical Applications.................................................................18
Available Ventilation Modes.....................................................................18
Mandatory Ventilation Monitoring............................................................ 19
Backup Ventilation With an Independent Manual Ventilation Device......19
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
17
Intended Use
Intended Medical Applications Restrictions of Use
Intended Medical Application
WARNING !
Babylog 8000 plus is solely intended for use by physicians or by allied health care profession­als on the order of a physician. All users must be properly trained and completely familiar with these Operating Instructions.
Babylog 8000 plus is a long term ventilator to be used in the intensive care of premature and newborn babies and of infants up to 10 kg (22 lbs).
Available Ventilation Modes
CMV Continuous Mandatory Ventilation,
Time-controlled, time-cycled, pressure­limited continuous flow ventilation.
A/C Assist Control Ventilation
Time-controlled, volume triggered, time­cycled, pressure-limited continuous flow venti­lation that is synchronized with each spontane­ous patient breath.
SIMV Synchronized Intermittent Mandatory
Ventilation,
Time-controlled, volume triggered, time­cycled, pressure-limited continuous flow venti­lation, synchronized with patient's spontane­ous breathing at the set ventilation rate.
PSV Pressure Support Ventilation
(available option), Time-controlled, volume-triggered, flow cycled, pressure-limited ventilation synchron­ized with each spontaneous patient breath.
CPAP Continuous Positive Airway Pressure,
Spontaneous breathing with positive airway pressure.
Restrictions of Use
CAUTION !
Restriction of Distribution
Federal Law and Regulations in the United States and Canada restrict this device to sale by or on the order of a physician.
WARNING !
This device is to be used only in rooms with line power installations complying with national safety standards for hospital patient rooms. (e.g., IEC 601.1, "Safety of Medical Equipment).
To maintain grounding integrity, connect only to a"hospital grade" receptacle.
Always disconnect supply before servicing.
WARNING !
DANGER, risk of explosion if used in the presence of flammable anesthetics.
This device is neither approved nor certified for use in areas where combustible or explo­sive gas mixtures with air or with nitrous oxide are likely.
WARNING !
Mobile telephones must not be used within 10 meters (33 feet) of the equipment. Mobile telephones can interfere with the function of electromedical equipment and therefore endanger the patient*!
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
Ventilation Mode Extensions
VG Volume Guarantee (available option)
Volume controlled ventilation. The ventilator controls inspiratory pressure in order to deliver the preset tidal volume. May be com­bined with A/C, SIMV, and PSV.
VIVE Variable Inspiratory, Variable Expiratory Flow
Separate continuous flow during expiration in mandatory ventilation modes.
18
WARNING !
Do not use ventilator in conjunction with nuclear spin tomography (MRT, NMR, or NMI)!
Equipment malfunction may result.
____________ *
Dräger medical equipment conforms to the interference immunity requirements laid down in product-specific standards or in EN 60601-1-2 (IEC 601-1-2). However, depending on the design of a mobile phone and the use situation, field strengths exceeding the values laid down in the specified standards may be generated in the immediate vicinity of mobile phones, thereby causing interference and malfunctions.
Intended Use
Mandatory Ventilation Monitoring Backup Ventilation With an Independent Manual Ventilation Device (Resuscitation Bag)
Mandatory Ventilation Monitoring
WARNING !
Sidestream monitors may cause negative pressures in the event of a blocked inspiratory circuit. Always connect monitoring lines of a suctioning sidestream monitor via adapter 84 12 448 with safety valve.
Babylog 8000 plus includes monitoring for: – inspiratory O2 concentration
– airway pressure – flow – tidal volumes – respiration rate (tachypnea monitoring)
WARNING !
Whenever a patient is connected to the venti­lator, constant attention by qualified medical staff is required in order to provide immediate corrective action in case of a malfunction.
WARNING !
Always use independent apnea monitor when the built-in apnea monitoring has been switched off. The operator of the ventilator must still assume full responsibility for proper ventilation and patient safety in all situations.
WARNING !
In case of malfunction of any of the built-in monitoring a substitute must be provided in order to maintain an adequate level of moni­toring. The operator of the ventilator must still assume full responsibility for proper ventilation and patient safety in all situations.
An optional communications interface for the Babylog 8000 plus ventilator is available for communicating measured data and ventilator settings to other devices, such as patient monitors or computers.
For administering pharmaceutical aerosols, the ventilator may be combined with a pneumatic nebulizer.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
Backup Ventilation With an Indepen­dent Manual Ventilation Device (Resuscitation Bag)
WARNING !
If a fault is detected in the ventilator and its life-support functions are in doubt, ventilation must be started without delay with an inde­pendent ventilation device (resuscitation bag)
- using PEEP and/or increased inspiratory O2 concentration where necessary and appro­priate. The unit should then be removed from use and serviced by an authorized service technician.
19
This page intentionally left blank
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
20
Operating Concept
Contents
Operating Concept
Contents
Design of the User Interface....................................................................22
Screen Layout ..........................................................................................23
Screen Menus ......................................................................................... 24
Ventilation Menus...................................................................................... 24
Monitoring Menus......................................................................................25
Calibration/Configuration Menus............................................................... 26
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
21
Operating Concept
Design of the User Interface
Design of the User Interface
The ventilator front end consists of a dial panel and a display/menu key panel.
Dial panel
The dial panel contains buttons for the main operating modes and dial knobs for the most important ventilator parameters.
1 Green LEDs next to the dials indicate mandatory
parameters in the respective operating modes. These LEDs will flash if a parameter has been internally limited or needs to be acknowledged.
2 »Vent. Mode« key switches to basic ventilation mode
menu
3 »Vent. Options« key switches to ventilation mode
extensions menu.
MT 550-97
111
111
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
2 3
22
Operating Concept
412
3
Design of the User Interface
Display/menu key panel
Main components of this panel are the LCD screen and a number of keys with fixed or variable function assignments.
1 » g « key silences audible alarms for 2 minutes. 2 »Confirm« key is used for acknowledging alarms or
to confirm settings.
3 »Man. Insp.« key initiates manual breaths. 4 »Cal. Config.« key switches to calibration menu. 5 In different screen menus, the 6 menu keys below the
screen are used to select monitoring functions and ventilation modes.
6 The illuminated bar graph above the screen displays
current airway pressure readings.
Screen Layout
1 The graphics display window shows pressure or flow
waveform or pressure-volume loops.
6
3
1 2
4
5
2 The measured values window numerically displays
measured values, e.g. for MV, FiO2, Peak, Mean, PEEP.
3 The status window displays the current mode of
ventilation and other status information.
4 The menu bar shows the respective function of the
menu keys as symbols or text labels. In some instances, text and measured values are
shown together in one larger window.
Messages are displayed as overlay windows on top of the current screen:
Display (example):
Apnea
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
23
Operating Concept
Screen Menus
Screen Menus
Ventilation Menus
24
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
Monitoring Menus
Operating Concept
Screen Menus
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
25
Operating Concept
Screen Menus
Calibration/Configuration Menus
26
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
Preparation
Contents
Preparation
Contents
Mounting Babylog 8000 plus on Mobile Stand........................................28
Mounting the Expiration Valve................................................................. 29
Installing an O2 Sensor Capsule..............................................................30
Connecting Supplies ...............................................................................31
Gas Supply............................................................................................... 31
Electrical Supply........................................................................................31
Connecting a Humidifer ..........................................................................32
Assembling Components.........................................................................33
Connecting a Patient Circuit......................................................................33
Installing an Optional Bacteria Filter...........................................................35
Installing the Wye and Flow Sensor...........................................................35
Before Using for the First Time............................................................... 37
Calibrations..............................................................................................38
Calibrating the O2 Sensor Manually...........................................................39
Calibrating the Flow Sensor...................................................................... 40
Exchanging a Flow Sensor Element...........................................................42
Checks of Readiness for Operation.........................................................43
Connecting a Test Lung............................................................................ 43
Testing Power Failure Alarm......................................................................43
Testing Gas Supply Failure Alarm..............................................................44
Testing CMV............................................................................................. 45
Testing PEEP............................................................................................ 45
Testing Alarm Limits.................................................................................. 46
Ventilator Checklist....................................................................................48
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
27
Preparation
Mounting Babylog 8000 plus on Mobile Stand
Mounting Babylog 8000 plus on
Mobile Stand
Turn stand until the two lockable casters are to the
right. NOTE: The following mounting instructions for
Babylog assume the stand is in this position:
1 Tilt Babylog forward approximately 45°. 2 Insert front latches into slots in platform. 3 Lower ventilator inserting back latches into slots in
platform. Secure at the back using knurled screws.
Place tray on the ventilator - the two latches fit into the
back slots in the enclosure.
CAUTION !
Do not place containers of liquids on top of the Babylog 8000 plus ventilator. Liquids penetrating the ventilator can cause equipment malfunction and damage.
1
3
2
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
28
Mounting the Expiratory Valve
WARNING !
Always install expiratory valve that has been cleaned and disinfected.
1 Flip lever in release position = upwards.
Push expiratory valve onto guide rod as far as it will
go.
Preparation
Mounting the Expiratory Valve
1
2 Lock expiratory valve = flip lever down. 3 Connect silencer to exhaust nozzle on expiratory
valve.
3
2
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
29
Preparation
Installing an O2 Sensor Capsule
Installing an O2 Sensor Capsule
– before first use – when calibration can no longer be performed.
Unscrew the two screws in cover on the right and
remove cover lid.
Pull used O2 sensor capsule.
Insert new capsule into opening - circular contacts
must point towards cover lid.
Press down cover and re-tighten the two screws.
Bab
yl
og
8
0
0
0
p
l
u
s
After installation:
Wait for 15 minutes for sensor to warm up before
calibration.
Calibrate sensor manually, see page 39.
Dispose of the used sensor, see page 104.
WARNING !
Treatment of batteries and O2-sensor capsules:
Do not throw into fire! Risk of explosion. Do not force open! Danger of bodily injury. Follow all local, state, and federal regulations
with respect to environmental protection when disposing of batteries and O2-sensor capsules.
Operating Instructions Babylog 8000 plus SW 5, USA, 3rd. ed.
30
Loading...
+ 118 hidden pages